HOVERTECH HMSLING-39-B Repositioning Sheet User Manual
- June 13, 2024
- HOVERTECH
Table of Contents
- HOVERTECH HMSLING-39-B Repositioning Sheet
- Product Information
- Product Usage Instructions
- Attaching Sling Support Straps
- Symbol Reference
- HoverSling® Split-Leg
- HoverSling® Repositioning Sheet
- Part Identification– HT-Air® 1200 Air Supply
- HT-Air® 1200 Air Supply Keypad Function
- Cleaning
- Warranty Statement
- Returns and Repairs
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HOVERTECH HMSLING-39-B Repositioning Sheet
Product Information
- Product Name: HoverSling
- Manufacturer: HoverTech International
- Website: https://manual-hub.com/
- Model Number : [not specified]
- Serial Number: [not specified]
- Weight Limit:
- 700 lbs / 318 kg
- 1000 lbs / 454 kg
- Required Accessories:
- For Use as a Transfer Mattress: [not specified]
- For Use as a Sling: Any patient lift with 2, 3, or 4-point loop style hanger bars
Product Usage Instructions
- Ensure that the sling support straps are stowed in the strap pockets and connect the four snaps located at the center leg andfoot sections.
- Preferably, position the patient in a supine position.
- Using a log-rolling technique, place the HoverSling Split-Leg underneath the patient and loosely secure the patient safety strap.
- Plug the HoverTech Air Supply power cord into an electrical outlet.
- Insert the hose nozzle into one of the two hose entries at the foot end of the HoverSling Split-Leg and snap it into place.
- Ensure that the transfer surfaces are as close as possible and lock all wheels.
- If possible, perform the transfer from a higher surface to a lower surface.
- Turn on the HoverTech Air Supply.
- Push the inflated HoverSling Split-Leg at an angle, either headfirst or feetfirst. Once halfway across, the opposite caregiver should grasp the closest handles and pull to the desired location.
- Ensure that the patient is centered on the receiving equipment before deflating.
- Turn off the Air Supply and employ the bed/stretcher rails. Unbuckle the patient safety strap.
Note: When using the 50″ HoverSling Split-Leg, two air supplies may be used for inflation.
Attaching Sling Support Straps
- Attach the outside leg support straps to the hanger bar.
- Release the inside leg snaps. Cross the inside leg support straps by taking one leg support strap and passing it through the other on the opposite side. Attach the inside leg supports to the hanger bar.
- Attach the shoulder straps to the hanger bar for the proper safety and comfort of the patient. The color-coded strap loops provide easy identification to match strap placement on the left and right side of the patient. The shorter strap loops at the shoulder (blue/beige) create a more upright sitting position, while the longer shoulder and leg strap loops (white) increase recline angle and reduce hip flexion.
Symbol Reference
DECLARATION OF CONFORMITY TO MEDICAL DEVICE DIRECTIVE
Intended Use and Precautions
INTENDED USE
The HoverSling® Split-Leg and HoverSling® Repositioning Sheet are combination
air-assisted transfer mattresses and lift slings. When used as an air-assisted
transfer mattress, the HoverTech Air Supply inflates the HoverSling to cushion
and cradle the patient, while air simultaneously escapes from the holes on the
underside, reducing the force needed to move the patient by 80-90%. In this
capacity, the HoverSling can be used to assist caregivers with lateral patient
transfers, positioning, turning and proning. The HoverSling can also be used
for vertical patient lifts with a loop style hanger bar.
INDICATIONS
- • Patients unable to assist in their own lateral or vertical transfer.
• Patients whose weight or girth poses a potential health risk for the caregivers responsible for repositioning, vertically transferring or laterally transferring said patients.
CONTRAINDICATIONS
Patients who are experiencing thoracic, cervical or lumbar fractures that are deemed unstable, unless using in conjunction with a spinal board on top of the HoverSling (follow your state’s protocol regarding use of spinal boards) for lateral transfers.
INTENDED CARE SETTINGS
Hospitals, long term or extended care facilities
PRECAUTIONS – FOR USE AS A TRANSFER MATTRESS
- Caregivers must verify that all brakes have been engaged prior to transfer.
- For safety, always use two people during patient transfers.
- Additional caregivers are recommended when moving a patient over 750 lbs./340kg.
- Never leave patient unattended on an inflated HoverSling.
- Only use the HoverSling for its intended purpose as described in this manual.
- Only use attachments and/or accessories that are authorized by HoverTech International.
- When transferring to a low air loss bed, set the bed mattress air flow to the highest level for a firm transfer surface.
- Never attempt to move a patient laterally on an uninflated HoverSling.
- Ensure patient is centered on HoverSling before inflating.
PRECAUTIONS – FOR USE AS A SLING
- Always use a minimum of two caregivers when operating the HoverSling.
- The HoverSling should only be used by trained personnel.
- Only use attachments and/or accessories that are authorized by HoverTech International.
- Before use, a risk assessment must be carried out to ensure that the correct size HoverSling is being used for the patient.
- A clinical assessment should be carried out by a qualified nurse and therapist before lifting patients with the HoverSling.
- Do not use the HoverSling lateral transfer handles to lift. Excessive force on the lateral transfer handles can cause the HoverSling to tear.
- Make sure the HoverSling is not attached to any object other than the loop style hanger bar.
- When connecting to loop style hanger bar, ensure shoulder strap loop colors match on the patient’s left and right side and leg strap loop colors match on the patient’s left and right side.
- Once the sling support straps are fully extended, make sure they are properly connected to the hanger bar before the patient is lifted.
- For seated lifts, make sure the patient is sitting securely in the HoverSling® Split-Leg before transferring or lifting.
- Never lift a patient higher than is necessary to complete the lift/transfer.
- Never leave the patient unattended while using the HoverSling.
- Use the HoverSling according to both the lift and HoverSling instructions.
- If the weight limit differs among the patient lift, hanger bar, and HoverSling, then the lowest weight limit applies.
WARNING: If using the HoverSling® Repositioning Sheet for bed to chair
transfer, the chair MUST be reclined. If the chair does not recline, then
using the HoverSling Repositioning Sheet is not recommended.
WARNING: If there is any indication of damage, remove the HoverSling from
service and discard.
WARNING: In the OR: To prevent the patient from slipping, always deflate
the HoverSling and secure the patient and HoverSling to the OR table prior to
moving the table into an angled position.
PRECAUTIONS – AIR SUPPLY
- Route the power cord in a manner to ensure freedom from hazard. Avoid blocking the air intakes of the air supply.
- When using the HoverSling in the MRI environment, a 25 ft. specialty MRI hose is required (available for purchase).
- CAUTION: Avoid electric shock. Do not open air supply.
- Reference product specific user manuals for operating instructions.
HoverSling® Split-Leg
Part Identification – HoverSling® Split-Leg
HoverSling® Split-Leg Product Specifications/Required Accessories
- Material:
- Top: Polyester knit
- Bottom: Nylon Twill
- Construction: Sewn
- Dimensions:
- Width:
- 34″ (86 cm) inflated
- 39″ (99 cm) inflated
- 50″ (127 cm) inflated
- Length: 75″ (190 cm) inflated
Model #: HMSLING-34 – 34″ W x 75″ L (10 per box)
Model #: HMSLING-34-B – 34″ W x 75″ L (10 per box)
WEIGHT LIMIT 700 LBS/ 318 KG
Model #: HMSLING-39 – 39″ W x 75″ L (5 per box)
Model #: HMSLING-39-B – 39″ W x 75″ L (5 per box)
Model #: HMSLING-50 – 50″ W x 75″ L (5 per box)
Model #: HMSLING-50-B – 50″ W x 75″ L (5 per box)
WEIGHT LIMIT 1000 LBS/ 454 KG
Breathable model
REQUIRED ACCESSORY FOR USE
AS A TRANSFER MATTRESS:
Model #: HTAIR1200 (North American Version) – 120V~, 60Hz, 10A
Model #: HTAIR2300 (European Version) – 230V~, 50 Hz, 6A
Model #: HTAIR1000 (Japanese Version) – 100V~, 50/60 Hz, 12.5A
Model #: HTAIR2356 (Korean Version) – 230V~, 50/60 Hz, 6A
Model #: AIR200G (800 W) – 120V~, 60Hz, 10A
Model #: AIR400G (1100 W) – 120V~, 60Hz, 10A
REQUIRED ACCESSORY FOR USE AS A SLING:
Any patient lift with 2, 3, or 4-point loop style hanger bars intended for use with:
- Mobile hoists with sling seats
- Hoist trolleys
- Stationary hoists fixed to the wall/walls, floor and/or ceiling
- Stationary free-standing hoists
HoverSling® Split-Leg Instructions for Use as a Transfer Mattress .
- Ensure sling support straps are stowed in strap pockets and the four (4) snaps located at the center leg and foot sections are connected.
- Patient should preferably be in a supine position.
- Place HoverSling Split-Leg underneath patient using a log-rolling technique and secure patient safety strap loosely.
- Plug the HoverTech Air Supply power cord into an electrical outlet.
- Insert the hose nozzle into one of two hose entries at foot end of HoverSling Split-Leg and snap into place.
- Ensure that transfer surfaces are as close as possible and lock all wheels.
- If possible, transfer from a higher surface to a lower surface.
- Turn on HoverTech Air Supply.
- Push inflated HoverSling Split-Leg at an angle, either headfirst or feet- first. Once half-way across, opposite caregiver should grasp closest handles and pull to desired location.
- Ensure patient is centered on receiving equipment prior to deflation.
- Turn off Air Supply and employ the bed/stretcher rails. Unbuckle patient safety strap.
NOTE: When using the 50” HoverSling Split-Leg, two air supplies may be used for inflation.
HoverSling® Split-Leg Instructions for Use as a Sling
INSTRUCTIONS FOR USE – PATIENT IN BED
Note: HoverSling Split-Leg has six (6) sling support straps that need to
be attached.
- Option 1 – Place the HoverSling Split-Leg underneath patient using log-rolling technique with the apex of the sling’s horseshoe at the coccyx of the patient. Elevate the head of the bed. Option 2 – Elevate the head of the bed. Lean patient forward and slide sling behind patient’s back until the apex of the horseshoe reaches the coccyx of the patient. Have patient lean back to hold sling in place.
- Release inside leg snaps. Cross inside leg support straps by taking one leg support strap and passing it through the other on the opposite side. Attach inside leg sup- ports to hanger bar.
- Attach outside leg support straps to hanger bar.
- Attach shoulder straps to the hanger bar for the proper safety and comfort of the patient. [Color coded strap loops provide easy identification to match strap placement on the left and right side of the patient. Shorter strap loops at shoulder (blue/beige) create a more upright sitting posi- tion, while longer shoulder and leg strap loops (white) increase recline angle and reduce hip flexion].
INSTRUCTIONS FOR USE – BACK TO BED
- Position patient above center of bed. Raise head of bed and lower patient on to bed.
- Detach strap loops from hanger bar.
- Place the sling support straps back into the strap pockets on the inside and outside edges of the HoverSling Split-Leg.
INSTRUCTIONS FOR USE – PATIENT IN CHAIR
-
Release leg snaps. Lean patient forward.
Place the apex of the sling’s horseshoe behind the patient until you touch the seat ensuring the sling is centered on the patient’s torso. Have patient lean back against the sling to hold it in place. -
Place the leg sections along the side of the hips and legs, then place each leg section underneath each respective leg.
-
Cross inside leg support straps by taking one leg support strap and passing it through the other on the opposite side. Attach inside and outside leg support straps to hanger bar.
-
Attach shoulder straps to the hanger bar for the proper safety and comfort of the patient. [Color coded strap loops provide easy identification to match strap place- ment on the left and right side of the patient. Shorter strap loops at shoulder
(blue/ beige) create a more upright sitting position, while longer shoulder and leg strap loops (white) increase recline angle and reduce hip flexion.]
HoverSling® Repositioning Sheet
Part Identification – HoverSling® Repositioning Sheet
Product Specifications/Required Accessories
- Material:
- Top: Polyester knit
- Bottom: Nylon Twill
- Construction : Sewn
- Width:
- 39″ (99 cm) inflated
- 50″ (127 cm) inflated
- Length: 75″ (190 cm) inflated
Model #: HMSLING-39RS-B – 39″ W x 75″ L (5 per box)
Model #: HMSLING-50RS-B – 50″ W x 75″ L (5 per box)
WEIGHT LIMIT 1000 LBS/ 454 KG
REQUIRED ACCESSORY FOR USE
AS A TRANSFER MATTRESS:
Model #: HTAIR1200 (North American Version) – 120V~, 60Hz, 10A Model #:
HTAIR2300 (European Version) – 230V~, 50 Hz, 6A
Model #: HTAIR1000 (Japanese Version) – 100V~, 50/60 Hz, 12.5A Model #:
HTAIR2356 (Korean Version) – 230V~, 50/60 Hz, 6A
Model #: AIR200G (800 W) – 120V~, 60Hz, 10A
Model #: AIR400G (1100 W) – 120V~, 60Hz, 10A
REQUIRED ACCESSORY FOR USE AS A SLING:
Any patient lift with 2, 3, or 4-point loop style hanger bars intended for use with:
- Mobile hoists
- Hoist trolleys
- Stationary hoists fixed to the wall/walls, floor and/or ceiling
- Stationary free-standing hoists
HoverSling® Repositioning Sheet Instructions for Use as a Transfer
Mattress
- Ensure sling support straps are stowed in strap pockets.
- Patient should preferably be in a supine position.
- Place HoverSling Repositioning Sheet underneath patient using a log-rolling technique and secure patient safety strap loosely.
- Plug the HoverTech Air Supply power cord into an electrical outlet.
- Insert the hose nozzle into one of two hose entries at foot end of HoverSling Repositioning Sheet and snap into place.
- Ensure that transfer surfaces are as close as possible and lock all wheels.
- If possible, transfer from a higher surface to a lower surface.
- Turn on HoverTech Air Supply.
- Push inflated HoverSling Repositioning Sheet at an angle, either head- first or feetfirst. Once half-way across, opposite caregiver should grasp closest handles and pull to desired location.
- Ensure patient is centered on receiving equipment prior to deflation.
- Turn off Air Supply and employ the bed/stretcher rails. Unbuckle patient safety strap.
NOTE: When using the 50″ HoverSling® Repositioning Sheet, two air supplies may be used for inflation.
HoverSling® Repositioning Sheet Instructions for Use as a Sling
INSTRUCTIONS FOR USE – PATIENT IN BED
Note: HoverSling® Repositioning Sheet has eight (8) sling support straps
that need to be attached.
- Place the HoverSling Repositioning Sheet underneath patient using log-rolling technique.
- Attach all straps to the hanger bar for the proper safety and comfort of the patient.[Color coded strap loops provide easy identification to match strap placement on the left and right side of the patient.] Lift patient according to lift manufacturer’s instructions.
INSTRUCTIONS FOR USE – BACK TO BED
- Position patient above center of bed. Lower patient on the bed.
- Detach strap loops from hanger bar.
- Place the sling support straps back into the strap pockets on the inside and outside edges of the HoverSling Repositioning Sheet.
INSTRUCTIONS FOR USE – LATERAL TURN
- With patient centered on the HoverSling® Repositioning Sheet, lower hanger bar to comfortable working height.
- Raise bed rails and attach all four (4) straps on opposite side of patient turn to loops on same side of hanger bar starting at foot end.
- As lift is raised, patient will turn towards the opposite side of the bed the straps are attached to. Use wedges to keep patient in position if desired. After the task is completed, lower hanger bar and remove sling straps.
INSTRUCTIONS FOR USE – HYGIENE TURN
- With patient centered on the the HoverSling Repositioning Sheet, lower hanger bar to comfortable working height.
- Raise bed rails and attach the sling strap that is closest to patient’s shoulder to hanger bar.
- As lift is raised, patient will begin to turn to the side opposite the connected strap. Leave sling in place to perform task. Once task is completed, lower hanger bar and remove sling strap.
Part Identification– HT-Air® 1200 Air Supply
WARNING: The HT-Air is not compatible with DC power supplies.
The HT-Air is not for use with the HoverJack Battery Cart.
HT-Air® 1200 Air Supply Keypad Function
-
ADJUSTABLE: For use with HoverTech air-assisted positioning devices. There are four different settings. Each press of the button increases the air pressure and rate of inflation. The Green Flashing LED will indicate the inflation speed by the number of flashes (i.e. two flashes equals the second inflation speed).
All of the settings in the ADJUSTABLE range are substantially lower than the HoverMatt and HoverJack settings. The ADJUSTABLE function is not to be used for transferring.
The ADJUSTABLE setting is a safety feature that can be used to ensure the patient is centered on HoverTech air-assisted devices and to gradually accustom a patient who is timid or in pain to both the noise and functionality of the inflated devices. -
STANDBY: Used to stop inflation/air flow (Amber LED indicates STANDBY mode).
-
HOVERMATT 28/34: For use with 28″ & 34″ HoverMatts and HoverSlings.
-
HOVERMATT 39/50 & HOVERJACK: For use with 39″ & 50″ HoverMatts and HoverSlings and 32″ & 39″ HoverJacks.
Air200G/Air400G Air Supplies
If using HoverTech’s Air200G or Air400G Air Supplies, press the grey button on
the top of the canister to initiate air flow. Press the button again to stop
air flow.
Cleaning
A HoverTech disposable sheet and/or absorbent chux can be placed on top of the HoverSling to keep it clean. The HoverSling is a single-patient use product and is not intended to be laundered. Discard if the device becomes grossly soiled.
PREVENTIVE MAINTENANCE
The HoverSling should be checked before each use to ensure there is no damage
that would cause it not to function as intended. It should have all safety
straps, sling support straps and handles intact. The Do Not Launder label
should be intact. There should be no tears, holes or loose threads. If such
damage is found, the HoverSling should be discarded. The HoverSling is
intended to be used for the duration of the patient’s stay at the
healthcare/nursing facility. If the stay exceeds three months, the HoverSling
should
be replaced.
INFECTION CONTROL
If the HoverSling is used on an isolation patient, the hospital should employ the same protocols/procedures it utilizes to discard other contaminated single patient use devices.
AIR SUPPLY CLEANING AND MAINTENANCE
See air supply manual for reference.
NOTE: CHECK YOUR LOCAL/STATE/FEDERAL/INTERNATIONAL GUIDELINES BEFORE
DISPOSAL.
Warranty Statement
The HoverSling is warranted to be free from defects in materials and work-
manship. In the unlikely event that a problem arises as a result of a defect
in materials or workmanship, we will promptly replace the HoverSling within
ninety (90) days of purchase or in-service.
This warranty is not an unconditional guarantee for the life of the product.
Our warranty does not cover product damage that may result from use contrary
to Manufacturer’s instructions or specifications, misuse, abuse, tampering, or
damage due to mishandling. Warranty specifically does not cover product damage
that may result from using an air supply that produces more than 3.5 psi to
inflate the HoverSling.
Equipment that has been neglected, improperly maintained, repaired or altered
by someone other than an authorized representative of Manufacturer, or
operated in anyway contrary to the operating instructions, shall void this
warranty.
This warranty does not cover normal “wear and tear”. Component parts,
particularly any optional equipment, valve caps, their attachments and cords,
will show wear with use over time and eventually may need to be refurbished or
replaced. This normal type of wear is not covered by our warranty, but we will
provide prompt, high quality repair service and parts at a nominal cost.
HoverTech International’s liability under this warranty and on any claim of
any kind for any loss or damage arising out of, connected with, or resulting
from the design, manufacture, sale, delivery, installation, repair or
operation of its products, whether in contract or tort, including negligence,
shall not exceed the purchase price paid for the product and upon expiration
of the applicable warranty period, all such liability terminates. The remedies
which this warranty provides are exclusive and HoverTech International shall
not be liable for any incidental or consequential damages.
There are no warranties, expressed or implied, which extend beyond this
warranty statement. The provisions of these warranty clauses are in lieu of
all other warranties, expressed or implied, and of all other obligations or
liabilities on HoverTech International’s part and they neither assume nor
authorize any other person to assume for HoverTech International any other
liability in connection with Manufacturer sale or lease of said products.
HoverTech International makes no warranty of merchantability or fitness for a
particular purpose. There is no warranty that the goods will be fit for a
particular purpose. By accepting the goods, the buyer acknowledges that buyer
has determined the goods are suitable for the buyer’s purposes.
MANUFACTURER’S SPECIFICATIONS ARE SUBJECT TO CHANGE.
Returns and Repairs
All products being returned to HoverTech International (HTI) must have a
Returned Goods Authorization (RGA) number issued by the company. Please call
800-471-2776 and ask for a member of
the RGA Team who will issue you an RGA number. Any product returned without an
RGA number will cause a delay in the repair time.
HTI’s warranty covers manufacturer’s defects in material and workmanship. If a
repair is not covered under warranty, a minimum repair fee of $100 per item
will be assessed, plus return shipping. A Purchase Order for the repair charge
should be provided by the facility at the time the RGA number is issued,
should the repair not be covered under warranty. Lead-time for repairs is
approximately 1-2 weeks, not including shipping time.
Returned products should be sent to:
HoverTech International
Attn: RGA # ___
4482 Innovation Way
Allentown, PA 18109
4482 Innovation Way Allentown, PA 18109
800.471.2776
Fax 610.694.9601
www.HoverMatt.com
Info@HoverMatt.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>