Pneumatic High Pressure Test Pump User s Manual

[Version number:1905V05]
Please download the latest version from www.additel.com

Warnings and cautions

  • Low pressure gauges can very easily be over presured if not careful.
  • Please take caution when applying pressures.
  • Do NOT exceed the safety pressure limit ( 3500 psi / 235 Bar ).
  • Do not switch between pressure and vacuum mode under pressure. Make sure you have vented the pressure and opened the isolation valve before switching to vacuum.
  • Over tightening connectors may cause damage.
  • Pump should be stored in a dry/clean environment.
  • Do NOT hold pressure above 1000 psi for more than 4 hours.
  • All moisture and contaminants should be cleaned out of the liquid isolator before creating a vacuum. This is accomplished by venting the pump when at high pressure.
  • For the least risk in contamination of the reference gauge, the DUT should be installed to the pressure port which is closest to the vent valve.
  • Additel is not liable for any safety problems or damages caused by misuse or incorrect operation.

Specification

  • Pressure range: 95% vacuum to 3000 psi (200 bar) (Based on an atmospheric pressure of 1 bar)
  • Temperature : 0 to 50 C/32 to 122 F
  • Humidity : <85%RH
  • Adjustment resolution : 0.001psi (0.1mbar, 10 Pa)
  • Overpressure Safety Limit : 3500 psi (235 bar)
  • Pressure media : Air
  • Weight : 14.4 lbs (6.5 kg ) for 920, 20 lbs (9.07 kg) for 920HV
  • Piston volume : 60 mi (3.7 in3) for 920,115 ml (7.02 in3) for 920 HV

Remark
If local atmosphere pressure is P, the vacuum can reach to – 0.95P bar.

Air routine

atecorp-ADT920-Pneumatic-High-Pressure-Test-Pump-fig-1

Configuration & Size

  1. Quick connector
  2. Hand pump
  3. Pressure/vacuum (P/V) selector (Positive pressure: pull it out; Vacuum: push it in)
  4. Do NOT switch while unit is under any pressure
  5. Pressure/Vacuum lock
  6. Fine adjust handle (Turn Clockwise to increase pressure)
  7. Maintenance cleaning port lsolation valve (isolates the calibration volume from the hand pump and check valve)
  8. Vent valve (clockwise to close/counterol ockw i se to release pressure)
  9. Vent port

Troubleshooting

atecorp-ADT920-Pneumatic-High-Pressure-Test-Pump-fig-4

0-Rings for pressure connector

atecorp-ADT920-Pneumatic-High-Pressure-Test-Pump-fig-5

Basic Operation

Connection.

atecorp-ADT920-Pneumatic-High-Pressure-Test-Pump-fig-6

Vent

Zeroing should be operated under this condition only.

Increase pressure process

Decrease pressure process

Vacuum process

Decrease vacuum process

Remark:

  1. Additel has made a concerted effort to provide complete and current information for the proper use of the equipment. The product specifications and other information contained this manual are subject to change without notice.