INTRODUCTION
The GSS CO2 sensor evaluation kit contains the following items.
- CO2 sensor
- USB cable with sensor connector
SOFTWARE INSTALLATION
The GSS evaluation board software is available direct from the GSS web site.
Download the .zip file to your computer. Unzip and click on Setup.exe and
follow the instructions on the screen. A short video showing how to get
started with your evaluation kit and start taking readings in less than 2
minutes is available here: https://www.gassensing.co.uk/videos/
For more technical information on the sensor, download the appropriate data
sheet from here: https://www.gassensing.co.uk/products/
CONNECTOR PIN-OUT DESCRIPTION
The cable has a USB-A connector at one end and a 4-way connector at the other end. The 4-way connector must be connected to the sensor UART pins.
PIN | NAME | TYPE | DESCRIPTION | COLOUR |
---|---|---|---|---|
1 | GND | Supply | Sensor ground | Black |
2 | VDD | Supply | Sensor supply voltage | Red |
3 | Rx_In | Digital Input | UART Receive Input | Orange |
4 | Tx_Out | Digital Output | UART Transmit Output | Yellow |
Ensure the connector is in the correct position before plugging the USB
connector into a USB port.
Full sensor pin-out details are provided in the appropriate product
datasheets.
USB DRIVER
The USB cable contains an integrated circuit that converts from the sensor
UART output to the PC USB input. Recent versions of Windows will automatically
identify and install the USB driver when you plug in the lead.
If you are prompted to locate a driver, download from the FTDI website:
https://www.ftdichip.com/Drivers/VCP.htm
Choose “VCP Drivers” and select the correct driver for your operating system.
SOFTWARE INSTALLATION
The GSS sensor software can be found on the GSS website in the resources section. Click on Setup.exe and follow the instructions on the screen.
RUNNING THE SOFTWARE
The software may start automatically after installation. If it does not, you
can start it from the “Program” Menu in the Start Menu.
The opening screen is below. To connect to the sensor, select the correct
COM port from the drop-down Connection list, then click the CONNECT button.
The drop-down list will only show ports with active devices attached.
SENSOR SELECTION
Select the correct sensor type and the full-scale range (CO2 ppm) of the
sensor. The sensor should automatically start to take CO2 measurements after
1-2 seconds. These measurements are shown graphically in the middle of the
screen, and in the dialogue box (red) at the bottom. The dialogue box contains
sensor serial data output (‘Z’ data). The CO2 ‘Z’ value needs to be scaled to
calculate the ppm value.
Ambient sensors measuring up to 1% (10000ppm) do not need a scaling factor,
for sensors capable of measuring 0-60%, the scaling factor is 10. For sensors
capable of measuring 0-100%, the scaling factor is 100.
CONTROLLING THE SENSOR
The software interface replicates the UART controls of the sensor. For full details of the functions, refer to the product data sheet. After power is applied, the sensor will automatically start to take CO2 measurements using the Mode 1 default settings, where the sensor is pre-programmed to send CO2 measurement data. This mode is labelled as ‘Stream (K1)’ mode.
MODE K1 STREAMING MODE
Click on the STREAM button. This is the factory default setting, where the
sensor is constantly taking measurements at a pre-programmed rate, defined by
the sensor type.
MODE K0 COMMAND MODE
Click on the STOPPED button. In this mode, the sensor is in a SLEEP mode,
waiting for commands. No measurements are made. There is no latency in command
responses. All commands that report measurements or alter the zero-point
settings are disabled in Mode 0. Mode 0 is NOT retained after power cycling.
MODE K2 POLLING MODE
Click on the POLLING button. In polling mode, the sensor only reports readings
when requested. The sensor will continue to take measurements in the
background, but the output stream is suppressed until data is requested. In
this mode, the user can programme the sensor to take a reading on request, or
at a pre-defined period.
ZERO-POINT SETTING COMMANDS
There are a several methods available to the user to set the zero point of the sensor. In all cases, the best zero is obtained when the gas concentration is stable, and the sensor is at a stabilised temperature. Note the zero-point settings are not cumulative and only the latest zero-point setting is effective. For example, there is no benefit in zeroing in nitrogen, and then zeroing in a calibration gas. The sensor will store only the latest zero point.
ZERO IN NITROGEN
Place the sensor in nitrogen gas and allow time for the sensor temperature to
stabilise and the gas to be fully diffused into the sensor. Click on the
NITROGEN (U) button. The sensor is zeroed assuming a 0ppm CO2 environment.
ZERO IN FRESH AIR
If there is no calibration gas or nitrogen available, the sensor zero-point
can be set in fresh air.
Ambient CO2 concentrations in fresh air are typically 400ppm. The CO2
concentration fresh air zero level is programmable over a range from 0ppm to
the full scale of the sensor.
Place the sensor in a fresh air environment and allow time for the sensor
temperature to stabilise, and for the fresh air to be fully diffused into the
sensor. Click on the AMBIENT (G) button.
The sensor can use the default fresh air CO2 concentration value (400ppm), or
the user can write a different fresh air value to the sensor if desired. To
write a different fresh air CO2 concentration value to the sensor, write the
new value in the ‘set to’ box.
ZERO IN A KNOWN GAS CONCENTRATION
Place the sensor in a known gas concentration, and allow time for the sensor
temperature to stabilise, and for the gas to be fully diffused into the
sensor. Set the CO2 concentration value to the appropriate value by writing
the value in the ‘set to’ box. Click on the KNOWN GAS (X) button.
ZERO-POINT ADJUSTMENT
If the CO2 concentration and the sensor reported concentration are known (both
in ppm), the zero point can be adjusted using the known concentration to fine
tune the zero point. For example, if the sensor has been in an environment
that has been exposed to outside air, and the sensor reading is known at that
time, the zero point can be fine-tuned to correct the reading. This is
typically used to implement automated zeroing routines.
The known CO2 concentration value should be entered into the ‘new’ box, the
reported CO2 value from the sensor into the ‘old’ box and click Finetune(F).
POLLING COMMANDS
Depending on the sensor type, various parameters can be output as a single
string of data in the dialogue box. If a temperature or humidity sensor is
fitted, the dialogue box will show the relevant data fields depending on user
requirements.
If running in polling mode, a single line of data is output each time the
‘POLLING (K2)’ button is clicked, or automatically within a set interval if
preferred by selecting the ‘Poll every seconds’ option.
INFO COMMANDS
ID
Clicking the ‘ID’ button will return the sensor serial number and firmware
version.
DEBUG
Clicking the ‘Debug’ button will return additional information about the
settings of the sensor. It is primarily intended to assist in remote debugging
by GSS Ltd.
CUSTOMISING SENSOR SETTINGS
The evaluation kit software allows the user to fully customise the settings of the sensor. Refer to the sensor data sheet for a full command and control description. All UART commands and functions of the sensor can be manually set by writing in the blank box with the correct syntax. To send an UART command or data to the sensor, write in the blank box and click the ‘SEND’ button.
DATALOGGER
You can record the sensor output to a text file. To start recording, click
‘Record’ button and follow the instructions. Click ‘Stop’ button to stop
recording.
The datalogger will record the same information as is displayed in the
dialogue box and will precede each line with a timestamp.
25/03/2016 14:07:51 Z 00426 z 00434
25/03/2016 14:07:51 Z 00426 z 00450
25/03/2016 14:07:52 Z 00428 z 00448
25/03/2016 14:07:52 Z 00428 z 00436
25/03/2016 14:07:53 Z 00428 z 00417
25/03/2016 14:07:53 Z 00428 z 00440
25/03/2016 14:07:54 Z 00428 z 00417
FAULT FINDING
Fault | Resolution |
---|---|
No output from sensor | If you do not see readings from the sensor after a few |
seconds, or if you get an error message indicating a problem, then follow
these steps in order:
• Check cable is connected correctly to the UART pins on the sensor (refer to
data sheet for pin-out)
• Check that you are clicking on the correct com port. If you cannot see the
USB lead in the comport list, there has been a problem installing the USB
driver. Check that you have downloaded the correct driver and follow the
instructions on the FTDI website.
• Select the COM Port corresponding to the USB Lead. Check the available ports
with the USB lead unplugged. Then plug in the USB
lead and note the new comport.
• Check the sensor is in ‘streaming’ mode by clicking the button marked
‘Stream (K1)’
Sensor reading is high/low| Press Zero-point setting – Ambient (air). Sensor
should output a reading of approximately 400ppm.
EVK EXAMPLES
IMPORTANT NOTICE
Gas Sensing Solutions Ltd. (GSS) products and services are sold subject to
GSS’s terms and conditions of sale, delivery and payment supplied at the time
of order acknowledgement. GSS warrants performance of its products to the
specifications in effect at the date of shipment. GSS reserves the right to
make changes to its products and specifications or to discontinue any product
or service without notice.
Customers should therefore obtain the latest version of relevant information
from GSS to verify that the information is current. Testing and other quality
control techniques are utilised to the extent GSS deems necessary to support
its warranty. Specific testing of all parameters of each device is not
necessarily performed unless required by law or regulation. In order to
minimise risks associated with customer applications, the customer must use
adequate design and operating safeguards to minimise inherent or procedural
hazards. GSS is not liable for applications assistance or customer product
design. The customer is solely responsible for its selection and use of GSS
products. GSS is not liable for such selection or use nor for use of any
circuitry other than circuitry entirely embodied in a GSS product.
GSS products are not intended for use in life support systems, appliances,
nuclear systems or systems where malfunction can reasonably be expected to
result in personal injury, death or severe property or environmental damage.
Any use of products by the customer for such purposes is at the customer’s
own risk.
GSS does not grant any licence (express or implied) under any patent right,
copyright, mask work right or other intellectual property right of GSS
covering or relating to any combination, machine, or process in which its
products or services might be or are used. Any provision or publication of any
third party’s products or services does not constitute GSS’s approval,
licence, warranty or endorsement thereof. Any third party trademarks contained
in this document belong to the respective third-party owner.
Reproduction of information from GSS datasheets is permissible only if
reproduction is without alteration and is accompanied by all associated
copyright, proprietary and other notices (including this notice) and
conditions. GSS is not liable for any unauthorised alteration of such
information or for any reliance placed thereon.
Any representations made, warranties given, and/or liabilities accepted by any
person which differ from those contained in this datasheet or in GSS’s
standard terms and conditions of sale, delivery and payment are made, given
and/or accepted at that person’s own risk. GSS is not liable for any such
representations, warranties or liabilities or for any reliance placed thereon
by any person.
ADDRESS
Gas Sensing Solutions Ltd.
Grayshill Road
Cumbernauld
G68 9HQ
United Kingdom
REVISION HISTORY
DATE | RELEASE | DESCRIPTION OF CHANGES | PAGES |
---|---|---|---|
13/05/2020 | 1.0 | First revision | All |
27/05/2020 | 1.1 | Re-write | All |
29/05/2020 | 1.2 | Added revision 2.1 of evaluation kit software | All |
28/01/2021 | 1.3 | Minor update to contents of kit | P.3 |
9/6/2021 | 1.4 | Software Installation | P.4 |
21/07/2022 | 1.5 | Added picture of SprintIR® -65/R | P.13 |
Gas Sensing Solutions Ltd.
For regular updates, sign up at https://gassensing.co.uk
Revision 1.5, 21 July 2022