Bio Compression SC 4004 DL Sequential Circulator Instruction Manual
- June 13, 2024
- Bio Compression
Table of Contents
SEQUENTIAL CIRCULATOR
Operating Instructions
Quality Medical Products Since 1983
Introduction
Congratulations on the purchase of your sequential circulator and garments.
Package Contents
- Sequential Circulator pump
- Power cord
- Instructions for use
- Blocker bar
- Garment(s) – possibly packed separately
Intended Use
Sequential Circulators are pneumatic compression devices intended for either
primary or adjunctive treatment of lymphedema, peripheral edema, lipedema,
venous insufficiency, and venous stasis ulcers. Sequential Circulators are
also intended for the prophylaxis. Intended for use in a home or healthcare
setting.
Contraindications
Compression IS NOT recommended in the following conditions:
- Infections in the limb, including cellulitis, without appropriate antibiotic coverage
- The presence of lymphangiosarcoma
- Suspicion or confirmation of the presence of Deep Vein Thrombosis (DVT)
- Inflammatory phlebitis or episodes of pulmonary embolism
- Congestive Heart Failure (CHF) unless controlled by medication
- Other indications as identified by the treating physician
Guidelines for Treatment
A physician is required to prescribe these settings, but general guidelines
are listed below:
- It is ultimately the physician’s responsibility to prescribe the setting and it should be written on the prescription upon referral. Every patient is unique and communication with the physician is important when setting the pressure.
- 50 mmHg works well for most patients. However, a different pressure might be prescribed for your personal need and tolerance.
- Presence of fibrotic tissue may require as much as 80 mmHg in order to soften the fibrotic tissue and achieve reduction. Once the tissue is soft, the compression can be readjusted to 50 mmHg.
- Patients with a history of Congestive Heart Failure (CHF), which is controlled with medication, should never be in a flat position while pumping. They should be in a reclined position with elevated legs during treatment. Their treatment regimen duration may be divided into twice a day 30 minutes per treatment.
- Patients with a history of Deep Vein Thrombosis with or without a filter may require less compression. These patients will generally tolerate 40 mmHg. Patients with a filter may need to divide their treatment into twice a day, 30 minutes per treatment. It is suggested that the provider obtain a negative Doppler study from the physician for their records.
Device Description and Operating Principle
Sequential Circulators provide gradient pneumatic compression for the
treatment of lymphedema and associated venous disorders. Sequential gradient
compression helps to increase blood flow and move excess lymph away from the
affected area for clearance from the body. This device can provide sequential
(distal to proximal) inflation/deflation cycles of compressed air at
prescribed pressures. Pneumatic compression devices are proven to reduce limb
swelling associated with lymphedema, help close up chronic ulcers, and act as
a prophylaxis for Deep Vein Thrombosis (DVT).
Adjustable Pressure
The pressure of the pump can be adjusted between 10 and 120 mmHg. Pressure
can be selected before treatment and adjustments can be made during treatment.
Adjustable Cycle Time
The cycle time is the time it takes the pump to inflate and deflate the
garment. The cycle time can be adjusted from 60 to 120 seconds in 15 second
intervals.
Timed Treatment
The pump can be run continuously or the treatment time can be set between 10
and 120 minutes in 5 minute increments.
Focus Therapy
Focus Therapy is used to help treat patients at the beginning of their session
in an area that requires more attention. Focus therapy doubles the inflation
time of two adjacent chambers (SC-4004-DL) or three adjacent chambers
(SC-4008-DL) during the first ten minutes of treatment.
Pre-Therapy
Pre-Therapy is only available in the SC-4008-DL. It is a setting that
allows a patient to inflate chambers 6, 7, and 8 for ten minutes prior to
running a full cycle of chambers 1 through 8.
Pause Feature
The Pause button gives the patient the option to stop the pump in the middle
of a treatment session allowing them to use the bathroom or attend to other
needs. Timed treatments will run for the full treatment time even if paused to
ensure patient receives appropriate therapy.
Treatment Compliance Meter
The pump records the hours of use.
Front Panel
- Touch Screen LCD Display
- Garment Connector Bar Port
- Auxiliary Connector Bar Port (shown with Blocker Bar)
The SC-4008-DL has two Garment Connector Bar Ports and two Auxiliary Connector
Bar Ports as shown below:
Warnings and Precautions
US federal law restricts this device to sale by or on the order of a
physician.
Electrical Medical Equipment
- To avoid the risk of electric shock, burns, fire, injury, or improper treatment, read the entire instruction manual before operating this device
- Use of accessories or a power cord not specified or provided by Bio Compression Systems could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation
- Portable RF communications equipment (including cell phones and peripherals such as antenna cables and external antennas) should be used no closer than 12″ (30 cm) to any part of the device including the power cord – otherwise, degradation of the performance of this equipment could result
- Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation
Do Not Use
- For any contraindicated condition
- If the pump, accessories, or power cord are damaged or have been immersed in water
- With any accessories or power cord not specified or provided by Bio Compression Systems
- In the presence of flammable anesthetics or in an oxygen rich environment
- In an MRI environment
- Near water, in a wet environment, or where aerosols are being sprayed
- While sleeping
- For any use not described in this manual
Ask A Doctor Before Use If You Have
- Insensitive, irritated, injured skin, or skin conditions in/around treat sites
When Using This Product
- Examine the device, accessories, and power cord for damage before using or cleaning
- Handle garments with care – do not fold or crease, use near a heat source, handle with sharp objects, clean with abrasive materials, or place in a washing machine or dryer
- Do not stand or walk while wearing garments as this can cause a fall
- Always wear clothing, bandages, or stockings underneath garments for hygienic reasons and to avoid irritation
- Never share garments or use someone else’s garments – single-patient use only
- Avoid folding, pinching, or kinking tubing as this can obstruct air flow
- Do not wrap tubing around limb as this can restrict blood flow
- Do not operate pump on a soft surface, under a blanket or covering, or near a heat source
- Never adjust pump settings unless directed by a physician
- Do not carry or suspend the device using tubing, valves, or the power cord as handles
- Do not submerge the device or allow liquids to enter the device
- Never attempt to open, repair, or modify the device – no modification of this equipment is allowed
Stop Use And Ask A Doctor If
- Changes in skin appearance occur such as color changes, blisters, welts, or increased swelling
- You feel burning, itching, increased pain, numbness, or tingling
In the event the pump stops working (e.g., power failure), release pressure by
disconnecting the garment.
Any serious incident that has occurred in relation to the device must be
reported to Bio Compression Systems. In the European Union (EU), incidents
should also be reported to the competent authority of the Member State in
which the user and / or patient is established.
Keep out of reach of children and pets.
Operating Instructions
The patient is the intended user and can safely use all functions.
Preparing the Device for Use
- Remove garment(s) and cardboard underneath from box
- Remove garment(s) from plastic bag, unroll, and spread flat
- Lift pump out of box and remove protective end caps – save packaging for transport and storage
- Place pump on a flat sturdy surface – pump must be close enough for the controls to be reached during use
- Attach power cord and plug into outlet
- Pump is set to run for one hour at 50 mmHg with 60 second cycle time – to change settings, see below
To Change Pump Settings
- Begin with the pump turned off
- Press the Touch Screen LCD Display (1) to wake up your device
- When the home screen lights up, press and hold the lower right corner of the Touch Screen LCD Display (1) for 3 seconds until you see the main setup screen
- You can review the pump’s current Pressure, Cycle Time, and Treatment Time settings on this main screen – if the current settings are acceptable, hit “Done”
- To adjust the settings, press “Setup” in the lower right corner
- The first screen that appears is the Pressure settings screen
- Press “Up” and “Down” to adjust the pressure, beginning with Chamber 1, and press “Chamber” to move to the next chamber
- Repeat the previous steps to set each chamber – please note that you cannot set the pressure in the next chamber higher than the previous chamber
- When all chambers have been set, press “Next”
- You are now on the Cycle Time screen – press “Up” and “Down” to adjust the cycle time and press “Next” when finished
- You are now on the Treatment Time screen – press “Up” and “Down” to adjust the treatment time
- To set on continuous mode, press “Up” one more time after reaching “120” and the word “Continuous” will appear – press “Next” when finished
- You are now on the Focus Therapy screen – press “Up” and “Down” to select chambers for focus therapy or “Off” to turn off
- At this point the SC-4008-DL will switch to the Pre-Therapy screen
- Press “On” to turn pre-therapy on and “Off” to turn off
- You can now review your settings on the screen. Hit “Done” if your new settings are correct, hit “Setup” if you want to make changes and follow directions above
Connecting the Garment(s)
- Locate the connector bar at the end of garment tubing
- Line up numbers on connector bar with numbers on pump
- Squeeze the sides and insert into pump – you will hear a “click” when connected
- (SC-4008-DL only) Locate the second connecter bar and repeat previous steps
- When using two garments, remove blocker bar(s) and repeat previous steps
Putting the Garment(s) On
- Leg sleeve: unzip, place foot in and use straps to guide garment on, pull up zipper to secure
- Arm sleeve: slide arm through larger opening
Operating the Device
- Sit in a comfortable reclined position with legs elevated within reach of controls
- Press the Touch Screen LCD Display (1) – to wake up your device – the pressure, cycle time, and treatment time appear at the top of the screen
- Press “Start” to begin treatment. If you do not press “Start” within 1 minute, your device will go back into sleep mode.
- At the end of a timed treatment, your pump will turn off
- To turn off the pump during a treatment, press “Stop” – “Please Wait” will appear while the pump returns to its resting position
- Once the pump has returned to its resting position, the sleeve will deflate and the LCD will turn off
- Squeeze sides of connector bar and pull to disconnect garment from pump
- Press garment to remove remaining air until loose enough to remove
- Unzip (if applicable) and remove garment
Pausing treatment
- To pause treatment, press “Pause” – “Please Wait” will appear while the pump returns to its resting position
- Once the pump has returned to its resting position, the sleeve will deflate and “Paused, Press to Continue” will appear (remove garment using directions above)
- Press “Press to Continue” to resume treatment
Changing pressure during treatment
- To change pressure during treatment, press and hold the lower right corner of the Touch Screen LCD Display (1) for 5 seconds – “Please Wait” will appear while the pump returns to its resting position
- Once the pump has returned to its resting position, the sleeve will deflate and “Change Pressure Setting” will appear – press “Up” and “Down” to change pressure and “Chamber” to select the next chamber
- Press “Set” when finished and the pump begins to operate again
To See Treatment Compliance Meter (Hours of Use)
- Wake up your device and wait for the “Start” screen to appear
- Press and hold the bottom center of the Touch Screen LCD (1) for 3 seconds until hours of use information appears
- Hours of use will appear for 5 seconds
- To reset the device to Factory Defaults, while the compliance meter is showing press and hold the bottom right corner of the LCD. A message will ask if you want to “Reset to Factory Defaults” – press “No” to return to the “Start” screen and “Yes” to reset.
Cleaning
The pump, garment, and tubing can be wiped down using a damp (not wet) soft
cloth while unplugged – if more thorough pump cleaning or garment disinfection
is desired, use the following directions.
Pump and tubing cleaning
- Unplug and wipe down using a damp (not wet) soft cloth with mild antibacterial soap as needed
- Do not use bleach
Garment disinfection
- Disconnect from pump and open to expose all sides
- Prepare a solution of 1/3 cup of laundry detergent per 1 gallon of warm water (20 mL laundry detergent per 1 L water) in a sink or container large enough to hold the garment
- Place garment in solution but do not submerge or place connectors in water as this will damage the device
- Soak for 30 minutes with mild agitation every 5-10 minutes – hard to remove soil may require hand washing using a soft clean cloth while garment is in solution
- Rinse with warm water and allow to air dry
- Repeat previous steps using a solution of 1 cup of bleach per 1 gallon of warm water (60 mL bleach per 1 L warm water)
Storing and Transporting
- Keep and reuse packaging for transporting the device
- Store in a dry location away from a source of heat and free of pests
Servicing and Repairs
- Contact Bio Compression Systems for servicing – there are no user serviceable parts
- Tampering, modifying, or dismantling this device in any way voids the warranty
- When contacting Bio Compression Systems, please have your model number and serial number ready
Troubleshooting
Pump does not turn on:
- Check to see if the pump is plugged in
- Unplug and examine power cord for damaged – if damaged, contact Bio Compression Systems
- Check circuit breaker to make sure outlet has power
- Contact Bio Compression Systems
Garment does not deflate:
- Check garment connection to pump
- Check garment hose for damage, kinks, or twists
- Check garment for damage
- Contact Bio Compression Systems
Pressure seems low:
- Check garment connection to pump
- Check garment hose for damage, kinks, or twists
- Check garment for damage
- Contact Bio Compression Systems
Device is loud or making strange noises:
- Make sure pump is on a stable surface
- Assure stable surface is free and clear of any loose object
- Contact Bio Compression Systems
Accessories (SC-4004-DL)
REF | Description |
---|---|
GS-3035-S | 4-Chamber Arm Sleeve – Small |
GS-3035-M | 4-Chamber Arm Sleeve – Medium |
GS-3035-L | 4-Chamber Arm Sleeve – Large |
GS-3035-SH-S-L | 4-Chamber Arm & Shoulder Sleeve – Small, Left |
GS-3035-SH-S-R | 4-Chamber Arm & Shoulder Sleeve – Small, Right |
GS-3035-SH-M-L | 4-Chamber Arm & Shoulder Sleeve – Medium, Left |
GS-3035-SH-M-R | 4-Chamber Arm & Shoulder Sleeve – Medium, Right |
GS-3035-SH-L-L | 4-Chamber Arm & Shoulder Sleeve – Large, Left |
GS-3035-SH-L-R | 4-Chamber Arm & Shoulder Sleeve – Large, Right |
GV-3000 | 4-Chamber Vest |
GS-3045-H | 4-Chamber Leg Sleeve – Half |
GS-3045-S | 4-Chamber Leg Sleeve – Small |
GS-3045-M | 4-Chamber Leg Sleeve – Medium |
GS-3045-L | 4-Chamber Leg Sleeve – Large |
GN-3045-S | 4-Chamber Narrow Leg Sleeve – Small |
GN-3045-M | 4-Chamber Narrow Leg Sleeve – Medium |
GN-3045-L | 4-Chamber Narrow Leg Sleeve – Large |
GW-3045-H | 4-Chamber Wide Leg Sleeve – Half |
GW-3045-S | 4-Chamber Wide Leg Sleeve – Small |
GW-3045-M | 4-Chamber Wide Leg Sleeve – Medium |
GW-3045-L | 4-Chamber Wide Leg Sleeve – Large |
GXW-3045 | 4-Chamber Extra Wide Leg Sleeve |
GA-3045-H | 4-Chamber Adjustable Leg Sleeve – Half |
GA-3045-S | 4-Chamber Adjustable Leg Sleeve – Small |
GA-3045-M | 4-Chamber Adjustable Leg Sleeve – Medium |
GA-3045-L | 4-Chamber Adjustable Leg Sleeve – Large |
GWA-3045-H | 4-Chamber Wide Adjustable Leg Sleeve – Half |
GWA-3045-S | 4-Chamber Wide Adjustable Leg Sleeve – Small |
GWA-3045-M | 4-Chamber Wide Adjustable Leg Sleeve – Medium |
GWA-3045-L | 4-Chamber Wide Adjustable Leg Sleeve – Large |
GXWA-3045 | 4-Chamber Extra Wide Adjustable Leg Sleeve |
Accessories (SC-4008-DL)
REF | Description |
---|---|
G8-3035-S | 8-Chamber Arm Sleeve – Small |
G8-3035-M | 8-Chamber Arm Sleeve – Medium |
G8-3035-L | 8-Chamber Arm Sleeve – Large |
G8-3035-SH-S-L | 8-Chamber Arm & Shoulder Sleeve – Small, Left |
G8-3035-SH-S-R | 8-Chamber Arm & Shoulder Sleeve – Small, Right |
G8-3035-SH-M-L | 8-Chamber Arm & Shoulder Sleeve – Medium, Left |
G8-3035-SH-M-R | 8-Chamber Arm & Shoulder Sleeve – Medium, Right |
G8-3035-SH-L-L | 8-Chamber Arm & Shoulder Sleeve – Large, Left |
G8-3035-SH-L-R | 8-Chamber Arm & Shoulder Sleeve – Large, Right |
G8-3045-S | 8-Chamber Leg Sleeve – Small |
G8-3045-M | 8-Chamber Leg Sleeve – Medium |
G8-3045-L | 8-Chamber Leg Sleeve – Large |
GN8-3045-S | 8-Chamber Narrow Leg Sleeve – Small |
GN8-3045-M | 8-Chamber Narrow Leg Sleeve – Medium |
GN8-3045-L | 8-Chamber Narrow Leg Sleeve – Large |
GW8-3045-S | 8-Chamber Wide Leg Sleeve – Small |
GW8-3045-M | 8-Chamber Wide Leg Sleeve – Medium |
GW8-3045-L | 8-Chamber Wide Leg Sleeve – Large |
GXW8-3045 | 8-Chamber Extra Wide Leg Sleeve |
A8-3045-S | 8-Chamber Adjustable Leg Sleeve – Small |
A8-3045-M | 8-Chamber Adjustable Leg Sleeve – Medium |
A8-3045-L | 8-Chamber Adjustable Leg Sleeve – Large |
GWA8-3045-S | 8-Chamber Wide Adjustable Leg Sleeve – Small |
GWA8-3045-M | 8-Chamber Wide Adjustable Leg Sleeve – Medium |
GWA8-3045-L | 8-Chamber Wide Adjustable Leg Sleeve – Large |
GXWA8-3045 | 8-Chamber Extra Wide Adjustable Leg Sleeve |
GBA-3045-S-2 | Bio Pants – Small |
GBA-3045-M-2 | Bio Pants – Medium |
GBA-3045-L-2 | Bio Pants – Large |
GBA-3045-S-L | Bio Abdominal – Small, Left Leg |
GBA-3045-S-R | Bio Abdominal – Small, Right Leg |
GBA-3045-M-L | Bio Abdominal – Medium, Left Leg |
GBA-3045-M-R | Bio Abdominal – Medium, Right Leg |
GBA-3045-L-L | Bio Abdominal – Large, Left Leg |
GBA-3045-L-R | Bio Abdominal – Large, Right Leg |
GV-3010-S-L | Elite 8 Bio Vest – Small, Left |
GV-3010-S-R | Elite 8 Bio Vest – Small, Right |
GV-3010-M-L | Elite 8 Bio Vest – Medium, Left |
GV-3010-M-R | Elite 8 Bio Vest – Medium, Right |
GV-3010-L-L | Elite 8 Bio Vest – Large, Left |
GV-3010-L-R | Elite 8 Bio Vest – Large, Right |
GV-3010-S-2 | Elite 8 Vest with Bilateral Arms – Small |
GV-3010-M-2 | Elite 8 Vest with Bilateral Arms – Medium |
GV-3010-L-2 | Elite 8 Vest with Bilateral Arms – Large |
Product Specifications
Models: SC-4004-DL, SC-4008-DL
Power Supply Rating: 120-240V, 50/60 Hz
Rated Input: 12VDC, 3A
Electrical Classification: Class II
Applied Part: Type BF
Ingress Protection: IP21
Mains Isolation: Unplug
Mode of Operation: Continuous
Essential Performance: The pump’s cyclical inflation and deflation of the
garment(s)
Cycle Time: 60-120 seconds in 15 second increments
Treatment Time: Continuous or 10-120 minutes in 5 minute increments
Pressure Range: 10-120 mmHg
Precision: 1 mmHg
Accuracy: ± 20%
Features: Adjustable Cycle Time, Compliance/Usage Meter, Individual Pressure
Adjustment in Each Chamber, Focus Therapy, Pause, PreTherapy, Timed Treatment
Warranty: Pump 3 years, garment 1 year
Expected Service Life: 5 years
Software Safety Class: A
Regulatory Classification: AU IIa, CA 2, BR II, EU IIa, US 2
Weight (SC-4004-DL) 3.5 lbs. (1.59 kg)
Weight (SC-4008-DL): 3.85 lbs. (1.75 kg)
Dimensions: 4.5” x 12” x 7.34” (114 mm x 304 mm x 186 mm)
Environmental Specifications
Consumables and Natural Resources Used During Care and Use
- Electrical energy for operation
- 70 mL laundry detergent and 250 mL bleach per 7.6 liters water for garment cleaning – only as needed
Emissions During Normal Use
- Compressed air
- Minimal acoustic energy – nearly silent
- Minimal electromagnetic emissions – see manufacturer’s declaration and related information below
Instructions for Minimizing Environmental Impact
- Unplug the pump after charging – unplugging electronic devices when they are not being used saves electricity
- Unplug the pump when not in use – unplugging electronic devices when they are not being used saves electricity
- Do not clean garment soiled – this minimized the consumables used
- Reused packaging for storing and transporting device
Operation Environment
- Intended for use in a healthcare or home environment
- Not intended of use in the presence of flammable anesthetics, an oxygen rich environment, or an MRI environment
- Altitude up to 6561 feet (2000 m)
- Temperature 50ºF – 100ºF (10ºC – 38ºC)
- Humidity 30-75% RH
- Atmospheric pressure 700-1060hPa
Transportation and Storage Environment
- Temperature -20°F – 110°F (-29°C – 43°C)
- Humidity 30-75% RH
- Atmospheric pressure 700-1060 hPa
End of Life Management
- There are no components which contain stored electrical energy after the device has been shut off
- Does not contain hazardous substances requiring special handling and treatment
- Dispose of in an environmentally responsible manner in accordance with regional requirements
- Contact Bio Compression Systems if you have questions or concerns regarding disassembly and disposal
Manufacturer’s EMC Declaration
Electromagnetic Emissions
E missions | Compliance | Electromagnetic Environment – Guidance |
---|---|---|
RF emissions CISPR 11 | Group 1 | The device uses RF energy only for its |
internal functions. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to the public low voltage power
supply network that supplies building used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3| Not applicable
Electromagnetic Immunity
Immunity Test | Immunity Test Level |
---|---|
IEC 61000-4-2 Electrostatic Discharge Immunity | ±8kV contact, ±2, 4, 8, 16kV |
air discharge
IEC 61000-4-3 Radiated RF Field Immunity| 80MHz — 2.7GHz, 10V/m, AM 80% at
1kHz
IEC 61000-4-3 Proximity Fields from RF Wireless Communications Equipment| IEC
60601-1-2, Section 8.10, Table 9
IEC 61000-4-4 Electrical Fast Transients| ±2kV/100kHz power, ±1kV/100kHz
signal
IEC 61000-4-5 Surge Immunity| ±0.5, 1kV line to line, ±0.5, 1, 2kV line to
ground
IEC 61000-4-6 Conducted RF Immunity| 150kHz – 80MHz, 3VRms in whole range,
6VRms in amateur radio and
ISM, AM 80% at 1kHz
IEC 61000-4-8 Magnetic Field Immunity| 30A/m, 50 or 60Hz
IEC 61000-4-11 Voltage Dips| 0% UT per 0.5 cycles, 0% UT per 1.0 cycle, 70% UT
per 25/30
cycles
IEC 61000-4-11 Voltage Interruptions| 0% UT per 250/300 cycles
Symbol Glossary
| Authorized Representative in the European Community
---|---
| Atmospheric pressure limitation
| Batch code (lot number)
| Catalog number
| Caution
| Class II equipment (protection against electric shock)
| Complies with the Waste Electrical and Electronic Equipment Directive
(WEEE Directive)
| Complies the European Medical Device Regulation
| Date of manufacture
| Fragile, handle with care
| Humidity limitations
IP21| Ingress protection (against solids up to 12.5 mm and dripping
water)
| Manufacturer
| Medical Device
| Keep dry
| Power on/off (stand-by)
| Refer to instruction mannual booklet
| Restricted to sale by or on the order of a physician
| Serial number
| Temperature Limit
| This way up
| TOV SOD Certification Mark (safety tested and production monitored)
| Type BF Applied Part
| Warning: Electricity
Information for Distributors and Healthcare Providers
Resetting the Pump
The pump remembers user settings and therefore it is important to reset the
pump to its original factory settings when providing the device to a new
patient. To reset the pump, refer to the directions under “To See Treatment
Compliance Meter (Hours of Use)” in the Operating Instructions section.
Authorized European Representative
Emergo Europe
Prinsessegracht 20
2514 AP, The Hague
The Netherlands
L-287 E EN 2022-02
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>