HTC INSTRUMENTS PDO-5000 Digital Storage Oscilloscope User Manual
- June 9, 2024
- HTC INSTRUMENTS
Table of Contents
PDO-5000 Digital Storage Oscilloscope
PDO-5000
Introduction
Dear users: Thank you for purchasing HTC product. To operate the instrument
correctly, please read this Operating Manual carefully and especially its
“Safety Notice” before use. Afer reading it, you are suggested to keep the
manual properly. Please keep it with the instrument together or place it in an
accessible location for future use.
General Safety Overview The instrument is designed and manufactured strictly according to “Safety Requirements of Electronic Measuring Instruments” (GB4793) and IEC61010-1 Safety Standard . It conforms to “Insulation Overvoltage Standard” (CAT 600V) and Safety Standard with Pollution Level You are expected to know the following safety precaution measures to avoid personal injury or product damage or any other product connected to it. Please use the product according to the regulations so as to avoid the possible damage. Only personnel with professional training can execute maintenance procedure to avoid fire or personal injury. Appropriate power wires should be used and special power wires must be approved by the state. Correct plug- in/out: Please do not plug in/out when connecting probe or testing conduits to voltage source.
Reliable grounding of products: The product is grounded by ground conductor of power supply. Ground conductor must be connected to earth to avoid electric shock. Please guarantee correct grounding before connecting input/output terminal of the product. Correct connection of oscillograph probe: Earth wire of probe is the same to earth potential. Please do not connect earth wire to high voltage. Review of rated values of all terminals: Please refer to all rated values and marks on the product surface to avoid fire or excessive current shock. Please refer to product manual before connecting products so as to know details of rated values.Do not open cover plate of machine box: It is not allowed to start or operate the product when opening external cover or panel.
Usage of appropriate fuse: It is only allowed to use fuse type and rated indicator designated by the product. To avoid exposed circuits: It is not allowed to touch exposed connector or element after power supply. Operation is not allowed if there is any doubt for product trouble: Please consult qualified maintenance personnel to inspect the product if there is any trouble doubt. It has to maintain appropriate ventilation. It is not allowed to operate in wet locations. It is not allowed to operate in combustible and explosive environment. It has to keep product surface clean and dry.
Safety Terms and Symbols Terms in the manual: The following terms may be
seen in the manual:
Product symbol: Following symbols may be shown on the product surface.
Warning: Warning statement means the conditions and behaviors that may
endanger lives.
Note: Notice statement means the conditions and behaviors that may cause other
property damage
Terms for products: The following terms may be Danger: It refers to direct
injury near the mark. Warning: It refers to potential injury near the mark.
Note: It refers to potential danger for the product and other properties.
High voltage
Caution! Refer to manual
Protective
Ground terminal Ground terminal
ground terminal
for chassis
for testing
Preface
The manual is about operation information of OPERATING MANUAL FOR PDO-5000
digital storage oscillograph. The manual includes the following chapters
Chapter 1: User Guideline: It is to introduce functions of digital storage
oscillograph easily and provide installation guidelines. Chapter 2: Instrument
Setup:Operation methods of OPERATING MANUAL FOR PDO-5000 digital storage
oscillograph. Chapter 3: Application example: Examples to solve various
measurement problems Chapter 4: System Prompt and Troubleshooting Chapter 5:
Appendix: Technical Indicators Appendix B: Fittings of OPERATING MANUAL FOR
PDO-5000 series digital storage oscillograph Appendix C: Maintenance and
Cleaning Maintenance
The manual includes 2 series of digital storage oscillograph as follows. Detailed models are shown in the following form:
PDO-5000
Series
PDO-5000E
PDO-5000S
Model
PDO-50200S PDO-5025S PDO-5070S PDO-50100S
Band width
200MHz 25MHz 70MHz 100MHz
Real-time
1GS/s 250MS/s
1GS/s 1GS/s
PDO-5000 series digital storage oscillograph provides simple front panel with clear functions to users for all basic operations. Scale and position of various channels can provide intuitional operation to meet usage habits of traditional instruments. Users can use it skillfully without learning and being familiar with operation for a long time. Users can also press AUTO key directly to accelerate adjustment and facilitate measurement. Appropriate waveform and gear setup can be shown on the instrument panel. PDO-5000 series digital storage oscillograph also has high-performance indicators and powerful functions to fulfill measurement tasks quickly besides easy usage. Quicker signals can be observed by PDO-5000 series digital storage oscillograph by real-time sampling of 500MS/s (or 1GS/s) and equivalent sampling of 25GS/s (or 50GS/s). It can capture and analyze waveform
easily by powerful trigger and analysis capacity. Users can also observe and analyze signal problems quickly and clearly by clear LCD display and mathematical calculation function. We can see that such series of digital storage oscillograph can meet your measurement requirements by following performance characteristics: Double analog channels High-resolution colorful LCD display system with resolution of 320 ×240 (or 800×480) To support plug and play USB storage instrument to communicate with computer Automatic waveform and status setup Waveform, setup, bitmap storage, waveform and setup reoccurrence
Elaborate view window extension function for
delicate analysis of waveform details and profile Automatic measurement of 28
kinds of
waveform parameters Automatic cursor tracing and measurement
function Unique waveform recording and playback
functions Embedded FFT Multiple mathematical calculation functions (including
+, -, ×,÷) for waveform Edge, video, pulse width and ALT TRIG
functions Multi-language menu display Fittings of PDO-5000 series digital
storage oscillograph 2 probes of 1.2m, 1:1/10:1; please refer to
manual of probe Fittings of probe details so as to meet EN61010-031:2008
standard.
1 power wire to meet national standards; 1 “Operating Manual”; 1 “Product Warranty” USB connecting wire: UT-D06 or UT-D05 Communication control software for PDO-5000 series oscillograph
Contents
Item
General Safety Overview Preface
Page
Chapter 1: User Guideline ……………………………………………………………………….. 1
General Inspection ………………………………………………………………………………. 5 Function Inspection
……………………………………………………………………………… 5 Probe Compensation
…………………………………………………………………………… 8 AutomaticSetup of Waveform
Display……………………………………………………. 9 Elementary understanding of Vertical
System………………………………………… 9 Elementary understanding of Horizontal
System……………………………………. 11 Elementary understanding of Trigger
System……………………………………. 14
Chapter 2: Instrument Setup…………………………………………………………….. 15
Setup of Vertical System………………………………………………………………. 16
Setup of Horizontal System……………………………………………………………. 27 Setup of Trigger
System………………………………………………………………… 31
Item
Page
Setup of Sampling System……………………………………………………….. 44 Setup of Display System………………………………………………………………………… 48 Storage and Callout ………………………………………………………………………………. 50 Setup of Auxiliary Functions …………………………………………………………………… 55
Automatic Measurement………………………………………………………………………… 60 Cursor Measurement …………………………………………………………………………….. 68 Operation Key ………………………………………………………………………………………. 70
Chapter 3: Application Examples …………………………………………………………. 72
Example 1: Measurement of Simple Signals …………………………………………….. 72 Example 2:
Delay Observation of Sine Signals by Circuit………………………… 73 Example 3: Capture
of Single Signal……………………………………………………. 75 Example 4: Reduction of Random Noise
for Signals……………………………….. 76 Example 5: Application Cursor
Measurement………………………………………… 78 Example 6: Application of X-Y
Function………………………………………………… 79 Example 7: Trigger of Video
Signals………………………………………… ………….. 81
Item
Page
ExampExample 8: Pass/Fail Detection ……………………………………………….. 83 ExampExample 9: Usage of Upgrading Procedure of U disk ………………… 84 Example 10: Usage of Storage Function ……………………………………………… 85
Chapter 4: System Prompt and Troubleshooting ………………………………….. 88
System Prompt Information ………………………………………………………………….. 88
Troubleshooting…………………………………………………………………………………… 89
Chapter 5: Technical Indicators…………………………………………………… 91
Appendix A: Technical Indicators………………………………………………. 91
Appendix B: Fittings of PDO-5000 Series Digital Storage Oscillograph…
104
Appendix C: Maintenance and Cleaning Maintenance………………………….. 105
Chapter 1: User Guideline
PDO-5000 series digital storage oscillograph is a small and handy desktop
product. Basic test can be done by providing convenient front panel to
facilitate operation for users. PDO-5000 series are to provide simple front
panel with clear functions to users for basic operation. Panel also includes
knob and function key. Knob functions are similar to those of other digital
storage oscillograph. The chapter is to show execution of following steps
General inspection Function inspection Probe compensation Automatic setup of
waveform display Elementary understanding of vertical system Elementary
understanding of horizontal system Elementary understanding of trigger system
General inspection Please know front operation panel firstly after getting PDO-5000 series digital storage oscillograph. The chapter is to introduce operation and functions of front panel for PDO-5000 products so as to be familiar with usage within the shortest period. 5 keys in a row in the right side of display panel are menu operation keys. (They are defined as F1 to F5 keys from top to bottom.) You can also set different options for current menu by them. Other keys are function keys. You can enter into different function menus or gain specific function application directly by them.
1
PDO-2102E
multi-purpose common Operation knob controller menu controls
trigger control horizontal control vertical control
interface
analog signal input
5.7-inch front panel of HTC
probe compensation signal output EXT TRIG INPUT
2
PDO-50100S
multi-purpose common Operation
knob controller menu
controls
trigger control horizontal control vertical control
interface
7-inch front panel of HCT
probe compensation signal output EXT TRIG INPUT analog signal input
3
display of horizontal trigger position
display of trigger status
time display of central scale line
mark of channel 1
mark of channel 2
Menu varies with different function keys.
waveform display window
vertical scale coefficient of display channel
display setup of main time base
Interface display diagram of digital storage oscillograph (i.e. PDO-5000S series)
4
General inspection It is suggested to inspect the instrument by following steps when getting a new PDO-5000 series digital storage oscillograph. 1. To check if there is damage due to transportation or not. Please replace carton or foaming plastic protection cushion with severe damage instantly. 2. Please inspect fitting details and shown it in “Fittings of PDO-5000 series digital Storage Oscillograph” shown in the manual. You can check if there are deficient fittings or not by referring to it. Please contact HTC distributor or local agent if there are deficient or damaged fittings. 3. Please inspect the entire machine and contact HTC distributor or local agent for damaged appearance, abnormal operation or failing to pass performance test. Please pay attention to packages and report to
transportation department and HTC distributor if machine damage is caused by
transportation. HTC will arrange maintenance or replacement. Function
inspection Please carry out once fast function inspection to check if the
machine operates normally or not.
Please operate according to following steps:
1. Connect the power supply You can supply power to the machine and voltage
is AC 100V to AC 240V. Frequency is 45Hz to 440Hz. It has to press [UTILITY]
menu and [F1] for self-correction after power supply 30 minutes after
operation of hot machine so that digital storage oscillograph operates under
the best status. As shown in Figure 14, please enter into next page and press
[F1] to call out ex-factory setup.
5
power switch
PDO-50100S
MULTIPUR POSE
M EA SU RE
A CQUIR E
S TO R AG E
S EL ECT
CURS OR
DISPLAY
UTILI TY
AUTO
VERTICAL
HORIZONTAL
TRIGGER
F1
P OSITIO N
P OS ITIO N
LEV E L
F2
CH 1
M AT H
HOR I MEN U
S ET
TRIG MEN U
F3
S CAL E
CH2
S CAL E
FORCE
F4
V
mV
VOLTS/D IV
F5
CH1
S
S
SEC /D IV
CH2
HE LP
function button
Figure 1-4
Warning: Please confirm safety grounding of digital storage oscillograph to
avoid dangers. 2. Signal access of digital storage oscillograph is double-
channel input for PDO-5000 series digital storage oscillograph. There is also
1 EXT TRIG INPUT channel. Please connect signals according to following steps:
(As shown in Figure 1-5), connect probe of digital storage oscillograph to CH1
input terminal and set attenuation ratio switch on probe to 10×Á
Figure 15 Setup of attenuation ratio switch on probe
6
.In digital storage oscilloscope to set the probe
attenuation factor. The attenuation coefficient of vertical gear ratio change
instruments, which makes measurement resultsSetup method of attenuation
coefficient of probe is as follows to show amplitude of measured signals
correctly. Please press F4 key for menu display of 10×
. Connect probe and grounding clip to
corresponding connection ends then press [AUTO] button. As shown in Figure
1-7, square wave (1kHz, about 3V and peak value) can be displayed within
several seconds. It has to inspect CH2 by the same method then press [OFF]
function button so as to power off CH1. It has to press [CH2] function button
to power on CH2 and repeat the second and third steps.
Probe ration
Figure 1-6 Setup of deflection coefficient of probe on digital storage
oscillograph
7
Figure 17
Probe compensation It needs to adjust it when connecting probe to any input channel for the first time so that probe can be appropriate to input channel. Probe without compensation correction can also bring in measurement tolerance or error. Please operate according to following steps for adjustment of probe compensation: 1. To set probe menu attenuation coefficient to 10× and place probe switch to 10× then connect digital storage oscillograph probe to CH1. Please guarantee reliable contact to probe. It has to connect probe end to signal output connector of probe compensator then connect grounding clip to earth wire connector of probe compensator. It also has to power on CH1 and press [AUTO].
2. Observation of displayed waveform
Exceeding
Correct
Insufficient
compensation compensation
compensation Figure 18 Compensation correction of probe 3. Adjust variable capacitance on the probe by
screwdriver of non-metallic handle for waveform
display of “insufficient compensation” or
“excessive compensation” shown in above figure
until “correct compensation” for waveform display
on the screen as above.
Warning: Please guarantee perfect insulation
conduits of probe and do not contact metallic part
of probe when connecting to HV source to avoid
electric shock of probe during HV measurement.
8
Automatic setup of waveform display PDO-5000 series digital storage oscillograph has automatic setup functions. It can adjust vertical deflection coefficient, scanning time base and trigger mode automatically according to input signal for the most appropriate waveform display. Frequency of measured signals shall be more than or equal to 50Hz and duty cycle shall exceed 1% according to automatic setup requirements. Automatic setup 1. Connect measured signals to signal input channel. 2. Press [AUTO] button so that digital storage oscillograph can set vertical deflection coefficient, scanning time base and trigger mode automatically. They can be adjusted again after careful observation and automatic setup so that waveform display can realize the best effect.
Elementary understanding of vertical system As shown in the following figure, there are a series of keys and knobs in vertical control area. Please introduce usage of vertical setup gradually as follows.
5.7-inch vertical
7-inch vertical control
control panel of screen panel of screen
Figure 19 Vertical control area on the panel
9
1. Use vertical position knob so that middle display of waveform signal in the window. Vertical display position of signal can be controlled by vertical position knob. Ground mark moves vertically with waveform when rotating vertical position knob. Measurement knack You can observe difference between waveform and signal ground to measure DC component of signal quickly if adopting DC channel coupling mode. DC components in the signal will be filtered if adopting AC coupling mode . You can also display AC component of signal by higher sensitivity conveniently.
Vertical position of double analog channels shall be restored to short-cut key
of zero pointnamely, SET TO ZERO. The key is to return vertical
displacement/horizontal displacement/hold off position to zero point (or
middle point). Of which: PDO-5000S series will trigger electric level Return
to 50% of position.
2. Alter vertical setup and observe alternation of status information. You
can determine variation of any vertical gear by status bar below waveform
window. Vertical gear of “volt/grid” can be altered by rotating vertical scale
knob to show corresponding variation for channel gear display which
corresponds to status bar. User can press [CH1], [CH2] , [MATH] and [REF] to
display operation menu, mark, and waveform and gear status information of
corresponding channels.
10
Users can also press [OFF] key to close currently selected channels for PDO- 5000E oscillograph. They can also click on CH1, CH2 and MATH keys twice to close corresponding channels for PDO-5000S oscillograph.
Elementary understanding of horizontal system As shown in following figure, there is 1 key and 2 knobs in horizontal control area. Please be familiar with setup of horizontal time base gradually as follows:
Horizontal control panel of 5.7-inch
Horizontal control panel of 7-inch
Figure 110 Horizontal control area on the panel
11
1. Use horizontal SCALE knob to alter horizontal time base gear then observe alternation of status information. To rotate horizontal SCALE knob to alter “s/div” time base gear so as to show corresponding variation of time base gear display for corresponding channels of status bar. Horizontal scanning speed steps from 2ns/div 50s/div by way of 125 mode. Note: Horizontal scanning time base gear varies for PDO-5000 series digital storage oscillograph due to different models. 2. Use horizontal POSITION knob to adjust horizontal position of signals in waveform window. To use horizontal POSITION knob to control trigger displacement of signals. Horizontal movement of waveform with knob can be observed when rotating horizontal POSITION knob for application for trigger displacement.
3. Please press [MENU] button for 5.7-inch oscillograph (and press [HORI MENU] button for 7inch oscillograph) to display Zoom menu. Please press [F3] to open view window extension under this menu then press [F1] to close view window extension and return to main time base. Hold off time can be set under this menu.
12
Trigger point can be restored to vertical middle point quickly by short-cut key of [SET TO ZERO] so as to restoring displacement of trigger point to horizontal zero point short-cut key. Users can also adjust horizontal position of signal in waveform window by rotating horizontal POSITION knob.
Elementary understanding of trigger system As shown in Figure 1 11, there is 1 knob and 3 (or 2) keys in control area of trigger menu. Please be familiar with setup of trigger system gradually as follows:
Term explanation Trigger point: It refers to position of actual trigger point
by comparing with middle point of storage instrument. Users can move trigger
horizontally by rotating horizontal POSITION knob. Hold off: It refers to
interval to start up trigger circuit again. Hold off time can be set by
rotating multi-purpose knob controller.
13
5.7-inch screen 7-inch screen Figure 1-12 Trigger menu
Figure 111 Trigger menu on the panel
1. To user trigger level knob to alter trigger level
in trigger signal amplitude.
and instruct trigger level line by seeing trigger mark 4.To press FORCE button: To produce 1 trigger
on the screen so that it moves vertically with knob signal forcefully for mainly application for normal
rotation. Variation of trigger level in lower part of
and single trigger mode.
screen can be observed when moving trigger level
at the same time.
2. (As shown in Figure 1-12,) TRIGGER MENU can
be used to alter trigger setup.
User can press [F1] key to select “edge” trigger.
They can press [F2]key to select “trigger source”
to CH1. They can press [F3] key to set “gradient” of edge
type to ascending.
They can press [F4 ]key to set “trigger mode” to
automatic. They can also press F5 key to set
“trigger coupling” to AC.
2. They can press [50%] button for 5.7-inch oscillograph
and press [SET TO ZERO] key for 7-inch oscillograph.
It also has to set vertical middle point of trigger level
14
Chapter 2: Instrument Setup
By now, you have been familiar with operation of vertical control area,
horizontal control area and trigger system menu for PDO-5000series digital
storage oscillograph preliminarily. You should be familiar with setup of
digital storage oscillograph by menu operation by introduction shown in
previous chapter. It is suggested to read Chapter 1 if you have not been
familiar with above operation and methods. The chapter is mainly to introduce
following themes Set vertical systems of [CH1], ]CH2],] MATH], ]REF] , ]OFF]
]VERTICAL POSITION] and ]VERTICAL SCALE]; Set horizontal systems of ]MENU] or
]HOR MENU], ]HORIZONTAL POSITION] or [HORIZONTAL SCALE]; Set trigger systems
of [TRIGGER LEVEL], [MENU], ]50%] and] FORCE]; Set sampling mode of [ACQUIRE];
15
Set display mode of [DISPLAY]; Store and call-out for [STORAGE]; Set auxiliary
system of [UTILITY]; Automatic measurement of [MEASURE]; Cursor measurement of
[CURSOR]; Use execution buttons of [AUTO] and [RUN/STOP]; It is suggested to
read the chapter carefully to know
many measurement functions for PDO-5000 and system operation method.
Setup of vertical system CH1 and CH2 channels and setup Every channel has its own vertical menu. Every item can be set individually according to different channels. As shown in Form 2-1, the system can display operation menu of CH1 or CH2 channel by pressing CH1 or CH2 function key.
Form 2-1
Function menu Setup
Description
Coupling
AC To obstruct DC component of input signal.
DC To cut off input signals by AC
Grounding and DC components of input signals.
Bandwidth limit
Volt/grid
ON
OFF Coarse tuning
Fine tuning
To limit band width to 20MHz so as to reduce display noise. Full band width To set vertical deflection coefficient for coarse tuning by 1-2-5 system. To refine within setup scope of coarse tuning for fine tuning so as to improve vertical resolution.
Probe
1× 10× 100× 1000×
To select one value according to probe attenuation
coefficient to keep correct vertical deflection coefficient.
There are totally 4 kinds, namely, 1×, 10×, 100× and 1000×.
ON Reverse phase
To realize reverse direction function for waveform;
OFF To display waveform normally. 16
1 Setup of channel coupling Measured signal is a sine signal with DC component by taking exerting signal to CH1 channel as an example. To press F1 to select AC and AC coupling mode. DC component contained by measured signal will be obstructed. Waveform display is shown in following figure.
Press F1 to DC to pass through DC and AC components of measured signals of CH1
channel. Waveform display is also shown in following figure:
DC coupling setup
Setup of AC coupling
Figure 21Jammed DC component of signal
Figure 2 2 Synchronous display of DC and AC components of signal Press F1 to select grounding and set channels to grounding mode. DC and AC components contained by measured signals will be obstructed. Waveform display is shown in following figure: displayed on the screen.
17
(Note: Input signal still shall be connected to channel circuit under such
mode although waveform is not
Ground coupling setup
2. Setup of channel bandwidth limit Input one sine signal of about 40MHz for CH1 as an example. It has to press [CH1] to open CH1 channel then press [F2] to set bandwidth limit to OFF. Channel band width is full band width to pass through high-frequency components contained by measured signal. Waveform display is also shown in following figure.
Figure 23 Synchronous obstruction of DC and AC components of signal
Bandwidth suppression 20MHz
Bandwidth suppression icon Figure 24 Waveform display when closing bandwidth
limit
18
Please press [F2] to set band width limitation to ON for great attenuation of
high-frequency component and noise of more than 20MHz contained by measured
signals. Waveform display is also shown in following figure.
Close bandwidth limit
3. Setup of multiplying factor for probe It has to set probe attenuation coefficient in operation menu of channel to coordinate with attenuation coefficient setup of probe. Probe coefficient in channel menu shall be set to 10× if probe attenuation coefficient is 10:1. It has to analogize to guarantee correct voltage. Volt/grid gears of vertical deflection coefficient can be divided into coarse tuning and fine tuning. Please refer to following figure for setup and vertical gear when using 10:1 probe.
Figure 25Waveform display when opening bandwidth limit
19
Probe attenuation coefficient
Figure 26 Probe attenuation coefficient shown in channel menu
4. Regulation setup of vertical volt/grid Volt/grid scope is 2mV/div 5 V/div (or 10V/div), or 1mV/div20V/div with step by way of 125. Deflection coefficient shall be altered within current vertical gear scope by less step during fine tuning so as to realize uninterrupted and continuous adjustment of vertical deflection coefficient within all vertical gears.
5. Setup of reverse phase of waveform Reverse phase of waveform: It is to show overturning degree of phase location of displayed signals. Please refer to Figure 28 for waveform without reverse phase. Please refer to Figure 29 for waveform with reverse phase.
Setup of fine tuning
Figure 27 Coarse tuning and fine tuning of vertical deflection coefficient
Without reverse phase Figure 28 Setup of reverse phase of vertical channel (or
without reverse phase)
20
I. Realization of mathematical calculation function Mathematical calculation
functions are to show +, -, ×, / as well as FFT calculation results for CH1
and CH2 channel waveform. Menu is shown as follows: Reverse phase of waveform
Figure 29 Setup of reverse phase of vertical channel
Mathematical calculation Figure 210 Mathematical calculation
21
Form 2-2: Introduction to mathematical calculation
Function Menu Setup
Type
Math
Description + , – , ×,÷
Information source
CH1 CH2
To set information source 1 as Ch1 channel waveform.
To set information source 1 as Ch2 channel waveform.
+ –
Operator
×
÷
Information CH1
source 2
CH2
Information source 1 + Information source 2 Information source 1 Information
source 2 Information source 1 × Information source 2 Information source 1 ÷
Information source 2
To set information source 2 as CH1 channel waveform. To set information source
2 as CH2 channel waveform.
Analysis of FFT frequency spectrum Time domain signal can be converted to frequency domain signal when using mathematical calculation of fast Fourier transtable (FFT). Following types of signals can be observed conveniently by using FFT Measure harmonic content and distortion within measurement system; Show noise characteristics in DC power; Analyze vibration;
22
Form 2-3: FFT menu Description
Function Setup
Description
Type
FFT
FFT mathematical calculation
Information CH1 To set CH1 and CH2 as
source
CH2 calculation waveform.
Hanning Set Hanning window
function
Hamming Set Hamming window
function
Window Blackman Set Blackman window function
Rectangle Set Rectangle window
function
Vertical unit
Vrms To set vertical unit to dBVrm Vrms or dBVrms.
FFT operation knack: Signals with DC component or deviation will bring in error or deviation of FFT waveform component. AC coupling mode can be selected to reduce DC component. Average obtainment mode of digital storage oscillograph can be set to reduce random noise and aliasing frequency component of repeated or single pulse event.
23
Selection of FFT window
Mathematical storage oscillograph will carry out FFT conversion for time record with limited length when
assuming continuous repetition of YT waveform. YT waveform will have the same amplitude in beginning
and ending locations for no interrupation. However, high-frequency transient interruption will be caused in
connector due to different waveform amplitudes for non-integral cycle of YT waveform. Such effect shall be
called as leakage in frequency domain. Therefore, value in forced beginning and ending locations is 0 by
multiplying original waveform by a window function to avoid leakage. Please refer to following form for
application of window function.
Form 2-4
FFT window
Characteristics
The most appropriate measurement content
Rectangle
The best frequency and the poorest
Transient or short impulse with basically equivalent
amplitude resolution to be basically
signal level; Sine with equivalent amplitude with similar frequency; Wide band random noise of wave
similar to circumstances without window; spectrum with slow variation;
With excellent frequency resolution and
Hanning
poor amplitude resolution by comparing Sine, cycle and narrow band random noise with rectangular window;
Hamming
Frequency resolution of Hamming window Transient or short impulse with large difference is better than that of Hanning window slightly. for signal level;
Blackman The best amplitude resolution and the poorest frequency resolution
Mainly used for single-frequency signal to search for higher order of harmonic wave.
24
Term explanation: FFT resolution: It is defined as factor of sampling and
calculation point. FFT resolution increases when sampling rate decreases for
fixed calculation point quantity. Nyquist frequency: Original waveform can not
be reset until using sampling rate of at least 2f for waveform with the
maximum frequency of “f”. It is also called as Nyquist guideline. “f” refers
to Nyquist frequency and “2f” refers to Nyquist sampling rate. Reference
waveform Please refer to waveform stored in nonvolatile memory (NVM) of
digital storage oscillograph or peripheral U disk with name of RefA and RefB
as follows. Please operate according to following steps to call out or close
reference waveform. 5. Please press REF menu key on the front panel for
5.7-screen oscillograph Please press [STORAGE]key and enter into the
25
second page to select call-out for 7-inch oscillograph (namely, PDO- 5000Sseries). 2. Please press [RefA] (RefA reference option) and select information source then select its position by rotating multi-purpose knob in upper part of front panel. 110 (or 120) can be selected for the position. Press “call-out” to call out waveform which is stored inthe position originally after selecting a certain position such as “1”. Please plug in U disk if waveform is stored in the U disk. There are 2 options for magnetic disk as follows: DSO/USB; Please press [F2] key and select USB. (The menu can not be activated until plugging in U disk.) Called out waveform is displayed on the screen for waveform called out and stored from U disk. Please press cancel key (F5) Return to previous menu after waveform call-out. 3. To press [RefB] (RefB reference option) and select
information source 2 which participates in calculation and method is the same to the secondstep. Please measure and observe related waveform by PDO-5000 series digital storage oscillograph during actual application. Please compare current waveform with reference waveform for analysis. Please press REF key to display reference waveform menu. Setup is also shown in following form. Form 2-5 Selection of storage position
Function menu Setup
Description
Storage position
1~20
1~20refer to positions of 20 groups of waveform respectively; There are 200 groups of waveform positions for storage to USB; (Storage position and quantity vary withmachine model.)
Magnetic disk
DSO To select internal storage position; USB To select external
storageposition
(until plugging inU disk);
Close Callout Cancel
To close called out waveform; To call out selected waveform; Return to previous menu;
It is allowed to select from 1 to 20 for internal
storage position. Please plug in U disk and press
F2 to select USB magnetic disk for external storage
instrument. Please refer to STORAGE menu for
waveform storage.
26
Setup of horizontal system Horizontal control knob Control knob on horizontal
surface can be used to alter horizontal scale (or time base) to trigger
horizontal position (or trigger position) shown in RAM. Vertical middle point
toward horizontal direction on the screen is time reference point for
waveform. Horizontal scale can be altered for waveform extension or shrinkage
by comparing with screen center. Horizontal position can also alter by
comparing with position of waveform trigger point. Horizontal position: To
adjust horizontal position of channel waveform (including mathematical
calculation). Resolution of such control key varies according to time base.
Horizontal scale: To adjust main time base, namely, s/div. Window width can
vary with delay scanning time base by rotating horizontal scale knob when
27
opening extension time base. Please refer to introduction to extension time base for details. (As shown in the following form), horizontal menu can be displayed for horizontal control key menu. Form 2-6
Function menu Setup
Description
1.To open main time base;
2.To close view window
Main time base —- extension according to main
time base after opening;
—-
View window extension
—-
— — To open extension time base;
Hold off
— — To regulate inhibition period;
Figure 2-11 Interface of horizontal system Mark Description: Mark represents position of current waveform view window in RAM Mark position of trigger point in RAM. Mark position of trigger point in current waveform view window. Display horizontal time base (or main time base), namely, s/div. Trigger horizontal distance of position by comparing with middle point of view window.
Term explanation: Y-T mode: Y axle refers to voltage and X axle refers to time
quantum under this mode. X-Y mode: X axle refers to CH1 voltage and Y axle
refers to CH2 voltage under this mode. Slow scanning mode: Instrument will
enter into slow scanning sampling mode when setting horizontal time base
control to 100ms/div or less. It is suggested to set channel coupling to DC
when observing low-frequency signal by applying for slow scanning mode. s-div:
Time base control can expand or compress waveform for horizontal scale (or
time base) unit if stopping waveform sampling (by using RUN/STOP key).
28
View window extension View window extension is to argument a section of
waveform to review image details conveniently. Setup of view window extension
can not be slower than that of main time base.
Horizontal extension of the waveform
Extension time base
Horizontally extended part of the waveform Figure 212 Screen display under
view window extension
29
(As shown in above figure), it can be divided into 2 display areas under extension time base. Original waveform is displayed for upper part. Selected area can move horizontally by rotating horizontal POSITION knob or increase or decrease by rotating horizontal SCALE knob. Waveform with horizontal extension in originally selected waveform area is shown in lower part. (As shown in above figure,) please pay attention that resolution increases for extension time base by comparing with main time base. Waveform displayed by the entire lower party corresponds to area selected for upper part so users can rotate horizontal SCALE knob to reduce selected area and increase extension time base, namely, horizontal extension multiple of waveform. The mode must be used for CH1 and CH2 at the same time. CH1 voltage will be displayed on X axle and CH2 voltage
will be displayed on Y axle after selecting X-Y display mode.
functions do not effect during X-Y display. Automatic testing mode Cursor measurement mode Reference or mathematical calculation waveform View window extension function Trigger control
Figure 213 Waveform display under X-Y mode
Note: Digital storage oscillograph can capture waveform by any sampling speed
under normal X-Y mode. Similarly, it is allowed to adjust sampling rate,
vertical gear of channels and time base gear under X-Y mode. Default sampling
rate is 100MS/s under X-Y mode. In general, sampling rate can reduce to gain
excellent Lissajou’s figure. Following
30
Setup of trigger system Trigger can determine data acquisition time and waveform display for digital storage oscillograph. It can convert instable display to meaningful waveform once correct setup. Digital storage oscillograph will collect sufficient data to map waveform in the left side of trigger point firstly. It will acquire data continuously and synchronously when waiting for trigger conditions. It also will acquire sufficient data continuously to map waveform in the right side of trigger point. Trigger control area of operation panel of digital storage oscillograph includes trigger level adjustment knob and trigger menu key of [TRIG MENU]; 5.7-inch oscillograph will set 50% of trigger level in vertical middle point of signal (and realize PDO-5000S by press [SET TO ZERO] key). It has to press forced trigger key of [FORCE].
Trigger level: Trigger level will set signal voltage which corresponds to trigger point. 50%: To set trigger level to vertical point of trigger signal amplitude. FORCE: To produce 1 trigger signal forcefully for mainly application for normal and single modes shown in trigger modes. TRIG MENU: To trigger and set menu key.
31
Trigger control Trigger mode: Edge, pulse width, video and ALT TRIG; Edge
trigger: Trigger when edge of trigger signal reaches a certain given level.
Pulse width trigger: Trigger when pulse width of trigger signal reaches
certain set trigger conditions Video trigger: Carry out field or line trigger
for standard video signals. (PDO-5025S seriesis no such function.) ALT TRIG:
It is appropriate to trigger signal without frequency relevance. Various kinds
of trigger menus are also shown as follows: Edge trigger: Edge trigger mode is
to trigger threshold value of input signal edge. That is, to trigger in
ascending/ descending edge of input signal when selecting “edge trigger”. Form
2-7
32
Function menu Type
Setup Edge
Description
Selection of information source
Gradient
Trigger mode
CH1 CH2 EXT EXT/5 Municipal power Alteration Ascend Descend Ascend/descend
Automatic Normal Single
Set CH1 as trigger signal of information source. et CH2 as trigger signal of information source. Set EXT TRIG INPUT channel as trigger signal of information source. Divide outer trigger source by 5 to expand outer trigger level scope. Set municipal power trigger ALT TRIG of CH1 and CH2 signals;. Set ascending edge trigger for signal; Set descending edge trigger for signal; Set ascending/descending edge trigger of signal; Set waveform acquisition without detecting trigger conditions; Set waveform acquisition after meeting trigger conditions; Set sampling of a waveform and shutdown after detecting once trigger;
Trigger coupling
AC DC High-frequency inhibition Low-frequency inhibition
Obstruct DC component of input signal; Pass through AC & DC components of
input signal; Inhibit high-frequency components of more than 80kHz for signal;
Inhibit low-frequency components of less than 80kHz for signal
33
Pulse width trigger Trigger time shall be determined for pulse width trigger according to pulse width. You can capture abnormal pulse by setting pulse width conditions. Form 2-8 (the first page)
Function menu Setup
Type
Selection of information source
Pulse width CH1 CH2 EXT EXT/5
Municipal power Alteration
Pulse width conditions
More than Less than Equal to
Pulse width setup
Next page 1/2
Description
To set CH1 as trigger signal of information source. To set CH2 as trigger
signal of information source. To set EXT TRIG INPUT channel as trigger signal
of information source. To divide outer trigger source by 5 to expand outer
trigger level scope. To set municipal power for trigger. ALT TRIG of CH1 and
CH2 signals To trigger when pulse width exceeds set value; To trigger when
pulse width is less than set value; To trigger when pulse width is equivale
To set pulse width of 20ns to 10s then regulate by multi-purpose knob in upper
part of front panel.
Enter into next page.
34
Form 2-9 (the second page)
Function menu
Setup
Type
Pulse width
Trigger polarity
Positive pulse width Negative pulse width
Trigger mode
Automatic Normal Single
Trigger coupling
Previous page 2/2
DC AC High-frequency inhibition
Description
To set positive pulse width as trigger signal; To set negative pulse width as
trigger signal; The system will acquire waveform data automatically and
display scanning base line on the screen without trigger signal input. It will
convert to trigger scanning automatically when producing trigger signal. To
produce trigger scanning if any trigger signal is produced. To produce once
trigger then stop during trigger signal input.
High-frequency inhibitionTo pass through AC & DC components of trigger signal.
To obstruct DC components of trigger signal. To obstruct pass through of high-
frequency components of signal and only low-frequency components can pass
through.
Return to previous page.
35
Video touch To trigger in field or row with NTSC or PAL standard video signals after selecting video trigger for PDO-5025S oscillograph shown in 5.7-inch or 7-inch screen. Trigger coupling is pre-set to DC. Please refer to following form for trigger menu. Form 2-10 Setup of video trigger
Function menu
Setup
Description
Type
Selection of information source
Pulse width CH1 CH2 EXT
EXT/5
Alteration
To set CH1 as trigger signal of information source. To set CH2 as trigger signal of information source. To set EXT TRIG INPUT channel as trigger signal of information source. To set 5 times of attenuation for outer trigger signal as trigger signal. ALT TRIG of CH1 and CH2 signals
Standard
PAL NTSC
Video signal to be appropriate to PAL system; Video signal to be appropriate to NTSC system;
Synchronization
All rows
To set synchronous trigger of video row.
Designated row To set synchronous trigger of designated video row for
adjustment by multifunctional knob in upper part of front panel.
Odd field
To set synchronous trigger for video odd field.
Even field
To set synchronous trigger for video even field.
36
Please refer to Figure 2-14 for screen display when selecting standard system of PAL and row synchronization as synchronization mode. Please refer to Figure 2-15 for field synchronization of synchronization mode.
Figure 214 Video trigger: Row synchronization
Figure 215 Video trigger: Field synchronization ALT TRIG mode ALT TRIGTrigger signal comes from 2 vertical channels during ALT TRIG. Such trigger mode can be adopted for synchronous observation of 2 signals without regard to signal frequency. Please refer to following figure for trigger alternation waveform display.
37
Setup of trigger coupling mode Please enter into trigger setup menu for setup of trigger coupling mode and gain the most sform synchronization. Please refer to following form for menu of trigger coupling mode.
Figure 216 Observation of 2 signals with
different frequency by.
Form 2-11 Trigger alternation menu
Function menu Setup Description
Type
Edge Set trigger mode to edge
Trigger source Gradient
ALT TRIG of CH1 and Alternation CH2 Ascending Set trigger gradient to
ascending edge.
Trigger mode
Automatic
Set trigger mode as automatic one
Trigger coupling
AC Set trigger coupling
modeto AC mode
38
Form 2-12
Function menu Type
Trigger source
Gradient Trigger mode
Setup —-
—- —- —-
Coupling
AC DC High-frequency inhibition
Low-frequencyinhibition
Description
Set pass-through obstruction of DC components. Set pass-through of all
components. Obstruct pass-through of high-frequency part of signal. Only
lowfrequency components can pass through. Obstruct pass-through of low-
frequency part of signal. Only highfrequency components can pass through.
39
Adjustment of hold off time Adjustment of hold off time can be used to observe complex waveform (such as pulse string series). Inhibition time refers to waiting period of digital storage oscillograph to start up trigger circuit again. Digital storage oscillograph will not trigger until inhibition time ends during inhibition period. For instance, as shown in Figure 2-17, the first pulse among a group of pulse series shall be triggered to set inhibition time to width of pulse string. Please refer to following form for hold off menu. Form 2-13
Function menu Main time base —-
Setup
Description
—-
1. To open main time base; 2. To close view window extension according to main time base after opening view window extension.
View window extension —- Hold off
—-
To open extension time base. To adjust inhibition time.
40
Trigger
Holdoff
Figure 2–17 Hold off used for synchronization of complex waveform Operation
Description
1. To select edge, trigger source and gradient shown in TRIG MENU according
to synchronization mode of normal signal then adjust trigger level for stable
waveform display as practicable as possible. 2. To press HORI MENU key to
display horizontal menu. 3. To adjust multi-purpose knob in upper part of
panel to alter inhibition time for stable waveform
41
display as practicable as possible. Term explanations 1. Trigger source: To gain trigger from many kinds of information sources, namely, input channel (CH1 & CH2), external trigger (EXT & EXT/5) and municipal power. Input channel: The most common trigger information source is (optional) input channel. Selected one is used as channel of trigger information source for normal operation regardless of input display. External trigger: Such trigger information source can be used for trigger on the third information channel during synchronous data acquisition on 2 channels. For instance, signals from external clock or circuit to measure can be used as trigger information source. EXT and EXT/5 trigger sources shall use external trigger signals
connected to EXT TRIG interface. Signal can be used for EXT and EXT can be used within signal trigger level scope of -3V to +3V. It is allowed to select EXT/5 for oscillograph of 5.7inch screen. Outer trigger shall be divided by 5 so that trigger scope can expand to -15V to +15V for trigger of digital storage oscillograph under large signal. Municipal power: Such trigger mode can be used to observe signal with regard to municipal power. For instance, it can provide relationship between lighting equipment and power supply equipment so as to gain stable synchronization. 2 . Trigger mode: It is to determine behavior mode of digital storage oscillograph without trigger event. This digital storage oscillograph can also provide 3 triggers, namely, automatic, normal and single trigger modes. Automatic trigger: The system will acquire
waveform data automatically so as to display scanning base line on the screen without trigger signal input. It will convert to trigger scanning automatically for signal synchronization when producing trigger signal. Note: Without trigger signal is allowed under ” automatic” mode when setting scanning waveform to 5-ms/div or less time base. Normal trigger: Digital storage oscillograph can not acquire waveform until meeting trigger conditions under ordinary trigger mode. Trigger scanning can be produced when producing trigger signal. Single trigger: User can press “operation” button once under single trigger mode for entry and trigger waiting for digital storage oscillograph. It will sample and display all acquired waveform then stop after digital storage oscillograph detects once trigger.
42
3.Trigger coupling: Trigger coupling can determine category of signal
component conveyed to trigger circuit. Coupling types include DC, AC,
lowfrequency inhibition and high-frequency inhibition. Pass through all
components of signals under DC mode. Obstruct DC components with signal
attenuation to be less than 400Hz under AC mode. Obstruct DC components with
attenuation of low-frequency component of less than 80kHz under “low-frequency
inhibition” mode. Attenuate high-frequency components of more than 80kHz under
“high-frequency inhibition” mode. 4. Pre-trigger/delay trigger: To trigger
data acquired in ahead/after event. Trigger position is often set in the
horizontal center on the screen. You can observe pre-trigger and delay
information of 5div (or 6div). You can also rotate horizontal position to
regulate horizontal displacement of waveform and
43
review more pre-trigger details. You can observe waveform in ahead of trigger by observing pretrigger data. For instance, burring reason can be shown by observation and analysis on pre-trigger data after capturing burring during circuit startup.
Setup of sampling system As shown in following figure, ACQUIRE in control area
is function key of sampling system.
Figure 218 Function key of sampling system User can use ACQUIRE key to pop out
sampling setup menu so as to adjust sampling mode by control button shown in
the menu.
44
Form 2-14 Sampling menu
Function menu Obtainment mode Average times
Setup
Description
Sampling
Activate ordinary sampling mode.
Peak-value detection Activate peak-value detection mode.
Average
Set average sampling mode and display average times.
2~256
Set average times for step by multiples of 2, namely, 2, 4, 8, 16, 32, 64, 128 and 256. As shown in Figure 2-18, average times can
alter by pressing left multi-purpose knob.
Sampling mode Fast acquisition
Real-time Equivalent
ON
Set sampling mode to real-time sampling mode. Set sampling mode to equivalent
sampling mode.
Acquire by higher screen refreshing rate to reflect dynamic effect of waveform
excellently. Close fast acquisition.
45
It has to observe waveform display variation by altering obtainment mode. Please refer to Figure 219 and 2-20 for sampled waveform display if signal contains large noise when adopting 32 times of average mode or without adopting average mode.
Figure 219 Waveform without adopting
Figure 2-20 Waveform by adopting 32 times of average mode Note
1. Please select real-time sampling mode for observation
of single signal. 2. Please select equivalent sampling mode for observation of
high-frequency periodic signal. 3. Please select peak-value detection mode to
avoid mixed signal envelope. Please select average sampling mode with step of
multiples of 2 for average value to reduce random noise of displayed signal.
Average times can be selected form 2 to 256.
46
Term explanation Real-time sampling: It refers to data required for once
acquisition. Equivalent sampling: It refers to repeated sampling mode. It can
facilitate careful observation of repeated periodic signals. It can be adopted
to gain horizontal resolution which is higher than real-time sampling
resolution greatly. Sampling mode: It can be classified into equivalent and
real-time modes. Peak-value detection mode: Digital storage oscillograph will
find out the maximum and minimum input signals within every sampling interval
for waveform display by using such values under such obtainment mode. It can
also obtain and display narrow impulse or they may be omitted for larger noise
under such mode. Average mode is as follows: Digital storage oscillograph may
obtain several waveform and calculate average value then display final
waveform under this obtainment mode. Such mode can be adopted to reduce random
noise.
47
Setup of display system As shown in following figure, DISPLAY in control area
is function key of display system
Figure 22 1 Function key of sampling system Please press DISPLAY button to pop
out setup menu shown in following figure. Display mode can be adjusted by
pressing menu control button.
48
Form 2-15 Display menu
Function menu Display type
Format
Setup
Vector point YT XY
Continuous
Close 1s 2s 5s Infinite
Descriptions
Only to display sampling points by wiring. Operation mode of digital storage
oscillograph X-Y display mode; X input for CH1 and Y input for CH2;
Real-time updating of screen waveform; To update after keeping waveform data
shown in the screen for 1s; To update after keeping waveform data shown in the
screen for 2s; To update after keeping waveform data shown in the screen for
5s; To always keep display of original waveform data shown in the screen until
this function is inactivated if new data is added into display continuously;
Waveform brightness 1% – 100% To set waveform brightness
Key points: Display type: To fill in blank between adjacent sampling points
during vector display. Only sampling points will be displayed for point
display.
49
Storage and callout As shown in following figure, STORAGE in MENU control area is function key of storage system.
Operation steps Please press STORAGE Enter into type menus. There are 3 types, namely, waveform, setup and bitmap. 2. Please refer to Figure 2-16 for selection of type waveform and entry into following waveform storage menu. Please call out REF (reference waveform shown in the second page) from Section 2 of the chapter after saving waveform. Form 2-16 Waveform storage menu (the first page)
Figure 222 Function key (storage) for sampling system STORAGE key can be used
to display storage setup menu. You can save oscillograph waveform or setup
status to internal storage area or U disk to call out saved waveform by RefA
(or RefB) or call out setup status by STORAGE key. Waveform display area of
oscillograph can be saved to DSO contents in U disk in a form of bitmap. Saved
bitmap can be read by PC computer.
50
Form 2-16
Function menu Type
Information source
Setup Waveform
CH1 CH2
Storage position
1~20
Save Next page 1/2
—- —-
Description Select waveform storage and callout menu Select waveform from CH1
channel. Select waveform from CH2 channel. 1~20 refer to positions to store 20
groups of waveform. There are 200 groups of waveform positions for storage to
USB. (Storage position and quantity vary with machine model.) Save waveform
Enter into next page.
Figure 2-23 Waveform storage
51
Figure 2-24 Waveform storage by U disk
Form 2-17 Storage menu (the second page)
Function menu Magnetic disk
Setup
DSO USB
Description To select internal RAM of digital storage oscillograph then select external U disk. (This function can not be used until plugging in U disk.)
Storage depth
Ordinary Long storage
To set storage depth to ordinary; (Callout only can be done in REF area when saving data to U disk.) To set storage depth to long storage. (Note: This function can not be activated until plugging in U disk. Callout can not be done until using loading function of computer communication software or waveform analysis load when saving data to U disk.)
Previous page 2/2
—-
Return to previous page.
52
2.As shown in Form 2-18, please select setup and enter into setup storage menu.
Figure 2-25 Storage setup Form 2-18 Setup of storage menu
Function menu Setup
Setup
Setup of (storage position)
120
Save Recall
53
Description To select panel setup menu. Operation setup to save 20 groups of
panels: To select by multipurpose knob in upper part of front panel. (Position
and quantity vary with machine model
Storage setup Callout setup
3. As shown in Form 2-19, please select bitmap for entry into bitmap storage menu. Remark: This function can not be called out until plugging in U disk. Form 2-19 Bitmap storage menu
Function menu Bitmap —-
Storage position
Setup 1200
Description Select bitmap menu
200 bitmap data can be saved by selecting multi-purpose knob in
upper part of front panel.
Save —-
Save bitmap data
Figure 2-26 Bitmap storage
54
Setup of auxiliary functions As shown in following figure, UTILITY in MENU control area is function key.
Function menu Setup Execution
Self-correction Cancel
Description
To carry out self-correction. To cancel self-correction and return to previous
page.
By detection
To set waveform Pass/Fail. Please refer (Only PDO-5000E and to Form 2-23 PDO-
5025S have this
function.)
Figure 227 Function key (and function) of sampling User can press UTILITY key to pop out setup menu of auxiliary system function. Form 2-20 (the first page)
Waveform recording
Language
Please refer Set waveform recording to Form 2-22 operation.
Multilanguage Select interface language.
55
Form 2-21-1 (the second page)
Function menu
Setup
Description
Fast correction
ON/OFF
Activate/inactivate fast correction function of channels within scope of 2mV/div to 10 mV/div. (Only PDO-5025S has this function.)
The first style
The secondstyle Interface style The third style Set 4 interface styles for digital storage oscillograph.
The fourth style
Gridding brightness 1% – 100%
Next page 2/3
—-
Form 2-21-2 (the third page)
Function menu System information
Frequency meter
The first page 3/3
Setup
Adjust gridding brightness of the screen by multi-purpose knob.
Enter into next page.
Description Display information of current oscillograph system.
Activate/inactivate function of frequency meter. Return to the first page.
56
Form 2-22 Waveform recording menu
Function menu
Setup
Information source
CH1 CH2 CH1+CH2
Operation
Save Recall Return
57
1 ~ 200 1 ~ 200
—-
Description Select CH1 as recording signal source. Select CH2 as recording
signal source. Select ct CH1+CH2 as recording signal source.
Record by pressing recording key for synchronous display of recorded quantity
in lower part of the screen.
1. Playback key; 2. Press the key for playback and display of current
playback screen No. in lower right corner of screen. Playback can be stopped
by rotating multi-purpose knob controller in upper part of panel. However,
waveform of a certain screen can be selected for repeated playback by
continuing to rotate it.
3.Pressto stop then press for playback of all items.
Stop recording Save newly recorded waveform for U disk plug-in. Storage
position can be selected by multi-purpose knob. Call out recorded waveform
from U disk then select recording position by multi-purpose knob.
Return to previous level of menu.
Form 2-23 By detection Function menu
Detection status
Setup Open Close
Information source
CH1 CH2 MATH RefA RefB
Output conditions
Pass Pass/Halt Fail/Halt Fail
Template setup
Refer to Form 2-24 for details
Return
Description
Open/close information of waveform pass/fail times
Select CH1 as detected signal source. Select CH2 as detected signal source.
Select MATH as detected signal source. Select RefA as detected signal source.
Select RefB as detected signal source. Output concurrency prompt when
detecting waveform pass; Output and pause when detecting waveform pass; Output
and pause when detecting waveform fail; Output concurrency when detecting
waveform fail;
Enter into template setup menu.
Return to previous menu.
58
Form 2-24 Template setup Function menu Setup
Template output
/
Description Set horizontal and vertical Pass/Fail tolerance scope according to adjustment
Horizontal Vertical Return
1-200Pixel Set horizontal tolerance scope
1-100Pixel Setvertical tolerance scope
Return to Pass detection
/
menu
Key note:Key note: Self-correction: Self-correction procedure is to correct measurement error caused by digital storage oscillograph due to ambient variation. You can operate the procedure according to demand.
Please open power of digital storage oscillograph then wait for instrument pre-heating for 30 minutes for more correct calibration. Please press UTILITY
key (as an auxiliary function) for operation
according to prompt shown in the screen. Language selection: There are many kinds of languages for PDO-5000 series digital storage oscillograph. Please press UTILITY menu button and select appropriate language.
59
Automatic measurement As shown in following figure, MEASURE is automatic
measurement function key. You can be familiar with powerful automatic
measurement function of PDO-5000 series digital storage oscillograph gradually
by following introduction.
Figure 2-28 Function key (automatic measurement) of sampling system
Application example Measurement menu of such series of oscillograph can be used to measure 28 kinds of waveform parameters. User can press MEASURE key Enter into parameter measurement display menu firstly. The menu includes 5 areas to display measurement value at the same time. They correspond to function keys of F1 to F5 respectively. Please press corresponding F keys Enter into measurement category selection menu if measurement category shall be selected for any area. Measurement category selection menu can be classified into voltage and time kinds Enter into voltage or time measurement category respectively. Please press F1 to F5 to select measurement category then return to parameter measurement display menu. In addition,user can
60
press F5 to select “All parameters” to display all voltage/time measurement
parameters. User can press F2 to select channel to measure. (Channel is
ineffective until opening.) User can press F1 Return to parameter measurement
display menu if not altering current measurement category. The first example:
Steps are as follows if displaying measured peak value of CH2 channel in F1
area. 1. To press F1 key Enter into measurement category selection menu; 2. To
press F2 to select CH2; 3. To press F3 to select voltage; 4. To press F5
(shown in the second page, 2/4) to see that F3 position is “peak value” 5. To
press F3 to select “peak value” and return to parameter measurement display
menu automatically; Peak value is displayed in F1 area on the homepage of
measurement menu.
61
The secondexample: Setup of delay measurement: Delay measurement function is
used for interval between ascending edges of 2 information sources, namely,
interval from ascending edge of the first cycle of a certain information
source to that of the first cycle of additional information source.
Measurement steps are as follows: 1. To refer to above example and select area
displayed by delay measurement value after entering into measurement menu
(shown in 3/3 page of time category); 2. To select reference information
source of CH1 then select delay information source of CH2;
3. To select reference information source of CH1
then select delay information source of CH2; 4. To press F5 to confirm that
delay measurement is displayed in selected area.
Automatic measurement of voltage parameters Voltage parameters which can be measured by PDO-5000 series oscillograph automatically are as follows: Peak value (Vpp): Voltage from the highest waveform point to the lowest waveform point; The maximum value (Vmax): Voltage from the highest waveform point to GND (ground); The minimum value (Vmin): Voltage from the lowest waveform point to GND (ground); Middle value (Vmid): Half of sum of waveform top and bottom voltage; Amplitude (Vamp): Voltage from waveform top to bottom; Middle value (Vmid): Half of sum of waveform top and bottom voltage; Top value (Vtop): Voltage from flat waveform top to GND (ground); Bottom value (Vbase): Voltage from waveform
bottom to GND (ground); Overshoot: Ratio of difference of the maximum waveform
value and top value and amplitude; Preshoot: Ratio of difference of the
minimum waveform value and bottom value and amplitude; Average value: Average
amplitude of signals within a cycle; Root-mean-square (Vrms): It is effective
value. Root-mean-square refers to DC voltage of converted energy within 1
cycle according to AC signal to correspond to equivalent energy.
62
Automatic measurement of time parameters PDO-5000 series oscillograph can
measure signal frequency, cycle, ascending time, descending time, positive
pulse width, negative pulse width, delay (of 9 combinations), positive duty
cycle, negative duty cycle and 10 time parameters automatically. Definitions
of such time parameters are as follows: RiseTime: It refers to period when
waveform amplitude increases from 10% to 90%; FallTime: It refers to period
when waveform amplitude decreases from 90% to 10%; Positive pulse width
(+Width): It refers to pulse width of positive pulse under 50% of amplitude;
Negative pulse width (-Width): It refers to pulse width of negative pulse
under 50% of amplitude; Delay (ascending edge): It refers to delay from
ascending edge to ascending edge; Delay (descending edge): It refers to delay
from
63
descending edge to descending edge; Positive duty cycle (+Duty): It refers to ratio of positive pulse width and cycle; Negative duty cycle (-Duty): It refers to ratio of negative pulse width and cycle; Measurement menu Operation Description: Please press MEASURE key so that the screen can display 5 measurement values. As shown in Form 2-25, user can press any key of F1 to F5 so that the screen enters into measurement selection menu.
Form 2-25 Function menu Return
Information source
Voltage kind Time kind All parameters
Setup
CH1 CH2
Description Return to parameter measurement display menu. Select channel of measurement parameters. Select channel of measurement parameters. Enter into voltage parameter menu. Enter into time parameter menu. Display/close all measurement parameters.
Of which: Menu included by “voltage kind” is shown in Form 2-26 to 2-29.
Form 2-26 Function/measurement item
Return
Preshoot
Description Return to menu shown in Form 2-25. Return to parameter measurement display menu and substitute original position parameter after selection.
Amplitude Overshoot
he same as above The same as above
Next page (1/4)
Page turning
64
Form 2-27 Function/measurement item Previous page Average value Peak value
Root-mean-square Next page (2/4)
Form 2-28 Function/measurement item Previous page Top value
Bottom value Middle value Next page (3/4)
65
Description Return to previous page. Return to parameter measurement display
menu and substitute original position parameter after selection. he same as
above he same as above Page turning
Description Return to previous page. Return to parameter measurement display
menu and substitute original position parameter after selection. The same as
above The same as above Page turning
Form 2-29
Function/measurement item Previous page
The maximum value
Description Return to previous page. Return to parameter measurement display menu and substitute original position parameter after selection.
The minimum value —-
Next page (4/4)
The same as above —-
Return to the first page (shown in Form 2-26)
Menu included by “time category” is shown in Form 2-30 to Form 2-32.
Form 2-30 Function/measurement item Return
Frequency
Cycle RiseTime Next page (1/3)
Description Return to menu shown in Form 2-25. Return to parameter measurement
display menu and substitute original position parameter after selection.
The same as above The same as above Page turning
66
Form 2-31 Function/measurement item Previous page Frequency Cycle RiseTime
Next page (2/3)
Form 2-32 Function/measurement item Previous page Delay Positive duty cycle
Negative duty cycle The first page (3/3)
67
Description Return to previous page. Return to parameter measurement display
menu and substitute original position parameter after selection. The same as
above The same as above Page turning
Description Return to previous page. Enter into delay option menu (shown in
Form 2-32a) after selection. Return to parameter measurement display menu and
substitute original position parameter after selection. The same as above
Return to the first page (shown in Form 2-30).
Form 2-32a
Function
Setup
Channel CH1/CH2 /MATH
Description Select measured channel.
CH1/CH2 Channel /MATH
—- —-
Enter
—- —-
Return to parameter measurement display menu and substitute original position parameter after selection.
Cursor measurement User can press CURSOR button to display measurement cursor and cursor menu then use multi-purpose knob controller to alter cursor position. As shown in following figure, CURSOR in MENU control area is cursor measurement function key.
Figure 2-29 Function key (or cursor) of sampling system
68
User can move cursor for measurement under CURSOR mode. There are 3 modes, namely, voltage, time and tracing. V can be measured by pressing SELECT and COARSE keys on the panel and multipurpose knob controller to adjust positions of 2 cursors respectively during voltage measurement. Similarly, it can be measured if selecting time. Cursor of digital storage oscillograph can also trace signal variation automatically under tracing mode with waveform display. Remark: For 5.7-inch oscillograph 1. Effect of SELECT key is to select the cursor. 2. Effect of COARSE key is to adjust movement speed of cursor. For 7-inch oscillograph: User can press multi-function knob key to select the cursor. Movement speed of cursor can also alter according to rotation speed.
- Voltage/time measurement mode: The first or
69
secondcursor will appear at the same time. Cursor position on the screen can be adjusted by multipurpose knob controller. User can press SELECT key (or multi-function knob key) to select cursor to adjust. Displayed reading is also time or voltage between 2 cursors. 2. Tracing mode: Horizontal and vertical cursors can intersect to become a crossing cursor. Crossing cursor can position on the waveform automatically to adjust horizontal position of crossing cursor on the waveform by rotating multi-purpose knob controller. Digital storage oscillograph can also display coordinate of cursor point at the same time. 3. Measurement value will be displayed in upper right corner of the screen automatically when opening cursor function.
Usage of operation key RUN/STOP key is in upper right corner of front panel of digital storage oscillograph. User can press the key so that green lamp is on to show operation status. Red lamp can be on after pressing the key to show stop status.
As shown in above figure, automatic setup is to simplify operation. Digital storage oscillograph can adjust vertical deflection coefficient and horizontal time base gear automatically for stable waveform display on the screen according to waveform amplitude and frequency when user presses AUTO key. System setup is shown in following form during automatic setup:
Figure 2-30 Operation/stop key
Automatic setup
70
Function Obtainment mode Display format Horizontal position
Second/grid Trigger coupling Hold off Trigger level Trigger mode
Trigger source
Trigger gradient Trigger type Vertical band width Volt/grid
Setup
Sampling Set to YT Automatic adjustment Adjustment according to signal
frequency AC The minimum value
Set to 50% Automatic Set to CH1; Set to CH2 when exerting on signal for it if
there is no signal for CH1; Ascending
Edge
All Adjustment according to signal amplitude
RUN/STOP: To acquire waveform or stop acquisition continuously; The key can be pressed for switchover of waveform sampling between operation and stop. User can press the key under operation status so that green lamp is on. “Auto” can be displayed in upper part of the screen. Old digital storage oscillograph can acquire waveform continuously while red lamp is on by pressing the key under stop status. Digital storage oscillograph can stop acquisition when displaying “Stop” in upper part of the screen.
71
Chapter 3: Application Example
2.Automatic measurement of signal voltage and time parameters
The first example: Measurement of simple signal It has to observe a unknown signal in the circuit for quick display and measurement of signal frequency and peak value. 1. Please operate according to following steps for quick signal display: 1. Please operate according to following steps for quick signal display: To set probe menu attenuation coefficient to 10 and set switch on the probe to 10. To connect CH1 probe to measured point of circuit. To press AUTO button. Digital storage oscillograph can set automatically for the best waveform display. You can also adjust vertical or horizontal gear on above basis so that
Digital storage oscillograph can measure most of display signals
automatically. Please operate according to following steps for measurement of
signal frequency and peak value. To press MEASURE key to display automatic
measurement menu; To press F1 to enter into measurement menu category for
selection; To press F3 to select voltage category; To press F5 to turn to 2/4
page then press F3 to select measurement type of peak value; To press F2 to
enter into measurement menu category for selection then press F4 to select
time category;
To press F2 to select measurement type of
frequency.
waveform display can meet related requirements.
72
Figure 3-1 Automatic measurement
73
The secondexample: Delay observation of sine wave signal by circuit The same as above example, please set attenuation coefficient of probe and digital storage oscillograph to 10*. Please connect CH1 channel of digital storage oscillograph to circuit signal input terminal and connect CH2 channel to output terminal. Operation steps: 1. Signal display of CH1 and CH2 channels: To press AUTO button. To continue to adjust horizontal or vertical gear so that waveform display can meet your test requirements. To press CH1 key to select CH1 then rotate vertical position knob to adjust vertical position of CH1 waveform. To press CH2 key to select CH2 then adjust vertical position of CH2 waveform to facilitate&
comparison without waveform overlapping for the first and second channels. 2. To measure delay of sine signal caused by circuit then observe waveform variation. Delay of automatic measurement channel To press MEASURE button to display automatic measurement menu; To press F1 key to enter into measurement menu category for selection; To press F4 key to enter into time measurement parameter list; To press F5 key twice to enter into 3/3 page; To press F2 key to select delay measurement; To press F1 key to select CH1 then press F2 key to select CH2 then press F5 as enter key You can see delay value from “CH1-CH2 delay” in F1 area. Observation of waveform variation (shown in following figure)
Figure 3-2: Waveform delay
74
The third example: Capture of single signal Advantages and characteristics of
digital storage oscillograph include convenient capture of aperiodic signals
such as pulse and burring, etc. It has to know this signal to set trigger
level and trigger edge to capture a signal signal. For instance: Trigger level
shall be set to about 2V and set trigger edge to ascending edge trigger if
pulse is a logic signal of TTL level. User can observe it by automatic or
ordinary trigger mode so as to determine trigger level and trigger edge for
unknown signals. Operation steps are as follows: 1. To set attenuation
coefficient of probe and CH1 channel like the previous example; 2. To set
triggers; To press MENU button in trigger control area to display trigger
setup menu. To set trigger type to edge, trigger source to CH1,
75
gradient to ascend, trigger mode of single mode and trigger coupling to AC by using menu operation keys of F1 to F5 respectively under this menu. To adjust horizontal time base and vertical gear to appropriate scope. To rotate TRIGGER LEVEL knob to adjust appropriate trigger level. To press RUN/STOP execution button to wait for signals which meet trigger conditions. Once sampling can be done for display on the screen if a certain signal reaches set trigger level. Occidental events can be captured easily by utilizing the function. For instance: Sudden burring with large amplitude: Please set trigger level to be higher than normal signal slightly then press RUN/STOP button for waiting. The machine will trigger automatically and record waveform before and after trigger during burring. Horizontal POSITION knob in horizontal
control area on the panel can be rotated to alter horizontal trigger position so as to gain negative delay trigger with different length to facilitate observation of waveform before burring.
The fourth example: To reduce random noise of the signal You can adjust setup of digital storage oscillograph for overlapping of random noise to tested signal so as to filter or reduce noise and avoid disturbance to ontology signal during measurement. (Please refer to following figure for waveform.)
Figure 3-3 Single signal
Figure 3-4 To reduce random noise of signals Operation steps are as follows:
- As shown in previous example, please set attenuation coefficient of probe
and CH1 channel.
76
2. To connect signals for stable waveform display on digital storage oscillograph. Please refer to previous example for operation. Please refer to previous chapter for adjustment of horizontal time base and vertical gear. 3. To improve trigger by setting trigger coupling. (1) To press MENU button in trigger area to display trigger setup menu. (2) To set trigger coupling to low-frequency inhibition or high-frequency inhibition. Lowfrequency inhibition is to set a high-pass filter to filter low-frequency signal component of less than 80kHz and allow pass through of high-frequency signal component. High- frequency inhibition is to set a low-pass filter to filter high-frequency signal component of more than 80kHz and allow pass through of low-frequency signal component. It can be set to inhibit low-frequency or high-frequency
noise respectively to gain stable trigger 4. To reduce display noise by setting sampling modes. To adopt average sampling mode to eliminate display of random noise to thinner waveform and facilitate observation & measurement if waveform is thickened by overlapping random noise to measured signal. Signal details can be observed conveniently by reduced random noise after gaining average value. Detailed operation is as follows: To press ACQUIRE button in panel menu area to display sampling setup menu. To press F1 key as menu operation key to set obtainment mode to average status. (As shown in following figure,) please adjust average times by pressing F2 key as menu operation key with step from 2 to 256 by multiples of 2 so that waveform display can meet observation & test equirements.
77
Figure 3-5 Inhibited signal noise To reduce display noise by reducing waveform brightness. Note: Average sampling mode can be used to slow down updating speed for waveform display as normal phenomenon.
The fifth example: Application cursor measurement This digital storage
oscillograph can measure 28 kinds of waveform parameters automatically. All
automatic measurement parameters can be measured by cursor. Waveform time and
voltage can be measured quickly by using cursor. To measure 1 staircase
voltage of staircase signal Please operate according to following steps to
measure 1 staircase voltage of staircase signal: 1. To press CURSOR key to
display cursor measurement menu. 2. To press F1 key as menu operation key to
set cursor type to voltage. 3. To rotate multi-purpose knob controller to
place the first cursor to 1 staircase of staircase signal. 4. To press SELECT
or PUSHSELECT to select the cursor then rotate multi-purpose knob controller
to place the secondcursor to another staircase of staircase signal.
78
V, namely, pressure difference can be shown in cursor menu automatically.
Please refer to following figure for details. Note: Please set cursor type to
time if onlying adopting the secondstep as above for time measurement by
cursor.
Figure 3-6 Voltage difference of signal by cursor measurement
The sixth example: Application of X-Y function To review phase location difference of signals of 2 channels. Example: Phase location variation caused by 1 circuit for tested signals. Please connect digital storage oscillograph to circuit to monitor input/output signal of circuits. Please operate according to following steps to review input/output terminal of circuits by way of X-Y coordinate diagram. 1. To set probe menu attenuation coefficient to 10 and set switch on the probe to 10. 2. To connect CH1 probe to network input then connect Ch2 probe to network output. 3. To press CH1 and CH2 menu keys and open 2 channels if they have not been displayed. 4. To press AUTO button. 5. To adjust vertical scale knob for approximate equivalent display amplitude of 2 loops of signals.
79
6. To press DISPLAY menu key to call out display control menu. 7. To press F2
to select X-Y. Digital storage oscillograph will display input/output
characteristics of circuit under Lissa Jous’s figure mode. 8. To adjust
vertical scale and vertical position knob to reach the best effect for
waveform. 9. (As shown in following figure), please observe and calculate
phase location difference by elliptic oscillograph figure method.
Signal must be in the horizontal center
sin=A/B or C/D.Of which: ” ” refers to phase
location difference angle between channels. Please
refer to above figure for definition of A, B, C and D.
Therefore, phase location difference angle is
calculated as follows: =±arcsin (A/B) or =±arcsin (C/D). Calculated phase location
difference angle shall be within Quadrant I or IV
(namely, (0~ 2 ) or ( 32~2 ) if main axle of ellipse
is within Quadrant I or III. Calculated value is within
(
2
~
)
or
(~
if
main
axle
of
ellipse
is
within
Quadrant II or IV. Additionally, frequency and phase
location relationship between 2 signals can be
calculated according to figure if frequency or phase
location difference of 2 measured signals is integral
multiple.
Figure 3-7
80
10.Form of X-Y phase location difference
Frequency ratio of signal
Phase location difference 0° 45° 90° 180° 270° 360°
1:1
The seventh example: Trigger of video signal Video trigger function can be used for all modes except PDO-5000S series. (Of which: PDO-5025S has video trigger function.) It has to adopt video trigger to gain stable video output signal display for observation of a certain video circuit. Trigger of video field Please operate according to following steps for trigger in video field.
- To press[ MENU ]button in trigger control area to display trigger menu. 2. To press[ F1] key as menu operation key to select video type. 3. To press[ F2] key as menu operation key to set trigger source to CH1. 4. To press[ F3] key as menu operation key to select video standard of PAL.
81
5. To press [F4] key as menu operation key to select synchronization of odd
field or even field. 6. To adjust horizontal time base by rotating horizontal
SCALE knob in horizontal control area so as to gain clear waveform display.
Figure 3-8 Trigger of video field
Trigger of video row Please operate according to following steps for trigger in video row: 1. To press [MENU] button in trigger control area to display trigger area. 2. To press[ F1] key as menu operation key to select video. 3. To press [F2] key as menu operation key to set trigger source to Ch1. 4. To press[F3] key as menu operation key to select video standard of PAL. 5. To press[ F4] key as menu operation key to select synchronization to designated row. 6. To rotate multi-purpose knob controller for trigger in any row. 7. To adjust horizontal time base by rotating horizontal SCALE knob in horizontal control area so as to gain clear waveform display.
82
Figure 3-9 Trigger of video row Pass/FailInterface Schematic
Rear panel of Pass / Fail output
83
The eighth example: Pass/Fail detection It has to check if input signal is within rule scope or not. Fail refers to out of scope and Pass refers to within the scope. Pass/Fail signal can be shown by output port of the rear panel. 1. To press[UTILITY] button then press[F2]Enter into Pass/Fail menu. 2. Setup of information source: Enter into Pass/Fail menu then press[F2] to select information source. 3. Template setup: To press[F4] Enter into template setup menu (Template). To press[F2] and[F3] shown in Template menu to set horizontal and vertical tolerance scope by multi-function knob, (namely, 1200Pixel toward horizontal direction and 1-100Pixel toward vertical direction.) Please press F4 finally Return to Pass/Fail menu. 4. Setup of output conditions: To press[F3] to set output conditions shown in Pass/Fail menu. 5. To begin detection: To press[F1] to open detection status shown in Pass/Fail menu.
The ninth example: Usage of updating procedure of U-disk Updating procedure of U-disk can be used more conveniently and flexibly. Please operate according to following steps if using this function: 1. To download procedure documents to update from Internet then save them in U-disk. (It is suggested to use U-disk suggested by us.) 2. To close oscillograph and insert U-disk then open oscillograph. 3. To turn to confirmation interface of whether to update it or not if there is only 1 procedure document in U-disk. Please press[F5] for exit or press F1 to exit procedure updating. Document selection interace will appear if there are at least 2 procedure documents in U-disk. Please press[F5] key for confirmation (or press[F1] to exit procedure updating). Please turn to confirmation interface of whether to update it or not
then press[F5] for updating). 4. To wait for updating for prompt of successful updating. Please power off oscillograph then open it to fulfill updating procedure. Note: 1. Please wait for updating for several minutes. 2. It is not allowed to power off oscillograph or plug out U-disk during updating to avoid updating failure or unforeseenable error. 3. Please power off oscillograph then open it for reupdating after updating failure.
84
The tenth example: Usage of storage function Storage functions of oscillograph include setup storage, waveform storage, bitmap storage and specific screen copy function of PDO-5000S series of oscillograph. 1. Setup of storage function: It has to select display channel, trigger channel, vertical and horizontal gears of oscillograph for setup. Please press[STORAGE] key to select torage type of “setup” by pressing[F1 ]key. As shown in following figure, it has to select position to save by multi-function button then press F3 for saving.
Figure 3-10 Saving setup interface It can be tested directly during next test
after saving
setup to reduce re-setup process. PDO-5000E and PDO-5000S series of
oscillograph can save 20 kinds of setup.
85
2. Waveform storage function: PDO-5000 series of oscillograph can provide waveform storage function. User can press[STORAGE]key and [F1] key to select storage type of “waveform”. User can press [F2]key to select channel to save waveform then press[F5]Enter into next page. Magnetic disk in F1 menu position can designate storage position and default storage position is digital storage oscillograph (DSO) without plugging in U-disk. Stored magnetic disk position can be [DSO] or [USB] after plugging in U-disk. Default storage depth shown by F2 is ordinary and saved waveform only can be displayed by oscillograph. Storage depth can be selected to long storage to save “.data” for exportation by waveform analysis software when plugging in U-disk and selecting magnetic disk position to [USB]. It is allowed to analyze content data of every point of saved waveform. User can also press F5 Return to previous menu then select storage position by multi-function button then press [F3] for
saving. Figure 3-11 Waveform storage Figure 3-12 Waveform storage by U-disk
86
As shown in following figure, saved waveform can be recalled by short-cut key of [REF] on the panel for PDO-5000E series of oscillograph. User can press [F3] shown in the second page of STORAGE main menu for recall for PDO-5000S series of oscillograph.
3. Bitmap storage function: As shown in following figure, user can press[ F1] menu key of STORAGE main menu to select storage type of “bitmap” after plugging in U-disk and successful connection of oscillograph. User can rotate multi-function knob to select saved bitmap No. Saved bitmap in U-disk with “.BMP” format can be read out by computer directly.
Figure 3-13 Vertical control panel of 5.7-inch screen
Figure 3-14 Bitmap storage
87
4. PrtSc copy & storage function There is 1 PrtSc key in upper left corner of
the screen for PDO-5000S series of oscillograph. User can press the key for
screen copy after plugging in Udisk. Displayed image can be saved in U-disk by
way of “.BMP” format.
Figure 3-15 Screen copy
Chapter 4: System Prompt and
Troubleshooting
Introduction to System Prompt Regulation to the extreme: It prompts that
multipurpose knob has been regulated to the greatest extent under current
status and continued adjustment is not allowed. The prompt will display when
regulating vertical deflection coefficient switch, time base switch, X
displacement, vertical displacement and trigger level to the terminal.
Successful connection of U-disk: The prompt will display on the screen after
connecting U-disk to digital storage oscillograph and correct connection.
Plug-out of U-disk: The prompt will display on the screen when plugging out
U-disk from digital storage oscillograph.Saving: The prompt will display on
the screen with schedule bar below it when saving waveform
88
Loading: The prompt will display on the screen with schedule bar below it when
calling out waveform. Troubleshooting 1. Please operate according to following
steps for black panel without display for digital storage oscillograph after
pressing power switch. To check if power plugs are connected with normal power
supply or not. To check if power switch is connected or not. To start up the
instrument again after above inspection. Please contact HTC for service if the
product still can not be used normally. 2. Please operate according to
following steps if there is no signal waveform in the image after signal
acquisition. To check if probe is connected to signal connection wire or not
normally. To check if signal connection wire is connected
89
to BNC (namely, channel connector) or not. To check if probe is connected to articles to measure or not normally. To check if signals are produced or not for articles to measure. (Channels with signal and those with any problem can be connected to find out reasons.) To acquire signals once again. 3. Measured voltage amplitude is 10 times larger or smaller than actual value. It has to check if channel attenuation coefficient is consistent with attenuation multiple of used probe or not. 4. With instable waveform display: To inspect setup of trigger source shown in trigger menu to check if it is consistent with input channel of actual signal or not. Inspection of trigger type: General signals shall be inspected by edge trigger mode while video trigger
mode shall be used for video signals. Waveform can not be displayed stably until setting correct trigger mode. Please strive to set coupling to high- frequency inhibition and low-frequency inhibition display to filter high/low- frequency noise triggered by interference. 5. To press RUN/STOP key without any display: To check if trigger mode of trigger menu is in normal or single gear or not. It has to check if trigger level is out of waveform scope or not. Trigger level shall be in the middle or trigger mode shall be set to AUTO gear if yes. To press AUTO key for automatic setup as above. 6. Display speed will reduce after opening average time sampling mode. In general, speed will reduce normally if average times exceed 32.
To reduce average times. 7. Staircase waveform display As normal phenomenon,
horizontal time base can increase under super-low horizontal time base gear so
as to increase horizontal resolution and improve display. Possible display
type is vector and sampling points can be connected for staircase waveform
display. Display type can be set to point display.
90
Chapter 5: Appendix
Appendix A: Technical Indicators Unless otherwise specified, all technical
dimensions are appropriate to probes by setting attenuation switch to 10* and
PDO-5000 series digital storage oscillograph. Digital storage oscillograph
must meet 2 conditions as follows to meet such dimension standards: Instrument
must operate under regulated operation temperature for more than 3 0 minutes
continuously. Be sure to open system function menu and execute “self-
correction” procedure if variation scope of operating temperature reaches or
exceeds 5. All dimensions shall be guaranteed except dimension with “typical”
pattern.
91
Technical Indicators
Sampling
Sampling rate
Sampling mode PDO-5000E PDO-5070S/PDO-50100S
Real-time 1GS/s 1GS/s
Equivalent 50GS/s 50GS/s
Average value
N can be selected from 2, 4, 8, 16, 32, 64, 128 and 256 after sampling all channels for N times at the same time.
Note: Real-time sampling rate is 250MS/s without equivalent sampling for , and PDO-5025S.
92
Input Input coupling
DC, AC, GND
Input impedance
PDO-5000E: 1±2% M for parallel connection to21±3pF PDO-5000S: 1±2% M for parallel connection to 24±3pF;
Probe attenuation coefficient The maximum input voltage (Typical) delay between channels
1×, 10×, 100×and 1000× 400V (DC + AC peak value and input impedance of 1 M) 150ps
93
Horizontal Waveform interpolation Horizontal resolution Record length Storage
depth
Scanning scope
Precision of sampling rate and delay time Measurement precision (full band
width) for interval (T)
Sin(x)/x PDO-500E:25points/div Sampling point of 2×512k 25k PDO-5000E 2ns/div-
50s/div (200MHz, 150MHz) PDO-5000S 2ns/div-50s/div ( PDO-5070S,PDO-50100S )
PDO-5000S 10ns/div-50s/div ( PDO-5025S )
±50ppm (for any interval of at least 1ms)
Single mode: ±(1 sampling interval+50ppm×reading+0.6ns) 16 average values: ±(
1 sampling interval+50ppm×reading+0.4ns)
94
Vertical
Analog digital converter (A/D)
8-bit resolution with synchronous sampling for 2 channels
Scope of deflection coefficient (V/div)
Displacement scope
PDO-5000E: 2mV/div~5V/div (in input BNC); PDO-5000S: 1 mV/div~20V/div (in input BNC) ±10div (Of which: ±5div for PDO-50200S)
(Typical) optional band width limitation 20MHz
Low-frequency response (AC coupling, -3dB)
10Hz (in BNC)
DC gain precision (by adopting sampling or sampling mode of average value)
DC measurement precision (by adopting average sampling mode)
PDO-5000E: ±4% if vertical sensitivity is 2mV/div or 5mV/div; ±3% if vertical sensitivity is 10mV/div to 5V/div; PDO-5000S: ±5% if vertical sensitivity is 1mV/div or 2mV/div; ±4% if vertical sensitivity is 5mV/div; ±3% if vertical sensitivity is 10mV/div to 20V/div; PDO-5000E: If vertical displacement is 0 and N16: ±(4%×reading+0.1 grid+1mV) by selecting 2mV/div or 5mV/div; ±(3%×reading+0.1 grid+1mV) by selecting 10mV/div to 5V/div; If vertical displacement is not 0 and N16:
95
DC measurement precision (by adopting average sampling mode)
Measurement precision of voltage difference (V) (by adopting sampling mode of
average value)
±( 3%×( reading+vertical displacement reading)+( 1%×vertical displacement
reading); To add 2mV when setting from 2mV/div to 200mV/div: To add set value
by 50mV when setting from 200mV/div to 5V/div; PDO-5000S: If vertical
displacement is 0 and N16: ±(5%×reading+0.1 grid+1mV) by selecting 1mV/div or
2mV/div; ±(4%×reading+0.1 grid+1mV) by selecting 5mV/div; ±(3%×reading+0.1
grid+1mV) by selecting 10mV/div to 20V/div; If vertical displacement is not 0
and N16: ± ( 3 % ×( reading+vertical displacement reading)+(1% ×vertical
displacement reading)+0.2div; To add 2mV when setting from 5mV/div to
200mV/div: To add set value by 50mV when setting from 200mV/div to 20V/div;
Voltage difference between any 2 points on the waveform after calculating
average value for at least 16 captured waveform under the same setup and
environment conditions: ±(3%×reading+0.05div)
96
band width series PDO-5000E
PDO-5000S
model PDO-50200S PDO-5025S PDO-5070S PDO-50100S
analog band width single band width RiseTime
200MHz
100MHz
1.8ns
25MHz
25MHz
14ns
70MHz 100MHz
70M H z 100M H z
5ns 3.5ns
97
Trigger Trigger sensitivity
Trigger level scope
(Typical) precsion of trigger level for signal of which RiseTime or
DescendTime is not less than 20ns
Pre-trigger ability
Inhibition scope
(Typical) level set to 50% Edge trigger
1div
Internal ±5div away from screen center
EXT EXT/5 *
Internal
±3V ±15V ±(0.3div×V/div) (within scope of ±4div from screen center)
EXT EXT/5 *
±(6% of set value+40mV) ±(6% of set value+200mV)
Normal mode/scanning mode, pre-trigger/delay trigger with adjustable pre- trigger depth
PDO-5000E: 100ns-1.5s PDO-5000S: 80ns-1.5s
Operation if frequency of input signal is not less than 50Hz
Edge type
Ascend, descend, ascend & descend
Pulse width trigger Trigger mode Pulse width scope
To be more than, less than or equivalent to positive/negative pulse width; 20ns-10s
98
Video trigger*
Trigger sensitivity (typical video trigger)
Signal system and row/ field frequency (video trigger type)
Internal EXT EXT/5*
Peak value of 2div 400mV 2V
To support standard NTSC and PAL; Scope of row quantity is 1-525 (NTSC) and 1-625 (PAL)
ALT TRIG CH1 trigger CH2 trigger
Edge, pulse width and video Edge, pulse width and video
Note*: PDO-5070S/50100S has no EXT/5 function and video trigger function.
99
Measurement
Cursor
Automatic measurement Mathematical operation
Manual mode Tracing mode
Voltage difference (V) between cursors; Time difference (T) between cursors;
Reciprocal of T ( 1 /T)
Voltage and time of waveform point
It is allowed to display cursor during automatic measurement.
Peak value, amplitude, the maximum value, the minimum value, top value, bottom value, middle value, average value, root-mean-square, overshoot, preshoot, frequency, cycle, RiseTime, DescendTime, positive pulse width, negative pulse width, positive duty cycle, negative duty cycle and delay
+, -,×, ÷
Stored waveform PDO-5000E/PDO-5000S: 20 groups of waveform and 20 kinds of setup
FFT Lissajous’s figure
Window Sampling point
Phase location difference
Hanning, Hamming, Blackman, Rectangle 1024 points
±3 degrees
100
Trigger frequency meter*
Reading resolution
6-bit
Trigger sensitivity
30Vrms
(Typical) precistion
±51ppm (+1 word)
General technical dimension
Display
PDO-5000E
Display type
Diagonal with diagonal of 145mm (5.7 inches)
Display resolution
320 (horizontal)×RGB×240 (vertical pixels)
Display color
Colorful
Waveform brightness* Adjustable (color)
(Typical) backlight 300nit strength
Category of display Multilingual selection language
PDO-5000S Diagonal with diagonal of 178mm (7 inches)
101
Probe compensator output (Typical) output voltage (Typical) frequency
About 3V if peak value is not less than 1M 1kHz
Interface function Standard configuration Optional fittings
PDO-5000E: 1 USB (D); 1 USB (H); PDO-5000S: 1 USB OTG
PDO-5000E series: LAN;
Power Power voltage 100-240VACRMS, 45-440Hz, CAT II
Power consumption To be less than 30VA
Fuse
F1.6AL 250V; Fuses of PDO-5000E series re on power board in the machine;
Fuses of PDO-5000S series are on power board in the machine.
Environment Temperature scope Cooling method Humidity scope
Altitude
Operation: 0~+40 Non-operation: -20~+60 Forced cooling by fan 35:90%RH ,
3540:60%RH
Operation elevation of less than 3,000m
Non-operation elevation of less than 15,000m
102
Machine dimension
Dimension Weight
Width Height Depth Not containing package Containing package
PDO-5000E
320mm 150mm 130mm 2.5kg 4.0kg
IP protection IP2 X
Adjustment interval Suggested calibration period is 1 year.
103
Appendix B: Fittings of PDO-5000 series digital storage oscillograph Standard
fittings: 2 inactive probes of 1.2m and 1:1 (10:1); Please
refer to manual of probe fittings of details to meet
EN61010-031:2008 standard. 150V CAT II grade if switch is in 1 position; 300V
CAT II grade if switch is in 10 position; 1 power wire to meet national
standards; 1 “Operation Manual”; 1 “Warranty Certificate”; Communication
control software of
PDO-5000 oscillograph; USB connecting wires:
PDO- 5000E: UT-D06
Please order all fittings (standard pieces and optional fittings) from local HTC dealer.
Optional fittings: PDO- 5000E series: LAN module: UT-M01
104
Appendix C: Maintenance and cleaning General maintenance It is not allowed to
store or place instruments in locations with direct sunshine for LCD for a
long time. Note: It is not allowed to touch instrument or probe by pressurized
spray, liquid or solvent to avoid damage. Cleaning: Instrument and probe can
be inspected constantly according to operation. Please clean outer surface of
instruments according to following steps: 1. Please remove floating dust out
of instrument and probe by soft cloth. Please do not scratch transparent LCD
protection screen during cleaning. 2. Please power off when cleaning the
instrument by wet soft cloth without dripping. It can be cleaned by soft
detergent or clean water. It is not allowed to use any corrosive chemical
detergent to avoid
105
damage to instrument or probe. Warning: Please confirm that instrument has been dried thoroughly before power re-supply to avoid electric short-circuit even personal injury due to moisture.
The information is subject to changes without prior notice if manual content
varies.
106
107
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>