CARE 55763 Two Point Hanger Bar
User Manual
Part no: 55763, 55751, 55764, 72754, 72755, 72757
55763 Two Point Hanger Bar
The “WARNING!” triangle is used when particular attention is required.
Read these instructions carefully before using the product. It is important
that the device is only used with the appropriate Human Care products.
Important instructions before use
Safety instructions
Before using the Hanger Bar for the first time, make sure that:
You have read and understood the instructions for use of the Hanger Bar, the
lift it is attached to and all of its accessories.
The Hanger Bar has been installed in accordance with the user manual.
The accessories which are intended for use with the Hanger Bar are correctly
fitted.
Before use, always check that the device is not damaged or worn. Especially
the lift hooks should be free of any cracks.
Warning: Product modification is not permitted.
Don’t drop the hanger bar. If dropped, perform a visual inspection of the
hanger bar and replace the complete hanger bar in case there are any cracks.
Important! The Hanger Bar is designed exclusively for lifting people, and
only one person at a time.
Human Care’s products are continuously being developed and updated to provide
our customers with the best products, we reserve the right to make design
changes without prior notice.
Preparations before use
Before and during use, always ensure that:
You read and ensure you understand this user manual.
Personnel, who use the equipment, have been given adequate training in the use
of the Hanger Bar and its accessories.
You should never leave the patient unattended during a lifting situation.
The lift belt is not twisted or worn and that it can move freely in and out of
the lift.
Accessories, e.g. loop slings and hanger bar, are showing no signs of wear or
other damages.
The lift is correctly suspended in the distance strap, rail roller and rail.
The right types, sizes, materials and designs of the accessories have been
selected with respect to the patient’s safety and needs.
The accessories have been correctly installed as described in the manual. The
sling’s strap loops and suspension must be checked once the loops are fully
extended when lifting but before the patient is lifted from the underlying
surface.
It is important to always check that the slings are at the bottom of all sling
hooks securely placed below the locking pin (see picture below).
Human Care provides suitable training in safe operation of the Hanger Bar and
other accessories.
Further information about slings and other accessories can be found at
www.humancaregroup.com.
Important! If the maximum load differs between lift, Hanger Bar and
slings, then the lowest maximum load shall always be used.
Important! To ensure that the lift belt will not be damaged, it is
important that the Hanger Bar is balanced and that the sling is suspended
correctly. Never hold or pull the lift belt and the Hanger Bar when moving a
patient. It is also important that the lift belt is tensioned as it is wound
in.
Technical specifications
| Hanger Bar for Altair/Roomer S| Hanger Bar for HeliQ
---|---|---
Art. No.:| 557635575155764| 727547275572757
Lifting capacity:| 300 kg/6611b
No. of possible connection
points| 2, 3 and 4| 2
Width:| 35cm/ 13,8 in| 45 cm / 17.7 in| 55cm/ 21.7 in| 35cm/ 13.8 in| 45cm/
17.7 in| 60 cm/ 23.6 in
Weight:| 1,0 kg/ 2,2 lb| 1,2 kg/ 2.6 lb| 1,5 kg/ 3.2 lb| 0,9 kg/ 2.0 lb| 1,1
kg/ 2.4 lb| 1,3 kg/ 6.6 lb
Available accessories: _| Full line of Human Care Loop Slings
Intended use of the sizes:
- 35 cm for children and slim patients.
- 45 cm for adults and appropriate for most of the lifting situations.
- 55 cm / 60 cm for patients who require more space in the sling because of broad-shouldered, obese or sensitivity to pressure.
Product description Hanger Bar for HeliQ
Item No. | Description |
---|---|
1 | Hook for lift |
2 | Locking Clip |
3 | Hanger Bar Profile |
4 | Hook for sling |
5 | Safety locking screw |
Product description Hanger Bar Altair/Roomer S
Item No. | Description |
---|---|
1 | Release button |
2 | Locking Clip |
3 | Hanger Bar Profile |
4 | Hook for sling |
Intended use
The Hanger Bar is designed exclusively for lifting people, only one person at a time, and shall never be held during lifting. The Hanger Bar has been tested for indoor use with Human Care loop slings only.
Product label
The product labels contain information about the hanger bars. One label is located underneath the hanger bar. The other label is located straight from the front of the hanger bar (72754, 72755, 72757) or on the top of the hanger bar (55763, 55751, 55764).
Symbols on the unit and product label
**** | Catalogue number symbol |
---|---|
EEEEEE | Item number |
Max. load | |
**** | Batch code symbol |
EAN | International article number |
UPC | Universal product code |
**** | Manufacturing date |
Manufacturing address | |
DDDDDD | Product name |
Label | Label item number |
CE mark | |
Human Care | |
Don’t drop the hanger bar | |
Perform a visual inspection | |
QR code to our webpage | |
Particular attention is required |
Attaching and detaching to HeliQ
Installation
The attachment shall only be conducted by trained personnel.
Attach the Hanger Bar to the Overhead Lift HeliQ
Feed out enough of the belt and connect the Hanger Bar to the loop on the lift
belt by threading the loop diagonally upwards until the tension of the belt is
released, see figure below. The product is now ready for use.
Detach the Hanger Bar from the Overhead Lift HeliQ
Feed out enough of the belt, until the tension of the belt is released, and
disconnect the Hanger Bar from the loop on the lift belt by threading the loop
diagonally downwards, see figure below. The product is now fully detached.
Attaching and detaching to Altair or Roomer S
Attach the Hanger Bar to the Overhead Lift Altair or Roomer S
Before connecting the Hanger Bar to the lift, make sure that both the shaft on
the lift and the Hanger Bar hole are properly cleaned. Connect the Hanger Bar
to the lift by pushing it (1) onto the shaft until you hear a click sound.
Make sure the Hanger Bar is fully attached by pulling it towards you and
pushing it away from you, and check that the release button is fully popped
out (2). The product is now ready for use.
Important! Never press the release button when performing a lift.
Detach the Hanger Bar from the Overhead Lift Altair or Roomer S
Disconnect the Hanger Bar from the lift by pressing the release button (1) and
then pulling the Hanger Bar out from the shaft, (2) see figure below. The
product is now fully detached.
Care instructions
Annual inspection
The Hanger Bar shall be inspected visually before every lift by the caregiver
and at least once a year by someo with Human Care qualifications and in
accordance with Human Care’s directive. Pay particular attention to pathat
show wear and which are critical for safety:
- plastic hooks should be without any deformation, cracks or damages
- safety locking screw (Hanger Bar HeliQ 72754, 72755, 72757) should be in place and thightend
- quick release button (Hanger Bar Altair/Roomer S 55763, 55751, 55764) should work properly:
when pressing in, it is possible to remove the hanger bar from the ceiling lift
when releasied, the hanger bar can’t be removed from the ceiling lift
Cleaning
When cleaning the surface and hole of the Hanger Bar only use acid- and
chloride-free detergents, like warm water or a soap solution, with a soft
cotton swab. Carefully clean the hole. The cleaning sessions shall always end
with water rinse and drying with a swab. The product shall be dry before
attaching to the lift.
Disinfecting
If the Hanger Bar needs to be disinfected, the surface of the entire product
can be cleaned using disinfectants, 70% ethanol, 45% isopropanol or similar
should be used.
Service
Installation, service and maintenance shall only be performed by personnel
with Human Care qualifications and in accordance with Human Care’s directive.
Only original spare parts may only be used.
Expected service life
The Hanger Bar is tested accordingly EN ISO 10535, which means it functions
for 11 000 lifting cycles at maximum lifting capacity or for 10 years with
correct service and maintenance.
Recycling
Once the product has reached the end of its technical life, it must be
disposed of in constituent parts according to local authorities. Metal parts
must be treated as metal waste for optimum recycling. Plastic parts can be
sorted as residual waste.
The latest version of the instructions of use is always available at our
website, www.humancaregroup.com.
Troubleshooting
Description | Action |
---|---|
The lifting hooks show a crack. | Replace the complete hanger bar for a new |
hanger bar. Discard the broken hanger bar.
Nothing happens when trying to lift a person.| Please go through the
troubleshooting in specific lift user manual.
The Hanger Bar cannot be detached from the Altair or Roomer S lift.| Make sure
the release button is fully pressed in and the Hanger Bar is unloaded before
trying to detach the Hanger Bar from Altair or Roomer S lift.
The loop slings cannot be detached from the Hanger Bar.| Make sure the locking
clip is folded up when removing the loop slings.
HUMAN CARE SWEDEN (HQ)
Årstaängsvägen 21B
117 43 Stockholm
Phone: +46 8 665 35 00
Fax: +46 8 665 35 10
E-post: info.se@humancaregroup.com
www.humancare.se| HUMAN CARE CANADA
10-155 Colonnade Road
Ottawa, ON K2E 7K1
Phone: +1 613 723 6734
Fax: +1 613 723 1058
Email: info.ca@humancaregroup.com
www.humancaregroup.com
---|---
HUMAN CARE UNITED STATES
8006 Cameron Street, Suite K
Austin, TX 78754
Phone: +1 512 476 7199
Fax: +1 512 476 7190
Email: info.us@humancaregroup.com
www.humancaregroup.com| HUMAN CARE
NETHERLANDS
Elspeterweg 124
8076 PA Vierhouten
Phone: +31 577-412171
Fax: +31 577-412170
Email: info.nl@humancaregroup.com
www.humancaregroup.com
997 16, Rev 05 , ML. User manual Hanger Bar , Januari 2022
References
- Safe Patient Handling Products & Walking Aids - Human Care USA
- Safe Patient Handling Products and Walking Aids - Human Care
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>