TEKBOX TBPCP2-3070 RF Pulse Current Monitoring Probe User Manual
- June 9, 2024
- TEKBOX
Table of Contents
TBPCP2-3070 RF Pulse Current Monitoring Probe
User Manual
Introduction
The TBPCP2-3070 is a RF pulse current monitoring probe, expanding the Tekbox
product range of affordable test equipment.
The probe has a very flat response from 30 Hz to 70 MHz and is characterized
over the frequency range from 1 Hz to 200 MHz. The TBPCP2-3070 is typically
used for surge or RF pulse current monitoring applications in the time domain,
in contrary to RF current monitoring probes designed for EMC applications,
which are used for measurements in the frequency domain.
The aperture of the RF current monitoring probe is 25 mm. Its transfer impedance measured in a 50Ω + 50Ω loop test fixture is -26 dB Ohm with a 3dB bandwidth from 30Hz to 70 MHz.
Specification
Characterized frequency range: 1 Hz to 200 MHz
3 dB bandwidth: 30 Hz to 70 MHz (measured in a 50 + 50 Ohm loop)
Transfer impedance into 50 Ohm load: -26 dB Ohm; 0.05 V/A
Transfer impedance into high Z: 0.1 V/A
Probe port impedance: 50 Ohm
Droop rate: < 20% / ms
Rise time: < 5 ns
Max. primary RMS AC current with respect to internal resistor dissipation
ratings: 40A
Max. primary single pulse current with respect to internal resistor pulse
current ratings: 200A@ 100ms pulse width; 400A @ 1ms pulse width
Max. current time product: 0.01 Ampere seconds
Max. RMS AC current, linear: 4 A
Max. core temperature: 80 °C
Aperture diameter: 5.5 mm
Outside diameter: 27 mm
Length including SMA connector: 40mm
Height: 19 mm
Weight: 60 g
Connector type: SMA female
Transfer impedance
Typical transfer impedance table
The table below shows typical transfer impedance data of a TBPCP2-3070 pulse
current probe. Each current probe is delivered with its corresponding
measurement protocol. The trans impedance is given for a 50 Ohm load.
Logarithmic to linear conversion:
Trans impedance [V/A] = 10^(dBOhm/20) @ 50 Ohm
Trans impedance [V/A] = 2*10^(dBOhm/20) @ High Z
Frequency| transfer impedance [dBΩ] 50 Ohm load| transfer
impedance [V/A] 50 Ohm load| transfer impedance [V/A] high Z load
---|---|---|---
1 Hz| -56.| 0,00| 0.00
2,5 Hz| -48.| 0,00| 0.01
5 Hz| -42.| 0,01| 0.02
7.5 Hz| -39.| 0,01| 0.02
10 Hz| -37.| 0,01| 0.03
12.5 Hz| -35.| 0,02| 0.04
15 Hz| -33.| 0,02| 0.04
17.5 Hz| –32.30| 0,02| 0.05
20 Hz| -32.| 0,03| 0.05
25 Hz| -30.| 0,03| 0.06
50 Hz| -28.| 0,04| 0.08
75 Hz| -27.| 0,05| 0.09
100 Hz| -27.| 0,05| 0.09
1 kHz| -26.| 0,05| 0.10
10 kHz| -26.| 0,05| 0.10
100 kHz| -26.| 0,05| 0.10
1 MHz| -26.| 0,05| 0.10
10 MHz| -26.| 0,05| 0.10
25 MHz| -26.| 0,05| 0.10
50 MHz| -26.| 0,05| 0.10
60 MHz| -25.| 0,05| 0.11
70 MHz| -25.| 0,06| 0.12
80 MHz| -24.| 0,06| 0.12
90 MHz| -23.| 0,07| 0.14
100 MHz| -23.| 0,07| 0.14
110 MHz| -22.| 0,08| 0.16
120 MHz| -21.| 0,09| 0.18
130 MHz| -20.| 0,10| 0.20
140 MHz| -19.| 0,12| 0.23
150 MHz| -17.| 0,14| 0.27
160 MHz| -16.| 0,16| 0.32
170 MHz| –14.40| 0,19| 0.38
180 MHz| -13.| 0,23| 0.46
190 MHz| -11.| 0,29| 0.57
200 MHz| -9.| 0,36| 0.73
Table 1: Transfer impedance: 1 Hz to 200 MHz, typical data
Accessory
Tekbox supplies a calibrator suitable for the TBPCP2-3070 current probe:
Ordering Information
Part Number | Description |
---|---|
TBPCP2-3070 | Pulse Current Probe 30Hz – 70 MHz |
TBPCP2-CAL | Calibration fixture for TBPCP2-3070 current probe |
History
Version | Date | Author | Changes |
---|---|---|---|
V 1.0 | 9.11.2022 | Mayerhofer | Creation of the preliminary document |
www.tekbox.com
TekBox Digital Solutions Vietnam Pte. Ltd.
Factory 4, F5, Lot I-3B-1, Saigon Hi-Tech Park,
Tan Phu Ward, District 9, Ho Chi Minh City, Vietnam
References
Read User Manual Online (PDF format)