BIRD 5000-EX Digital Power Meter Instruction Manual
- June 9, 2024
- Bird
Table of Contents
5000-EX Digital Power Meter
Digital Power Meter Model 5000-EX
The Digital Power Meter Model 5000-EX is a device designed to
measure the power of RF transmission lines. It is manufactured by
Advanced Test Equipment Corp. and can be purchased through their
website www.atecorp.com or by calling their customer service
hotline at 800-404-ATEC (2832).
Safety Precautions
Before using the Digital Power Meter Model 5000-EX, it is
important to understand and follow the safety precautions listed in
the user manual:
-
Always keep away from live circuits.
-
Do not attempt to remove the RF transmission line while RF
power is present as it can be a potential health hazard. -
Do not service or adjust alone. Always have someone capable of
rendering aid nearby. -
Avoid inhalation of fumes or prolonged contact with skin when
using dry cleaning solvents for cleaning parts as they may be
potentially dangerous. -
Ensure that an uninterruptible earth safety ground is supplied
from the main power source to test instruments. Grounding one
conductor of a two conductor power cable is not sufficient
protection and can lead to serious injury or death if not properly
supplied. -
Personnel working with or near high voltages should be familiar
with modern methods of resuscitation.
Usage Instructions
Follow these steps for proper usage of the Digital Power Meter
Model 5000-EX:
-
Ensure that the device is properly grounded and connected to a
power source. -
Connect the RF transmission line to the device.
-
Turn on the device and follow the instructions on the screen to
select the appropriate settings and measure the power of the RF
transmission line. -
When finished, turn off the device and disconnect the RF
transmission line. -
Always disconnect from external power before any disassembly to
avoid electric shock. -
If using the TPS or TPS-EF, only turn the connector nut when
connecting to avoid damage to the sensor body. Discharge all static
potentials before connecting the TPS(-EF) to prevent electrostatic
shock from causing sensor damage.
Advanced Test Equipment Corp.
www.atecorp.com 800-404-ATEC (2832)
INSTRUCTION BOOK
DIGITAL POWER METER MODEL 5000-EX
©Copyright 2008 by Bird Electronic Corporation Instruction Book P/N
920-5000-EX Rev. G THRULINE® is a Registered Trademark of Bird Electronic
Corporation
Microsoft® and Windows® are registered trademarks of the Microsoft Corporation
I am not blank
Safety Precautions
The following are general safety precautions that are not necessarily related
to any specific part or procedure, and do not necessarily appear elsewhere in
this publication. These precautions must be thoroughly understood and apply to
all phases of operation and maintenance.
Keep Away From Live Circuits
Operating personnel must at all times observe normal safety regulations. Do
not replace components or make adjustments inside the equipment with high
voltage turned on. To avoid casualties, always remove power.
Shock Hazard
Do not attempt to remove the RF transmission line while RF power is present.
Radiated RF power is a potential health hazard.
Do Not Service or Adjust Alone
Under no circumstances should any person reach into an enclosure for the
purpose of service or adjustment of equipment except in the presence of
someone who is capable of rendering aid.
Chemical Hazard
Dry cleaning solvents for cleaning parts may be potentially dangerous. Avoid
inhalation of fumes or prolonged contact with skin.
i
Bird Model 5000-EX Digital Power Meter
Safety Earth Ground
An uninterruptible earth safety ground must be supplied from the main power
source to test instruments. Grounding one conductor of a two conductor power
cable is not sufficient protection. Serious injury or death can occur if this
grounding is not properly supplied.
Resuscitation
Personnel working with or near high voltages should be familiar with modern
methods of resuscitation.
Safety Symbols
WARNING Warning notes call attention to a procedure which,
if not correctly performed, could result in personal injury.
CAUTION Caution notes call attention to a procedure which,
if not correctly performed, could result in damage to the instrument.
This symbol indicates that a shock hazard exists if the precautions in the
instruction manual are not followed.
The caution symbol appears on the equipment indicating there is important
information in the instruction manual regarding that particular area.
ii
Note: NOTE: Calls attention to supplemental informationWarning Statements
The following safety warnings appear in the text where there is danger to
operating and maintenance personnel, and are repeated here for emphasis.
WARNING Never attempt to connect or disconnect RF equipment from the
transmission line while RF
power is being applied. Leaking RF energy is a potential health hazard.
WARNING Disconnect from external power before any disassembly. The potential
for electric shock exists.
WARNING RF voltage may be present in RF element socket.
Keep element in socket during operation.
Caution Statements
The following equipment cautions appear in the text whenever the equipment is
in danger of damage, and are repeated here for emphasis.
CAUTION When connecting the TPS or the TPS-EF, only turn
the connector nut. Damage may occur if torque is applied to the sensor body.
iii
Bird Model 5000-EX Digital Power Meter
CAUTION Discharge all static potentials before connecting
the TPS(-EF). Electrostatic shock could damage the sensor.
CAUTION Only use the ac adapter provided with the DPM. Do not use the adapter
with the batteries removed.
CAUTION Do not exceed 2 W average or 125 W peak power for
5 µs when using the TPS or the TPS-EF. Doing so will render the sensor
inoperative.
CAUTION Do not use harsh or abrasive detergents for cleaning.
CAUTION Replace only with Ni-MH rechargeable A batteries.
Nominal Voltage 1.2V; Capacity 2700mAhr. Use Sanyo P/N HR-AUX or equivalent.
Safety Statements
USAGE ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED BY THE MANUFACTURER
MAY IMPAIR THE INSTRUMENT’S SAFETY PROTECTION. USO EL USO DE ESTE INSTRUMENTO
DE MANERA NO ESPECIFICADA POR EL FABRICANTE, PUEDE
iv
ANULAR LA PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO. BENUTZUNG WIRD DAS GERÄT
AUF ANDERE WEISE VERWENDET ALS VOM HERSTELLER BESCHRIEBEN, KANN DIE
GERÄTESICHERHEIT BEEINTRÄCHTIGT WERDEN.
v
Bird Model 5000-EX Digital Power Meter
UTILISATION
TOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS EXPLICITEMENT PRÉVUE PAR LE
FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L’INSTRUMENT.
IMPIEGO
QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN MODO DIVERSO DA COME
SPECIFICATO DAL PRODUTTORE LA PROZIONE DI SICUREZZA POTREBBE VENIRNE
COMPROMESSA.
SERVICE
SERVICING INSTRUCTIONS ARE FOR USE BY SERVICE – TRAINED PERSONNEL ONLY. TO
AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS QUALIFIED
TO DO SO.
SERVICIO
LAS INSTRUCCIONES DE SERVICIO SON PARA USO EXCLUSIVO DEL PERSONAL DE SERVICIO
CAPACITADO. PARA EVITAR EL PELIGRO DE DESCARGAS ELÉCTRICAS, NO REALICE NINGÚN
SERVICIO A MENOS QUE ESTÉ CAPACITADO PARA HACERIO.
WARTUNG
ANWEISUNGEN FÜR DIE WARTUNG DES GERÄTES GELTEN NUR FÜR GESCHULTES
FACHPERSONAL.
vi
ZUR VERMEIDUNG GEFÄHRLICHE, ELEKTRISCHE SCHOCKS, SIND WARTUNGSARBEITEN
AUSSCHLIEßLICH VON QUALIFIZIERTEM SERVICEPERSONAL DURCHZUFÜHREN.
ENTRENTIEN
L’EMPLOI DES INSTRUCTIONS D’ENTRETIEN DOIT ÊTRE RÉSERVÉ AU PERSONNEL FORMÉ AUX
OPÉRATIONS D’ENTRETIEN. POUR PRÉVENIR UN CHOC ÉLECTRIQUE DANGEREUX, NE PAS
EFFECTUER D’ENTRETIEN SI L’ON N’A PAS ÉTÉ QUALIFIÉ POUR CE FAIRE.
RF VOLTAGE MAY BE PRESENT IN RF ELEMENT SOCKET – KEEP ELEMENT IN SOCKET DURING
OPERATION.
DE LA TENSION H.F. PEAT ÊTRE PRÉSENTE DANS LA PRISE DE L’ÉLÉMENT H.F.
CONSERVER L’ÉLÉMENT DANS LA PRISE LORS DE L’EMPLOI.
HF-SPANNUNG KANN IN DER HF-ELEMENTBUCHSE ANSTEHEN – ELEMENT WÄHREND DES
BETRIEBS EINGESTÖPSELT LASSEN.
PUEDE HABER VOLTAJE RF EN EL ENCHUFE DEL ELEMENTO RF – MANTENGA EL ELEMENTO EN
EL ENCHUFE DURANTE LA OPERACION.
IL PORTAELEMENTO RF PUÒ PRESENTARE VOLTAGGIO RF – TENERE L’ELEMENTO NELLA
PRESA DURANTE IL FUNZIONAMENTO.
vii
Bird Model 5000-EX Digital Power Meter
About This Manual
This instruction manual guides users through the operation and maintenance of
the Bird 5000-EX Digital Power Meter (DPM), as well as the Bird 5010B
Directional Power Sensor (DPS), Bird 5011 and 5011-EF Terminating Power
Sensors (TPS), and Bird 5012 Wideband Power Sensor (WPS).
Typography
There are two types of keys on the DPM. A hard key has a specific function
which is indicated on the key. The key names for hard keys are set in bold
typeface, e.g. Press the ON key.
Speed keys, which appear under the display, have a different label depending
upon the function selected. The names appear at the bottom of the display,
directly above the corresponding key. The key names for speed keys are set in
bold italic typeface, e.g. Press the SCALE Key. In this manual, speed keys
will also be referred to by number, with speed key 1 being the key on the
left. Figure 1 on page 3 shows these speed key numbers.
Chapter Layout
Introduction — Identifies the parts of the DPM, describes the functions of the
various keys, and explains the meaning of the indicators which may be
displayed. Also lists the items supplied and optional equipment available.
viii
Installation — Gives directions for connecting the DPM, and discusses the
various power sources. Operation — Explains how to make measurements with the
DPM, and the special functions used with specific sensors. Maintenance — Lists
routine maintenance tasks for the Digital Power Meter, and troubleshooting
tips for common problems. Specifications and battery information are also
included. Changes to this Manual We have made every effort to ensure this
manual is accurate at the time of publication. If you should discover any
errors or if you have suggestions for improving this manual, please send your
comments to our factory. This manual may be periodically updated, when
inquiring about updates to this manual refer to the part number and revision
level on the title page.
ix
Table of Contents
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . viii
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 1
Component Description . . . . . . . . . . . . . . . . . . . . . . 3 Display
Description . . . . . . . . . . . . . . . . . . . . . . . . . 5 Speed Keys . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Installation. . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Directional Power Sensor (DPS) . . . . . . . . . . . 11 Wideband Power Sensor
(WPS) . . . . . . . . . . . 11 Terminating Power Sensor (TPS) . . . . . . . .
. . 12 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 13 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 19
x
Chapter 1
Introduction
Items Supplied
1. Bird Digital Power Meter (DPM) 2. AC Mains Adapter 3. Instruction Book 4.
9-pin Serial Cable
Optional Equipment
Directional Power Sensor (DPS)
Bird 5010B Thruline Sensor, requires two 43 or APM/DPM elements. The DPS
measures both forward and reverse power, so VSWR and other match measurements
can be calculated and displayed.
Note: The 5010B replaces the Bird 5010.
The 5010 was only compatible with APM/DPM elements, and lacked peak-reading
capability. The 5000-EX will still recognize the 5010.
Terminating Power Sensors (TPS)
Highly accurate terminating sensors; do not require elements. The Bird 5011
frequency range is 40 MHz 4 GHz. The 5011-EF frequency range is 40 MHz 12
GHz.
1
Bird Model 5000-EX Digital Power Meter
Wideband Power Sensor (WPS)
Bird 5012 Thruline Sensor that does not require elements. Can measure average,
peak, or burst power, VSWR, crest factor, and Complementary Cumulative
Distribution Function (CCDF). A status LED on the front lights when the WPS is
powered, and blinks when the WPS is connected to the DPM.
– Automobile Cigarette Lighter Adapter
(P/N 5A2238-1) Connects the DPM or WPS to a standard 12V automotive cigarette
lighter jack.
– Virtual Power Meter
(VPM) For the WPS only. PC software that can interface directly with the WPS
without requiring a DPM or `black box’. Can also log data for analysis,
printing, or storage.
System requirements: IBM PC or equivalent; Windows 95 or later, 4 MB free hard
disk space; available Serial or USB Port. Includes interface cable.
– Serial to USB Adapter
(P/N DC-DB9-U) Converts the serial cable to USB. Connects the DPM to a PC’s
USB port if the serial port is unavailable.
– Attenuators & Accessories
A variety of attenuators and connectors for measuring large powers with the
TPS. For a complete list, see page 26.
– Soft Carry Case
(P/N 5000-030) Convenient and protective. Cutouts allow operation while in the
case.
– Hard Carry Case
(P/N 5000-035) Protective case holds DPM, sensor, and accessories.
2
Component Description
11
10
Introduction
12 13
1
Model 5000-EX 2
9
SHIFT
8
7 6
5
Digital Power Meter
4
3 Figure 1Component Description
3
Bird Model 5000-EX Digital Power Meter
1. Speed Keys
Activates the command displayed above it.
2. Backlight Key Turns the backlight on momentarily.
3. On Key
Turns the DPM on or off.
4. +/- Key
Toggles between positive and negative numbers.
5. Enter Key
Completes data entry.
6. Escape Key
Aborts data entry without accepting changes. For WPS, in second set of softkeys, selects previous measurement mode.
7. Offset Key
Enter offset values in dB.
8. Shift Key
Toggles between different sets of speed key functions.
9. Numeric Keys Input numeric values.
10. Sensor Port Connection for power sensors.
11. RS-232 Port
Connection for an optional PC 9-pin RS-232 (DB9) connector, compatible with PC serial port. 9600 baud, 8 data bits, 1 stop bit, and no parity.
12. LCD Display Backlit liquid crystal display.
13. External DC Connector
Connect either the ac adapter or the cigarette lighter adapter. External supplies power the unit and charge the internal battery.
4
Display Description
1
2
Introduction
3
10
.0.0.0.0.0 Over
Fwd Under
No Sensor Error
4
9
MHz kHz Average Peak Burst Crest CCDF dBm µW mW kW
.0.0.0.0.0 00 Over Offset
Rfl Under Setup
% Duty
5
8
MHz kHz Average Rho VSWR RL dBm µW mW kW
6
Alarm
Type
Filter Frequency
7
Limit
Fwd
Rfl
Factor
ALARM
Scale
Units
Units
Setup
Zero
Figure 2Display Description
Note: Numbers in parentheses (1) refer to
the labeled parts of the display above.
– Display (1)
The DPM-EX has two measurement displays. The top displays forward power, crest
factor, and CCDF. The bottom displays reflected power and reflection
measurement (VSWR).
– Display Units (2)
Shows the measurement mode and units for the display directly above.
5
Bird Model 5000-EX Digital Power Meter
– Battery Level Indicator (3)
When the external adapter is connected, the indicator blinks until the battery
is fully charged. When using the internal batteries, the indicator is on
continuously and the black bars show the battery change remaining. When the
battery is low, the indicator will be empty and will blink. Recharge the
batteries. – No Sensor Indicator (4)
Turns on when no sensor is connected to the sensor port. – Duty Cycle
Indicator (5)
For burst power measurements, shows the signal’s duty cycle. – Analog Bar
Graph (6)
Proportional to the forward average power. Minimum at 0 W or 20 dBm. Max at
full scale power. – Speed Key Labels (7)
Press the speed key below the label to activate the displayed function. –
Setup Indicator (8)
Turns on when a stored setup is being used. – Offset Indicator (9)
Blinks when an offset is in use. – Power Indicator (10)
If the power exceeds 100% of full scale, “Over” will be displayed. If the
power is less than zero, “Under” will be displayed. If the power exceeds 120%
of full scale, the display will show a line of dashes and the bar scale will
be filled.
6
Introduction
Speed Keys
– Scale
For element-based sensors, sets the forward full scale power. This is listed
on the forward element’s nameplate. Reflected full scale power is
automatically set to 10 dB below the forward full scale. This speed key is
disabled for non-element-based sensors.
– Fwd Units
Selects the units for the top display.
– Rfl Units
Selects the units for the bottom display. If the bottom display is not in use,
or does not have selectable units this key will be disabled.
– Setup
For the DPS, enter a setup number to load a stored setup. 0 returns to the
default setups. For the WPS enter the threshold power, in Watts, for CCDF
mode.
– Zero
For the TPS and the WPS, does a zero-power calibration. This speed key is
disabled for all other sensors.
– Type
For the DPS, selects either average (APM/DPM) or peak-reading (43) elements.
For the WPS, selects the measurement mode. In the filter screen, selects the
filter bandwidth. – Filter
For the WPS, opens the filter screen for selecting the filter bandwidth. This
speed key is disabled for all other sensors.
7
Bird Model 5000-EX Digital Power Meter
8
Chapter 2
Installation
Power Supply
The DPM uses a rechargeable Nickel-Metal Hydride battery pack. Charge life is
about 20 hours with the Bird WPS, 50 hours with other sensors.
The DPM can use an external power source. Using the DPM with the ac adapter or
the 12V cigarette lighter adapter will also charge the battery. Charging time
from full discharge is 8 hours using the ac adapter. When using the cigarette
lighter adapter, charge time will depend on the car battery charge. When the
external adapter is connected, the battery level indicator will blink until
the battery is fully charged. When using the internal batteries, the indicator
will be on continuously and the black bar will show the battery change
remaining.
Note: For optimum battery life, only
charge the batteries after the battery level indicator empties completely and
begins to blink.
CAUTION Only use the ac adapter provided with the DPM. Do not use the adapter
with the batteries removed.
” AC Mains Adapter — To use the ac adapter. insert the adapter’s barrel
connector into the DPM’s external dc connector (See Figure 1 on page 3).
Insert the adapter plug into a wall receptacle.
” Automobile Cigarette Lighter Adapter — Insert the adapter’s barrel connector
into the DPM’s external dc connector. Insert the adapter plug into a cigarette
lighter jack.
9
Bird Model 5000-EX Digital Power Meter
Connections
WARNING Never attempt to connect or disconnect RF equipment from the
transmission line while RF
power is being applied. Leaking RF energy is a potential health hazard.
Figure 3DPM Connections 10
Directional Power Sensor (DPS)
Installation
WARNING RF voltage may be present in RF element socket.
Keep element in socket during operation.
Connect the Bird DPS to the “Sensor” port on the DPM using the sensor cable
provided.
Connect the DPS to the RF line so that the arrow on the sensor points towards
the load. The arrow on the forward element should point towards the load. The
arrow on the reflected element should point towards the transmitter. The
forward element’s power rating must be 10x the reflected’s rating. Both
elements must be either APM/DPM or 43 types, do not mix elements.
Figure 4 DPS
Element Orientation
Wideband Power Sensor (WPS)
Connect the DPM port on the Bird WPS to the “Sensor” port on the DPM using the
sensor cable provided. Connect the WPS to the RF line so that the arrow on the
sensor points towards the load.
Note: The WPS is not compatible with the
Bird PCTool. Use the VPM software instead. Refer to the VPM Quickstart for
connection instructions.
11
Bird Model 5000-EX Digital Power Meter
Terminating Power Sensor (TPS)
CAUTION Discharge all static potentials before connecting
the TPS(-EF). Electrostatic shock could damage the sensor.
CAUTION When connecting the TPS or the TPS-EF, only turn
the connector nut. Damage may occur if torque is applied to the sensor body.
CAUTION Do not exceed 2 W average or 125 W peak power for
5 µs when using the TPS or the TPS-EF. Doing so will render the sensor
inoperative.
Connections are the same for the Bird TPS and TPS-EF. Connect the TPS to the
“Sensor” port on the DPM using the sensor cable provided.
In most cases, an attenuator or directional coupler should be used with the
TPS. For example, for an RF source with output between 0.1 and 50 W, use a 40
dB, 50 W attenuator. Put the attenuator between the TPS and the source.
Note: Only connect the TPS directly to a
source if the RF power will be less than 10 mW.
12
Chapter 3
Operation
The Bird Digital Power Meter is very easy to operate. Once a sensor is
connected, turn the DPM on and take a reading. Additional commands are
available, depending on the sensor used.
Normal Operation
” Connect the sensor.
” Turn on the DPM.
” Use the speed keys to set the measurement and measurement units. Refer to
the sections following for instructions specific to each sensor.
” If you know the system loss or are using an attenuator, add the losses (in
dB) of all components in the system. For attenuators and other
frequencydependent components, use the loss at the measured frequency. Then,
press OFFSET and enter the total loss in dB. This will allow you to read the
actual line power. The DPM accepts offsets from 10 to 100 dB, depending on
the sensor.
” Turn on the RF source.
” Take a reading.
Note: The analog bar graph will respond
immedi-ately to changes in the RF power. The major and minor displays will
respond after a delay of 1 to 13 seconds, depending on the power level.
13
Bird Model 5000-EX Digital Power Meter
DPM Shutdown
When Powered by Internal Batteries
To shut down the DPM when it is powered from its internal batteries, press the
On key one time.
When Powered by an External Source
To shut down the DPM when it is powered from an external source connected to
the External DC Connector jack at the side of the unit, make sure that the
backlight is extinguished then press the ON key one time.
Note: If you turn off the DPM while the
backlight is lighted, you will need to disconnect the external power source
for at least 30 seconds before turning the DPM on.
Directional Power Sensor (DPS)
WARNING RF voltage may be present in RF element socket.
Keep element in socket during operation.
Setting Scale
The Bird DPS uses Bird Plug-In Elements. These are labeled with a max power
and a frequency range. The transmitter frequency should be within the element
range, and the forward element power should be 10x the reverse element power.
Forward full scale power must be entered manually. Reflected full scale is
automatically set to 1/10 of the forward full scale.
14
Operation
” Press Scale. ” Change the units with the Fwd Units key until the
DPM units match the units on the forward element. ” Use the numeric keypad to
enter the maximum
power of the element in the forward element socket.
Note: The element’s max power is listed on
the element nameplate.
” Press ENTER.
Terminating Power Sensor (TPS)
CAUTION Do not exceed 2 W average or 125 W peak power for
5 µs when using the TPS or the TPS-EF. Doing so will render the sensor
inoperative.
Zeroing Sensor
Over time, the sensor’s “zero value” (reading with no applied RF power) can
drift, making all readings inaccurate by this value. For example, if the zero
value is 2 µW, measuring a 5 mW signal will give a reading of 4.998 mW, a
0.04% error. Measuring a 50 µW signal will give a reading of 48 µW, a 4%
error. If the drift would be a significant error, rezero the sensor: ” Make
sure the sensor has been connected to the
DPM and the DPM turned on for at least 5 minutes. ” Make sure no RF power is
applied to the sensor. ” Press Zero. “CAL 0″ will be displayed and
calibration will begin. ” Zeroing the sensor takes 60 seconds. The bar graph
will display calibration progress. When complete, “0 CAL PASS” should be
displayed. Press ENTER to return to normal operation. If “0 CAL FAIL” is
displayed, make sure no RF power is applied to the sensor, press ENTER and
zero the sensor again.
15
Bird Model 5000-EX Digital Power Meter
TPS-EF
The Bird TPS-EF uses frequency correction factors to allow more accurate
measurements. Look at the label on the side of the sensor and find the
correction factor for the frequency being measured. Add the correction factor
to the other attenuation or coupling factors and enter this as an offset.
Wideband Power Sensor (WPS)
Zeroing Sensor
Over time, the sensor’s “zero value” (reading with no applied RF power) can
drift, making all readings inaccurate by this value. For example, if the zero
value is 0.02 W, measuring a 50 W signal will give a reading of 49.98 W, a
0.04% error. Measuring a 1 W signal will give a reading of 0.98 W, a 2% error.
If the drift would be a significant error, rezero the sensor: ” Make sure the
sensor has reached a stable
operating temperature. ” Make sure no RF power is applied to the sensor. ”
Press Zero. Calibration will begin. ” Calibration will take about 30 seconds.
Do not
interrupt the calibration! A bar on the screen will display calibration
progress. ” After a successful calibration, “Cal Pass” or “Calibration
Complete” will be displayed. Press any key to return to normal operation. ” If
calibration fails, “Cal Fail” will be displayed. Press a key to return to
normal operation, then check that the WPS is properly connected, and that the
RF is off. Rezero.
16
Selecting Measurement
Operation
The Bird Wideband Power Sensor can measure Average, Peak, and Burst power as
well as Crest Factor and CCDF. To select the measurement:
” Press SHIFT to display the second set of softkeys. Type should be key 2 and
Filter should be key 3.
” Press Type to cycle through the possible modes.
Video Filter
Except for average power and VSWR measurements, all measurements use a variable video filter to improve accuracy. This filter can be set to either 4.5 kHz, 400 kHz, or Full bandwidth. It should be as narrow as possible while still being larger than the demodulated signal bandwidth (video bandwidth). Narrowing the filter limits the noise contribution from interfering signals. Listed below are some common modulation schemes and the appropriate video filter.
Video Filter 4.5 kHz 400 kHz Full Bandwidth
Modulation Type
CW Burst (Burst width > 150 µs), Voice Band AM, FM, Phase Modulation, Tetra
CW Burst (b.w. > 3 µs), GSM, 50 kHz AM, DQPSK (¼, symbol rate < 24 k/s)
CW Burst (b.w. > 200 ns), CDMA, WCDMA, DQPSK (¼, symbol rate < 200 k/s),
DAB/DVB-T
17
Bird Model 5000-EX Digital Power Meter
Average Mode
This mode displays the average forward and reflected power. In average mode,
the VSWR or system match can be displayed instead of the reflected power. ”
Press SHIFT to display the first set of softkeys.
Press Fwd Units or Rfl Units to cycle through the possible units.
Peak Mode
This mode displays the peak envelope power. ” Press SHIFT to display the first
set of softkeys. Press
Fwd Units to cycle through the possible units.
Burst Mode
This mode displays the average power in a burst. The burst’s duty cycle is
measured by the DPM. ” Press SHIFT to display the first set of softkeys. Press
Fwd Units to cycle through the possible units.
Crest Factor Mode
The crest factor is the ratio of the forward peak power and the forward
average power. It is measured in dB. There are no additional controls in this
mode.
CCDF Mode
The CCDF measures the percentage of time the power level is above a threshold.
To set the power threshold: ” Press Setup. Enter the threshold power (in
Watts)
and press ENTER.
18
Chapter 4
Maintenance
Cleaning
CAUTION Do not use harsh or abrasive detergents for cleaning.
Clean the Bird Digital Power Meter and its display with a soft cloth dampened
with mild detergent and water only. Clean sensors with a dry cleaning solvent
that leaves no residue.
Charging Batteries
Fully charged batteries provide about 20 hours of continuous operation with
the Bird WPS, 50 hours with other sensors. Charging time is typically 8 hours
using the ac adapter. The batteries charge whenever the DPM is connected to ac
or dc power sources, using either the ac Mains adapter or the automobile
cigarette lighter adapter. When the external adapter is connected, the battery
level indicator will blink until the battery is fully charged. When using the
internal batteries, the indicator will be on continuously and the black bar
will show the battery change remaining. The unit will charge with its power
turned either on or off.
Note: For optimum battery life, only
charge the batteries after the battery level indicator empties completely and
begins to blink.
19
Bird Model 5000-EX Digital Power Meter
Troubleshooting
PROBLEM
POSSIBLE CAUSE
REMEDY
Nothing shown on display
Display shows dashes; bar scale is blank
Display shows “Bat Error”; bar scale is blank Display shows dashes and
“Overrange”; bar scale is full WPS Status LED is not on
Unit is off
Batteries are drained
No sensor connected Sensor communication has failed Batteries are drained
Unit is overrange
WPS not powered
Momentarily press the ON key. Use external power source. (page 9) Replace the
batteries. (page 22) Connect a sensor.
Use a different cable. Use a different sensor.
Use external power source. (page 9)
Use higher power elements (if applicable), or reduce RF power.
Check connection to DPM Connect WPS to a power supply using the ac adapter
20
Maintenance
PROBLEM Erratic power readings
Unit will not turn ON
POSSIBLE CAUSE
REMEDY
Element contact out of alignment (DPS)
Align the contact. It must be far enough out to make good contact with the element, but must not restrict entry of the element body.
Damaged
Replace element.
element (DPS)
Sensor has lost Rezero sensor. its zero (TPS)
Sensor is damaged
Replace sensor.
Unit is using ac power adapter and was turned OFF while the back light was illuminated.
Disconnect the ac adapter connector from the unit and wait for at least 30 seconds, then turn the unit ON.
21
Bird Model 5000-EX Digital Power Meter
Battery Replacement
The Nickel-Metal Hydride (Ni-MH) batteries do not normally need to be
replaced. If necessary, however, follow these instructions (see Figure 5).
WARNING Disconnect from external power before any disassembly. The potential
for electric shock exists.
CAUTION Replace only with Ni-MH rechargeable A batteries.
Nominal Voltage 1.2V; Capacity 2700mAhr. Use Sanyo P/N HR-AUX or equivalent.
Battery Connector Figure 5Back Cover Removal
22
Maintenance
” Lay the DPM, display side down, on a clean surface. ” Use a small
screwdriver to remove all six screws
from the back cover. ” Taking care to not disconnect the battery connector,
lift the back cover off, flip it over, and lay it, battery side up, next to
the front cover. ” Remove the old batteries. ” Install the new batteries
checking the orientation of the positive and negative terminals. ” Make sure
the battery connector is connected. ” Taking care to not pinch the connector
wires, put the covers back together and screw into place.
23
Bird Model 5000-EX Digital Power Meter
Customer Service
Any maintenance or service procedure beyond the scope of those in this chapter
should be referred to a qualified service center. If you need to return the
unit for any reason, contact the Bird Service Center for a return
authorization. All instruments returned must be shipped prepaid and to the
attention of Bird Service Center. Bird Service Center 30303 Aurora Road
Cleveland (Solon), Ohio 44139-2794 Phone:
440-519-2298 Fax:
440-519-2326 E-mail: bsc@bird-
technologies.com For the location of the Sales Office nearest you, give us a
call or visit our Web site at: http://www.bird-electronic.com
24
Parts List
Part Name
Digital Power Meter
AC adapter 120V ac 230V ac
Cigarette Lighter Adapter
Directional Power Sensor
DPM Elements See P/N 871-DPM-019-901, the DPM Element Guide, for a complete
list of elements
Terminating Power Sensor 40 MHz 4 GHz 40 MHz 12 GHz
Wideband Power Sensor
Soft Carry Case
Hard Carry Case
PC Interface Virtual Power Meter (WPS)
Serial to USB Adapter
Instruction manual
Maintenance
Part Number 5000-EX
5A2229 5A2226 5A2238-1 5010B Various
5011 5011-EF 5012 5000-030 5000-035
VPM DC-DB9-U 920-5000-EX
25
Bird Model 5000-EX Digital Power Meter
Attenuators & Accessories
N(F) N(M) Attenuators (RF power range with TPS)
30 dB (10 mW 10 W) 40 db (100 mW 50 W)
8353A03010 8353A04050
DC Block
5011A0351
N(F) N(M) Test Cable, 1.5 m
TCMNFN1.5G
N(F) N(M) Armored, Phase Stable Test Cable
1.5 m 3.0 m
TCMNFN1.5 TCMNFN3.0
Calibration Data
5011CALDATA
Recommended for attenuators, test cables, dc block, and right angle adapter
N(F) N(F)
N(F) SMA(F) N(F) 7/16 DIN(F)
Adapters
4240-500-1 N(F)N(M) Right Angle
4240-500-4 N(F)SMA(M)
PA-FNFE
N(F) 7/16 DIN(M)
4240-500-3
4240-500-5 PA-FNME
26
Specifications
Maintenance
Bird 5000-EX Digital Power Meter
Display
Backlit LCD with major and minor digital displays, analog bar graph, and speed key labels
Analog Bar Graph
20 segment, horizontal. Tracks the major display
Measurement Modes Determined by the sensor
Sensor Interface
9-pin RS-232 serial port. Sensor is powered from the DPM.
PC Interface
9-pin RS-232 serial port
Dimensions
8.0″H x 4.625″W x 2.0″D (203 x118 x 51 mm)
Weight
2 lbs. (0.9 kg) nominal
Battery
Rechargeable Nickel-Metal Hydride (NiMH) battery. 8 hours for full charge
AC Power
Input Output
Operates from ac mains using supplied adapter 120 ± 10% Vac @ 50/60 Hz 12 Vdc, < 400 mA
Operating Temp.
0 to +50 °C (32 to 122 ° F)
Storage Temp.
20 to +50 °C (4 to +122 ° F)
Humidity, Max.
95% (non-condensing)
Altitude
15,000 ft. (4500 m) operating
Mechanical Shock and In accordance with MIL-T-28800D
Vibration
Class 3
27
Bird Model 5000-EX Digital Power Meter
Bird 5010B Directional Power Sensor
Sensor Type
Bird Thruline directional twoelement line section.
Elements Frequency Range*
APM/DPM or 43 series elements. Select two from the same series, with RFL power
1/10 of FWD power
2 3600 MHz
Average Power Measurement, APM/DPM Elements, Forward or Reflected Direction
RF Power Range* 0.1 W to 1 kW
Uncertainty
± 5% of reading (95% c.l.)
Peak/Average Ratio, Max
10 dB
Average Power Measurement, 43 Elements, FWD or RFL
RF Power Range* 0.1 W to 10 kW
Uncertainty
± 5% of full-scale power (95% c.l.)
Peak Power Measurement, 43 Elements, FWD only
Pulse Width, Min
2 25 MHz 15 µs 25 100 MHz1.5 µs > 100 MHz 800 ns
Rep. Rate, Min Duty Cycle, Min Uncertainty
15 pps 1 x 104
± 8% of full-scale peak envelope power (95% c.l.)
28
Maintenance
Match Measurement
Measurement Range
Return Loss Rho () VSWR
0 to 20 dB 0.1 to 1 1.22 to 99.99
Uncertainty
Calculated from forward and reflected uncertainty. Twice the Avg Power Uncertainty
Settling Time, Max
2.5 seconds
Impedance, Nominal 50 ohms
Insertion Loss, Max 0.05 dB up to 1 GHz
Input VSWR, Max.
1.05:1 up to 1 GHz
Directivity, Typical* RF Connectors
30 dB QC Type, N(F) normally supplied
Power Supply
From host instrument via cable
Mechanical Shock and In accordance with MIL-T-28800D
Vibration
Class 3
CE Temp, Operating
CE compliant. Refer to DOC for specific standards.
10 to +50 °C (+14 to +122 °F)
Temp, Storage
40 to +75 °C (40 to +167 °F)
Humidity, Max
95% (non-condensing)
Altitude, Max
3,000 m (10,000 ft.)
Dimensions, Nominal 2.5″ x 5.0″ x 2.0″ (59 x 127 x 51 mm)
Weight, Nominal
1.25 lb. (0.6 kg)
- Exact value depends on element selected Above 35 °C or below 15 °C add 2%
29
Bird Model 5000-EX Digital Power Meter
Bird 5011 and 5011-EF Terminating Sensors
General
Sensor Type
Frequency Range 5011 5011-EF
RF Power Range
Max Power (Damage Level) Peak/Average Ratio Uncertainty*
RF Connector 5011 5011-EF
Impedance, Nominal Input VSWR: 5011
Typical Maximum 5011-EF Typical Maximum Power Supply
Corrected terminated average power Diode based terminated true average power
measurement
40 MHz to 4 GHz 40 MHz to 12 GHz 10 µW to 10 mW (20 dBm to +10 dBm) 2 W avg.
125 W peak for 5 µs 12 dB max ±(5% of reading ± 1 µW) (95% c.l.) (excluding
mismatch uncertainty) (with correction factors for -EF)
N Male Precision N Male 50 ohms
1.03 (36.6 dB return loss) 1.20 (20.8 dB return loss)
1.05 (32.0 dB return loss) 1.25 (19.1 dB return loss) From host instrument via
cable
30
Maintenance
Temp, Operating
10 to +50 °C (14 to 122 °F)
Temp, Storage
40 to +80 °C (40 to +176 °F)
Mechanical Shock
IAW MIL-PRF-28800F class 3
Vibration
IAW MIL-PRF-28800F class 3
Humidity, Max
95% (non-condensing)
Altitude, Max
4,500 m (15,000 ft.)
CE
CE compliant
ROHSS
ROHSS compliant
Dimensions, Nominal 6″ long x 1.5″ diameter
Weight, Max
0.75 lb. (0.35 kg)
Recommended Calibration Interval
12 Months
- 5011: Above 40 °C, when making measurements at frequencies between 40 and 100 MHz, add 1%.
5011-EF: Above 40 °C or below 10 °C, add 1%.
31
Bird Model 5000-EX Digital Power Meter
Bird 5012 Wideband Sensor
Sensor Characteristics
Frequency Range
RF Power Range
Maximum Power
Impedance, Nominal
Insertion Loss, Max 0.35 1 GHz 1 4 GHz
Input VSWR, Max 0.35 2.5 GHz 2.5 4 GHz
Directivity, Min 0.35 3 GHz 3 4 GHz
RF Connectors
Interface: DPM
PC Serial Port
PC USB Port
Power Supply: DPM USB Port DC Connector
350 MHz to 4 GHz 0.15 W 150 W Average, 4.0 400 W Peak See Figure 7 on page
34 50 ohms
0.05 dB 0.1 dB
1.05:1 1.10:1
30 dB 28 dB N Female
Male DB-9, EIA-232, 9600 baud, no parity, 8 data bits, 1 stop bit Fem. DB-9,
EIA-232, 9600 baud, no parity, 8 data bits, 1 stop bit USB 1.1 interface
From host instrument via cable less than one low-power USB load 7 18 Vdc, <
100 mA
32
Average Power
Maintenance
RF Power Range Peak/Average Ratio, Max Measurement Uncert.
0.15 150 W 12 dB ± (4% of reading + 0.05 W)*
- Above 35 °C or below 15 °C add 3%
Match Measurement
Measurement Range Return Loss Rho () VSWR
Forward Power, Min
Measurement Uncert.
0 to 23 dB 0.07 to 1.0 1.15 to 99.9
0.5 W
See Figure 6 on page 34
Peak Envelope Power
RF Power Range
Measurement Uncert. burst width > 200 µs 1 µs < b.w. < 200 µs burst width < 1
µs burst width < 0.5 µs
4.0 400 W* ± (7% of reading + 0.2 W) ± (10% of reading + 0.4 W) ± (15% of reading + 0.4 W) ± (20% of reading + 0.4 W)
- Max. power depends on frequency and system VSWR. See Figure 7 on page 34
Above 35 °C or below 15 °C add 3% For D < 0.1 add 0.1 W For period > 0.1s add (1.5% + 0.15 W)
33
Bird Model 5000-EX Digital Power Meter
Uncertainty (dB)
Match Measurement Uncertainty
4 Above 3 GHz
2
Below 3 GHz
0
2
4
6
0
5
10
15
20
25
Return Loss (dB)
Figure 6Match Measurement Uncertainty
Maximum Peak Power
1000
VSWR = 1
VSWR = 1.5
VSWR = 3
400
Peak Power (Watts)
100 0.3
34
1
4
Frequency (GHZ)
Figure 7Max. Peak Power
Maintenance
Burst Average Power
RF Power Range Burst Width Repetition Rate, Min Duty Cycle (D)
Measurement Uncert.
2 150 W average
1 µs 50 ms
15 Hz
0.001 1 D = Burst Width / Period ± (6% of reading + 0.05/D W)*
- Above 35 °C or below 15 °C add 3%
Crest Factor
RF Power Range Measurement Uncert.
0.15 150 W
Linear sum of peak and average power uncertainty
Complementatry Cumulative Distribution Function (CCDF)
Measurement Range Measurement Uncert. Threshold Level Range Level Set Accuracy
0.1 100% ± 0.2% 2 400 W As peak power uncert. + 2%
Physical and Environmental Specifications
Temp, Operating Temp, Storage Mechanical Shock and Vibration Humidity, Max
Altitude, Max Dimensions, Nominal
Weight, Max
10 to +50 °C (+14 to +122 °F) 40 to +80 °C (40 to +176 °F) MIL-PRF-28800F
class 3
95% (non-condensing) 15,000 ft. (4,500 m) 4.75″ x 4.6″ x 1.3″ (121 x 117 x 33
mm) 1.2 lb. (0.55 kg)
35
Bird Technologies
Limited Warranty
All products manufactured by Seller are warranted to be free from defects in
material and workmanship for a period of one (1) year, unless otherwise
specified, from date of shipment and to conform to applicable specifications,
drawings, blueprints and/or samples. Seller’s sole obligation under these
warranties shall be to issue credit, repair or replace any item or part
thereof which is proved to be other than as warranted; no allowance shall be
made for any labor charges of Buyer for replacement of parts, adjustment or
repairs, or any other work, unless such charges are authorized in advance by
Seller. If Seller’s products are claimed to be defective in material or
workmanship or not to conform to specifications, drawings, blueprints and/or
samples, Seller shall, upon prompt notice thereof, either examine the products
where they are located or issue shipping instructions for return to Seller
(transportation charges prepaid by Buyer). In the event any of our products
are proved to be other than as warranted, transportation costs (cheapest way)
to and from Seller’s plant, will be borne by Seller and reimbursement or
credit will be made for amounts so expended by Buyer. Every such claim for
breach of these warranties shall be deemed to be waived by Buyer unless made
in writing within ten (10) days from the date of discovery of the defect. The
above warranties shall not extend to any products or parts thereof which have
been subjected to any misuse or neglect, damaged by accident, rendered
defective by reason of improper installation or by the performance of repairs
or alterations outside of our plant, and shall not apply to any goods or parts
thereof furnished by Buyer or acquired from others at Buyer’s request and/or
to Buyer’s specifications. Routine (regularly required) calibration is not
covered under this limited warranty. In addition, Seller’s warranties do not
extend to the failure of tubes, transistors, fuses and batteries, or to other
equipment and parts manufactured by others except to the extent of the
original manufacturer’s warranty to Seller. The obligations under the
foregoing warranties are limited to the precise terms thereof. These
warranties provide exclusive remedies, expressly in lieu of all other remedies
including claims for special or consequential damages. SELLER NEITHER MAKES
NOR ASSUMES ANY OTHER WARRANTY WHATSOEVER, WHETHER EXPRESS, STATUTORY, OR
IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS, AND NO PERSON IS
AUTHORIZED TO ASSUME FOR SELLER ANY OBLIGATION OR LIABILITY NOT STRICTLY IN
ACCORDANCE WITH THE FOREGOING.
36
References
- technologies.com
- Test Equipment Rentals, Sales, Calibration | ATEC
- Bird | RF Test & Measurement | Precision & Calibration | RF Recorders & Playback
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>