RADWAG 5Y Series UYA 5Y Ultra Microbalances User Manual
- June 12, 2024
- RADWAG
Table of Contents
- RADWAG 5Y Series UYA 5Y Ultra Microbalances
- Product Information
- Product Usage Instructions
- GENERAL INFORMATION
- TRANSPORT AND STORAGE
- UNPACKING AND INSTALLATION
- PROGRAM STRUCTURE
- WEIGHING RESULT WINDOW
- OPERATING THE MENU
- ADJUSTMENT
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
RADWAG 5Y Series UYA 5Y Ultra Microbalances
Product Information
The 5Y Series Balances are a range of precision weighing devices manufactured by RADWAG. The series includes the following models:
- UYA 5Y Ultra-Microbalances
- MYA 5Y Microbalances
- XA 5Y.M.A Microbalances
- XA 5Y.A Analytical Balances
- XA 5Y.F Analytical Balances for Filters
- PM 5Y Precision Balances
This user manual, with the code IMMU-111-04-03-22-ENG, provides detailed instructions for the operation and maintenance of the 5Y Series Balances. The manual is available at www.radwag.com.
Product Usage Instructions
Transport and Storage
Before using the balance, perform the following:
- Check the delivery to ensure all components are present.
- Inspect the packaging for any signs of damage.
Profiles
The balances support profile creation for personalized settings. Follow
these steps to create a profile:
- Navigate to the profile creation section.
- Configure the settings according to your preferences.
- Select the working modes required.
- Specify the readout and units.
Weighing
To perform weighing using the 5Y Series Balances:
- Prepare the sample to be weighed.
- Place the sample on the balance pan.
- Wait for the balance to stabilize and display the weight.
Parts Counting
If you need to count parts using the balance:
- Prepare a known quantity of parts as a reference.
- Select the parts counting mode on the balance.
- Place the reference parts on the balance pan and press the count button.
- Add the parts to be counted and let the balance calculate the quantity.
Checkweighing
To use the checkweighing feature:
- Set the desired checkweighing thresholds in the balance settings.
- Select the checkweighing mode.
- Place the item to be checked on the balance pan.
- Check if the weight falls within the specified thresholds.
Dosing
If you need to dose a specific product:
- Utilize the product database during dosing operations.
- Configure the mode-related settings for dosing.
- Follow the dosing procedure specified in the user manual.
Percent Weighing Against Reference Sample Mass
To perform percent weighing against a reference sample mass:
- Compare the sample mass with the reference sample mass.
- Utilize the checkweighing and dosing features in percent weighing mode.
- Refer to the bar graph for visual representation of the comparison.
5Y Series Balances
UYA 5Y Ultra-Microbalances MYA 5Y Microbalances XA 5Y.M.A Microbalances XA
5Y.A Analytical Balances XA 5Y.F Analytical Balances for Filters PM 5Y
Precision Balances
USER MANUAL
IMMU-111-04-03-22-ENG
www.radwag.com
Congratulations and thank you for selecting RADWAG product. You have purchased
a device that has been designed and manufactured to give you years of service.
Please read this user manual carefully, this shall guarantee reliable
operation.
MARCH 2022
– 2 –
– 8 –
GENERAL INFORMATION
Dimensions
UYA 5Y Ult ra- Microbalances and MYA 5Y Microbalances
– 9 –
I ndicator XA 5Y.M.A Microbalances
– 10 –
XA 5Y.M Microbalances XA 5Y.A Series Balances
– 11 –
XA 5Y Series Balance XA 5Y.F Series Balances
– 12 –
PM xx.4Y, d=0.01g PM xx.4Y, d=0.1g
– 13 –
PM xx.4Y, d=0.5g, d=1g
– 14 –
1.2. Connectors Top view
1 Button used to hard restart or to switch 10 Power supply socket USB type C
the balance off/on.
2 Left speaker
11 HDMI socket
3 Right speaker
12 Media box socket
4 USB type A
13 RFID sensor
5 USB type A
14 Proximity sensors
6 Microphone
15 Camera
7 Kensington Lock
16 Camera diode
8 Ethernet socket
17 Signalling diode
9 Power supply socket
1.3. Intended Use
5Y series balances have been designed to enable precise determination of mass in laboratory.
1.4. Precautions
· Prior to the first use, carefully read this user manual. Use the device only
as intended. · Do not operate the touch panel using sharp-edged tools (knife,
screwdriver, etc.). · Place loads in the centre of the weighing pan. · Load
the weighing pan with loads of gross weight which does not exceed the maximum
capacity. · Mind not to leave heavy loads on the weighing pan for longer
periods of time. · In case of damage, immediately unplug the device from the
mains. · The balance to be decommissioned must be decommissioned in accordance
with valid
legal regulations. · Do not use the balance in areas endangered with
explosion. The balance is not intended
to be operated in hazardous areas.
– 15 –
1.5. Warranty Conditions
A. RADWAG is obliged to repair or exchange all elements that appear to be
faulty by production or by construction.
B. Defining defects of unclear origin and means of their elimination can only
be realized with assistance of the manufacturer and user representatives.
C. RADWAG does not bear any responsibility for damage or loss resulting from
unauthorized or inadequate performing of production or service processes.
D. The warranty does not cover: · mechanical damage caused by product
exploitation other than intended, damage of thermal and chemical origin, and
damage caused by lightning, overvoltage in the power grid or other random
event, · damage caused by product exploitation other than intended, damage,
when security seal stickers protecting balance housing against unauthorized
access are removed or broken, · damage caused by liquids and natural wear, or
by overloading the measuring mechanism, · damage caused by inappropriate
setting or by electrical system failures, · inappropriate cleaning habits.
E. Loss of warranty takes place if: · a repair is carried out outside RADWAG
authorized service point, · service claims intrusion into mechanical or
electronic construction by unauthorized people, · the balance does not feature
security stickers.
F. For detailed warranty conditions read the warranty certificate.
1.6. Supervision Over Metrological Parameters
Metrological parameters need to be verified in determined time intervals.
Inspection frequency depends on ambient conditions in which the balance is
operated, type of carried out processes and adopted quality management system.
1.7. User Manual Significance
Even if you are experienced and have already worked with this type of balance,
you are still obliged to read this user manual carefully prior to switching
the device on. This user manual provides all necessary operation-related
information. Following the user manual guidelines guarantees correct weighing
instrument performance.
1.8. Operator Training
The balance must be operated and supervised only by personnel who have been
trained to use such type of a weighing instrument.
– 16 –
TRANSPORT AND STORAGE
Delivery Check
Upon delivery, it is necessary to check the package and the device. Make sure
that your package bears no signs of damage. Otherwise, contact the
manufacturer’s representative.
2.2. Packaging
Keep all package elements should your device be transported in the future.
Remember that only original packaging can be used for shipping purposes. Prior
to packing, uncouple any cables, remove any separable components (weighing
pan, shields, inserts). The balance components must be packed into an original
packaging, this is to protect them against potential damage during
transportation.
UNPACKING AND INSTALLATION
Workstation
· Operate the device in a workroom free of vibrations and shakes, where there
are no air drafts nor dust.
· Make sure that the ambient temperature ranges between: +10 °C ÷ +40 °C, ·
Make sure that the relative humidity is not higher than 80%. · Make sure that
in the course of balance operation, the room temperature changes
slowly and gradually. · Place the balance either on a robust-design table or
on a wall bracket which is both
distant from heat sources and insusceptible to vibrations. · Take special
precaution while weighing magnetic objects, part of the balance is a
strong magnet. · Ground the balance’s base should static electricity affect
the weighing result. The
grounding screw is located at the back of the balance base. · Ultra-
microbalances and other devices installed in the place of use by RADWAG
employees cannot be moved to another location. Only authorized RADWAG employee
can move the device to another location.
3.2. Unpacking
Cut the adhesive tape. Take the balance out of the packaging. Open the
accessory box, take the balance components out of it.
3.3. Standard Delivery Components
· Balance · Bottom weighing chamber shield (XA) · Centring ring (XA) ·
Weighing chamber shield/lid (UYA, MYA) · Weighing pan · Weighing pan shield ·
Bottom ring (MYA, model C) · Power supply
– 17 –
UYA 5Y, MYA 5Y Microbalances
CAUTION: Make sure there is no protective sticker 1 while carrying out the measurement (see the above diagrams). It is obligatory to remove the sticker prior to chamber assembly. Install components following the above diagrams.
A standard balance (UYA and MYA)
2. Draft shield 3. Weighing pan 4. Glass lid for the weighing
chamber
B MYA 0.8/3.5Y microbalance Balances are equipped with standard weighing pan
(A) and weighing pan for filters.
9. Weighing pan for filters 4. Glass lid for the weighing chamber
– 18 –
C balance for filters
12. Bottom ring 10. Weighing pan for filters 11. Standard weighing pan
C balance for filters with automatic weighing chamber
2. Draft shield 3. Standard weighing pan 4. Glass lid for the weighing
chamber 9. Weighing pan for filters
D balance for pipettes calibration 2. Draft shield 3. Weighing pan 4. Glass lid for the weighing chamber 5. Glass vessel 6. Evaporation ring 7. Glass lid for the weighing chamber, with an opening 8. Additional glass lid
Pipette calibration:: Run the balance, replace the weighing pan with a set of a weighing pan and a glass vessel (5). Wait for a stable indication, next press TARE button.
XA 5Y.M.A Microbalances
Install: – draft shield (1), – weighing pan (2).
3 – diode: ionizer on/off.
– 19 –
XA 5Y.M Microbalances
Install: – draft shield (1), – weighing pan (2).
XA 5Y.A Series Balances
A balance with d = 0.01 mg B balance with d = 0.1 mg
Remove transport lock (9) – press the transport lock down gently, turn it in
direction pointed by
– centring ring {embossing side upwards} (4), – draft shield (3) or (1), –
weighing pan (2).
5 – diode: ionizer on/off
Balance for pipette calibration Replace the draft shield and the weighing pan
with the chamber for pipette calibration, inside install: – weighing pan –
other components (see description below)
– 20 –
XA 5Y Series Balances XA 5Y.F Series Balances PM 5Y Series Balances
A balance with d = 0.01 mg B balance with d = 0.1 mg
Remove transport lock (9) – press the transport lock down gently, turn it
direction pointed by
Balance for pipettes calibration Replace the draft shield and the weighing pan
with the chamber for pipette calibration, inside install: – weighing pan –
other components (see description below)
Install all standard components:
– bottom weighing chamber shield (4), – bottom ring (3), – standard weighing
pan (1), – draft shield (2). Install additional accessories. Connect the
balance to the mains. To do it, use power supply that comes standard with the
balance.
Prior to filters weighing, remove draft shield (2), standard weighing pan (1),
and bottom shield (3). Next open the weighing chamber, inside place glass
shield (6), install pan-stand (5) intended for filter weighing .
Tare the balance and start filter weighing.
Install the weighing pan 1 for PM with d=0.01g or pan 2 for other PM.
Make sure that the pan is installed properly on the pan receivers: it cannot
touch the housing and must be stable.
Connect the indicator first (the connector is to be found at the back of the
balance housing), next connect the supplementary equipment.
Now, connect the weighing instrument to the mains (the power supply socket is
located on the left side of the indicator).
– 21 –
Caution: Be careful while installing the components so as not to cause any
damage to the balance mechanism. When connecting the indicator to the balance,
pay special attention to proper insertion of the plug into the socket and its
secure fastening in the socket by carefully screwing the securing nut.
3.4. Maintenance Activities
Electric shock.
Some balance models are equipped with an ionizer. Make sure to take special
safety measures during maintenance.
1. The balance must be disconnected from the mains before maintenance. 2.
Ensure that no liquid comes into contact with the power supply. 3. Never open
the balance housing, terminal, ionizer or power supply they contain no
internal components that can be cleaned, repaired or replaced by the user. 4.
Do not connect the balance to the mains until the cleaned parts are dry. 5.
Any maintenance activities or cleaning must be performed 5 minutes after
disconnection of the device from the mains. 6. The emitter pins are sharp. Do
not touch them directly as this may cause body injury.
The ionizer covers have to be cleaned at least once a month using a lint-free
cloth soaked with alcohol-based substance.
CAUTION: Cleaning the weighing pan while still installed may cause damage
of the measuring system.
1. Uninstall the weighing pan and other detachable components (the components
differ depending on a balance type see section: UNPACKING AND INSTALLATION).
Be careful while installing the components so as not to cause any damage to
the balance mechanism. Remove the weighing pan of a microbalance using
tweezers.
2. Using handheld vacuum cleaner remove dust from the weighing chamber
(recommended for the microbalance).
3. In order to ease cleaning of some of the weighing instrument components,
it is permissible to remove them following the below instruction.
Cleaning anti-draft chamber panes:
Select dissolvent depending on the type of contamination. Never soak the glass
panes in alkaline solutions since they interact with glass and may cause
damage. Do not use cleansers containing abrasive substances.
For organic dirt use acetone first, next use water or detergent. For other
than organic dirt use diluted acid solutions (soluble salts of hydrochloric or
nitric acid) or base solutions (ammonium or sodium base).
To remove ACIDS use protophilic solvent (sodium carbonate), to remove BASE use
protogenic solvent (mineral acid of various concentration).
– 22 –
In case of heavy contamination, use brush and detergent, nevertheless avoid detergents containing large and hard molecules which could potentially scratch glass panes.
At the end of the cleaning process rinse the pane using distilled water.
Use soft brush with wooden or plastic handle exclusively to avoid risk of scratches. Do not use wire brush.
Rinsing is a necessary cleaning process stage allowing to remove remaining soap, detergents and other cleansers from the panes prior to their reinstallation.
After the preliminary cleaning process stage, rinse the pane using running water first, distilled next.
Avoid drying the panes either using paper towel or forced air circulation since some fibres, grains or contamination of other type could permeate into the panes thus causing weighing errors.
We do not recommend using driers when drying measuring glass tools.
It is a frequent treatment to leave glass components on a rack to dry.
Cleaning powder-coated components:
For preliminary cleaning process stage you need running water or wet sponge with large pores, this will help you to remove loose, heavy dirt.
Do not use cleansers containing abrasive substances.
Next, using cloth and cleanser-water solution (soap, dishwashing liquid) gently rub the surface.
Avoid using cleanser without water since it may result in damage of the cleaned surface, please mind that large amount of water mixed with cleanser is a must.
Cleaning aluminium components:
While cleaning aluminium components use products acid by nature, e.g. spirit vinegar, lemon. Do not use cleansers containing abrasive substances. Avoid using hard brush, this may cause scratches. It is recommended to use microfibre cloth.
While polishing the surface use circular movements. Use clean, dry cloth to make the surface shine.
Cleaning stainless steel components When cleaning stainless steel components, the following table should be followed, which lists the types of contamination and ways of removing it.
Fingerprints
Clean with alcohol or dilutant. Rinse with clean water and wipe dry.
Oils, fats, greases Wash with organic solvents and then clean with warm water with soap or mild detergent. Rinse with clean water and wipe dry.
– 23 –
Temperature stains and discolorations Strong discoloration Traces of rust
Paints
Scratches on the surface
Wash with a gentle abrasive cleaner, clean lightly according to the direction
of the surface structure. Rinse with clean water and wipe dry.
Clean lightly according to the direction of the surface structure. Rinse with
clean water and wipe dry.
Moisten with oxalic acid solution and leave for about 15-20 minutes, then wash
with warm water with soap or mild detergent. Rinse with clean water and wipe
dry.
Wash with paint solvent and then rinse with warm water with soap or mild
detergent. Rinse with clean water and wipe dry.
Gently polish the surface with a non-woven fabric (iron-free) according to the
direction of the surface structure. Wash with a gentle abrasive cleaner. Rinse
with clean water and wipe dry.
Cleaning ABS components
To clean dry surfaces and avoid smudging, use clean non-colouring cloths made of cellulose or cotton. You can use a solution of water and detergent (soap, dishwashing detergent, glass cleaner). Gently rub the cleaned surface and let it dry. Repeat the cleaning process if necessary.
In the case of hard to remove contamination, e.g. residues of adhesive, rubber, resin, polyurethane foam etc., you can use a special cleaning agents based on a mixture of aliphatic hydrocarbons that do not dissolve plastics. Before using the cleanser for all surfaces, we recommend carrying out tests. Do not use cleansers containing abrasive substances.
Disassembly steps: MYA Series Balance
Open the chamber. · Carefully remove weighing pan (2). Remove draft shield
(3). · Undo draft shield sleeve (4). · Remove internal ring assembly (5). ·
Remove snap-in holder (6), lever it using e.g.
tweezers. · Close the chamber, press clamp (8) down
and move ring assembly (7) to the front and up.
– 24 –
Disassembly steps: MYA F balances
· Open and remove chamber lid (1). Remove weighing pan (2). · Press the
buttons marked with arrows to
release chamber for filter weighing (3) and remove it.
CAUTION: Pay special attention while assembling the chamber. Guide the pins so
that they are evenly inserted into the base openings (inaccurate inserting of
the pins will cause incorrect assembly and may cause balance damage).
Disassembly steps for XA 5Y.A, XA 5Y:
Remove the weighing pan and the draft shield. Clean the components when detached. With this, your balance mechanism is protected against accidental damage.
Undo and remove the top pane protection, next slide the pane out of a guide bar.
– 25 –
Remove the left and the right side pane. Before removing the panes completely, pull away the protection, see the picture on the left. The side panes are not interchangeable, therefore, remember which is left and which is right. Upon cleaning procedure completion, the panes must be installed on the correct sides.
Undo and remove the top frame protection, next disassembly the top frame.
Remove the front pane.
Clean the weighing chamber and the panes. All the operations have to be done
carefully. Pay special attention to the spot where the weighing pan is
installed; dirt and other small elements may enter the balance construction
through this opening. This may cause incorrect operation of the instrument.
Upon maintenance is completed, carry out the same set of actions in a reverse
order. Pay special attention to the left and right side panes, assembly them
on the correct side of the balance. While sliding the pane back onto its
place, remember to pull away the protection (like before, when disassembling
the anti-draft chamber).
3.5. Connecting the Balance to the Mains
The weighing device can be connected to the mains only with a power supply
that comes standard with the particular model. Nominal voltage of the power
supply (specified on the power supply data plate) has to be compatible with
the mains nominal voltage. Plug the balance to the mains connect the power
adapter to the socket, next connect its connector to the interface located at
the back of the balance housing.
3.6. Temperature Stabilization Time
Prior to the measuring processes start, it is necessary to wait until the
balance reaches thermal stabilisation. Balances stored in much lower
temperatures, than the workroom temperature, before being plugged to the mains
must be subjected to thermal stabilisation. The thermal stabilisation period
takes about 12 hours. During the thermal stabilization, the indications on the
screen can change. Potential workroom temperature change shall be
insignificant and occur slowly. Thermal stabilization also applies to the
monitoring system and automatic balance levelling.
3.7. Connecting Peripheral Equipment
Use only accessories and peripheral equipment recommended by the manufacturer.
The balance must be unplugged from the mains before connecting or
disconnecting any peripheral equipment (printer, PC, computer keyboard of USB
type or extra display). Upon connecting the peripherals, plug the balance to
the mains.
– 26 –
3.8. Balances with Integrated Ionizer
Safety notes contain important information on safety issues. Ignoring the
safety notes may lead to personal injury, damage to the instrument or
breakdown. Safety notes are marked with the following signal words and warning
symbols:
Do not touch ions source. This may cause electric shock (during operation) or
injury (components generating ions are sharp). Should you notice any signs of
incorrect operation (smoke, burnt material smell, irregular ionizer warming),
disconnect the device from the power supply immediately. Should water or any
foreign bodies get inside the ionizer, behave likewise, i.e. disconnect the
ionizer from the mains immediately.
Due to high voltage, typical for the ionizer, remain extremely careful while
operating it.
It is not allowed to either disassemble or modify the ionizer. Use the device
only as intended.
The antistatic ionizer, a built-in component of the XA 5Y.M.A.I microbalance
and the XA.5Y.A.I balance, neutralizes electric charges inside the weighing
chamber upon placing the sample in it. The lower the value of relative
humidity and a reading unit of the balance, the more evident influence of
static electricity. The ionizer neutralises any uncompensated electric charge
accumulated on the weighed object. Affected operation results in greater
repeatability error, often also greater error of indication. Measurement
errors, due to the harmful effects of electrostatics, often reach the value of
thousands of reading units and more. The problem concerns non-conductive
samples, e.g. glass flasks, paper filters, plastics, etc. Implemented in the
XA 5Y.M.A.I microbalance, the antistatic ionizer is a modern technology
solution which compensates both cations and anions. The ionizer activates when
the balance door opens, and deactivates when the door is closed. It is
possible to control the ionizer using quick access button.
Press to start/stop ionizer operation.
Press to activate.
Press to deactivate.
The operation is signalled by blue LED diode (installed at the back of the
weighing chamber) which emits light when the ionizer is on.
– 27 –
4. START-UP · Power the balance. CAUTION: There are two sockets on the
balance for connecting to the power supply. The indicator has one socket, and
the weighing mechanism housing has the other. The balance can be powered from
both the indicator and the weighing mechanism, but not simultaneously from
both sides, using two power supplies. Such a connection may result in damage
to the balance and power supplies. · After a while, the procedure of loading
the operating system with RADWAG software gets started. During start-up, the
signalling diodes and LEDs at the front of the indicator start blinking. ·
Upon balance start-up, the home screen is displayed, levelling process runs
automatically. During the automatic levelling, the feet engines operation is
heard. Respective messages are displayed.
Levelling range is limited. In the case of too great tilt of the surface
(counter, table top) where the balance rests, the levelling gets aborted and
the following message is displayed:
displayed automatically. · The balance runs with no user logged in. In order
to start operation, it is necessary
to log in (for detailed login procedure, read later sections of this user
manual).
– 28 –
Caution: Prior to the balance start, make sure that there is no load on the weighing pan. In accordance with the EN 45501 standard, verified balances cannot display mass value below -20e. If the indication value is below -20e,
Press
key to zero the balance.
If the software crashes during operation, carry out a hard reset. To do this, press and hold for about 5s the button on the indicator. The software will reset and restart.
PROGRAM STRUCTURE
Program menu is divided into function groups. A function group is a group of
interrelated parameters. For detailed description of each group, read later
sections of this service manual.
Function groups:
The main menu is accessed by pressing display.
button located in the bottom part of the
The menu comprises parameters related to balance settings, functions and profiles.
– 29 –
WEIGHING RESULT WINDOW
The home screen features 5 sections: · Top section displaying data on active
working mode, logged-in user, date, time, computer connection status. ·
Section presenting the weighing result and a current level status.
· Section comprising supplementary information on currently performed
operations (grey background).
CAUTION: Data contained in this workspace is freely configurable. For detailed
information, read section 13.4 of this user manual. · Section comprising on-
screen function buttons: CAUTION: The operator can define the on-screen
function buttons. For detailed information, read section 13.3 of this user
manual. · Section comprising on-screen function buttons:
– 30 –
7. LOG IN In order to access operator-related parameters and edit databases,
you need to log in as an operator with
First Log in: · Pull down the top menu, press
· Press
· The default password for
· Enter the password and press
button to confirm.
· Press
button to unmask the entered password.
· The home screen is displayed,
CAUTION: After first logging in, enter operators and assign appropriate permission levels and individual access passwords to them (for detailed procedure, read sections 10 and 11). While logging in again, select an operator from the list and enter the password, the program initiates operation with permission level set for the selected operator.
Log out operation: · Pull down the menu at the top of the display.
· Press < > field. · The home screen is displayed, logged in operator name is
replaced with <Not logged
in> entry.
– 31 –
OPERATING THE MENU
Operation of the program menu is intuitive. The touch panel makes the software
operation easy. Pressing a function button or an area on the display initiates
an assigned function or a process.
8.1. Operation Panel
Press to enter the main menu.
Press to confirm modification. Press to return to the previous view without
changing the parameter value.
Press to add a database record.
Press to search for a particular database entry by date.
Press to search for a particular database entry by name.
Press to search for a particular database entry by code. Press to print database entries.
Press to select printout template variables from the list.
Press to go back.
8.2. Voice Commands
The balance software allows several operations to be performed using voice
commands. These are: tare, zero, print/save measurement. The commands must be
spoken in English, in the correct sequence. Below is the structure of the
individual commands.
Tare: ellipsis [please] (tare | tar | terre) [the] device
Zero: ellipsis [please] zero [the] device
Print/Save: ellipsis [please] save [the] (mass | measurement | mass
measurement)
How to say the command: black text is mandatory, green – may or may not be
spoken, red one of the texts is mandatory to be spoken e.g: for Print command:
ellipsis save mass.
Procedure:
– 32 –
1. Place the < > button for activating voice commands on the button bar (for
how to add buttons, read further down this manual).
2. Press < >. 3. The voice command recognition procedure is initiated and the
user is informed about
it through the displayed messages.
4. Say the command as described above. 5. If the command is spoken correctly
and the balance program recognizes the
command, a message is displayed and the command is automatically executed.
– 33 –
8.3. Return to Weighing
Modifications introduced into balance memory are automatically saved upon return to weighing operation.
Procedure: · Press · Press
button repeatedly, keep pressing it until you see the home screen. button (top bar), the home screen is displayed immediately.
– 34 –
ADJUSTMENT
The 5Y series balances feature automatic internal adjustment system which
ensures correct measurement accuracy.
9.1. Internal Adjustment
Internal adjustment is carried out by means of an internal adjustment weight.
Caution: Balance adjustment procedure requires stable conditions (no air drafts, no vibrations). In the course of adjustment the weighing pan must be unloaded.
9.2. External Adjustment
External adjustment is carried out using external mass standards of the right accuracy and weight value, which value depends on balance model and capacity. The process is carried out semi-automatically, successive process stages are signalled with prompts.
Caution: External adjustment is possible for balances that are not subject to the conformity assessment (verification).
Procedure: · Enter
button. The balance determines start mass, message: Start mass determination is displayed. · Upon completed start mass determination, message ordering you to load the weighing pan is displayed. Load the weighing pan with weight of respective mass
value, and press
button.
· For some balance types, the next message is displayed. Load the weighing pan with
weight of subsequent mass value, and press
button.
· Upon completed procedure, unload the weighing pan and press confirm. Balance returns to the weighing process.
button to
9.3. Adjustment Test
– 35 –
9.4. Automatic Adjustment
Enter this menu to specify conditions initiating an automatic adjustment.
Options: · None automatic adjustment inactive. · Time adjustment takes
place in time intervals declared in
Caution: This parameter’s settings can be modified only for balances that are
not subject to the conformity assessment (verification).
9.5. Automatic Adjustment Time
9.7. GLP Project
GLP is one of many methods for documenting work in accordance with adopted quality system. Data selected for printing is printed on each release of an adjustment report. Information and signs to be used in a GLP report:
adjustment (adjustment type) date balance type operator level status current mass nominal mass 2 difference 2 dashes signature
working mode (working mode name) time balance S/N first and last name nominal mass difference current mass 2 temperature empty line non-standard printout
– 36 –
9.8. Adjustment History
Adjustment history contains data on all carried out adjustment processes. The
record is carried out automatically. Each adjustment record comprises basic
data on completed process. This menu enables displaying the list of completed
adjustment processes. Each report is printable. In order to print an
adjustment report, enter
<
print button.
Caution: If the balance memory is full, then the oldest record gets automatically deleted.
If internal procedures of an organization require maintaining complete documentation concerning all carried out adjustment processes, then the list with adjustment records has to be printed and archived periodically.
Searching for adjustment record It is possible to search for a specific completed adjustment record. To do that,
press <
button and enter date of an adjustment process that is to be found.
Exporting data on completed adjustment process
Connect a USB flash drive to the balance USB port. Press <
button located in the
upper right corner of the display. The process is fully automatic and on its completion a file
with .db. extension is saved to a USB flash drive connected to the USB port.
10. OPERATORS
Operators database features list of users permissioned to operate the balance. The following data can be defined for each balance operator:
ID Code First and last name Active account Default profile Theme
Name Password Permissions Language Card number
CAUTION: Only a user with Administrator status can add new operators or delete operators from the database.
In order to add an operator:
· Enter
button.
– 37 –
Caution: It is possible to search for an operator in a database of operators
by code or name. Log in using a transponder card: Caution: The RFID reader
mounted in the indicator operates at a frequency of 13.56MHz and complies with
ISO/IEC 14443 Type A. Only cards of this standard will be recognised by the
reader.
· Enter Operator settings. · Select
· Apply the card to the RFID reader. · The card number is automatically
entered in the editing field.
– 38 –
· Press
button, the card number is added to the operator settings.
· Exit to the home screen. From now on, applying this card to the RFID reader
will automatically log the user in.
Log in using a fingerprint: Caution: Only the fingerprint reader listed as a
dedicated accessory for these balances can be connected to the balance. A list
of accessories can be found on RADWAG website.
· When the fingerprint reader is connected to the USB type A socket, the
– 39 –
· Scan the fingerprint and repeat the procedure 7 times (as described in the
message box).
· When the procedure is completed correctly, a summary window is displayed.
· Confirm the procedure by pressing
button.
· From now on, if a fingerprint reader is connected to the USB port, a pictogram of the
active fingerprint login option is displayed in the login window.
· After placing a finger on the reader and confirming the correctness of the
fingerprint with the stored pattern, the pictogram briefly turns green, the
user is automatically logged in, and the software switches to the home screen.
The name of the logged-in user is displayed in the top bar of the window.
· If the scanned fingerprint does not match the stored pattern, the pictogram
briefly turns red, the user is not logged in, and the software displays the
login window.
Log in using a face profile: · To add a face profile for a user, go to the
option presented below
– 40 –
· After starting the option, a message is displayed and the face profile is read; you must position yourself correctly in relation to the camera so that the program can save the profile.
· Once the profile has been saved, a message is displayed along with a picture from the camera.
· Press
button to close the widow.
· The option description in the user settings changes to <Delete face profile>,
indicating that the profile is stored in the balance memory.
· From now on, if the user for whom the face profile was added logs in, after
entering the logging window, the program will automatically read the profile,
– 41 –
which will be signalled by the pictogram on the left side of the window
· The program automatically reads a photo of the user’s face profile, and
after the photo is read, the camera turns on (which is indicated by the
lighting of a diode next to the camera) and compares the saved photo with the
profile read by the camera. If these images match, the main program window
will open and the name of the logged-in user is displayed in the top bar of
the window.
Editing data on an operator: · Press field with operator name. · Data on a
specific operator is displayed. · Select and change necessary data.
To delete an operator: · Press and hold operator name. · Database menu is
displayed. · Select
– 42 –
11. ADMINISTRATOR PANEL This menu enables to determine: activities to be
carried out by an operator with certain permissions, password strength and
permissions for unlogged operator. CAUTION: Only operator with Administrator
permissions can modify permissionrelated parameters.
11.1. Password Settings This submenu is designed to set password
sophistication rate.
Minimum password characters quantity Lower case and upper case letters required Digits required Special characters required Password validity period
Determining quantity of characters in operator’s password. For ‘0’ value set
any quantity of characters. Parameters specifying requirements regarding
password characters type and format.
Determining time interval in days after which the password has to be changed.
For ‘0’ entry password change is not required.
11.2. Operator Accounts Settings
– 43 –
Unlogged operator permissions Administrator can assign an unlogged balance
operator (so called anonymous operator) with permissions level.
Procedure: Enter
11.3. Permissions Management
Caution: When the
– 44 –
Parameter values:
Databases
Default balance settings enable an operator who is logged as Administrator to
modify settings of particular databases. The values can be changed depending
on the needs. The editing permissions for the other options available in this
menu can be changed on a similar basis. 12. PROFILES A profile is a data pack
that determines:
particular function operation, e.g. ‘Parts counting’, ‘Percent weighing’,
etc., data to be displayed during operation, function keys to be active,
measuring units to be accessible, criteria determining speed of operation and
measurement stability, Balance software allows you to create numerous
profiles, this enables: designing an individual working environment by each
operator, customized balance operation, i.e. activating selected function keys
and information
(improved operation performance). For ease of operation, 4 default balance
profiles have been designed. The profiles settings provide optimal weighing
process for particular requirements and conditions.
– 45 –
Profile types: · Fast profile set to enable fast measurement of mass of any
size regardless of the selected working mode. Fast profile gets activated
automatically at the first balance start-up. For Fast profile, the parameters
settings enable obtaining the final result as fast as possible. · Fast dosing
profile intended for dosing, it is set to enable fast dispensing of product.
For this profile selected, the DOSING mode gets activated automatically, ·
Precision profile set to enable precise measurement of mass of any size
regardless of the selected working mode. For this profile the weighing process
takes a lot of time, but this guarantees the most precise weighing result. ·
User basic profile, filters settings provide quite fast and precise
measurement. For this profile selected, the WEIGHING mode gets activated
automatically.
Caution: As an operator, you can modify all profile settings only for User
profile.When it comes to the remaining default profiles (Fast, Fast dosing and
Precision), they can only be modified partially. Each additional profile you
created can be modified in terms of all settings.
12.1. Creating a Profile
The Administrator can create new profiles by: · Copying an already existing
profile and modifying it, · Creating a new profile.
Copying an existing profile
Procedure:
· Enter the main menu, to do it press
button.
· Enter <
submenu.
· Press and hold the profile that is to be copied.
· Menu referring to the profile record is displayed. Select
· A new
are identical.
· Modify necessary data: (name, etc.).
Adding a new profile
Procedure:
· Enter the main menu, to do it press
button.
· Enter < > submenu.
· Press
button.
· A new profile is added automatically, it is opened in an editing mode.
– 46 –
Deleting a profile Procedure:
· Enter the main menu, to do it press
button.
· Enter < > submenu. · Press and hold the profile that is to be deleted. ·
Menu referring to a profile record is displayed. Select
· Press
button to confirm, the profile gets deleted.
Caution: In order to operate profiles, it is necessary to log in as an Administrator.
12.2. Profile Structure
Profile parameters list:
Settings
Enter this menu to set an individual profile name (an order of
alphanumeric characters), and to declare a default working mode (the
selected mode is activated as default upon selecting the profile).
Working modes Readout
Units
Option to set specific functions for individual working modes. Mode
submenus: Mode related settings Buttons Information Printouts
Readout submenus: Filter Value release Autozero Autozero: Dosing Last digit
Last digits quantity Ambient conditions
Enter this menu to declare the start unit, the supplementary unit, 2
custom units, to enter the value of gravitational acceleration force in
balance’s operation place and to declare availability of individual units.
12.2.1. Settings
Name Upon entering this parameter, a message box with an on-screen keyboard is
displayed.
Enter the name of a profile and press to the profile.
button to confirm. The name is assigned
Default working mode Upon entering this parameter, you can select a specific
working mode that is to be set as default. For
– 47 –
12.2.2. Working Modes
Upon entering Working Modes parameter, a window containing all accessible
working modes is displayed. Parameters are set for each working mode
independently. Particular profile when activated runs with a given set of
parameters.
List of parameters common for each working mode: Settings: specific parameters
relating to a working mode, and universal settings such as: result control,
tare mode, automatic footer printout, printout mode, printout. Functions of
quick access buttons: declaring quick access buttons to be displayed.
Information: declaring information to be displayed. Printouts: declaring type
of a printout or defining a non-standard printout.
12.2.3. Readout
You can adjust the balance to ambient conditions (filter level) or to your own
needs.
FILTER (function disabled for Fast, Fast dosing, Precision profiles) Each
measurement signal, before being displayed, is electronically processed in
order to provide correct parameters specific for a stable measurement result
(ready to be read). You can determine signal processing rate, there are five
different filters: Range:
· very fast, fast, average, slow, very slow. While setting the filter,
consider the actual operating conditions. In the case of very good operating
conditions, select average or fast filter, in the case of rough conditions,
select slow or very slow filter.
CAUTION: For precision balances, it is recommended to select very fast, fast
or average filter÷, For analytical balances and microbalances, it is
recommended to select average,
slow or very slow filter÷.
Value release (option disabled for Fast, Fast dosing, Precision profiles) The
parameter has been designed to determine when the stability marker is to be
displayed.
There are three different settings: · fast, fast and reliable, reliable.
Caution: Both parameters, filter and value release, determine how long it
takes to obtain stable result.
Autozero function
The function has been designed to enable automatic control and correction of zero
indication.
If the function is enabled, the subsequent measurement results are compared to each other
in constant time intervals. If the results differ less than declared AUTOZERO range, e.g. 1
division, the balance is automatically zeroed, and the pictograms of stable indication
and precise zero
are displayed. Active Autozero function means, that each
measurement starts with the precise zero. There are, however, some cases when this
function can be a disturbing factor for the measuring process, e.g. very slow placing of a load
– 48 –
on the weighing pan (load adding, e.g. pouring, filling). In such case, the system correcting
zero indication may also correct the actual indication of a load placed on the weighing pan.
Values:
NO – autozero function disabled.
YES – autozero function enabled.
Autozero function: Dosing (option disabled for Fast, Fast dosing, Precision) The function has been designed to enable/disable autozero function for Dosing. Options: NO – autozero function disabled automatically when Dosing mode is run. YES – autozero function enabled automatically when Dosing mode is run.
Last digit The function has been designed to enable/disable visibility of the
last decimal place of measurement result. Values:
– Always: all digits are displayed. – Never: last digit disabled. – When
stable: last digit displayed only when the result is stable.
Last digits quantity This function enables declaring the number of hidden last
digits of the result. The function has three settings and works in connection
with the
– 1: the last digit of the result. – 2: the last two digits of the result. –
3: the last three digits of the result. Ambient conditions Parameter relating
to ambient and environmental conditions of the workstation. For unstable
ambient conditions, it is advisable to set the parameter to ‘Unstable’ value.
By default, the parameter is set to ‘Stable’ value. ‘Stable’ value results
with faster operation, i.e. weighing takes much less time than in the case of
‘Unstable’ value set. – Unstable; Stable.
12.2.4. Units
For a selected profile, you can declare a start unit, supplementary unit and
two custom units.
Units menu enables you to enter the value of gravitational acceleration force
characteristic for a particular place of use. It is obligatory should the
balance be used to determine mass in [N].
– 49 –
Custom unit features: o Formula (custom unit recalculation formula):
Coefficient * Mass or Coefficient / Mass. o Coefficient (coefficient for
custom unit recalculation) o Name (unique unit name to be displayed next to
the result – max 3 characters)
Newly designed custom unit name is added to the list of accessible measuring
units (refers to non-verified balances exclusively). 13. WORKING MODES
General Information 5Y series balances feature the following working modes:
Weighing Means of operation: weight of a load is determined through an
indirect measurement. The balance measures gravitational force which attracts
the load. The obtained result is processed and displayed on the balance screen
in a digital format. Checkweighing Means of operation: control of sample’s
mass carried out with reference to specified thresholds, it is necessary to
give the value of min threshold .
Parts counting Means of operation: based on a weight value of a single piece,
it is possible to count pieces quantity; necessary conditions: the weight
value of a single piece is determined with sufficient accuracy, the mass of
all the pieces is the same.
Dosing Means of operation: it is necessary to specify sample’s target mass to
be obtained by dispensing (pouring, filling, etc.).
– 50 –
Percent weighing Means of operation: control of percent ratio of a test sample in relation to a reference sample. Obtained data provides info on how much the test sample differs from the adopted reference sample. Animal weighing Means of operation: mass measurement carried out with use of filters which dampen animal moves on a weighing pan; the filters enable obtaining correct measurement result. Statistics Means of operation: carried out measurements are used to calculate statistical data such as Min, Max, deviation, etc.
Density Means of operation: based on Archimedes principle, it is possible to
determine density of solids and liquids. Density mode requires use of an
optional density determination kit.
Formulations Means of operation: by mixing number of ingredients you can
obtain particular mixture; in order to program a given formulation, it is
necessary to give mass values for respective ingredients. Pipette calibration
Means of operation: calibration of pipettes in accordance with ISO 8655
standard procedure or with regard to user requirements.
Differential weighing
Statistical quality control
Means of operation: analysis of change of SQC is a mode useful in the course of packing
sample mass over time.
process. It aims at monitoring and/or
controlling product packing. It enables
detecting product deficiency and excess.
Peak hold
Means of operation: freeze of max weight
value on a display (max. indication
reflecting the greatest force exerted on a
pan).
Packaged goods control
Mass control
Control of packaged goods in accordance Fast statistical control of samples in
with Packaged Goods Control regulation. accordance with requirements of quality
(function disabled for standard balance
systems and/or internally adopted standards.
models).
(function disabled for standard balance
models)
Particular working modes feature specific settings. The settings enable adapting mode
operation to your individual needs. The specific settings are activated upon selecting a
respective profile. For detailed description of specific working mode settings, refer to the
given working mode section.
13.1. Running Working Mode
To run working mode other than currently operated one: · Slide down the menu at the top of the display.
· Press
– 51 –
· Select the mode that is to be operated. 13.2. Working Mode Parameters
Each working mode has programmable parameters determining its functioning. To
access particular working mode parameters:
1. Slide out the menu on the left.
2. The following menu is displayed: ·
Another way to add or change a button is to press and hold a field in the
button bar for about 3s.
A window with the available buttons is displayed automatically.
Press the respective button; the software returns to the home screen, and the
selected button appears in the place where it was activated.
Caution: You can view the user manual saved to the balance memory. In order to take
advantage of this function, add Help
button to the tool bar.
– 53 –
13.3.1. Proximity Sensors The balance is equipped with two proximity sensors
which enable touch free control. The program detects four motions performed
around the sensors:
1. Hand in a close vicinity to the
List of operations to be assigned: None; Profile; Adjustment; Zero; Tare; Set
tare; Disable tare; Enable tare; Packaging; Print; Header printout; Footer
printout; Unit; Universal variable 1; Universal variable 2; Universal variable
3; Universal variable 4; Universal variable 5; Accept; Abort; Operator; Left
door; Right door; Open/close door; Parameters; Product; Warehouse; Customer.
Upon completed configuration procedure, the software runs function assigned to
a particular proximity sensor, having detected motion around it. To provide
correct operation, it is necessary to set respective proximity sensors
sensitivity and delay (read section 33 of this user manual).
13.3.2. Automatic Door Settings for devices equipped with automatic door
opening and closing mechanism. The settings concern quick access buttons and
proximity sensors.
MYA 5Y series balance: Open/close door parameter for changing door state;
e.g. door closing when the weighing chamber is open, and door opening when the
weighing chamber is closed. Left door changing position (open/close state)
of the door on the left side of the weighing chamber. Right door changing
position (open/close state) of the door on the right side of the weighing
chamber.
XA 5Y.A series balance: Open weighing chamber door opening. Close weighing
chamber door closing. Open/close door parameter for changing door state,
e.g. door closing when the weighing chamber is open, and door opening when the
weighing chamber is closed.
To swap from automatic to manual door opening mode in the XA 5Y.A balance: 1.
Turn the knob by 180 degrees into the position as shown in the picture below
(with the groove milled in the upper position).
To swap from manual to automatic door opening mode in the XA 5Y.A balance: 1.
Turn the knob by 180 degrees into the position as shown in the picture below
(with the
groove milled in the bottom position).
2. Move the left door with your hands back and forth to couple with the
mechanism.
Door coupling is easy to be sensed (the pawls get engaged with the notch).
– 54 –
13.4. Information
Information related to the weighing process is displayed on the left side of the the display.
The workspace can contain up to six data. Mind that if you select more, only the first six of
them are displayed.Each parameter features two options:
–
data enabled.
–
data disabled.
13.5. Printouts
Printout menu comprises two separate sections: standard printouts and non-
standard printouts.
Standard printout
It comprises three internal sections containing different variables. For each variable, set
variable printout enabled or
variable printout disabled.
Procedure:
1. Press name of a selected template (Header GLP Printout Footer) and select variables that are to be printed.
2. If a printout is a non-standard one, create it.
– 55 –
HEADER – Dashes – Working mode – Date – Time – Balance type – Balance S/N – Operator – First and last name – Level status – Customer – Warehouse – Product – Packaging – Universal variable 1 … 5 – Empty line – GLP report – Non- standard printout
GLP PRINTOUT – N (measurement number) – Date – Time – Level status – Customer – Warehouse – Product – Packaging – Universal variable 1 … 5 – Net – Tare – Gross – Current result – Supplementary unit – Mass – MW-Method – Reference tare – Minimum sample – Minimum sample status – Mass value for a titrator – ID – Non-standard printout
HOW TO OPERATE PRINTOUTS
FOOTER – Working mode – Date – Time – Balance type – Balance S/N – Operator – First and last name – Level status – Customer – Warehouse – Product – Packaging – Universal variable 1 … 5 – Dashes – Empty line – GLP report – Signature – Non-standard printout
1. In order to print variables of the GLP PRINTOUT section, press button on the
operation panel; only variables with
attribute can be printed (see the above
list).
2. Variables with
attribute, present in the HEADER and FOOTER, are printed
ONLY upon pressing PRINT HEADER and PRINT FOOTER pictograms. The
pictograms have to be added to the bottom bar of the display as quick access buttons.
(For detailed instruction on how to add quick access buttons to the bottom bar of the
display, go to section 13.3)
Print header data
Print footer data
CAUTION: Measuring units for mass indication:
– Net the main measuring unit (adjustment unit) – Tare the main measuring
unit (adjustment unit) – Gross the main measuring unit (adjustment unit) –
Current result currently displayed measuring unit – Supplementary unit
supplementary measuring unit – Mass the main measuring unit (adjustment
unit) Non-standard printout Non-standard printout may contain: TEXTS and
VARIABLES (acquired from the software during printout). Each non-standard
printout is a separate project, featuring a specific name by which it is
identified and saved to the database of printouts.
– 56 –
Procedure: 1. Press
2. Press
variables.
CAUTION: It is possible to add a printout by import of configured texts from
flash drives connected
to balance’s USB port. Printout’s name is ONLY a NAME, and it is not part of
its content. For information on how to design a non-standard printout
template, read section 39.19
Printouts.
13.6. Profiles
Profiles function has been described in section 12 Profiles.
14. WEIGHING
Load the weighing pan. Wait for a stable indication (stability marker, ) and read the measurement result.
Record/printout of the measurement result is available upon pressing < PRINT> key: · for verified balances only stable measurement result is saved or printed (stability marker visible on the balance’s display), · for non- verified balances stable and unstable measurement results are saved or printed (regardless of the stability marker absence). For unstable measurement result, a question mark <?> is printed next to the mass value.
14.1. Weighing Unit Selection
Change of the weighing unit is carried out by pressing the weighing unit pictogram visible
next to the measurement result. Upon pressing the weighing unit, a list of available units is
displayed. Select the unit you need, the software automatically recalculates the indicated
value.
Options:
unit
symbol
verified balance
unit
symbol
verified balance
gram
[g]
yes
Taele Singapore
[tls]
no
milligram
[mg]
yes*
Taele Taiwan
[tlt]
no
carat
[ct]
yes*
Taele China
[tlc]
no
pound
[lb]
no
Momme
[mom]
no
ounce
[oz]
no
Grain
[gr]
no
ounce Troy
[ozt]
no
Tical
[ti]
no
pennyweight
[dwt]
no
Newton
[N]
no
Taele Hongkong
[tlh]
no
Mesghal
[msg]
no
- – units accessibility is conditioned by the balance type. For non-verified balances, all non-SI units are available.
– 57 –
14.2. Good Weighing Practice
To assure long-term operation and correct mass measurements, follow the rules
presented
below: Start the balance with no load on the weighing pan (permissible load
value on the
weighing pan at start is ±5% of the maximum capacity).
YES
NO
YES
NO
Load the weighing pan steadily avoiding
shocks: Go to Setup/Misc./Vibrations
detection parameter and enable control of
weighing pan loading (for detailed
information, read section 33.8).
NO
NO
Place weighed loads centrally on the weighing pan.
Avoid side loading, in particular side shocks.
Prior to weighing or in the case of ambient conditions change, it is necessary to carry out balance adjustment.
· Prior to the measurement, it is recommended to load the weighing pan several
times with mass close to the maximum capacity. In the case of instruments
offering two reading units (d), the initial load of the weighing pan depends
on the reading unit value. Example: For the XA 82/220.5Y balances with d1=
0.01mg and d2=0.1mg, load the weighing pan as follows: 50 g weight for
measurements with d1=0.01mg and 200 g weight for measurements with d2=0.1mg.
· Make sure that the unloaded weighing instrument indicates zero ( pictogram)
and check whether the measurement is stable ( button.
pictogram); if not, press
· Select weighing unit: press the unit next to the result
or
press
w button at the bottom of the display.
Caution: In accordance with the EN 45501 standard, verified balances cannot display mass value below -20e. If the indication value is below -20e, <Lo mass> information is displayed.
– 58 –
Press
button to zero the balance..
Do not unplug the balance from the mains between measurement series. It is recommended to turn off the display by pressing the button located in the top menu.
To turn the balance on, press the button on the indicator. Once the start-up procedure is complete, the balance is ready for next weighings.
14.3. Levelling
Ultra-microbalances and microbalances, and the A- class microbalances and
analytical balances are equipped with Reflex Level System, which allows
monitoring of the balance level and its automatic levelling.
It guarantees: · ground tilt test, · fully automatic balance levelling, ·
balance level control, · confirmation of balance levelling printed on
adjustment reports and weighing reports.
Upon switching the balance on, the levelling takes place automatically. Each
time before the internal adjustment, the level is checked. Wwhen the balance
is not levelled, the internal adjustment takes place only upon prior
levelling.
Procedure of user-performed balance levelling:
· Press level status pictogram< >, it is displayed in the upper right corner
of the operation panel.
· Control panel of levelling function is displayed. Along with the level
indicator, balance view is presented.
· Press <
button levelling process runs automatically; during the
automatic levelling, the feet engines operation can be heard.
· To abort the levelling process, press <
button.
· When the marker takes central position, its colour turns from red to green <
the balance is correctly levelled.
– 59 –
· Press
button to return to the home screen.
Remaining balances are equipped with AutoLEVEL System. This solution enables continuous monitoring of the balance level during operation, which is signalled in the upper right corner of the display. If the level changes, the system signals it by changing the position of the level indicator and/or by triggering an alarm and going to the balance levelling screen.
14.4. Balance Zeroing
To zero mass indication, press
key. Zero indication and the following pictograms are
displayed: and . Zeroing operation means determining a new zero point, recognized
by the balance as precise zero. The balance can be zeroed only when the indication is
stable.
Caution: Indication can be zeroed only within ±2% range of the maximum capacity. If the zeroed value is above ±2% of the maximum capacity, then the software indicates a respective error message.
14.5. Balance Taring
To determine net weight value, load the weighing pan with a packaging, wait for a stable
indication and press
key. Zero indication and the following pictograms are displayed:
Net and . Upon weighing pan unloading, the sum of tared masses with minus sign is
displayed. You can assign a tare value to a particular product stored in the database. In such
a case, the assigned tare value is automatically acquired upon selection of the given
product.
Caution: It is impossible to tare negative values. When you try to tare negative values, the weighing instrument responds with an error message. In such a case, zero the indication and repeat taring procedure.
Entering Tare Value Manually Procedure:
· Press
quick access button. The button is available regardless of the working
mode you operate.
· An on-screen numeric keyboard is displayed.
· Enter tare value and press
button.
· The balance returns to weighing mode. Tare value with ,,” sign is displayed.
Deleting Tare The displayed tare value can be deleted by pressing ZERO key on
the operation panel, or by using programmable
Procedure 1 – upon taking the tared load off the weighing pan:
· Press
button,
· the
– 60 –
Procedure 2 – with tared load resting on the weighing pan:
· Press
button,
· the
· when the tare value exceeds 2% of the maximum capacity, -Err- message is displayed
in order to inform you about the fact that operation cannot be carried out.
Procedure 3 – when the tared load is on the weighing pan or upon taking the tared load off the weighing pan:
· press <
Deactivate tare> programmable button,
· the NET marker is deleted,
· the display indicates tare value,
· to restore the last tare value press <
Restore tare> button.
Use of Air Buoyancy Compensation Factor
The application enables correcting mass measurement errors in the course of:
- weighing loads of density considerably different from the density of a standard that has been used to adjust the balance. Typically, the balance is adjusted with a steel mass standard of ~8.0g/cm3 density or a brass mass standard of ~8.7g/cm3 density. If loads made of other substances are weighed, then ratio indicated on below graph applies;
Error value versus density of weighed sample
The graph presents correction values for mass depending on the density of weighed load, assuming the air density to be a constant value of 1.2 kg/m3..
2. testing sample mass change within a few hours long period of time for a
case when the sample mass is relatively constant (minor changes); it is
assumed that air density changes affect the final result considerably, wherein
the density is influenced by changes of pressure, temperature and humidity.
For measurements to be reliable, the density of the air in the laboratory and
the density of the material to be weighed must be determined (depending on the
ambient conditions, the measurement method and the properties of the material
under test).
– 61 –
FUNCTIONING The air buoyancy compensation procedure is to be used as follows:
1. Method 1: Enter air density value and the density value of the weighed
sample to the balance memory. Wait, the software automatically calculates the
correction factor for weighed mass. Next, the corrected value of weighed mass
is displayed. To avoid mistakes, the <!> symbol is presented next to the
corrected mass value (on a display and a printout).
2. Method 2: Determine air density value using the balance program (semi-
automatic operation) and enter the value of the weighed sample density to the
balance memory. Determination of air density is carried out using a set of two
mass standards, one made of stainless steel, and the other one of aluminium.
Based on mass indications of the two standards, the program automatically
calculates the air density and saves the value in the balance memory on
operator’s confirmation. Now, enter the density value of the weighed sample to
the balance memory. Wait, the software automatically calculates the correction
factor for weighed mass. Next, the corrected value of weighed mass is
displayed. To avoid mistakes, the <!> symbol is presented next to the
corrected mass value (on a display and a printout).
In order to enable/disable the air buoyancy compensation procedure, go to the
user menu. The procedure only works in the weighing mode.
AIR BUOYANCY COMPENSATION START-UP
Procedure: · Enter home screen and press grey workspace. · Select
Parameter values: VALUE select to open a window to enter a specified air density value (e.g.: determined by means of other methods). The entered value is used in the course of compensation. It is assigned automatically upon completed air density determination (to complete air density
determination, press <
button).
ONLINE select to make the balance acquire air density value from a connected THB
sensor at a real time, if such is connected to the balance, or from an internal sensor installed
in the balance.
If your balance features both types of sensors (THB and internal), then data from the THB
supersedes the other. In such a case, go to Setup/Ambient conditions/Ambient
conditions module parameter and set SAVE or SAVE & ALERTS value.
If your balance features only one set of sensors recording temperature, humidity and
pressure, then correct ONLINE operation requires setting the following parameters:
– internal sensors exclusively go to Setup/Ambient conditions/Ambient conditions
module parameter and set NONE value,
– 62 –
– external THB sensor exclusively go to Setup/Ambient conditions/Ambient
conditions module parameter and set SAVE or SAVE & ALERTS value. For correct
balance cooperation with the THB sensor, it is necessary to set port’s
transmission parameters. They must be compatible with those of the THB sensor.
The THB sensor’s transmission parameters are detailed on its data plate.
On returning to the weighing mode, the display indicates an additional symbol
<!>, as
indicated in the figure below. From now on, the value of indicated mass is
corrected, wherein the air buoyancy compensation factor and sample density are
referred to.
For the result to be corrected by the correct value, the correct values of air
density and density of the weighed sample must be entered into the balance
memory.
Caution: If the
14.6. Additional Parameters for a Weighing Process
Modification of the weighing functions enables programming balance operation.
Procedure: · Slide out the menu. · The following submenus are displayed:
Settings, Buttons, Information, Printouts. · Enter particular submenu and
select the component that is to be modified
· AUTOTARE Means of operation: Each first stable measurement result is tared. The NET sign is displayed. Now you can determine the net mass of the weighed load. Upon unloading the weighing pan and autozeroing, the tare value is automatically deleted.
– Automatic footer printout Available options: MODE None manual footer printout, Sum of measurements footer is printed when mass value entered in
– Printout / Release mode
· PRINT/RELEASE BUTTON (manual operation)
Never
printout disabled.
First stable
first stable measurement is recorded.
Each stable
all stable measurements are recorded.
Each
printout of all measurements (stable and unstable); for
verified balance, only stable measurements are printed out (as
in
· AUTOMATIC MODE
Never
printout disabled.
First stable
first stable measurement is recorded upon loading of the
weighing pan with a load; record of the next stable
measurement result takes place only if the following conditions
are met: weighing pan gets unloaded, measurement value
lower than the value of the set threshold is displayed, a new
load is placed on the weighing pan.
Last stable
last stable measurement, registered prior to unloading, is
recorded. The record takes place only if the following
conditions are met: weighing pan gets unloaded, measurement
value lower than the value of the set threshold is displayed.
With interval
select this value to start automatic printout and record of
indications to the WEIGHINGS database, carried out
periodically in a specified time intervals. The interval is set in
[min] in
min 59 s.
Caution: Each weighing result is printed and recorded (stable and unstable for a non-
verified balance; stable for a verified balance).
Automatic operation with interval starts at the moment of switching the function on. The
first stable weighing result of a value greater than the THRESHOLD value is printed and
recorded as the first measurement. The following measurements are printed with
frequency set in INTERVAL parameter. To stop the automatic operation with interval,
switch the option off.
– 64 –
THRESHOLD:
Mass value obligatory for operation of automatic printout, set in grams.
INTERVAL
Record frequency for automatic operation with interval
–
Printout
Printout type related to a working mode. Printout is activated by pressing PRINT
key on the balance operation panel.
Available options:
· STANDARD PRINTOUT:
Enables declaring content of the following printouts: HEADER, GLP
PRINTOUT and FOOTER.
Select
printout-activating function key.
· NON-STANDARD PRINTOUT:
Enables selecting one of database-stored non-standard printouts from a list
of
automatically added to the database.
Caution: For detailed instruction, refer to section 13.5 of this user manual.
– Air buoyancy compensation Parameters allowing you to switch on the compensation and to enter weighed sample density data and air density data.
Caution: Function valid for the weighing mode exclusively. For detailed information regarding the operation and settings, refer to the previous section of this user manual.
– Minimum Sample
Weighing mode settings comprise
– 65 –
Available options: · METHOD Option allowing you to designate used quality
standard. Press METHOD field to open a window with a list of database-stored
methods, by means of which the minimum sample were determined. In order to add
a new method, go to <Databases /Minimum sample>. · MODE Block select to
display pictograms informing you on status of weighed mass (weighed mass
below/above minimum sample) and to disable accepting the measurement result if
it is below the minimum sample limit. Warn select to enable display of
respective pictograms informing you about the measured mass (whether it is
lower than the MS value or higher than the MS value). You can accept a
measurement that is lower than the MS value. Mind that on a printout it will
be preceded by an asterisk ().
Caution: For information on how to enter a method for minimum sample, go to
section 29.14 <Databases /Minimum sample> of this user manual. New minimum
sample values can be entered by an authorized RADWAG personnel exclusively.
Weighing with use of
1. Press grey workspace area. 2. The following submenus are displayed:
Settings, Buttons, Information, Printouts,
Profile. 3. Select
a window with a list of database-stored methods opens. 6. Select respective
method. 7. Previous menu is displayed. 8. Press
Block select to display pictograms informing you on the status of weighed
mass (weighed mass below/above minimum sample) and to disable accepting the
measurement result if it is below the minimum sample value. Warn select to
enable display of respective pictograms informing about the measured mass
(whether it is lower than the MS value or higher than the MSvalue). You can
accept the measurement that is lower than the MSvalue. Mind that on a printout
it will be preceded by an asterisk (
– 66 –
Pictograms providing MS-related information: Weight value lower than the
specified MS value;
Weight value higher than or equal to the specified MS value. Weight value
lower than the specified MS value. A pictogram with a clock informs on expiry
of the minimum sample validity period (it is displayed two weeks before the
determined expiry date). Weight value higher than the specified MS value. A
pictogram with a clock informs on expiry of the minimum sample validity period
(it is displayed two weeks before the determined expiry date). The validity
period of the minimum sample has expired. The minimum sample settings need to
be modified. The minimum sample settings can be modified by an authorized
RADWAG personnel exclusively.
Caution: If more than one reference tare value is programmed (with
corresponding minimum sample value) then the indicated value automatically
moves to a range corresponding to the weight of the tared container. At the
same time, the required minimum mass is changed.
14.7. Cooperation with Titrators
In order to provide correct cooperation with TITRATORS, go to standard
printout content settings and set
Usually the vibrations are dampened with use of anti-vibration tables. A
signalling device, i.e. ambient conditions module equipped with a vibration
sensor, is used to support the process of vibrations dampening. Such device
comes standard with each 5Y series balance. With use of the said signalling
device, it is possible to evaluate workstation conditions and, if necessary,
to find more convenient place of use. It is often impossible to find a place
where no vibrations occur, however the signalling device allows the operator
to determine when and where to expect vibrations of the lowest-frequency. Due
to the following 3 factors: specified values of free vibrations, resolution,
and permissible weighing instrument errors, the analysis must be carried out
with regard to the right effective value (rms value) and bandwidth. Such
analysis is performed on each 5Y series balance in real time. In the case of
the module that is an integrated part of the balances manufactured by RADWAG,
the filters analyse those frequencies to which the weighing devices are
sensitive. In the case of balances, effective value (rms value) measured from
3 axes and for a particular time interval is given, such value provides
information on amplitude and history of disturbances. Radwag has designed an
analysis method and selected permissible vibration thresholds respectively to
make signalling device operation both effective and convenient (not
troublesome). Balance screen displays balance-registered vibration level in
percent. Vibration threshold cannot be changed, however the operator can
modify settings in a way ensuring that alerts are displayed in the most
convenient way, depending on the needs. By default, the set value is 100%.
Example of settings modification:
– 68 –
For better understanding, analyse the following example (the values used below
are exemplary ones, they were not taken from the balance). 1. Assumed
vibration threshold set by default on 5Y series balance: 100 mm/s2. 2. Max
vibrations value: 100% (threshold for alert pictogram colour change, from
green to
red). Pictogram colour is green for vibrations within 0 100 mm/s2 range, for
vibrations value of 101 mm/s2, the pictogram changes to red. 3. Max vibrations
value: 150%. Pictogram colour is green for vibrations within 0 – 150 mm/s2
range, for vibrations value of 151 mm/s2, the pictogram changes to red.
Vibrations are signalled by a respective pictogram displayed in the weighing
result window. A window with the current sensor values is presented in the
middle of the display on the right. Displayed vibration value (Vibrations) is
the effective value registered within the most recent 2 seconds. This value is
refreshed every 2 seconds. If the screen displays 100%, this means that
vibrations have reached the threshold set by RADWAG.
Vibrations are recorded in the database, this allows to analyse the history of
ambient conditions quality.
– 69 –
The effective value of vibrations registered by the balance over a declared
time interval is saved (from the moment of the last record) Setup/Ambient
conditions/Ambient conditions recording interval.
By default, the interval is set to 10 min; this value may range between 1 – 60
min. Record analysis allows the user to select the optimal time of the day for
the measurements to be carried out, i.e. the time when the ambient conditions
influence the vibrations the least. Mind that there are places where the most
convenient working conditions occur after-hours. Along with each measurement,
data on ambient conditions alerts and level of vibration that occurred in the
course of the measurement are saved.
– 70 –
15. PARTS COUNTING
determined using a reference quantity of parts subjected to weighing, acquired
from the database of products, entered manually as a numeric value. Mode
activation procedure: · Go to the home screen, slide down the top menu and
press < > field. <Working
modes> submenu with list of working modes opens. · Select
the top bar of the screen features working mode name.
– 71 –
Parts Counting mode screen
15.1. Mode Related Settings The supplementary settings enable you to adjust
the working mode to your needs and requirements. Procedure:
1. Slide out the menu on the left, the available options are displayed. 2.
Select
displayed:
Functions: – ACAI, Automatic Accuracy Correction YES, weight value of a single
piece to be updated. NO, weight value of a single piece not to be updated.
Means of operation of ACAI function: 1. pieces quantity upon adding must
increase, 2. pieces quantity upon adding cannot increase more than twice,
– 72 –
3. pieces quantity upon adding must be comprised within ± 0.3 tolerance of
the total value.
4. the weighing result must be stable.
– Minimum Reference Sample Mass 1 unit, 2 units, 5 units, 10 units; minimum
mass value of a single piece. If this condition is not met, the counting
process will not start.
– Result control YES, set to print and save only those measurements which are
comprised within the Lo and Hi thresholds. NO, set to print and save all
measurements.
– Other functions in
15.2. Parts Counting Quick Access Buttons
Each working mode features automatically displayed set of default on-screen
quick access buttons. The set can be modified, i.e. you can add out-of-list
quick access buttons to it. Such operation requires particular permission
level. You can add special buttons enabling quick access to the function of
determination of a single piece mass, where the single piece mass is
determined on the basis of the reference mass, given for a particular quantity
of pieces:
15.3. Setting Reference Sample Mass by Entering Mass of a Single Part Procedure:
· Press <
Set part mass> button,
featuring an on-screen keyboard is displayed.
· Enter respective value and press
button for confirmation, <Parts
counting> mode is run with automatically set mass of a single part.
CAUTION: If the value of entered single part mass is lower than 0.1 of the
reading unit, then the following message is displayed:
15.4. Setting Reference Sample Mass by Determining Mass of a Single Part
Procedure: · Put the container on a weighing pan and tare it.
· Press
button (Determine part mass),
featuring an on-screen keyboard is displayed.
– 73 –
· Enter respective value and press
button to confirm. The following message
is displayed:
value).
· Load the weighing pan with the declared amount of parts. When the indication is stable
( pictogram is displayed), press
button to confirm the mass.
· Single part mass is calculated automatically,
the weighing device displays quantity of single parts (pcs).
CAUTION: Remember:
· Total weight value of all parts loaded onto the weighing pan cannot be
greater than the max capacity.
· Total weight value of all parts loaded onto the weighing pan cannot be lower
than the value determined by
· Single part mass value must be equal to or greater than 0.1 of the reading
unit. Unless this condition is met, the weighing instrument displays a
message:
15.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part from the Database
Each product recorded in the database of products features a set of information which allows identifying it. Among the information there is weight value used during part counting process.
Procedure: Enter
Entering reference sample mass to the balance memory Procedure for entering
single part mass to the products database:
1. Press
Procedure for entering single part mass to products database when there is no
data entered yet:
a) Press
b) Press < Add> button. c) Fill in all product-related fields and modify
15.6. Single Part Mass Update in the Database
Determined single part mass can be assigned to a product in the database of products. This option is applied while using ACAI (Automatic Accuracy Correction) function in order to determine part mass with high accuracy.
– 74 –
Procedure: 1. Determine single part mass. 2. Press
product is recorded as
15.7. Parts Counting Procedure
First, it is necessary to obtain weight value of a single piece. Select one of
the available options:
– Enter the weight value for a single part (section 15.3); next, put parts to
be counted on the weighing pan. Parts quantity is displayed.
– Determine mass of a single part from a given quantity of parts (section
15.4.). ACAI function pictogram is displayed (providing that the function is
active). Next, put parts to be counted on the weighing pan. Parts quantity is
displayed.
– Acquire weight value for a single part from the database (section 15.5). To
do it, you have to select respective product. Now put parts to be counted on
the weighing pan, parts quantity is displayed.
Caution: It is necessary to tare any additional elements (packaging) prior to
counting process start.
15.8. Checkweighing for Parts Counting Mode
Parts counting process can be aided by checkweighing function, i.e. control
whether indication is within the set thresholds. Checkweighing requires
setting values for two thresholds:
– Min threshold [Min= … pcs.] – Max threshold [Max= … pcs.] and setting YES
value for a Bar graph. The Bar graph displays relation: CURRENT PARTS
QUANTITY/CHECKWEIGHING THRESHOLDS. In order to define Min and Max thresholds’
values, open Products database and edit the
Product record or use < Procedure:
Checkweighing thresholds> quick access button.
· Press grey workspace, next select
– 75 –
· Press <
Checkweighing thresholds> button and enter Min and Max thresholds
values. Return to parts counting mode.
· Bar graph is displayed, its colour corresponds to the status of parts counting process.
o yellow: current parts quantity below the set Min threshold value.
o green: current parts quantity within the set thresholds.
o red: current parts quantity above the set Max threshold value.
15.9. Dosing in Parts Counting Mode
Parts counting process can be aided by dosing function, i.e. control whether
indication does not exceed the set target value. Dosing requires setting a
target value, e.g. 100 parts, and a tolerance in %. The bar graph
displays the target value as a bracket. In order to define target mass value,
press < Target value> quick access button. Procedure:
· Press grey workspace, next select
· Press <
Target Value> button and enter the quantity of parts to obtain.
– 76 –
· If the tolerance option is used, determine the tolerance value (0 100%). ·
Bar graph is displayed. It provides the following information:
o current parts quantity on the weighing pan (colour signalling see section
15.7.), o target mass value (black marker).
Caution: The checkweighing and dosing functions can operate simultaneously in
the parts counting mode. In such a case, the dosing tolerance is controlled by
Min and Max thresholds of the checkweighing function.
16. CHECKWEIGHING
16.1. Making Use of Checkweighing Thresholds
There are two methods for making use of the checkweighing thresholds: –
selecting Product with Min and Max threshold declared,
– entering numeric value for the thresholds, < not refer to any product.
– 77 –
; in this case, the thresholds do
Procedure 1 selecting product from Products Database:
· Press
button (Product database), list of products is displayed.
· Select the product that is to be weighed.
· Threshold values are displayed automatically within an Info field.
· Bar graph is displayed, its colour corresponds to the current status of mass:
o yellow: mass value lower than Min threshold value,
o green: mass value higher than Min threshold value and lower than Max threshold
value,
o red: mass value higher than Max threshold value.
Procedure 2 entering checkweighing thresholds manually:
· Press
button (Checkweighing thresholds).
· Press Min threshold
button and enter its value.
· Press
button to confirm.
· Press Max threshold
button and enter its value.
· Press
button to confirm.
Caution: Max threshold value has to be greater than Min threshold value.
16.2. Mode Related Settings
To find out more about the remaining functions, refer to section 14.6 of this user manual, Additional Parameters for a Weighing Process.
– 78 –
17. DOSING
<
Dosing> is a working mode that facilitates dispensing process, wherein the
dispensing takes as long as it is necessary to obtain a pre-defined target weight.
Dosing mode screen:
17.1. Making Use of Products Database in the Course of Dosing Operation
You can use target value assigned to a product or values recorded in the
Products Database. Additionally, it is possible to define temporary target
values. In the case of the Product Database, the target value is the value
entered in the mass field.
Procedure 1 selecting product from Products Database:
· Press
button (Product database), a list of products is displayed.
· Select product that be weighed.
· Target and tolerance values are displayed automatically within the Info field.
· Target value is displayed with minus sign.
– 79 –
· Bar graph is displayed, its colour corresponds to the current mass status : o yellow: mass value lower than ‘target value – tolerance value’, o green: mass value within permissible tolerance range: ‘target value +/- tolerance’ o red: mass value greater than ‘target value + tolerance value’.
Procedure 2 entering checkweighing thresholds manually:
· Press <
Target value> button.
· Window for entering the target value and tolerance is displayed, enter the values.
· Return to weighing.
Caution: For a product acquired from the Products Database, the Target Value
and Tolerance data are already given. The data can be edited and modified.
17.2. Mode Related Settings
To find out more about the remaining functions, refer to section 14.6 of this
user manual, Additional Parameters for a Weighing Process.
– 80 –
18. PERCENT WEIGHING AGAINST REFERENCE SAMPLE MASS
< Percent weighing> is a working mode for comparison of the measured sample’s weight with the reference mass. The result is expressed in [%]. PERCENT WEIGHING can be aided by DOSING and CHECKWEIGHING. Mind that these are not started automatically, neither is the bar graph. Percent Weighing mode screen:18.1. Sample Mass and Reference Sample Mass Comparison Comparison methods:
– Entering reference mass, to do it, press <
Set reference mass> button.
– Accepting mass of a load on the weighing pan as a reference mass; to do it press
button (Set as 100%). – selecting a product from the products database for which the mass parameter has been
determined; to do it, press
button (Products database).
Procedure 1 entering reference mass manually
· Press
button (Set reference mass).
· Enter the reference mass value and press
button to confirm.
· The mass of weighed products is compared to the reference mass, the difference is
displayed in [%].
– 81 –
Procedure 2 accepting mass of a load on the weighing pan as a reference mass · Load the weighing pan with the sample, wait for a stable indication.
· Press
button (Set as 100%).
· 100.000% indication is displayed, weight of the product on the weighing pan is
accepted as the reference mass; it is automatically recorded and displayed in the
reference mass field.
· Unload the weighing pan.
· From now on, the mass of weighed products is compared to the reference mass, the
difference is displayed in [%].
Procedure 3 selecting product from the Products Database:
· Press
button (Products database). Select product to be weighed.
· Info field automatically displays modified information concerning reference mass.
· Product-related reference mass is automatically recorded and displayed in the
reference mass field. Press
button to open the reference mass field.
· 0.00 % indication is displayed (for unloaded pan).
· From now on, the mass of weighed products is compared to the reference mass, the
difference is displayed in [%].
18.2. Checkweighing and Dosing in Percent Weighing Mode
PERCENT WEIGHING can be aided by DOSING and CHECKWEIGHING. In order to make
them accessible, assign them to an on-screen button.
Remember to provide percent values for both of the functions.
Procedure: 1. Press grey workspace area. 2. The following submenus are displayed: Settings, Buttons, Information, Printouts, Profile. 3. Select
1. Press
button (Checkweighing thresholds).
2. Press Min threshold button and enter its value [%].
3. Press
button to confirm.
4. Press Max threshold button and enter its value [%].
5. Press
button to confirm.
Caution: Max threshold value has to be greater than Min threshold value.
– 82 –
DOSING Dosing function requires the target value to be specified (in [%])
along with the dosing tolerance, i.e. permissible +/- surplus or shortfall.
Procedure:
1. Press
button (Target value).
2. Enter the target value in [%].
3. Set tolerance value, if required.
4. Press
button to confirm.
5. Press Min threshold button and enter its value [%].
6. Press
button to confirm.
18.3. Bar Graph
The DOSING and CHECKWEIGHING functions are aided by a graphic indication, i.e. a bar graph. An example of simultaneous operation of DOSING and CHECKWEIGHING functions is presented below.
· < Checkweighing thresholds> settings: Min = 90%, Max = 110%.
· reference mass = 2.500 g <
.
18.4. Mode Related Settings To find out more about the remaining functions,
refer to section 14.6 of this user manual, Additional Parameters for a
Weighing Process.
– 83 –
19. DENSITY
<
Density> is a working mode for determining the density of: solids, liquids, air
(available for XA 5Y), various substances using pycnometer.
In order to operate this mode, a density determination kit (supplementary equipment) is
required.
Density mode screen
19.1. Density Determination Kit Kit intended for XA balances:
– 84 –
1 Beaker base
2 Stand
3 Sinker
4 Beaker
5 Thermometer holder
6 Thermometer
7 Flexible connector for sinker
8 Hanger
9
Top weighing pan for solids density determination
10
Flexible connector for weighing pans
11
Bottom weighing pan for solids density determination
12 Additional sinker
13
Additional stand for a set of pans or a sinker
Additional set of weighing pans
14
for density determination of solids with density lower that
density of water
CAUTION: – Components of density determination kit shall be stored in a respective box. – Weighing pans and sinker must not be put directly on a table, this might cause their damage. Weighing pans and sinker, if not in use, must be put on an additional stand.
19.2. Assembly XA balances
CAUTION:
If the kit has been installed and -NULL- message is displayed, load the
balance with weight set (12) and zero or tare the indication. Thus prepared
balance can be used for density determination.
19.3. Solid Body Density Determination
Prior to the procedure start, it is necessary to set process-related
parameters: – Liquid type:
o Distilled water o Ethanol o Other liquid with determined density – Liquid
temperature (required if either distilled water or ethanol is used) – Liquids
Density The parameter is set automatically after the temperature has been
entered, if either distilled water or ethanol is used as liquid. In the case
of using liquid other than distilled water or ethanol, it is necessary to
enter its density manually.
The density of solids is calculated using the following formula:
A
=
o
A-B
– density of a sample
A – sample mass measured in the air
B – sample mass measured in liquid
o- liquid density
Procedure: 1. Install the density determination kit.
– 85 –
2. In order to record density value for a particul
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>