FISHER PAYKEL 950N82J 950 SLE Ventilator Circuit Kit Instructions
- June 3, 2024
- Fisher & Paykel
Table of Contents
FISHER PAYKEL 950N82J 950 SLE Ventilator Circuit
INSTALLATION
PRECAUTION
F&P 950™ SLE Ventilator Circuit Kit
Indications for use
The breathing set is an accessory to the F&P 950 Respiratory Humidifier.
It is intended for delivery of heated humidified respiratory gases to
neonatal, infant and child patients, within the limits of its stated technical
specifications.
Contraindications
There are no contraindications for heated respiratory humidification.
Side effects
There are no known side effects of heated respiratory humidification.
Technical specifications
Compatible with the F&P 950 Respiratory Humidifier. Refer to humidifier, accessories, and patient interface for additional warnings, cautions, contraindications, side effects, and system information.
-
Tidal volume: <300 mL
-
Interface connections: ISO 5356-1 Conical Connectors
Breathing tube length: -
Inspiratory: 1.75 m
-
Expiratory: 1.60 m
Gas leakage (60 cmH₂0, BTPS): -
Dual limb: <30 mL/min
-
Resistance to flow:
Without SLE Adaptor (15 L/min, BTPS): -
Inspiratory: 1.05 ± 0.06 cmH₂O
Expiratory: 1.11 ± 0.09 cmH₂O
With SLE Adaptor (2.5 L/min, BTPS): -
Inspiratory: 2.70 ± 0.18 cmH₂O
-
Expiratory: 0.07 ± 0.03 cmH₂O
includes measurement uncertainty of 0.02 cmH2O Compliance (60 cmH₂0, BTPS): -
Dual limb: 1.01 ± 0.13 mL/cmH₂O
With accessories: -
Dual limb: 1.02 ± 0.13 mL/cmH₂O
includes measurement uncertainty of 0.08 mL/cmH₂O -
Tube minimum internal diameter: 11.2 mm
-
Maximum tube surface temperature: 44 °C (<25 cm from patient port)
-
Maximum operating pressure: 80 cmH₂O
-
Operating flow range (Ventilator): 0.5–40 L/min
-
Chamber maximum volume of water: 160 mL
-
Room temperature: 20–26 °C
-
Maximum incubator temperature: 37 °C
Symbol definitions
Follow instructions for use – safety.
Consult instructions for use. fphcare.com/950IFU
Single use.
Correct water level in water chamber.
Incorrect water level; do not use water chamber.
Component can be disposed of in setup.
Type BF applied part.
Ventilator inspiratory port (flow to patient).
Ventilator expiratory port (flow from patient).
LATEX This product is not made with
natural rubber latex.
Temperatures limitations.
Manufacturer.
Date of manufacture.
Date of expiration.
Warnings, cautions, and notes
WARNINGS
- Appropriate patient monitoring (e.g. oxygen saturation) must be used at all times. Failure to monitor the patient (e.g. in the event of an interruption to gas flow) may result in serious harm or death.
- Single use. Do not reuse this product. Reuse may result in infection of the airway.
- Do not use product near any ignition source. Exposure to oxygen increases the risk of fire that may cause serious injury or death.
- The use (or modification) of breathing circuit/chamber combinations not recommended by Fisher & Paykel Healthcare may result in poor humidification, ventilator malfunction, or serious harm.
Failure to comply with the following warnings may impair performance of the device or compromise safety (including potentially cause serious harm):
- This product is designed for the delivery of air, oxygen, and/or nitric oxide. It is not suitable for the delivery of flammable anesthetic gas mixes or Heliox gas.
- Do not crush, stretch, or milk the tubing.
- Do not clean or sterilize this product. Avoid contact with chemicals, cleaning agents, or hand sanitizers.
- Set appropriate ventilator or flow source alarms to monitor therapy delivery.
- Perform a pressure and leak test on the breathing system before connecting to a patient.
- Monitor circuit condensate to prevent occlusion or build-up of fluid. Drain as required.
- The humidifier should always be level and positioned lower than the patient.
- Check all connections are tight before use.
- Do not use the chamber if it has been visibly damaged or dropped.
- Do not use the chamber if any of the sensor covers are dislodged or missing from the chamber.
- Do not use the chamber if the water level rises above the maximum water level line.
- Do not fill the chamber with water in excess of 37 °C.
- Do not use medications containing Tyloxapol (such as Tacholiquin) as this may damage the tubing and lead to a loss of ventilation pressure.
- Do not use beyond the 14-day maximum duration of use.
CAUTIONS
- Do not cover the circuit with materials such as towels, pillows, or bed linen. Failure to comply may result in skin burn.
- Do not touch the hot surface of the chamber base. Failure to comply may result in a skin burn.
- Avoid leaving tubing in prolonged contact with patient’s skin. Failure to comply may result in a skin burn.
- Do not operate the chamber without water or gas flow. Failure to comply may result in low levels of humidity.
- he water source must be at least 50 cm higher than the chamber. Failure to comply may result in low levels of humidity.
NOTES
- Use USP Sterile Water for Irrigation, or equivalent. Adding other substances may have adverse effects.
- User may be exposed to breathing tract fluids during disposal. Breathing circuits must be treated as contaminated. Dispose of according to local, state or country specific environmental laws and regulations.
USA/Canada Tel: 1800 446 3908 or +1 949 453 4000
Fax: +1 949 453 4001
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>