TOSHIBA SSA-550A Ultrasound Owner’s Manual

June 13, 2024
Toshiba

TOSHIBA SSA-550A Ultrasound

TOSHIBA-SSA-550A-Ultrasound-PRODUCT

Product Information

The product is called Nemio, with model number SSA-550A. It is a medical imaging device manufactured by Toshiba Medical Systems Co., Ltd. The product has various features and components, including a monitor, magneto-optical disk drive, floppy disk drive, transducer  cable hanger, peripheral unit rack, black-and-white printer,transducer connectors, handle, rear panel, power panel/main switch, VCR (optional), and color printer (optional).

Product Usage Instructions

Special Menu and Passwords

To access the special menu and passwords:

  1. Go to the User Function screen: [PRESET D..] -> Preset -> [Shift] and [password oimen] -> [OK]
  2. For the Service function main menu: [PRESET D..] -> Preset -> Maintenance -> [Shift] and [password oimeN] (last N capital letter) -> [OK]
  3. For the Special Configuration preset menu: [PRESET D..] -> Preset -> System Configuration -> [Shift] and [Preset] -> Exam preset -> [Condition Set] -> [Shift] and [hummer mark (Utility)] -> Configuration Utility -> SERVICE -> password pc-ws

Minimum System Check

To perform the minimum system check:

  1. Ensure the following PCBs are functioning: SYSC PWB, ECDC PWB, SUBC PWB, RPNL PWB
  2. Use the RPNL JIG (55400ARPJG) to display the image on a TV screen (order to SA-P if required)
  3. Note: The Power Supply, Panel (CTPN), and Monitor must be functioning properly for the minimum system check.

For more detailed information and instructions, please refer to the complete user manual.

Overview

TOSHIBA-SSA-550A-Ultrasound-FIG-1 \(1\) TOSHIBA-SSA-550A-Ultrasound-FIG-1 \(2\)

Special menu and Passwords

  • User Function screen
    [PRESET D..] -> Preset -> [Shift] and [<] -> password “oimen” -> [OK]

  • Service function “Main menu” ( see below photo ) [PRESET D..] -> Preset -> Maintenance -> [Shift] and [<] -> password “oimeN” (last “N” capital letter) -> [OK]

  • Special Configuration preset menu
    [PRESET D..] -> Preset -> System Configuration -> [Shift] and [<]

  • Special Preset menu
    [PRESET D..] -> Preset -> Exam preset -> [Condition Set] -> [Shift] and [<]

  • DMU “Service” menu
    [APPLI] -> hummer mark (Utility) -> Configuration Utility -> SERVICE -> password “pc-ws”

Minimum System check

The minimum system (with panel operation and overlay display)

  1. SYSC PWB
  2. ECDC PWB
  3. SUBC PWB
  4. RPNL PWB <when JIG is installed in Nemio back side ; left side on photo>
  5. + the following jig to display the image on TV screen. RPNL JIG : 55400ARPJG ( order to SA-P if required

Note : Power Supply, Panel (CTPN), and Monitor must be okay for the above minimum system check.

User function settings

How to operate

  • [PRESET D..] -> Preset -> [Shift] and [<] -> password “oimen” -> [OK]
  • Select the target switch from the Preset Key window and press the [SET] switch.
  • The Function List window is displayed.
  • Select the function to be set and press [OK]. Repeat these steps to assign the desired functions.
  • When [OK] on the Preset Key window is pressed, the settings are saved in the system.TOSHIBA-SSA-550A-Ultrasound-FIG-1 \(6\)

Panel Overview function ( available on & after V3.0)
It is possible to view panel key overview.TOSHIBA-SSA-550A-Ultrasound-FIG-1
\(7\)

IQ Report

Saving image on TV screen

  • Press [Shift] and [F1] keys
  • (IQ Report Error may occur depends on the system settings. but, IQ report image is stored even with the error message.)

Viewing IQ report image

  • Service function “Main menu”
  • [PRESET D..] -> Preset -> Maintenance -> [Shift] and [<] -> password “oimeN” (last “N” capital letter) -> [OK]
  • Maintenance -> Report -> IQ report -> Image Data Viewer. Press any key to exit.

Download the IQ report data into MO or FD (Nemio10)

  • Service function “Main menu”
  • [PRESET D..] -> Preset -> Maintenance -> [Shift] and [<] -> password “oimeN” (last “N” capital letter) -> [OK]
  • Insert blank & properly formatted MO disk into left MO drive. Write protect must be set to OFF.
  • Select Service function “Main menu” -> (in Short-Cut Menu) select “Backup / Restore” ->
  • check command “Backup” -> check the back-up item ;
  • Select (put check mark on) “IQ Report” at the bottom OR
  • select “All Packages” , Select Drive to MO ( or FD )
  • Press [Go], then system will ask to format MO disk, select OK to format and start back-up.
  • If MO disk format type does not match with Nemio MO, the system will display error message ; “Disk Error Check to see if disk is present”
  • Check MO disk format type (FLOPPY TYPE (Super Floppy Type) FAT16) according to the section 7 “MO Format” for the proper format type.
  • Smart back-up will also back-up all user data.
  • Set Write protect to ON for safety. And mention system identification ( such as customer and system serial number ) on MO disk label.. This is because one MO disk can store the back-up data only for one particular system.

SCSI printer settings ( CP-770DW)

  • Nemio (SSA-550A) plus Mitsubishi SCSI Color Printer CP770DW (CP770D)
  • The following you can find the procedures and check points for recommended SCSI color printer
  • CP770DW (CP770D) connection On Nemio SSA-550A.

Special Configuration preset settings ;

  • Select [PRESET D..] on the panel, select “Preset” on the full-down menu. Select “System Configuration”,
  • Press [SHIFT] plus [<]. The Special Configuration menu is now displayed.
  • Set “SCSI Color Printer” to ON.
  • Set “DMU Record Sel” according to the following table for the other Special Configuration settings if
  • System has DMU (Data Management Unit = UIDM-550A).
  • Reboot (turn off / on) the system to activate the Special Configuration preset setting change.TOSHIBA-SSA-550A-Ultrasound-FIG-1 \(8\)
  • ( “PRINT” in User Function is for B/W Printer.)

Set UF (User Function) switch accordingly.

  • (the attached pages 2/5 to 5/5 for the UF setting procedures.)

Confirm the following setting.

  • Below (5) : The Slide switchmust be set to OPTION (SCSI).

Use SCSI cable included in Nemio package.

  • Be careful for the SCSI cable length because SCSI cable length must be with-in 3 meters in total
  • This 3 meters also includes SCSI cablings inside.
  • SCSI printer, MO drives, Nemio and DMU operation will be unstable if SCSI cable length is longer
  • than the limited length = 3 meters.

SCSI terminate

  • Use SCSI terminator connector to terminate SCSI printer end.
  • SCSI printer, MO drives, Nemio and DMU operation will be unstable if no SCSI terminate.
  • Use “Active terminator” but do not use “Passive terminator”. Passive terminator may cause
  • unstable SCSI device operation and DMU operation such as some errors on MO

MO drive / format

Nemio (SSA-550A): Cautions for MO Disk format

  • On Nemio (SSA-550A), the system may not recognize MO Disk due to It’s formatting type. ( or if not yet formatted)
  • If MO disk format type does not match with Nemio MO, the system will display error message ;
  • “Disk Error Check to see if disk is present”
  • Check MO disk format type (FLOPPY TYPE (Super Floppy Type) FAT16).
  • There are the following notes mentioned in the SSA-550A Operation manual
  • Application volume 2B730-611E. ( below extracted from section 10.3 .)
  • Bitmap (BMP) or JPEG (JPG) data can be saved on the MOD. The preset “Image File Format” is used to set one of the two formats for saving.

CAUTION:

  • The lossy data compression technique is always applied when JPG images are recorded onto a Magneto-Optical disk.
  • Although this technique helps increase the number of images that can be stored, it can cause image deterioration.
  • The amount of compression, therefore, must be restricted so that the image quality is maintained at a level which does not adversely affect diagnosis.

CAUTION:

  • Do not display BMP and JPG data that have been edited or registered by other devices. Such data may be damaged.
  • Do not display BMP and JPG data that have been edited or registered by other devices. Such data may be damaged.

NOTE: Use MO disks that meet the following specifications:

  • Capacity: 640 MB/230 MB
  • Format: Windows format (Windows 2000 format is recommended)
  • Unformatted disks cannot be used.
  • Macintosh format disks cannot be used.
  • The supplied MO disks are not included in the warranty. With the factory setting, data is saved in a folder by the patient ID.

It is possible to change the factory setting to create a folder for each examination date. Contact your Toshiba service representative for the procedures to make this change.
Caution: On Nemio FILE UTILITY (can be registered on UF key), there is MO Format menu. However, this is only QUICK FORMAT and this is not PHYSICAL FORMAT function. So, if the MO Disk cannot be detected caused By it’s format style, Nemio MO Format function does not help. Because of this situation, Use only recommended type MO Disk, but Do not use non-recommended type MO Disks. Just in case if non-recommended type MO Disk must be used, follow The procedures shown on the next page.

Action: Format the MO Disk by “Formatter” available at internet.
There are several Formatters available. some are freeware. Below you can find the MO Formatter download site. http://mo.fujitsu.com/global/download/index2dlwin.html.TOSHIBA-SSA-550A-
Ultrasound-FIG-1 \(10\)

MO drive switch settings

( MO drive: Fujitsu MCE3064SS : Toshiba Part No. BSM31-3086 )
The following switch settingsTOSHIBA-SSA-550A-Ultrasound-FIG-1
\(11\)

In case if termination and SCSI ID’s are not set properly, the system operation maybe un-stable, such as SCSI errors or MO drive malfunctions. ( cannot access MO drive, etc..)

FD drive switch settings (Nemio 10)

( MO drive: Fujitsu MCE3064SS : Toshiba Part No. BSM31-3086 )

Note See Section 12 “Line Command” to activate FDD and MOD by some line commands. See Section 38 “Image format” for the image data can be saved into MO drive or FD drive.

Nemio Option Software installation

Refer to section 37 “SYSC PWB… ” for some notes about SYSC PWB replacement related with optional software !
Installation Procedure

  1. Turn ON the power supply of the SSA-550A.
  2. Remove write protection of the option key MO disk.
  3. Insert the option key MO disk into the right MO drive.
  4. Press [Preset D…] The Exam Select menu is displayed.
  5. Select [Preset] in the Exam Select menu. The Preset screen is displayed.
  6. Select [Maintenance] from the Preset screen. The “User Maintenance” menu is displayed.
  7. Press the [Shift] + [<] keys and enter the dedicated password “oimeN” on the password input screen..  After that, press [OK] . The menu for service engineer is displayed.
  8. Select [Short-cut Menu] [Software Install/Remove].
  9. Select the “Install” in Command.
  10. Select “MO” in Drive.
  11. Place a check in the check box for the option you will install in Option Software.
  12. Click [Go]. (Confirm that there is no error display.) Do not double click GO, it will cause error. Installation is completed when “Done!” is displayed to the right of the Progress status bar.
  13. Remove the option key disk from the MO drive and write-protect it. If key MO disk cannot be ejected out, exit (select Home ) from the service menu and try it again.
  14. Attach the kit label to SSA-550A.

Major Nemio options

Model name| Service part number

(only key MO disk)

| Function
---|---|---
USTH-550A| BSM31-3210-01| THI
USTH-551A (Nemio 10)| BSM31-3211-01| THI
USHI-550A| BSM31-3212-01| CHI / FEI
USHI-551A (Nemio 10)| BSM31-3789-01| CHI / FEI
USDA-550A| BSM31-3213-01| DCA
USTD-550A| BSM31-3214-01| TDI
USHP-550A| BSM31-3215-01| HPRF
USDF-550A| BSM31-3792-01| Dynamic Flow
USAC-550A| BSM31-3217-01| ACM
USTI-550A| BSM31-3216-01| TIC

  • The above service part number is only for trouble repair, but not for sale.
  • The above service part number does not include option case, manuals.
  • Refer to the systems data for the option software details for your sales purpose.

Note

  1. It is recommended to un-install optional software for the following cases. a) Replacing SYSC PWB, b) Re-install Nemio system software, c) System clear by ALL CLEAR function. SYSC PWB has the system serial No., if SYSC is replaced, all optional software will be automatically removed.
  2. The optional key MO disks can be used as anew key MO disks after un-install.
  3. If particular MO disk was already installed in another system, it will be not possible to un-install it in this system which has another serial number. So that it is very important to remember which MO disk was installed for which system (serial number).

DMU Option Software installation

Installing/Uninstalling the Optional Application Software (Install/Uninstall)
The optional application software can be installed or uninstalled using the Configuration Utility window.
<>

  1. Click [Utilities] in the toolbar of the main window and select Configuration Utility.
  2. Click [Service] and enter the utility password “pc-ws”. <>

Insert the optional application software installation FD in the drive and click [Install/Uninstall]. After installation, reboot the system. Once the MO is used for installation, it cannot be used in other UIDM-550A systems. The MO is designed for exclusive use by one UIDM-550A system, and it is required for uninstalling the installed optional application software. The MO is also required for reinstalling optional application software that is uninstalled. Therefore, the MO must be retained.TOSHIBA-SSA-550A-Ultrasound-FIG-1
\(13\)

<> Insert the MO that was used to install optional application software in the drive and click [Install/Uninstall] . After un-installation, reboot the system. The MO used for un-install is initialized and can be used for other UIDM-550A systems.

CAUTION:
Do not write any other files on the installation MO. Doing so may destroy the data. <> When [Installed Options] is clicked, the installed options are listed.

Note

  1. It is recommended to un-install optional software for the following cases.
    • a) Re-install DMU application software, d) Any other DMU troubleshooting.
  2. The optional key MO disks can be used as anew key MO disks after un-install.
  3. If particular MO disk was already installed in another system, it will be not possible to un-install it in this system which has another serial number. So that it is very important to remember which MO disk was installed for which system (serial number).

Major DMU options

Model name| Service part number

(only key MO disk)

| Function
---|---|---
UDPA-550A| BSM31-3231-01| Panoramic View
UDFE-550A| BSM31-3232-01| Fetal Face
UDTI-550A| BSM31-3245-01| TIC
UDAT-550A| BSM31-3247-01| ACT
UDTD-550A| BSM31-3248-01| TDI
UDAM-550A| BSM31-3249-01| ACM
UDSE-550A| BSM31-3250-01| Stress Echo
UDFR-550A| BSM31-3251-01| 3D
UDFR-551A| BSM31-3252-01| Fusion 3D
UDDS-550A| BSM31-3253-01| DICOM
UDTM-550A| BSM31-3254-01| Tele Medicine (Live View)
UDSR-550A| BSM31-3255-01| Summery Report

  • The above service part number is only for trouble repair, but not for sale.
  • The above service part number does not include option case, manuals.
  • Refer to the systems data for the option software details for your sales purpose.

Line Command

How to set Line Command

  • a) Press [PRESET D..], then select PRESET in the pull-down menu.
  • b) Select “System Configuration”. Then press [SHIFT] and [<].
  • c) Scroll down the menu, Find “Line Command” and set it to ON. Press OK when system ask you “Are you sure?”
  • d) System will ask you for power off. So, power off the system and power on the system.
  • e) Press [SET] key.
  • f) Press [Ctrl] and [Alt] and [D] keys simultaneously. There is no beep sound, but don’t worry, it is on Line Command. (Also note that the above is not [Ctrl] and [Alt] and [Del] keys, but [Ctrl] and [Alt] and [D] keys.) Now it is ready to input the line command. Below picture one example when input version display commandTOSHIBA-SSA-550A-Ultrasound-FIG-1 \(14\)
  • g) After accessing Line Commands, please set Line Command to OFF and power off / on the system. It is not very sure, but it is better set Line Command off ( and re-boot the system) for stable system operation.
  Command Purpose
1 optionchange1 Preparation to enter FDD$ON and MOD$ON commands.
2 FDD$ON Make FDD enable.
3 MOD$ON Make MOD enable.
4 version#= Version and TV system checking. If only for Version, or you may

use HW/SW
5| tveia$on| Set screen format to NTSC.
6| tvpal$on| Set screen format to PAL.
7| usaop$on| Only for USA (TAMS). It is to set acoustic power to 80% from max.
8| usaop$off| Except for USA (TAMS). It is to set acoustic power 100%.
9| txbfispmo flttbl5| To add new probe PIT data into TXBF via MO.
10| txbfispfd fittbl5| To add new probe PIT data into TXBF via FD.
11| SetSystemConfig=| Preparation to enable FDD and MOD on Nemio10 only.

When Line Command required ?

: System : Nemio30 with DMU installed. SYSC PWB was replaced for some kind of troubleshooting. After replacing SYSC PWB, DMU screen (APPLI) menu is shifted to upwards. **< Action required>** SYSC PWB in the service parts stock is usually set to NTSC TV system. If DMU has PAL TV settings, SYSC PWB (Nemio TV system) will not match to the TV system settings on DMU. So that only DMU screen has vertically shifted image. So, set TV system by Line Command function shown below. Note that Nemio Observation Monitor is PAL/NTSC universal type. So that it will be no problem on Monitor image.
  1. Select [Preset] in the Exam Select menu. The Preset screen is displayed.
  2. Select [System Configuration] from the Preset screen.
  3. Press the [Shift] and [ < ] keys. The Special Configuration screen is displayed.
  4. Set the “Line Command” to On and press [OK].
  5. When the message “Are you sure” is displayed, press [OK].
  6. When the message “Please Power OFF” is displayed, reboot the system.
  7. Press [SET] switch.
  8. Press the [Ctrl]+[Alt]+[D] keys simultaneously.
  9. Enter “version#=” using the keyboard and press [ENTER].
  10. The system software version and TV system will be displayed for a while. Check if TV system is set wrongly.
  11. Enter “tvpal$on” using the keyboard and press [ENTER] to set the Nemio TV system to PAL. “tveia$on” for NTSC.
  12. Set the Line Command to OFF then, Turn off / on the system, and check the problem. Note This will be the same problem if DMU side TV setting is wrong. ( DMU TV setting is on DMU service menu.)
: System : Nemio10 only with FDD installed. After replacing SYSC PWB, it is not possible to access Floppy Disk Drive. There is no FDD Even in File Utility menu (System does not recognize FDD.)

< Action required>
SYSC PWB in the service parts stock is usually set for Nemio30 and Nemio20 which has only MO drive. To activate FDD, set the following Line Commands.

  1. Step -1 to -8 are the same as above.
  2. Enter “SetSystemConfig=L” using the keyboard and press [ENTER].
  3. Enter “optionchange1” using the keyboard and press [ENTER].
  4. Enter “FDD$ON” using the keyboard and press [ENTER].
  5. Set the Line Command to OFF then, Turn off / on the system, and check the problem.
: You borrow SYSC PWB from Nemio 10 temporary for service checking purpose. System does not recognize MO drive.

< Action required>

  1. Step -1 to -8 are the same as above.
  2. Enter “optionchange1” using the keyboard and press [ENTER].
  3. Enter “FDD$ON” using the keyboard and press [ENTER].
  4. Enter “MOD$OFF” using the keyboard and press [ENTER].
  5. Set the Line Command to OFF then, Turn off / on the system, and check the problem.

Nemio Data Back-up & Restore

BACK-UP Operation procedures

  • Service function “Main menu”
  • [PRESET D..] -> Preset -> Maintenance -> [Shift] and [<] -> password “oimeN” (last “N” capital letter) -> [OK]
  • Insert blank & properly formatted MO disk into left MO drive. Write protect must be set to OFF.
  • Select Service function “Main menu” -> (in Short-Cut Menu) select “Backup / Restore” -> check command “Backup” -> check the back-up item ;
  • Select (put check mark on) “IQ Report” at the bottom OR
  • select “All Packages” , Select Drive to MO ( or FD )
  • Press [Go], then system will ask to format MO disk, select OK to format and start back-up.
  • If MO disk format type does not match with Nemio MO, the system will display error message ;
  • “Disk Error Check to see if disk is present”
  • Check MO disk format type (FLOPPY TYPE (Super Floppy Type) FAT16) according to the section 7 “MO Format” for the proper format type.
  • Smart back-up will also back-up all user data.
  • Set Write protect to ON for safety. And mention system identification ( such as customer and system serial number ) on MO disk label.. This is because one MO disk can store the back-up data only for one particular system.

RESTORE Operation procedures

  • Insert MO disk which has back-up data stored.
  • [PRESET D..] -> Preset -> Maintenance -> [Shift] and [<] -> password “oimeN” (last “N” capital letter) -> [OK]
  • Select Service function “Main menu” -> Backup / Restore -> check command “Restore” -> check the back-up item
  • select “All Packages. Select Drive to MO ( or FD )
  • Press [Go], then system will ask to format MO disk, select OK to format and start back-up.

DMU Data Back-up & Restore

Data backup

  1. Using the backup function of the application software
    • Place the test MO in the MO drive and copy the files to the MO.
    • It is possible to use the copy function of the application software installed in the system.
    • Note that if no problems are found in operational checks after finished, it is not necessary to restore the backup files.
    • Data backup procedures – Using the copy function of the application software. Click the Patient List on the toolbar.
    • Select the study to be saved and click the [Send] button. Select “MOD Drive”.
  2. Using Windows Explorer
    • If the system application cannot be started up, use Windows Explorer. In this case, also back up the following file.
    • The backup file must be restored together with the data.
    • C:\ Dicomit\db\tpatient.mdb
    • Note that if no problems are found in operational checks after finished, it is not necessary to restore the backup files.
    • Data backup procedures – Using Windows Explorer. Launch Windows Explorer.
    • Copy all of the files in the following directories to the MO.
    • D:\ Images D:\Advanced D:\Measure Copy the following file to the MO. C:\Dicomit\db\tpatient.mdb

System configuration backup
Perform the following procedures to back up the settings of the system. Backup of the system settings. Click [Utilities] on the toolbar of the main menu and select “Configuration Utility”. Click [Service] and enter the utility password [pc-ws]. Place a formatted MO into the MO drive and click [Backup]. If the system application cannot be started up, use Windows Explorer to copy the following files to the MO.

  • c:\dicomita\dicomit.cfg
  • c:\dicomit\ini\ToFile.cfg
  • c:\dicomit\ini\Toshiba.cfg
  • c:\dicomit\dtk\dtk32.mdb
  • c:\winnt\system32\drivers\etc\Lmhosts
  • c:\tomtec\*.ecp (all files with extension “ecp”)
  • c:\tomtec\*.dct (all files with extension “dct”)
  • Note that if no problems are found in operational checks after finished, it is not necessary to restore the backup files.

RESTORE

  1. Restoring the system. Click [Utilities] in the toolbar of the main window, and select Configuration Utility. Click [Service] and enter the utility password “pc-ws” .Insert the backup MO disk in the MO drive and click [Restore]. Select the folder corresponding to “AE Title” of the PC (example: “AE Title_TOSHIBAxxxxxxx”). Click [OK].

Restoring the DB (database)

  1. Using the application function
    • When backup was performed using the application function, follow the procedures below to restore the database.
    • Restoring the database – Using the application function. Click the Patient List button on the toolbar.
    • Select “MO Images” in the “Source” selection column. Select the study to be restored and click the [Retrieve] button.
  2. Using Windows Explorer
    •   * When Windows Explorer was used for backup because the system application could not be started up, follow the procedures below to restore the database.
    • Restoring the database – Using Windows Explorer. Launch Windows Explorer. Copy all of the files from the
    • MO to the following directories.
    • D:\Images D:\Advanced D:\Measure
    • Copy the “tpatient.mdb” file from the MO to the following directory.
    • C:\Dicomit\db
    • After restoration is completed, start up the application software.
    • Restoring the database – Using Windows Explorer. Launch Windows Explorer.
    • Copy all of the files from the MO to the following directories.
    • D:\Images D:\Advanced D:\Measure
    • Copy the “tpatient.mdb” file from the MO to the following directory
    • C:\Dicomit\db
      After restoration is completed, start up the application software.

Nemio system software re-installation

Back-up User data and un-install Nemio option software before you do the following. See the section 10 Option software installation and section 13 Data back-up and restore.

SOFTWARE INSTALLATION

  1. Insert the System Disk in the FD drive.
  2. Turn ON the power and hold down the FREEZE switch.
  3. When the initial screen is displayed and the message “Detect FREEZE key” is displayed at the bottom of the screen, release the FREEZE switch.
  4. The 2D screen is displayed when system software installation is completed. Misalignment between the 2D image and characters may occur after 2D image display. This can be solved by rebooting the system. If the problem persists even after the system is rebooted, go to step (17).
  5. Remove the System Disk from the FD drive.
  6. Reboot the system.
  7. Press [Preset D..] . The Exam Select menu is displayed.
  8. Select [Preset] in the Exam Select menu. The Preset screen is displayed.
  9. Select [System Configuration] from the Preset screen. The ” System Configuration ” screen is displayed.
  10. Press the keys. The Special Configuration screen is displayed.
  11. Set the “Line Command” to On and press [OK].
  12. When the message “Are you sure” is displayed, press [OK].
  13. When the message “Please Power OFF” is displayed, reboot the system.
  14. Press the [SET] switch.
  15. Press the [Ctrl]+[Alt]+[D] keys simultaneously.
  16. When this kit is to be installed in the USA, enter “usaop$on” using the keyboard and press [Enter] key. When this kit is to be installed in countries other than the USA, enter “usaop$off” and press [Enter] key.
  17. If the TV format remains abnormal even after the system is rebooted, follow the steps below.
    • (a) For the NTSC TV format, enter “tveia$on” and press [Enter] key.
    • (b) For the PAL TV format, enter “tvpal$on” and press [Enter] key.
    • (c) Reboot the system.
  18. Turn OFF the power and disconnect all transducers.
  19. Turn ON the power.
  20. When the 2D screen is displayed on the system monitor, insert the System Maintenance Disk in the FD drive.
  21. Make acoustic power minimum and press [SET] switch.
  22. Press the [Ctrl]+[Alt]+[D] keys simultaneously.
  23. Enter ” txbfispfd flttbl7″ (the characters other than (7) in this command are alphabetical characters) using the keyboard and press [Enter] key.
  24. Following comment is displayed in about one minute. ISP terminated with error 0 Error zero means successfully terminated.
  25. Remove the System Maintenance Disk from the FD drive.
  26. Set the “Line Command” in the Special Configuration screen to Off and press [OK].
  27. When the message “Are you sure” is displayed, press [OK].
  28. When the message “Please Power OFF” is displayed, turn OFF the power.
  29. Connect transducers.

DATA RECOVERY
Confirm that the data items that were recorded in section 2 ” DATA RECORDING” have not changed. If some settings are to be changed or appear to be incorrect, change or correct them.
CONVERSION OF USER-REGISTERED MEASUREMENT DATA

  1. Start the system and press PRESET D. . The Exam Select menu is displayed.
  2. Select [Preset] from the Exam Select menu. The Preset screen is displayed.
  3. Select [Measure Registration] on the Preset screen. The Measure Registration screen is displayed.
  4. If user-registered measurement data exists, an error message is displayed. Press [OK] several times until the error message screen disappears.
    • On this screen, “Error” is displayed in place of the user-registered measurement name.
  5. Press the [Shift] and [<] keys. The [Convert] switch is displayed.
    • If the [Convert] switch is grayed out, the following steps need not be performed. Press [Exit] to close the Measure Registration screen.
  6. Press the [Convert] switch., press [OK].
  7. When conversion is completed,. Press [OK].
  8. Confirm that the measurement names that were displayed as “Error” are now displayed correctly.
  9. Press [Exit] to close the Measure Registration screen.

DMU Application software re-installation

For Version is V1.1 or later version

  • Remove all probes. Remove front cover – only one screw at the upper part of front cover. ( below two points are fixed by “magic tape” tightly ; just pull it.)
  • Find mouse and keyboard connectors. Remove internal ones originally connected and connect external mouse and keyboard supplied with the system.
  • Turn on the system. Select APPLI switch on Nemio to display DMU screen after boot-up. Set Write protect for MO disk.
  • Insert MO disk )(or CD-ROM) into right side MO disk. (left side is also okay, but right one is basically for DMU)
  • Select Utility in tool bar – Configuration – (password : pc-ws) – select service – EXIT TO OS. ( or Enter [Ctrl]+[Alt]+[Delete] to start the task manager and terminate the TDICOM10 application)
  • Start Windows Explorer, select (CD-ROM or) Removable Disk [E:] (means MO) , and double-click .
  • You can find “UpV12” directory (or UpV11). open this directory and find batch file named ;
  • UpV11.bat ( or UpV12.bat). Double click this batch file.
  • Installation will be started.
  • During installation, there will be the message on page 6/6 may appear : saying
  • Windows 2000 hotfix setup… [OK] This is the message regarding red-code virus.
  • Select OK to continue. Then proceed further software Installation according to the messages.
  • The installation will be finished with-in a few seconds.
  • Remove the system software disk from the drive. Select DMU application “Nemio” from the start bar.
  • After DMU application starts, de-select APPLI switch. Turn off the system. Restart the SSA-550A

For Version V1.0 DMU( will be very rare case)

  • Remove all probes. Remove front cover – only one screw at the upper part of front cover. ( below two points are fixed by “magic tape” tightly ; just pull it.)
  • Find mouse and keyboard connectors. Remove internal ones originally connected and connect external mouse and keyboard supplied with the system.
  • Turn on the system.
  • Select APPLI switch on Nemio to display DMU screen after boot-up.
  • Set Write protect for MO disk.
  • Insert MO disk )(or CD-ROM) into right side MO disk. (left side is also okay, but right one is basically for DMU)
  • Select Utility in tool bar – Configuration – (password : pc-ws) – select service – EXIT TO OS. ( or Enter [Ctrl]+[Alt]+[Delete] to start the task manager and terminate the TDICOM10 application)
  • Start Windows Explorer, select (CD-ROM or) Removable Disk [E:] (means MO) , and double-click .
  • You can find “APP” directory. open APP directory and select all files – copy.. then select C drive, (right click) COPY. ( This means copy all the files / directory under “APP” directory in MO disk to DMU “C” drive.)
  • Overwrite all the files and directory.
  • Remove the system software disk from the drive.
  • Select DMU application “Nemio” from the start bar.
  • After DMU application starts, de-select APPLI switch. Turn off the system.
  • Restart the SSA-550A

DMU Network settings

Network Settings for UIDM-550A
The UIDM-550A Workstation, or DMU, is a Windows 2000 based system, so the method for making the network settings is slightly different than that used for the PowerView, which is a Windows NT based system.
The following is offered as an aid in navigating the newer system.

  1. To get to the Network Settings area, use the same steps as those used on the PowerView. Click on the Hammer on the toolbar, select Configuration, and click on the Service button and enter password pc-ws.
  2. This will bring you to the Service Configuration screen. Click on the Network button as before.
  3. You will now see the Windows 2000 Network Connection screen.
  4. Click on Advanced on the menu bar, and select Network Identification from the drop down menu.
  5. You will see the System Properties screen. Click on Properties to rename the DMU to your desired name.
  6. On the Identification Changes screen, type in the name selected for your DMU.
  7. Click on OK for the name change, and OK for the System Properties.
  8. That will bring you back to the Network Connections screen. Double click on Local Area Connection, below Make New Connection.
  9. That will bring up the Local Area Connection Properties screen. Select Internet Protocol (TCP/IP) and then click on Properties.
  10. That will bring up the Internet Protocol (TCP/IP) Properties screen where you can enter your assigned IP Address.
  11. After entering the IP address, click on OK on this screen, and OK on the next screen. You will get a message. asking if you want to restart now. Click on Yes to accept the restart.
  12. After the DMU restarts, you can continue with the configuration the same as you do with the PowerView systems; DICOM setup, Install Options, etc.

Note If the system start asking you for LOG-ON password after changing some Network settings, check the next section 18 “DMU LOG-ON password” to remove the LOG-ON message.

DMU LOG-ON password Nemio DMU (UIDM-550A) log-on password
[ TROUBLE ]
DMU ask you for log-on password during boot-up.
[ OUTLINE ]
For DMU password, this password must be “system”. We suppose the registry file was changed by some improper re-works for network settings.
[ SOLUTION ]
The following you can find the solution. This work is to setup Windows auto log-on with default password. So that after trying the following procedures “1)” to “8)”, Windows will not asking you for log-on but it will be automatically log-on.
BE CAREFUL DO NOT ACCESS ANY OTHER REGISTRY FILES EXCEPT THE ONE WE MENTIONED BELOW !! Connect keyboard and mouse to the DMU. Switch on.

  1. logon with the “system” password.
  2. With the “Window” key on your keyboard go to the “start bar” and select Start –>
  3. RUN…. type “regedit”
    • In the path : Mycomputer \ HKEY_LOCAL_MACHINE \ software \ Microsoft \ windowsNT \
    • CurrentVersion \ WinLogOn
    • You will see that the “AutoAdminLogon” is set to “0”
  4. Double click on “AutoAdminLogon” and change the value to “1”
  5. Then with the mouse go to Edit –> New –> String Value
  6. Type “DefaultPassword”
    • Right click on it and select “modify”
    • Type as value “system”
    • Now exit and save the new settings.
  7. Shutdown and disconnect your keyboard and mouse and connect the panel and trackball again.
  8. Switch back on and the unit is running without asking for Log-on message.

Nemio Reset (without turn off / on)

Nemio can be restarted instead of Power off / on by the following.
Press the combination of ; FREEZE & SELECT (NEXT) & NEW PATIENT switches.

DMU Reset (without turn off / on)

Preparation
External Keyboard External Mouse: Remove rear cover to see DMU PC inside to access PC power switch afterwards.Remove front cover by removing only one screw. ( bottom of front cover is fixed by “magic tape” strongly.) Refer to service manual for those mechanical dis-assembly procedures. At the system front side, you can find DMU PC rear side. So that remove standard Keyboard and Mouse connectors and connect External Keyboard and External Muse to the proper connectors on DMU PC rear. Also prepare MO disk.

Operation procedures-1

  • Turn on Nemio.
  • After boot-up, select [APPLI] switch.
  • Now DMU can be operated by External Keyboard and External mouse.
  • Select hummer mark on TOOL BAR. Select Configuration Utility.
  • Select [SERVICE] then key-in “pc-ws” as a password to enter Service menu.
  • If optional software were installed, un-install those options by INSTALL/UNINSTALL.
  • For the procedures for un-installation, you of course need the original optional software key MO (or FD).
  • Un-install all optional software accordingly.
  • If any important patient data in HDD, we also recommend you back-up those data into MO disks.
  • Those patient data will be not deleted by the following process, but as usual ( generally speaking ), it is better back-up those user data into MO disks for safety back-up.

Select [EXIT TO OS]

  • The message will appear. Select [OK] to exit to the Windows OS.
  • Then move the courser to START, you can find the DMU application software “Nemio” at the top of START bar,
  • Select “Nemio” in the START bar, then Nemio DMU application will be re-started.

Operation procedures-2

  • If the above, you exit to Windows OS via service menu. But you can also exit to OS by simple way.
  • After you go to APPLI, press the keys : [Ctrl] + [Alt] + [Del], then Windows Security screen will appear.
  • Select “Task Manager”, Task Manager screen will appear. Select “End Task” to end all the activating Tasks.
  • After ending all those Tasks, it will be Windows screen. Then the procedures will be the same as the above.

Total Blockdiagram

Genral Blockdiagram
TOSHIBA-SSA-550A-Ultrasound-FIG-1 \(15\)

Power Supply Section TOSHIBA-SSA-550A-Ultrasound-FIG-1
\(16\)

“Power Monitor” function is available in Service menu – Test Tools – Diagnostics [PRESET D..] -> Preset -> Maintenance -> [Shift] and [<] -> password “oimeN” (last “N” capital letter) -> [OK] Service menu items : For the details, refer to 2D730-1511E Service manual [Maintenance Menu Volume]