BK PRECISION 390B Series True RMS Handheld Digital Multimeters User Manual

June 13, 2024
BK PRECISION

BK PRECISION 390B Series True RMS Handheld Digital Multimeters

BK-PRECISION-390B-Series-True-RMS-Handheld-Digital-
Multimeters

Product Information

The product models 390B, 391B, and 393B comply with the following specifications:

  • EMC Directive 2014/30/EU
  • EN 61010 Safety requirements for electrical equipment for measurement, control, and laboratory use (applicable for domestic and professional use, capable of measuring mains voltage)
  • EN 61326 Electrical equipment for measurement, control and laboratory use – EMC requirements
  • EN 55011:2009+A1:2010 Industrial, scientific and medical equipment. Radio-frequency disturbance characteristics. Limits and methods of measurement

The product model 394B complies with the following specifications:

  • EMC Directive 2014/30/EU
  • EN 61010 Safety requirements for electrical equipment for measurement, control, and laboratory use (applicable for domestic and professional use, capable of measuring mains voltage)
  • EN 61326 Electrical equipment for measurement, control and laboratory use – EMC requirements
  • EN 55011:2016+A1:2017 Industrial, scientific and medical equipment. Radio-frequency disturbance characteristics. Limits and methods of measurement

The product also meets the requirements of IEC 61000 for electromagnetic compatibility (EMC) and voltage variations immunity tests.

The product is classified into different Measurement Categories (CAT) based on the types of mains circuits it can be connected to:

  • Measurement Category other than II, III, or IV: Circuits not directly connected to the mains supply.
  • Measurement Category II (CAT II): Test and measuring circuits connected directly to utilization points (socket outlets and similar prints) of the low-voltage mains installation.
  • Measurement Category III (CAT III): Test and measuring circuits connected to the distribution part of a building’s low-voltage mains installation.
  • Measurement Category IV (CAT IV): Test and measuring circuits connected at the source of the building’s low-voltage mains installation.

The product is designed for measurements performed on circuits not directly connected to a mains supply (Mains Isolated).
Pollution Degree 2 (P2) indicates that only non-conductive pollution occurs except for occasional temporary conductivity caused by condensation.

Product Usage Instructions

  1. Ensure that the product is used in accordance with thes pecified Measurement Category (CAT) for your application. Refer to the product’s user manual for guidance on selecting the appropriate CAT.
  2. Before using the product, make sure it complies with the necessary safety requirements for electrical equipment, as specified in EN 61010.
  3. The product is suitable for both domestic and professional use and is capable of measuring mains voltage. Use it for measurement, control, and laboratory purposes.
  4. Follow the EMC requirements outlined in EN 61326 to ensure proper electrical equipment performance, especially in terms of electromagnetic compatibility.
  5. To avoid any potential harm to the environment or human health, it is important to handle the product properly at the end of its life.
  6. Recycling is encouraged to reduce the use of natural resources and prevent the release of harmful substances into the environment. Dispose of the product at an appropriate recycling facility in accordance with Directive 2012/19/EU (WEEE).
  7. Do not dispose of the product as unsorted municipal waste. Utilize local WEEE collection facilities for proper disposal.

Compliance Information

390B, 391B, & 393B
EC Declaration of Conformity – EMC
Compliance was demonstrated to the following specifications listed in the Official Journal of the European Communities: EMC Directive 2014/30/EU.
EN 61010 Safety requirements for electrical equipment for measurement, control, and laboratory use
­ Part 1: General requirements ­ Part 2-033: Particular requirements for hand- held multimeters and other meters, for
domestic and professional use, capable of measuring mains voltage
EN 61326 Electrical equipment for measurement, control and laboratory use – EMC requirements
­ Part 1: General requirements ­ Part 2-1: Particular requirements – Test configurations, operational conditions
and performance criteria for sensitive test and measurement equipment for EMC unprotected applications ­ Part 2-2:Particular requirements – Test configurations, operational conditions and performance criteria for portable testing, measuring and monitoring equipment used in low-voltage distribution systems
EN 55011:2009+A1:2010 Industrial, scientific and medical equipment. Radio- frequency disturbance characteristics. Limits and methods of measurement

Compliance Information 3
IEC 61000 Electromagnetic compatibility (EMC)
­ Part 3 ­ 2: Limits — Limits for harmonic current emissions (equipment input current 16 A per phase)
­ Part 3-3: Limits – Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current 16 A per phase and not subject to conditional connection
­ Part 4-2: Testing and measurement techniques – Electrostatic discharge immunity test ­ Part 4-3: Testing and measurement techniques – Radiated, radio-frequency,
electromagnetic field immunity test ­ Part 4-4: Testing and measurement techniques – Electrical fast transient/burst
immunity test ­ Part 4-5: Testing and measurement techniques – Surge immunity test ­ Part 4-6: Testing and measurement techniques – Immunity to conducted disturbances,
induced by radio-frequency fields ­ Part 4-8: Testing and measurement techniques – Power frequency magnetic field
immunity test ­ Part 4-11: Testing and measurement techniques – Voltage dips, short interruptions
and voltage variations immunity tests

394B

Compliance Information 4

EC Declaration of Conformity – EMC
Compliance was demonstrated to the following specifications listed in the Official Journal of the European Communities: EMC Directive 2014/30/EU.
EN 61010 Safety requirements for electrical equipment for measurement, control, and laboratory use
­ Part 1: General requirements ­ Part 2-030: Particular requirements for testing and measuring circuits ­ Part 2-033: Particular requirements for hand- held multimeters and other meters, for
domestic and professional use, capable of measuring mains voltage
EN 61326 Electrical equipment for measurement, control and laboratory use – EMC requirements
­ Part 1: General requirements ­ Part 2-1: Particular requirements – Test configurations, operational conditions
and performance criteria for sensitive test and measurement equipment for EMC unprotected applications ­ Part 2-2:Particular requirements – Test configurations, operational conditions and performance criteria for portable testing, measuring and monitoring equipment used in low-voltage distribution systems
EN 55011:2016+A1:2017 Industrial, scientific and medical equipment. Radio- frequency disturbance characteristics. Limits and methods of measurement

Compliance Information 5
IEC 61000 Electromagnetic compatibility (EMC)
­ Part 4-2: Testing and measurement techniques – Electrostatic discharge immunity test ­ Part 4-3: Testing and measurement techniques – Radiated, radio-frequency,
electromagnetic field immunity test ­ Part 4-4: Testing and measurement techniques – Electrical fast transient/burst
immunity test ­ Part 4-5: Testing and measurement techniques – Surge immunity test ­ Part 4-6: Testing and measurement techniques – Immunity to conducted disturbances,
induced by radio-frequency fields ­ Part 4-8: Testing and measurement techniques – Power frequency magnetic field
immunity test

Compliance Information 6
IEC Measurement Category & Pollution Degree Definitions
Measurement Category (CAT) – classification of testing and measuring circuits according to the types of mains circuits to which they are intended to be connected.
Measurement Category other than II, III, or IV : circuits that are not directly connected to the mains supply.
Measurement Category II (CAT II) : test and measuring circuits connected directly to utilization points (socket outlets and similar prints) of the low- voltage mains installation.
Measurement Category III (CAT III) : test and measuring circuits connected to the distribution part of a building’s low-voltage mains installation.
Measurement Category IV (CAT IV) : test and measuring circuits connected at the source of the building’s low-voltage mains installation.
Mains Isolated : is for measurements performed on circuits not directly connected to a mains supply.
Pollution – addition of foreign matter, solid, liquid, or gaseous (ionized gases) that may produce a reduction of dielectric strength or surface resistivity.
Pollution Degree 2 (P2) – only non-conductive pollution occurs except that occasionally a temporary conductivity caused by condensation is expected

Product End-of-Life Handling

Compliance Information 7

The equipment may contain substances that could be harmful to the environment or human health if improperly handled at the product’s end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we encourage you to recycle this product to an appropriate system that will ensure that most of the materials are reused or recycled appropriately.

This product is subject to Directive 2012/19/EU of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product.

Terms and Symbols

Terms

A caution statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in damage to or destruction of parts or the entire product.
A warning statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in injury or death to personnel.
A note statement calls attention to an operating procedure, practice, or condition, which, should be noted before proceeding.

Symbols

WARNING – Risk of electric shock.

Compliance Information 8

CAUTION ­ Statements or instructions that must be consulted in order to find out the nature of the potential hazard and any actions which must be taken.
Equipment protected by double or reinforced insulation

Battery

AC measurement

DC measurement

Fuse

Earth (ground) TERMINAL – Refer to the instructions accompanying this symbol in this manual.

Safety Notices
The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Before applying power to this instrument:
· Read and understand the safety and operational information in this manual.
· Apply all the listed safety precautions.
· Verify that the voltage selector at the line power cord input is set to the correct line voltage. Operating the instrument at an incorrect line voltage will void the warranty.
· Make all connections to the instrument before applying power.
· Do not operate the instrument in ways not specified by this manual or by B&K Precision.
Failure to comply with these precautions or with warnings elsewhere in this manual violates the safety standards of design, manufacture, and intended use of the instrument. B&K Precision assumes no liability for a customer’s failure to comply with these requirements.
Electrical Power
This instrument is intended to be powered from a CATEGORY II mains power environment. The mains power should be 115 V RMS or 230 V RMS. Use only the power cord supplied with the instrument and ensure it is appropriate for your country of use.
Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual for this instrument.
You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument’s category rating to maintain the instrument’s category rating. Failure to do so will lower the category rating of the measuring system.

Ground the Instrument

Safety Notices 10

To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical safety ground. This instrument is grounded through the ground conductor of the supplied, threeconductor AC line power cable. The power cable must be plugged into an approved three-conductor electrical outlet. The power jack and mating plug of the power cable meet IEC safety standards.

Do not alter or defeat the ground connection. Without the safety ground connection, all accessible conductive parts (including control knobs) may provide an electric shock. Failure to use a properly-grounded approved outlet and the recommended threeconductor AC line power cable may result in injury or death.

Unless otherwise stated, a ground connection on the instrument’s front or rear panel is for a reference of potential only and is not to be used as a safety ground. Do not operate in an explosive or flammable atmosphere.

Environmental Conditions

Safety Notices 11

This instrument is intended to be used in an indoor pollution degree 2 environment. The operating temperature range is 0C to 40C and 20% to 80% relative humidity, with no condensation allowed.
Measurements made by this instrument may be outside specifications if the instrument is used in nonoffice-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields.

Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finely-divided particulates.

The instrument is designed to be used in office-type indoor environments. Do not operate the instrument
· In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finely-divided particulates.
· In relative humidity conditions outside the instrument’s specifications.
· In environments where there is a danger of any liquid being spilled on the instrument or where any liquid can condense on the instrument.
· In air temperatures exceeding the specified operating temperatures.
· In atmospheric pressures outside the specified altitude limits or where the surrounding gas is not air.
· In environments with restricted cooling air flow, even if the air temperatures are within specifications.
· In direct sunlight.

Do not operate instrument if damaged

Safety Notices 12

If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument’s power cord, remove the instrument from service, label it as not to be operated, and return the instrument to B&K Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument.

Hazardous voltages may be present in unexpected locations in circuitry being tested when a fault condition in the circuit exists.

Clean the instrument only as instructed
Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual. Not for critical applications.

Do not touch live circuits

Safety Notices 13

Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified service-trained maintenance personnel who are aware of the hazards involved when the instrument’s covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed.

To avoid injuries, always disconnect the power cord from the instrument, disconnect all other connections (for example, test leads, computer interface cables, etc.), discharge all circuits, and verify there are no hazardous voltages present on any conductors by measurements with a properly-operating voltagesensing device before touching any internal parts. Verify the voltage- sensing device is working properly before and after making the measurements by testing with known-operating voltage sources and test for both DC and AC voltages.
Do not attempt any service or adjustment unless another person capable of rendering first aid and resuscitation is present.

General Safety

Do not insert any object into an instrument’s ventilation openings or other openings.

This instrument is not authorized for use in contact with the human body or for use as a component in a life-support device or system.

Servicing

Safety Notices 14
Do not substitute parts that are not approved by B&K Precision or modify this instrument. Return the instrument to B&K Precision for service and repair to ensure that safety and performance features are maintained.
Fuse replacement must be done by qualified service-trained maintenance personnel who are aware of the instrument’s fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specified in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a fire. Failure to use the specified fuses will void the warranty.

For continued safe use of the instrument
· Do not place heavy objects on the instrument. · Do not obstruct cooling air flow to the instrument. · Do not place a hot soldering iron on the instrument. · Do not pull the instrument with the power cord, connected probe, or connected test lead. · Do not move the instrument when a probe is connected to a circuit being tested.

Contents

Compliance Information

2

390B, 391B, & 393B

2

394B

4

IEC Measurement Category & Pollution Degree Definitions

6

Product End-of-Life Handling

7

Terms and Symbols

7

Safety Notices

9

1 General Information

17

1.1 Product Overview

17

1.2 Contents

18

1.3 Front Panel Overview

19

1.3.1 Push Buttons

20

2 Making Basic Measurements

21

2.1 Measuring AC /DC Voltage

22

2.1.1 High Frequency Reject Mode

23

2.1.2 Making dB Measurements AC mV only

24

2.2 Measuring Resistance

25

2.2.1 Testing for Continuity

26

2.2.2 Testing Diodes

27

2.2.3 Measuring Capacitance

28

2.3 Measuring AC / DC Current

29

2.4 Measuring Frequency

31

2.5 Measuring Temperature C F

32

3 Functions

33

3.1 Store and Recall Functions

34

3.2 MAX MIN AVG

35

3.3 Relative Mode

35

3.4 Digit

36

3.5 Auto Hold

36

3.6 Peak Hold

36

3.7 Data Log and Log Rate

37

3.8 Auto Backlight

37

3.9 Power Up Option

37

4 Process Multimeter Functions

38

4.1 DC Current Output

39

16

4.1.1 Adjustable DC Current Output

40

4.1.2 Auto DC Current Output

40

4.2 Source Mode

41

4.3 Simulation Mode

42

4.4 Loop Power

43

4.4.1 250 HART

43

5 Battery and Fuse Replacement

44

6 Specifications

45

7 Service Information

49

8 LIMITED THREE-YEAR WARRANTY

50

General Information

Product Overview
The 390B Series True RMS multimeters offer a comprehensive solution for general purpose measurement applications. Bundled with a complete set of accessories, these multimeters provide technicians and engineers with accurate results, data logging capabilities, and measurement features for evaluating a variety of electronics or electrical systems.
Advanced features like data logging records measurements to the meter’s internal memory at a userspecified sampling interval.

Features True RMS DCV Accuracy Display Count AutoV/LoZ Auto Backlit, Dual Display Analog Style Bar Graph Relative Mode (% and ) Min/Max/Avg dBm/dB measurement Auto Hold, Peak Hold Data Log Capacity Duty Cycle Temperature Probe Optical Isolated USB Interface

390B

391B

393B

AC, AC + DC voltage and current

± (0.03% + 10 digits) ± (0.03% + 20 digits) ± (0.015% + 20 digits)

40,000

40,000

100,000

40,000 readings –

20,000 readings

20,000 readings

Table 1.1 390B Series Models

General Information 18
Figure 1.1 390B
1.2 Contents
Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items from the shipping carton, and check for any obvious signs of physical damage that may have occurred during transportation. Report any damage to the shipping agent immediately. Save the original packing carton for possible future reshipment. Every instrument is shipped with the following contents: · 390B series DMM · test leads · K-type thermocouple (not included with 394B model) · magnetic hanger strap · USB communication cable
Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision.

1.3 Front Panel Overview

General Information 19

Item
1 2 3 4
5
6 7 8

Figure 1.2 Front Panel

Name
Light Sensor

Description
Auto backlight sense point.

Display

40,000 /100,000 count dual display

Push Buttons

Push buttons. Refer to table 1.3 for more information.

Rotary Switch Input Terminal Return Terminal

Switches between the available operation modes.
Input terminal for voltage, frequency, resistance, continuity, diode, capacitance and temperature measurements.
Return terminal for all measurements.

mA Input Terminal

Input terminal for 0 to 400mA current measurements.

A Input Terminal

Input terminal for 0 to 10A current measurements.

Table 1.2 Front Panel Overview

1.3.1 Push Buttons

General Information 20

Button
Shift RANGE
HFR Link Digit Hz ENTER CANCEL A-HOLD
P-HOLD
REL

Description
Select measurement function. Select measurement range. Press > 1 sec to enter auto range mode. Enable/Disable the High Frequency Reject mode for AC measurements. Enable/Disable the link mode. (390B only) Selects the display digit. Measures the frequency for voltage or current. (394B only) Enter menu function in pointer position. Cancel current menu function. Enable/Disable the Auto-Hold mode. Enable the Peak-Hold mode in the AC or DC measurements. In this mode, press button to select Peak-Hold MAX or MIN. Press > 1 sec to disable the Peak-Hold mode. Enable/Disable the relative mode. (394B only)
Table 1.3 Push Buttons

Making Basic Measurements

2.1 Measuring AC /DC Voltage

22

2.1.1 High Frequency Reject Mode

23

2.1.2 Making dB Measurements AC mV only

24

2.2 Measuring Resistance

25

2.2.1 Testing for Continuity

26

2.2.2 Testing Diodes

27

2.2.3 Measuring Capacitance

28

2.3 Measuring AC / DC Current

29

2.4 Measuring Frequency

31

2.5 Measuring Temperature C F

32

When connecting the test leads to the DUT (Device Under Test) connect the common (mA) test lead before connecting the live lead; when removing the test leads, remove the test live lead before removing the common test lead.
The figures on the following pages show how to make basic measurements.

2.1 Measuring AC /DC Voltage

Making Basic Measurements 22

Set up your multimeter to measure voltage as shown in figure 2.1. Probe the test points and read the display.

Figure 2.1 Voltage Measurement
Press the SHIFT button to select manually select one of the available measuring function (AC/DC/AC+DC). Press and hold the SHIFT button to enter manual test mode. In manual test mode the measuring function will automatically selected based on the input signal.

2.1.1 High Frequency Reject Mode

Making Basic Measurements 23

The 390B series is equipped with an AC LPF (low-pass filter) to help reduce unwanted electronic noise when measuring AC voltage or AC frequency. The LPF can improve measurement performance on composite sine waves that are typically generated by inverters and variable frequency drives.
Set up the multimeter to measure AC voltage. Press the HFR button to enable the LPF. The multimeter will continue measuring in the chosen AC mode, but now the signal diverts through a filter that blocks voltages above 800 Hz.

Figure 2.2 High Frequency Reject
To avoid possible electric shock or personal injury, do not use the HFR option to verify the presence of hazardous AC voltages. AC voltage values greater than what are indicated may be present when the LPF is enabled. First, make an AC voltage measurement without LPF to detect the possible presence of hazardous voltages. Then, enable the LPF if required for measurement stability and response speed.
The HFR function is only available for AC voltage measurements. This function is not supported for AC + DC voltage measurements.

2.1.2 Making dB Measurements AC mV only

Making Basic Measurements 24

The multimeter is capable of displaying voltage as a dB value, either relative to 1 milliwatt (dBm) or a reference voltage of 1 volt (dB).
To set the multimeter to display values in either dBm or dB, first set up your multimeter to measure AC voltage as shown in figure 2.1. Once AC voltage is selected use the navigation keys to select either dB or dBm which are located near the bottom of the display. Once you are hovering over the desired display option press the ENTER button.
The dB or dBm display will appear above the voltage reading. You can now probe the test points, and read the display in the selected unit.

Figure 2.3 Display dB

A dBm measurement must use a reference impedance (resistance) to calculate a dB value based on 1 milliwatt. The reference impedance is set to 50 by default.
­ A dB measurement uses a 1 volt reference voltage to compare the present measurement against a stored relative value. The difference between the two AC signals is displayed as a dB value. The reference impedance setting is not part of a dB measurement.

=

20

0.7746

=

20

1

2.2 Measuring Resistance

Making Basic Measurements 25

Set up your multimeter to measure resistance as shown in figure 2.5. Press the SHIFT button to navigate through the available measurements shown in figure 2.4. Probe the test points and read the display.

Figure 2.4 Available Measurement Modes

Figure 2.5 Resistance Measurement

Making Basic Measurements 26
To avoid possible damage to your multimeter or to the equipment under test, disconnect the circuit power and discharge all highvoltage capacitors before measuring resistance.
The test leads can add 0.1 to 0.2 of error to resistance measurements.

Testing for Continuity
Set up your multimeter to test continuity as shown in figure 2.5. Press the SHIFT button once to select test mode 2, continuity, as shown in figure 2.4. Probe the test points and listen for a beep.
To avoid possible damage to your multimeter or to the equipment under test, disconnect the circuit power and discharge all highvoltage capacitors before testing for continuity.
The continuity test features a beeper that sounds as long as a circuit is not
broken. The audible alert allows you to perform quick continuity tests without
having to watch the display.
In continuity, a short means a measured value is less than the set threshold resistance values.

Model 390B
391B & 393B

Range 400.0 400.00

Resolution 0.1 0.01

Accuracy 0.2 + 2 0.2 + 30

Table 2.1 Threshold Resistance Values
· Open Circuit Voltage: -1.2 V · Short Test Current: -0.3 mA · Threshold: Adjustable 10 to 50 · Buzzer Response Time: < 10 ms
The threshold is adjustable under the SETUP menu > Cnt in.

2.2.2 Testing Diodes

Making Basic Measurements 27

Set up your multimeter to test diodes as shown in figure 2.5. Press the SHIFT button twice to select test mode 3, Diode, as shown in figure 2.4. Probe the test points as shown in figure 2.6 and read the display.
Use the diode test to check diodes, transistors, silicon controlled rectifiers (SCRs), and other semiconductor devices. A good diode allows current to flow in one direction only.
This test sends a current through a semiconductor junction and then measures the junction’s voltage drop. A typical junction drop is 0.3 V to 0.8 V.
Connect the red test lead to the positive terminal (anode) of the diode and the black test lead to the negative terminal (cathode).
The multimeter can display diode forward bias of up to approximately 3.1 V. The forward bias of a typical diode is within the range of 0.3 V to 0.8 V; however, the reading can vary depending on the resistance of other pathways between the probe tips.

Reverse the probes (as shown in figure 2.6) and measure the voltage across the diode again. Assess the diode according to the following guidelines:
­ A diode is considered good if the multimeter displays 0L in reverse bias mode. ­ A diode is considered shorted if the multimeter displays approximately 0 V in both forward and reverse
bias modes, and the multimeter beeps continuously. ­ A diode is considered open if the multimeter displays 0L in both forward and reverse bias modes.

Forward Bias

Reverse Bias

Figure 2.6 Testing Diodes

2.2.3 Measuring Capacitance

Making Basic Measurements 28

Set up your multimeter for capacitance measurement as shown in figure 2.5. Press the SHIFT button three times to select test mode 4, Capacitance, as shown in figure 2.4. Probe the test points as shown in figure 2.7 and read the display.
To avoid possible damage to the multimeter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the DC voltage function to confirm that the capacitor is fully discharged.

The multimeter measures capacitance by charging the capacitor with a known current for a known period of time, measuring the resulting voltage, and then calculating the capacitance.
For measuring capacitance values greater than 1000 F, discharge the capacitor first, then select a suitable range for measurement. This will speed up the measurement time and also ensure that the correct capacitance value is obtained.

Figure 2.7 Capacitance Measurement

2.3 Measuring AC / DC Current

Making Basic Measurements 29

Set up your multimeter to measure AC or DC current as shown in figure 2.8. Press the SHIFT button to toggle between AC and DC mode. Open the circuit path to be tested, and probe the test points as shown in figure 2.8 then read the display.

Figure 2.8 Current Measurement
Before attempting any current measurement, switch off the power source of the circuit and measure the AC or DC voltage to ensure that the power source has been switched off.
Never attempt an in-circuit current measurement where the opencircuit potential to earth is greater than 1000 V. Doing so will cause damage to the multimeter and possible electric shock or personal injury.

Making Basic Measurements 30
To avoid possible damage to the multimeter or to the equipment under test:
­ Check the multimeter’s fuses before measuring current. ­ Use the proper terminals, function, and range for your measurement. ­ Never place the probes across (in parallel with) any circuit
or component when the leads are plugged into the current terminals. ­ Current can be measured continuously for 10 minutes at mA inputs, and 1 minute at A inputs. ­ After measuring current continuously, allow for a minimum cool down of 20 minutes. ­ Burden Voltage: 2 mV/mA at mA inputs and 60 mV/A at A inputs ­ Overload protection: AC/DC 400 mA at mA inputs, AC/DC 10 A at A inputs

2.4 Measuring Frequency

Making Basic Measurements 31

Set up your multimeter to measure frequency as shown in figure 2.9. Press the SHIFT button to toggle between period and duty.

Figure 2.9 Frequency Measurement
Never measure the frequency where the voltage or current level exceeds the specified range. Manually set the voltage or current range if you want to measure frequencies below 20 Hz.

2.5 Measuring Temperature C F

Making Basic Measurements 32

Set up your multimeter to measure temperature as shown in figure 2.10. Press the SHIFT button to toggle between Celsius and Fahrenheit.

Figure 2.10 Temperature Measurement
Never measure the frequency where the voltage or current level exceeds the specified range. Manually set the voltage or current range if you want to measure frequencies below 20 Hz.

Functions

The 390B series offers the following functions:

3.1 Store and Recall Functions

34

3.2 MAX MIN AVG

35

3.3 Relative Mode

35

3.4 Digit

36

3.5 Auto Hold

36

3.6 Peak Hold

36

3.7 Data Log and Log Rate

37

3.8 Auto Backlight

37

3.9 Power Up Option

37

These functions can be enabled / disabled using the push buttons and the navigation keys. The navigation keys can be used to navigate the Function menu located at the bottom portion of the display. The currently selected Function will blink. To enable or enter the menu of the function press the ENTER button.

Figure 3.1 Function Menu

3.1 Store and Recall Functions

Functions 34

The store function records the measured value to the internal memory. A total of 1000 measurements can be recorded.

Figure 3.2 Store Measurement
Recorded measurements can be recalled from memory and displayed one at a time or imported to a PC in spreadsheet format using the provided software

Figure 3.3 Recall Measurement

Figure 3.4 Import measurements with software

3.2 MAX MIN AVG

Functions 35

The MAX/MIN/AVG mode records the minimum and maximum input values. When the input is measured below the record minimum value or above the record maximum value, the meter records the new value. The MAX/MIN/AVG mode can also calculate the average of the maximum value and the minimum value.

Figure 3.5 Max
3.3 Relative Mode
The relative mode records the input value as reference and appears on the upper display. The following
displayed measurements will be the calculated difference () of the reference value and the input value
or the difference percent (%) of the reference value and the input value. One possible application is to increase the accuracy of a resistance measurement by nulling the test lead resistance. Nulling the leads is also particularly important prior to making capacitance measurements.
Relative mode can be set for both auto and manual range settings, but an overload reading cannot be stored as a null value.

Figure 3.6 Relative Mode

3.4 Digit
Press the DIGIT button to select the display digit as shown in figure 3.7.

Functions 36

Figure 3.7 Digit Configuration
3.5 Auto Hold
The Auto Hold function, holds the last measured reading, and the current reading appears on the upper display. When the difference of the hold value and the current value is above 20 counts, the meter beeps and holds the new value. To use the Auto Hold mode, press the A-Hold button as shown in figure 3.8.

Figure 3.8 A-HOLD Function
3.6 Peak Hold
The Peak Hold function records the peak maximum value and the peak maximum value. When the input goes below the recorded peak minimum value or above the recorded peak maximum value, the meter records the new value. To use the Peak HOLD mode, press the P-HOLD button as shown in figure 3.9

Figure 3.9 P-HOLD Function

3.7 Data Log and Log Rate

Functions 37

The Data Logging function provides the convenience of recording test data for future review or analysis. Since data is stored in the nonvolatile memory, the data remains saved even when the multimeter is turned OFF or if the battery is replaced.
The Data Logging feature collects measurement information at the specified log rate. The log rate can be configured between 1 to 600 seconds. Table 3.1 shows the logging capacity for each model in the series.

Model Display Count DC Voltage Basic Accuracy Wireless Data-logging Data Log Capacity

390B 40,000 count
0.03%
40,000 readings

391B 40,000 count
0.03% –
20,000 readings

Table 3.1 Log Capacity

393B 100,000 count
0.015% –
20,000 readings

A measurement can manually be logged into the nonvolatile memory using the Store Function.

3.8 Auto Backlight
The multimeter’s automatic backlight features uses a photoresistor to determine when to automatically enable / disable the backlight.
The auto backlight can be disabled in the Setup Menu.

3.9 Power Up Option
Press the Enter or Cancel button while turning the meter ON from the OFF position to call the corresponding function;
Enter : Displays the firmware version Cancel : Clear all stored data.

Process Multimeter Functions

The 394B Process Multimeter combines the capabilities of a mA loop calibrator with a full-featured True RMS multimeter in one package.

The following functions are exclusive to the 394B Model:

4.1 DC Current Output

39

4.1.1 Adjustable DC Current Output

40

4.1.2 Auto DC Current Output

40

4.2 Source Mode

41

4.3 Simulation Mode

42

4.4 Loop Power

43

4.4.1 250 HART

43

Process Multimeter
Current Output Ranges Current Output Adjustment Modes
Loop Power Supply 250 HART® Mode
General Purpose Multimeter
True RMS Basic DCV Accuracy
Display

Key Specifications
0-20 mA or 4-20 mA, using internal batteries or external loop supply Slow ramp, fast ramp, 25% step > 24 V

AC, AC+DC voltage and current
± 0.05% 5 digit / 50,000 count

Table 4.1 394B Key Specifications

4.1 DC Current Output

Process Multimeter Functions 39

To use the DC current output function, turn the rotary knob to the output position (Adjustable DC output or Auto DC output) as shown in figure 4.1.

Figure 4.1 DC Current Output
The DC current output function has both Source Mode & Simulate Mode. The output mode has both types: 0-20mA & 4-20mA.

4.1.1 Adjustable DC Current Output

Process Multimeter Functions 40

To use the adjustable DC current output function, turn the rotary to the adjustable output position. In this function, you can adjust the DC current output.

­ %STEP: 0% / 25% / 50% / 75% / 100% / 120% / 125%

­ Fast Setup: 0% / 50% / 100%

­ Fine Setup : Minimum resolution 1uA, 0mA to 24mA

% STEP
0% 25% 50% 75% 100% 120% 125%

0-20mA Mode
0mA 5mA 10mA 15mA 20mA 24mA N/A Table 4.2 Adjustable DC Current Output

4-20mA Mode
4mA 8mA 12mA 16mA 20mA N/A 24mA

4.1.2 Auto DC Current Output

To use the auto DC current output function, turn the rotary knob to the auto output position.
In this function, you can press the SHIFT button to select 4 kinds of the auto DC current output. Press the HOLD button to pause / continue the output.

Mode

Type
Linear

Action
0% to 100% to 0% per 40 sec

Linear

0% to 100% to 0% per 20 sec

25% Step

0% to 100% to 0%, a step per 15 sec

25% Step

0% to 100% to 0%, a step per 5 sec

Table 4.3 Auto DC Current Output

4.2 Source Mode

Process Multimeter Functions 41

When the meter is in source mode it provides an internal power supply (> 4.5V) to drive the DC current output. To operate in the source mode, place the positive probe in the A terminal and the negative probe in the mA terminal. This will cause the meter to automatically enter the source mode.
Source mode works in both modes Adjustable DC output and Auto DC output. In auto DC current output mode, you can press the A-HOLD button to pause / continue the output.

Figure 4.2 Source Mode Connection
Do not turn the rotary knob while the probe is in the A terminal. This action maybe caused > 30mA to pass through the loop circuit.

4.3 Simulation Mode

Process Multimeter Functions 42

In Simulation mode an external power supply (12V to 48V) is needed to drive the DC current output.
To operate in the simulate mode place the positive probe in the mA terminal and the negative probe in the COM terminal. This will cause the meter to automatically enter the simulation mode.
Simulation mode works in both modes Adjustable DC output and Auto DC output. In auto DC current output mode, you can press the A-HOLD button to pause / continue the output.

Figure 4.3 Simulation Mode Connection
Do not turn the rotary knob while the probe is in the A terminal. This action maybe caused > 30mA to pass through the loop circuit.

4.4 Loop Power

Process Multimeter Functions 43

To operate in the loop power function, place the positive probe in the mA terminal and the negative probe in the COM terminal.

Figure 4.4 Loop Power Mode Connection
Do not turn the rotary knob while the probe is in the A terminal. This action maybe caused > 30mA to pass through the loop circuit.
4.4.1 250 HART
Switch the rotary knob to the loop power position, then press the SHIFT button to enable / disable the 250 HART.

Battery and Fuse Replacement
Refer to figure 5.1 to replace fuse and the batteries :
Figure 5.1 Battery and Fuse Replacement

Specifications

Specifications are based on the following conditions/assumptions:
n Accuracy specifications: ± (% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH One year calibration cycle n Temperature coefficient is 0.1 x (specified accuracy)/°C for T < 18 °C, T > 28°C n AC Voltage and AC Current specifications are AC coupled, true RMS n For non-sinusoidal waveforms:
– Add 1.0% to accuracy specification for Crest Factor 1.0 to 2.0 – Add 2.5% to accuracy specification for Crest Factor 2.0 to 2.5 – Add 4.0% to accuracy specification for Crest Factor 2.5 to 3.0 n For best accuracy use REL (delta) function to compensate the offsets

DC Voltage

DC Current

Model

Range

40.00 mV

400.0 mV

4.000 V 390B
40.00 V

400.0 V

1000 V

40.000 mV

400.00 mV

4.000 V 391B
40.000 V

400.00 V

1000.0 V

100.000 mV

1000.00 mV

393B

10.0000 V

100.000 V

1000.00 V

– Input Impedance: 10 M, < 100 pF – Overload Protection: AC/DC 1000 V

Continuity

Model

Range

390B 391B & 393B

400.0 400.00

– Open Circuit Voltage: -1.2 V
– Short Test Current: -0.3 mA – Threshold: Adjustable 10 to 50 – Buzzer Response Time: < 10 ms

Diode Test

Model

Range

390B

2.000 V

391B & 393B

2.0000 V

– Open Circuit Voltage: 2.5 V – Short Test Current: 1 mA

Resolution 1 µV 10 µV
100 µV 1 mV 10 mV 100 mV 1 µV 10 µV 100 µV 1 mV 10 mV 100 mV 1 µV 10 µV 100 µV 1 mV 10 mV
Resolution 0.1 0.01
Resolution 1 mV 100 µV

Accuracy 0.03 + 3
0.03 + 1
0.040 + 40 0.035 + 20 0.030 + 20 0.025 + 40 0.020 + 20 0.015 + 20
Accuracy 0.2 + 2 0.2 + 30
Accuracy 1.5 + 2 1.5 + 20

Model

Range

Resolution

Accuracy

40.00 mA

1 µA

400.0 mA

10 µA

390B

4.000 A

100 µA

10.00 A

1 mA

40.000 mA

1 µA

400.00 mA

10 µA

391B

4.0000 A

100 µA

10.000 A

1 mA

10.0000 mA

0.1 µA

393B

100.000 mA

1 µA

10.0000 A

100 µA

0.2 + 1 0.2 + 2 0.2 + 40 0.2 + 80 0.1 + 40 0.1 + 80

– Burden Voltage: 2 mV/mA at mA inputs and 60 mV/A at A inputs – Max. Continuous Measuring Time: 10 minutes at mA inputs, 1 minute at A inputs – Min. Rest Time: 20 minutes after continuous measuring – Overload protection: AC/DC 400 mA at mA inputs, AC/DC 10 A at A inputs

Resistance Model 390B
391B
393B

Range
400.0 4.000 k 40.00 k 400.0 k 4.000 M 40.00 M 400.00 4.0000 k 40.000 k 400.00 k 4.0000 M 40.000 M 1000.00 10.0000 k 100.000 k 1000.00 k 10.0000 M 40.000 M

Resolution
0.1 1 10 100 1 k 10 k 0.01 0.1 1 10 100 1 k 0.01 0.1 1 10 100 1 k

Accuracy 0.2 + 2
0.2 + 1
1.0 + 1 2.0 + 20
0.2 + 30
0.3 + 30 1.0 + 30 1.5 + 30 0.050 + 30 0.025 + 30 0.3 + 30 1.0 + 30 1.5 + 30

Specifications 46

True RMS AC Voltage

Model

Range

Frequency

Accuracy

390B 391B 393B

40.00 mV(1), 400.0 mV(1), 4.000 V, 40.00 V, 400.0 V(1), 1000 V(2) 40.000 mV, 400.00 mV 4.0000 V, 40.000 V 400.00 V, 1000.0 V 100.000 mV, 1000.00 mV 10.0000 V, 100.000 V 1000.00 V

40 Hz to 70 Hz(3) 70 Hz to 1 kHz(3) 1 kHz to 5 kHz(3) 5 kHz to 100 kHz(4) (5) 40 Hz to 65 Hz 65 Hz to 1 kHz 1 kHz to 3 kHz 40 Hz to 45 Hz 45 Hz to 65 Hz 65 Hz to 1 kHz 1 kHz to 10 kHz 10 kHz to 50 kHz 50 kHz to 100 kHz 40 Hz to 45 Hz 45 Hz to 65 Hz 65 Hz to 1 kHz 40 Hz to 65 Hz 66 Hz to 1 kHz 1 kHz to 3 kHz 40 Hz to 45 Hz 45 Hz to 65 Hz 65 Hz to 1 kHz 1 kHz to 10 kHz 10 kHz to 20 kHz 20 kHz to 50 kHz 50 kHz to 100 kHz 40 Hz to 45 Hz 45 Hz to 65 Hz 65 Hz to 1 kHz

0.5 + 2 1.5 + 4 3.0 + 4 5.0 + 20 1.00 + 50 3.00 + 50 5.00 + 50 1.50 + 50 0.70

  • 50 1.50 + 50 3.00 + 50 5.00 + 50 10.0 + 50 1.50 + 50 0.70 + 50 1.50 + 50 0.70 + 50 1.50 + 50 3.00 + 50 1.00 + 50 0.40 + 50 1.00 + 50 2.00 + 50 3.00 + 50 5.00 + 50 10.0 + 50 1.00 + 50 0.40 + 50 1.00 + 50

(1) The bandwidth is 40 Hz to 5 kHz (2) The bandwidth is 40 Hz to 1 kHz

(3) Below 10% of range, add 2 digits to accuracy (4) Below 10% of range, add 10 digits to accuracy, < 50 kHz

(5) Below 10% of range, add 20 digits to accuracy, > 50 kHz

True RMS AC Current

Model

Range

Frequency

Accuracy

40 Hz to 70 Hz

390B

40.00 mA, 400.0 mA, 4.000 A, 10.00 A

70 Hz to 1 kHz

1 kHz to 10 kHz

40 Hz to 65 Hz

391B

40.000 mA, 400.00 mA, 4.0000 A, 10.000 A

65 Hz to 1 kHz

40 Hz to 65 Hz

393B

10.0000 mA, 100.000 mA, 10.0000 A

65 Hz to 1 kHz

– Below 5% of range, add 70 digits to accuracy – Max. Continuous Measuring Time: 10 minutes at mA inputs, 1 minute at A inputs

– Min. Rest Time: 20 minutes after continuous measuring – Overload Protection: AC/DC 400 mA at mA inputs, AC/DC 10 A at A inputs

0.8 + 2
2.0 + 4
0.8 + 80 3.0 + 80 0.7 + 80 2.0 + 80

Frequency Model 390B
391B & 393B

Range
400.0 Hz 4.000 kHz 40.00 kHz 100.0 kHz 40.000 Hz 400.00 Hz 4.0000 kHz 40.000 kHz 100.00 kHz

Resolution
0.1 Hz 1 Hz 10 Hz 100 Hz 0.001 Hz 0.01 Hz 0.1 Hz 1 Hz 10 Hz

Specifications 47
Accuracy 1d (3 3/4-digit mode) 5d (4 3/4-digit mode)
0.1 + 10

Temperature

Model

Range

390B 391B & 393B

-200 °C to 1200 °C -328 °F to 2192 °F -200 °C to 10.0 °C -10.1 °C to 1200 °C -328 °F to 50.0 °F -50.1 °F to 2192 °F

– Specified after 60 minutes of warm-up time – Specification does not include error of the thermocouple probe

Resolution
0.1 °C 0.1 °F 0.1 °C 0.1 °C 0.1 °F 0.1 °F

Accuracy
1.0 + 10 1.0 + 18 1.0 + 2 °C 1.0 + 1 °C 1.0 + 4 °F 1.0 + 2 °F

Supplementary Functions

Model

Function

AC + DC

390B

HFR

Peak Hold

AC + DC

HFR

391B & 393B

Peak Hold

dB dBm (600 )

Range
Same as voltage and current
Same as AC function Same as AC function 391B: Same as AC function x 1.25 393B: Same as AC function x 0.125
120.00 dB 120.00 dBm

Accuracy AC accuracy ± 1.0% AC accuracy ± 1.0% for 40 Hz to 400 Hz AC accuracy ± (3.0% + 200 digits) for 40 Hz to 1 kHz AC accuracy ± 1.0% AC accuracy ± 1.0% for 40 Hz to 400 Hz
AC accuracy ± (3.0% + 100 digits) for 40 Hz to 500 Hz
Not specified Not specified

Capacitance Model 390B
391B & 393B
(1) Available in manual range selection only

Range
40.00 nF 400.0 nF 4.000 µF 40.00 µF 400.0 µF 4.000 mF 40.00 mF 4.0000 nF 40.000 nF 400.00 nF 4.0000 µF 40.000 µF 400.00 µF 4.0000 mF(1) 40.000 mF(1)

Resolution
10 pF 100 pF
1 nF 10 nF 100 nF 1 µF 10 µF 0.1 pF 1 pF 10 pF 100 pF 1 nF 10 nF 100 nF 1 µF

Specifications 48
Accuracy 0.9 + 20 0.9 + 10
0.9 + 2
0.9 + 10 0.9 + 20 Not specified 1.2 + 200
0.8 + 20
1.2 + 200 1.2 + 400

General

Model Display

Measurement Speed

Data Log Capacity

Connectivity Power
Battery Life (typical) Auto Power Off
Low Battery Indicator Overrange

Temperature

Operating Storage

Safety Dimensions (W x H x D), without case
Weight Warranty

Standard Accessories

390B

391B

393B

4,000 / 40,000 count

10,000 / 100,000 count

10 times per second

3 times per second

40,000 measurements

20,000 measurements

IR-USB and Bluetooth (class 2)

IR-USB

4 x 1.5 V AA size batteries

50 hours

100 hours

Adjustable up to 30 minutes or never

OL or -OL is displayed

14 ºF to 122 ºF (-10 °C to 50 °C) at 80% relative humidity

-4 ºF to 140 ºF (-20 °C to 60 °C)

EN61010-1 to 600 V CAT IV / 1000 V CAT III

3.8″ x 8.2″ x 2″ (95 mm x 207 mm x 52 mm)

1.4 lbs (630 g)

3 Years

Test leads, K-type thermocouple adapter, protective case, optical-isolated USB cable, magnetic hanging kit, alkaline batteries

Service Information

Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device.
Non-Warranty Service: Please go to the support and service section on our website at bkprecision.com to obtain an RMA #. Return the product in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Customers not on an open account must include payment in the form of a money order or credit card. For the most current repair charges please refer to the service and support section on our website. Return all merchandise to B&K Precision Corp. with prepaid shipping. The flat-rate repair charge for Non-Warranty Service does not include return shipping. Return shipping to locations in North America is included for Warranty Service. For overnight shipments and non-North American shipping fees please contact B&K Precision Corp.
Include with the returned instrument your complete return shipping address, contact name, phone number and description of problem.
B&K Precision Corp. 22820 Savi Ranch Parkway
Yorba Linda, CA 92887 bkprecision.com 714-921-9095

LIMITED THREE-YEAR WARRANTY

B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To help us better serve you, please complete the warranty registration for your new instrument via our website www.bkprecision.com Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed. B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you. This warranty gives you specific rights and you may have other rights, which vary from state-to-state.
B&K Precision Corp. 22820 Savi Ranch Parkway
Yorba Linda, CA 92887 www.bkprecision.com
714-921-9095
Version: September 21, 2023

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals