AND Weighing MS-70 Moisture Analyzer Instruction Manual
- June 9, 2024
- AND Weighing
Table of Contents
MS-70 Moisture Analyzer
Product Information:
The product is a set of four testing equipment models: MS-70,
MX-50, MF-50, and ML-50. The company manufacturing the product is
A&D Company, Limited. The product is available for purchase at
Test Equipment Depot located in Woburn, MA, USA. The contact number
for Test Equipment Depot is
781-665-1400 and their toll-free number
is 1-800-517-8431. Customers can also
visit the website
www.TestEquipmentDepot.com to buy the product.
The model numbers of the four testing equipment are:
- MS-70
- MX-50
- MF-50
- ML-50
Product Usage Instructions:
Before using the product, customers should read the user manual
carefully and familiarize themselves with the product features and
usage instructions.
To use the testing equipment, follow these steps:
-
Turn on the device by pressing the power button.
-
Select the appropriate testing mode from the options
provided. -
Connect the device to the object being tested using the
appropriate cables or probes. -
Start the testing process by pressing the start button.
-
Observe the results displayed on the device’s screen.
-
If necessary, adjust the settings or parameters to get accurate
results. -
Once the testing is complete, turn off the device and
disconnect it from the object being tested.
Note: Customers should follow all safety instructions provided
in the user manual while using the product.
MS-70 MX-50 MF-50 ML-50
5 Commonwealth Ave Woburn, MA 01801 Phone
781-665-1400 Toll Free
1-800-517-8431 Visit us at
www.TestEquipmentDepot.com
1WMPD4000477F
A&D Company, Limited
WARNING o Do not use a sample that could make a dangerous chemical reaction
and cause an explosion or poisonous gas, when the sample is dried. o Keep
flammables away from the analyzer. Parts of the analyzer become very hot.
Materials placed near it might catch fire. o Do not use the analyzer in
ambient ignitable gas. It may cause explosion and fire. o Use a power source
(voltage, frequency, outlet type) adapted to the specification of the
analyzer. If excessive voltage is used, the analyzer may overheat and be
damage or cause a fire. o Turn off the power switch and remove the power cord
from the power outlet, when replacing the halogen lamp. Touching an electrode
of the halogen lamp connector carelessly, it may cause to receive an electric
shock. o Do not disassemble the analyzer. It may cause an error, damage,
receiving an electric shock or fire. If the analyzer needs service or repair,
contact the local A&D dealer. o Avoid getting the analyzer wet. It is not a
water-resistant structure. If there is leakage of liquid into the analyzer, it
may cause damage to the analyzer or receiving electric shock. o Do not look at
the active halogen lamp to protect your eyes from damage. o Do not drop, hit
or crack the glassware including the halogen lamp, to avoid an injury. o When
the halogen lamp is used beyond 5000 hours, we recommend replacing the lamp
with a new one to avoid trouble. o When discarding a halogen lamp, do not
break it to avoid scattering glass and injury.
CAUTION o Do not touch the heater cover, the halogen lamp, glass-housing, pan
handle, sample pan and sample without adequate protection, it could cause a
burn or scald. Parts of the analyzer are very hot when a measurement finishes.
For operation, use the specified grips of the heater cover and pan handle. Use
the standard accessory tools. o Do not touch parts affixed with the & mark,
because they may get very hot and dangerous. o When the analyzer is used in a
room where hot air does not diffuse, it may unexpectedly overheat. In this
case, adjust the drying temperature or move the analyzer to a place with
adequate ventilation. o Avoid leaving the analyzer in direct sunlight, as that
could cause discoloration of the case or a malfunction.
Test Equipment Depot – 800.517.8431 – 5 Commonwealth Ave, Woburn, MA 01801 TestEquipmentDepot.com
MS-70 / MX-50 I MF-50 / ML-50
3
,:”,;”J >· H “,,1′”; r_'”:11 l<:·”-,””1/””,-z7′
,”0V’-;’/,f;:”;’ 1′:,1); “::’J.c-9-,c;z>;;;”)J:”‘:
.· ‘!J;. .Nn,easmn1:He,rn Bliieg,11ams· ·, · ><: ,.. ·. :, :;;,
‘.’
-l· .
‘
‘
‘-
“” o/” ‘ ‘
it;Qffl’.tvlJttSl£{c,Jmiij,Kti’l1/il!lllll§,:.::.::.;.;;:.· ,.,
” ,-
. ,;.,:;
j::!’ ‘) ‘
,:
There are five analysis modes.
Measurement Programs
Drying Program
Analyzing mode to Finish Measurement
Heating pattern
MS-70, MX-50, MF-50
ML-50
Drying Temperature
MX-50 MS-70 MF-50
ML-50
Standard Mode Std
Measurement condition is automatically selected by ACCURACY and minimum
Standard drying Ramp drying Step drying
Standard drying
value of% display.
Quick Mode
When drying rate is less than preset termination value,
9uc
measurement is completed
Quick drying
automaticallv.
Automatic Mode
U-R
Timer Mode LJ-l::
When drying rate is less than preset termination value, measurement is completed automaticallv. Sample is dried for a preset time. 1 min. to 480min.
Standard drying Ramp drying Step drying
Standard drying
I I Manuau”·l n-Mode
Measurement is completed by the STOP key. Max.
480 min.
Drying rate: Change of moisture content per one minute [%/min]
30°c to
200°c
50°c to
200°c
Test Equipment Depot – 800.517.8431 – 5 Commonwealth Ave, Woburn, MA 01801 TestEquipmentDepot.com
26
MS-70 / MX-50 I MF-50 / ML-50
. J ra.ra., ,i;JJJD”11,aJjrn..:i,1ffi.i1@;tirn, ,fs.rns.!l).:. ..,,
“/”t/'”;n.·”IJ” ,,R’0′::,-/;r-,q~r,f,0″;,J’> r “‘””‘-“i, ,t”;: -.;;-, “”‘s”J”‘””‘”A ,,”” ;; ”ccf”h'” ::-:r. , .t,;>
A .s. :.···,;. ,/;,
o A standard mass of 20g or 50g can be used for the calibration o A 20g
standard mass (AX-MX-41) is recommend. Caution o Avoid vibration and drafts
that affect the calibration. If affected, the analyzer
may be unable to calibrate the weighing sensor. o Use a 20g mass for the
calibration, because the height between the weighing
pan and glass-housing is 26 mm. If a tall mass is used, open the glass-housing
and avoid external influence.
12.2.1. Operation
1. Display the gram unit (of the weighing mode).
2. Install the weighing pan, pan support and pan handle. Close the heater
cover.
3. Press and hold the I PROGRAM I key to display I CH I.
!Preparation I
IPress and hold”IO”D'”
4. Press the I SELECT I key to display I CAL I.
I [H
[RL
5. Press the I ENTER I key to display I CAL O I.
0
I I 6. If 20g mass is used, …… Press the ENTER key.
:,-·11 LrR,L un
I
I I Proceed to step 8.
If 50g mass is used, …… Press the SELECT key.
Proceed to step 7.
7. Press the [jJ or IT] key to select 50.000g.
I I Press the ENTER key to store it. I I [I;[], CAL O are displayed in order.
Sun.ununung
0
I End
I
I I, I I 8. When displaying CRL O press the ENTER key to input “Data to display ZERO”. The standard mass value is displayed (Example: 20g).
9. Open the heater cover and put the standard mass
I I on the center of the pan and press the ENTER I I key to input “mass
value”. End is displayed.
Standard mass
10. Remove the mass to return to the weighing mode
I I If the report for GLP, GMP and ISO is to be
output (Refer to page 61), GLP is displayed. The output condition for the
report is selected in the function table.
Remove mass
CLP
0.000g
Test Equipment Depot – 800.517.8431 – 5 Commonwealth Ave, Woburn, MA 01801
TestEquipmentDepot.com
MS-70 I MX-50 I MF-50 / ML-50
57
IJ,dII____ 616 ID number ____56, 61
Indicators
13
,nFo
1
I SO
4, 57, 60, 61
Key mode
1
Keys Leveling
foot
___ 11,
13 15
LF
52
L ,51:
6
LO
19,22,27
Manual mode ___ 9, 26, 28, 37, 39
Mass
57, 66
nd’:i
16
Measurement program _____26
MMIEDM. ____
25, 54, 67 19,22,27
MOIST MOIST
/D /W
____2299
Moisture content ____29, 56, 61
oul:
55
Output format ___52, 61
Pan handle
11, 12, 15
Pan support ___11, 12, 15
PPPaoorlwaiteryirtyc ord____ 11_, 551212
Power input
11
Power supply voltage label——- 6, 15, 66
Power switch
11, 15
Pre-heating process
17
Printer
4, 70
PROG PROGRAM
key___2154
1 P”lru/c: ____26 Quick drying28, 37, 39 Quick mode 9, 26, 28,37, 39 Ramp drying 9, 28, 33, 35, 37, 39
Ratio
29
RRRAAefTTeIIrOOenceWD/c/WDard_ __ __ __221919
Repeatability
17
Report
58, 60
RESET key
14
RS-232C
11, 50, 51
RTS
51
RXD __ 51
Sample mass _ ___ 26
5S-admple pan
7, 11, 12 52,53,61
SELECTkey
14
5SS,Go[dj i_ umtar trat e d ihydrat e _ ___
51 46 43
Spoon Standard
accessories
11,
12 12
Standard drying——-9, 28, 30, 37, 39, 43
Standard mod
9, 26, 28, 37, 39
STARTkey
51:d
Step drying_____9,
28,
14, 37,
14
26
39
Stop bit
51
1SS:-Ttr[eORaPLmkmeyo de _
14 61 59
Temperature
6
Temperature sensor __59
Terminator
51, 52
Test sample ____ 11, 43
Thermogravimetric analysis
10
Thermometer
59
Timer mode
9, 26, 28, 37, 39
1T:-eUstPsample
12 67
Tweezer s __ 12
TXD
U-R _____
14,
51 26
U-n __ 14, 26
UU-n1i:t
29, 14,
52 26
Weighing sensor__ 56, 57
Wet sample mass
29
WinCT WinCT-Moisture
9,9,,
50 50
‘:ind Zero
display
16 14
Test Equipment Depot – 800.517.8431 – 5 Commonwealth Ave, Woburn, MA 01801 TestEquipmentDepot.com
72
MS-70 / MX-50 I MF-50 / ML-50
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>