Insulin Pump System Overview
The Personal Diabetes Manager (PDM)
- Power button
- Home Key (Software Key)
- Charging Port
- Sound Hole
- Indicator light
Pump Base
Reservoir Patch
- Your ultra-thin Patch Pump is composed of a reusable Pump Base and a disposable Reservoir Patch which stores up to 200 U of insulin and lasts for up to 3 days.
- The Pump Base stores delivery settings and sends data to the PDM via Bluetooth Low Energy.
- The Patch Pump (Base and Patch connected) is waterproof at 8 feet (2.5 meters) for 60 minutes (IPX8).
- Store the Pump Base and Reservoir Patch at temperatures between -10°C (14°F) and 55°C(131°F).
Charge PDM
The PDM requires an AC adapter with an output of DC 5.0V
PDM Settings
-
Turn on/off the PDM
• Turn-on – Long-press the power button, a green light will flash.
• Turn-off – Long-press the power button for about 2 seconds, then slide to power off. Or long-press the power button for about 6 seconds -
Wake the PDM
When the screen goes black, press the power button and then slide to unlock -
Select language and country
-
Set time and date
Note: You can only change the date and time when there is no active Reservoir Patch. -
Bolus Calculator
Refer to Page 5 -
Set audio options
Main Menu ➔ Settings ➔ General ➔ Audio Options
If you set the audio option to Audio off /Vibrate off, your PDM only vibrates when a serious Alarm occurs. -
Set display settings
Main Menu ➔ Settings ➔ General ➔ Display
It is recommended for new users to set backlight time as 2 minutes -
Find Pump Base SN
You can find the Pump Base SN on the product box or on the back of the Pump Base. -
Enter Pump Base SN
Main Menu ➔ Patch ➔ Pump Base SN
Tap —— to enter SN manually
Basal Settings
Main Menu ➔ Settings ➔ Insulin Pump ➔ Basal Setup
Note: Recheck all the numbers you entered to make sure that they are the
intended values.
1. Max Basal
Tap the Max Basal rate to edit it
2. Edit Basal Pattern
-
Tap Edit Basal, then select the Pattern to edit it
-
Set basal rate
• Tap blue U/H field to set the Basal rate.
• You can set a basal rate between 0 and the
Max Basal Rate with an increment of 0.05 U/H. -
Add Time Segment
• Tap +Add Time Segment to add a new segment.
• You can set up to 48 segments for each day. -
Delete Time Segment
• Slide to left on each segment, tap Delete. The start time of next segment will be adjusted -
Scroll up and down to review all segments
-
Tap to save the settings
3. Selecting a Basal Pattern
Main Menu ➔ Basal ➔ Select Pattern
Main Menu ➔ Settings➔ Insulin Pump ➔ Bolus Setup
-
Max Bolus
Tap the Max Bolus to edit it -
Set up the Bolus Calculator
• Turn on/off the Bolus Calculator
• Set IC ratios
• Set Insulin Sensitivity
• Set BG target
• Set IOB Time (Active Insulin Time) -
Normal, Extended and Combo Bolus
Start a new Patch
-
Select the infusion site
• Clean, less hair, no sweat, uncovered, enough fat (at least 5mm thick), away from navel.
• A new infusion site should be at least 2.5 cm (1 inch) away from the last site.
• If you choose an infusion site on your abdomen, back or buttocks, apply the Patch horizontally.
• If you choose an infusion site on your upper arm or thigh, apply the Patch vertically. -
Prepare the infusion site
• Wash your hands with soap and water.
• Clean the infusion site with an alcohol wipe to enhance adhesion.
• Let the skin air dry completely. -
Connect Pump Base and Reservoir Patch
Main Menu ➔ Patch ➔ New Patch Follow the on-screen instructions
-
Fill the New Reservoir Patch
With the help of your healthcare provider, decide on the amount of insulin (70 U – 200 U) you need to insert into the Reservoir Patch.
Warning: Do not inject air into the fill port. Doing so may result in unintended or interrupted insulin delivery.
-
Prime the Pump
-
Place the Patch
1. Remove the button cover| 2. Peel off the adhesive liners| 3. Press the Pump against the skin| 4. Press the needle button with one quick motion to completely insert the needle
---|---|---|---
-
Start Insulin Delivery
Read Pump
Home Screen
- Alert/Alarm
- PDM Battery
- Easy Loop Icon
- Insulin Delivering Status
- IOB
- Time
- Pump RF Signal
- Audio and Vibration
Pump Status
Slide on the Home Screen from left to right to go to status screen
Main Menu
Slide on the Home Screen from right to left to go to main menu History
Main Menu ➔ History
Delivery Status
Ieon | Shape and Color | Description |
---|---|---|
Grey ring | Grey ring indicates that there is no activated Pump. | |
Green ring | Green ring represents the basal delivery. | |
Green and dark green ring | Green ring with dark green part represents Temp |
Basal. the dark green part indicates the progress of Temp Basal delivery.
| Blue and dark blue ring| Blue ring represents the Normal Bolus. the dark
blue part indicates the progress of delivery.
| Purple and dark purple ring| Purple ring represents the extend Bolus. the
deep purple part indicates the progress of delivery.
| Red ring| Red ring represents the delivery suspend status.
Temp Basal
Activate a Temp Basal
Main Menu ➔ Basal ➔ Temp Basal
With a Temp Basal you can adjust your Basal rate for a short time Select a
Basal type, rate or percent and duration
Cancel a Temp Basal
Main Menu ➔ Basal ➔ Cancel Temp Basal
Set Preset Temp Basal
Main Menu ➔ Settings➔ Insulin Pump ➔ Basal Setup ➔ Preset Temp Setup
Activate a Preset Temp Basal
Main Menu ➔ Basal ➔ Preset Temp Basal
Manual Bolus
1 Main Menu ➔ Bolus ➔ Manual bolus 2 Set Bolus dose 3 Set Bolus type 4-1
Normal Bolus
Slide to deliver 4-2 Extended Bolus
Set duration
Slide to deliver 4-3 Combo Bolus
Tap blue/purple plus sign to set proportion of Normal part and
Extended part Set duration Slide to deliver
Food Bolus Calculator
1 Main Menu ➔ Bolus ➔ Bolus Calculator 2 Enter BG or Carb, or both 3 Set Bolus
dose and Bolus type 4-1 Normal Bolus
Slide to deliver 4-2 Extended Bolus
Set duration
Slide to deliver 4-3 Combo Bolus
Tap blue/purple plus sign to set proportion of Normal part and
Extended part Set duration
Slide to deliver
Deactivate the Patch
1 Deactivate the Patch
Menu➔ Patch➔ Deactivate Patch
2 Remove the Patch
- Press and hold the needle release button on the side of the Reservoir Patch, and slide it in the direction of the arrow to retract the needle.
- Gently remove the entire Patch Pump. Use a medical adhesive remover or baby oil if necessary.
- Bend the adhesive pad and peel off the double-sided tape.
- Press and hold the two release buttons on opposite sides of the Pump Base, and pull the Pump Base outward to remove it from the Reservoir Patch.
Alerts & Emergency Kit
Alert
PDM Message | Actions to Take |
---|---|
END OF SUSPEND | Check BG. Resume basal delivery if necessary. |
LOW RESERVOIR | Change Patch soon. |
PATCH EXP IN 1 HOUR | Change Patch soon. |
PATCH BATTERY LOW | Change Reservoir Patch soon. |
Alarm (Medium Priority)
PDM Message | Actions to Take |
---|---|
PATCH EXPIRED | Change Patch. |
Check blood glucose.
EMPTY RESERVOIR| Change Patch.
Check blood glucose.
EXCEEDS MAX TDD| Check BG. Resume basal delivery.
Check bolus history and reevaluate your need for insulin.
Alarm (High Priority)
PDM Message | Actions to Take |
---|---|
OCCLUSION DETECTED | Change Patch. |
Check blood glucose.
PATCH ERROR| Change Patch.
Check blood glucose.
PUMP BASE ERROR| Remove Pump.
Contact customer support
Check blood glucose
Emergency Kit
Keep an emergency kit with you at all times to make sure you have necessary
supplies. Inform a family member, co-worker, and/or friend where this
emergency kit is kept.
This kit should include but is not limited to:
- Fast-acting glucose tablets or gel
- BG monitoring supplies
- Urine ketone testing supplies
- Insulin syringe
- Rapid-acting U-100 insulin
- Extra Medtrum 2.0 mL Reservoir Patches
- Power Bank
- Instructions from your healthcare provider about how much insulin to inject if pump delivery is interrupted
- Alcohol wipes
- Glucagon emergency kit
- Emergency contact phone numbers
- If traveling, a copy of a letter from your healthcare provider for airline security
Your Personal Settings
Max Basal Rate: ____ U/H
Basal Pattern Name: __
Start Time| Rate (U/H)| Start Time| Rate (U/H)
:| | :|
:| | :|
:| | :|
:| | :|
:| | :|
:| | :|
:| | :|
:| | :|
Name| Rate| Duration
---|---|---
Heavy Exercise| __ % or U/H| :
Medium Exercise| __ % or U/H| :
Light Exercise| __ % or U/H| :
Sick| __ % or U/H| :
Temp 1| __ % or U/H| :
Temp 2| __ % or U/H| :
Temp 3| __ % or U/H| :
I:C Ratio
Start Time| I:C (g Carb)
:| 1 U| : g
:| 1 U| : g
:| 1 U| : g
:| 1 U| : g
:| 1 U| : g
:| 1 U| : g
:| 1 U| : g
:| 1 U| : g
BG Target (mmol/L)
Start Time| Low Limit| High Limit
:| |
:| |
:| |
:| |
:| |
:| |
:| |
:| |
Insulin Sensitivity
Start Time| ISF (mmol/L)
:| 1 U| : mmol/L
:| 1 U| : mmol/L
:| 1 U| : mmol/L
:| 1 U| : mmol/L
:| 1 U| : mmol/L
:| 1 U| : mmol/L
:| 1 U| : mmol/L
:| 1 U| : mmol/L
Max Bolus: ____ U
Active Insulin Time: : .
Medtrum
| |
---|---|---
Medtrum Technologies Inc.
7F, Building 8, No. 200, Niudun Road
Shanghai 201203, China
Tel: +86-21-50274781
Fax: +86-21-50274779
www.medtrum.com| Medtrum B.V.
Nijverheidsweg 17
5683 CJ Best
The Netherlands
Tel: +31 (0) 499745037| This product complies with
Directive 93/42/EEC (MDD) and
Directive 2014/53/EU (RED).
MD-SY-011C
IM882111WW-002
348367
Version:1.4