TEKBOX TBPCP2-3070 RF Pulse Current Monitoring Probe User Manual

June 9, 2024
TEKBOX

TBPCP2-3070 RF Pulse Current Monitoring Probe
User Manual

Introduction

The TBPCP2-3070 is a RF pulse current monitoring probe, expanding the Tekbox product range of affordable test equipment.
The probe has a very flat response from 30 Hz to 70 MHz and is characterized over the frequency range from 1 Hz to 200 MHz. The TBPCP2-3070 is typically used for surge or RF pulse current monitoring applications in the time domain, in contrary to RF current monitoring probes designed for EMC applications, which are used for measurements in the frequency domain.

TEKBOX TBPCP2 3070 RF Pulse Current Monitoring Probe -
Introduction

The aperture of the RF current monitoring probe is 25 mm. Its transfer impedance measured in a 50Ω + 50Ω loop test fixture is -26 dB Ohm with a 3dB bandwidth from 30Hz to 70 MHz.

Specification

Characterized frequency range: 1 Hz to 200 MHz
3 dB bandwidth: 30 Hz to 70 MHz (measured in a 50 + 50 Ohm loop)
Transfer impedance into 50 Ohm load: -26 dB Ohm; 0.05 V/A
Transfer impedance into high Z: 0.1 V/A
Probe port impedance: 50 Ohm
Droop rate: < 20% / ms
Rise time: < 5 ns
Max. primary RMS AC current with respect to internal resistor dissipation ratings: 40A
Max. primary single pulse current with respect to internal resistor pulse current ratings: 200A@ 100ms pulse width; 400A @ 1ms pulse width
Max. current time product: 0.01 Ampere seconds
Max. RMS AC current, linear: 4 A
Max. core temperature: 80 °C
Aperture diameter: 5.5 mm
Outside diameter: 27 mm
Length including SMA connector: 40mm
Height: 19 mm
Weight: 60 g
Connector type: SMA female

Transfer impedance

TEKBOX TBPCP2 3070 RF Pulse Current Monitoring Probe - Transfer
impedance

Typical transfer impedance table

The table below shows typical transfer impedance data of a TBPCP2-3070 pulse current probe. Each current probe is delivered with its corresponding measurement protocol. The trans impedance is given for a 50 Ohm load.
Logarithmic to linear conversion:
Trans impedance [V/A] = 10^(dBOhm/20) @ 50 Ohm
Trans impedance [V/A] = 2*10^(dBOhm/20) @ High Z

Frequency| transfer impedance [dBΩ] 50 Ohm load| transfer impedance [V/A] 50 Ohm load| transfer impedance [V/A] high Z load
---|---|---|---
1 Hz| -56.| 0,00| 0.00
2,5 Hz| -48.| 0,00| 0.01
5 Hz| -42.| 0,01| 0.02
7.5 Hz| -39.| 0,01| 0.02
10 Hz| -37.| 0,01| 0.03
12.5 Hz| -35.| 0,02| 0.04
15 Hz| -33.| 0,02| 0.04
17.5 Hz| –32.30| 0,02| 0.05
20 Hz| -32.| 0,03| 0.05
25 Hz| -30.| 0,03| 0.06
50 Hz| -28.| 0,04| 0.08
75 Hz| -27.| 0,05| 0.09
100 Hz| -27.| 0,05| 0.09
1 kHz| -26.| 0,05| 0.10
10 kHz| -26.| 0,05| 0.10
100 kHz| -26.| 0,05| 0.10
1 MHz| -26.| 0,05| 0.10
10 MHz| -26.| 0,05| 0.10
25 MHz| -26.| 0,05| 0.10
50 MHz| -26.| 0,05| 0.10
60 MHz| -25.| 0,05| 0.11
70 MHz| -25.| 0,06| 0.12
80 MHz| -24.| 0,06| 0.12
90 MHz| -23.| 0,07| 0.14
100 MHz| -23.| 0,07| 0.14
110 MHz| -22.| 0,08| 0.16
120 MHz| -21.| 0,09| 0.18
130 MHz| -20.| 0,10| 0.20
140 MHz| -19.| 0,12| 0.23
150 MHz| -17.| 0,14| 0.27
160 MHz| -16.| 0,16| 0.32
170 MHz| –14.40| 0,19| 0.38
180 MHz| -13.| 0,23| 0.46
190 MHz| -11.| 0,29| 0.57
200 MHz| -9.| 0,36| 0.73

Table 1: Transfer impedance: 1 Hz to 200 MHz, typical data

Accessory
Tekbox supplies a calibrator suitable for the TBPCP2-3070 current probe:

Ordering Information

Part Number Description
TBPCP2-3070 Pulse Current Probe 30Hz – 70 MHz
TBPCP2-CAL Calibration fixture for TBPCP2-3070 current probe

History

Version Date Author Changes
V 1.0 9.11.2022 Mayerhofer Creation of the preliminary document

www.tekbox.com
TekBox Digital Solutions Vietnam Pte. Ltd.
Factory 4, F5, Lot I-3B-1, Saigon Hi-Tech Park,
Tan Phu Ward, District 9, Ho Chi Minh City, Vietnam