BROMIC MED0140GD-NR Medical And Vaccine Fridges
Product Information
The Instruction Manual is for the Medical and Vaccine Fridges model MED0140GD- NR, manufactured by Bromic Refrigeration. The manual contains important information about the installation, operation, and maintenance of the fridge. It also includes important safety information and warnings.
Here are some key details about the product:
- Part Number: MED0140GD-NR
- External Dimension: Height – mm, Width – mm, Depth – mm
- Capacity: Litres
- Weight: Kg
- Temperature Range: °C
- Ambient Temperature Range: °C
- Climate Class: (not specified)
- Energy Consumption: kWh/24h
- Refrigerant: (not specified)
- Air Circulation: (not specified)
- Current: (not specified)
- Electrical Voltage: (not specified)
- Frequency: Hz
- Plug Location: (not specified)
- Color: (not specified)
- Internal Light: Yes/No
- Number of Doors: (not specified)
- Door Swing Dimensions: (not specified)
- Total Depth of Door Open: mm
- Number of Shelves base per Door: (not specified)
- Shelf Dimensions: (not specified)
- Temp, Door, Audible & Visual Alarm: Yes/No
Product Usage Instructions
Installation
To install the Medical and Vaccine Fridges:
- Locate the cabinet in a suitable location.
- Ensure proper ventilation by leaving at least 100mm of space around the fridge on all sides.
- Connect the power supply to the fridge, following local electrical codes.
Operation
To operate the fridge:
- Before operating, make sure the fridge is properly installed and connected to the power supply.
- Press the power button to start up the fridge.
- To set the temperature, press the temperature button. The current set temperature will be displayed. Use the up and down buttons to adjust the temperature. Press the temperature button again to exit. If no further buttons are pressed within 6 seconds, the cabinet temperature will be displayed.
- To display the product sensor temperature, press the temperature button. The product temperature will flash and display. After 6 seconds, the cabinet temperature will be displayed again.
- To turn the internal LED lights on or off, press the light button.
- It is important to record the maximum and minimum temperatures of the fridge twice a day. Recordings should be taken before the first door opening of the day.
Maintenance
For maintenance of the fridge, refer to the maintenance section of the
Instruction Manual.
Troubleshooting
If you encounter any issues with the fridge, refer to the troubleshooting
section of the Instruction Manual. If problems persist, contact the
manufacturer or an approved service agent.
Note:
Bromic Refrigeration reserves the right to make changes to specifications,
parts, components and equipment without prior notification. This Installation,
operation and service manual may not be reproduced in any form with prior
written consent from Bromic Refrigeration.
PREFIX
This manual contains important information about the installation, operation, and maintenance of Bromic MediFridgeTM. Please pay close attention to the important safety information shown throughout this instruction manual. Any safety information will be accompanied by the following safety alert symbols:
- DANGER,
- WARNING,
- IMPORTANT
- READ THIS MANUAL CAREFULLY before installing or servicing this product.
- Improper installation, operation, or maintenance can result in death, severe injury, or property damage.
- This appliance is intended for 230-240 Volt power supply.
- Installation MUST be carried out by a licensed and authorized technician in accordance with local electrical codes.
IMPORTANT NOTES AND WARNINGS
- Read all instructions before installing or using this appliance
- Retain this manual for future reference
- Use this appliance only as described in this manual. Any other use not recommended by the manufacturer may cause fire, electric shock or injury to persons.
- Improper installation, adjustment or alteration and failure to follow the warnings and instructions in this manual could result in severe personal injury, death or property damage
- The manufacturer is not responsible for any damage that could happen from improper use. The manufacturer emphasizes that this appliance should be used in a responsible manner and that all procedures, warnings, and safety instructions contained in this booklet be followed strictly.
- Check the appliance for damage. Do not operate a damaged unit.
- Check for damage to the appliance regularly. If damage to the appliance is suspected, discontinue use immediately and contact the supplier or qualified person to repair it.
- This appliance is not intended for use by persons (including children) with reduced physical, sensory, or mental capabilities, or lack of experience and knowledge unless they have been given supervision or instruction concerning use of the appliance by a person responsible of their safety.
- Keep packaging material out reach of children
- Children should be carefully supervised to ensure that they do not play with the appliance
- Do not attempt to alter the appliance in any manner.
- Please pay attention to the maximum load limit when filling the shelves.
- All work on electrical equipment must be carried out by a qualified and licensed technician only.
- Do not perform maintenance until the appliance has been turned off and power disconnected.
- Never detach any cover except where specifically instructed in this manual. Doing so might expose live electrical parts.
- Do not operate the appliance with panels, covers or guards removed.
- Keep the air intake and outlets free of obstructions.
- Do not use the appliance outdoors.
- If the supply cord becomes damaged, it must be replaced by an authorized service agent or similarly qualified person in order to avoid a hazard.
- If the appliance has not been used, or will not be used for a long period of time, disconnect the power supply.
- Ensure the appliance has adequate ventilation.
- The top of the cabinet should not be used for the storage of any objects and must be clean and dry at all times.
PRODUCT OVERVIEW
MED0140GD-NR
- Part Number 3736257
- External Dimension
- Height, mm 820
- Width, mm 595
- Depth, mm 595
- Capacity Litres 225
- Weight Kg 40
- Temperature
- Range +2°C to +8°C
- Type Digital
- Ambient Temperature Type 32°C
- Climate Class Rating 4M2
- Energy Consumption kWh/24h 1.396kWh/24h
- Refrigerant
- Refrigerant R600a
- Charge, g 100W
- Air Circulation Fan Forced
- Current 1.2A
- Electrical Voltage 220-240V
- Frequency Hz 50
- Plug Location Rear, bottom right
- Colour Default White
- Internal Light Type LED
- Number of Doors Quantity 1
- Door
- Type Glass
- Glass Type Double glazed Low-E Glass + Argon Gas
- Style Hinged
- Lock Yes
- Self Closing Yes
- Shape Flat
- Door Swing Dimensions mm 590
- Total Depth of Door Open mm 1990
- Number of Shelves base per Door Quantity 4
- Shelf Dimensions mm 500(w) x 355(d) mm
- Temp, Door, Audible & Visual Alarm Y/N Yes
INSTALLATION
NOTE:
The maximum ambient temperature for the refrigerator is 30°C. Conditions
outside this temperature will result in cabinet temperatures outside the
required 2 to 8°C range. Bromic recommends that the refrigerator is installed
at an ambient temperature of 20 to 25°C.
Locating the Cabinet
Ensure the following instructions are observed as this is vital for proper and
trouble-free operation.
- Do not install the cabinet outdoors.
- Locate the appliance in a dry and well ventilated place.
- The unit must be positioned on a level surface.
- Do not install in the vicinity of heating radiators or other heat sources.
- Do not install the appliance where it is exposed to direct sunlight.
- Do not install the appliance where it is exposed to warm draughts or open doors.
- Keep the air intake and outlets free of obstructions.
- Adequate allowance should be made for door openings.
Ventilation
Ensure the following instructions are observed as this is vital for proper and
trouble-free operation.
- Allow 100mm clearance between walls and the appliance.
- Ensure that the condenser fan is free to take in ambient air from the room.
- Adherence to minimum clearances is necessary
Power Supply
- The appliance must not be plugged into a multiple outlet power board.
- The appliance is supplied with a flexible power cord fitted with a 3-pin plug.
- The socket used must be properly earthed and fuse protected.
- See the rating label inside the cabinet for supply voltage, frequency and current draw.
- Power cable is rated 70°C. It is recommended that the cable should not come into contact with parts over 50°C.
- Do not turn off power before consulting the person responsible.
- Mark the power source clearly, so the refrigerator is not unplugged or turned off accidentally.
Before Operating
- If the appliance has been sharply tilted or transported in the horizontal position, it must be left to stand in the upright position and not plugged in for 1 hour.
- Remove all the packaging material before using the appliance.
- Start the appliance only if it has been installed as described.
OPERATION
Start-up
- Connect the appliance to the mains power supply and check that the compressor, electronic controller, lighting and all fans are functioning correctly.
- The compressor and condenser fan should operate should operate 3 minutes after the unit has been connected to the power supply.
- The evaporator fan should operate after the unit has been plugged in.
- The electronic controller will display the current cabinet temperature after the unit has been connected to the power supply.
- The LED indicates that refrigeration circuit is operating.
- During the start-up delay (3 minutes), the LED is flashing
- The LED indicates that the refrigeration circuit is defrosting
Setting the temperature
NOTE:
The controller is factory set at 4°C. The controller is programmed to control
cabinet temperatures between 2 to 8°C.
- Press the button.
- The set temperature will be displayed.
- Press or to adjust the temperature.
- Press the button to exit. If no further buttons are pressed within 6 seconds, the cabinet temperature will be displayed.
Display the Product Sensor Temperature
- Press the button.
- The product temperature will flash and display.
- After 6 seconds, the cabinet temperature will be displayed.
Operating the Light
- Press the button to turn internal LED lights ON or OFF
MAX/MIN Temperature Recording
IMPORTANT:
Record the maximum and minimum temperatures two (2) times a day. Recordings
should be taken before the first door opening of the day.
- Press the button once
- The highest temperature recording will be displayed.
- Press the button again while the highest temperature recording is displayed.
- The lowest temperature recording will be displayed.
- To clear the MAX/MIN recording – Hold the button for 6 seconds
- The display will flash for 3 seconds and the buzzer will sound – this will clear the previous highest and lowest temperature recording.
- The current cabinet temperature will be recorded.
Controller Alarms
If alarms persist, contact the manufacturer or an approved service agent.
Code | Alarm Type | Corrective Action |
---|---|---|
HI | Cabinet high temperature alarm | • Check if there is adequate |
ventilation
• Check if there is free circulation of air inside the cabinet
• Check and clean condenser coil and fan
• Check if door is shut
LO| Cabinet low temperature alarm
1H| Cabinet temperature sensor short circuit| • Unplug the unit for 1
minute then reconnect to power
1L| Cabinet temperature sensor open circuit
2H| Product temperature sensor short circuit
2L| Product temperature open circuit
H2| Product high temperature alarm| • Check if there is adequate
ventilation
• Check if there is free circulation of air inside the cabinet
• Check and clean condenser coil and fan
• Check if door is shut
dR| Door open alarm| • Close the door
PRODUCT LOADING AND STORAGE
- WARNING: Failure to follow the guidelines below will result in cabinet temperatures above 8°C
- IMPORTANT: Run the refrigerator for at least 2 hours before loading.
Start-up
- Clean the interior surface with a soft cloth and warm water before loading.
- Ensure there is free circulation of air between the products.
- Do not block the evaporator fan.
- Do not overstock the refrigerator.
- Use cold mass (i.e. water bottle) on unused shelves.
- Keep refrigerator door openings to a minimum.
- Ensure products are between 2 and 8°C before loading – if there is a breach in the cold-chain, isolate the affected products. Do not use or discard until advise is given.
- Some products are sensitive to ultraviolet (UV) light, please see storage guidelines from the manufacturer before removing from packaging.
- Do not store food and other goods in the refrigerator.
- Maximum recommended shelf load is 20kg.
NOTE: Bromic cannot be responsible for loss of stock or damage to stock.
Please refer to the ‘National Vaccine Storage Guidelines: Strive for 5’ for more information on vaccine storage and handling or contact your local state or territory health department. (http://www.health.gov.au/internet/immunise/publishing.nsf/content/IMM77-cnt).
MAINTENANCE
- WARNING: Disconnect the appliance from the power supply before performing any maintenance. Never unplug the appliance by pulling on the cord. Always grasp the plug firmly and pull straight out of the outlet.
- WARNING: Exposure to significant water ingress will void warranty and cause potential safety issues. If the unit should be exposed to water, please contact the manufacturer immediately.
- IMPORTANT: Warranty may be void if there is no evidence of preventative maintenance is performed on the appliance.
Cleaning the Cabinet
- Remove the products and transfer to a suitable location prior to cleaning the inside of the cabinet.
- Do not use any aggressive or abrasive cleaning agents.
- Never use a water hose or a high-pressure jet to clean.
- Do not use sharp or pointed objects for cleaning.
- Clean the inside and outside of the appliance with a clean damp cloth or with a neutral cleanser.
- Dry all surfaces with a dry cloth.
Condenser Coil
IMPORTANT:
The condenser coil must be cleaned with a vacuum or brush once a month. The
condenser coil must be cleaned regularly to maintain the performance and
minimise power consumption. A poorly maintained condenser coil can lead to
failure of the unit and cabinet temperatures above 8°C.
Extended Defrosting
The evaporator (cooling) coil of the refrigerator may build up a layer of ice
over time. This will reduce the performance of the unit – inspect the
evaporator coil for ice build-up regularly. For extended defrosting, turn off
the unit with the door opened for a period of 24 hours.
TROUBLESHOOTING
IMPORTANT:
Repairs on the unit must be carried out by an authorised and licensed person
only. Contact the manufacturer or an approved service agent for any repairs
needed.
Symptom| Possible Cause(s)|
---|---|---
Unit does not work| A. No power supply| A. Check power supply
Cabinet Temperature Too High| A. Ambient temperature too high
B. Inadequate ventilation
C. Wall clearances below minimum distance
D. Dirty condenser coil
E. Ice build-up on evaporator coil
| A. Relocate unit
B. Keep ventilation slots free
C. Adjust to minimum clearance
D. Clean
E. Perform extended defrosting
Cabinet Temperature Too Low| A. Set temperature too low| A. Adjust set
temperature
Cabinet lights not working| A. Light not switched on
B. Light is faulty
| A. Switch on
B. Replace light
Drip tray overflowing| A. Extreme operating condition| A. Relocate unit
Odour inside cabinet| A. Cabinet dirty| A. Clean
SCHEMATIC DIAGRAM
ENVIRONMENTAL AND DISPOSAL NOTES
These refrigerated cabinets are designed to high standards of environmental protection and recycling convenience. Cyclopentane, an environmentally friendly foaming agent, has been used. Cyclopentane is distinguished by zero ozone depletion potential (ODP) and zero global warming potential (GWP). The refrigerant used and contained within a hermetically tight circuit is environmentally friendly and does not contain chlorine and zero ODP. The device must be disposed of in accordance with applicable regulations. The WEEE regulations must be observed.
In the interest of preventing environmental nuisance and promoting re-use of the high quality materials employed, used commercial refrigeration equipment must be collected for expert disposal and recycling. Therefore, observe the following disposal regulations:
- The cabinet must not be discarded with regular waste.
- The refrigeration circuit must not be damaged.
- Before disposal, make it unserviceable by disconnecting the power plug and cutting off the power cable.
- Consult your local authorities or authorised dealers for details of collection and disposal facilities.
- Children are often unable to recognise the dangers associated with commercial refrigeration equipment. Proper supervision must be maintained and children must be prevented from playing with discarded refrigerated cabinets.
WARRANTY
- All Bromic products are manufactured to the highest quality standard and are supported by our Extra Care Service Warranty, including guarantees against any material or workmanship defects; guarantees that cannot be excluded under Australian Consumer Law.
- Warranty Terms vary across the range. Please ensure you are familiar with the Terms and Conditions that apply to the purchase of your Bromic product by referring to the Warranty Policy included with your purchased product.
- Alternatively, you can also download the Warranty Policy at https://bromicrefrigeration.com.au/extra-care-warranty.
HOW TO PLACE A SERVICE REQUEST
- Please call 1300 276 642 to place a warranty service request and let the operator know whether you need Priority A or B service. More information can be found below.
- Ensure you have the serial number of your product ready before contacting us.
- The serial number of your product can be found on a compliance plate located on the inside right and at the rear of your cabinet.
SERVICE REQUEST PRIORITY
-
PRIORITY A SERVICE
When the unit is NOT performing its primary function such as cooling or heating, or presents an OH&S or food safety risk -
PRIORITY B SERVICE
When the unit is performing its primary function, and presents no OH&S or food safety risk. i.e. lights, gaskets. locks, door adjustment -
SERVICE CALL PLACED BEFORE MIDDAY
- Priority A – Technician to attend site within 4 hours
- Priority B – Technician to attend site within 8 business hours
-
SERVICE CALL PLACED AFTER MIDDAY
- Priority A – Technician attends site before midday on the following business day
- Priority B – Technician attends site before 4pm on the following business day
BROMIC REFRIGERATION
- 10 Phiney Place, Ingleburn (Sydney) NSW 2565 Australia
- T. 1300 276 642
- For further product information bromicrefrigeration.com.au.