SD1 Ultrasonic Pocket Doppler User Manual
About this Manual
P/N: 01.54.458123
Version 1.5
MPN: 01.54.458123015
Release Date: Mar. 2023
Statement
This manual will help you understand the operation and maintenance of the
product better. It is reminded that the product shall be used strictly
complying with this manual. User’s operation failing to comply with this
manual may result in malfunction or accident for which the manufacturer cannot
be held liable.
The manufacturer owns the copyrights of this manual. Without prior written
consent of the manufacturer, any materials contained in this manual shall not
be photocopied, reproduced or translated into other languages.
Materials protected by the copyright law, including but not limited to
confidential information such as technical information and patent information
are contained in this manual, the user shall not disclose such information to
any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or
implicitly, any right or license to use any of the intellectual properties of
the manufacturer.
The manufacturer holds the rights to modify, update, and ultimately explain
this manual.
Responsibility of the Manufacturer
The manufacturer only considers itself responsible for any effect on safety,
reliability and performance of the equipment if:
Assembly operations, extensions, re-adjustments, modifications or repairs are
carried out by persons authorized by the manufacturer, and
The electrical installation of the relevant room complies with national
standards, and
The instrument is used in accordance with the instructions for use.
The manufacturer will make available on request circuit diagrams, component
part lists, descriptions, calibration instructions, or other information that
will assist service personnel to repair those parts of the equipment that are
designated by the manufacturer as repairable by service personnel.
Product Information
Product Name: Ultrasonic Pocket Doppler
Model:SD1
Terms Used in this Manual
This guide is designed to give key concepts on safety precautions.
WARNING
AWARNING label advises against certain actions or situations that could result
in personal injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage
equipment, produce inaccurate data, or invalidate a procedure.
NOTE
A NOTE provides useful information regarding a function or a procedure.
Safety Precautions
CAUTION
Federal (U.S.) Law restricts this device to sale by or on the order of a
physician.
NOTE:
This user manual is written to cover the maximum configuration. Therefore,
your model may or may not have some of the parameters and functions described,
depending on what you have ordered.
This unit is internally powered equipment, and it is an IEC/EN 60601-1 Type BF
applied part.
Type BF protection means that the connection between the equipment and
personnel complies with permitted leakage currents and dielectric strength of
IEC/EN 60601-1.
WARNING and CAUTION messages must be observed. To avoid the possibility
of injury, observe the following precautions during the operation of the
device.
WARNING
- It is to be used by health care professionals on the order of a physician.
- The Doppler is a tool to aid the healthcare professional in hospitals, clinics and should not be used in place of normal fetal detection. It is not intended for treatment or use during labor and delivery.
- Placement of the ultrasound transducer on the abdomen is critical to obtaining the fetal heart beat as opposed to maternal heart beat or other abdominal noise. The user should be trained in proper placement techniques either through acceptable Ob/Gyn training and individual state accreditation, or as being prescribed by such a trained clinician and trained in device placement.
- This device is not explosion-proof and cannot be used in the presence of flammable anesthetics.
- Magnetic and electrical fields are capable of interfering with the proper performance of the device. For this reason, make sure that all external devices operated in the vicinity of this device comply with the relevant EMC requirements. X-ray equipment and magnetic resonance imaging (MRI) devices can emit high levels of electromagnetic radiation.
- We recommend that exposure to ultrasound should be kept as low as reasonably achievable. This is considered to be good practice and should be observed at all time.
- Do not use the device with HF surgical equipment and do not use it in an MRI environment.
- The device is not protected against defibrillation.
- SHOCK HAZARD – Do not attempt to replace batteries with wet hands.
- The device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
- The medical electrical equipment needs to be installed and put into service according to the EMC Information provided in this user manual.
- Portable and mobile RF communications equipment can affect medical electrical equipment; refer to section Recommended Separation Distances.
- Do not service or maintain the device or any accessory which is in use with a patient.
CAUTION
-
Refer servicing to qualified personnel.
-
Keep the device in a clean environment and avoid vibration during storage.
-
Do not sterilize the Doppler.
-
Electromagnetic Interference – Ensure that the environment in which the device is operated is not subject to any source of strong electromagnetic emissions, such as radio transmitters, mobile telephones, etc.
-
Prior to examination using the Doppler, check for visible damages of the main unit and the probe that may endanger the patient/operator or machine performance. If the damage is found, replace them with good ones at once.
-
The following safety checks should be performed once every two years or as specified in the institution’s test and inspection protocol by a qualified person who has adequate training, knowledge, and practical experience to perform these tests.
◆ Inspect the equipment for mechanical and functional damage.
◆ Inspect the safety relevant labels for legibility.
◆ Inspect the equipment for mechanical and functional damage.
◆ Inspect the safety relevant labels for legibility.
The leakage current should never exceed the limit. The data should be recorded in an equipment log. If the device is not functioning properly or fails any of the above tests, the device has to be repaired. -
The device and accessories are to be disposed of according to local regulations after their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose them together with house-hold garbage.
Introduction
Intended Use/Indications for Use
The SD1 is a pocket Doppler device used for detecting the fetal heartbeat from
the 10th week of gestation. It is intended to be used by medical professionals
only.
Features
◆ FHR detection and display
◆ FH sound
◆ Switching off when no signal received for 2 Min
◆ Sound volume adjustment
◆ Bluetooth connection (Optional)
◆ FH signal intensity indicator
◆ FH icon
◆ Battery indicator
◆ Low battery warning
◆ Sound volume levels
Appearance (Above pictures are just for reference)
LCD Display & Touch Keys
Item | Description | |
---|---|---|
1 | Fetal heart icon |
heart beat.
2| Fetal heart signal intensity indicator| | This indicator displays on the
left side of the screen and has three status: empty, half empty and full,
which respectively represents low, medium and high fetal heart signal
intensity.
3| FHR numeric| | Displays fetal heart rate within the range from 50 bpm to
240 bpm. When fetal heart rate is out of the range, it displays “—”.
Sound volume numeric| | Sound volume numeric is displayed in the center
of the screen, the same area as the FHR numeric. When you adjust sound volume,
the sound volume numeric will display for 0.5 second before switching back to
display FHR numeric. Sound volume ranges from level 0 to 7.
4| Battery indicator| | Battery indicator displays on the right side of
the screen. There are 5 battery levels, represented by 0 to 4 panes in the
icon. When battery is empty, battery empty icon will be displayed and
flickering, and the battery needs replacing.
5| Sound volume increase touch key| | Touch the key to increase sound volume.
**** 6| Sound volume decrease touch key| | Touch the key to decrease sound
volume.
7| On/Off touch key| | When the Doppler is off, touch this key for a little
while to switch it on; When the Doppler is on, touch this key for a little
while to switch it off.
Battery
SD1 is powered by two AA alkaline batteries. Battery specification: LR6, AA,
1.5 V;
Note:
You can use AA alkaline batteries of the same specification purchased locally.
Basic Operation
NOTE:
To ensure that the Doppler works properly, please read this chapter and
Chapter Safety Precautions before operation; follow the steps when connecting
all the components.
Opening the Package and Checking
Open the package; take out the Doppler and accessories carefully. Keep the
package for possible future transportation or storage. Check the components
according to the packing list.
◆ Check for any mechanical damage.
◆ Check all the cables and accessories.
If there is any problem, contact us or your local distributor immediately.
Installing the Battery
a) Unscrew the screw with a cross screwdriver and remove the battery
compartment cover.
b) Insert the battery into the compartment carefully. Ensure its anode and
cathode terminals are aligned with the anode and cathode marks on the
compartment.
c) Install the compartment cover and secure it with the screw.
Removing/ Replacing the Battery
a) Unscrew the screw with a cross screwdriver and remove the battery
compartment cover.
b) Take out the used battery. You can also replace it with a new one. Ensure
the new battery’s terminals are placed in the right direction as indicated by
the anode and cathode marks.
c) Install the compartment cover and secure it with the screw.
WARNING
- Turn off the Doppler before removing or replacing the battery.
- Replace alkaline batteries with those of identical specifications provided by the manufacturer or purchased locally. See Chapter Product Specifications for details about battery specifications.
- If the batteries have been inserted incorrectly, the Doppler will not function or it will be damaged.
- Do not disassemble or short-circuit batteries.
- Do not recharge batteries.
- Do not dispose of batteries in fire or water.
- Do not allow metal objects to contact the battery terminals.
- Do not mix with used or other battery type (such as alkaline with carbon zinc).
- Do not solder the batteries directly. If soldering or welding connection to the battery is required, consult our engineer for proper methods.
- Do not over-discharge batteries.
- To install or remove batteries, follow the equipment manufacturer’s instructions.
- Keep battery away from small children. If swallowed, consult a physician at once.
- Store the battery in cool, dry place before use. Do not keep batteries at temperature of 45°C or above, or at humidity of 75% or above.
- Dispose the battery according to the local regulations. Refer to IEC61429 for standard disposal when necessary.
- Remove the battery and store it at a cool and dry environment if the Doppler is not used for a long time.
- Batteries have life cycles. If the time that the Doppler using battery becomes much shorter than usual, the battery life is at an end. Replace the battery with a new one of the same specification as the one provided or recommended by the manufacturer.
Switching On
Touch the On/Off touch key for about 1second when the Doppler is off, and the
Doppler will display the switching on interface before switching to display
the test interface .
Switching Off
Touch the On/Off touch key for about 1second when the Doppler is on, and the
Doppler will be switched off.
If the Doppler is not in operation or no signal is received for 2 minutes, the
Doppler will switch off automatically.
FHR Detection
Before applying the Doppler to inspect FHR, you should always check whether
the Doppler is in good condition and whether there is evident damage that
might affect patient’s safety and the device’s function.
If evident damage is found, stop using it at once and replace it with a good
one.
Procedures to Detect FHR:
a) Have the patient lie on her back.
b) Apply appropriate amount of coupling gel to the ultrasonic transducer head
of the Doppler and switch on the Doppler.
c) Palpate the patient’s abdomen gently to confirm the fetus’s position.
d) Place the Doppler on the patient’s abdomen, and move it around the fetus’s
position or tilt it until a clear and rhythmic heart sound is heard and FHR
numeric is stably displayed.
Note:
-
Do not mistake the maternal heart rate for fetal heart rate.
The fetal pulse should be different from the maternal pulse, which can be measured at the wrist or neck. -
Do not wear gloves to touch the keys. If there’s water and coupling gel on the fingers, please clean them first or the touching effect will be influenced.
How to Find the Best FH Signal:
- The easiest way: take the position the doctor last detected for FHR as a reference and move the Doppler around the position slowly until the best FH signal is found.
- The fetal heart position may change as the fetus moves inside the uterus. You can confirm the fetal position first according to the position of the uterus fundus in different gestational weeks.
The clearest and loudest fetal heart sound is generally obtained when the Doppler is placed on the fetus’s back. Fetal movement is usually the movement of fetal limbs. So, if frequent fetal movement occurs at the right side of the abdomen, the fetus’s back is probably at the left side and vice versa. You can find the fetus’s back according to fetal movement’s position.
If the fetus is in cephalic delivery position, the fetal heart is either on the right side or on the left side below the navel; if the fetus is in breech delivery position, the fetal heart is either on the right side or on the left side above the navel.
Steps to Find Fetal Heart:
Have the patient lie on back and relax >> confirm fetal position by hand >>
apply coupling gel to the Doppler>> place the Doppler on patient’s abdomen and
start looking for the fetal heart >> the fetal heart is found when the Doppler
gives out a continuing thumping sound “boom-boom-boom”.
CAUTION
- The Doppler’s degree of protection against harmful ingress of water is IP22. Do not immerse it in water.
- The Doppler is delicate and sensitive. Please handle it with care and try to avoid dropping on to the ground or any hard surfaces. Any damage caused by dropping is not covered by the warranty.
- Keep the coupling gel away from children. If swallowed, consult a physician at once.
Note:
- The best quality of fetal heart signal is obtained only when the Doppler is placed on the best detection position.
- Do not place the Doppler near positions where placental sound or umbilical blood flow sound is loud.
- If the fetus is in the cephalic position and the mother is supine, the clearest heart sound will normally be found on the midline below the navel. During detection, the pregnant woman’s prolonged lying in the supine position should be avoided to reduce the possibility of supine hypotension. Putting a pillow or cushion under the patient’s head or feet can be of help.
- It is not possible to obtain accurate FHR unless a clear fetal heart signal is detected. If the calculated FHR is not in accordance with the beat of the fetal heart sound, the fetal heart sound auscultation result shall prevail.
- When applied to the patient, the Doppler may warm slightly (less than 2°C (35.6°F) above ambient temperature). When NOT applied, the Doppler may slightly (less than 5°C (41°F) above ambient temperature).
After Detection
- Switch off the Doppler.
- Wipe the remaining gel off the patient and the probe with a clean soft cloth or tissue.
Mobile Application Software (APP)
SD1 can connect to mobile phones with its Bluetooth function (optional). The
SD1 Care APP has both Android and iOS versions.
iOS APP operating environment:
A) hardware environment
Processor: dual-core Apple A6 RAM: ≥1GB
B) software environment: iOS 8.0 and above operating system
C)network environment: support Bluetooth How to use SD1 Care APP
Android APP operating environment:
A) hardware environment
CPU: frequency≥1.0GHz
RAM: ≥1GB
B)software environment: Android 4.3 and above operating system
C)network environment: support Bluetooth
- Download and install software
Scan either of the following QR codes to download the SD1Care APP, and install and run it as prompted.|
---|---
https://itunes.apple.com/cn/app/sd1-care/id1360969195?mt=8| http://cloud.edan.com.cn/downloads/sd1National/us/sd1.apk
Note:
1 Your mobile phone may prohibit the installation of “applications from
unknown sources”. Enter Settings to allow the installation first.
2 For normal functioning of the APP, please give the APP function-related
permissions.
3 For how to use the APP, read the instructions in the About sub-interface
under the Settings interface of the APP.
-
Activate the device
Open the APP and go to Settings>Activation and input SD1 activation code (14 numbers after 01). -
Pair device
Open Bluetooth function of the mobile to automatically pair the SD1. -
Start detection
Put the coupling gel on SD1 and position the probe to the optimal place of maternity’s abdomen. And click the “start” key. After pressing start, confirm that the data on the APP and the SD1 probe match. As with any Bluetooth communication, it is important to make sure the connection is not compromised. -
Adjust the fetal heart beat sound volume
When using mobile phone to play the fetal heart beat sound, you can adjust the volume with the volume keys of the mobile phone. When using SD1 to play the heart beat sound, touch ‘volume+’ or ‘volume -’ to adjust the volume. -
Finish the detection
When the detection is finished, click ‘Stop’ touch key and the detection data will be saved automatically.
Note: Please make sure your mobile phone has enough battery power, and avoid killing the process directly or switching to other applications during the fetal heart detection. -
Real time detection mode and DEMO mode
We provide DEMO mode for users’ reference. You can turn on DEMO key in Setup and enter fetal heart detection interface to watch the DEMO. The word ‘DEMO’ is displayed in the interface to distinguish from real time detection.
WARNING
SD1 complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
- this device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
NOTE:
- This equipment (SD1) has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
-Consult the dealer or an experienced radio/TV technician for help. - Any changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Maintenance and Cleaning
Maintenance
Before each use, check if the equipment has visible evidence of damage that
may affect the patient and the operator’s safety or the Doppler’s functioning.
If the damage is evident, contact the manufacturer for service or replace it.
The overall check of the Doppler, including safety check and function check,
should be performed by qualified personnel every 12 months, and each time
after service. And safety check must include current leakage test and
insulation test. Besides the above requirements, comply with local regulations
on maintenance and measurement.
The accuracy of FHR is determined by the Doppler and cannot be adjusted by
user. If you have doubt concerning the accuracy of FHR, verify it with other
methods such as using a stethoscope, or contact local distributor or the
manufacturer for help.
The Doppler is frangible and must be handled with care. Wipe the remaining gel
off the Doppler after each use. These measures can help prolong the Doppler’s
life.
Replace the accessories such as the battery according to use. If any of the
accessories are damaged, refer to chapter Ordering Information for details and
order new ones.
Please check the label for the date of manufacture, the service life is 5
years (The service life is limited to the Doppler, not including the
replaceable accessories. The only replaceable accessory of SD1 is battery. The
frequency of usage is 8 hours/day).
Cleaning
Before cleaning, switch off the Doppler.
Keep the exterior surface of the device clean and free of dust and dirt.
Clean the exterior surface of the Doppler with a dry, soft cloth. If
necessary, clean it using a soft cloth dampened with mild near neutral
detergent, ethanol (75%) or isopropanol (70%), and then wipe it dry with a dry
cloth immediately.
CAUTION
- Do not use strong solvent, such as acetone.
- Never use an abrasive such as steel wool or metal polish.
- The Doppler’s degree of protection against harmful ingress of water is IP22. Do not immerse it in water.
- Do not remain any solution on the surface after cleaning.
Disinfection
In normal use the Doppler does not need disinfection. In case of being soiled,
clean the main unit case and then disinfect it by wiping it with a soft cloth
dampened with ethanol (75%) or isopropanol (70%). Then wipe it dry with a dry
cloth.
CAUTION
Do not immerse the Doppler into the disinfector.
Product Specifications
Product Information
Product Name | Ultrasonic Pocket Doppler |
---|---|
Model | SD1 |
Complied Standards
IEC 60601-1:2005/A1:2012, EN 60601-1:2006/A1:2013, IEC 60601-1-2:2014, IEC
60601-2-37:2015,IEC 61266:1994
Classification
Anti-electric Shock Type: | Internally powered equipment |
---|---|
Anti-electric Shock Degree: | Type BF equipment |
Degree of Protection against Harmful Ingress of Water: | IP22. Protection |
against vertically falling water drops when ENCLOSURE tilted up to 15°
Degree of Safety in Presence of Flammable Gases:| Equipment not suitable for
use in presence of flammable gases
Working System:| Continuous running equipment
EMC:| CISPR 11 Group 1 Class B
Physical Specifications
Size: | LengthWidth Height: (48±2) mm× (39±2) mm× (147±3) mm |
---|---|
Weight: | < 180g |
**** LCD: | Size: |
Display: | ◆ FHR |
◆ Battery level
◆ Signal intensity| ◆ Sound volume level
◆ FH icon
Coupling Gel:| pH: 5.5~8.0
Acoustic Impedance: 1.5×106 Pa.s/m ~1.7×106Pa.s/m (35°C/95ºF)
Environment
Working:| Temperature:+5°C ~ +40°C ( +41ºF ~ +104ºF) Humidity:15% RH ~ 95% RH
(non-condensing) Atmospheric Pressure:70kPa ~ 106 kPa
---|---
Transport and Storage:| Temperature:-25°C ~ +70°C (-13ºF ~ +158ºF)
Humidity:15% RH ~ 95% RH (non-condensing) Atmospheric Pressure:70 kPa ~106 kPa
Note: The time required for the Doppler to warm from the minimum storage temperature between uses until it is ready for intended use is at least 2 hours; the time required for the Doppler to cool from the maximum storage temperature between uses until it is ready for intended use is at least 2 hours Performance Specifications
FHR (Essential Performance):| FHR Measuring Range: 50 bpm ~ 240 bpm
Accuracy: ±2 bpm
Note: FHR measurement result may not be accurate if the equipment is measuring
beyond its measuring range.
---|---
FHR Resolution:| 1bpm
Audio Output:| Output Power: 2w
Background noise: <45dBA
Auto Power-off:| Power off when the Doppler receives no signal or operation
for 2 minutes.
Bluetooth:| Transmission Range (Without Obstacles) :>5m (Indoor range depends
on the building’s structure and material.)
Ultrasound:| Nominal Frequency: 3MHz
Working Frequency: 3MHz
p_<1 MPa
Iob<10 mW/cm2
Ispta<100 mW/cm2
Isata<10 mW/cm2
Isppa.3<190 W/cm2
Ispta.3<94 mW/cm2
Effective Radiating Area: 490mm2 ± 15%
Working Mode: pulse wave
Battery Specifications
Specification: | Two AA 1.5V alkaline batteries (AA, LR6, 1.5V) |
---|---|
Working Duration: | ◆ ≥6h |
Bluetooth Specifications
FCC ID | SMQSD1 |
---|---|
Modulation: | GFSK |
Frequency: | 2400-2483.5MHz |
Tolerance Frequency: | ≤ 20ppm |
RF output power: | ≤ 20dBm (EIRP) |
Occupied Channel Bandwidth: | ≤ 3MHz |
Transmitter Unwanted Emissions: | ≤﹣30dBm |
Low Output Summary Table
(For systems whose global maximum value does not exceed 1.0)
System: SD1 Ultrasonic Pocket Doppler
Model (MHz) | Ispta.32(mW/cm) | TI Type | TI Value | MI | Isppa.32(W/cm) |
---|---|---|---|---|---|
SD1 CD3.0 | 5.69 | TIS | 0.05 | 0.01 | 0.02 |
TIB | 0.01 |
Ordering Information
CAUTION
Only the parts supplied by the manufacturer should be used with the Doppler.
Parts | Part Number |
---|
Main Unit
SD1 Doppler(Non-Bluetooth version)| 02.06.262535
SD1 Doppler(Bluetooth version)| 02.06.262639
Accessories
AA Alkaline Battery| 01.21.064086
Normal Carry Case| 01.56.466428
Coupling Gel| 01.57.078170
Ultrasound Intensity and Safety
Ultrasound in Medicine
The use of diagnostic ultrasound has proved to be a valuable tool in medical
practice. Given its known benefits for non-invasive investigations and medical
diagnosis, including investigation of the human fetus, the question of
clinical safety with regards to ultrasound intensity arises.
There is no easy answer to the question of safety surrounding the use of
diagnostic ultrasound equipment. Application of the ALARA (As Low As
Reasonably Achievable) principle serves as a rule-of-thumb that will help you
to get reasonable results with the lowest possible ultrasonic output.
The American Institute of Ultrasound in Medicine (AIUM) states that given its
track record of over 25 years of use and no confirmed biological effects on
patients or instrument operators, the benefits of the prudent use of
diagnostic ultrasound clearly outweigh any risks.
Ultrasound Safety and the ALARA Principle
Ultrasound waves dissipate energy in the form of heat and can therefore cause
tissue warming. Although this effect is extremely low with Doppler, it is
important to know how to control and limit patient exposure. Major governing
bodies in ultrasound have issued statements to the effect that there are no
known adverse effects from the use of diagnostic ultrasound, however, exposure
levels should always be limited to As Low As Reasonably Achievable (the ALARA
principle).
Explanation of MI/TI
MI (Mechanical Index)
Cavitations will be generated when ultrasound wave passes through and contacts
tissues, resulting in instantaneous local overheating. This phenomenon is
determined by acoustic pressure, spectrum, focus, transmission mode, and
factors such as states and properties of the tissue and boundary. This
mechanical bioeffect is a threshold phenomenon that occurs when a certain
level of ultrasound output is exceeded. The threshold is related to the type
of tissue. Although no confirmed adverse mechanical effects on patients or
mammals caused by exposure at intensities typical of present diagnostic
ultrasound instruments have ever been reported, the threshold for cavitation
is still undetermined.
Generally speaking, the higher the acoustic pressure, the greater the
potential for mechanical bioeffects; the lower the acoustic frequency, the
greater the potential for mechanical bioeffects. The AIUM and NEMA formulate
mechanical index (MI) in order to indicate the potential for mechanical
effects. The MI is defined as the ratio of the peak-rarefactional acoustic
pressure (should be calculated by tissue acoustic attenuation coefficient 0.3
dB/cm/MHz) to the acoustic frequency.
MI = Pr, α
fawf ×CMI
CMI = 1 (MPa / MHz )
TI (Thermal Index)
Heating of tissues is caused by absorption of ultrasound when the ultrasound
energy is applied. The temperature rise is determined by the acoustic
intensity, exposed area and thermo physical properties of the tissue.
In order to indicate the potential for temperature rise caused by thermal
effects, the AIUM and NEMA formulate thermal index (TI). It is defined as the
ratio of the total acoustic power to the acoustic power required to raise the
tissue temperature by 1ºC (1.8°F).
According to different thermo physical properties of the tissue, TI is divided
into three kinds: TIS, TIB and TIC.
TIS (Soft Tissue Thermal Index): It provides an estimate of potential
temperature rise in soft or similar tissues.
TIB (Bone Thermal Index): It provides an estimate of potential temperature
rise when the ultrasound beam passes through soft tissue and a focal region is
in the immediate vicinity of bone.
TIC (Cranial Bone Thermal Index): It provides an estimate of potential
temperature rise in the cranial bones or superficial bones.
Measurement Uncertainties
The uncertainties in the measurements were predominantly systematic in origin;
the random uncertainties were negligible in comparison. The overall systematic
uncertainties were determined as follows:
-
Hydrophone Sensitivity: ± 12percent for intensity, ± 6 percent for pressure. Based on the hydrophone calibration report by ONDA. The uncertainty was determined within ±1 dB in frequency range 1-15 MHz.
-
Digitizer: ±0.3 percent for intensity. ± 0.15 percent for pressure.
Based on the stated accuracy of the 8-bit resolution of the Agilent DSO6012 Digital Oscilloscope and the signal-to-noise ratio of the measurement. -
Temperature:±2.4 percent for intensity uncertainty, ±1.2 percent for pressure uncertainty.
Based on the temperature variation of the water bath of ± 1ºC (1.8°F). -
Spatial Averaging: ± 3.5 percent for intensity, ± 1.75percent for pressure.
-
Non-linear Distortion: N/A.
No effects of nonlinear propagation were observed.
Since all the above error sources are independent, they may be added on an RMS
basis, giving a total uncertainty of ± 12.73 percent for all intensity values
reported, ± 6.37 percent for all the pressure values,,± 12.6 percent for the
Mechanical Index, uncertainty of ±12.73% percent for power,±0.15 percent for
center frequency, ±6.87%for the MI.
Prudent Use Statement
Although no confirmed bioeffects on patients caused by exposure from present
diagnostic ultrasound equipment have ever been reported, the potential exists
that such bioeffects may be identified in the future. Therefore, the
ultrasound should be used prudently. High levels of acoustic output and long
exposure time should be avoided while acquiring necessary clinical
information.
Reference for Acoustic Output and Safety
- “Bioeffects and Safety of Diagnostic Ultrasound” issued by AIUM in 1993
- “Medical Ultrasound Safety” issued by AIUM in 1994
- “Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, Revision 3” issued by AIUM/NEMA in 2004
- “Standard for real-time display of thermal and mechanical acoustic output indices on diagnostic ultrasound equipment, Revision 2” issued by AIUM/NEMA in 2004
- “Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers” issued in 2008.
- “Medical electrical equipment—Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and detection equipment” issued by IEC in 2007.
Acoustic Output Reporting Table for Track 1 Acoustic output reporting table
for IEC60601-2-37(IEC60601-2-37, Edition 2.1, 2015-0, table 201.103)
Transducer Model: SD1, Operating Mode: PW mode
Index label| MI| TIS| TIB| ****
TIC
---|---|---|---|---
| At Surface| Below Surface| At surface| Below Surface
Maximum index value| 0.01| 0.05| 0.01| N/A
Index component value| | N/A| 0.05| NA| 0.01|
Acoustic Parameters| pr.αat zMI (MPa)| 0.02| | | |
|
P (mW)| | 7.35| 7.35| N/A
P1x1 (mW)| | N/A| N/A|
zs(cm)| | | 3.50| | |
zb(cm)| | | | | 3.70|
zMI(cm)| 3.70| | | | |
zPII.α (cm).α| 3.70| | | | |
fawf (MHz)| 3.00| 3.00| 3.00| N/A
Other information| prr (Hz)| 5000| | | | |
srr(Hz)| N/A| | | | |
npps| 1| | | | |
Ipa.α at zPII.α (W/cm2)| 0.02| | | | |
Ispta.α at zPII.α or zSII.α(mW/cm2)|
5.69
| | | | |
Ispta at zPII or zSII (mW/cm2)| 12.26| | | | |
pr. at zPII
(MPa)
| 0.04| | | | |
| | | | | | |
Operating control conditions| Fixed
Acoustic Output Reporting Table for Track1(Non-autoscanning Mode)
Transducer Model: SD1 ,Operating Model: PW
Acoustic Output | MI | ISPTA.3 (mW/cm^2) | ISPPA.3 (W/cm^2) |
---|---|---|---|
Global Maximum Value | 0.01 | 5.69 | 0.02 |
Associated Acoustic Parameter | Pr.3 | (MPa) | |
W0 | (mW) | ||
fc | (MHz) | 3.00 | |
Zsp | (cm) | 3.70 | |
Beam dimensions | X-6 (cm) | 2.50 |
Y-6
(cm)
| |
2.50
|
2.50
PD| | (usec)| 72.25| | 72.25
PRF| | (Hz)| 5000| | 5000
EBD| Az. (cm)| | 2.50|
Ele. (cm)| | 2.50|
Operating Control Conditions| Fixed
Standard Parameter Equal Contrast List
Parameter| Note| | Parameter| Note
p r.α| Attenuated
Peak-rare-factional Acoustic
Pressure| | f awf| Center Frequency, Acoustic Working Frequency
p r| Peak-rare-factional Acoustic Pressure| | X| -12dB Output Beam
Dimensions
P| Output Power| | Y
z s| Depth for Soft Tissue Thermal Index| | t d| Pulse
Duration
P α (Z s )| Attenuated Output Power| | prr| Pulse Repetition
Frequency (Pulse Repetition Rate)
I ta.α (Z s )| Attenuated Temporal-average Intensity| | d
eq| Equivalent Beam Diameter
z bp| Break-point Depth| | I pi.α at max MI| Attenuated Pulse-
average Intensity at the point of Maximum MI
z b| Depth for Bone Thermal Index| | Aaprt| -12dB Output Beam Area
I pi.α| Attenuated Pulse-intensity Integral| | MI| Mechanical Index
I pi| Pulse-intensity Integral| | TIS| Soft Tissue Thermal Index
d eq (Z b )| Equivalent Beam Diameter at the point of Zsp| | TIB|
Bone Thermal Index
TIC| Cranial-bone Thermal Index
EMC Information
Electromagnetic Emissions
Guidance and manufacturer’s declaration – electromagnetic emission
The SD1 Ultrasonic Pocket Doppler is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should
assure that it is used in such an environment.
Emission test| Compliance| Electromagnetic environment –
guidance
RF emissions CISPR 11| Group 1| The SD1Ultrasonic Pocket Doppler uses RF
energy only for its internal function. Therefore, its RF emissions are very
low
| | and are not likely to cause any interference in nearby electronic
equipment.
---|---|---
RF emission CISPR 11| Class B| ****
The SD1 Ultrasonic Pocket Doppler is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC/EN61000-3-2| Not
applicable
Voltage fluctuations /flicker emissions IEC/EN61000-3-3| Not applicable
Electromagnetic Immunity
Guidance and manufacture’s declaration–electromagnetic immunity
The SD1 Ultrasonic Pocket Doppler is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test| IEC 60601 test level| Compliance level|
Electromagnetic environment-guidance
Electrostatic discharge (ESD) IEC 61000-4-2| ±8kV contact
±15kV air| ±8kV contact
±15kV air| Floors should be wood, concrete or ceramic tile. If floor are
covered with synthetic material, the relative humidity should be at least 30%.
Electrical Fast Transient/Burst IEC/EN61000-4-4| ±2kV for power supply lines
±1kV for input/output lines| ****
Not applicable
| ****
Not applicable
Surge IEC/EN61000-4-5| ± 1 kV line(s) to line(s)
± 2 kV line(s) to earth| Not applicable| Not applicable
Voltage dips, short interruptions, and voltage variations on power supply
input lines IEC/EN61000-4-11| <5%UT(>95% dip inUT) for 0.5cycle 40%UT(60%dip
in UT) for5 cycles 70%UT(30%dip in UT) for25 cycles <5%UT(>95% dip inUT) for
5s| Not applicable| Not applicable
Power frequency (50Hz/60Hz) magnetic field IEC61000-4-8| 30 A/m| 30 A/m| Power
frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
IMMUNITY to proximity magnetic fields| 8A/m, Modulation: CW Test
frequency:30KHz; 65A/m, Modulation: Pulse modulation, 2.1KHz Test
frequency:134.2KHz; 7.5A/m, Modulation: Pulse modulation, 50KHz Test
requency:13.56MHz;| 8A/m, Modulation: CW Test frequency:30KHz; 65A/m,
Modulation: Pulse modulation, 2.1KHz Test frequency:134.2KHz; 7.5A/m,
Modulation: Pulse modulation, 50KHz Test
frequency:13.56MHz;|
Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
The SD1 Ultrasonic Pocket Doppler is intended for use in the electromagnetic
environment specified below. The customer or the user of the device should
assure that it is used in such an environment.
Immunity test| IEC 60601
test level| Compliancelevel| Electromagnetic environment-
guidance
---|---|---|---
Conducted RF
IEC61000-4-6 Radiated RF
IEC61000-4-3| 3 Vrms
150 kHz ~ 80 MHz
6Vrmsc)in ISM bands
between 0,15 MHz and80 MHz 10V/m 80 MHz ~ 2.7
GHz| Not
Applicable 10 V/m
80 MHz to
2.7 GHz| Portable and mobile RF communications equipment should be used no
closer to any part of the SD1 Ultrasonic Pocket Doppler, including cables,
than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 0.35 √P 80 MHz to 800 MHz
d = 0.7 √P 800 MHz to 2.7 GHz
d =6 √P/E at RF wireless communications equipment bands (Portable RF
communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part
of the SD1 Ultrasonic Pocket Doppler, including cables specified by the
manufacturer).
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance level in
each frequency range. b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE1:At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the SD1 Ultrasonic Pocket
Doppler is used exceeds the applicable RF compliance level above, the
SD1Ultrasonic Pocket Doppler should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the SD1 Ultrasonic Pocket Doppler.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
c The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80
MHz are 6,765 MHz to6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283
MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and
80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz
to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17
MHz,21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and
50,0 MHz to 54,0 MHz.
Table-Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
Test Frequenc y (MHz)| Brand a) (MHz)| Service a)|
Modulatio n b)| Maximu m Power(W )| Distanc e (m)|
IMMUNIT Y TEST LEVEL (V/m)
---|---|---|---|---|---|---
385| 380-390| TETRA 400| Pulse modulationb) 18Hz| 1.8| 0.3| 27
450
| 430-470| GMRS 460, FRS 460| FM C)
±5 kHz deviation 1kHz sine| 2| 0.3| 28
710| 704-787| LTE Brand 13,17| Pulse modulation b) 217 Hz| 0.2| 0.3| 9
745
780
810| 800-960| GSM 800/900,TETR A 800,iDEN 820, CDMA 850, LTE Band 5| Pulse
modulationb) 18 Hz| 2| 0.3| 28
870
930
1720| 1700-1990| GSM 1800;
CDMA 1900;
GSM 1900; DECT; LTE
Band 1, 3, 4,25;UMTS| Pulse modulationb) 217 Hz| 2| 0.3| 28
1845
1970
2450| 2400-2570| Bluetooth, WLAN,802.11 b/g/n, RFID 2450, LTE Brand 7|
Pulse modulation b) 217 Hz| 2| 0.3| 28
5240| 5100-5800| WLAN 802.11 a/n| Pulse modulation b) 217 Hz| **** 0.2| 0.3|
9
5500
5785
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between
the transmitting antenna and the ME EQUIPMENT or ME SYSTEM maybe reduce to 1m.
The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative FM modulation, 50% pulse modulation at 18 Hz may be used
because while it does not represent actual modulation, it would be worst case
Recommended Separation Distances
Recommended separation distances between portable and mobile RF communications
equipment and the SD1 Ultrasonic Pocket Doppler
The SD1 Ultrasonic Pocket Doppler is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or
the user of the SD1 Ultrasonic Pocket Doppler can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the SD1
Ultrasonic Pocket Doppler as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of transmitter
(W)
| Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2 √ P| 80 MHz to 800 MHz
d = 0.35 √P| 800 MHz to 2.7GHz
d = 0.7 √P
0.01| /| 0.035| 0.07
0.1| /| 0.11| 0.22
1| /| 0.35| 0.7
10| /| 1.11| 2.21
100| /| 3.5| 7
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Overall Sensitivity
D | d | A | B | V S | V n | C | S |
---|---|---|---|---|---|---|---|
∑Ba | B w | B |
1.58
3MHz
| 50| 40.9| T| 5#4#| 0| 77.0| 140| 75| 5.42| 123.32
Ba| 77.0
75| 44.4| T| 5#3#| 0| 68.4| 80| 40| 6.02| 118.82
Ba| 68.4
100| 46.9| T| 5#3#| 0| 69.6| 180| 90| 6.02| 122.52
Ba| 69.6
200| 52.9| T| 5#3#| 0| 68.4| 83| 42| 5.92| 127.22
Ba| 75.1
2.38
3MHz
| 50| 39.0| T| 5#4#| 0| 77.0| 130| 69| 5.50| 121.50
Ba| 77.0
75| 42.5| T| 5#4#| 0| 77.0| 115| 55| 6.41| 125.91
Ba| 77.0
100| 45.0| T| 5#3#| 0| 68.4| 130| 65| 6.02| 119.42
Ba| 68.4
200| 51.0| T| 5#4#| 0| 77.0| 78| 43| 5.17| 133.17
Ba| 75.1
Doppler Frequency (Hz)| 505| Velocity of Target (cm/s)| 10
Note| D: Diameter of Target
Reflector(mm)| A: Attenuation A(dB))| S:Overall Sensitivity (S=A+B+C)dB
d: Distance (d)(mm)| VS: Signal RMS (mV)| C:Signal to Noise Ratio (dB)
B:Two-wayAttenuation(dB) B=∑Ba+Bw| Vn: Noise RMS (mV)
Troubleshooting
Problem | Possible Cause | Solution |
---|---|---|
**** Fail to power on, or shut down shortly after switching on | Battery level | |
is very low. | Replace the battery. | |
Battery is not installed properly. | Re-install the battery. | |
Fail to switch on the Doppler as instructed. | Touch the On/Off touch key for a |
while to power on the Doppler.
The Doppler has malfunctions.| Contact service personnel.
Loudspeaker does not work.| Sound volume has been turned down to the lowest
level.| Adjust sound volume to appropriate level.
If the Doppler is configured with Bluetooth, fetal heart sound can be played
by mobile phone.| Set to play fetal heart sound by mobile phone or the Doppler
on
the APP.
The Doppler has malfunctions.| Contact service personnel.
FHR cannot be displayed stably.| There is strong interference source such as
high frequency machines and mobile phones nearby.| Use the Doppler away from
strong interference sources.
The fetal heart position has changed because of fetal movement.| Relocate the
Doppler to the best fetal heart rate detection position.
Friction between the Doppler and patient’s abdomen causes false displaying.|
Find the best fetal heart rate detection position.
Sensitivity is low and noise is too much.| There is strong interference source
such as high frequency machines and mobile phones nearby.| Use the Doppler
away from strong interference sources.
The Doppler is not applied with coupling gel.| Apply coupling gel to the
Doppler.
The Doppler is not placed at the best detection position.| Relocate the
Doppler to the best
fetal heart rate detection position.
The Doppler has malfunctions.| Contact service personnel.
Doppler cannot be connected to mobile phone.| The Bluetooth of mobile is not
open.| Open the Bluetooth of mobile.
The Doppler used is not configured with Bluetooth function.| Use the Doppler
with Bluetooth function.
The Bluetooth function of Doppler has malfunctions.| Use the FHR and sound
detected and displayed on the SD1 itself, and contact service personnel.
Warranty and Service
Warranty
The manufacturer warrants that the manufacturer’s products meet the labeled
specifications of the products and will be free from defects in materials and
workmanship that occur within warranty period.
The warranty is void in cases of:
A. damage caused by mishandling during shipping.
B. subsequent damage caused by improper use or maintenance.
C. damage caused by alteration or repair by anyone not authorized by the
manufacturer.
D. damage caused by accidents.
E. replacement or removal of serial number label and manufacture label.
If a product covered by this warranty is determined to be defective because of
defective materials, components, or workmanship, and the warranty claim is
made within the warranty period, the manufacturer will, at its discretion,
repair or replace the defective part(s) free of charge. the manufacturer will
not provide a substitute product for use when the defective product is being
repaired.
Contact Information
If you have any question about maintenance, technical specifications or
malfunctions of devices, contact your local distributor.
EDAN INSTRUMENTS, INC.
Address: #15 Jinhui Road, Jinsha Community, Kengzi Sub-District,
PingshanDistric,
518122 Shenzhen, P.R. China
Email: info@edan.com
Tel: +86-755-2689 8326
Fax: +86-755-2689 8330
www.edan.com
EC REPRESENTATIVE
Shanghai International Holding Corp. GmbH
Eiffestrasse 80, 20537 Hamburg Germany
Tel. : +49-40-2513175
E-mail: shholding@hotmail.com
Definition of Symbols
No. | Symbol | Definition | No. | Symbol | Definition |
---|---|---|---|---|---|
1 |
| CE marking| 11|
| Authorized Representative in the European Community
2|
|
The products marked with this symbol apply to the European WEEE directive. This symbol Indicates this equipment contains electrical or electronic components that must not be disposed of as unsorted municipal waste, but collected separately. Contact an authorized representative of the manufacturer for information for the decommissioning of your equipment.
| 12|
| General symbol for recovery/recyclable
3| ****| Operating instructions| 13|
| Refer to User Manual (Background: Blue; Symbol: White)
4|
| Caution| 14| ****| MR Unsafe–Keep away from magnetic resonance imaging
(MRI) equipment
5|
| Type BF applied part| 15| ****| Non-ionizing electromagnetic radiation
6|
P/N
| Part Number| 16|
**IP22**
| IP22 Protected against solid foreign objects of 12,5 mm ∅ and greater.
Protection against vertically falling water drops when ENCLOSURE tilted up to
15°
7|
| Serial Number (Start with H on battery compartment cover)| ****
17
|
**Rx Only**
| Caution: Federal (U.S.) law restricts this device to sale by or on the
order of a physician.
8| | Date of Manufacture| 18| | Medical device
9| FCC ID: SMQSD1| Federal Communications Commission: FCC ID: SMQSD1|
19
| | Unique Device Identifier
10| | Manufacturer| | |
References