ZEAGLE
DEFENDER EBS
EMERGENCY BREATHING SYSTEM
USER MANUAL
Huish Outdoors Part #
PN 350-9200 ZEAGLE DEFENDER EBS 207 BAR SYSTEM
PN 350-9210 ZEAGLE DEFENDER EBS 207 BAR SYSTEM w/ ON/OFF
COPYRIGHT/TRADEMARK NOTICES
COPYRIGHT NOTICE
This user manual is copyrighted, all rights are reserved. This material may
not be reproduced, displayed, modified or distributed without the express
prior written permission of the copyright holder. It may not be distributed
through the Internet or on a computer without prior written consent.
MANUFACTURER
The Emergency Breathing System (EBS) is manufactured by Huish Outdoors, Inc
1540 2200 W, Salt Lake City, UT 84116, USA
TRADEMARK NOTICE
HUISH OUTDOORS and the ZEAGLE logo are registered or unregistered
trademarks of Huish Outdoors. All rights are reserved.
EC TYPE EXAMINATION CONDUCTED BY:
Products carrying the CE mark have the EC Type Examination conducted by:
SGS Fimko Oy, Takomotie 8, Fl-00380 Helsinki, Finland Notified Body0598
All products sold by Zeagle in the EU (European Union) comply with the
following requirements where applicable.
EN12021: This standard specifies the allowable contaminates and component
gases that make up compressed air. This standard is the equivalent of the USA
Compressed Gas Association’s Grade E air. Both standards allow very small
amounts of contaminants that are not harmful to breathe, but can cause a
problem if present in systems using gases with a high percentage of oxygen.
EN4856: The EBS is considered a Emergency Self Contained Emergency
Underwater Breathing Apparatus. A such, the EBS is designed for user
deployment and operation with sufficient notice given to the user to deploy
the device.
A “Declaration of Conformity” is on the Huish Outdoors website at:
https://www.huishoutdoors.com/eu-declarations/declarations-of-conformity/
The Zeagle Defender EBS is certified as an underwater escape breathing apparatus for use to a maximum depth of 4m in water temperatures of 4°C or greater. It is certified that the EBS meets the Essential Health and Safety Requirements in Annex Il of Regulation (EU) 2016/425 Personal Protective Equipment. This PPE satisfies the requirements for a shallow water underwater escape breathing apparatus based on relevant principles and elements of EN4856:2023 – Rotorcraft Emergency Breathing Systems (EBS), requirements, testing and marking. Also in compliance to harmonized standard EN250:2014 – Respiratory Equipment open circuit, self contained, compressed air diving apparatus, requirements, testing and marking.
INTRODUCTION
The Defender Emergency Breathing System (EBS) is a device intended to assist
an individual to escape from an emergency situation involving involuntary
submergence below water, such as that which might occur with a downed aircraft
in a body of water.
It is a self-contained system that features a robust design with particular
attention to simplicity, reliability and convenient compact size to enable
ease of use and maintenance.
Additional features that include adjustable mouthpiece positions, a flexible
hose with swivel connections and convenient port connections provide desirable
options for the user to adapt the device to personal preferences.
While the EBS is not difficult to use, specific training and practice is
required for the user to use it effectively in an emergency situation.
WARNINGS, CAUTIONS AND NOTES
This manual contains information that could affect your safety. Certain items
are emphasized with symbols and signal words to indicate a potential hazard.
WARNING
The Defender EBS is intended to be used as a device to assist an individual in
self-rescue to escape from an emergency situation involving submergence in
shallow water for a brief period. It is not designed nor tested for any other
purpose.
Proper training is essential for your safety. Training experience must include
in- water survival, breathing compressed air underwater and emergency egress
from aircraft.
You must be familiar with and trained in the use of this particular device to
maximize your effective use of it. Failure to respond to this warning could
result in a situation leading to serious injury or death.
WARNING
In order to prepare for, use and maintain this device properly, it is
essential that you understand all of these aspects through reading this manual
and mastering the skills necessary to operate, inspect and maintain it
properly. Failure to do so could result in a situation whereby the EBS would
not be an available resource to assist with self rescue and possibly lead to
serious injury or death.
GENERAL DESCRIPTION
The EBS is comprised of a cylinder of compressed air with an integrated cylinder valve and pressure regulator designed to interface with the user and deliver air at a reduced pressure suitable for breathing. When fully charged to maximum operating pressure of 207 BAR (3000 PSI), the cylinder has the capacity to deliver 56.6L (2.0 Cubic Feet) of air at atmospheric pressure.
WARNING
The EBS has a limited capacity to deliver air for only a brief period in
shallow water (4m;13 ft)
The time duration for supplying air will vary between individuals and
breathing rates. This is an indicator supporting the need for training to
maximize effective use of the device.
CYLINDER
An aluminum alloy refillable seamless Catalina cylinder is used to store
compressed air for delivery to the user. It complies with ISO 7866 standards
for certification and has a maximum working pressure of 207 bar (3000 PSI).
All cylinders shipped with the Defender EBS are EU approved and labeled as
follows in compliance with EU standards.
EBS FIRST STAGE REGULATOR
The EBS integrated first stage regulator and valve is threaded into the
cylinder to provide a compact assembly. It is provided with an onboard
pressure gauge, an ON/OFF port or knob a assembly, fill port and low-pressure
hose with a swivel connection attached to the second stage.
1 | LP HOSE SWIVEL |
---|---|
2 | LP HOSE |
3 | PRESSURE GAUGE |
4 | FILL PORT |
5 | ON/OFF PORT |
The integrated cylinder valve/regulator is available in two different
configurations with either a manually-operated ON/OFF port or an option that
requires a 3 mm hex tool to operate the ON/OFF selection.
Selection of the OFF position applies to maintenance activity only and should
never be selected when the EBS is ready and available for use.
EBS FIRST STAGE WITH ON/OFF KNOB INSTRUCTIONS
The first stage regulator with an integrated on/off knob is illustrated below.
The unit must be in the on position or turned fully in the counterclockwise
direction until aligned with green zone and arrow.
To turn off, rotate clockwise until the red zone is aligned with the arrow.
See instructions for use in EBS Deployment section of this manual.
WARNING The ON/OFF port of the integrated valve and first stage must
always be in the ON position any time the EBS is in service and ready to be
deployed. If the valve has been closed to the OFF position at any time, it
must be fully reopened to the ON position by rotating counterclockwise at
least one full revolution before it is acceptable for use. DO NOT place the
unit into service without ensuring it is in the ON position.
Failure to ensure unit is in the ON position will render the EBS incapable of
delivering air to the user.
EBS SECOND STAGE REGULATOR
The second stage regulator with an integrated mouthpiece is connected to the
first stage by means of a flexible hose with a swivel connection. Its purpose
is to provide a means to connect with the user’s mouth and supply air on
demand for breathing. See instructions for use in EBS Deployment section of
this manual.
The second stage assembly also provides a location for storage of a nose clip
to assist with breathing control and a purge cover cap to prevent accidental
operation of the purge function when the EBS is in transit or otherwise not in
service and ready to use.
ITEM # | DESCRIPTION |
---|---|
1 | NOSE CLIP |
2 | MOUTHPIECE |
3 | NOSE CLIP TAB |
4 | EXHAUST VALVES |
5 | PURGE COVER CAP |
EBS PREPARATION FOR IN-SERVICE USE
CYLINDER PRESSURE – The EBS is delivered from the factory fully charged
with the valve in the ON position ready for immediate use. Cylinder pressure
must not be less than 207 bar (3000) PSI to be fully charged. If charging is
necessary, refer to the Cylinder Filling Procedure contained in this manual.
FIRST STAGE ON/OFF PORT – The port must be in the ON position at all
times when the EBS is in service and ready to use.
SECOND STAGE FEATURE – The mouthpiece orientation is adjustable but
should only be changed by a trained service technician. It may be indexed to
the user’s preference, but only as a maintenance option and may not be changed
while the unit is in service for immediate use.
SECOND STAGE PROTECTIVE COVER – This device is intended for EBS transit
or storage only. It MUST NOT be installed anytime the EBS is placed in service
and ready for use.
DEPLOYMENT OF EBS IN EMERGENCY SITUATION
-
Use one hand to secure your position at the nearest exit escape route. Use your free hand to grasp the EBS second stage regulator and pull out and away from your body to ensure hose is free from entanglement.
-
Place your mouth over the mouthpiece and exhale providing a puff of air to clear residual water that might be inside the second stage housing. Then inhale, lightly at first to be sure residual water is not present, and then continue to breathe normally. If there is water present after first exhalation, block the air path in the mouthpiece with your tongue and press the purge feature on the front of the second stage housing to clear residual water and then continue to breathe normally, recognizing that the air supply is limited and expedient travel to the surface is required.
WARNING DO NOT HOLD YOUR BREATH. BREATHE NORMALLY. Failure to exhale lightly and continuously or breathe uninterrupted as you ascend from even shallow depths can create a hazardous pressure imbalance in the lungs risking respiratory damage, with the possibility of serious injury or death. -
Immediately after you have control of the mouthpiece in your mouth, use your free hand to locate and grasp the Nose Clip that is secured to the second stage housing. Pinch the Nose Clip tabs together as shown in the illustration below to open it and place
it over the nose and release. The purpose of the Nose Clip is to block entry of water into the nose while breathing. Do not try to breathe through the nose while submerged. -
Note that the EBS is designed to deliver air only when you attempt to inhale. When you stop inhaling, air flow ceases allowing you to comfortably exhale. Air flow to you can be manually activated anytime by pressing the purge feature as long as there is a remaining air supply in the cylinder.
-
After you reach the surface and have access to air, discontinue use of the EBS. Once the EBS has been deployed for either an emergency or training it must be removed from service, no longer available for emergency use until after it has been recharged or received periodic maintenance.
CYLINDER FILLING GUIDELINES
Connection to a compressed air supply to charge the EBS requires a specific Filler Adapter. It is a compressor accessory for the air supply facility and is not supplied * as part of the EBS assembly.
Filler Adapter Assembly (includes Male QD Connector) is available from Zeagle. Part Number 330-9420. Note: Fill Adapter DIN connection option for attaching to the HP supply is acceptable The yoke type (CGA) connector option is limited to a maximum pressure of 232 bar (3364 PSI) and is not suitable for greater pressure required to fully charge the cylinder.
WARNING Do not expose a Yoke type (CGA) Fill Adapter to a pressure
greater than 232 bar (3364 PSI). It is not rated for the higher pressure and
could lead to damage exposing the operator to the risk of serious or death.
WARNING Do not attempt to fill the EBS cylinder with any gas other than
clean air that conforms to EN12021 or the equivalent CGA Grade E standard.
Introduction of other gases including Enriched Air Nitrox (EAN) is prohibited.
Handling of oxygen- enriched air requires specialized treatment of involved
equipment and if not performed properly, could create a hazard with a possible
risk of serious injury or death.
NOTICE
Before connecting the Fill Adapter to the Male QD observe the following
information.
When compressed air is introduced into the cylinder heat is generated. The
amount of heat is related to how rapidly the cylinder is filled. In order to
prevent overheating the cylinder the fill rate must be limited.
Iuminum cylinders such as that used with the EBS can tolerate a limited amount
of heat exposure if they are well maintained and undamaged, but the presence
of heat interferes with reaching the full volumetric capacity of the cylinder
at 207 bar with the cylinder at 21°C (70°F)
If the cylinder is filled to 207 bar at a higher temperature, after filling
has stopped and the cylinder cools, the pressure will reduce and the charge
will not be at full capacity. Slow filling is preferred and is limited to a
recommended rate of not greater than 28 bar (400 PSI) per minute. This will
greatly reduce the generation of heat and will require about 11 minutes to
reach full capacity.
WARNING Do not exceed the maximum working pressure of the EBS at 207 bar
(3000 PSI) for any reason. Use the fill station pressure gauge to monitor
cylinder pressure as it is generally more precise than the EBS on board gauge.
Cylinder pressure and filling rate must be constantly monitored to ensure the
maximum working pressure is not exceeded.
Slow filling at a rate not to exceed 28 bar per minute is preferred.
According to the cylinder manufacturer, fast filling is acceptable provided
the cylinder is well maintained and undamaged and the fill rate does not
exceed 30 L per minute. For the EBS this is equivalent to a maximum
compression rate of 94 bar (1350 PSI) per minute, requiring less than 3.5
minutes to achieve a full charge.
Exceeding the maximum pressure and fill rate could result in damage to the EBS
and create a hazardous condition with the possibility of serious injury or
death.
To Fill the EBS cylinder perform the following procedure:
- Remove the Fill Port Cap turning counterclockwise.
- Inspect the fill port passage and remove any moisture, corrosion or debris.
- Insert the Male QD adapter of the Filler Adapter assembly into the EBS fill port turning clockwise by hand until snug.
- Operate the valve located on the compressor fill whip to begin introduction air into the cylinder. Slowly fill the cylinder.
- After filling is complete, close the fill whip control valve, open the bleed screw to remove residual pressure and operate the Female QD to separate the Fill Adapter from the EBS.
- Remove the Male QD Adapter from the Fill Port. Replace the Fill Cap and tighten until snug.
CARE AND MAINTENANCE
Transport and Storage
Whenever possible the EBS should be dried and placed in a padded carrying or
equipment bag for transport and storage. Care should be taken to avoid
exposure of the EBS to sharp or heavy objects that might scratch or otherwise
damage the unit. Whenever the EBS is removed from service and after drying,
install the Second Stage Protective Cover. Installation of the cover protects
against accident purge operation and signifies it is out-of service and not
available for use.
Cleaning after use and exposure to submergence in water
-
If possible, immerse the entire EBS in a warm, clean freshwater bath and soak for one hour, preferably while still pressurized and with the ON/OFF port in the ON position.
If pressure is not available, DO NOT operate the purge function while the unit is submerged or being rinsed. Such action can cause water and other contaminates to enter the EBS requiring further service. -
Flush the exterior and accessible interior with clean freshwater to remove sea salt and other contaminates. Be sure mouthpiece and exhaust valves are included in flushing operation.
-
Following rinsing, dip the second stage in a sanitizing bath as prescribed below.
Dissolve one Edwards-Councilor Steramine Sanitizing Tablet per 3.8 L (1 Gal) of freshwater per manufacturer’s instruction. Miltons tablets are an acceptable alternative provided that use is consistent with Milton’s instruction and cleaning protocol. -
Do not rinse again after exposing unit to sanitizing bath. Just allow it to dry. If possible, place the entire EBS Assembly on a cool, flat surface not exposed to direct sunlight to dry.
NOTICE Do not inject or spray lubricants or other chemicals into the EBS
first or second stages or on to external surfaces. It is not necessary and may
attract contaminants that may later interfere with proper operation.
WARNING Do not remove any parts of the EBS for routine rinsing, including
removal of the purge cover. Improper reassembly could result in interference
with proper operation of the EBS and could result in unexpected loss of air
delivery while underwater.
Prior to storing the EBS:
- Ensure that the unit is completely dry and clean.
- If circumstances prevented cleaning of the EBS before transport, or if it became exposed to wet or unclean conditions it must be subjected to the cleaning procedure described above before being returned to storage.
PERIODIC MAINTENANCE AND SERVICE
As with any equipment that may be called upon for assistance in an emergency,
the readiness of the equipment to perform as intended is crucial. Therefore it
is extremely important that your organization have and follow a plan that
prescribes the frequency of inspection and servicing to ensure the EBS
condition has not been compromised such that its ability to perform is in
jeopardy. Needs vary from one organization to another and the proper interval
for inspections is at the discretion of your safety officer. However, at a
minimum, the following should be performed.
At any time during the handling or observation of the EBS there is evidence of
loss of pressure or physical damage to the unit, it must be taken out of
service and inspected by an authorized service technician.
NOTICE According to the aluminum cylinder manufacturer, the EBS cylinder
if less than 2″ in diameter and less than 12″ in length is excepted from
volumetric expansion test.
However the cylinder must be taken out of service if it has been exposed to a
temperature greater than 176°C (350°F) or has received a scratch or scuff on
the cylinder body with a depth greater than 0.127mm (0.005 inches).
Operational units must be serviced at least annually. Service may include
comprehensive inspection and disassembly, cleaning and adjustment.
WARNING Disassembly of the EBS is to be done only by a technician that
has received training and authorization to service the EBS, using Zeagle or
Huish Outdoors supplied parts only.
Do not attempt to disassemble the unit without training and authorization.
Such activity could lead to a serious malfunction of the unit resulting in a
loss of air delivery in a critical situation, possibly leading to serious
injury or death.
Training units must be serviced at least annually, after not more than 200 uses or after any observation of improper operation or physical damage to the EBS unit. If it is suspected that water or other contaminants has entered the unit, it should be taken out of service and submitted for maintenance. If there are any questions as to the need for maintenance, contact your distributor.
ZE.01.05.0031 REV. A
Documents / Resources
|
zeagle DEFENDER EBS Emergency Breathing
System
[pdf] User Manual
PN 350-9200, PN 350-9210, DEFENDER EBS, DEFENDER EBS Emergency Breathing
System, Emergency Breathing System, Breathing System, System
---|---
|
Zeagle DEFENDER EBS Emergency Breathing
System
[pdf] User Manual
ZE.01.05.0030, DEFENDER EBS, DEFENDER EBS Emergency Breathing System,
Emergency Breathing System, Breathing System
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>