STRAVA Smart Splint Pan User Guide

June 12, 2024
STRAVA

STRAVA-logo Smart Splint Pan

Product Information

The Splint Bath is a device designed for cleaning and maintaining splints. It features a water basin, a water sensor, and a thermocouple inline filter. The unit operates at a temperature of 135°F/57°C and requires a cleaning solution consisting of vinegar or a coffee machine/commercial descaler. The unit also has a glass surface that may accumulate calcium over time, which can be easily cleaned with vinegar or a commercial glass cleaner. The unit is compatible with tap water, but distilled water can be used in hard-water environments with the provided proportions.

Product Usage Instructions

  1. Evacuate the water from the Splint Bath. Refer to the user manual for the procedure.
  2. Turn the unit off and add water and cleaning solution. Fill the unit just above the water sensor with a 2:1 ratio of tap water to the cleaning solution.
  3. Turn the unit back on and run it for 10 minutes at a temperature of 135°F/57°C.
  4. Evacuate the water from the Splint Bath. Refer to the user manual for the procedure.
  5. Turn the unit off, unplug it, and allow it to cool down before proceeding.
  6. Clean the inside water basin and the inline filter. Remove the thermocouple inline filter found inside the water basin beneath the grate to clean debris. See the quick tips section for removal and cleaning. Return the cleaned inline filter to the thermocouple.
  7. Refill the Splint Bath with tap water to the operating level. See the quick tips section for distilled water usage.
  8. Run the Splint Bath for 5 minutes. Connect the power, turn the unit on, and prime it to ensure circulation. Refer to the quick tips section for priming instructions.
  9. If necessary, repeat the cleaning procedures.

Note: For technical service, please contact Strava Solutions via email at technicalsupport@stravasolutions.com or by phone at 423.667.2325.

Cleaning Your Splint Bath

Frequency – Two or more times monthly. Strava recommends each facility determine its cleaning frequency based on usage, water grade, visual inspection, and sanitation protocols that apply to your facility and/or JACHO.

  1. EVACUATE WATER
    See the user manual procedure.

  2. TURN THE UNIT OFF-ADD WATER & CLEANING SOLUTION
    Fill the unit just above the water sensor with a 2:1 ratio of tap water to the cleaning solution.
    Cleaning Solution =
    Vinegar -OR- Coffee Machine/Commercial Descaler

  3. TURN THE UNIT BACK ON
    Run for 10 minutes @ 135oF/57oC.

  4. EVACUATE WATER
    See the user manual for the procedure.

  5. TURN UNIT OFF
    Unplug and allow the unit to cool down before proceeding.

  6. CLEAN INSIDE WATER BASIN
    Using a Scotchbrite® Scour Pad or cloth, wipe the inside tub with warm water. To remove scum, waterlines, mineral deposits, etc., add a mild detergent. Be sure to rinse thoroughly and wipe dry. Do NOT use bleach or extra abrasive products such as steel wool.

  7. CLEAN INLINE FILTER
    Remove the thermocouple inline filter found inside the water basin beneath the grate to clean debris. See the “quick tips” section for removal and cleaning. Return the cleaned inline filter to the thermocouple.

  8. REFILL WITH WATER
    Fill the unit to operating level with tap water. See the “quick tips” section if you use distilled water.

  9. RUN FOR 5 MIN.
    Connect power, turn unit on, and prime unit (see “quick tips” section) to ensure circulation.
    Repeat procedures if necessary

Cleaning The Glass

Over time, water vapor may cause calcium to build upon the glass and in the hinges. This is perfectly normal and easily cleaned with vinegar on a soft cloth or a commercial glass cleaner.

Replacing the UV-C Bulb

Annually – Recommend changing the bulb annually even when if is still working.
Purchase – Replacement bulbs can be purchased from a Strava dealer.
Replacement – Bulb replacement is simple and can be done by anyone, however, we recommend a qualified service technician if your organization has one.

Check out the bulb replacement video at: www.stravasolutions.com

DISTILLED WATER

Tap water is required for the unit to run correctly. Follow these proportions if you are in a hard-water environment and require distilled water.

CLEAN INLINE FILTER

Note: there are two filter versions depending on the model date.

Loosen the filter with supplied Allen key. Wiggle the filter off of the thermocouple. Run under water and scrub out debris from the screen with a hard bristled brush utensil.
Fit the cleaned filter back on the thermocouple by wiggling it into place (screw heads facing out). Tighten the filter, ensuring that it is properly aligned and close to the hole.

PRIMING

Priming removes any trapped air allowing the pump to engage. Insert priming bulb into the priming port. Squeeze and hold until the bulb is fully inflated. Release from the port. Once the water bubbles, the unit is primed. Repeat if necessary. ****

FOR TECHNICAL SERVICE
email: technicalsupport@stravasolutions.com tel: 423.667.2325
PART NUMBER 400.610 STRAVA SPLINT BATH CARE CARD ©Copyright 2022 Strava Solutions, LLC All rights reserved

References