NEURO TRIGGER NTB023 Powered Muscle Stimulator Instructions
- June 9, 2024
- NEURO TRIGGER
Table of Contents
NeuroTrigger Basic
Powere d Muscle Stimulator
Rx only
Instructions
Safety Information
Validity
The information and technical data contained in this document relates to the
NeuroTrigger Basic power muscle stimulator provided with this manual. Each
NeuroTrigger unit is attributed a serial number which is located on the back
of the unit. The information and technical data disclosed i n this document
are proprietary to NeuroTrigger Ltd. and may only be used and disseminated for
the purposes and to the extent specifically authorized in writing by the
company.
Disclaimers
All items of equipment manufactured and sold by NeuroTrigger Ltd are
rigorously checked and tested prior to shipment. However the use of these
units is beyond the area of the company’s control. NeuroTrigger Ltd only
accepts responsibility for the safety, reliability and performance of the
equipment when it is operated in accordance with the instructions herein and
within the given specifications.
Therefore, the user must bear full responsibility for any actions arising out
of the use or misuse of this equipment. Any modifications, repairs or
servicing must be undertaken by authorized NeuroTrigger Ltd personnel. This
manual is intended for the guidance of the administrator, patient, and
clinician. Patients should not set up their own stimulator and must visit
their physician for setup. This device conforms with IP67 Standard.
Intended Use
NeuroTrigger Basic (NTB) is a powered muscle stimulator intended for medical
purposes that repeatedly contracts muscles by passing electrical currents
through electrodes contacting the affected body area.
Indication for use
NeuroTrigger Basic is intended to stimulate healthy muscles in order to
improve or facilitate muscle performance for maintaining/increasing muscle
range of motion and for the prevention or retardation of disuse atrophy.
WARNINGS, CAUTIONS, AND NOTES The terms warning, caution, and note have
specific meanings in this manual.
WARNING advises against certain actions or situations that could result
in personal injury or death.
CAUTION advises against actions or situations that could damage
equipment, produce inaccurate data, or invalidate a procedure, although
personal injury is unlikely.
NOTE provides useful information regarding a function or procedure.
CAUTION
-
Read, understand, and practice the precautionary and operating instructions. Know the limitations and hazards associated with using any electrical stimulation. Observe the precautionary and operational decals placed on the unit.
-
DO NOT operate this unit when connected to any accessories other than NeuroTrigger Y-shaped Electrodes specifically described in user or service manuals. There are two types o f electrodes that differ in shape: wide & narrow versions.
-
DO NOT use sharp objects such as a pencil point or ballpoint pen to operate the buttons on both the NeuroTrigger Cradle and Stimulator.
-
DO NOT disassemble, modify, or remodel the NeuroTrigger Cradle and Stimulator. This may cause unit damage, malfunction, electrical shock, or personal injury.
-
DO NOT permit foreign materials, liquids or cleaning agents to enter the unit, including, but not limited to, inflammables, water, and metallic objects, to prevent unit damage,
malfunction, electrical shock, fire, or personal injury. -
DO NOT operate the NeuroTrigger Basic within the vicinity or environment of any therapeutic microwave or RF shortwave diathermy system in operation.
-
Device is designed to comply with electromagnetic safety standards. However, there i s no guarantee that interference will not occur i n a particular installation. Harmful interference to other devices can be determined by turning this equipment on and off.
-
Inspect NeuroTrigger Electrodes, cables and associated connectors before each use.
-
No distance is required between the equipment and all persons given that the device comes i n contact with patient skin.
-
This unit should be operated at 5°C to 40°C and 20% to 90% Relative Humidity.
-
The unit should be transported and stored at -10°C to 50°C and 20% to 90% Relative Humidity.
-
Allow the patient to assume a comfortable position during NeuroTrigger therapy session.
-
Failure to use and maintain the NeuroTrigger Basic and its accessories in accordance with the instructions outlined i n this manual will invalidate the warranty.
-
If you have difficulty operating the unit after carefully reviewing this user manual, contact NeuroTrigger Ltd or authorized distributor for assistance.
-
DO not use any parts or materials other than those supplied by NeuroTrigger.
-
Safe use of electrotherapy during pregnancy has not been established.
-
Caution should be used for patients with suspected or diagnosed heart problems. Application of electrode near the thorax may increase the risk of cardiac fibrillation.
-
Caution should be used for patients with suspected or diagnosed epilepsy.
-
Caution should be used in the presence of a tendency to hemorrhage following acute trauma or fracture, following recent surgical procedures when muscle contraction may disrupt the healing process and over areas o f skin which lack normal sensation.
-
Some patients may experience skin irritation or hypersensitivity due to electrical stimulation or electrical conductive medium. The irritation can usually be reduced by moistening the skin following the stimulation
-
Inspect NeuroTrigger Electrode and associated connectors for signs of damage or tear before use. Replace damaged electrode immediately with another new before any treatment is applied.
-
Electrode placement and stimulation settings should be based on the guidance of the prescribing practitioner or other licensed health professional.
-
Portable powered muscle stimulators should not be used while driving, operating machinery, or during any activity i n which involuntary muscle contractions may put the user at undue risk of injury.
-
Always check the stimulation controls before treating a patient. The stimulation amplitude/intensity should always be adjusted gradually.
-
All of the materials exposed directly to skin are Biocompatible.
WARNING
-
U.S.A. Federal Law restricts these devices to sale by, or on the order of, a physician or licensed practitioner. This device should be used only under the continued supervision of a physician or licensed practitioner.
-
Be sure to read all instructions for operation before treating patient.
-
The stimulation should not be applied across or through the head, directly on the eyes, covering the mouth, on the front of the neck, (especially the carotid sinus), or from electrodes placed on the chest and the upper or crossing over the heart.
-
The device i s not designed to be attached in a sleeping position.
-
Patients with an implanted electronic device (for example a cardiac pacemaker) should not be subjected to stimulation unless specialist medical opinion has first been obtained.
-
Care must be taken when operating this equipment around other equipment. Potential electromagnetic or other interference could occur to this or to the other equipment. Try to minimize this interference by not using other equipment in conjunction with it.
-
Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when electrical stimulation is i n use.
-
Do not drop the unit on hard surfaces or submerge i n water. These actions will damage the unit. Damage resulting from these conditions is not covered under the warranty.
-
This device should be kept out of the reach of children.
-
Use only cables and accessories that are specially designed for the NeuroTrigger unit. Do not use accessories manufactured by other companies on the NeuroTrigger unit. NeuroTrigger is not responsible for any consequence resulting from using products manufactured by other companies. The use of other accessories or cables may result in increased emissions or decreased immunity of the NeuroTrigger Basic unit.
-
Contaminated sticker electrode, and gel can lead to infection.
-
DO NOT use electrode if the gel circles shifted, moved, or
detached from the electrode. -
DO NOT use electrode if the underlaying silver layer is exposed.
-
Use of electrode with degraded hydrogel can result in burn to the skin.
-
DO NOT operate this unit in an environment where other devices are being used that intentionally radiate electromagnetic energy in an unshielded manner.
-
Stop treatment immediately if patient experiences discomfort or pain.
-
Long term effects of chronic electrical stimulation are unknown.
-
Stimulation should not be applied trans cerebrally.
-
Stimulation should not be applied over swollen, infected, and inflamed areas or skin eruptions, e.g., phlebitis, thrombophlebitis, varicose veins, etc.
-
Stimulation should not be applied over, or in proximity to, cancerous lesions.
-
Stimulation should not be applied over the carotid sinus particularly in patients with a known sensitivity to the carotid sinus reflex.
-
Using the supplied stimulation electrodes, the current density will not exceed 2mA/cm?. Special caution is to be exercised when adjusting the current level beyond levels set by the clinicians, as the greater current values and small electrode area may possibly cause skin irritation or burns.
-
The NeuroTrigger Basic accessories are designed for use only with the NeuroTrigger Basic.
-
Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
-
Medical electrical equipment needs special precautions regarding EMC. Portable and mobile RF communication equipment can be affected by other medical electrical devices.
-
Common RF emitting devices (e.g., RFID) and electromagnetic security systems (e.g., metal detectors) may produce instability in the NeuroTrigger Basic output.
-
The NeuroTrigger Basic has been tested in the presence of Equipment Under Test monitors, which caused the stimulation to stop.
-
Before administering any treatment to a patient you should become acquainted with the operating procedures for each mode of treatment available, as well as the indications, contraindications, warnings and precautions. Consult other resources for additional information regardin g the application of each mode of treatment.
-
Simultaneous connection of a patient to high frequency surgical Medical Electrical equipment may result i n burns at the site of the stimulator electrodes and possible damage to the stimulator.
-
Keep Y – shaped electrode’s ends separated during treatment.
Electrodes in contact with each other could result in improper stimulation or s kin burns. -
Electrodes should be inspected before each use. ( i.e.hydration level, tack, discoloration and impurities) Follow the manufacturing guidelines on electrode packaging.
-
Stimulator includes a current monitor that checks whether stimulation is delivered correctly to the electrodes; the stimulator also has a self-test mechanism that is used to detect whether the stimulator hardware modules are properly functioning. This feature is available in the application is automatically activated.
-
Extra security measures should be taken to avoid cybersecurity threats such as keeping the android device up to date, and checking for any future updates for the firmware.
-
The password for each user should be kept hidden and changed periodically for confidential measures.
-
Passwords for the application login and NeuroTrigger Stimulator are provided in the kit on a separate paper slip.
-
This device contains lithium batteries, which are classified as Class 9 hazardous material UN3480. Must be handled with care by qualified personnel. Do not replace battery.
DANGER
- Use only electrodes and accessories designed specifically for use with the NeuroTrigger Basic. Use of other accessories and/or techniques not approved under the NeuroTrigger certification training may result in death, injury, or adverse effects to patient or undesirable and ineffective.
DESCRIPTION OF DEVICE MARKINGS:
The marking on the unit are assurance of its conformity to the highest applicable standards of medical equipment safety and electromagnetic compatibility. One or more of the following markings may appear on the device.
Symbol | Title |
---|---|
**** | Serial Number |
Caution: consult Accompanying Documents | |
**** | DO NOT THROW IN TRASH. |
Separate treatment from general waste at end of life
| Lithium battery
| Certification the electromagnetic interference from the device is under
limits approved by the Federal Communications Commission
| Manufacturer
| Date of Manufacture
---|---
| Use-by Date
| Authorized Representative in the European Community
| Batch Code
| Refer to Instruction Manual/Booklet
| Quantity
| Neuromuscular Stimulation (STIM) and sEMG. Stimulation should not be used by
Patients fitted with demand style cardiac pacemakers
| Keep dry
| Type BF applied part. Have a conductive contact with the patient
PRODUCT DESCRIPTION
The NeuroTrigger Basic is a powered muscle stimulator system consisting of a
stimulator, disposable electrodes, a cradle, a stimulator holder, application,
and charging accessories. The stimulator transfers electrical stimulations
through pre-gelled Y-shaped electrodes.
NOTE: This equipment is to be used only under the prescription and
supervision of a licensed medical practitioner.
Components
The NeuroTrigger Basi c powered muscle stimulator system is composed of the
following components as shown below.
Stimulator
Narrow Y- Shaped hydrogel electrode wide (Disposable, Single Use)
Step By Step Treatment Guide
NeuroTrigger Application Android Installation
****https://2hac.page.link/6M7g
- Scan the QR code.
- Copy the link to your browser.
- Download the APK file from the link.
- Install the application on your device.
- NeuroTrigger Application will request to access some permissions for the app, such a s Bluetooth, device location, photos and media, and network & WiFi.
NeuroTrigger Electrode Instruction for Use
- Using a mild soap and water, clean the skin thoroughly where you will be placing the electrodes, and wait until the area i s perfectly dry. The electrodes do not adhere well if any dirt, oils, creams or other cosmetics are still on the skin.
- Take one electrode pouch from the NeuroTrigger Electrode box. Be careful not to bend the pouch.
- Open the pouch from the tear notch.
- Carefully extract the NeuroTrigger Electrode sticker from the pouch, making sure not to bend it.
- Remove the liners from the adhesive side and position the sticker on the skin, with the Yheads directly over the muscle for stimulation. The electrodes should be placed as directed by your physician. The electrode’s patch connector should be free.
- For setting up stimulation, follow the steps outlined in NeuroTrigger Stimulator Instruction for Use.
Please note the following points:
- Aclinician must provide instructions on exact electrode placement.
- The safety information provided in this manual must be followed.
- The complete surface of the sticker electrode should be in contact with the skin.
NeuroTrigger Stimulator Instruction for Use
-
Remove the NeuroTrigger Charging Cradle from its b o x and unwarp its cover.
-
Open the cradle and take out the NeuroTrigger Stimulator.
-
Turn on the stimulator by pressing the ON/OFF button (circled in red) as shown in the image to the right until a green light flashes near the electrode connection side of the stimulator (see green circle).
-
Make sure that your smartphone has Bluetooth turned on and pair the stimulator to the NeuroTrigger App via Bluetooth connection, by choosing the serial number of the device. The device will flash blue for one minute once connection is obtained.
-
Depending on the required area of application, a holder may be used to attach the stimulator to the body. The stimulator can be clipped into the holder and held in place, and the holder can be attached to the body using pre-cut DST tape supplied with the NeuroTrigger Basic package.
-
Clasp the stimulator into the holder and ensure it is sitting securely. Remove the DST sticker cover on both side of the holder, and attach one side of the DST tape to the holder and the other side to the desired region.
-
Please note that the stimulator must align with the electrode that is already attached to clean skin.
-
Align a good connection between both magnet poles of the NeuroTrigger stimulator and electrode sticker. All the sticker’s magnetic pins must be attached to their pairs on the NeuroTrigger Stimulator with no metal pin visible.
-
The stimulator can be used up to 30 hours. Check for a flashing red light as a sign of low battery, or the application battery indicator.
For recharging the stimulator: -
Place the stimulator back inside the charging cradle with the stimulator button facing the same side of the USB C connector (shown in a circle).
-
Attach the electrical adapter to the electrical outlet – may use the European converter if necessary. Plug the USB side of the charging cable to the charger, and the USB C side to the cradle.
-
Plug the charger into the electrical socket, and check the cradle light for charging signal, shown in the box.
a. An orange light indicates that there is no cradle-stimulator coupling/ stimulatoris not in cradle/improper coupling of the two components
b . Ared light indicates that the stimulatoris its correct position and is charging.
c. Agreen light indicates that the stimulator battery is fully charged.
NeuroTrigger Application Instruction for Use
After downloading the application, a password is required to enter the
application, as well as when pairing the stimulator to the application. Both
4-digit passwords are provided in the kit. Passwords can be changed after the
first log in to custom 4-digit password, or if permissible and supported by
the mobile application, biometric access fingerprints and face recognition.
The application supports three types of users; patient, clinician, and
technician. Please follow the instruction for your user type. The application
will save the last user login and automatically pair to the last used
stimulator when launched, and can be activated through the setting tab.
Patient
Patients are capable o f connecting the NeuroTrigger device and controlling
functionality, which needs BLE pairing, choose your device from the available
device list. You will need to provide a pairing password
The patient is then directed to the operation dashboard as shown on the next
page (left).
-
Device status: Battery Level, BLE Connection Quality, Device Functionality, Electrode Quality, Running Scene.
-
Stimulation control can be done using the Turn On/off button.
Turn ON will start the simulation, the system allows user to control parameters (Pace 0-9 sec & Current Strength 0-9mA). -
The running scene is the stimulation parameters pre-defined by the clinician (custom scenes) or built-in on the app (fixed scenes). The clinician has four fixed pre-set programs (Indoors, Mild, Office, and Outdoors) that can be renamed & two custom programs, which can be shared by an authorized person ( see image on the top right). Device will run the default scene once it starts.
-
The default scene can be chosen by clicking on the green star icon on the left of each scene.
-
Tochoose a different running scene double click the desired scene.
The clinician can share a file through email or WhatsApp. Download the file to
your documents, for devices with Android 10 software and above, move the
shared file from downloads to Internal Storage — Android — Media. Not
applicable for lower than 10 Android devices.
Warning: Please do not add info personal information to the database. It
is at the user’s discretion to add their personal information such as: first
and last name, and patient ID.
Administrator
The administrator i s the eligible technician, responsible for entering
system parameters, exert system data, and perform remote maintenance.
The administrator account can navigate between 1) the programs and 2) settings
screens. There will be 3 tabs placed on screen bottom.
The patient tab indicates that the administrator will have access to a list of
all their patients, as well as:
- Creating/ adding new patient profiles.
- Editing patient details by doubled-clicking the patient’s name.
- Searching for patients by their name or ID #.
- Deleting patients.
The program tab will allow the administrator to control the general settings, saved Scene #1, and saved scene #2
- General settings allow you to search for the patient i n the search bar (shown at the top of the page in the image on the previous page).
- Saved scene #1, and saved scene #2, which are customized scenes that the user can insert the desire configuration parameters, such as stimulation pace, strength, polarity, pulse width, and burst frequency. Also, these parameters can be shared by sending encrypted file to patient’s smartphone via standard sharing application (WhatsApp, email, Telegram, etc.).
Clinician
The clinicians are the medical professionals instructed and trained by
NeuroTrigger to modify and optimize the settings of the stimulation bursts for
each individual patient, and save the stimulation parameters in the
NeuroTrigger Stimulator.
The clinician can carry out the same functions as the administrator, which are
mentioned in the previous section.
Maintenance & Troubleshooting
Maintenance
The unit should be cleaned regularly using a soft cloth, lightly dampened with
soapy water. Do not allow the interior of the NeuroTrigger Stimulator or
Cradle to get wet during cleaning.
The NeuroTrigger electrode i s a single-use electrode, and can be disposed
after use. If the electrodes are dirty or no longer adhere properly, they need
to be replaced. Replace the sticker if the connecting base is damaged or the
sticker liner is cut.
Repair, Service & Modification
Access to the interior i s not required for maintenance purposes. Repair,
service and modifications may not be carried out by anyone other than
qualified service personnel authorized by NeuroTrigger Ltd. Do not use the
unit if it is defective. Please return it to NeuroTrigger. NeuroTrigger Ltd
will not accept any responsibility where the guidelines and instructions are
not followed.
Troubleshooting
Problem | Possible Cause | Solution |
---|---|---|
BLE stimulator signal is not detected by application | Battery depleted or | |
other unknown device have connected to stimulator | Check battery indicator, |
and recharge battery according to instructions in this manual.
Restart stimulator, pairing the device in close proximity to the stimulator
The unit is switched on but does not respond to commands| Stimulator device
malfunction or crash, Cybersecurity threat| Restart device and recheck if
problem still occurs.
Restart stimulator, pairing the device in close proximity to the stimulator
Stimulation stops immediately| Stimulator is sensing high impedance;| Try to
press firmly on electrodes to attach
thoroughly on skin.
| Electrodes are not set correctly on skin| Detach electrodes, clean skin and
reattach the electrodes. Replace electrodes.
---|---|---
Increasing intensity causes unpleasant
sensation.| Electrodes are not attached fully on skin Stimulator malfunction|
Try to press firmly on electrodes to ttach thoroughly on skin. Detach
electrodes, clean skin and reattach the electrodes. Replace electrodes.
Technical Information
General Specifications
Product Type: | Electrical Muscle Neuro Stimulator |
---|---|
No. of Channels: | 1 |
Waveform: | Biphasic Symmetrical current pulse |
Environmental Specifications:
Operation:
Storage:| Temperature 5° to 40°C
Humidity 20 —90 % RH
Temperature -25° to 70°C
Humidity 20 —90 % RH
Physical Specification:| Nonrectangular folded 10 cm X 1cm X
Unit Dimensions:| 8mm length
Weight| 11 grams
Nominal Output Voltage / Power
Parameter| 500 Ω| 2 KΩ| 10 KΩ
Max output voltage| 7.5V| 30V| 90V
Max output current| 15mA| 15mA| 9mA
Output frequency| 10-300Hz| 10-300Hz| 10-300Hz
DC Component (charge)| 0C| 0C| 0C
Pulse Width| 80-400μs| 80-400μs| 80-400μs
Current Intensity Range (Per Pulse)| 0-15mA| 0-15mA| 0-9mA
Burst length| 0.1-0.2sec| 0.1-0.2sec| 0.1-0.2sec
Burst frequency| 2-10sec| 2-10sec| 2-10sec
Safety Features
There are two types of NeuroTrigger Electrodes: narrow and wide span. Both
electrodes provided with the device have the same surface area; the
difference between the narrow and wide i s the distance and divergence angle
between the two Y heads of the electrode. Your assigned caregiver will specify
which electrode to use depending on your application.
The surface area of less than 2 cm? can cause current densities in excess of
2mA/cm? above certain intensities. This might cause unpleasant feelings, pain,
or leave red marks i n some patients; In this case, contact the manufacturer
or your clinician.
Product Information
- Manufacturer: Senso Medical LTD
- Customer: NeuroTrigger LTD.
- Customer’s Address: Nirim 3, Tel Aviv 67060, Israel
- Equipment Model No.: NTB-KT-0003
- For labels explanation look into NT-11 labels file.
- FCCID: 2AS65NEUROTBASIC
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential in stallation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used i n accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on , the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Declaration – Electromagnetic Emissions
Emissions test| Compliance| Electromagnetic environment –
guidance
RF Emissions CISPR 11| Group1 Class A| The NTB System uses RF energy only
for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
Harmonic Emissions IEC 61000-3-2| Class A|
The NTB System is suitable for use in all establishments other than domestic, and may be used in domestic establishments and those directly connected to the public low- voltage power supply network that supplies buildings used for domestic purposes, provided the following warning is heeded: Warning: This equipment/system is intended for use by healthcare professionals only. This equipment/ system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the NTB System or shielding the location.
Voltage Fluctuations And Flicker IEC 61000-3- 3:2013| Complies
IMMUNITY test| IEC 60601 test level| Complian ce level|
Electromagnetic environment – guid ance
---|---|---|---
Electrostatic Discharge (ESD)
IEC 61000-4- 2| 8 kV contact 2, 4,8, 15kV air| 8 kV contact 2, 4, 8,
15kV air| Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at least 30
%.
Electrical Fast Transient/B urst
IEC 61000-4- 4| 2 kV for power supply lines
1 kV for input/outp ut lines| 2 kV for power supply lines N/A| MainS power
quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5| 1 kV line(s) to line(s)
2 kV line(s) to earth
2 kV Signal input/outp ut) to earth| 1 kV line(s) to line(s)
2 kV line(s) to earth N/A| Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips, Short interruption s and voltage variations| 0% UT;
0.5cycle at 0°, 45°,
90°,
135°,180°,| 0% UT;
0.5cycle at 0°,
45°, 90°,
135°,180
°, 225°,| Mains power quality should be that of a typical commercial or
hospital environment. If the user of the NT-11 System requires continued
operation during
on power supply input lines IEC 61000-4- 11| 225°, 270° and 315°
0% UT; 1cycle and 70% UT; 25/30 cycles Single phase at 0° 0% UT; 250/300
cycle| 270° and 315° 0% UT;
1cycle and 70% UT;
25/30 cycles Single phase at 0° 0% UT; 250/300 cycle| power mains
interruptions, it is recommended that the NT- 11 System be powered from an
uninterruptible power supply or a battery.
---|---|---|---
Power Frequency (50/60 Hz)
Magnetic Field
IEC 61000-4-8| 30 (A/m)| 30 (A/m)| Power frequency magnetic fields should be
at levels characteristic of a typical location in a typical commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test
level.|
Conducted RF
IEC 61000-4-6| 3V, 6V| 3Vrms, 6V| Portable and mobile RF communications
equipment should be used no closer to any part of the NT-11 System ,
including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter. Recommended
separation distance
Radiated RF
IEC 61000-4-3| 3V/m 3V from 0.15 to 80MHz; 6V from 0.15 to 80MHz and 80% AM at
1kHz 3V/m from 80MHz
to 2.7GHz| 3V/m 3V from 0.15 to 80MHz; 6V from 0.15 to 80MHz and 80% AM at
1kHz 3V/m from 80MHz
to 2.7GHz| where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
Should be less than the compliance level in each frequency range.
D Interference may occur in the vicinity of equipment marked with the
following symbol
---|---|---|---
Separation Distance Table
Recommended Separation Distances Between Portable And Mobile RF
Communications Equipment And The NT-11
System
Rated maximum output power of transmitte r W| Separation distance
according to frequency of transmitter
m
150 kHz to 80 MHz outside ISM bands| 150 kHz to 80 MHz in
ISM bands| 80 MHz to 800 MHz| 800 MHz to 2,5 GHz
0.01| 0.12| 0.2| 0.4| 1
0.1| 0.37| 0.64| 1.3| 2.6
1| 1.17| 2| 4| 8
10| 3.7| 6.4| 13| 26
100| 11.7| 20| 40| 80
Enclosure Port Immunity Test Specifications
Test Specifications For ENCLOSURE PORT IMMUNITY To RF Wireless Communications Equipment
Test frequency
(MHz)| Band a)
(MHz)| Service a)| Modulation b)| Max power (W)| Distanc e
(m)| IMMUNITY TEST LEVEL (V/m)| Complian ce level (V/m)
385| 380 – 390| TETRA 400| Pulse modulation
b) 18 Hz| 1.8| 0.3| 27| 27
450| 430 – 470| GMRS 460,
FRS 460| FM c) ± 5 kHz deviation 1 kHz sine| 2| 0.3| 28| 28
710| 704 – 787| LTE Band 13, 17| Pulse modulation
b) 217 Hz| 0.2| 0.3| 9| 9
745
780
810| 800 – 960| GSM 800/900, TETRA 800,
iDEN 820,
CDMA 850,| Pulse modulation
b) 18 Hz| 2| 0.3| 28| 28
870
930
| | LTE Band 5| | | | |
---|---|---|---|---|---|---|---
1720| 1 700 – 1 990| GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4,
25; UMTS| Pulse modulation b) 217 Hz| 2| 0.3| 28| 28
1845
1970
2450| 2 400 – 2 570| Bluetooth, WLAN, 802.11 b/g/n,
RFID 2450, LTE Band 7| Pulse modulation
b) 217 Hz| 2| 0.3| 28| 28
5240
5500
5785| 5 100 – 5 800| WLAN 802.11 a/n| Pulse modulation
b)217 Hz| 0.2| 0.3| 9| 9