ASSEMBLY GUIDE (REV 00)
M162 LCR METER DIY KIT
Tech M162 LCR Meter Diy Kit
MODEL | M162 |
---|---|
PCB VERSION | MAIN: 109-16202-0013 ANALOG: 109-16201-00F |
FIRMWARE | 113-16202-060 |
REQUIRED TOOLS
- 25W Soldering Iron
- Soldering Wire (Rosin-core Recommended)
- Digital Multimeter
- Screwdriver – Philips f1
- Flush (wire) cutter
- Tweezers
- Micro-USB power cable
Getting Started
TEST THE MAIN BOARD
- Connect the USB Cable to the Micro-USB port on the Main Board and a USB power source.
- Ensure your M162 main board boots up correctly. When powered on. the LED (D1) will blink twice. The display will then go through two splash screens and enter into a state similar to the below illustration.
- The values of the numbers are random. Please ignore them and just ensure the display is normal.
If your M162 main board does not power up. or powers up with a blank screen. please contact us at support@jyetech.com. Do not solder any parts onto the board if you encounter any issues as this will void the warranty.
Main Board Assembly
STRIP SOCKETS
- J4 12×1 strip socket
- J8. J9 2×1 strip socket
Analog Board Assembly
RESISTORS
Use a multimeter or a resistor color code to identify the resistor values.
SLIDE SWITCH
- SW4 SK-22003T
PIN HEADERS
- J5 12×1 pin header
- J8. J9 2×1 pin header
CONTACT ARMS
- El JI3. J14. J15. JI6. J17. J18
- RI. R2. R4. P-R8. R11. R15. R16. RI9. R20 ………………1KΩ
- R3, R12 …………………………………100Ω
- R6 ………………………………………..560Ω
- R10……………………………………. 100KΩ
- R14 ……………………………………………5.1KΩ
- R17. R21. R23. R24 ………………. 6.2KΩ
- R18 …………………………………………47Ω
- R27. R28…………………………………….. 10Ω
CAPACITORS
- C9. CIO. C1I. 100μF, 16V C12. C13
Ensure the polarity of the capacitors are installed correctly. The positive
pole goes into the holes of the square pads.
PUSH BUTTONS
-
SWI. SW2. SW3 6.2×6.2×9.5 mm tact switch
-
J11 XH2.54 2-pin header. right-angled
-
BOBY JYEI18 battery charger breakout board (with 5xI strip pin header)
CLOSE JP1
(Note: skip this step if the battery charger board is installed.)
JPI
Verify Voltages
- Apply 5V power supply via the micro-USEI connector on the analog board.
- Spot check the voltages on the board at the test points labeled to ensure they measure to the voltages in the chart below.
REFERENCE
| Voltage
---|---
Test Points| Min| Typical| Max
V+| 3.60V| | 5.50V
VS+| 3.60V| | 5.50V
AV+| 310V| 3.30V| 3.50V
V-| | | -2.50V
VS-| | | -2.50V
AV-| -2.20V| -2.00V| 1.80V
V1| | 0.0V|
V2| | 0.0V|
V3| | 0.0V|
V4| 1.40V| 1.65V| 1.80V
Quick Test
-
Attach the main board to the analog board
-
Apply power supply via the micro-USEI connector on the analog board and slide the power switch to ON position.
-
Once the device starts up. hold down RU. button to bring up the menu. Select the ‘Default’ option to set all the parameters to their default values
-
Test measurement with the provided resistors (10 & 1140). capacitor (0.10F). and inductor ilmH). or with your own com-ponents.
Note: The provided resistors are 1% tolerance. The capacitor and inductor tolerance can be 10% or higher. If the measured values are away from their labeled value they might not reflect a problem of the meter. Further confirmation is required.
Enclosure Assembly
-
Identify the ends of the standoffs.
-
Assemble the two side pieces and back piece. Ensure the standoffs are at the correct orientation.
-
Insert the assembled main board-analog board with the analog board fitted into the standoff slots near the bottom ends.
-
Assemble the front piece. Ensure the standoffs are at the corret orientation.
-
Assemble the top and bottom pieces with screws. Remember to attach the 3 caps to the pushbuttons before securing the top piece
-