**PDC UV-C LED Bore Disinfection System
**
IMPORTANT INSTRUCTIONS
Step 1: Watch the 6-minute UV-C Quick Start video and read below STEPS
before proceeding (scan QR code above or go to http://bit.ly/3FBbSDy)
Step 2 : Position UV-C cart near end of patient table farthest from
magnet bore. Ensure it’s at least 5 ft. from bore
Step 3: Remove SECTION 1 from cart and place it on patient table with its
support legs near magnet bore
Step 4: Remove SECTION 2 from cart and place it on patient table with its
support legs near end of table farthest from bore
Step 5: Connect both sections together as shown in the video
Step 6: Move the patient table into the bore
Step 7: Plug the GREY UV-C cable into SECTION 2 as shown in the video
Step 8 : Plug the BLACK power cord into a wall outlet – BLUE light starts
flashing* for 60 seconds while UV-C System initializes (during steps 8-11)
Step 9: Position UV-C cart so lights are visable from viewing window
Step 10: Turn key to ENABLE SYSTEM – key must be turned on while BLUE
light is flashing
Step 11: Push Power button beside key switch – while BLUE light is
flashing UV-C System will not start until motion sensors detect no motion in
the room
Step 12: Walk out of room and close the door
– Refer to UV-C System LIGHT KEY on back
– If motion is detected before cycle begins, GREEN light will flash rapidly
until no motion has been detected for 30 seconds
Step 13: Disinfection cycle begins and lasts 3 minutes
– BLUE light is ON, GREEN light is flashing slowly
– If cycle is interrupted, turn key to DISABLE and start at Step 10
– Once ALL 3 lights on the cart are ON – Disinfection cycle is COMPLETE
Step 14: Enter room and turn key to DISABLE – BLUE light starts flashing
Step 15: Unplug the BLACK power cord and hang it on UV-C cart
Step 16: Unplug the GREY UV-C cable and hang it on UV-C cart
Step 17: Return patient table to Home position
Step 18: Disconnect SECTIONS 1 and 2 and hang each section on UV-C cart
– Support legs near bottom of cart facing in
*If an error condition occurs, the RED lights will flash intermittently. See back of card.
LIGHT KEY: Flashing Constant ON
ERROR CODE PER FLASHING SEQUENCE OF RED LIGHTS
ERROR CODE| FOLLOW THESE STEPS
RED LIGHT| Turn key OFF
FLASHES| Go to Step 10 on Quick Start Guide. Begin at Step 10 and
continue
2 Slow Flashes| until complete
1 Fast Flash| If condition repeats contact PDC at
262-369-1929
CONNECTION ERROR CONDITION
BLUE LIGHT AND RED LIGHTS FLASHING| Verify both SECTION 1
and SECTION 2 are connected and together at connecting point Verify UV-C cable
is plugged securely into the end of SECTION 2
KEY – UV-C SYSTEM LIGHTS
FLASHING BLUE
| UV-C System Initializing
SOLID BLUE
| UV-C System Ready
SOLID GREEN
| Disinfection Cycle Started – verifying no motion
FAST FLASHING **GREEN
**
| Motion Detected before UV-C Disinfection Cycle starts, UV-C Disinfection Cycle will start 30 seconds after last motion is detected
SLOW FLASHING **GREEN
| UV-C System Running
ALL 3 SOLID
| Disinfection Cycle Complete
FLASHING RED
| Error (refer to error code guide above)
Technical Support: 1-262-369-1929 •
pdcbiz.com