1216486PA Back Biowrap User Guide

June 9, 2024
BIOWRAP

1216486PA Back Biowrap

Compression Electrode Garment

Utilizes Advanced Silver Fabric Electrode Technology

NOTE: ONLY for use with BioWaveHOME & BioWavePRO

Questions? Call us!

This Package Contains

  • One SILVER Back BioWrap Compression Electrode Garment
  • One 4 oz. Bottle of BioWrap Conductive Cream

Inside View

1216486PA-Back-Biowrap-FIG-2

Outside View

1216486PA-Back-Biowrap-FIG-3

Quick Set Up and Use

See video at biowave.com/support/biowrap

  1. The 4” Snap Cables should already be snapped onto each of the 2 snaps on the outside of the BioWrap.

  2. Plug the blue connectors at the other end of each BioWrap cable into your BioWave Leadwire Cable.

  3. Clean your skin with soap and water using a washcloth in the locations that the silver fabric will touch your skin.

  4. Squeeze a pea size amount of BioWrap Conductive Cream onto your finger tip and rub into your skin thoroughly in the locations that the silver fabric will touch your skin.

  5. See diagrams on opposite side to put on the BioWrap and velcro it closed firmly so there is strong but comfortable compression over the silver fabric locations.

  6. Plug the single end of the leadwire cable into the BioWave device.

  7. Press the power button to turn on the BioWave device.

  8. Start the treatment by pressing the PLUS (+) button to increase intensity so the sensation felt is strong but still comfortable. As your body adapts to the electrical field and the sensation fades, keep pressing the PLUS
    (+) button to increase the intensity to maintain a strong but comfortable sensation at the treatment location.

  9. When the 30-minute treatment is over, disconnect the BioWave Leadwire cable from the 4” BioWrap cables, remove the BioWrap from your body and store in a cool dry place.

Directions for Use

1216486PA-Back-Biowrap-FIG-4

  1. Squeeze a pea size amount of BioWrap CONDUCTIVE CREAM onto your finger tip and rub into your skin thoroughly in the two locations that the silver fabric will touch.
  2. Both 4” Sanp Cables should already be snapped onto the 2 snaps on the back side of the BioWrap. Attach your BioWave LEADWIRE CABLE to the two connectors on the back of the BioWrap Line up SILVER FABRIC RECTANGLES on inside of BioWrap over painful areas on your low back.
  3. Wrap LEFT SIDE around stomach. Wrap RIGHT SIDE around stomach, pull tight, and attach Velcro closure over left side.1216486PA-Back-Biowrap-FIG-5
  4. Unhook SECONDARY STRAPS on front of BioWrap.
  5. PULL FORWARD and VELCRO DOWN on front of BioWrap to further tighten BioWrap over your low back. BioWrap should provide TIGHT BUT COMFORTABLE compression across your lower back.

TROUBLESHOOTING

My BioWave shows a connection error but my BioWrap is plugged in If your BioWave device shows a connection error when you are fully connected to the BioWrap, remove the BioWrap and follow these steps:

If you forgot to apply BioWave Conductive Cream:

  • CLEAN YOUR SKIN. Remove BioWrap and clean your skin with soap and water in the region that the silver fabric electrodes touch your skin. Shave any excessive hair in the locations that the silver fabric electrodes touch your skin.
  • APPLY A PEA-SIZE AMOUNT OF BIOWAVE CONDUCTIVE CREAM directly to skin areas that will be against the silver fabric inside the BioWrap. Gently massage BioWave Conductive Cream into the skin allowing it to absorb, before putting on the BioWrap.

If you applied too much BioWave Conductive Cream (more than a pea-size amount):

  • CLEAN YOUR SKIN. Remove BioWrap and use a dry towel to remove all excess Conductive Cream from your skin and any excess cream from the silver fabric in the BioWrap. Next, reattach the BioWrap and begin the treatment.

The treatment doesn’t feel strong enough

  • If the intensity is turned up to the maximum setting and the treatment doesn’t feel strong under the electrodes, pause the treatment, remove the BioWrap and apply more conductive cream to the skin in the areas the silver fabric touches the skin.
  • Tighter Compression and/or More Conductive Cream = Stronger Treatment Sensation.
  • Looser Compression and/or Less Conductive Cream = Less Treatment Sensation.

Questions? Call us!

IMPORTANT

Contraindications When Not to Use BioWrap

  • DO NOT use if you have an implanted cardiac pacemaker.
  • DO NOT place the silver fabric electrodes in the BioWrap over open wounds, broken skin, rashy skin or sensitive skin areas like sunburned skin.

NOTE:

  • BioWrap Wearables MAY BE PLACED over or in the proximity of implanted orthopedic hardware, including total joint replacements, anchors, plates, rods, screws, pins and clips.
  • Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.

Cleaning

  • BioWraps may be HAND WASHED using mild soap. Disconnect the two 4” BioWrap cables from their snaps before hand washing the BioWrap.
  • AIR DRY.
  • DO NOT wash in a washing machine.
  • DO NOT place in a dryer.

Need help?

The information, including but not limited to text, graphics, images and other material contained herein are for informational purposes only. Learn more at BioWave.com/terms-of-use

Agenoria Medical Ltd. Blair House Upper O’Connell St. Ennis, Co Clare Ireland BioWave Corporation 8 Knight St., Suite 201 Norwalk, CT 06851 © 2022 BioWave Corporation Rev 3 – 221021

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

BIOWRAP User Manuals

Related Manuals