KULZER Venus Supra Polisher Flame Instruction Manual
- June 8, 2024
- KULZER
Table of Contents
KULZER Venus Supra Polisher Flame Instruction Manual
SYMBOLS
All used symbols and pictograms according EN ISO 15223.
WARNINGS
- Observe the manufacturer information on material compatibilities for cleaning desenfection and sterilisation/
- All instrument are delivered unsterile and must go through the indicated cycle before and after each use.
- Strong acids and strong bases may oxidise the stainless steel shank.
- Avoid temperatures > 150°C.
- Instruments that have not completely dried after cleaning and disinfection must be dried again (e.g. with medical compressed air)
- Instructions of cleaning and/or disinfecting solutions must specifically state “suitable for rubber polishers or synthetics/silicones”. The exposure time and concentration specified by the manufacturer must be followed.
RESTRICTION OF REPROCESSING:
Repetitive reprocessing can slightly change both the look and feel of the
product, but does not interfere with the instrument’s function.
RISK ASSESSMENT AND CLASSIFICATION OF MEDICAL DEVICES BEFORE
REPROCESSING:
The type and scope of reprocessing is determined by the use of the medical
device. Therefore, the operator is responsible for the correct classification
of the medical devices and thus for the definition of the type and scope of
reprocessing (see KRINKO/BfArM recommendation, point 1.2.1 Risk assessment and
classification of medical devices prior to reprocessing). On the basis of this
user-dependent classification, the operator can determine which of the
reprocessing methods listed in this preparation and reprocessing instruction
needs to be applied
PLACE OF USE:
No special requirements
STORAGE AND TRANSPORT:
It is recommended to transport the contaminated instruments in a closed
container. It is recommended that instruments be reprocessed as soon as
possible, within 2 hours after use at the most. Intermediate storage of used
instruments with contamination such as blood residues can lead to corrosion
damage.
PREPARATION:
Wear personal protective equipment (durable gloves, water-repellent coat, face
protection mask or goggles and protection mask).
PRE-TREATMENT:
Pre-clean under running water with a brush (plastic) directly after use.
Equipment: Plastic brush (e.g. lnterlock, #09084), tap water (20± 2 °C)
(at least drinking water quality) 1.access areas of the head (bristles,
silicone bristle tips).
CLEANING: MANUAL
Note: Coarse surface contamination on the instruments must be removed before manual reprocessing (see pre-treatment) Equipment: Multi-stage enzymatic cleaner (e.g. Dürr Dental, ID 215), tap water/flowing water (20± 2 °C) (at least drinking water quality), ultrasonic bath (e.g. Sonorex Digital 10P).
- Prepare the cleaning solution according to the manufacturer‘s instructions (Dürr Dental ID 215 2% solution was validated) and fill into an ultrasonic bath.
- Completely immerse the polishers in the solution.
- Expose the products for 1 minute to the ultrasonic bath.
- Remove the polishers from the cleaning solution and rinse them each thoroughly (30 seconds) under running water.
- Check for cleanliness. If contamination is still visible, repeat the above specified steps.
DISINFECTION:
MANUAL (with subsequent sterilisation)
Equipment: At least limited virucidal instrument disinfectant (VAH listed
– or at least listed in the IHO with testing according to DVV)
e.g. based on quaternary ammonium compound(s), alkylamine(s)/alkylamine
derivative(s), guanidine(s)/guanidine derivative(s) (e.g. Dürr Dental, ID
212), preferably fully deionised water (deionised water, according to
KRINKO/BfArM recommendation free of facultatively pathogenic microorganisms),
ultrasonic bath (e.g. Sonorex Digital 10P), lint-free sterile cloth.
- Place into an ultrasonic bath.
- Completely immerse the polishers in the disinfectant solution.
- Expose the products for 2 minutes to the ultrasonic bath.
- Further exposure time to the disinfectant solution for 5 minutes according to the desinfectant manufacturer’s instructions.
- Rinse the products with deionised water for 30 seconds.
- Wipe with a single use sterile lint-free cloth or, if necessary, dry with medical compressed air.
CLEANING AND DISINFECTION: AUTOMATIC
Note: Coarse surface contamination on the instruments must be removed prior to automatic reprocessing (see pre-treatment) Equipment: Cleaning and disinfection unit according to DIN EN ISO 15883-1+2 with thermal programme (temperature 90 °C to 95 °C), detergent: mildly alkaline detergent (e.g. Dr. Weigert neodisher MediClean Dental).
- Place the instruments in a suitable small parts tray or on the load carrier such that all surfaces of the instruments are cleaned and disinfected.
- Close WD and start programme, see table below for programme sequence.
- Remove the instruments at the end of the programme.
- Check that the load is dry and, if necessary, dry with medical compressed air.
- Visual inspection for cleanliness is performed after removal from the WD. If contamination is still visible, reclean medical devices again manually. Subsequently, the recleaned medical devices must again be reprocessed automatically
PROG. STEP | WATER | DOSAGE | TIME | TEMPERATURE |
---|---|---|---|---|
Pre-rinse | CW | 5 min | ||
Dosage of detergent | According to manufacturer‘s instructions | According |
to manufacturer‘s instructions
Clean| Fully deionised water| | 10 min| 55 °C
Rinse| Fully deionised water| | 2 min|
Disinfect| Fully deionised water| | 3 min| Ao-value > 30001(e.g. 90 °C, 5 min)
Drying| | | 15 min| up to 120 °C
1 Authorities may issue other operational regulations (disinfection
performance parameters) in their area of competence.
MAINTENANCE, INSPECTION AND CHECK:
Equipment: Illuminated magnifying glass (3-6 dioptres)
All instruments must be inspected visually for cleanliness, integrity and
functionality, if necessary by using an illuminated magnifying glass (3-6
dioptres). All instruments are to be checked for damage and wear. Damaged
medical devices may no longer be used and must be sorted out.
PACKAGING:
Equipment: Film-paper packaging (e.g. steriCLIN, art. no. 3FKFB210112 and
3FKFB210140), sealing device (e.g. HAWO, type 880 DC-V) A suitable method
(sterile barrier system) is to be used to package the instruments. Packaging
according to DIN EN ISO 11607 A sterile barrier system (e.g. film-paper
packaging) according to DIN EN ISO 11607 is to be used, which is intended for
steam sterilisation by the manufacturer. The instruments are double packed.
The packaging must be large enough to avoid stressing the sealing seam.
Note: After the heat sealing process, the sealing seam must be checked
visually for any defects. In case of defects, the packaging must be opened and
the instrument repacked and sealed.
STERILISATION
Device: Steriliser according to DIN EN 285 or small steam steriliser
according to DIN EN 13060, type B process
Process : Steam sterilisation with fractionated pre-vacuum, 134 °C,
holding time min. 3 min (in Germany according to KRINKO/BfArM recommendation
134 °C min. 5 min) or 132 °C min. 3 min (parameter of validation). Longer
holding times are possible.
- Place the packaged products in the sterilisation chamber
- Start the programme.
- Remove the products at the end of the programme and allow to cool down.
- Then check the packaging for possible damage. Faulted packaging must be regarded as being non-sterile. The instruments must be repacked and sterilised.
STORAGE
Duration of storage according to own specifications. It is recommended to store instruments packed and protected from recontamination in proven suitable sterile packaging, cassettes or retainers.
ADDITIONAL INFORMATION:
CONTACT TO THE MANUFACTURER:
EVE Ernst Vetter GmbH, Neureutstr. 6, 75210 Keltern, Germany
Telephone: +49 (0) 72 31 97 77 -0,
Fax: +49 (0) 72 31 97 77 -99,
Email info@eve-rotary.com,
Web www.eve-rotary.com
THE INSTRUCTIONS FOR USE ON THE LABELS MUST BE OBSERVED.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>