HomeBrace MyEcc Control Wheelchair With Your Eyes
Purpose
The special wheelchair control MyEcc Control is a manufacturer-independent power wheelchair control in combination with an eye gaze control system (like ALEA, Tobii), With this special control it is not only possible to drive the wheelchair (and to stop it by closing the eyes), it is also possible to adjust seat-settings of the wheelchair or to move attachments (robot arms, etc.). Any other use is NOT in accordance with intended use and leads to exclusion of liability and loss of warranty (optional/ mandatory interfaces are described in section 9). The same applies to unauthorized retrofits/modifications. Further warnings and safety instructions are provided in the manual.
Indication and Contraindication
Indication
Indication The product is indicated for anyone unable to perform voluntary
muscle movements (such as using a joystick (hand/foot/ chin or other body
part) or pressing a button (with a body part)). The eyes / the eye can
consciously and safely navigate to a target visually.
Possible medical conditions (this is not a complete list):
Amyotrophic lateral sclerosis (ALS), Multiple sclerosis (MS), Duchenne’s
muscular dystrophy, Infantile cerebral palsy (ICP), Joint contractures, …
Contraindication
- Complete loss of vision in both eyes
- Insufficient cognitive abilities to be able to control the environment/wheelchair in a targeted and safe manner
Warnings
This manual contains warnings to protect you from damage to property and
physical injury. Always read and observe these warnings. Follow all the
instructions in this operating manual and comply with all the actions marked
with a warning symbol and word of warning. Read this manual carefully before
starting up your MyEcc Control. Keep this manual. You may need to consult it
later. If you give the device to anybody else for their use, share this
operating manual as well.
Do not open the device. It does not contain any parts requiring user maintenance. Opening the device may give you an electric shock and damage the components. Avoid exposure to moisture, dust, sunlight, and other sources of heat.
Warning: Risk of personal injury
Non-observance results in death or serious injury.
Product liability
Pursuant to the liability of manufacturers for their products as defined in
the Product Liability Act, the instructions provided in this brochure must be
observed (product information and intended use, improper use, product
performance, product maintenance, information and instruction obligations).
Non-observance releases the manufacturer from any liability.
MyEcc Control is only meant to be an interface between the power chair and
MyEcc Control. Any other use is improper and will result in loss of warranty
and liability. The same applies to any alterations and modifications.
Important safety instructions
Only experts authorized by HomeBraceGermany may perform assembly, startup, and
maintenance. This in particular means the individuals referred to in the
following as “installers”; for further details, please consult item 13 of the
operating manual. HomeBraceGermany will not be liable for any damage or loss
resulting from unauthorized changes to the product. HomeBraceGermany accepts
no liability or warranty if MyEcc Control is combined with other brands, in
particular any improper use of MyEcc Control with other devices.
Observe the maximum power supply your peripherals can use (USB port 5VDC;
100mA). The use of the mounted model is permitted only by individuals with the
appropriate mental capacity and cognitive function. Non-observance of this
precaution may endanger the user or other people.
In addition, please observe the information for users in item 16.
HomeBraceGermany accepts no liability or warranty in the event of
infringement.
Maintenance and cleaning
- The product does not require regular maintenance. We recommend regular cleaning and disinfection with a disinfectant and cleaning agent intended for use in hospitals. The cleaning should only be done with a
- wipe. Wipe all exterior surfaces and allow them to dry by themselves.
- For any repairs, consult an expert. If a malfunction occurs, send your device to our after-sales service.
Mandatory review of the mounted system by the installer
Suitable measures should be taken to check for and protect against crushing,
impact, shearing, and pinching. The following areas require special review and
monitoring:
- Check that the software feature to stop the wheelchair by closing the eyes is working properly (the software must self-deactivate)
- Speed
- Seating options and settings
- Environmental hazard check!
The following monitoring mechanisms are actively monitored and will stop the wheelchair in the event of malfunction
- Camera contact with the eye
- Running ability of the eye control, monitoring of the Windows operating system
- Connection from the eye control to the MyEcc Control
- Cables and connectors from the eye control to the MyEcc Control
MyEcc Control requires the following interfaces with the wheelchair
- Up to 8 jacks, 3.5 mm module, e.g. switch module
- 9-pin D-sub connector, e.g. Omni Display
Wheelchair procedure
The power chair needs to be prepared for switch control by the installer
(authorized specialists only!). This step is not included with MyEcc Control;
please contact your contractual partner (system installer or manufacturer) for
this purpose. There is no need to install software, just copy a folder. Window
size and relay activation are easily changed using a file included in the
pack. We have taken great pains to maximize safety and reliability.
For the optimal use of the wheelchair with our MyEcc product group we
recommend a corresponding driving stability module of the wheelchair
manufacturer (ESP / Gyro).
Installation on the wheelchair
The eye gaze control system’s position on the wheelchair needs to be chosen
properly to suit the user’s needs. For eye gaze control of seating options
such as changing the backrest position, the eye gaze control system needs to
be mounted on the back of the wheelchair. Ensure that the wheelchair has a
dynamic length adjustment feature. This is essential in order to make sure
that the chair can be operated by the eyes in any wheelchair position.
TherThere is a risk that the user may e is a risk that the user may go out of the eye gaze control out of the eye gaze control system’s scan range systems scan range. .
- The user needs to have the wheelchair display in sight. The best option is to mount a second display (e.g. Omni) at eye gaze control level, which will allow the user to switch between modes with ease.
- If you intend to set the seating positions directly via the eye gaze control system’s 4 option buttons, this second display is not strictly necessary.
Startup
When wiring and assembling, please check the wheelchair routes, given the risk that poorly run cables may be torn off. Attach the MyEcc Control control box to the wheelchair in a (waterproof) bag if possible, preferably to the backrest or armrest. The cords to the eye gaze control system are detachable and should also remain removable after assembly so that carers can disassemble them easily when transferring the eye gaze control system.
WARNING: You must check the MyEcc Control safety features.
Preparing the eye gaze control system
The eye gaze control system has to move the mouse pointer on the Windows
interface. The camera driver needs to be addressed directly to ensure safe
operation. No third-party software such as Grid may be used for camera control
while driving the wheelchair.
WARNING: You must check the MyEcc Control safety features.
Software and interface for wheelchair control compatible with eye gaze control on Windows PCs
Installation
The application runs on Windows PCs with Windows 7, Windows 8, and Windows 10
and uses a USB connector (USB
To install, copy the MyEcc folder from the USB stick supplied with the pack to
the tablet PC’s program path, e.g. to C:/Program Files. The folder contains
the manual and 3 files: usbiodll.dll, matrix32.dll, eccapp.ini, and
MyEccApp.exe. Depending on requirements, you can copy a link from the
MyEccApp.exe executable file to the desktop and/or quick launch bar and to the
auto startup folder. Connect the control box to a USB port on the tablet. The
power supply is via USB (current consumption at 5V max. 200mA –> power
consumption max. 1W when the relays are activated, approx. 0.3W at rest. 8
LEDs on the control box show which relays are switched on.
Connecting to the wheelchair
Functions 1 – 5 (drive forward / right/back/left + function 5) are on a 9-pin
D-sub connector as indicated in R-Net I/O module SK78814/1. (D-sub pins 1, 2,
3, 4, 6, common: pin 8). Functions 6 / 7 / 8 are available on separate jacks.
- up arrow
- right arrow
- down arrow
- left arrow
- center button 1
- center button 2 7= center button 3 8= center button 4
Parameter settings
The eccapp.ini file contains parameters to adjust the application’s appearance
and behavior:[GENERAL]
[GENERAL]
- SIZEX= 600 Initial window size, horizontal (pixels)
- SIZEY=6 00 Initial window size, vertical (pixels)
- OFFSETX =700 Initial window position, horizontal (top left corner) on screen (pixels)
- OFFSETY=3 00 Initial window position, horizontal (top left corner) on screen (pixels) on screen (pixels)
- OPACITY Opacity (inversely related to transparency) of screen dialogue. Values 10 – 100
- SWITCHOPACITY Opacity for short-term feedback when switching on a button, values 10 – 100
[BUTTON]
- HOLDFOCUSTOACTIVE =400 Time to activate button when focusing (ms)
- FOCUSTIMEOUT= 600 Time between exiting/defocusing a button or the entire application (mouse downtime) until the active button is deactivated (ms)
- DRIVEDELAYGAP =150 Timeout when switching between drive buttons so that drive relays do not switch off briefly when driving the gap between buttons. (ms)
[RELAIS]
- PULSELENGTH= 25 Pulse length to activate drive and switch relays (ms). This setting should not need to be changed.
- PULSEINTERVAL= 120 120 Pulse interval to activate drive and switch relays (ms). This setting should not need to be changed.
- SWITCHONTIME= 500 Mode button 1 – 4 switch on time (ms),
- SWITCHWAITTIME= 800 Mode button 1 – 4 initial pause time to 1st switch (ms),
- SWITCHPAUSETIME =600 Pause between repeated switching to switch 1 through 4 (ms)
[MAPPING]
FUNCTION1= 1
FUNCTION8 =8 Here you can swap the mapping of functions to the output
relays. Default- mapping is 1:1. To swap functions 2 (right) and 4 (left), for
instance, enter:
- FUNCTION1= 1
- FUNCTION2 =4
- FUNCTION3 =3
- FUNCTION4= 2
[BUTTON_POSITIONS]
- M_DUP_LEFT_USED=1
- M_DUP_LEFT_OFFSETX=11
- M_DUP_LEFT_OFFSETY=6
- M_DUP_LEFT_WIDTH=283
- M_DUP_LEFT_HEIGHT=222
- M_DUP_LEFT_IMAGEFILE=
- M_DUP_USED=1
- M_DUP_OFFSETX=307
- M_DUP_OFFSETY=6
- M_DUP_WIDTH=283
- M_DUP_HEIGHT=222
- M_DUP_IMAGEFILE=
- M_DUP_RIGHT_USED=1
- M_DUP_RIGHT_OFFSETX=602
- M_DUP_RIGHT_OFFSETY=6
- M_DUP_RIGHT_WIDTH=283
- M_DUP_RIGHT_HEIGHT=222
- M_DUP_RIGHT_IMAGEFILE=
- M_DRIGHT_USED=1
- M_DRIGHT_OFFSETX=676
- M_DRIGHT_OFFSETY=236
- M_DRIGHT_WIDTH=208
- M_DRIGHT_HEIGHT=220
- M_DRIGHT_IMAGEFILE=
- K1_USED=1
- K1_OFFSETX=228
- K1_OFFSETY=274
- K1_WIDTH=111
- K1_HEIGHT=135
- K1_IMAGEFILE=
- K2_USED=1
- K2_OFFSETX=558
- K2_OFFSETY=274
- K2_WIDTH=111
- K2_HEIGHT=135
- K2_IMAGEFILE=
- M_F1_USED=1
- M_F1_OFFSETX=359
- M_F1_OFFSETY=236
- M_F1_WIDTH=190
- M_F1_HEIGHT=222
- M_F1_IMAGEFILE=
- M_F2_USED=0
- M_F2_OFFSETX=455
- M_F2_OFFSETY=236
- M_F2_WIDTH=89
- M_F2_HEIGHT=100
- M_F2_IMAGEFILE=
- M_F3_USED=0
- M_F3_OFFSETX=361
- M_F3_OFFSETY=356
- M_F3_WIDTH=89
- M_F3_HEIGHT=100
- M_F3_IMAGEFILE=
- M_F4_USED=0
- M_F4_OFFSETX=455
- M_F4_OFFSETY=356
- M_F4_WIDTH=89
- M_F4_HEIGHT=100
- M_F4_IMAGEFILE=
- M_DDOWN_LEFT_USED=1
- M_DDOWN_LEFT_OFFSETX=11
- M_DDOWN_LEFT_OFFSETY=464
- M_DDOWN_LEFT_WIDTH=283
- M_DDOWN_LEFT_HEIGHT=222
- M_DDOWN_LEFT_IMAGEFILE=
- M_DDOWN_USED=1
- M_DDOWN_OFFSETX=307
- M_DDOWN_OFFSETY=464
- M_DDOWN_WIDTH=283
- M_DDOWN_HEIGHT=222
- M_DDOWN_IMAGEFILE=
- M_DDOWN_RIGHT_USED=1
- M_DDOWN_RIGHT_OFFSETX=602
- M_DDOWN_RIGHT_OFFSETY=464
- M_DDOWN_RIGHT_WIDTH=283
- M_DDOWN_RIGHT_HEIGHT=222
- M_DDOWN_RIGHT_IMAGEFILE=
- M_DLEFT_USED=1
- M_DLEFT_OFFSETX=13
- M_DLEFT_OFFSETY=236
- M_DLEFT_WIDTH=208
- M_DLEFT_HEIGHT=220
- M_DLEFT_IMAGEFILE=
The values in the section „BUTTON_POSITIONS“ are the result data of the button editor. See section 18-05. It is not recommended to change these values in the file eccapp.ini manually.
Important for users
Operation
Please note all the information in this operating manual before starting operations (especially items 01 to 17)! The application will run only if the control box is connected. Messages will appear if the control box is not recognized at the USB port. You can make the application window larger or smaller but these changes will not be saved in eccapp.ini.
Activation
Eye focus must first capture Key1 and then Key2. The feedback is all buttons
green (=active mode). Eye focus must continue to be inside the ECCApp range,
otherwise, the buttons will time out after a defined interval (FOCUSTIMEOUT).
Activating wheelchair operating modes
modes To switch wheelchair operating mode, activate Key1 and Key2, then hold
focus on one of the four central switch buttons, which will then automatically
switch the switch relay in the set interval (SWITCHPAUSETIME) and set switch-
on time (SWITCHONTIME) until the focus leaves the button. Drive commands can
now be given directly without re-activating codes.
Driving
Following activation of Key1 and Key2, the chosen drive buttons can be focused
on to give drive commands. When the eye focus leaves the drive buttons, the
wheelchair stops.
Customizing the user interface (Version 3.0 and higher)
-
Adjust button size
Press „Shift“, „Control“ or „Alt“ on the keypad to switch to Edit mode (the exit button now shows the word „Edit“). Press the „Edit“ button and drag the edges to resize the buttons. To quit Edit mode, press Exit Edit. -
Change image content
Right-click to change button content (default: arrow, key, or number). -
Show/Hide: Show/hide button (hidden: no red frame in Edit mode)
-
New Image: Load a new PNG or JPG
-
Original Image: reloads original image (arrow/key/number) Images are stored in an automatically generated „Pictures“ subfolder with the program. All non-default data are stored in eccapp.ini, [BUTTON_POSITIONS] section. To reset everything to „original“, just delete this entire section from eccapp.ini
Safety features
- If for some reason (eye focus loss or other eye-camera problems, tablet PC overload) the eye-camera software does not transmit any new eye-mouse coordinates for a specific time (FOCUSTIMEOUT), all drive commands will stop.
- If for any reason (Windows problems, the program crashed, etc.) the application is not transmitting activation pulses to the relays in the control box, all relays will shut down immediately and the wheelchair will stop.
WARNING: You must check the MyEcc Control safety features
Troubleshooting
A “Joystick in center position” error message can have the following causes: If you operate the joystick before or immediately after switching on the control system, a “Joystick displaced” message appears in the display. This message can also appear if you give multiple commands simultaneously by moving your eyes rapidly in different directions. To reset the display, switch the control system off and back on again.
WARNING: Any claim under the warranty in accordance with our General Terms and Conditions of Business shall lapse if the customer independently remedies the defect, in particular by opening the device
Technical Data
Connections
- 1 × USB B
- 8 × 3,5-mm-jack
- 1 × 9-pin D-sub connector
Operating/storage environment
- Maximum ambient temperature during operation: 0°C (32°F) to 55°C (131°F)
- Maximum humidity during operation: 0% – 85% relative humidity
- Maximum ambient temperature during storage: -20°C (-4°F) to 75°C (167°F)
- Maximum humidity during storage: 5% – 85% relative humidity
- MyEcc Control is splash-proof (IP 22)
Power requirements
USB to 5VDC connector (100mA)
Contents of the pack
- MyEcc Control
- MyEcc software supplied as an executable app, no installation required, contained on a USB memory stick only
- Connecting cable; 3 × jack cables, 9-pin D-sub cable, USB cable
- HomeBrace bag
Weight/dimensions
MyEcc Control 0,325 kg 120 × 95 × 50 mm (Length × Width × Depth)
Service Life
With proper use and because of its components and design, MyEcc Control has a service life of 5 to 8 years.UDI-DI 4260737430006
MyEcc Control disposal
MyEcc Control consists of recyclable materials that do not belong in household waste and should be recycled.
That is why you as a consumer are obliged under the Electrical and Electronic Equipment Act („Gesetz über das
Inverkehrbringen, die Rücknahme und die umweltverträgliche Entsorgung von Elektro- und Elektronikgeräten“)
in force in Germany to ensure that electrical and electronic equipment is disposed of properly and
separately from household waste. Please help protect the climate and environment by returning the MyEcc
Control to your recycling center or an appropriate collection point.
QR SCAN
Contact
HomeBraceGermany UG (haftungsbeschränkt)
CEO: Thomas Ro
Postal and delivery addrsneress
Stuttgarter Str. 11
73614 Schorndorf
Headquarters and billing address
Birkenweg 12 | 73660 Urbach
+49 7181 20741-0
info@homebrace.com
Edition English 02/2022
© HomeBraceGermany UG (haftungsbeschränkt)
All rights reserved. No part of this manual may be reproduced in any form or
copied or processed by electronic, mechanical or chemical means without the
publisher‘s written consent.
This manual may contain printing flaws or errors. However, the information
provided in this manual is reviewed regularly and corrections will be
incorporated in the next edition. We accept no liability for any technical or
printing errors or the consequences thereof.
All trademarks and intellectual property rights are acknowledged.
Printed in Germany.
The information provided may change without notice to keep pace with technical
progress.
02/2022 // medienmaler.de
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>