Summit Appliance DL2B USB LED Digital Data Logger User Manual
- September 23, 2024
- SUMMIT APPLIANCE
Table of Contents
DL2B USB LED Digital Data Logger
“`html
Specifications:
-
Data logger with temperature display
-
Visual and audio alerts for high and low temperatures
-
Min/max feature to monitor and store highest and lowest
readings -
Glycol-filled bottle for temperature sensor protection
-
Low battery alert function
-
Temperature display in oC or oF
-
Measuring temperature range: -45 ~ 120 oC or -49 ~ 248 oF
-
Operating conditions: -10 ~ 60 oC or -50 ~ 140 oF, 20% to 90%
humidity -
User-defined logging interval
-
6.5 ft NTC probe-connecting cable
-
Rechargeable Li-ion battery for up to 8 hours of data recording
during power failure -
Powered by a 12VDC power adapter
-
Compatible with USB 3.0 Extension Cable
-
Mounting hole dimension: 71.5mm(W) x 133mm(L)
Product Usage Instructions:
Installing the Data Logger:
-
Install the backup battery by following the polarity
diagram. -
Connect the temperature sensor and power adapter plugs without
using force.
To Use:
Before use, remove the clear plastic protective film from the
screen (LCD). Place the temperature sensor in the desired location,
such as inside a refrigerator.
Parts and Controls/Features:
- USB port
- LCD display
- Buttons for operations
LCD Display Description:
-
MEM: Memory capacity; full when it reaches the 10th grid, can
hold data for a year. -
BAT: Battery capacity.
-
Current temperature value displayed.
-
REC/STOP: Press to stop or record data; press for 3 seconds to
erase min and max temperature history.
FAQ (Frequently Asked Questions):
Q: Can I change the temperature display between Celsius and
Fahrenheit?
A: Yes, you can select either oC or oF for temperature
display.
Q: How long does the battery last during a power failure
event?
A: The rechargeable Li-ion battery can record data for up to 8
hours during a power failure.
“`
TEMPERATURE DATA LOGGER (DL2B)
USER MANUAL
Features
Data logger simultaneously displays minimum, maximum and current
temperatures
The unit will provide a visual and audio alert when temperature rises above or falls
below the high and low set points.
The min/max feature is designed to monitor and store the highest and lowest
readings until the memory is cleared, or removal of battery.
The temperature sensor is enclosed in a glycol-filled bottle, protecting it from rapid
temperature changes when refrigerator/freezer door is opened.
Low battery alert function ( battery symbol flashes) User can select oC or oF temperature display Measuring temperature range -45 ~ 120 oC ( or -49 ~ 248 oF) Operating conditions: -10 ~ 60 oC ( or -50 ~ 140 oF) and 20% to 90%
non-condensing (relative humidity) Accuracy : ± 0.5 oC (-10 ~ 10 oC or 14 ~ 50 oF), in other range ± 1 oC ( or ± 2 oF)
1
User defined logging interval
6.5 ft (2 meters) NTC probe-connecting cable
Rechargeable Li-ion battery to record data up to 8 hours during a power- failure
event
Powered by a 12VDC power adapter
Compatible with a USB 3.0 Extension Cable for durability and easy data transfer
Large LED lit LCD screen Dimensions137mm(L)×76mm(W)×40mm(D)
Mounting hole dimension: 71.5mm(W) x 133mm(L)
Felix Storch, Inc. Bronx, NY 10474 Tel:
718-893-3900 Fax:
844-478-8799
SAVE THIS MANUAL FOR FUTURE REFERENCE
AccuCold, Div of Felix Storch, Inc. | R005-07102018
READ ALL INSTRUCTIONS BEFORE USE
Package Contents
Data logger Temperature sensor (NTC)
in a glycol-filled bottle
Instructions manual Rechargeable x2 AA
batteries (1.5Volts)
4 GB Memory stick [FAT 32] Power adapter Antistatic bag
NIST-traceable calibration certificate
Installing the Data logger
1. Install the backup battery Unscrew the battery compartment cover located
on the back of the unit and install the battery. Follow polarity (+/-) diagram
below. Replace the battery cover. The unit will beep and all segments of the
LCD will be activated.
2. Connect the temperature sensor and power adapter plugs Do not use force to connect the probe or the power adapter plugs. The power adapter plug is different from the probe plug.
Power
2
supply port
Temperature sensor port
DL2B side view
AccuCold, Div of Felix Storch, Inc. | R005-07102018
DL2B rear view
To Use
NOTE: Before use, remove and discard the clear plastic protective film from
the screen (LCD). Place the temperature sensor (in glycol bottle) in the
location to be monitored, such as inside a refrigerator or
freezer. The data logger may be placed on top of unit with LCD display easily
visible and the alarm audible. Data Logger displays internal temperature of
unit being monitored, as well as maximum and minimum temperatures reached. The
Data Logger’s maximum and minimum readings reflect the highest and lowest
temperatures since the unit was powered or since the MIN/MAX history was
cleared. If the temperature measurement rises above or falls below the set
temperature range, the alarm will sound. To silence the alarm, press any key
ONCE. Clear MIN/MAX history once the unit is stable.
Parts and Controls/Features
USB port
LCD
Buttons
LCD Display Description
Icon
3
AccuCold, Div of Felix Storch, Inc. | R005-07102018
Description 1 MEM is memory capacity.
When MEM reaches the 10th grid, MEM is full. MEM can hold data for one year.
2 BAT is battery capacity.
3 Current temperature value is displayed 4 When “REC” is displayed, it means
the logger is
logging temperature; When “STOP” is displayed, it means the logger is not
logging.
5 Maximum temperature value during a data logging
session is displayed
6 Minimum temperature value during a data logging
session is displayed
7 High temperature alarm status 8 Low temperature alarm status 9 °C: Celsius
temperature units;
°F: Fahrenheit temperature units
10
Time icon: HH:MM:SS (24-hour clock)
11
Date icon: YYYY-MM-DD
Buttons Description
REC/STOP Press REC/STOP to STOP or RECORD data.
Press for 3 seconds to ERASE the MIN and MAX temperature MAX/MIN
history.
DL
Copy the recorded data (CSV file) to USB
SET
Hold the SET button to cycle through configuration settings.
Up/Down keys to change settings. Press and hold either key to advance the values rapidly.
Default Data Logger Settings
Code
Function
Range
Default Setting
*Please enter the correct temperature units oF / oC
C1
High temp. alarm C2 ~ 100oC /212 oF
C2
Low temp. alarm
-45oC /-49 oF ~ C1
C3
Alarm hysteresis
0.120.0oC
0.236 oF
8.0 oC 2.0 oC 1.0 oC /2.0 oF
C4
Alarm delay
0090 min
0 min
C5
Start delay
0090 min
CF
Temperature Unit
oC =Celsius
oF =Fahrenheit
E5
Offset temperature
-2020oC
-3636 oF
0 min oC
0.0 oC/ oF
L1
Logging interval
00240 min
05 min
PAS
Password
00 ~99
50
4
Programming the Data Logger
Password Input
High Alarm Temperature Setting
From the main display screen: Hold the SET key for 3 seconds. Use the up and down arrows to adjust the password to the correct password. By default the correct password is 50. Press SET key ONCE to confirm the settings. By default, the high and low alarm settings are 8 oC and 2 oC respectively. To reset high alarm and low alarm temperature settings, follow instructions below. From the main display screen: Hold the SET key for 3 seconds. Enter the correct password then press the SET key ONCE to enter the HI Temp Alarm setting mode. Use the up and down arrows to adjust the temperature accordingly. Press SET key ONCE to confirm the settings.
AccuCold, Div of Felix Storch, Inc. | R005-07102018
Programming the Data Logger (continued)
Low Alarm Temperature Setting
Alarm Hysteresis
From the main display screen: Hold the SET key for 3 seconds. Enter the
correct password then press the SET key 2x to enter the LO Temp Alarm setting
mode. Use the up and down arrows to adjust the temperature accordingly. Press
SET key ONCE to confirm the settings.
Hysteresis is the tolerance band to prevent alarm chattering. For instance, if
the high alarm temperature is set at 8 oC with a hysteresis of 1 oC, once the
alarm is activated, it will not return to normal until the temperature goes
below 7 oC. By default, the alarm hysteresis is set at 1 oC. To reset, follow
instructions below.
When the temperature is higher than (Low Alarm Temperature setting + Alarm Hysteresis) it will exit low temperature alarm.
When the temperature is lower than (High Alarm Temperature setting Alarm Hysteresis), it will exit high temperature alarm.
When an alarm condition occurs, HI-ALARM and LO-ALARM icons will appear on the
display along with a beeping sound to alert user. The beeping sound will stay
ON until the unit gets back in range. Press any key ONCE to stop the beeping
sound. The high and low indicators will remain on display even when the unit
gets back in range. Press for 3 seconds to clear the HI and LO alarm icons.
Note- The HI and LO alarm icons will only clear when the unit is back in range.
5 Alarm Delay
Start Delay Temperature Unit
From the main display screen: Hold the SET key for 3 seconds. Enter the correct password then press the SET key 3x to enter the Alarm Hysteresis setting mode. Use the up and down arrows to adjust the temperature accordingly. Press SET key ONCE to confirm the settings. The alarm delay is used to avoid unnecessary alarms when the temperature exceeds the set high and low alarm limits. This feature will delay alarm activation by the amount of time entered. By default, the alarm delay is set at 0 minutes. To reset, follow instructions below. From the main display screen: Hold the SET key for 3 seconds. Enter the correct password then press the SET key 4x to enter the Alarm Delay setting mode. Use the up and down arrows to adjust the time accordingly. Press SET key ONCE to confirm the settings. From the main display screen: Hold the SET key for 3 seconds. Enter the correct password then press the SET key 5x to enter the Start Delay setting mode. Use the up and down arrows to adjust the time accordingly. Press SET key ONCE to confirm the settings. From the main display screen: Hold the SET key for 3 seconds. Enter the correct password then press the SET key 6x to enter the Temperature unit setting mode. Use the up and down arrows to adjust the temperature units accordingly. Press SET key ONCE to confirm the settings.
AccuCold, Div of Felix Storch, Inc. | R005-07102018
Programming the Data Logger (continued)
Offset temperature
Logging/Record Interval
The offset temperature feature is useful for customers who require a positive or negative temperature offset to be applied to the temperature sensor reading. By default, the offset temperature is preset to 0 oC. To change the setting, follow instructions below: From the main display screen: Hold the SET key for 3 seconds. Enter the correct password then press the SET key 7x to enter the Offset temperature setting mode. Use the up and down arrows to adjust the temperature accordingly. Press SET key ONCE to confirm the settings. This setting tells the logger how frequently to take and store readings. The unit has a logging interval of 10 s to 240 minutes. By default, the logging interval is preset to 5 minutes. To change the setting, follow instructions below: From the main display screen: Hold the SET key for 3 seconds. Enter the correct password then press the SET key 8x to enter the Record Interval setting mode. Use the up and down arrows to adjust the time accordingly. Press SET key ONCE to confirm the settings.
Date and Time Setting
Press MIN/MAX and SET keys simultaneously and hold for 3 seconds to enter the date and time setting mode. Use the up and down arrows to adjust the year accordingly. Press SET to confirm and move to the month setting mode. Repeat the same steps to set MONTH/DAY/HOUR /MINUTE & SECOND
Other Functions
CLEAR the high and low
alarm temperature
Press
for 3 seconds to clear visual alarm (Lo-Alarm and Hi-Alarm)
indicators.
indicators from the display.
Delete all data history
Press REC/STOP and DL keys simultaneously for 3 seconds to delete all data
record 6
history. DLT will display on the screen when the data is deleted successfully, and MEM capacity display will be empty.
Delete max and min temperature history
Press MIN/MAX key for 3 seconds to clear max and min temperature history. Clear will display on the screen if the data is deleted successfully.
Copy the recorded data in CSV to USB
Using USB 3.0 Extension Cable
First step: Insert the USB Flash Drive. USB will display on the screen when
the logger detects the flash drive. Second step: Press DL button for 3 seconds
to download the data. CPL will display on the screen when the data is
successfully transferred to the flash drive. Third Step: When CPL displays on
the screen, the flash drive can be removed. Always clear the MEM/Data
logger’s internal memory before taking new readings. Otherwise, it will take a
long time to transfer large sets of data. Connect the male end of the cable
to the USB port then connect the flash drive to the female end of the cable.
Male
Female
AccuCold, Div of Felix Storch, Inc. | R005-07102018
Please note:
When the MEM is full, the unit overwrites old data If the temperature sensor
is loose or not inserted, “NP” will be displayed and the NP alarm will
be activated. When PAS is 0 there is no password. User can enter parameter
setup directly. When the logging interval (LI) =0, the record interval is 10
seconds. To modify the factory settings: Press the SET key for 3 seconds to
enter parameter setup state.
After adjusting parameters, press SET key button again for 3 seconds. “COP”
will be displayed. The modified and stored set temperature and parameters will
be new default settings. To resume original factory settings, press DL and SET
keys simultaneously for 3 seconds, “888” will display when the parameters are
reset to factory settings. To resume customers default settings, press and
keys simultaneously for 3 seconds, “888” will display when the parameters are
reset to customers default settings.
CSV File
To download data, USB drive is ejected safely and connected to a computer.
Open file(s) in Microsoft Excel or any .CSV compatible program.
Data results will be displayed in tabular form as shown below:-
Date
Time
Temp
6/12/2018 16:33:27 24.9C
6/12/2018 16:32:27 24.9C
6/12/2018 16:31:27 24.9C
7
6/12/2018 16:30:27 24.9C
6/12/2018 16:29:27 24.9C
6/12/2018 16:28:27 24.9C
6/12/2018 16:27:19 24.9C
Hi Alarm
0 0 0 0 0 0 0
Lo Alarm
0 0 0 0 0 0 0
Hi Alarm Setting
30.0C 30.0C 30.0C 30.0C 30.0C 30.0C 30.0C
Lo Alarm Setting
-10.0C -10.0C -10.0C -10.0C -10.0C -10.0C -10.0C
Date
Time (24 hour clock)
Temperature (oC)
High Alarm & Low Alarm temperature status
0 = No alarm event 1= Alarm event
Low Alarm & High Alarm Temperature Setting in Degrees Celsius
AccuCold, Div of Felix Storch, Inc. | R005-07102018
Troubleshooting
Displays ,,NP Display screen not working “Low battery” indicator flashing
Logger is not logging
Logger is taking too long to copy data to a flash drive The date sequence of
the logged data is NOT accurate Recorded data is corrupted
The logger does not record data when the AC power is OFF
Recalibration
The temperature sensor is not installed correctly. Make sure the AC adapter and batteries are installed correctly.
Battery may need to be recharged.
Press the
key and make sure the REC symbol appears on the
display. The logger will stop logging if AC power is removed and rechargeable
battery is not connected or not charged.
The logger internal memory should be cleared
Reset the date and time on the logger
Make sure the unit is not installed in an area with strong electromagnetic
interference.
Is the rechargeable battery inserted correctly? Please note the negative and
positive poles of the battery during replacing a new battery.
The rechargeable battery has not been charged prior to the power failure. The
battery needs to be charged for a minimum of 2 days.
It is recommended that Felix Storch, Inc., temperature monitoring products be
calibrated every two years to ensure proper function and accurate
measurements; however, your quality system or regulatory body may require more
frequent calibrations. To schedule your recalibration, please contact our
ISO/IEC 17025:2005 calibration laboratory 8 accredited by Perry Johnson
Laboratory Accreditation, Inc., at calibration@summitappliance.com
Care & Maintenance
Do not disassemble the product, as product damage may result. Store the
product where it will not be exposed to direct sunlight, dust or high
humidity. Do not wash or expose the product to water or other liquids. Clean
the product by wiping with a soft, dry cloth. Never use volatile or abrasive
liquids or cleaners to clean the product. Do not drop the product or subject
it to sudden shock or impact. The sensor cable leads must be kept away from
main voltage wires in order to avoid high frequency
noise. Separate the power supply of the loads from the power supply of the
Logger. When installing the sensor, place it with the head upward and the wire
downward. The logger must not be installed in an area where water drops may be
present. The logger must not be installed in an area where corrosive materials
or a strong electromagnetic
interference may be present.
AccuCold, Div of Felix Storch, Inc. | R005-07102018
Battery Handling and Usage:
WARNING To reduce the risk of serious personal injury: Keep batteries away
from children. Only adults should handle batteries. Follow the battery
manufacturer’s safety and usage instructions. Never throw batteries into a
fire. Dispose of or recycle spent/discharged batteries in compliance with all
applicable laws.
CAUTION To reduce the risk of personal injury: Always use the size and type of
battery indicated. Insert the battery observing the proper polarity (+/-) as
indicated.
Customer Support
For technical support, please call
800-932-4267 (U.S. and Canada) or email
info@summitappliance.com
For calibration services, please email calibration@summitappliance.com
For more information, please call ACCUCOLD at
718-893-3900 (U.S. and Canada). See us
on the web at www.summitappliance.com
9 Limited Warranty
ACCUCOLD products have a limited warranty period of 1 year against defects in
materials and workmanship from the date of purchase. Accessory items and
sensors have a limited warranty of 3 months. Repair services have a limited
warranty period of 3 months against defects in materials and workmanship.
ACCUCOLD shall, at its option either repair or replace hardware products that
prove to be defective, if a notice to that effect is received within the
warranty period. ACCUCOLD makes no other warranties or representations of any
k ind whatsoever, expressed or implied, except that of title, and all implied
warranties including any warranty of merchantability and fitness for a
particular purpose are hereby disclaimed.
WARNING: This product can expose you to chemicals including Nickel (Metallic)
which is known to the State of California to cause cancer. For more
information go to www.P65Warnings.ca.gov
Note: Nickel is a component in all stainless steel and some other metal
components.
AccuCold, Div of Felix Storch, Inc. | R005-07102018
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>