Liquid Instruments Moku:Pro Laser Lock Box User Manual
- June 9, 2024
- Liquid instruments
Table of Contents
Moku:Pro Laser Lock Box
User Manual
Moku:Pro Laser Lock Box
The Moku:Pro Laser Lock Box enables you to stabilize a laser’s frequency to a
reference cavity or atomic transition using high-performance modulation
locking techniques. The Laser Lock Box is designed with the concept of staged
locking and a “Lock Assist”, enabling a user-defined locking process to
quickly lock to any zero-crossing on the demodulated error signal. It also
features an integrated four-channel Oscilloscope, allowing you to observe
signals at any point in the signal processing chain at up to 1.25 GSa/s.
Additionally, the built-in Data Logger enables long-term recording of signals.
Ensure Moku:Pro is fully updated. For the latest information, visit:
liquidinstruments.com
Introduction
Laser locking systems are widely used to control and match a laser’s frequency
to an optical frequency reference, such as an optical reference cavity or
atomic transition. Such systems are vital for high-resolution interference
measurement, spectroscopy, and time and frequency standards.
Locking a laser by forcing the laser and reference frequency to be equal
allows for two scenarios:
- The locking system steers the laser frequency to be equal to the reference frequency, which is referred to as frequency stabilization; and
- The locking system forces the reference frequency to follow the laser frequency, which is referred to as frequency tracking.
Whether used for frequency stabilization or frequency tracking, the Moku:Pro Laser Lock Box is designed to assist in high-performance, high-gain laser locking systems. It offers advanced setup, acquisition, and diagnostic features that makes it easier and quicker to set up and characterize laser locking systems.
User Interface
The main user interface is divided into upper and lower screen sections. The upper user interface displays the processing chain and principal controls of the Laser Lock Box.
The lower half is readily set to display one of three parameter control panels: locking stages, Oscilloscope, and Data Logger.
ID | Description |
---|---|
1 | Locking stages |
2 | Oscilloscope |
3 | Data Logger |
Main Menu
The main menu can be accessed by pressing the icon, allowing you to:
Preferences
The preferences pane can be accessed via the main menu. In here, you can reassign the color representations for each channel, connect to Dropbox, and more. Throughout this manual, the default colors (shown in the figure below) are used to present instrument features.
ID | Description |
---|---|
1 | Tap to change the color associated with input channels. |
2 | Tap to change the color associated with output channels. |
3 | Tap to change the color associated with maths channel. |
4 | Indicate touch points on the screen with circles. This can be useful for |
demonstrations.
5| Change the currently linked Dropbox account to which data can be uploaded.
6| Notify when a new version of the app is available.
7| Moku:Pro automatically saves instrument settings when exiting the app, and
restores them again at launch. When disabled, all settings will be reset to
defaults on launch.
8| Moku:Pro can remember the last used instrument and automatically reconnect
to it at launch. When disabled, you will need to manually connect every time.
9| Reset all instruments to their default state.
10| Save and apply settings.
Upper User Interface
ID | Description |
---|---|
1 | Main menu |
2a-c | Tap to drop oscilloscope probe points to examine signals along the |
processing chain
3| Tap to configure the digital filter
4a| Tap to configure fast PID controller
4b| Tap to configure slow PID controller
5a| Connect fast PID chain to output
5b| Connect slow PID chain to output
6a| Apply output offset to fast PID chain
6b| Apply output offset to slow PID chain
7a| Output limiter on fast PID chain
7b| Output limiter on slow PID chain
8| Configure demodulation modes
9a| Turn output 1 on/off
9b| Turn output 2 on/off
9c| Configure output 3 gain
10| Configure local oscillator demodulation
11| Configure modulation frequency, amplitude and output channel
12| Configure scan frequency, shape, amplitude and output channel
Signal input
Tap the icon to configure the input settings for the signal input. Similar
configurations can be made on input 2.
Scan configurations
Tap the icon to configure the settings for the scan signal.
Modulation configuration
Tap the icon to configure the settings for the modulation signal.
Digital filter configuration
Immediately after the demodulation function, there is a digital filter
designed to remove unwanted signal components. This is highly configurable;
just tap the lowpass filter icon.
ID | Description |
---|---|
1 | Configure lowpass filter corner frequency: tap to enter frequency or touch |
and drag
2| Toggle between magnitude or phase plots
3| Tap to select filter shape
4| Fixed sample rate of 78.125 MHz
5| Tap to select filter type
6| Tap to select filter order
7| Toggle view of poles/zeroes of filter
8| Tap to enter corner frequency
Filter shapes
The shape of the filter can be selected by tapping the filter icon. There are
two pre-defined filter shapes and a fully customizable filter option.
PID controllers
The Moku:Pro Laser Lock Box implements two cascaded PID controllers: a fast
controller and a slow controller. The input of the slow PID controller is the
output of the fast PID controller.
Both the fast and slow PID controllers can be configured graphically by
dragging interactively on the magnitude chart or by tapping on cross-over tabs
and entering frequency or gain on the soft keypad.
The PID controller provides full control over proportional, integral, and
derivative gain profiles with saturation levels available for the integral and
derivative components. The PID’s transfer function is updated in real time.
ID | Description |
---|---|
1 | Drag or tap to enter integrator crossover frequency |
2 | Drag or tap to enter differentiation crossover frequency |
3 | Toggle between magnitude or phase plots |
4 | Drag or tap to enter integrator saturation |
5 | Tap to set differentiation saturation |
6 | Close the PID controller interface |
7 | Proportional, integrator, differentiation, double integrator (only on the |
fast controller), integrator saturation, and differentiation saturation
settings
8| Drag or tap to enter proportional gain
9| Swipe to adjust selected PID parameter
Local oscillator
The demodulation signal source can be configured in the settings dialog.
Demodulation
The demodulation mode determines which reference oscillator is used to
demodulate the input signal.
Modulation
The input signal can be demodulated with the modulation signal, which is user-
configured here. This local oscillator is locked to the modulation signal with
a frequency range from 1 mHz to 300 MHz.
Internal
The input signal can be demodulated with an internally generated reference
signal. This local oscillator is derived from the Moku:Pro internal clock and
thus shares the same time base. The frequency range of the internal reference
is 1 mHz to 600 MHz.
External
The input signal can be demodulated by a direct external reference, permitting
the use of non sinusoidal demodulation using the input signal applied on input
2.
External (PLL)
External (PLL) mode enables the Laser Lock Box to lock to an externally
sourced demodulation reference applied to input 2. This mode uses a digitally
implemented phase-locked loop (PLL) to track the phase of the external
reference with a user configurable bandwidth. To configure the bandwidth of
the PLL, tap the PLL icon to select the bandwidth between 1 Hz to 1 MHz. The
PLL will automatically lock to the strongest harmonic of the external
reference in the range of 10 Hz to 600 MHz with a manually configurable local
phase shift. The PLL can be manually set as low at 10 Hz. The PLL can be
frequency multiplied up to 250x or divided down to 1/8x with the Multiplier
for use as local oscillator, minimum Multiplier step size is 1/8 th. The
reacquire button can be used to re lock to the external reference.
None
The demodulation step can be bypassed by selecting “None.” This enables
modulation-free locking techniques such as DC locking, fringe-side locking,
and tilt locking.
Lower User Interface
Use the lower user interface to either configure the locking stages or display
the half screen Oscilloscope or Data Logger.
Locking stages
ID | Description |
---|---|
1 | Lock stage 1 of (up to) 3 |
2 | Remove a certain action in the locking stage |
3 | Add a new locking stage (allows up to 3 stages at maximum) |
4 | Remove a certain locking stage |
5 | Engage the actions defined in the locking stage |
6 | Add a new action in the locking stage |
7 | Disengage the actions defined in the locking stage |
8 | Tap to connect the slow PID to Output 1to enable the slow feedback |
9 | Tap to connect the fast PID to Output 2 to enable the fast feedback |
10 | Tap to enable/disable the scan signal |
11 | Tap to switch to the Oscilloscope to view the error signal, where the user |
can select the locking point
Add action
Choose any of the available actions to add to each locking stage. When
engaged, all of the actions associated with that stage will be enabled
simultaneously.
Oscilloscope
The Moku:Pro Laser Lock Box includes a built-in Oscilloscope, enabling you to
observe and record data of up to four signals at a time in the Laser Lock
Box’s processing chain.
ID | Description |
---|---|
1 | Close Oscilloscope panel |
2 | Select Lock Assist |
3 | Share Oscilloscope data |
4 | Reveal/hide settings sidebar |
5 | Tap to select channel tab |
6 | Tap to select time base tab |
7 | Tap to select trigger tab |
8 | Tap to select measurements tab |
9 | Tap to set Oscilloscope to full screen |
10 | Pause/run Oscilloscope |
11 | Tap to add time/voltage cursors, or drag right or drag up to create a |
cursor
The Oscilloscope will appear automatically when a probe point is activated.
You can hide the Oscilloscope by pressing the icon and reveal it by pressing
the icon.
Lock Assist feature
The Lock Assist can only be enabled after the locking stage is configured.
When the Lock Assist is enabled, the trigger settings will be configured to
synchronize the oscilloscope traces with the scan waveform. Selecting one of
the error signal’s zero crossings will then adjust the output offset to the
corresponding voltage and engage the first locking stage which is not already
fully engaged.
Additional details about the Oscilloscope can be found in the Moku:Pro
Oscilloscope manual.
Data Logger
The built-in Data Logger allows you to acquire data from up to four probe
points at a time at a maximum sampling rate of 2.5 MSa/s with three or four
channels, 5 MSa/s with two channels and 10 MSa/s with one channel. To access
the data acquisition menu, press the icon. More details about Data Logger are
available in the Moku:Go Data Logger manual.
ID | Description |
---|---|
1 | Select the sampling rate at which your measurement is recorded |
2 | Upload saved data |
3 | Select between Normal, Precision, or Peak Detection acquisition modes |
4 | Change filename and comments for the logged file |
5 | Record a new measurement |
6 | Configure measurement duration |
7 | Configure when to begin recording data |
The embedded Data Logger can stream over a network or save data on Moku:Pro.
For details, refer to the Data Logger user manual. More streaming information
is in our API documents at
apis.liquidinstruments.com.
Ensure Moku:Pro is fully updated. For the latest information, visit:
liquidinstruments.com
© 2023 Liquid Instruments. All rights reserved.
Documents / Resources
|
Liquid Instruments Moku:Pro Laser Lock
Box
[pdf] User Manual
Moku Pro Laser Lock Box, Moku Pro, Laser Lock Box, Lock Box, Box
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>