SARAD Radon Scout eXpert Pressure Sensor User Manual
- October 30, 2023
- SARAD
Table of Contents
SARAD Radon Scout eXpert Pressure Sensor
Radon Scout eXpert
The Radon Scout eXpert is a highly sensitive radon monitor used for reference measurements. It has a real-time clock and accepts alkaline, NiCd, or NiMH batteries for power supply.
Controls
The instrument has three LED indicators:
- One green LED flash each four seconds indicates that the measurement is running.
- Four red LED flashes each four seconds indicate a radon alarm.
- One red LED flash each four seconds indicates that the battery is low.
Power Supply
Before changing the batteries, ensure that the instrument has entered standby mode. The battery compartment can be accessed by unscrewing the round battery cover. When inserting new batteries, pay attention to their correct polarity and use batteries of the same type. Do not use Lithium ion batteries because their cell voltage is 3V or 3.6V.
The instrument accepts alkaline, NiCd, or NiMH batteries. For maximum operational period, it is recommended to use alkaline batteries optimized for low loads. If the cell voltage drops below 1.1V, the red LED on the side panel will blink once every four seconds. The measurement will continue until the voltage drops below 0.95V. At this point, the measurement will stop, and the instrument will enter standby mode.
Battery Replacement
Replace both batteries at the same time to avoid failures caused
by different charging levels. The real-time clock must be reset by
software after changing the batteries.
Selection of the Right Location for Exposure
The instrument should be placed in a spot representative of the measurement, avoiding direct light incidence and exposure to strong heat sources. For permanent installation, attention should be paid to building materials because the wall itself could be the origin of the radon inside a room. Comparative measurements at the preferred placing position and in the middle of the room under similar weather conditions can determine if this is the case.
Selection of the Right Integration Interval
Short sampling intervals, such as five minutes, are recommended if the instrument is used as a personal dosimeter. Note that the running sampling interval will be discarded when the measurement is stopped, so the period from the last completed interval to this moment will not be included in the dose calculation.
Highly sensitive Radon Monitor for reference measurements
Version June 2022
Referenced documents Software-Manual Radon Vision
SARAD GmbH Wiesbadener Straße 10 01159 Dresden Germany
www.sarad.de
info@sarad.de
The instrument
Controls
Tab. 1. – LED signaling.
Power supply
The power supply of the Radon Scout eXpert is realised by a commercial 5V
AC/DC adapter (Fig. 1). The instrument should not be used without batteries.
If the AC/DC adapter is connected, the batteries will not be discharged. They
will work as a buffer in case of mains power interruption. During power supply
failure the instrument will continue to measure Radon for a number of
consecutive days without interruption. If the instrument works only with
batteries the Wi-Fi interface and the CO2 Sensor (when present) will be
deactivated.
Make sure that the instrument has entered the standby mode before changing the
batteries (Fig. 3). The battery compartment can be accessed after opening
(unscrewing) the round battery cover (Fig. 4). Please pay special attention to
the correct polarity of the batteries when inserting them (Fig. 5). Change
both batteries at the same time because different charging levels may lead to
failures. Use always batteries of the same type.
The instrument accepts alkaline as well as NiCd or NiMH batteries. ATTENTION!
Do not use Lithium ion batteries because their cell voltage is 3V or 3.6V.
For maximum operational period we recommend the usage of alkaline batteries
which are optimized for low loads.
If the cell voltage drops below 1.1V, the red LED at the side panel starts
blinking (one single flash each four seconds). The measurement will be
continued until the voltage drops below 0.95V. Then, the measurement will be
stopped and the instrument enters in standby mode.
The real time clock of the instrument must be set by software after changing
the batteries. Corresponding reminder will be displayed. The stored data
inside of the instrument are not affected by battery replacement.
Battery replacement
Selection of the right location for exposure
The instrument, depending on case scenario e.g.: laboratory, test chamber or
other locations, should be placed at a spot which is representative for the
measurement. Direct light incidence and exposure to strong heat sources should
be avoided.
For permanent installation attention has to be paid to building materials
because the wall itself could be the origin of the Radon inside a room. This
would result in an increased concentration in the surrounding of the surface.
Some comparative measurements at the preferred placing position and in the
middle of the room (each at least a few days under similar weather conditions)
will show if this is the case or not.
Selection of the right integration interval
It is possible to adjust the sampling interval of the Radon Scout eXpert
between 1 and 255 minutes. If the expected Radon concentrations are in the
order or less than the statutory reference level of 300 Bq/m³, an interval of
60 minutes should be used. Frequent zero readings for individual sampling
intervals indicate that the chosen interval is too short.
Short sampling intervals (for example 5 minutes) are recommended if the
instrument is used as personal dosimeter. In that case only the dose as
accumulated value for the entire sampling period is of interest. Note that the
running sampling interval will be discarded in the moment when the measurement
will be stopped. Therefore, the period from the last completed interval to
this moment would not be included in the dose calculation.
Operation
Start of a measurement
To start a measurement, the toggle button (2 in Fig.1) must be pressed to wake
up the instrument from the standby mode (display turns on). Now, the sliding
switch (10 in Fig.2) at the side panel can be shifted to the right into the
“Start” position. The display shows the time remaining to the presentation of
the first measurement result. Newly acquired sampling results will be appended
periodically to the previously stored results and shown on the display.
If the sliding switch was already in “START” position during battery
replacement it is necessary to shift it back to “STOP” (left) at first and set
it to “START” again.
The display turns off automatically after fifteen seconds and may be re-
activated by pressing the toggle button (2). After finishing the first
sampling interval this button is used to toggle between two display pages:
- Date and time
- Recent Radon concentration
- Average Radon concentration (“AVG”) or accumulated equivalent dose (“DOS”) since start of the measurement
- Temperature
- Relative Humidity
- CO2 concentration or atmospheric pressure (if available)
- Battery voltage
All results will be presented in the physical unit system which has been
selected by software (SI or US).
Pressing light button (3 in Fig.1) turns the display backlight ON.
Stopping of a measurement
To stop a running measurement, the sliding switch (10 in Fig.2) must be set to
the position “STOP”. This is not possible if the lock function is enabled (see
chapter “Lock function”). The instrument cancels the sampling without
finishing of the actual sampling interval. The last stored sampling interval
is the last one which has been completed.
After stopping the measurement, the display shows either the average
concentration or the accumulated dose of the entire sampling period.
Furthermore, the time stamps of the beginning and the end of the sampling
period are shown. If the first sampling interval was not completed “Radon-
Scout eXpert” with device serial number appears on the display.
Carrying out a measurement
Prior to each new measurement campaign, the voltage of the batteries should be
checked. This can be done by software in standby mode or during the first
sampling interval (when the display shows no results yet). Older data in the
memory which shall not be written into the data file together with the new
data can be deleted by software.
If the Radon concentration exceeds the adjusted alert threshold, the red LED
(1 in Fig.2) at the side panel starts blinking (four consecutive flashes each
four seconds). The contacts of the switching output become closed (see chapter
“Alert switch”).
Lock function
To avoid a sudden or unintended interruption of a measurement campaign by
unauthorised staff, the toggle switch as well as the display output can be
locked by software setup. It is possible to activate the lock function during
sampling.
Dose calculation
The Radon Scout eXpert offers the possibility to show the equivalence dose
derived from the Radon exposure. The dose value on the display means the
accumulated dose from the start of a measurement until the end of the last
completed sampling interval. To calculate the dose, a dose conversion
coefficient must be defined with respect to the local statutory regulations.
The factory setting of the dose conversion coefficient follows the recent ICRP
2018 recommendations for indoor workplaces (predominantly sedentary activity;
F * 16.8mSv/MBqhm-3). An equilibrium factor F of 0.5 is assumed resulting in a
pre-set value of 8.4mSv/MBqhm-3 (8.4nSv/Bqhm-3).
To disable the presentation of equivalence dose on the display the dose
conversion coefficient must be set to zero. To change the dose conversion
coefficient (or to set it to zero), the initialisation software for the
instrument is required.
CO2-Sensor
The optional integrated CO2 sensor uses the non-dispersive infrared (NDIR)
operational principle. This requires an infrared source which cannot be
supplied by the internal batteries due to the increased power consumption.
Therefore, an external supply (AC/DC adapter) must be connected to the
instrument.
Continuous operation
The automatic calibration procedure of the sensor uses the CO2 concentration
of fresh air (400ppm) as reference. The instrument must be exposed to this
“fresh air concentration” at least once during the last 24 hours. This can be
achieved by short ventilation of the room. If there are no persons inside a
room for a while (for example overnight), the concentration goes back to
400ppm.
ATTENTION! If the reference concentration cannot be reached within the past 24
hours, the sensor interprets the lowest measured result as 400 ppm. This
results in a systematically increasing display value.
Occasional operation
If the CO2-sensor is used for sampling periods less than 24 hours no
calibration procedure as described above is carried out. Therefore it is
recommended to operate the sensor at least once per week for a period longer
than 24 hours to force a calibration. Of course, the instrument must be
exposed to the fresh air concentration of 400ppm during that time.
Communication interfaces
USB
The communication port according USB standard (11 in Fig. 2) is used to read
out measurement data and to adjust the measurement parameter of the Radon
Scout eXpert. The interface requires a software driver which can be downloaded
from the SARAD website. Click on the downloaded exe-file to install the
driver. The communication path appears as an additional COM port in Radon
Vision software.
The display is permanently activated while the instrument remains connected
with a computer or any other power source trough the USB port. At the same
time, as long as the USB connection persists, the Wi-Fi interface will be
disabled.
Wi-Fi
The Radon Scout eXpert is equipped with Wi-Fi interface. This feature allows to integrate the device within Local Area Network. After connecting the device to the local Wireless Access Point data download and device configuration can be performed by means of Radon Vision software. Following steps are required to obtain wireless connection with the device:
- PC with Radon Vison software installation has to be present in the same LAN as Wireless Access Point.
- IP address of the PC has to be known.
- You have to connect the device with USB cable to the PC and access the setup in Radon Vision.
- Subsequent parameters have to be entered in the Wi-Fi section:
- SSID: Name of the local Wi-Fi
- PASS: Wi-Fi password
- IP: IP address of the PC
- Port: 50002
- After cancelling the USB connection and restarting the device through detaching and connecting the power supply unit once more, the device will be selectable in the device list of Radon Vision.
After completing the above steps the device will be accessible form all PCs present in the same LAN with Radon Vision installation. The only requirement is that the PC, which IP address is written into the Radon Scout eXpert, has to be active. It is also possible to access the device with means of internet connection from outside of the LAN. This is additional service. Please, contact SARAD GmbH to acquire the conditions!
Alert switch
The Radon Scout eXpert is equipped with a switch output which can be used for
ventilation control or alert indication. The potential-free contacts of the
switch will be closed at the end of the sampling interval if the measured
Radon concentration exceeds the threshold. The output will be deactivated one
minute before the subsequent interval ends. Therefore, the sampling interval
must be set at least to two minutes. The alert threshold is factory-set to
300Bq/m³ (statutory reference value of the Radiation Protection Act) and may
be changed by the user with respect to local regulations (instrument setup in
Radon Vision). The output connector for the alert switch can be found on the
side panel (8 in Fig. 2).
As an accessory, we offer a wireless power switch which can be used for direct
ventilation control. The switch includes a transmitter unit which has to be
connected to the alert switch output of the instrument, and a plug adaptor
(receiver) which can be placed between the wall outlet and the power cable of
the ventilation unit. Optionally, a flush-mounting power switch is also
available. The plug adaptor offers an additional timer function to define the
duration of the ventilation period independent from the sampling interval of
the instrument (for example 15 minutes on while the sampling interval is 60
minutes). The wireless switch is coded – that means, several transmission
lines can be established within one building. The assignment between
transmitter and receiver is easily done by a push button.
The sampling interval should be set to a value between 30 to 60 minutes if the
threshold level is lower than 300Bq/m³.
Technical data
Radon Scout eXpert
- Operational principal Lucas cell with gross alpha counting
- Sampling method Diffusion
- Sensitivity > 25.0 cpm/(kBq/m³)
- Range 1 Bq/m³ … 1,000,000 Bq/m³
- Uncertainty < 2% over the entire range
- Sample interval 1 … 255 minutes adjustable
- Ambient conditions -10°C … 50°C, 0%rH … 100%rH non-condensing
- Temperature sensor 0°C … 100°C
- Humidity Sensor 0%rH … 100%rH
- Pressure sensor 1) 700 mbar … 1200 mbar
- CO2 Sensor 1) 400 ppm … 5000 ppm
- Switch output Optical relay with potential-free contacts; max. switch current 0.2A; max. switch voltage 40V; 2 pin socket Binder Series 712
- Display 20×4 characters LCD with backlighting
- Staus LEDs Red and green LED for alert, battery and cycle status.
- Memory 16383 data records (approx. 2 years at 60 min sampling interval)
- Power supply 5V AC/DC adapter required for operation with CO2 or Wi-Fi, otherwise 2 x AA batteries
- Autonomous operation Approx. 1 month at 20°C depending on battery type and usage profile
- Interface USB
- Wi-Fi Switch output (potential-free contacts)
- Controls Start/stop with sliding switch (with lock function); two push buttons (backlight and toggle between result pages)
- Dimensions 168 mm x 240 mm x 36 mm + Wi-Fi antenna
- Weight ca. 1.240 g incl. Batteries
- Options
Wireless switch
Transmitter| 2 inputs| Additional input for a manual switch for
example
---|---|---
Frequency| 433 MHz| With coding for safe transmission
Range| approx. 30 m|
Stromversorgung| Battery CR2032| Good for 20,000 switching cycles
Load switch (plug adapter)| 220 V/50 Hz| 6 programmable groups (e.g.
to connect more than one Radon monitors)
Max. switching load| 3500 VA|
Adjustable timer| 7 s, 5 min, 30 min, 1 h,
2 h, 4 h, 8 h
| If the timer function is not used, the power
switch follows the state of the switch contact at the instrument
Documents / Resources
|
SARAD Radon Scout eXpert Pressure
Sensor
[pdf] User Manual
Radon Scout eXpert Pressure Sensor, Radon Scout eXpert, Pressure Sensor,
Sensor
---|---
References
- SARAD Radon Thoron Gas Monitoring - Strahlenschutz
- SARAD Radon Thoron Gas Monitoring - Strahlenschutz
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>