USABlueBook 28392 Differential pH and ORP Probes Instruction Manual
- June 12, 2024
- USABlueBook
Table of Contents
OPERATING INSTRUCTION MANUAL
MODEL 28392/28393, 28388/28389, 28218/28219
DIFFERENTIAL PH & ORP PROBES
GENERAL INFORMATION
This manual covers USABlueBook differential measurement pH and ORP probes.
All mounting configurations are described.
Both those probes with encapsulated preamplifiers and those with encapsulated
two-wire
transmitters are covered. If your probe has five wires it has an encapsulated
preamplifier (28388, 28389, 28392, 28393). If your probe has two wires, or in
the case of an ORP probe four wires, it has an encapsulated two-wire
transmitter (28218 and 28219).
The output from a two-wire transmitter type is non-isolated and un-calibrated.
The system must provide 24 Vdc, with “low” isolated from earth ground, and a
means of calibrating for offset and span.
NOTE: Do not discard the protective cap(s) that came with the sensor. If
the sensor is removed from the process for an extended period of time,
thoroughly clean the sensor, put a piece of cotton ball with few drops of
water into the protective cap and replace it on the sensor. This keeps the
junction from drying out which causes slow response when put back into
operation or causes permanent damage to the sensor. Sensors should not be left
in dry lines or empty tanks for extended periods.
Do not store the sensors in a dry or humid location. When storing, check the
protective cap(s) regularly to make sure the cotton ball remains moist.
Improper storage of sensors voids the warranty.
SPECIFICATIONS
MEASURING RANGES
pH: 0 to 14.00 pH
ORP: 28389/28393: +1999 mV
28219: 0-1000 mV / +500 mV field selectable
FLOW RATE
10 ft./sec max. Flow should be as low as possible in low conductivity water
and in solutions with high suspended solids
WETTED MATERIALS
28388/28399/28218/28219: Ryton®, Kynar, glass, titanium palladium alloy and
Viton® (Platinum for ORP probes). 28392/28393: Ryton (PPS) body, ceramic
junction, glass electrode, Viton®, titanium, (Platinum for ORP probe).
TRANSMISSION DISTANCE
5-wire Probes: 900 meters (3000 ft.) 2-wire Probes: Limited only by cable
resistance and voltage of power supply
SENSITIVITY
pH: Less than 0.005 pH
ORP: Less than 0.5mV
STABILITY
0.03 pH / 2mV per day, non-cumulative
PRESSURE LIMIT
100 psig @ 65°C
40 psig @ 95°C
TEMPERATURE LIMITS
-5 to 95°C (23 to 203°F)
POWER SUPPLY LIMIT
24VDC ± 4V
PROBE CABLE
15 ft. (4.6 m)
AUTOMATIC TEMPERATURE COMPENSATION
-5 to 95°C (23 to 203°F) 28218:
OUTPUT SPAN: 1.14mA / pH
OUTPUT OFFSET: 12mA @ 7pH
+1.14mA
MAX LOAD: 450 Ohms
28219:
OUTPUT SPAN:
0-1000mV: 1.6mA / 100mV
+500mV: 1.6mA / 100mV
OUTPUT OFFSET:
0-1000mV: 12mA @ 500mV
+1mA
+500mV: 12mA @ 0mV +1mA
MAX LOAD: 450 Ohms
INSTALLATION
3.1. General Instructions
Specific instructions for each type of probe are given in the following pages.
Common to all probes are the following instructions:
a) If the distance between the probe and the instrument is such that a direct
connection is not possible, the probe cable should be routed to a junction box
with a terminal strip. The box should be well sealed and away from corrosion
danger. Be sure that you have sufficient slack cable to allow for probe
removal for calibration and servicing.
b) Route the interconnect cable from the junction box to the instrument,
preferably in metal conduit. Do not run the power cable or control cables in
the same conduit with the probe interconnect cable.
c) Remove the protective plastic caps from the end of the probe before placing
in service.
d) For best results probes should always be mounted vertically with electrodes
down. If this is not possible, the probe must be at least 15° above
horizontal.
3.2. ORP 4-20 mA Probes
a) ORP 4-20 mA probes (28219) have four wires; black, red, green and white.
The red wire is to be connected to the +24 Vdc terminal and the back wire to
the 24 Vdc common terminal via the load in the loop.
b) For an instrument with a range of 0 to 1000 mV the green and white wires
are to be shorted.
c) For an instrument with a range of –500 to 500 mV the green and white wires
are to be isolated from each other.
3.3. Submersion Mounting Differential Probes
a) Apply a thread sealant to the thread on the cable end of the probe and
screw a 1-1/2” x 1” NPT reducer onto the probe, supplied from your own
hardware. Route the sensor cable through an appropriate length of 1” pipe and
using thread sealant, screw the pipe into the reducer. 28392 and 28393 have
1” threads on the probe itself and do not require the use of a reducer. The
cable end of the probe should not be exposed to the process. A cable strain
relief fitting should be used on the upper end of the pipe as well as a wire
bracket to aid in supporting the assembly.
3.4. Flow-through Tee Mounting Differential Probes
a) Apply pipe sealant to the electrode end of the probe and screw it into any
standard 1-1/2” NPT tee (28388/28389, 28218/28219) or 1” NPT tee
(28392/28393). Be careful to not over-thread the probe into the tee, causing
the glass at the end of the probe to become damaged from contacting the
inside wall of the tee.
SERVICE AND MAINTENANCE
4.1. Probe Cleaning
a) The probe should be kept reasonably clean to avoid measurement errors.
Frequency of cleaning can only be determined by experience. To clean proceed
as follows:
b) Rinse with clean warm water.
c) Soak the end of the probe in warm water and dish detergent for 3 or 4
minutes.
d) Brush the end of the probe, particularly the three electrodes with a soft
bristle brush such as a tooth brush. Take care not to scratch the glass
electrode.
e) If the probe is still not clean, it may have to be cleaned with acid.
CAUTION: Do not acid clean probes used in processes containing cyanide
solutions. Some experimentation may be required to determine the most suitable
acid for your process. Use the most dilute acid which is effective. Normally
10 parts of water to one part muriatic acid is sufficient. Do not use
hydrofluoric acid.
f) Soak the probe for not more than 5 minutes in the chosen acid; then rinse
thoroughly with clean warm water and soak in water for 3-5 minutes.
g) Calibrate the system in accordance with the instrument instruction manual.
4.2. Replacement of Salt Bridge for Differential Probes
a) If the system cannot be calibrated after cleaning the probe, it may be
necessary to replace the standard cell solution (7pH buffer). Salt bridges are
also available (Part No. 28391 for 1-1/2” body probes, Part No. 28361 for 1”
body probes). Proceed as follows:
b) Hold the probe vertically with the sensor face up. Using a small hex wrench
(or phillips screwdriver for 1” body probes), remove salt bridge by turning
counter-clockwise, taking care not to damage the glass electrode. Discard the
used salt bridge.
c) Up-end the probe and pour out the contents of the standard electrode
chamber. Flush the chamber with a small amount of pH 7 buffer or distilled
water.
d) Refill the chamber with 7pH buffer solution up to the tip of the electrode
inside the chamber. DO NOT OVERFILL. It is important to leave space for the
salt bridge thread and a small amount of air.
e) Screw the new salt bridge into the cavity until finger tight. Now perform a
¼ turn with wrench (or phillips screwdriver).
4.3. Storage
a) Do not discard the protective cap(s) that came with the sensor. If the
sensor is removed from the process for an extended period of time, thoroughly
clean the sensor, put a piece of cotton ball with few drops of water into the
protective cap and replace it on the sensor. This keeps the junction from
drying out which causes slow response when put back into operation or causes
permanent damage to the sensor. Sensors should not be left in dry lines or
empty tanks for extended periods.
b) Do not store the sensors in a dry or humid location. When storing, check
the protective cap(s) regularly to make sure the cotton ball remains moist.
Improper storage of sensors voids the warranty.
TROUBLESHOOTING AND SERVICE
5.1. Checking 5-wire Differential Probes (28388/28389, 28392/28393)
The probe can be checked by a few simple measurements. Two pH buffer
solutions, 7 pH and either 4 pH or 10 pH, and a multimeter are required. For
ORP probes two calibration solutions, 200 and 600 mV, are required. Millivolt
solutions may be ±20% of nominal value. Actual value is noted on the bottle.
5.1.1. pH probes (28388, 28392)
a) Clean the probe as described in Section 4.1. If the system cannot be
calibrated, replace the salt bridge and 7pH buffer solution as described in
4.2. If the system still can’t be calibrated check the probe as follows:
b) Disconnect red, green, yellow and black wires at the junction box. If you
are not using a junction box, disconnect at the instrument after shutting off
the power.
c) Place the probe in 7 pH buffer. Allow enough time for the temperature of
the probe and buffer to stabilize at room temperature.
d) Measure the resistance between the yellow and black wires to check the
probe’s temperature compensator. The resistance should be between 250 and 350
ohms at 25°C. If the resistance is within specifications the probes thermistor
is functioning correctly.
e) Reconnect the yellow and black wires and restore power to the instrument.
f) Measure the voltage between the red and green wires. If it is not within
–50 to +50 millivolts with the probe in 7 pH buffer, the probe is defective.
If the voltage is OK proceed to the next step.
g) Rinse the probe and place it in 4 pH or 10 pH buffer. Allow it to stabilize
then check the voltage again between the red and green wires. If the voltage
is between 100 and 230 millivolts (negative in 10 pH buffer, positive in 4 pH
buffer) the probe is within specifications.
5.1.2. ORP probes (28389, 28393)
a) Clean the probe as described in Section 4.1. If the system cannot be
calibrated, replace the salt bridge and 7pH buffer solution as described in
4.2. If the system still can’t be calibrated check the probe as follows:
b) Disconnect red, green, yellow and black wires at the junction box. If you
are not using a junction box, disconnect at the instrument after shutting off
the power.
c) Place the probe in 200 mV solution. Allow enough time for the temperature
of the probe and solution to stabilize at room temperature.
d) Measure the resistance between the yellow and black wires to check the
probe’s temperature compensator. The resistance should be between 250 and 350
ohms at 25°C. If the resistance is within specifications the probes thermistor
is functioning correctly.
e) Reconnect the yellow and black wires and restore power to the instrument.
f) Measure the voltage between the red and green wires. The reading should be
between 160 and 240 mV; otherwise, the probe is defective. If the voltage is
OK proceed to the next step.
g) Rinse the probe and place it in 600 mV solution. Allow it to stabilize then
check the voltage again between the red and green wires. If the voltage is
between 560 and 640 mV, the probe is within specifications.
5.2. Checking Probes 28218/28219 (4-20 mA output)
The operation of the 2-wire, 4-20 mA, probe can be checked by a few simple
measurements. Two pH buffer solutions, 7 pH and either 4 pH or 10 pH, and a dc
milliammeter are required.
5.2.1. 28218 pH Probe
a) Disconnect the red wire at the instrument or power supply and connect it to
the milliammeter (-) black.
b) Connect the milliammeter (+) red to the instrument or power supply red wire
output terminal.
c) Rinse the probe and place it in 7 pH buffer. Allow the temperature of the
buffer and probe to stabilize at room temperature.
d) Check the offset of the probe by reading the milliammeter. The reading
should be between 11 and 13 mA. If not the probe is defective. If the offset
is OK, note the exact reading and proceed to the next step.
e) Rinse the probe and place it in 4 pH or 10 pH buffer. Allow the temperature
of the probe and buffer to stabilize at room temperature. Now check the span
of the probe by reading the milliammeter.
• If the probe is in 4 pH buffer, the reading should be between 2.85 and 3.99
lower than the reading obtained in (d).
• If the probe is in 10 pH buffer, the reading should be between 2.85 and 3.99
higher than the reading obtained in (d).
o If this test is not satisfied, the probe is defective.
• If you wish to check the temperature compensator, proceed to step (f).
• If the span of the probe drops below 2.85mA than the probe still can be used
adjustments will have to be made to the receiving device to compensate for the
low span.
f) To check the operation of the temperature sensor in the probe, heat the
buffer used in step (e) with the probe in it to about 50°C. The milliammeter
reading should be within ±0.15 mA of the reading observed in step (e).
5.2.2. 28219 ORP Probe
a) Disconnect the red wire at the instrument or power supply and connect it to
the milliammeter (-) black. Connect the milliammeter (+) red to the instrument
or power supply red wire output terminal.
b) For a probe with range of –500 to 500 mV (white and green wire open):
1. Rinse the probe and place it in the 200 mV solution. The milliammeter
should read between 13.7 and 16.7 mA.
2. Rinse the probe and place it in the 600 mV solution. The meter should read
between 19.44 and 23.8 mA.
c) For a probe with range of 0 to 1000 mV: (white and green wire joined):
1. Rinse the probe and place it in the 200 mV solution. The milliammeter
should read between 6.4 and 8.0 mA.
2. Rinse the probe and place it in the 600 mV solution. The meter should read
between 12.2 and 15.0 mA.
5.3. Customer Service
If a problem has not been resolved with the above procedures, a telephone
consultation with USABlueBook will provide the answer.
USABlueBook
800-548-1234
tech@usabluebook.com
5.4. Parts and Accessories
Description | Part # |
---|---|
pH 7, 500 mL Buffer Solution | 40470 |
pH 7, 4L Buffer Solution | 40475 |
pH 4, 500 mL Buffer Solution | 40460 |
pH 4, 4L Buffer Solution | 40465 |
pH 10, 500 mL Buffer Solution | 40480 |
pH 10, 4L Buffer Solution | 40477 |
200mV ORP Standard Solution, 500 mL | 40423 |
400mV ORP Standard Solution, 500 mL | 40424 |
600mV ORP Standard Solution, 500 mL | 40425 |
Salt bridge kit for 1″ probes (Package of 3 salt bridges; pH7 buffer fill
solution not included)| 28361
Salt bridge kit for 1.5″ probes (Package of 3 salt bridges; pH7 buffer fill
solution not included)| 28391
Junction box with terminal strip| 22950
50 ft. 5-wire Interconnect cable dressed both ends| 28400
DRAWINGS
WIRING
DIFFERENTIAL PROBE MOUNTING OPTIONS
INSTALLATION OF SALT BRIDGE IN A DIFFERENTIAL PROBE
USABlueBook Differential pH and ORP Probes
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>