ThetaStation
User Guide Rev C Theta
Station
Chapter 1 Warning Symbol Definition
Below is a list of warning symbols you may encounter in this manual or on your device.
| NOTICE
Information considered important but not hazard related
---|---
| CAUTION
Instructions for use that, if disregarded, might result in product damage.
| WARNING
Instructions for use that, if disregarded, might result in personal injury or
death
| Caution: Risk of Electric Shock
| Warning: Laser Radiation
Chapter 2 Safety
| NOTICE
All statements regarding safety of operation and technical data in this
instruction manual will only apply when the unit is operated correctly.
---|---
| WARNING
Do not open housing. Do not operate without cover installed.
| WARNING
Do not operate in wet or damp conditions.
| WARNING
Light source must be turned off before connecting it to ThetaStation.
Chapter 3 Description
ThetaStation is an opto-mechanical tool designed to perform site-selective
light delivery with OptogeniX tapered fibers1. ThetaStation can be used with
any visible light source. Selection of the emitting sub-region of a Lambda
fiber (or the active optical window of a Sigma fiber) is achieved by manual
operation of a micrometric screw.
ThetaStation can also be used in combination with OptogeniX Magic Box to
perform site-selective fiber photometry. Figure 1 ThetaStation front view with Input
Port (left) and rear view with Output Port (right)
1
https://www.optogenix.com/applications/
Chapter 4 Configuration Guide
4.1 Definitions and abbreviations
NA – numerical aperture
Input Patch cord – Patch cord connecting the light source to the ThetaStation
Input Port.
Input Power – Optical power at ThetaStation input (measured at the output of
the input patch cord connected to light source).
Output Patch Cord – Patch cord connecting the tapered fiber to the
ThetaStation output port
Output Fiber – Common fiber type (NA/core size) for both the output patch cord
and the tapered fiber.
Output Power – Optical power at the output of the output patch cord.
Configuration – A specific combination of light source, input patch cord
(NA/core size), output fiber (NA/core size), and rotary joint (if employed).
Transmission efficiency – Ratio between the output and the input power.
Working range – Range of positions of the micrometric screw within which site-
selective operation is effective.
4.2 Use with Lambda fibers
The full set up for site-selective light delivery with ThetaStation and a
generic light source is shown in Figure 3.LASER sources are recommended for Optogenetic
stimulation. Transmission efficiency decreases with the position of the
micrometric screw from 0mm to end of working range2 . Values of maximum
transmission efficiency achieved with a specific laser source and recommended
input patch cord and output fiber types (NA/core size) are reported in TABLE 1
and can be used for estimating the transmission efficiency achievable with
other laser sources with wavelength in the 400 – 490nm interval.
Source Type| Input Patch Cord| Output Fiber| Transmission
Efficiency (0mm)| Input Power for Optogenetic Stimulation
---|---|---|---|---
LASER (Cobolt 06-MLD 473nm)
| Single mode (SM)| .22NA/105µm| 1%| 30mW
.39NA/200µm| 5%| 6mW
Multimode (MM)
.22NA/50µm| .22NA/105µm| 0.8%| 40mW
.39NA/200µm| 4.5%| 7mW
TABLE I
The last column in TABLE I reports for each listed configuration the input
power required for 1mW/mm2 emission out of a Lambda fiber having the
2 Minimum value of transmission efficiency (typically 70% of maximum) depends
on the actual configuration.
same NA/core size of the output fiber, and at 0mm position of the micrometric
screw3.
General recommendations
- Use a Single Mode (SM) input patch cord for best transmission efficiency;
- The use of a low NA/small core input patch cord gives acceptable transmission efficiency;
- Transmission efficiency increases with both NA and core size of the output fiber. Therefore, small diameter tapered fibers should be preferred only for their reduced invasiveness.
N.B.: Site-selective operation is compatible with the use of a rotary joint.
4.3 Use with .66NA Lambda fibers
With ThetaStation is possible to operate site-selective light delivery with
Lambda fibers based on Plexon high efficiency .66NA fiber4. Values of
transmission efficiency listed in TABLE I for the case of the .39NA fiber can
be used as a reference for a .66NA output fiber type.
3 Assuming 15% power loss at the ferrule-ferrule joint. Calculated from M.
Pisanello et al., Scientific Reports (2018)8:4467
4 Available at https://plexon.com/products/lambda-fibers/
Chapter 5 Setup and Operation
5.1 Connection Guide
- Unbox the ThetaStation and remove all packaging material.
- Connect all the setup components as in Figure 3 of Chapter 4.
N.B.: Connection of the output patch cord is easier when the micrometric screw
is at 0mm position.
N.B.: For FC port type, to ensure a good connection pay attention that the
connector key is oriented to enter within the receptacle slot.
WARNING
The output port and attached patch cord connector move with the rotation of
the micrometric screw. Do not touch them during operation.
5.2 Calibration
Position and size of light emitting sub-regions of Lambda fibers (or windows
emission properties of Sigma fibers), and transmission efficiency need to be
calibrated as a function of the micrometer screw position for:
- every different configuration
- every different-active-length Lambda fiber or different geometry Sigma fiber
Sub-regions position and size (Lambda Fibers)
Immerse the tapered fiber in a solution of fluorophore suitable for the
wavelength in use and place it under a microscope equipped with the
corresponding emission filter5 (Figure 4).With a microscope camera, image the
fluorescence emission as a function of micrometric screw position. With the
help of an image analysis software 6 , for each collected image take the
intensity profile along a line close to the taper edge. The sub-region width
can be estimated as the FWHM of the intensity profile. The distance from the
taper tip of the FWHM middle point is then conventionally defined as the sub-
region position.
5 e.g. PBS:fluorescein solution and FITC filter for blue light (450nm-480nm).
6 e.g. ImageJ (free software)
NOTICE
Sub-regions definition is part of the design of the experiment.
As the active sub-region varies continously with the position of the
micrometric screw, the number and position of the active sub-regions are
defined arbitrary by the end-user during calibration.
Optical Windows activation and emission profiles (Sigma Fibers)
Light emission properties of Sigma fibers as a function of the micrometric
screw position can be verified with the same method described in the previous
paragraph.
Transmission Efficiency
Transmission efficiency as a function of the micrometric screw position can be
easily collected with a light intensity sensor (and related power meter) of
the appropriate wavelenght range, as shown in Figure 5. 5.3 Operation
Use with Lambda Fibers
Select an emitting sub-portion by setting the micrometric screw at the
calibrated position.
For each position, set the light source power to achieve the desired subregion
output power as:
output power = (input power) x (transmission efficiency) x (F-F joint
efficiency)
where F-F joint efficiency accounts for the power loss in the ferrule-ferrule
joint between the output patch cord and the tapered fiber. The exact value of
the F-F joint efficiency depends on the status of connectors (facet
cleanliness and smoothness) and should be measured for any tapered fiber
stub/output fiber combination. Assuming 85% as average value gives a good
estimation of the output power.
Use with Sigma Fibers
Select the active optical window by setting the micrometric screw at the
calibrated position.
The optical power emitted by the active window for a given position of the
micrometric screw must be directly measured with a power meter as a function
of the light source power before fiber implantation.
Chapter 6 Patch cords
Recommended patch cables:
– for 0.22NA and 0.39NA Lambda fibers:
Thorlabs patch cable
configurator
(part numbers: FG105UCA or FT200UMT, respectively)
– for 0.66 Lambda-B fibers:
Plexon starter kit configurator
or Plexon contact form
Chapter 7 Specifications
Weight (kg): | 5 |
---|---|
Enclosure material: | Aluminum 5052-H32 |
Input port: | FC/PC |
Output port: | FC/PC or SMA |
Working range (105/.22NA output fiber) | 0-8mm |
Working range (200/.39NA output fiber) | 0-14mm |
IEC protection class: | Classe III |
Max Input Optical Power | 500mW |
Chapter 8 Troubleshooting
Problem | Solutions |
---|
Irregular light emission from the tapered fiber (e.g. contemporary light
emission from two or more sub-regions of a Lambda fiber)| Output Patch Cord
can be defective.
Verify light emission from the Output Patch Cord by directing it toward a
flat, non-reflecting object (like a piece of paper). Signature of regular
emission as a function of the micrometric screw position is a filled circle at
0mm that transforms in increasing diameter rings until the end of the working
range.
Examples of irregular light emission:
– separate rings emitted at the same position of the micrometric screw;
– rings emitted independently from the micrometric screw position.
Transmission efficiency is significantly lower than the typical value reported
in TABLE I for the same configuration.| Check patch cords connection to the
device. Check for any abnormal operation of the light source. If the problem
persists contact OptogeniX.
Chapter 9 Mechanical Drawing
Chapter 10 Cleaning and Maintenance
ThetaStation can be cleaned using a soft cloth. If needed, the cloth can be
dampened with some isopropyl alcohol.
DANGER
Disconnect the system from accessories and from the light source before any
cleaning and maintenance operation.
Chapter 11 EU Declaration of Conformity
EU Declaration of Conformity
Optogenix S.r.1.
Via Barsanti c/o Istituto Italiano di Tecnologia
73010, Arnesano (LE) — ITALIA
VAT n°: 1T04644390751
This declaration of conformity is issued for:
Opto-mechanical instrument model name/number | ThetaStation |
---|---|
Serial Number of ThetaStation covered by this EU DoC | YYMMDDTS1XX (YY = year, |
MM = month DD = day, TS1 = product code, XX = serial number)
We, Optogenix Srl, declare under our sole responsibility that the above named product conform to the essential requirements of the following European Union directives and normative documents:
-
Direttive 2014/35/UE, Low Voltage (LVD)
-
Directive 2014/30/UE, Electromagnetic Compatibility (EMCD)
-
Direttiva 2011/65/UE, 2015/863/UE Restriction of the use of certain
Hazardous Substances -
Direttiva 2001/95/CE, General Product Safety
-
Directive 2012/19/UE, “RAEE”
-
EC 61511-1:2016 Functional safety – Safety instrumented systems for the process industry sector – Part 1: Framework, definitions, system, hardware and application programming requirements
-
IEC 61882:2016 risk analysis by the HAZOP method
Issued On: April irst, 2020 Leonardo Sileo
Sole administrator
Chapter 12 Warranty
ThetaStation (the Product) is guaranteed to perform per advertised
specifications and is covered against material, manufacturing or design
defects for two (2) years following the date of delivery to buyer. If, prior
to the expiration of the Warranty Period, the Buyer informs OptogeniX in
writing of any breach of this limited warranty, then OptogeniX may repair or
replace the Product that gave rise to the breach or, in OptogeniX’s sole and
exclusive discretion, refund the amounts that Buyer paid for the Product.
Buyer will bear the costs of access, de-installation, re-installation and
transportation of the Product to OptogeniX and back to Buyer. Any repair or
replacement pursuant to this limited warranty will not extend the Warranty
Period. OptogeniX does not warrant the Product, or any repaired or replacement
parts, against normal wear and tear or corrosion. This limited warranty and
remedy are expressly conditioned upon: (i) Buyer’s payment of the purchase
price in full, (ii) Buyer giving written notice of the defect, reasonably
described, to OptogeniX within ten (10) days of the time when Buyer discovers
or ought to have discovered the defect, (iii) the storage, installation,
operation, use, and maintenance of the Product in compliance with the
Instructions, (iv) the existence of proper records of Buyer’s operation and
maintenance of the Product during the Warranty Period, (v) Buyer providing
OptogeniX with a reasonable opportunity to examine the Product and the
aforementioned records, and (vi) the absence of any unauthorized modification
or repair of the Product, including without limitation the removal or
alteration of any serial numbers or warranty date decals.
Before any test may be used to evaluate the Products, Buyer will: (i) provide
OptogeniX with reasonable written notification of the test, (ii) allow
OptogeniX to be present during the test, and (iii) receive OptogeniX’s consent
to the conditions of the test, which consent will not be unreasonably
withheld. If a test is performed on the Products, and OptogeniX has not
consented to the conditions of the test, then this limited warranty will be
void.
Chapter 13 Regulatory
As required by the (Waste Electrical and Electronic Equipment Directive
2012/19/UE) of the European Community (EC) and the corresponding national
laws, OptogeniX offers all end users in the EC the possibility to return “end
of life” units without incurring disposal charges.
This offer is valid for OptogeniX equipment:
- Sold after June 26, 2020
- Marked correspondingly with the crossed out “wheelie bin” logo (see right)
- Sold to a company or institute within the EC
- Currently owned by a company or institute within the EC ****
- Still complete, not disassembled and not contaminated
If you wish to return an OptogeniX unit for waste recovery, please contact
OptogeniX or your nearest dealer for further information.
Waste Treatment
If you do not return an “end of life” unit to OptogeniX, you must hand it to a
company specialized in waste recovery. Do not dispose of the unit in a litter
bin or at a public waste disposal site.
Chapter 14 OptogeniX Contacts
For technical support or sales inquiries, please write us at info@optogenix.com or visit www.optogenix.com.
References
- Home – Optogenetics and Fiber Photometry - OPTOGENIX
- Home – Optogenetics and Fiber Photometry - OPTOGENIX
- Create Your Starter Kit - Plexon
- Plexon Sales Support | Plexon
- Plexon Lambda Fibers | Plexon
- Applications - OPTOGENIX
- Custom Fiber Optic Patch Cables
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>