Lifeloc Technology LT7 Breath Alyzer Instruction Manual

September 17, 2024
Lifeloc Technology

LT7 Breath Alyzer

“`html

Specifications

  • Product Name: Lifeloc LT7

  • Manufacturer: Lifeloc Technologies Inc.

  • Power Source: Rechargeable NiMH battery pack / 3 AA Alkaline
    batteries

  • Location of Manufacture: Wheat Ridge, Colorado

  • Quality Management System: ISO 9001

Product Information

The Lifeloc LT7 is an advanced Breath Alcohol Detector designed
for law enforcement, corrections professionals, workplace, and
school markets. It offers features such as calibration/accuracy
check lockouts and multiple test modes. Manufactured by Lifeloc
Technologies Inc. in Wheat Ridge, Colorado, the LT7 is known for
its product innovation, precision instrumentation, ease of use, and
customer care.

Product Usage Instructions

Preparation

Rechargeable Battery Pack Option: The LT7 comes
with an internal rechargeable NiMH battery pack. Charge the battery
pack using the LT7 charger and cable included in the unit for
approximately two hours before initial use.

AA Batteries Option: If using the AA batteries
option, load 3 AA Alkaline batteries into the unit before use.

Powering On and Off: Press and hold the Power
button located on the bottom of the unit until the 3 center buttons
light up blue. The LT7 will perform an automatic internal
diagnostic check and display the Blowing Man Icon when ready.

Breath Testing Modes

Testing Modes Explained: Understand the
different testing modes available on the LT7.

Conducting Tests:

  • Automatic Test: Follow on-screen instructions
    for conducting an automatic test.

  • Manual Override Test: Perform a manual
    override test as needed.

  • Passive Test: Conduct a passive test using the
    appropriate procedure.

Calibration Settings

Calibration Menu: Access the calibration
settings to ensure accuracy.

Accuracy Checks:

  • Wet Bath Simulator Setup: Perform accuracy
    checks using a wet bath simulator.

  • Dry Gas Setup: Alternatively, use dry gas for
    calibration checks.

User Settings

Lockouts Explained: Understand how lockouts
work and adjust settings as necessary.

Test Settings: Configure test settings based on
your requirements.

Display Settings

Data Entry:

  • Data Entry Explained: Learn about data entry
    fields and how to configure them.

  • Date/Time Security: Set up date/time security
    settings for accurate recordings.

On-Screen Messages

Messages Explained: Interpret and respond to
on-screen messages effectively.

Frequently Asked Questions (FAQ)

Q: How do I know when to recalibrate the Lifeloc LT7?

A: It is recommended to recalibrate the LT7 if the calibration
as found result is outside the accuracy range or if there are any
accuracy concerns during testing.

Q: Can I use the LT7 without a mouthpiece?

A: No, a mouthpiece is required for conducting breath tests with
the LT7 for accurate results.

“`

Lifeloc LT7
Operations Manual
Unlock the Power of Alcohol Testing

2

Table of Contents

Introduction Front View Features
Preparation Rechargeable Battery Pack Option AA Batteries Option Powering the LT7 On and Of Navigating the Menu Observing the Subject Attaching a Mouthpiece Removing the Mouthpiece
Breath Testing Modes Testing Modes Explained Conducting an Automatic Test Automatic Test Screen Displays Conducting a Manual Override Test Conducting a Passive Test Manual Override Passive Test Passive Test Screen Displays
Operation History
Calibration Settings Calibration Menu Accuracy Checks Wet Bath Simulator Setup Performing a Wet Bath Accuracy Check or Calibration Dry Gas Setup Performing a Dry Gas Accuracy Check or Calibration Calibration Calibration “As Found” Result outside Accuracy Range Cal Settings
User Settings Lockouts Explained Settings Test Settings

www.lifeloc.com | 303.431.9500

5 6
7 7 7 8 8 8 9
9 10 10 11 12 12 13
13
14 14 14 15 15 16 16 17 17
18 18 18 19

Display Settings Data Entry Data Entry Explained Confguring a Data Field Date/Time Security Security Settings Explained System System Settings
On-Screen Messages Messages Explained
Service Service Extended Service Plans Notices
General Usage Specifcations and Technical Data
Menu Chart Settings Chart
Warranty Lifeloc Factory Warranty

Table of Contents

3

20 20 20 21 21 21 22 22 22
23
24 24 24
25
26
27 29

www.lifeloc.com | 303.431.9500

4

Introduction

Congratulations on your purchase of a Lifeloc LT7 advanced Breath Alcohol Detector For over 30 years Lifeloc Technologies has been providing technologically advanced alcohol detection instruments and training to Law Enforcement and Corrections’ Professionals and internationally, workplace and school markets We are a leader in product innovation, precision instrumentation, ease of use and Five Star Customer Care
The LT7 is manufactured in Wheat Ridge, Colorado by Lifeloc Technologies Inc and produced under our ISO 9001 quality management system Lifeloc ofers premium quality products combined with exceptional service and support
The LT7 incorporates cutting edge features including: · Advanced digital calibration protocol · Wide operating temperature range 14° to 130° F ( 10° to 55°C) · Full color graphic display visible in daylight or dark · Easy intuitive menu system for operational efciency · Two levels of security for added fexibility · Up to 10,000 test memory complete with relevant data felds · Fast charge NiMH batteries or optional AA alkaline batteries · 360° Rotational Mouthpiece
The LT7 includes many other features including calibration/accuracy check lockouts and multiple test modes
In the following pages this manual will assist you in understanding all that the LT7 has to ofer and how to operate this cutting edge instrument

www.lifeloc.com | 303.431.9500

LT7 Front View

Test Number
Display Navigation Buttons

LT 7

Introduction

5

360° Rotational Mouthpiece
Time
Battery Indicator

Power Button www.lifeloc.com | 303.431.9500

Execute Button

6

Introduction

Features1
Large Color Display: 2 2 inch full color display readable in daylight or dark Digital Calibration: Complete record keeping of calibration events including as found and as left accuracy measurements2 The LT7 can be calibrated with wet bath simulators or dry gas cylinders Lockouts: User selectable settings for calibration and accuracy check reminders and lockouts to prevent using an out of compliance instrument L360 Mouthpiece: Mouthpiece rotates to the position you feel safest, right or lefthanded Security: Password protection for critical functions and settings Auto Test Mode: Easiest way to take a test Unit automatically samples at proper time of breath exhalation Manual Test Override: Enables the operator to control exact point of breath sample Passive Test Mode: An automatic sample of exhaled breath without a mouthpiece, used to check for the presence of alcohol in breath Passive Test Override: Used to detect alcohol in a cup or open container User Selectable Test Order: Choose between Auto Test or Passive Test upon start up Memory: Stores up to 10,000 tests in memory Power: Choose between a fast charge rechargeable battery pack or optional three AA alkaline batteries Data Entry: “Subject ID” and “Operator” data felds standard History: Keeps a log of all the tests, calibrations and accuracy checks in memory Can be viewed by date User Interface: Easy to navigate menu driven software design Language: Select between languages without reprograming frmware

1 Note: all features may not be available in your model 2 See Specifcations for more information
www.lifeloc.com | 303.431.9500

“AA” LR6 USB C
ALKALINE

Rechargeable Battery Pack Option
The LT7 comes with an internal rechargeable NiMH battery pack that should be charged before initial use When used with the LT7 charger and cable included in your unit the battery pack will fully charge in approximately two hours

Preparation

7

LT 7

AA Batteries Option
If you have the LT7 battery option, you will need to load 3 AA Alkaline batteries into the unit before use

­

Powering the LT7 On and Of
Press and hold the Power button located on the bottom of the front of the unit until the 3 center buttons light up blue The LT7 then performs an automatic internal diagnostic check and will display the Blowing Man Icon when ready
To turn the unit of, press and hold the Power button until you hear the tone The unit will shut down
Note: Momentarily pressing the Power button, when the unit is on, will return you to the main menu

­

LT 7

Power Button

www.lifeloc.com | 303.431.9500

8

Preparation

Navigating the Menu
The LT7 is designed with an intuitive menu system to make operation easy
LT 7
There are 4 Navigation buttons used to move throughout the menu system With these you can move through all the screens and internal keyboards
There is one central Execute button to select and enable features and functions
Selecting the menu icon takes you to the “Menu” screen From here you can choose which process you need

Observing the Subject
The LT7 delivers a highly accurate reading of breath alcohol acquired by sampling deep lung air (alveolar) Readings can however be afected by residual mouth alcohol
To prevent mouth alcohol from afecting a test, make certain that the subject is not allowed to put anything in their mouth for 15 minutes prior to taking a test This will ensure all residual alcohol has completely dissipated and test results will be valid

Attaching a Mouthpiece
Remove the mouthpiece from its wrapper, making sure not to touch the end which the subject will blow into
Attach the mouthpiece to the port on the top of the LT7

Press the mouthpiece onto the port to snap in place Ensure it is securely attached
www.lifeloc.com | 303.431.9500

Breath Testing Modes

9

Removing the Mouthpiece
Eject the mouthpiece by pressing up with your fnger underneath the tab

Breath Testing Modes Explained
The LT7 has 4 test functions
Auto Test is the typical way to take a deep lung breath test The LT7 monitors the subject’s breath and automatically takes the sample near the end of the breath fow
Manual Override Test is used only when the subject is unable to provide a sufcient breath sample for the automatic test
Passive Test is a quick screen, without a mouthpiece, to detect alcohol but is not designed to quantify the results Passive results are reported as “POS” if alcohol is detected, “NEG” if alcohol is not detected In this mode, a passive cup may be used
Manual Override Passive Test is used to determine if alcohol is in liquid in a cup or container

www.lifeloc.com | 303.431.9500

10

Breath Testing Modes

Conducting an Automatic Test
Turn the LT7 on
Verify the display reads “AUTO TEST” and you see the Blowing Man icon
Attach the mouthpiece to the top of the unit and rotate to the position desired
Instruct the subject to blow into the mouthpiece frmly and steadily for as long as they can (But not necessarily as hard as they can ) The unit calculates volume and will give an error message if the subject cannot reach 1 5 liters of breath
Read the result
After taking a test, the LT7 will display the results in large numbers on the display
The result is stored in memory and available for viewing or downloading at a later date
Momentarily press the Power button to return to the test mode

Automatic Test Screen Displays
Blowing Man Screen Breath Flow Screen

Alcohol Curve Screen

Result Screen

The LT7 is ready to take an AUTO TEST

As the subject blows into the mouthpiece, the LT7 will show a graph of the breath fow on the display

If the LT7 detects alcohol, the alcohol level is graphed
and will be displayed before the result

After the alcohol is graphed, the test result is displayed The result will remain on the screen until the Power button is pressed or until shut of
time is reached

www.lifeloc.com | 303.431.9500

Breath Testing Modes

11

Conducting a Manual Override Test
Note: This feature allows the completion of a test in the occasional instance when the subject may have diminished lung capacity and cannot activate the Auto Test
Turn the LT7 on LT 7
Attach a mouthpiece to the top of the unit and verify the display reads “AUTO TEST” with the Blowing Man icon
Instruct the subject to blow into the mouthpiece frmly and steadily for as long as they can
While they are blowing and when they are near the end of their breath, press the Execute button to take the sample
Read the result
Note: Manual Override testing is pressure activated If the subject does not blow air into the mouthpiece, the test cannot be conducted

www.lifeloc.com | 303.431.9500

12

Breath Testing Modes

Conducting a Passive Test
The LT7 monitors pressure at the sample port and automatically takes a passive test when it senses the subject’s breath fow
Passive testing can be done either with or without a cup, which makes passive testing on multiple subjects easier
Turn the LT7 on
From the Blowing Man icon, press the right arrow key to enter Passive Test mode
Verify the display shows “Passive Test” and there is no Blowing Man icon
Hold the LT7 sample port on top of the unit about between 1 6 inches from the subject’s mouth
Have the subject blow toward the port
The LT7 will display a graph of the subject’s breath fow and when high enough, take the sample
Read the result The unit will only display “POS” or “NEG” It will not display the actual numerical result

Manual Override Passive Test
When using the LT7 to detect alcohol in an open container, position the unit over the end of the container and press the Execute button
Read the result The unit will only display “POS” or “NEG” It will not display an actual numerical result

1″­ 6″
1″­ 6″

LT 7

www.lifeloc.com | 303.431.9500

Passive Test Screen Displays
Passive Test Screen Breath Flow Screen

Alcohol Curve Screen

Operation

13

Results Screen

When taking a passive test, the Blowing Man icon is
not displayed and Passive Test is displayed

As the subject blows into the mouthpiece, the LT7 will show a graph of the breath fow on the display

If the LT7 detects alcohol, the alcohol level is graphed
and will be displayed before the result

The unit will only display “POS” or “NEG” It will not
display the actual numerical result

History

The History menu is where all tests, calibrations and accuracy checks are stored The LT7 can store up to 10,000 tests in memory Any test, calibration or accuracy check in memory can be viewed or downloaded
Results sorted by date Select a date to see all the tests from that date in test numerical order All View any test, calibration or accuracy check in memory
Tests View any test in memory, sorted by date
Calibrations View any calibration in memory, sorted by date
Accuracy Checks View any accuracy check in memory, sorted by date
Clear Test Memory Deletes test results in memory, does not delete accuracy checks or calibrations stored in memory
www.lifeloc.com | 303.431.9500

14

Calibration Settings

Calibration Menu

Use this menu to choose type of calibration, concentration of reference standards, and to calibrate or perform accuracy checks on your LT7 The LT7 can be calibrated or have its accuracy checked using either a wet bath simulator or an Ethanol/Nitrogen cylinder (Dry Gas Tank )
Accuracy Checks
An Accuracy Check is a test on a known standard to verify proper function of the LT7 Accuracy checks are used to verify the LT7 is calibrated correctly and is within the acceptable accuracy range Accuracy Checks are also referred to as “Verifcations” or “Cal Checks” See Specifcations for recommendations regarding regularity of accuracy checks
Wet Bath Simulator Set-Up
Pour a bottle of certifed alcohol solution into the simulator jar and hand tighten lid Connect 11 12″ long tube from input port to output port so no alcohol escapes while the simulator heats up
LT 7
Plug in the simulator and turn it on The simulator automatically heats the solution to 34° C (93 2° F) in about 5 to 10 minutes Proper operating temperature is important for accuracy so be certain to check temperature before proceeding
www.lifeloc.com | 303.431.9500

Calibration Settings

15

Performing a Wet Bath Accuracy Check or Calibration

From the LT7 menu choose either Accuracy Check or Calibration and verify your standard concentration
When you see this image on the LT7 connect the tube from the simulator input to a pump or a calibration mouthpiece, then use a calibration adapter to connect the simulator output to the LT7 mouthpiece
When you see this image on the LT7, turn on the pump or blow into the simulator at a rate that creates about 0 5 inches (12 mm) bubbles on the surface of the solution and press the Execute button to begin the LT7 sample countdown
After 3 seconds the LT7 will take the sample, after 6 seconds you can turn of the pump or stop blowing
Remember when using a wet bath you do not adjust for altitude

Dry Gas Tank Set-Up
Attach the regulator to the dry gas tank Attach the mouthpiece adapter to the output port on the regulator Securely ft the mouthpiece to the mouthpiece adapter

LX9

500 35
00 PSI BAR

1000 70

www.lifeloc.com | 303.431.9500

16

Calibration Settings

Performing a Dry Gas Accuracy Check or Calibration

From the LT7 menu choose either Accuracy Check or Calibration and verify your standard concentration
When you see this image, connect a mouthpiece from the LT7 to the Calibration Adapter on the regulatory and press the Execute button
When you see this image start the gas fow by pressing the button on the dry gas regulator, then press the Execute button to start the LT7 sample countdown
The LT7 will sample after 3 seconds and you can shut of the gas after 6 seconds
Remember, with the LT7 you do not adjust for altitude when using dry gas standards
Calibration
A calibration is a test on a known standard to verify accuracy of the LT7 The calibration process compares the LT7 accuracy against the standard and if needed adjusts the internal measurement to correct the instrument accuracy
The initial calibration test records the condition of the instrument when calibration is started commonly referred to as As Found If an adjustment is made, then the following accuracy check will record the As Left condition
Calibrations are typically performed by trained individuals and to calibrate the LT7 internal temperature must be between 68° and 95° F (20° 35° C)
See Specifcations for recommendations regarding regularity of calibrations
Note: The LT7 has an internal barometric pressure sensor to simplify the use of dry gas when performing a calibration Do not adjust the values of dry gas standards for altitude as this is done automatically by the unit
www.lifeloc.com | 303.431.9500

Calibration Settings

17

Calibration “As Found” Result outside Accuracy Range
The results will show on the display with a red X When ready to proceed with adjustment, choose Execute The LT7 will adjust based upon the initial calibration test A 2 minute timer will start to ensure the instrument is ready before proceeding
The LT7 will then display instructions to connect it to a cylinder or simulator Choose Execute and start your gas fow A timer will display showing 6 seconds that calibration gas should fow The LT7 will take the sample at the 3 second mark
This accuracy check will record the condition of the instrument As Left
Note: If the As Left accuracy check is outside of the specifed accuracy range of the instrument, it will proceed with adjustment again To cancel this process, momentarily press the Power button but understand that the instrument is not yet properly calibrated

Cal Settings
Calibration Type selects between wet bath simulators “Wet” and dry gas ethanol in nitrogen tanks “Dry”
Cal Standard selects between the 4 default standard concentrations or allows selection of a custom standard value
Note: The LT7 has an internal barometric pressure sensor to simplify the use of dry gas when performing a calibration or accuracy check Do not adjust the values of dry gas standards for altitude as this is done automatically by the unit

www.lifeloc.com | 303.431.9500

18

User Settings

Lockouts Explained
Lockouts are powerful features to ensure conformity to your program standards Lockouts are set separately for calibrations and accuracy checks and include the following options:

Lockout Enable/Disable lockout feature

Type

Choose to lock testing so the device cannot be used or just warning allowing usage

Time

Use the Execute button to choose between days, number of tests or both days and tests The right Navigation button will allow you to set time (days) or number of tests

Reminder

Used to set up a reminder that your unit will need a calibration or an accuracy check in a selected time frame This will give the operator notifcation in a selectable time or test number range that the LT7 will need a calibration or accuracy check

Changes to calibrations types, standards and lockouts are retained in memory and do not have to be changed unless the method used to perform calibrations or accuracy checks changes

Settings
The settings menu contains the functional confgurations of the LT7
Test Settings

Mode Selects test order Pressing Execute will change between test orders www.lifeloc.com | 303.431.9500

User Settings

19

Test Order 1 (Default)

Test Order 2

Auto Test

Auto Test Passive Test

Note: By selecting test order 1 you will disable passive testing

Test Order 3 Passive Test
Auto Test

Manual Sample Enables Manual Override of Automatic and Passive Tests Press Execute to turn manual sampling on or of

Results Test results for Auto and Manual Override tests can be displayed in a “NUMERIC” or “PASS/ WARN/FAIL” (PWF) format Levels will only be available when PWF is selected

Numeric results are in a 3 digit format

PWF results display as either PASS, WARN or FAIL Example, if “Pass” is set at 040 and “Fail” is set at 080 (the default settings on your LT7), then:

Test results from 000 040 will display PASS Test results from 041 079 will display WARN Test results from 080 and above will display FAIL Pass/Fail Level Used to select the levels for “Pass” and “Fail” If the “Pass” and “Fail” level are next to each other then you will not see “Warn”

Note: you cannot set the “Pass” level higher than the “Fail” level

www.lifeloc.com | 303.431.9500

20 User Settings
Display Settings
Auto Shut Of sets the time the unit will stay on without a keypress Press Execute to select time Brightness sets the level of display brightness Choose from 0% to 100% or select “Auto” for automatic adjusting Note: higher backlight levels will consume more battery power Language (Optional) if factory enabled allows selection of language for the operation of the LT7
Data Entry
Data Entry Explained
The LT7 contains 2 data felds set up at the factory as Subject ID and Operator Each data feld can hold up to 30 alpha numeric characters and can be set up as either “Fixed” with static data that does not change per test or “User Input” (dynamic) that will be entered for each test Subject ID ­ Used to identify a unique test subject If enabled the LT7 will prompt you to enter name or number each time you prepare to administer a test The “Subject ID” will then appear with that test retained with the test record Operator ­ Used to identify an operator If enabled the LT7 will allow you to enter the operator’s name or number and will print that data with each test and retain it with each test record
www.lifeloc.com | 303.431.9500

User Settings

21

Confguring a Data Field
Data Entry Displays all data felds enabled in the LT7 Select a data feld to modify settings for that feld Data Field Displays settings for the selected data feld Enabled Turns a data feld on or of Label Changes the name used to identify the data feld Type Choose between “User Input” or “Fixed” Data If “Type” is “Fixed” data is entered here

Date/Time

Date Sets the date for the LT7 clock Date Format Sets the date format as either “mmddyyyy” or “ddmmyyyy” Time Sets the time of day Time Format Selects between 12 hour (AM/PM) and 24 hour time display Daylight Savings Moves the clock ahead or back one hour
Security
www.lifeloc.com | 303.431.9500

22

User Settings

Security Settings Explained
The LT7 comes with the ability to use password security to protect access to calibration, calibration settings and user settings
If you enable password protection the factory default password will be 1 It is highly recommended that you do not use the factory default password
Without a password users are still able to conduct Automatic, Manual Override, Passive and Open Container tests; check battery status and temperature
Note: Record and store your password in a safe place. Lifeloc does not have access to your password. If password is lost, the only way to reset the unit is to call Lifeloc Tech Support.
System

System Settings
Temperature ­ Choose between Celsius and Fahrenheit Battery NiMH rechargeable or Alkaline, these settings are determined at the factory Button Sounds Enables audible feedback of button presses
www.lifeloc.com | 303.431.9500

Messages Explained
Insufcient Breath >0.600 Cal/Accuracy Check Expired Cal/Accuracy Check Needed XX days Invalid Test Low Battery Temperature

On-Screen Messages

23

Subject breath fow ended early Instruct the subject to try again or use manual override
BAC is above the instrument’s maximum measurement range Subject may require medical attention
Perform a Calibration or Accuracy Check
The LT7 is within XX days of Calibration or Accuracy Check lockout time
No alcohol was detected during a Calibration or Accuracy Check
Battery voltage is too low to take a test, recharge or replace batteries
When calibrating or taking a test while the LT7 is outside of temperature limits Calibration range = 68° ­ 95° F (20° ­ 35° C) Testing range is 14° to 130° F ( 10° to 55°C)

www.lifeloc.com | 303.431.9500

24 Service
Service
If your LT7 should require repairs or maintenance, Lifeloc is here for you Just an email or phone call will put you in contact with our technical support personnel To reach our service department call 303 431 9500 or email techsupport@lifeloc com
Note: To prevent delays in service or additional charges make certain you receive an authorization number before sending your equipment to Lifeloc
International Service Contact your local factory authorized service center or email Lifeloc Technologies at techsupport@lifeloc com
Extended Service Plans
In the U S , Extended Service Plans are available for your LT7 These provide complete coverage for an additional year at a reasonable cost and include free factory diagnostic checks Call Lifeloc for details
Notices
The LT7 is a professional device designed to be used by trained operators in conjunction with a specifed, periodic maintenance and calibration/calibration check regimen Use by untrained operators or without periodic calibration or accuracy checks may result in invalid results or incorrect interpretation of results. The LT7 is not to be used by children under 12 years of age.
DO NOT DRINK AND DRIVE Lifeloc strongly recommends that no vehicle or machinery be operated after alcohol consumption Even small quantities of alcohol can result in driving impairment
The LT7 is not waterproof and should not be immersed in or exposed to excessive water The LT7 is not suitable for use in a potentially explosive environment The LT7 cannot be used inside an oxygen tent
If the LT7 is equipped with Alkaline batteries and will not be used for more than 6 months remove batteries to avoid damage caused by leaking battery acid
Disposal of Instrument
At the end of the instrument’s service life: · Do not dispose of the LT7 as unsorted municipal waste · Dispose of the LT7 in accordance with national waste disposal regulations
www.lifeloc.com | 303.431.9500

General Usage

25

Specifcations and Technical Data

Sensor

Electrochemical Fuel Cell Meets or exceeds DOT specifcations of +/ 005 BAC up to 100 5% from 100 BAC to 400 BAC

Accuracy

BAC, BrAC, g/210L 100 BAC +/ 005 BAC 100 BAC +/ 5% of measured value

Measurement Range Analysis time Display Memory Communications Power Supply Calibration Frequency
Accuracy Checks
Calibration Adjustment
Size Weight [w/ batteries] Operating temperature

mg/L 0 50 mg/L +/ 0 02 0 50 mg/L +/ 5% of measured value 000 to 600 BAC Approximately 6 seconds 2 2″ 240 x 320 color TFT display visible in daylight or dark
Up to 10,000 tests with all associated data USB
Fast charge NiMH battery pack or three AA Alkaline batteries
Lifeloc recommends calibrating the LT7 at least once a year or more often if specifed by your internal policies and procedures
Lifeloc recommends performing accuracy checks on your device once a month or more often if specifed by your internal policies and procedures to ensure proper operation between calibrations
Lifeloc recommends that you adjust your LT7 during calibration process if the as found condition is greater than +/ 003 BAC of your calibration standard (if your calibration standard is less than or equal a 100 BAC)
~2 75″ X 5″ X 1 3″ (70 X 127 X 33 mm)
8 oz (227 grams)
14° to 130° F ( 10° to 55°C)

Storage Temperature

0° to 130° F ( 18° to 55°C)

www.lifeloc.com | 303.431.9500

26 Menu Chart
Settings Chart
www.lifeloc.com | 303.431.9500

Warranty

27

Lifeloc Factory Warranty
The LT7 comes with a one year limited parts and labor warranty, efective on the date of purchase by the end user
The Warranty covers: · Parts and labor on covered repairs · Software updates, as applicable · Air freight back to the customer after the unit is repaired (U S only)
The Warranty does not cover: · Freight to the Lifeloc factory · Misuse, abuse, negligence or accidents
Lifeloc Technologies, Inc (“Lifeloc”) warrants to the buyer that at the time of shipment from Lifeloc’s facilities, all new Lifeloc Equipment purchased through Lifeloc or one of our authorized distributors will be free from defects in material and workmanship, under normal use and service, provided that the buyer gives Lifeloc written notice of any defect within twelve (12) months from original invoice date (the “Warranty”) The term “Lifeloc Equipment” includes all [portable hand held and fxed station breathalyzers manufactured by Lifeloc], as well as all printers, keyboards, cables, cases, and power supplies purchased as part of a new Lifeloc kit confguration Software is warranted to perform substantially in accordance with Lifeloc user manuals and to be free from defects in materials and workmanship under normal use and service for a period of twelve (12) months from original invoice date
The Warranty does not apply if the product has been altered, customized, repaired, reported stolen or modifed by someone other than a Lifeloc factory authorized technician, or if parts other than Lifeloc approved parts are used in replacement or repair
The buyer’s exclusive remedy and Lifeloc’s sole liability for breach of the Warranty shall be repair, replacement or, at Lifeloc’s option, refund of the original purchase price paid by the buyer to Lifeloc for the Lifeloc Equipment that is shown, to Lifeloc’s reasonable satisfaction, to be defective in breach of the Warranty (“Defective” and, the defect causing the Lifeloc Equipment to be Defective, a “Defect”) Under no circumstance shall Lifeloc be liable for an amount that exceeds the lesser of the cost of replacement or the original purchase price paid by the buyer to Lifeloc for the Lifeloc Equipment Lifeloc shall not be responsible for any customer software, customer settings or confguration data or customer test records resident in any products returned for service, repair, warranty or recertifcation
Lifeloc assumes no risk for damage in transit If Lifeloc determines that a Defect was outside of or not covered under the Warranty, Lifeloc will estimate repair and service charges and obtain the buyer’s
www.lifeloc.com | 303.431.9500

28

Warranty

authorization prior to conducting any work to repair the Defective Lifeloc Equipment Following repair, the Lifeloc Equipment will be returned to the buyer via standard ground transportation prepaid Express charges, if authorized by the buyer, will be invoiced at the diference between the express charges and the standard return shipping charges
Lifeloc shall not incur costs related to loss, damage or incomplete or inaccurate paperwork of returned product regardless of origination point
THIS WARRANTY IS THE PURCHASER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED LIFELOC SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY
NOTWITHSTANDING ANYTHING TO THE CONTRARY IN THE FOREGOING WARRANTY, OR IN ANY AGREEMENT BETWEEN THE BUYER AND LIFELOC PERTAINING TO ANY LIFELOC EQUIPMENT, OR AT LAW OR IN EQUITY:
THE WARRANTIES, OBLIGATIONS, AND LIABILITIES OF LIFELOC, AND THE REMEDIES OF THE BUYER SET OUT IN THE FOREGOING WARRANTY, ARE SOLE AND EXCLUSIVE, AND ARE MADE AND ACCEPTED BY THE BUYER IN LIEU OF, AND THE BUYER WAIVES AND RELEASES, ALL OTHER WARRANTIES, OBLIGATIONS, AND LIABILITIES OF LIFELOC, AND ALL OTHER CLAIMS AND REMEDIES OF THE BUYER, EXPRESS OR IMPLIED, ARISING BY STATUTE OR OTHERWISE, WITH RESPECT TO ANY DEFECT IN LIFELOC EQUIPMENT, INCLUDING BUT NOT LIMITED TO THE WAIVER AND RELEASE BY THE BUYER OF THE FOLLOWING: (i) ANY STATUTORY OR IMPLIED WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR ANY PURPOSE, (ii) ANY IMPLIED WARRANTY ARISING FROM COURSE OF DEALING, COURSE OF PERFORMANCE OR USAGE OF TRADE, AND (iii) ANY OTHER RIGHT, CLAIM, OR REMEDY WHATSOEVER OF THE BUYER OR OF ANY PERSON OR ENTITY CLAIMING BY, THROUGH, OR UNDER THE BUYER AGAINST LIFELOC, WHETHER ARISING PURSUANT TO THE FOREGOING WARRANTY, PURSUANT TO ANY AGREEMENT PERTAINING TO LIFELOC EQUIPMENT, OR IN CONTRACT, IN INDEMNITY, IN TORT (INCLUDING, BUT NOT LIMITED TO, NEGLIGENCE), IN PRODUCTS LIABILITY, IN STRICT LIABILITY, OR OTHERWISE BY ACCEPTING OR USING LIFELOC EQUIPMENT, THE BUYER FURTHER ACKNOWLEDGES, UNDERSTANDS, AND AGREES THAT NO STATEMENT OR REPRESENTATION HAS BEEN MADE BY LIFELOC, OR RELIED UPON BY THE BUYER, THAT IS INCONSISTENT WITH THE FOREGOING WARRANTY
Without prejudice to the foregoing, Lifeloc shall not have any obligation or liability, and the buyer acknowledges that Lifeloc shall not have any obligation or liability whatsoever, to the buyer, or to any person or entity claiming by, through, or under the buyer, whether arising pursuant to the foregoing Warranty, pursuant to any agreement pertaining to Lifeloc Equipment, or in contract, in indemnity, in tort (including, but not limited to, negligence), in products liability, in strict liability, or otherwise,
www.lifeloc.com | 303.431.9500

Warranty

29

(i) for any transportation, installation, removal, reinstallation, adjustment, or other expenses related to any Lifeloc Equipment covered by the Warranty or to other property, (ii) for any damage or loss to any property other than the Lifeloc Equipment covered by the Warranty, or (iii) for any special, indirect, incidental, or consequential damage or loss, even though such expenses, damages, or losses may be foreseeable, including, but not limited to: loss of profts or revenues, loss of use or equipment, cost of capital, cost of substitute equipment, repairs, or facilities, cost of downtime, or cost of purchased or replacement equipment or parts
The limited remedies of the buyer set forth above shall be exclusive even though they may fail of their essential purpose No agreement varying or extending the foregoing Warranty, no remedies, no exclusions, or no limitations shall be efective unless in a writing signed by an executive ofcer of LIFELOC The correction of any Defect shall in no way extend the duration of the Warranty; only the unexpired warranty term of the Lifeloc Equipment applies to any repaired or replacement Lifeloc Equipment that is provided under the Warranty
The Warranty is non transferable and is efective on all Lifeloce Equipment purchased from and after [September 1, 2014] Rights and recourse may vary by country

www.lifeloc.com | 303.431.9500

12441 West 49th Avenue, Suite 4 Wheat Ridge, CO 80033 303.431.9500 | 800.722.4872 fax: 303.431.1423 www.lifeloc.com
Part #51124 Rev 1 0 | ©2010 Lifeloc Technologies, Inc

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals