KEMIDOSE pH+ORP Pump Seko Instruction Manual
- June 13, 2024
- KEMIDOSE
Table of Contents
- KEMIDOSE pH+ORP Pump Seko
- Product Information
- Product Usage Instructions
- PACKAGE CONTENT
- INSTALLATION INSTRUCTIONS
- ELECTRICAL CONNECTIONS
- TECHNICAL SPECIFICATIONS
- SETTING PROGRAM
- CALIBRATIONS
- DOSING METHOD
- INTERNAL WEB SERVER
- ALARMS
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
KEMIDOSE pH+ORP Pump Seko
Product Information
The KemiDose Double is a system that allows for precise measurement and injection of chemicals into a water system. It is designed to ensure accurate pH and ORP levels in water treatment processes. The package includes various components such as hoses, valves, probes, filters, and mounting brackets.
The KemiDose Double system is available in two configurations:
KemiDose pH – ORP: This configuration includes 2 pumps, 2 suction hoses, 2 delivery hoses, 2 ball valves, 4 probe-holders, 1 tapping saddle, 1 reducer, 1 foot filter, 1 mounting bracket kit, 5 filter minors, 1 probe holder with chlorine probe, 1 pH probe, 1 redox probe, 1 cleaning brush for chlorine probe, and 1 set of balls for chlorine probe.
KemiDose pH – ORP – CL: This configuration includes additional components for chlorine measurement and injection. It includes all the components of the KemiDose pH – ORP configuration, plus an additional probe holder with chlorine probe, pH 4 and pH 7 buffer solutions, a calibration solution, a filter wrench, and a PVC hose for probes holder.
Note: The WiFi model of the KemiDose Double may include an extra piece of certain components.
Product Usage Instructions
Warning
Before carrying out any work inside the control panel of the KemiDose device, ensure that it is disconnected from the power supply. Failure to comply with this instruction could result in injury or damage to the appliance and system.
Installation Instructions
Probe Positioning: For optimal probe reading, position it perpendicular to the tubing.
Use with Salt Chlorinator: When using the KemiDose system with a salt chlorinator, follow these instructions to prevent malfunctioning or damage:
- Position the pH measuring probe prior to the chlorinator cell.
- To eliminate eddy currents, connect the pool water to an electrical ground point.
- Position the product injection point after the chlorinator cell.
Electrical Connections
- Refer to the connection label for the corresponding input and output connections. Ensure proper connections are made to the power supply, pumps, probes, and other components as indicated in the label.
WARNING
- Before carrying out ANY work inside control panel of the KemiDose device, make sure you disconnect it from the power supply.
- Failure to comply with the instructions contained in this manual could cause injury to people and/or damage to the appliance and the system.
PACKAGE CONTENT
| Double pump
---|---
System
**Item***
| KemiDose pH – ORP| KemiDose pH – ORP – CL
A| 2| 2
B| 2| 2
C| 2| 2
D| 2| 2
E| 4(1)| 4(1)
F| 2| 2
G| 2| 2
H| 1| 1
I| –| 1
J| –| 1
K| 1| 1
L| 1| 1
M| –| 1
N| –| 1
O| 1| 1
P| 1| 1
Q| 1| 1
R| 1| 1
S| –| 1
T| –| 1
U| 1 (2)| 1 (2)
-
- The values from the table represent the number of items inside the package.
- (1 One piece more for the WiFi model only), (2 One piece for the WiFi model only)
WARNING: These products are DANGEROUS (IA) and require special precautions during use, handling and storage.
- NEVER mix chemical products.
- NEVER allow children or people who have not read this manual to use or tamper with KemiDose or any of its peripheral components (including chemical products).
pH chemical products:
- ABSOLUTELY not recommended => pure sulphuric acid
- Recommended for lowering pH => negative pH (with a sulphuric acid base)
- Recommended for raising pH => positive pH (sodium carbonate or bicarbonate)
ORP chemical products:
- ABSOLUTELY not recommended => all types of organic chlorine
- Liquid chlorine or 12% bleach can be used neat. If the product has a concentration of 48%, it is necessary to dilute it in water in a 1:3 ratio.
The pH / ORP probes are subject to wear and tear and therefore are not covered by the warranty.
INSTALLATION INSTRUCTIONS
Warning
- Make sure that the injection pressure is below 1.5 bar
Probe positioning
Installation plant
Warning: Use with salt chlorinator:
For the pH systems, to prevent the risk of system malfunctioning or damage, observe the following instructions:
- Position the pH measuring probe prior to the chlorinator cell.
- To eliminate eddy currents, connect the pool water to an electrical ground point
- Position the product injection point after the chlorinator cell.
ELECTRICAL CONNECTIONS
| Double pump system
---|---
Clamp| Description| KemiDose pH · ORP| KemiDose pH · ORP ·
CL
1| Input Probe| ORP| ORP
2| Input Probe| pH| pH
3| Input Temperature| TEMP (PT100)| TEMP (PT100)
4| Input Probe| Not used| Free Chlorine
5| Input Freq. signal| Flow Rate (Freq. Input)| Flow Rate (Freq. Input)
6| Level (product tank)| pH Level probe| pH Level probe
7| Level (product tank)| Chlorine (ORP) level probe| Chlorine level probe
8| Level (product tank)| Flow (REED sensor)| Flow (REED sensor)
9| Serial Port| Not present| Not present
10| Trigger Input| Circulation Pump (220Vac input)| Circulation Pump (220Vac
input)
11| Output Relay| RL1 AUX1 pH| RL1 AUX1 pH
12| Output Relay| RL2 AUX2 OPR/Chlorine| RL2 AUX2 OPR/Chlorine
13| Output Relay| RL3 Alarm| RL3 Alarm
14| Earth connector| Earth| Earth
15| Power Supply| 220-240 Vac 50-60 Hz (F/N)| 220-240 Vac 50-60 Hz (F/N)
C1| Pump connection| pH| pH
C2| Pump connection| Chlorine (ORP)| Chlorine
SekoNet| WiFi Module| WiFi card (dedicate code)| WiFi card (dedicate code)
Connection label
TECHNICAL SPECIFICATIONS
Specifications| KemiDose Double pH/ORP| KemiDose Double
pH/ORP/Chorine
---|---|---
Dimensions (H–W–D)| H :196 x W :205 x D :171 mm| H :196
x W :205 x D :171 mm
Weight| 6 Kg| 6 Kg
Pump state| Pause – Supply| Pause – Supply
Probe calibration| Automatic| Automatic
Power supply| 220-240 VAC 50-60 Hz| 220-240 VAC 50-60 Hz
Consumption (W)| 32 Watt| 32 Watt
Device precision| ± 0.1 pH; ±10mV; ±1°C| ± 0.1 pH; ±10mV; 0.1 ppm; ±1°C
Accuracy| ±0.02pH, ±3mV; ±0,5°C| ±0.02pH, ±3mV; 0.05 ppm; ±0.5°C
Range| 0-14pH; -99 -1000mV; 0…+55°C| 0-14pH; -99 -1000mV; 0-5 ppm;
0…+55°C
Flow rate pump (l/h)| 5 l/h| 5 l/h
Max. back-pressure| 5 bar| 5 bar
Relay contact (number 3)| 250 Vac 10A (resistive load)| 250
Vac 10A (resistive load)
Fuse| 500 mA (fast)| 500 mA (fast)
Frequency pump dosing| 160 strokes/minute| 160 strokes/minute
SETTING PROGRAM
- Button to increase the value
- Button to decrease the value
- Button Menu/Esc
- Button Cal/OK
- Button to set the setpoint
- Digital display
Program Setup
Press for 5 seconds
- At the entry of each menu item, the parameter can be directly modified using the arrow keys ( + and – ).
- Confirmation of the current setting and switching to the next item is done by pressing the button.
- The menu has a circular structure: once you arrive at last item, the confirmation of the parameter set by pressing , determines the return to the first menu item.
Note: Timeout setting menu, after 120 seconds without action the controller escape itself without saving parameters.
View Parameters
Calibration Menu
- Press (3 seconds) and calibrate probe pH, Chlorine, Temperature, OR
CALIBRATIONS
pH CALIBRATION
Note: If you have selected the “1 point cal.” function, the calibration will be made only in 1 point using the 7 pH buffer solution.
Reference calibration
ORP CALIBRATION
Reference calibration
CHLORINE CALIBRATION
- Press ESC to exit the calibration menu.
TEMPERATURE CALIBRATION
Set menu
- Press (3 seconds) and adjust set point value and press to confirm.
Calibration Menu
- Press (3 seconds) and calibrate probe pH, Chlorine, Temperature, ORP
System StandBy
- Press + – (5 seconds) the system sets in StandBy mode; all functions are disabled.
Reset OFA Timer
- Press (3 seconds) to reset OFA Alarm or press + – (5 seconds) to reset OFA Alarm.
Priming pumps
- Only while the pump is in “stand-by mode” press down + to reset flow totalizer, – to run pH pump, press to run ORP/Chlorine pump, press to run Aux1 relay, press to run Aux2 relay
To restore the default parameters, follow the steps below:
- Power off the KemiDose unit
- Keep + and – pressed and power on the unit
- The unit will flash
- Select the unit to reset – WiFi module or dosing system.
- Press +
- Press to restore the default parameters.
Default parameters
- Language = EN
- Set Point value = 7.5 pH; 700 mV; 1.2 ppm
- Dosing method = Acid (pH); Low (Redox); Low (Cl)
- OFA Time = OFF
- Calibration = Full
- Flow Input = OFF (recirculation pump)
- Dosing type = PROP; ON/OFF Relay Aux1 and Aux2
- IN Freq. = OFF
- Reed = NC (normally closed)
- P.ON (On delay) = OFF
- Flow delay = OFF
DOSING METHOD
-
Setpoint = 7.5 pH
-
Dosing mode = Acid
-
Prop. Band= 1.0 pH
-
Setpoint = 700 mV
-
Dosing mode = Low
-
Prop. Band = 250mV
-
Setpoint =1.2ppm Free Chlorine
-
Dosing mode = Low
-
Prop. Band = 0.8ppm
INTERNAL WEB SERVER
Download SekoLink
Register your account
Using the QrCode, login to the internal webpages
Set:
- User = ADMIN
- Password = 0000
Set your WiFi LAN name and password and confirm
- Complete your device registration
Thanks to your registration it is possible to use sekolink and sekoweb.
Thanks to sekolink it is possible to mange your pool:
- Monitoring and limited management
- Smartphone app compatible with iPhone or Android
- For end users
Use sekoweb address linkwww.sekoweb.com or APP to manage your pools with professional webportal:
- Monitoring and complete management
- Internet portal accessible via online login or by scanning a product’s QR code
- For pool and spa installers, technicians and engineers
ALARMS
(*1 Ranges Measure alarms)
n | Item | Limits |
---|---|---|
1 | Temp. Measure min | +10°C |
2 | Temp. Measure Max | +38°C |
3 | pH Measure min | 6 pH |
4 | pH Measure Max | 8 pH |
5 | ORP Measure min | +600 mV |
6 | ORP Measure Max | +800 mV |
7 | CL Measure min | 0,50 ppm |
8 | CL Measure Max | 2 ppm |
Connection label
Wire connection
Clamp | Description | KemiDose pH · ORP | Details |
---|---|---|---|
1 | Input Probe | ORP | |
2 | Input Probe | pH |
3
| ****
Input Probe
| ****
TEMP (PT100)
A= two wires sensor B= three wires sensor
|
4
| ****
Input Free Chlorine sensor
| ****
Input free chlorine probe: Pt: Platinum sensor Cu: Cupper sensor
| ****
5
| ****
Input Freq. signal
| ****
Flow Rate (Freq. Input) A= Mechanical Reed
B= Padwheel Hall sensor
|
6| Level (product tank)| pH Level probe| Level probe for
chemical tank
7| Level (product tank)| Chlorine (ORP) level probe| Level
probe for chemical tank
8| Level (product tank)| Flow (REED sensor)| Flow Sensor
9| Serial Port| Not present| None
10| Trigger Input| Circulation Pump (220Vac input)|
Line/Neutral wires
11| Output Relay| RL1 AUX1 pH| Dry contact
12| Output Relay| RL2 AUX2 OPR/Chlorine| Dry contact
13| Output Relay| RL3 Alarm| Dry contact
14| Earth connector| Earth| —
15| Power Supply| 220-240 Vac 50-60 Hz (F/N)| —