Cellaca MX High Throughput Automated Cell Counter
Contents of Shipping Container
- Cellaca MX Instrument
- Power Supply and Power Cord
- USB 3.0 Connector Cable
- Nexcelom-provided Laptop with Matrix Software (pre-installed)
- Cellaca MX User Manual (PDF file on Laptop)
- Matrix Software User Manual (PDF file on Laptop)
- Cellaca MX Quick Start Guide (PDF file on Laptop − this document)
- Cellaca MX Plate Loading Template − Graphic sheet indicating mixing/loading wells for both 12×2 and 8×3 plate layouts
- Focus Guide − Graphic sheet to help with adjusting instrument focus
- Nexcelom Counting Plates (24-well, 12×2 layout)
- AO/PI Viability Reagent
- Fluorescent Beads
Unboxing the Instrument
Unpack and visually inspect the Cellaca MX to ensure no damage has occurred during shipping. For assistance in setting up the instrument, visit the Cellaca MX page on our website, click the Resources tab and scroll down to find training Videos on unboxing and getting started. As the Cellometer MX instrument is run using the Matrix Software, the logo distinguishes it from earlier versions.
Site Preparation
Instrument must be placed on a level surface and plugged directly into a surge
protector (recommended) or power outlet. Ensure all cables are free from
tangles prior to starting the Cellaca MX.
Follow all equipment safety protocols and keep the area around instrument
clean both during and post operation. Do not position the device so that it is
difficult to disconnect from power main.
Setting Up the System
The Cellaca MX System is comprised of the instrument connected via USB cable
to an Operating Computer that is used to run the Matrix software. The
Operating Computer can be linked to a network for accessing external files,
printers and for storing count results.
As the Operating Computer provided by Nexcelom is a touch-screen device, you
can interact directly with the Matrix software GUI by tapping gently on screen
elements (e.g., tabs, drop-downs or buttons) using a finger or stylus. If you
choose to use touch-screen functionality, the term “click” as it appears in
this guide may be replaced with “tap” interchangeably. As an alternative, a
USB mouse may also be connected.
Connect the Power Supply/Power Cord to instrument and plug Power Cord into
electrical outlet. Connect Cellaca to Nexcelom-provided Laptop using USB 3.0
Connector Cable. Turn instrument Power Switch to ON position and confirm that
light bar appearing across the front is lit.
From desktop of Operating Computer, double-click the Matrix icon to launch the
software. The instrument will run through a startup sequence that includes
connecting to the database and initializing the calibrations.
Click the icon located in upper right corner of the screen to confirm that
text in the dialog box displayed indicates both the instrument and camera are
connected. Click the icon once again to close the dialog box.
Calibrating Your Instrument
After initial setup of an instrument, you must calibrate it using the Matrix
software prior to first use. The calibration process takes a background image
that will be used to normalize the cell counter for each installed filter pair
without a consumable counting plate loaded in the instrument. Contact Nexcelom
Support for assistance.
Matrix Screen Elements
Upon launching the Matrix Software you are presented with the Acquire tab >
Setup screen by default. Basic screen elements are described below.
Note: If the Matrix 21 CFR Part 11 module has been enabled for your
system, users must log in before they can begin using the software.
The Header Bar appearing across the top of the screen contains plate
Eject/Load buttons that control movement of instrument stage (for loading and
unloading of counting plates) and instrument status icons. It will also
display assay/plate names as you proceed through the count process.
The Navigation Bar appearing on left side of the screen is always displayed. Functionality associated with each tab is described below:
-
Home
Displays the logo distinguishing the software from prior versions. Also contains the About Matrix button which displays version details and Nexcelom contact information. -
Acquire
Displays screen in the Data Acquisition Process to which you have progressed. For example, when clicking the Acquire tab you will enter Setup details and then Preview the sample to confirm focus and fluorescent exposure. -
Data
Displays screen in the Data Analysis Process to which you have progressed. For example, when clicking the Data tab you will Select a scan result and then view count Results. You can also perform a Recount by updating parameter settings used in analyzing the scan to create a new result -
Manage
Displays options for customizing assays, cell types and report templates used in the counting/analysis workflow.
Staining Solution Guidelines
Use the following staining guidelines to prepare cell samples for Cellaca MX.
See Evaluating Viability Methods on page 7 for details on how to choose a staining method based on the cell sample and selected assay type.
Preparing Sample Plates
Although the Cellaca MX does not require any routine testing or calibration,
counting beads are available to verify instrument functionality. Nexcelom
counting beads CCBM-011-2ML are recommended for use with Cellaca MX.
To prepare sample plates containing counting beads or cell samples:
-
From desktop of Operating Computer, launch the Matrix software by double-clicking the Matrix software icon.
-
Place the Cellaca MX Plate Loading Template on the lab bench with the appropriate layout facing up and set a counting plate on top of the template graphic, aligning the notched corners of the plate. You should be able to view and identify the mixing/loading wells through the plate.
-
Invert counting bead solution or tube containing cell sample a total of 10 times. If using counting beads, vortex bead solution for 10 seconds. Do not vortex cell samples.
-
If using bead solution or a cell sample that does not need to be mixed, set pipette to 50 μL and then pipette bead solution/cell sample up and down ten times (10x) to break up any potential clumps. Load 50 μL of bead solution into Loading Well of counting plate.
If using a cell sample that needs to be mixed prior to loading, use the Mixing Well of counting plate to prepare the sample and then load 50 μL of cell sample into Loading Well of counting plate.
Note: Liquid will move from the Loading Well into the Imaging Area. -
From the desktop of Operating Computer, click the Eject button located at the top of the Acquire tab to open the instrument stage. Place the prepared plate onto the stage taking care to align the notched corners of the plate with the stage accordingly (i.e., well A1 is in top left corner).
-
Click the Load button to close the instrument stage.
-
In the Setup Details area of the screen, enter a Plate Name, select an appropriate assay from the drop-down and enter a Dilution Factor.
-
In the Well Details area, ensure the displayed Plate Type is correct, then click on the individual wells used to load the sample.
To select or clear multiple wells, click a well and hold/drag your mouse to encompass other wells. To select or clear all wells, click the button. -
Click the Preview button to view the sample.
Note: Cell concentrations of 1.0 x 105 − 1.0 x 107 cells/mL can be analyzed on the Cellaca MX, with a concentration of 1.0 x 106 cells/mL being optimal. Do not shake or vortex the sample as this may damage cell membranes.
Performing Image Clipping Check
- Click the Acquire tab and select Brightfield Channel Verification assay.
- Select a well in a corner of the plate and click the Preview button.
- Visually verify that no clipping has occurred around the edges.
- Repeat this step for wells in remaining corners of the plate.
Understanding Instrument Focus
Live cells should have a bright center and dark, crisp clearly defined edges.
Counting and Analysis Workflow
Performing a cell count consists of Entering Setup Details and Previewing the Sample Image before Reviewing Count Results. Upon launch of the Matrix software, the Acquire tab > Setup screen is displayed by default.
- Home Tab – Displays Home screen with instrument/software details.
- Acquire Tab – Use to enter Setup Details and Preview the sample image.
- Data Tab – Use to view current scans/select saved scans for re-analysis.
- Manage Tab – Displays Assay, Cell Type and Report Template libraries.
- Eject Button – Controls instrument stage (toggles between Eject/Load).
- Setup Details – Use to enter plate name, select an assay, add a tag (to group scan results for custom reporting) and dilution factor.
- Well Details – Use to enter plate type.
- Well Selection – Displays a plate map for use in selecting wells.
- Well Names – Use to define well names for selected wells.
- Auto Export Area – Use to define auto export location and options.
- Preview Button – If plate is loaded and Setup details entered, displays sample in Preview mode to adjust focus and fluorescence signals.
Entering Setup Details
- In the Acquire tab, Setup screen, enter a plate name and select an assay. If you do not enter a plate name, a date/time stamp is appended to the “New Sample” default (e.g., New Sample 2021/03/01-11:58:09).
- Select plate type (e.g., 12×2 or 8×3 formats) and click on loaded wells in the plate map. To select wells individually, click on each well to select or de-select it accordingly. To select a group of wells, click on a well at the beginning of the group and hold the mouse button down while dragging your mouse to the end of the group before releasing it. To select or de-select all wells, click the All Wells icon.
- If desired, enter Auto Export options including a default export location and image file types for automatically saving scan results.
- Click the Preview button.
Previewing Live Images
The Preview screen displays live images for selected wells. Current Zoom
magnification is displayed in the bottom right corner of viewing pane.
-
In the Wells area, click on highlighted wells to view live images of samples in the plate. As you move from well to well, the live image being displayed changes per the selected well (i.e., which is indicated in the plate map by an outline surrounding the well). To zoom in and out of the image, turn the scroll wheel on your mouse or, if using the touch screen, apply universal gestures (e.g., touch the center of the image with two fingers and then slowly spread them apart to zoom in and reverse this action to zoom out.) Current Zoom magnification is displayed in bottom right corner of the viewing pane. To move around an image, click once inside the viewing pane and hold your mouse button down while dragging the image to a new location.
-
In the Preview area, the channels displayed for viewing (e.g., Channel 1|Brightfield and Channel 1|Fluorescent, and Channel 2| Brightfield and Channel 2|Fluorescent) will depend on configuration of your instrument and Imaging Mode of the assay. The Channel 1| Brightfield image is displayed by default as shown below.
-
To view other available channels (i.e., channel display is based on selected assay), click the FL buttons accordingly. For example, sample Channel 1|FL and Channel 2|FL images are shown below.
-
Adjust the focus for brightfield image and verify fluorescent exposure.
Note: If a fluorescent image is available for Channel 1, use the BR 1 image to adjust instrument focus and then select the FL 1 image to confirm exposure. If a fluorescent image is available for Channel 2, you need only to confirm fluorescent exposure. Keep in mind that a slight offset exists between the BR 1 and BR 2 images due to the distance the camera has to travel between the two filter sets.
To adjust focus of the live image being previewed, use the Focus Controls as indicated in the following table. Obtaining good focus is key to ensuring accurate cell counts.-
Auto Focus
Allows instrument to determine the best focal position for the selected well. -
Position
Allows users to enter a numerical value for vertical (Z) position of the objective lens. -
Fine Focus Manual Offset
Allows users to finely adjust vertical (Z) position of the objective lens for optimal focus (in μm). Click the up/down single arrows to adjust focus. -
Coarse Focus Manual Offset
Allows users to coarsely adjust vertical (Z) position of the objective lens for optimal focus (in μm). Click the up/down double arrows to adjust focus.
-
-
Click the Count button.
Reviewing Count Results
As the system acquires sample images, the colors used to mark selected wells
change to indicate well status (i.e., from Selected to Acquired to Counted) as
shown in the legend displayed below the plate map. As soon as a well is
Counted, you can click on that well to display its results below the viewing
pane. Count results are displayed in the Well View tab using the assigned
Display report template. Additional tabs may be enabled for the scan result
(e.g., Summary and .csv tabs for Matrix v4.0; All Wells tab for v3.0). Count
results will be displayed, printed and exported based on Reporting options
defined for the assay.
When viewing count results, a single image is displayed representing the sample in the outlined well. Click on other highlighted wells to view them.
- In the Wells area, click on a well to display its image in the viewing pane.
- At the top of the Well View tab, click the brightfield (BR), Fluorescence (FL1) or Dual Fluorescence buttons (FL2) to change the presentation of the image. Channel views are overlaid on top of each other.
- Click the Zoom button to enable/disable display of the current Zoom magnification in the bottom right corner of viewing pane.
- Click the Counted button to show/hide a graphic overlay that highlights Counted cells by circling them with color-coded outlines (e.g., Green for counted/live cells, Red for dead cells and Yellow for cells not counted because they were larger than the specified cell diameter).
- At the bottom of the Well View tab is a report containing well-level details for the sample image in the selected well.
- Click the Summary tab (displayed for a Matrix v4.0 assay) or All Wells tab (displayed for a v3.0 assay) to view a full page containing count results for all selected wells.
Exporting/Printing Scan Result Data
To update export options for scan result files (Auto Export options may be
defined in the Setup Screen) as well as custom report files, click the Export
button located at the bottom of the Results screen.
- Confirm the current export location. Click Browse, navigate to a folder on your Operating Computer or network and click OK.
- Select the scan result files (e.g., Raw Images, Colorized Images, Data Set) to be exported. Data sets contain all images, results, assays, cell types and report templates associated with the scan result and are stored as .SCANRESULT files.
- For reports, select file types (i.e., Excel, CSV, PDF, Word) to be exported and if files are to be automatically opened/printed upon export.
- Click the Export and Print button.
Viewing the Data Tab
The Data tab contains the Select, Results and Recount screens which must be
completed in sequence when analyzing scan results (i.e., you must select a
scan result before you can view its data and then fine-tune assay/cell type
parameter settings before you can perform a recount).
To open a scan result contained in the Results List of the Select screen you can either double-click the result, or click it once (to highlight it in the list) and then click the View button. Once a scan result is displayed you can analyze data by clicking channels and varying the presentation of images displayed. To fine-tune assay parameter settings, click the Recount button. If it becomes necessary to select another scan result while you are in the Results screen, click the Select option in the Navigation Bar to choose another scan result. To fine-tune assay parameters for the scan result, you can click the View button for either the Last Used assay or a selected new assay, and edit its parameters as necessary. To perform a recount, click the Recount button. As you navigate through each screen, the Navigation Bar is automatically updated a to reflect your progression. After a recount is performed, the Navigation Bar returns to the Data tab > Results screen.
Viewing the Manage Tab
The Manage tab lists all Assays, Cell Types and Report Templates loaded in
your system. From this screen you can import/export, rename, delete or
show/hide any of these entities in their respective lists.
The State column displays icons that identify the current state of an entity indicating whether they are Locked, Unlocked or a Nexcelom System default. Although a locked/system assay, cell type or report template cannot be edited, you can select it to use as a source for creating a new entity and click the Save As button to save it using a new name.
Nexcelom provides extensive assay, cell type and report template libraries as
defaults in the Matrix software. If you would like help creating a new assay,
customizing cell type parameter settings for an assay, or creating a new cell
type, send saved raw images to Nexcelom Support and we will optimize a custom
assay/cell type for you. For help with customizing report templates, share
with us the goal you are trying to accomplish and we will create a new
template to accommodate your needs.
For a complete list of available assays, cell types and report templates,
contact Nexcelom Support.
Evaluating Viability Methods
When evaluating viability methods, it is critically important to use a single
aliquot from the stock cell culture to perform all testing. The cell sample
should be evaluated for concentration on the Auto 2000 prior to staining.
If comparing the Trypan Blue and AO/PI methods, a portion of the sample should
be stained with trypan blue and another portion stained with AO/PI.
Using Trypan Blue Viability Method
Brightfield imaging used in conjunction with Trypan Blue Viability may be used
to determine the number, concentration and percentage of live cells for cell
lines and cultured primary cells. Brightfield imaging with trypan blue is not
recommended for samples containing debris, platelets or red blood cells as
fluorescence is required to accurately differentiate nucleated cells.
Preparing Cell Sample for Trypan Blue Viability Determination
Cell concentrations of 1.0 x 105 − 1.0 x 107 cells/mL can be analyzed on the
Cellaca MX, with a concentration of 1.0 x 106 cells/mL being optimal.
Invert the tube containing cells ten times (10x) and pipette up and down 10x
to generate a homogeneous cell sample and reduce cell clumps. Do not shake or
vortex the sample as this may damage cell membranes.
For viability measurement, stain cells by combining 50 μL of cell sample with
50 μL of a 0.2% trypan blue staining solution (for a final concentration of
0.1% trypan blue). Gently mix by pipetting up and down 10x.
Using AO/PI Viability
Dual-fluorescence methods have been developed to accurately determine
nucleated cell concentration and viability in primary cell samples containing
debris and non-nucleated cells, including platelets and red blood cells.
In the AO/PI viability method, acridine orange (AO) enters all cells and
stains their DNA causing nucleated cells to fluoresce Green, while propidium
iodide (PI) enters only dead cells with compromised membranes and stains their
DNA causing dead nucleated cells to fluoresce Red.
- Cells stained with both AO and PI fluoresce Red due to quenching.
- Live nucleated cells are easily identified in the Green FL channel.
- Dead nucleated cells are easily identified in the Red FL channel.
As a result, there is no interference from debris and non-nucleated cells when using the AO/PI viability method.
Preparing Cell Sample for AO/PI Viability Determination
Cell concentrations of 1.0 x 105 − 1.0 x 107 cells/mL can be analyzed on the
Cellaca MX, with a concentration of 1.0 x 106 cells/mL being optimal.
Dilution or concentration of a cell sample may be required based on the
initial concentration.
Invert the tube containing cells ten times (10x) and pipette up and down 10x
to generate a homogeneous cell sample and reduce cell clumps. Vortexing may
help for some samples, but may generate bubbles that make it difficult to
pipette when working with small sample volumes.
If necessary, dilute original sample with PBS. Stain cells by combining 50 μL
of cell sample with 50 μL of AO/PI staining solution. For whole blood and
other viscous samples, draw sample in and out of the pipette tip prior to
transfer. Gently mix by pipetting up and down 10x before adding sample to
counting plate wells.
The table below shows the recommended dilution when preparing cell samples for
AO/PI viability analysis and the final Dilution Factor to enter into Matrix
software for a variety of sample types.
Contacting Nexcelom Support
Nexcelom Support is available from 8:30 am to 5:30 pm EST and can be reached
at +1 978-327-5340 or by sending email:
support@nexcelom.com
Trained specialists are available to assist your team with sample analysis and
optimization of assay/cell type imaging parameters.
See the following documentation for additional instrument information:
- 8002958 Cellaca MX User Manual for instrument operation, care and maintenance details (available as a PDF on desktop of the Nexcelom-provided Laptop shipped with your instrument).
- 8003394 Matrix Software User Manual for complete details on using software functionality (available as a PDF on desktop of Nexcelom-provided Laptop shipped with your instrument).
- 8002845 Cellaca MX Focus Guide for assistance with optimal focus.