PST GPR-1200 GPR Portable Oxygen Analyzer User Guide

June 12, 2024
PST

PST logo Portable Oxygen Analyzer Quick Start Guide
First span calibration and measurement
PST-QSG-3203-01PST GPR 1200 GPR Portable Oxygen
Analyzer

ProcessSensing.com
06/2023

GPR-1200 GPR Portable Oxygen Analyzer

Welcome to the Quick Start Guide for first span calibration and first measurement using your ppm portable analyzer.
Here, you will find information covering gas connection in section A, and first span calibration in section B.
Please read the safety information below.

Start here

Safety information

  • Avoid covering the vent for the test flow indicator when gas is flowing to the sensor. This can pressurize the sensor causing damage.
  • If your analyzer has a pump, open the metering valve completely before switching on the pump.
  • To remove moisture and particulates, open the sensor housing and either blow on the sensing surface or gently wipe the surface with a damp cloth. Ensure ppm sensors have minimal exposure to air.
  • Only charge your analyzer in a safe (non-Ex) area, and do not charge for more than 24 hours.
  • You must connect the analog signal output to a recording device in accordance with local safety directives. Ensure the recording device does not generate a voltage greater than 12 V DC.

The first calibration is of utmost importance as all subsequent calibrations are based on the initial one.
NOTE: Your portable analyzer is supplied with an oxygen sensor installed.
We recommend you use certified span gas for calibration to ensure the best measurement readings.
The GPR-series of portable oxygen analyzers is compliant with the following safety approvals and directives:PST GPR 1200 GPR Portable Oxygen Analyzer -
portable

A unit with a blue display outline is for general purpose, red is for hazardous area, as shown above.

User Interface (UI)

Button

| Function
---|---
| On/Off
| Menu
| Enter
| Previous (decrement)
| Next (increment)

Gas connection

NOTE: The calibration span gas should be 50…80% of the required measurement range or one range above. E.g. For a 1 ppm measurement, span gas should be 5…8 ppm or 50…80 ppm.

GPR-1200 / GPR-3500 MOVR
The GPR-1200 and GPR-3500 MOVR portable analyzers are equipped with a bypass valve. Always ensure it is in the Bypass position before connecting to gas.

  1. With the bypass valve in the Bypass position, connect your span gas line to the Sample In port, using the fitting provided.
  2. Set the flow rate to 1…2 SCFH (refer to Figure 1 on page 4), and allow the span gas to flow for 2…3 minutes to purge the bypass valve.
  3. Now turn the bypass valve to the Sample position and continue to section B.

GPR-1100
The GPR-1100 is equipped with quick disconnect fittings. Please ensure the system is not pressurized before connecting your span gas.

  1. Set your flow rate to 1…2 SCFH before connecting gas.
  2. Connect fitting to one of the ports. Ensure gas is venting to the atmosphere.
  3. Start the flow of your span gas, then connect your process gas line to the other port and allow gas to flow for 2…3 minutes. This will purge the system. Continue to section B.

GPR-2000

  1. Set your flow rate to 1…2 SCFH before connecting gas.
  2. Connect your span gas line to the port, using the fitting provided.
  3. Start the flow of gas and allow it to flow for 2…3 minutes. Continue to section B.

First span calibration

  1. After connecting the calibration span gas, press on your analyzer flow gas for 10…15 minutes to allow the sensor to stabilize near the span gas value.

  2. Now press and use and to navigate to Calibration > Span Calibrate.

  3. Use and to enter the span gas value. Ensure the reading has stabilized before continuing.
    NOTE: When a Span or Zero Cal starts, only “Abort” with is shown until the reading is stable, then “Accept” with appears.

  4. Use to Accept, and to Abort. Your portable analyzer is now ready to use.

Gas disconnection

GPR-1200 / GPR-3500 MOVR

  1. Turn the bypass valve to the Bypass position, then stop the flow of gas.
  2. Disconnect your span gas lines from the Sample In port.

GPR-1100

  1. Disconnect your span gas line using the quick disconnect fitting.
  2. Stop the flow of gas.
  3. Now disconnect the gas line venting to the atmosphere using the quick disconnect fitting.

GPR-2000

  1. Stop the flow of gas.
  2. Disconnect your gas line by loosening the fitting.

Making your first measurement

To make your first measurement, connect your process gas lines by following the procedure relevant to your analyzer model in section A.
Figures

PST GPR 1200 GPR Portable Oxygen Analyzer - Figures

Useful links
Scan below for more information.

https://www.processsensing.com/en-us/oxygen/portable-analyzers//

PST logoProcessSensing.com

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

PST User Manuals

Related Manuals