MULTI PROTOCOL ANALYZER
LE-3500XR
LE-2500XR
Quick Start Guide
Thank you for your purchase of LE-series.
This booklet tells you only the basic operation. For more detailed
information, please refer to the instruction manual (PDF) in the utility CD
attached to the product.
When you unpack the product, confirm following items
□ Protocol Analyzer | 1 |
---|---|
□ Interface Sub-board (attached to the analyzer) | 1 |
□ DSUB 25pin Monitor Cable (Model: LE-25M1) | 1 |
□ DSUB 9pin Monitor cable (Model: LE-009M2) | 1 |
□ DSUB 25-9 conversion adapter | 1 |
□ 5wire TTL probe (Model: LE-5LS) | 1 |
□ Micro USB Cable | 1 |
□ Carrying Bag (Model: LEB-01) | 1 |
□ USB Battery Charger (Model: LE-P2USB) | 1 |
□ Utility CD | 1 |
□ Quick Start Guide (this booklet) | 1 |
□ Registration Card, Warranty | 1 |
USER LIMITATION
This product has been developed for the purpose of using as an analyzer only.
This product has not been developed for the use that needs exclusively high
reliability and safety:
aerospace apparatus, trunk communication apparatus, nuclear control apparatus,
medical apparatus related with life maintenance etc. Therefore, do no use for
those purposes.
This product has Wi-Fi function (IEEE 802.11b/g/n) and emits radio wave.
Please do not use near a medical device, microwave, high-level electronics,
TV, radio, wireless station for mobile communications, or specified low power
radio station.
Where an administrator limits the use of radio devices, follow the instruction
of the administrator.
Note : The Wi-Fi module used for this product conforms to SRRC(China),
FCC(USA), CE(EU), TELEC(Japan), KCC(Korea), ISED(Canada), NCC(Republic of
China), however, this product with Wi-Fi function is allowed to use in Japan,
USA, Canada, and EU nations (RE directive, 2014/53/EU).
The Wi-Fi function of this product is set to be invalid depending on the
country where it is shipped. Please contact the sales department for details.
Representation of the Operating Procedure
-
Key operation is described in [ ].
-
Continuous key operation is described as follows.
ex) Press [MENU] and then [0] → [MENU],[0]. -
Screen display is described in “ ”.
-
Reference page is displayed in (Pxxx).
It is prohibited to reprint or duplicate any part of this manual without prior
permission of LINEEYE.
The content of this manual and specification of the product is subject to
change without any notice.
Safety Information
Please do not use the analyzer in the following conditions.
[Description of the symbol and mark] Warning There is a possibility of
getting hurt, such as a death or a serious injury.
Caution There is a possibility of getting injured or damaging the
product.
Warning
- Do not disassemble, modify or repair the analyzer.
This may result in an injury, electric shock, and ignition. - Turn off the power and unplug the analyzer immediately when emanating smoke or odor.
Continuous use may result in an electric shock, burn and ignition.. - Do not use the analyzer if there is inflammable gas.
This may result in ignition and explosion. - Turn off the power and unplug the analyzer immediately when liquid or foreign substance gets into the analyzer.
Continuous use may result in ignition, electric shock and malfunction. - Do not touch the analyzer with wet hand.
This may result in an electric shock and malfunction. - Do not use the batteries other than the Lithium-Ion batteries specified by LINEEYE.
Do not short the electrode of battery, disassemble or modify.
This may result in generation of explosion and ignition.
Caution
-
Do not give a strong impact to the analyzer.
-
Do not place the analyzer in following conditions.
・ Not flat or vibrated place.
・ Temperature or humidity is above the specification.
・ Change the temperature rapidly.
・ Have a direct sun or near the fire.
・ Magnetic field. Have static electricity. -
Do not use the analyzer near the following devices.
・ Medical device, such as a heart pacemaker.
・ Automatic control devices easily affected by radio waves.
・ Devices controlled by the radio waves.
Panel Description
N a m e | D e s c r i p t i o n |
---|---|
① | Power Switch |
➁ | SD Card Slot |
③ | USB Host Port |
④ | USB Device Port |
⑤ | Interface Sub-Board |
⑥ | RS-232C Por t |
⑦ | TTL, External Signal I/O port |
external signal.
⑧| RS-422/485 Port| Measurement port for RS-422/485.
⑨| 4.3 Inch Color Display| Capacitive touch panel.
⑩| Line State LED| Indicating the logical status of signal lines on the target
interface.
⑪| Power LED| Lighting in green when using. Blinking in red when charging
battery.
⑫| Keypad| Entering commands and other operation.
⑬| Battery Cover| Open/close when replacing the battery.
Power Supply and Battery
This analyzer drives by the battery or bus power from a USB port.
-
USBport
Charge the Battery
Lithium ion battery is charged while running it by the USB bus power.
Charging time
When turning off the power: Take about 5 hours for full charge.
When turning on the power: Take several hours for full charge.
→ Change the charging type from the system menu.
Charge the battery under the condition of 5 °C to 40 °C. If the temperature is below 5, it cannot charge the battery.
Charge the battery full if you are not going to use it for a long term. -
Replace the Battery
If the analyzer can no longer drive by the battery, or the service time after charging becomes extremely short, replace the battery.
Remove the battery cover and replace the battery.
Battery is necessary for saving data in the storage (USB flash drive etc)when having the power shortage. Insert the battery all the time.
Warning < Notice for battery >
- Do not short the terminal of the battery.
- Use the battery and battery charger specified by LINEEYE.
- Do not disassemble or modify the battery.
- Do not use the analyzer near the heater or leave it in the place with high temperature.
It can cause heat, fire, explosion, electric shock, liquid spill and malfunction.
Touch Panel and Keys
Touch panels enable to select settings by soft touch or to scroll displays by swipe.
- Keys
Operations | Function |
---|---|
Turn ON/OFF the power. |
Press it longer to turn off the power.
[RUN]| Start monitoring, measuring or testing operation.
[STOP]| Stop monitoring, measuring or testing operation.
[MENU]| Return to the top menu.
[ESC]| Return to the previous display. Stop renewing data display while
measuring.
[ ▲ ] , [ ▼ ]| Scroll one data line. Move the cursor on the condition setting.
| Scroll one character of data.
Change the setting on the condition setting.
[ENTER]| Apply selected settings.
[0] ~ [F]| Enter numerical values. Select the data send table number.
[TOP/DEL]| Jump to the top of data. Delete the entry indicated by the cursor.
[END/X]| Jump to the end of the screen.
Enter the data of “Don’t Care”.
[SHIFT]| Press to use the expanded alternate function of each key.
[SHIFT]+[MENU]| Displays the operation menu while displaying measurement
data
[SHIFT]+[ESC]| Save a screenshot or print it by a printer * 1
[SHIFT]+[RUN]| Brighter the LCD back light.
[SHIFT]+[STOP]| Darker the LCD back light.
[ENTER]+[TOP/DEL]+| Initialize settings. Clear all.
[SHIFT]+[STOP]+| Start the firmware loader.
*This booklet uses [+] mark, when two keys need to be pressed at the same time.
- 1: Specify at the “Screen shot destination” in the Display/Print tab of the system settings.
Initial Settings
Press to turn on the power. Select language and set data & time.
*The top screen displays the model name.
Press [MENU] to set the initial settings. Swipe the screen to set all
settings.
Touch the setting item or press [ENTER] key after selecting the item by [▲] [▼] keys.
Touch “File operation” to save the settings or measured data.
The left top of the screen indicates the storage device.
The filename extension “.DT” represents measured data file and “.SU”
represents settings file.
Touch [ ] to change the storage device (SD or USB) to filer saved files,
or to filer the display by selecting the file type.
Double tap on the folder to check the files in the folder.
Interface Select a measurement port of analyzer.
For Simulation, select DTE or DCE mode and polarity.
For Monitoring, polarity has to be set as well.
Function Select a function.
→[Monitor Function] (P13)
→[Simulation Function] (P16)
→[BERT Function] (P19)
Select a mode for each function.
Select “ONLINE” for normal use of monitoring.
Depending on the function, touch
“Advanced settings” or press
[ENTER] to configure the operational condition.
LE-2500XR does not support TREND mode of the monitor function, BUFFER mode
and
PROGRAM mode of the simulation function.
Touch “Tx-data registration” to register the data to be used in the simulation
function.
Configuration Select basic communication conditions.
Protocol :
Select Asynchronous (ASYNC), Character synchronous (SYNC, BSC), Bit
synchronous (HDLC, SDLC), I2C or SPI.
Speed :
Touch [ ] and select the communication speed. It is possible to type any speed
(4 digits) by numerical key. ([C] represents “decimal point”. [D] represents
“kilo”. [E] represents “mega”.
Other settings such as data code, data bit, parity, stop bit need to be set
for some protocols.
BCC :
Select BCC type, begin code and end Code etc. If selecting transparent
mode,select DLE (Data Link Escape) code.
Bit sequence :
Select “LSB first” for normal use.
Frame end time :
Set the time of non-communication state which is recognized as a frame end.
Frame end code:
Frame end code can be set up to 2 characters.
Frame end code is necessary to record time stamp.
Necessary settings are different in each protocol.
Auto configuration is available (P25).
Record control Set recording conditions of measured data.
Shortcut key [MENU] ,[3] Touch the tab or press, SHIFT]+[▶] or [SHIFT]+[◀].
-
Additional info.
Idle time :
Select ON/OFF of idle time (noncommunication time) and time resolution.
Time stamp :
Select ON/OFF of time stamp and time resolution.
Line state :
Mark on the check box, then status of RTS, CTS, DCD, DTR, DSR, RI and TRG(trigger input) are recorded along with data. -
Capture buffer :
Recording area :
“BUF0” means full buffer. “BUF1/BUF2” means capture buffer is divided into two and use half of them for recording.
Protected :
Prevent data stored in the buffer memory from being inadvertently overwritten.
Write control :
Select Ring buffer (endless recording) or Full stop when the capture buffer is full.
-
Auto save
Set conditions of auto save function to record data in the storage device.
→[Auto Save Function] (P23)
→[Auto Backup] (P24) -
Auto run
Start measuring automatically at the specific period of time.
→[Auto Run Function] (P24)
Trigger Set the trigger conditions.
→[Trigger Function] (P20)
Wave monitor Set the conditions of logic analyzer function.
→[Timing Waveform Measurement function (Wave Monitor)] (P22)
System Setting and Version
Press [MENU] and then touch “System Settings”.
Touch the tab or press, [SHIFT]+[▶] or [SHIFT]+[◀].
- Display / Print
It is able to adjust the brightness of LCD back light, set the power saving mode,
Press [SHIFT]+[ESC] to save or print the screen images.
- Power
Set ON/OFF of Auto Power Off function and select the charging type of Lithium ion battery.
Auto Power Off function does not work during measurement.
-
Wireless LAN
To enable this function, touch “Setting change” and select “Stations” mode or “Access Point” mode.
(Factory setting of Wireless LAN is Off.)
Some models without this function do not have this setting. -
Version
Serial number and firmware version are displayed.
Connect to the Target Devices
-
Connect to RS-232C
Connect RS-232C connectors of the target devices to the RS-232C port of analyzer using monitor cable. -
For monitoring
If the connector of either target device is Dsub 25 pin, use attached Dsub 25
pin monitor cable (LE25M1) to connect to the analyzer.
If the connectors of both target devices are Dsub 9 pin, use attached Dsub 9
pin monitor cable (LE-009M2) and Dsub 25-9 conversion adapter to connect to
the analyzer.
- For Simulation
Connect analyzer and target device one by one. Connect as following, concerning the specification of target device (DTE/DCE) and RS-232C cable.
DTE device —–Straight cable —–Analyzer (DCE setting)
DCE device —–Straight cable —–Analyzer (DTE setting)
DTE device —— Cross cable ——-Analyzer (DTE setting)
DCE device ——Cross cable ——-Analyzer (DCE setting)
- Connect to RS-422/485
Connect appropriate signals of target device to the RS-422/485 port.
Connect TXD+ and TXD-, and RXD+ and
RXD- using twisted pair cable.
*This is the removable terminal.
Remove it from analyzer first, connect cables and return it to the analyzer.
- RS-422 Full-duplex
- RS-485 Half-duplex
Monitoring/Simulating
If the analyzer is located as the terminal of communication line, set the Terminal resistance 120 Ω jumper pin to be “1” on the interface sub board. (JP1: TXD, JP2: RXD)
- Connect to TTL interface
Connect appropriate signals of UART, I2C and SPI to the TTL port of analyzer using attached probe cable.
MIL box type 10pin header *1 | Cable color |
---|---|
Signal Name | Pin |
Output
| LE-5LS
TXD, SDA of I2C, SDO of SPI| 1| I/O2| Brown
RXD, SDI of SPI| 3| I| Red
RTS, SS(CS) of SPI| 5| I/O2| Orange
CTS, SCL of I2C, SCK of SPI| 7| I/O2| Yellow
Signal ground| 9| –| Green
Signal ground| 2| –|
IN: Trigger input| 4| I|
OT1: Trigger output 1| 6| O|
OT2: Trigger output 2| 8| O|
PWR: Power for external3| 10| –|
-
1: 2.54mm pitch. Equivalent to “HIF3FC-10PA-2.54DS(71)” of HIROSE Electric.CO.,LTD.
-
2:It is used for output of simulation, except SPI slave simulation (SS and SCK are inputs).
-
3: TTL voltage (max. 30mA) is outputted when simulating.
- External Trigger Terminal
External trigger input level is LVTTL (3.3V). External trigger output is open collector output type and pulled up in 5V by 10KΩ in the analyzer.
Attached 5-wire TTL prove can be used for connection of external trigger. To use the TTL proves for both TTL and external trigger, prepare additional one.
→ [Trigger Function] (P20)
- External Trigger Terminal
5wire TTL probe (LE-5LS)
If the target devices are easily affected by radiation noise from the analyzer, try to use the shielded cable and add the ferrite core on the cable.
Monitor Function
- Start and Stop measurement
Press [RUN] to start monitoring. Measured data is displayed at real time and recorded in the capture memory. Send (TXD) data and receive (RXD) data are displayed in two lines as a pair. The latest data is displayed on the left side of “ ▢” marker. Press [STOP] to finish measuring.
When TXD and RXD are generated at the same time, they are displayed on the
same column.
Press [ESC] to pause the display.
It is able to stop measuring automatically by setting the trigger function and
record control (full stop setting on capture memory).
Errors and breaks are displayed in unique characters.
Character | Description |
---|---|
Parity Error (parity bit error). | |
Framing Error (stop bit is “0”). | |
Parity & Framing Error . | |
Break (start bit, character bit, (parity bit), stop bit are all “0”.). | |
Abort (7bits or more of “1” is continuously detected). | |
Shart Frame (frame length is short). | |
Block check code normal (BCC or FCS is normal). | |
Block check code abnormal (BCC or FCS is abnormal). | |
HDLC starting flag pattern (7Eh) is detected. | |
HDLC ending flag pattern (7Eh) is detected. | |
Multi processor bit or acknowledge bit of I2C is “1”. (displayed in blue |
back).
| I2C (re)start sequence is detected.
| I2C stop sequence is detected.
| Overrun Error.
| Overload (Cannot save all data by auto save function ).
① | Measuring 〔 〕 , Pause 〔 〕 . |
---|---|
➁ | Position of data/ all data. Able to type the position of data. Speed and |
character framing while measuring (e.g. B8-PO-S1). (1)
③| Selected interface.
➃| Selected Mode. [Mon]: Monitoring, [DTE/DCE/Sim]:Simulating.
⑤| Status of USB device port (Bright: Connected).
⑥| Status of USB host port (Bright: Connected). (2)
⑦| Status of SD card slot (Bright: Connected). (2)
⑧| Level of remained battery. 〔 〕 indicates “USB bus power”.
⑨| Display more setting items.
⑩| Change displayed mode. –>Normal–>Line state–>Each frame/ Translated (3) —
——————Logic analyzer<—————————
⑪| Selected data code. Change data code.
⑫| Change display. HEX↔Data code.
⑬| Find specific data and errors.
⑭| Start retrieval.
-
1: [B: Data length], [P: Parity (E: Even, O: Odd, N: None)], [S: Stop bit] * 2: Become red while accessing to the storage device.
-
3: Translated display is difference by selected protocol.
Scroll data by swiping on the screen.
Swipe in upward to look older data.
Swipe in downward to look newer data.
Press [SHIFT]+[▲ ]or [SHIFT]+[ ▼]to go to next/ previous page.
Do not remove the USB flash drive and SD card while measuring and displaying “under cleanup”.- Signal Voltage and Delay Time
Press [MENU]->“Function”->“MONITOR”->“DELAY Function” and then touch “Advanced settings”.
Set the start/ stop conditions of delay time for target signals.
- Signal Voltage and Delay Time
Press [RUN] to display the delay time and voltage (max./ min.).
Use “Wave monitor” for delay time less than 0.1ms.
- Statistic Function (TREND) [LE-3500XR only] Press [MENU]->“Function”->“MONITOR”->“TREND” and then touch “Advanced settings”.
EVENT :
Select “Data”, “Frame” or “Trigger” to calculate.
Trigger: The number of the events which satisfies the trigger conditions set
in factor of “TRIGGER 0” and “TRIGGER 1” are counted.
Press [RUN] to start the statistic. As unit time of statistical processing
passes, the results of calculated value will be updated on a bar graph.
Simulation Function
You can execute a communication test for a device in development by sending/receiving data by the analyzer depending on the selected mode.
- Registration of Transmission Data
Select the simulation function at “Operation mode” of [MENU], and then touch
“Tx-data registration” to display the transmission data table.
Up to 160 types of data can be registered in the 160 tables of 0 to 9 groups
(each group has 16 tables). Touch the table and group to register the data in
it.
Input the data to be transmitted in HEX by [0] to [F].
Copy and paste of the data are available by [ENTER], [ ▶ ], and [ ◀ ] .
By touching “PE/MP”, the data on the cursor turns to be parity error or
multiprocessor bit, and by touching it again it returns to be the original
one.
You can input text by touching “Text input”.
Input the text by capital letter or symbol and touch [ENTER] or “OK”, then the input text line is inserted to the cursor position.
Touch “ ” to display menus such as fill data set, copy table, buffer copy. Touch each item to execute.
-
Start and Stop Measurement
Touch [RUN] to start the specified test function and touch [STOP] to stop it.
When the analyzer is set to DTE, The TXD line means the data is transmitted from the analyzer. When the analyzer is set to DCE, the RXD line means the data is transmitted from the analyzer. -
MANUAL Mode
Checking the data from the connected device on the display, you can transmit the preset data which you registered in advance by touching [0] to [F] or [SHIFT]+[0] to [SHIFT]+[D].
While putting [SHIFT], the control guide for preset data is displayed.
It can output break signal by [END/X].
You can specify the interval of each data, on/off of repeated transmission, and the interval of data frame.
-
FLOW Mode
This unit becomes the transmission side or reception side, and stats the simulation of flow control using X-ON/OFF and control lines.
When “Send test” is set, it can display the data of transmission between the start of transmission and the interruption request for 16 times.
In “Receive test”, you can specify the number of reception data until the transmission interruption request is issued and the time until the restart request of transmission is sent.
-
ECHO Mode
Received data will be returned inside the analyzer and returned to the sender. This mode is used for testing of devices such as display terminals and communication terminals.
When “Loop back” is set, data can be returned in bit units.
-
POLLING Mode
When the master mode is set, polling messages are sent to 32 types of slave address stations, and the returned data is checked and displayed for each slave station. When the slave mode is set, the specified data is responded to the polling of the own station address, and the number of times and the presence/absence of an error are checked and displayed.
-
BUFFER Mode [LE-3500XR Only] From the send/receive data stored in the memory by the monitor function, it transmits the data of send or receive line set in advance as it is.
This is effective for a reproduction test using the same data as the communication status monitored on site.
-
PROGRAM Mode [LE-3500XR Only] You can easily test a communication protocol that involves condition judgment using a test program created using dedicated commands.
You can register four types of test programs A to D in advance by touching “Edit”.
-
Contents of setting example
000 : Label 000
001 : Waits for new received frame
002 : Transmits data of table 08
003 : Branches to label 000 when receiving N, G, CR, LF
004 : Increments the register 0
005 : Branches to label 000
BERT (Bit Error Rate Test) Function
- Measurement Start and Stop
Select the test mode, test pattern, and on/off of RTS-CTS flow control in advance. By pressing [RUN] it starts transmission of the specified test pattern and compares the received data with the test pattern. Then it measures the bit error rate etc. in compliance with ITU-T G.821 and displays them on the screen. It stops measurement by [STOP].
Set the synchronization method (ASYNC or SYNC/BSC) and the communication speed at the “protocol” and “speed” in the configuration. For the other items, only the data bit and stop bit settings are valid for ASYNC, and only the clock settings is valid for SYNC/BSC.
Item | Contents | Range of measurement |
---|---|---|
Savail | Effective time since the first syn-chronization was established | 0 ~ |
9999999(sec)
Receive bit| Number of received bits during synchronization establishment| 0 ~
9999999 ~ 9.99E9
Error bit| Number of bit errors| 0 ~ 9999999 ~ 9.99E9
Bit error rate| Bit error rate| 0.00E-0 ~ 9.99E-9
Sync loss| Number of loss of synchronization| 0 ~ 9999
Receive block| Number of received blocks during synchronization establishment|
0 ~ 9999999 ~ 9.99E9
Error block| Number of blocks with bit errors| 0 ~ 9999999 ~ 9.99E9
Block error rate| Block error rate| 0.00E-0 ~ 9.99E-9
Error second| Time when bit error was detected during Savail| 0 ~ 9999
%E.F.S.| Errored second rate (%)| 0.000 ~ 100.000(%)
Useful Functions
- Trigger Function
By this function it executes a special measurement processing (“Action”) triggered by specific conditions (“Factor”) such as the occurrence of a communication error during the measurement or reception of the specified data.
〔Factor〕
Error | Communication error, break, multiprocessor bit on |
---|---|
Character | Communication data of up to 8 characters |
Line state | Logic state of interface signal line and external trigger input |
Timer/Counter | Timer/Counter value match |
Idle time | Idle time exceeding the specified time |
〔Action〕
Buzzer | Buzzer sound |
---|---|
Stop measurement | Stops the measurement/test (offset to stop can be |
specified.)
Save data| Save data before and after the trigger to the storage device
Timer control| Start/stop/restart of the timer 0/1 for trigger
Counter control| Count up/clear of the counter 0/1 for trigger
Trigger control| Enables/disables/inverts the state of trigger condition
Transmit data| Transmits the specified character string (in MANUAL simulation)
OT2 pulse output| Outputs an L pulse of about 1 ms to external trigger
terminal OT2
Press [MENU] and touch “Trigger” to display the current trigger settings.
The trigger number with check mark [ ] is valid. If you check more than one,
it will be OR condition.
Touch or [SHIFT] + [0] to [3] to turn the check mark on/off. Touch the
factor or action to be changed and configure it.
Touch “Timer/Counter” settings to set the comparison values for timer/counter
used in the trigger function.
These timers and counters are cleared at the start of measurement, and the
timer will be stopped state.
Setting example 1:
When you want to notify an external measuring instrument that a block check
error has occurred.
Turn on the check mark of the trigger to be used and touch the trigger factor.
Select “Error” on the trigger factor setting and check the BCC error you want
to detect. Press [ESC] to return to the trigger setting screen, then touch the
corresponding trigger action and select “OT2 pulse output”. With this setting,
when a BCC error is detected, it outputs an L pulse of about 1 ms to the
external trigger output terminal OT2. Observe the OT2 signal by connecting
measurement device such as an oscilloscope.
Regardless of the trigger action setting, it outputs outputs an L pulse of
about 1 ms to the external trigger output terminal OT1 when any trigger factor
matches.
Setting example 2:
When you want to stop the measurement immediately when receiving the character
string of 41h, 42h, 30h or 41h, 42h, 31h on the transmitting side.
On the setting screen of the checked trigger factor configuer as shown on the
right. For the third byte of the character string to be detected we use the
bit mask W0 to set bit 0 to don’t care (X).
Press [ESC] to return to the trigger setting screen and touch the
corresponding trigger action.
Select “Stop measurement” on the trigger action setting screen.
You can specify whether to stop measurement immediately or stop after
receiving data by “Stop”.
- Timing Waveform Measurement Function
This function periodically records the logical state of the communication line and displays the timing waveform like a logic analyzer.
Press [MENU] and touch “Wave monitor”.
Mark on the box of Enable wave monitor and configure the sampling clock (50ns
to 1ms), trigger position, and trigger pattern.
Set the sampling clock to approximately 5 times the communication speed as a
guide.
“↓” and “↑” can also be entered by [SHIFT] + [0] or [SHIFT] + [1].
When measurement is started by [RUN], the status of the communication line is
recorded in the waveform recording memory at the specified cycle while normal
operation is executed.
will be displayed at the upper left of the screen until the trigger is
established and the waveform recording ends.
After the measurement is stopped, press “Change display” for several times to
switch to the following screen.
Touch the magnification to zoom in and out, and swipe left or right to scroll
to the part you want to see.
The red vertical line is the trigger point. At first, cursor C (blue dotted
line) and marker M (red dotted line) overlap the trigger point.
〇 Time measurement between two points
The time between the cursor and the marker is displayed at “C-M:”.
The cursor can be moved by [▶ ] and [◀ ], and becomes a marker by touching
[ENTER] or “Mark”.
〇 Search waveform
Touch “Search” and enter the state of the waveform you want to search by [ ▶]
and [ ◀ ] to move the cursor to the first matching waveform in that direction.
Touch “Search” again to exit from the search mode.
- Auto Save Function
With this function, it automatically saves the communication data as a measurement log file in the optional SD card or USB flash drive while measuring. Since long time communication data can be recorded, it is useful to clarify rare communication failures of unknown cause.
Press [MENU], and then touch “Record control” and “Auto save” in order.
Save device:
Set the save destination of the log file.
Insert the SD card or USB flash drive as selected.
Auto save :
Restart Continuous ring recording up to the maximum number of files.
Max-stop Saves up to the maximum number of files and stops measurement.
Append Continuous ring recording from the continuation of the existing auto
save file.
Please note that except for “Append” all the existing auto save files will be
deleted at the start of measurement.
Set the main unit memory to “Ring buffer”.
→ [MENU] ,“Record control”, and then “Write control” of “Capture buffer”.
File size :
Set the size of the communication log file to be automatically saved.
Max files :
Set the maximum number of files to be recorded.
The communication log file name is # XXXXXXX.DT (XXXXXXX is a sequential
number that increases by 1 from 0000000).
When you start the measurement by [RUN], a confirmation message for
overwriting the file will be displayed. Touch [RUN] or “Continue” to start
measurement. If you do not want to lose the communication log file and want to
save it to another device, press
[STOP] and save it to a PC or other device.
If the power is turned down during auto save because of low battery charged,
it may cause damage to the saved files and the storage device. Start
measurement with the sufficiently charged battery.
- Auto Backup Function
Since the capture memory is DDR3-SDRAM which loses data when the power is turned off, at the end of measurement the analyzer automatically backs up the about 512 KB of the latest part of the measurement data into the SRAM backed up by the battery. To automatically save the entire measurement data or to erase the data when the power is turned off, change the setting of the auto backup function.
Press [MENU], and then touch “Record control” and “Auto save” in order.
Auto backup:
Off No automatic backup.
Save to SRAM Save about 512 Kbytes of the latest data into internal SRAM.
Save to File Save all the measurement data to the storage device
The data saved in the SRAM is automatically loaded into the capture memory
when the power is turned on.
The file automatically saved in the storage device, @AUTOBUn.DAT (n is the
number of the used capture memory – BUF0, BUF1, or BUF2), needs to be loaded
manually to use.
- Auto Run/Stop Function
By specifying the date and time for the start and end of measurement, it automatically executes measurement for the specified period.
Press [MENU], and then touch “Record control” and “Auto run” .
Mode:
Select the repeat condition for automatic measurement.
Touch to activate the checked item.
Set the automatic measurement period for “Run time” and “Stop time”.
With the settings on this screen, measurement starts at 12:30 every day and
ends automatically at 13:00.
If you want to start the measurement in conjunction with the power on of the
target line, check “External power on Run”.
- Communication Condition Auto Setting Function
The monitor processing of the analyzer estimates the communication conditions
of unknown communication lines for measurement.
By touching “Auto Configuration” in the [ ] menu, “Analyzing transmission
speed…” is displayed and the analysis process of communication condition
starts.
When the communication conditions are determined, the details will be displayed. A few seconds after that, the current communication condition settings are automatically changed to the determined communication conditions and it starts the measurement.
If you touch “Cancel” or press [STOP] or [ESC] before starting measurement,
the communication condition analysis process will be stopped and the current
communication condition settings will not be changed.
The following conditions are required for the line to be measured in order to
judge the communication conditions correctly.
- Communication speed is 460.8Kbps or less.
- Protocols are ASYNC, SYNC/BSC, or HDLC/SDLC.
- On the line communication data without error is flowing at a certain frequency.
- On the line communication data including bit patterns of ‘101’ or ‘010’ is flowing.
- Conversion and saving, and print function for the measurement data
Measured data can be converted to text and saved to a storage de vice, or printed continuously by a dedicated printer. (Version 1.05 or later)
In the [ ] menu at the bottom left of the measurement data display screen, touch “Print out”, specify the output destination and the target range from the current display screen, and click “OK”. When the output destination is the storage device the text file is saved, and it is a dedicated printer the measurement data is printed out continuously.
Measurable Interface Expansion
By using the option for measurement expansion, you can expand the communication target to be measured.
- Exchange the board
Remove the standard interface sub board and replace it with the optional interface sub board.
- Loading and switching of extended firmware
The firmwares for expansion options are available on the LINEEYE website. Load it to the analyzer using le8firm.exe included in the attached CD. Once written, the corresponding firmware starts up simply by replacing the sub- board.
When using the optional high-speed firmware OP-FW10XR, you can switch between standard firmware and high-speed firmware by turning on the power pressing [SHIFT] and [0] or [SHIFT] and [3].
PC Link Function
You can remotely operate the unit from a PC and update the unit’s firmware. Connect this unit to a PC via USB or Wi-Fi.
-
USB connection
Install the USB driver in the “Driver” folder in the attached CD or on the LINEEYE website to your PC.
Execute setup.exe without connecting the analyzer to the PC. -
For Wi-Fi connection
At the system setting of the analyzer select the station mode or the access point mode and set the SSID and password required for Wi-Fi connection. -
Check with your network administrator about the network settings.
- Installation of PC link software
The PC Link software offers remote monitoring and conversion of measurement data to text files on a PC. The “lite version” with some restrictions can be downloaded free of charge from the LINEEYE website.
Unzip the downloaded file and execute setup.exe in the folder.
- Installation of PC link software
-
See the online help of the PC link software for how to use it.
[Example of Key emulation Screen]
- How to update the firmware
The latest firmware is available on the LINEEYE website below.
https://www.lineeye.com/html/download_update.html
Download the firmware and update it from your PC using the firmware transfer software le8firm.exe included in the “Utility” folder of the attached CD.
Specifications of Function and Hardware
Item | LE-3500XR | LE-2500XR |
---|---|---|
Interface | RS-232C, RS-422/RS-485, TTL (1.8V, 2.5V, 3.3V, 5V level) | |
Protocol | ASYNC (Asynchronous), ASYNC-PPP, Character synchronous SYNC/BSC, Bit |
synchronous HDLC/SDLC/X.25, I2C, SPI, Burst
Modbus| –
Expansion measurement option1| CAN/CAN FD/LIN, CAN/CAN FD/CXPI, X.20/21,
RS-449, V.35, RS-530, Current loop communication
Capture memory| DDR3-SDRAM 100MB, 2-split usage is available
Baud rate (bps)| 50bps ~ 3.150Mbps2| 50bps ~ 1.544Mbps2
Speed setting range| Freely set separately for transmission and reception.
(Margin of er-ror: ±0.01% or less)
Data format| NRZ, NRZI, FM0, FM1
Data code| ASCII, EBCDIC, JIS, Baudot, Transcode, IPARS, EBCD, EBC-DIK, HEX
Measurement function| Monitor function, Simulation function, BERT function
Buffer transmission| 〇| –
Program mode| 〇| –
Statistic analyze function| 〇| –
External trigger| LVTTL input 1, Open collector output 2
Signal voltage measurement| ±18V input range3, 0.1V resolution
LED| Green/red two-color light, 11 for line status, 1 for power supply
Display| 4.3 inch TFT color display (480x272dot)
Operation part| Capacitance touch panel and 29 buttons
USB 2.0 device port| Micro B connector, High speed
USB 2.0 host port| Standard A connector, High speed 4
SD card slot| Supports SDHC card5. Conforms to the SD association standard.
Wi-Fi connection
| IEEE 802.11 b/g/n
Frequency range : 2400MHz-2483.5MHz
Transmission power : +20dBm(802.11b), +17dBm(802.11g),
+14dBm(802.11n)
Power| USB bus power 5V/1A Lithium-ion rechargeable battery operation time:
about 7 hours *6
Temperature| In operation : 0 to 40 degree Celsius, In storage : -10 to 50
degree Celsius
Humid| 5 to 85% RH (no condensation)
Standard| CE (class A)
Size & weight| 190(W)x153(D)x38(H)mm, Approx 550g
- 1 : Optional measurement board or optional cable is required depending on the target.
- 2 : Firmware Ver.1.04 or later.
- 3 : SD (TXD), RD (RXD), ER (DTR), and CD (DCD) of RS-232C and 4 communication lines of TTL port can be measured.
When measuring the TTL port, the input voltage range is -1 to 6V. - 4 : For a USB memory or dedicated printer. Not all USB memory devices and printers are supported.
- 5 : For optional SDHC card sold by LINEEYE.
- 6 : Based on our measurement conditions.
For any technical issues
Read “FAQ” in our website or email us.
https://www.lineeye.com/html/contactus.html
LINEEYE CO., LTD.
4F., Marufuku Bldg., 39-1, Karahashi Nishihiragaki-cho, Minami-ku, Kyoto,
601-8468, Japan
Tel : 075(693)0161 Fax : 075(693)0163
URL https://www.lineeye.com
Printed in Japan
M-53235XQE/LE
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>