STERLINK U510 Veterinary Equipment User Guide

August 11, 2024
STERLINK

STERLINK U510 Veterinary Equipment

STERLINK-U510-Veterinary-Equipment-product

Specifications

  • Power Source: Electric
  • Preheating Time: Approximately 15 minutes
  • Sterilization Modes: Standard Mode, Advanced Mode
  • Sterilization Duration: Varies based on selected mode

Product Usage Instructions

  1. Power On
    Turn on the power switch located on the side of the sterilizer. Ensure the power line is plugged in and the power button is switched on.

  2. Preheating
    Preheating process starts automatically when the power is turned on. It takes about 15 minutes to complete. The device will enter Ready mode when preheating reaches 100%.

  3. Load Cartridge (Step 1)
    Open the lower compartment hatch handle on the front of the product. Load the sterilant consumables by sliding them onto the lower rails until they click into place.

  4. Load Cartridge (Step 2)
    Hold the hatch handle and lift it up to open the compartment. Ensure the consumables are correctly loaded, then close the hatch completely. Failure to close it properly may result in an error.

  5. Sterilization
    Open the chamber door and use the provided tray to load the object to be sterilized. Close the door and select the desired sterilization process using the dial. Press the dial to initiate the process.

  6. Sterilization Complete:
    When the sterilization process is finished, open the door and retrieve the sterilized object.

Frequently Asked Questions (FAQ)

  • Q: What should I do if I encounter an error during the sterilization process?
    A: Refer to the user manual for troubleshooting steps. If the issue persists, contact the manufacturer or service center for assistance.

  • Q: How should I clean the sterilizer?
    A: Use aqueous alcohol for cleaning both the exterior and interior. For dust, wipe with a wet tissue. For stubborn stains, use a cleaning sponge with water.

Quick Guide

  1. Power onSTERLINK-U510-Veterinary-Equipment- \(1\)

    • Turn on the power switch on the side of the sterilizer.
    • Make sure the power line is plugged in.
    • Make sure the power button is on.
  2. Preheating

    • When the power is turned on, the preheating proceeds automatically.
    • Takes about 15 minutes maximum.
    • Automatic entry to the Ready screen when 100% is reached.
  3. Load Cartridge (Step 1)

    • Lower compartment hatch handle on the front of the product.

    • Hold the hatch handle and lift it up to open the compartment.

  4. Load Cartridge (Step 2)

    • Load the sterilant consumables by sliding puck onto lower rails until click
    • Align it with the rail and slide it in until it clicks into place.
    • After checking that the sterilant consumables are loaded in the correct position, pull hatch handle down and completely close hatch.
    • Error will occur if it is not closed all the way
  5. Sterilization

    • Hold the door, open it
    • Use the dedicated tray provided to load the object to be sterilized
    • After loading, close the door and turn the dial to select the desired sterilization process.
    • Press the dial to perform the sterilization process.
  6. Sterilization CompleteSTERLINK-U510-Veterinary-Equipment- \(9\)

    • When the sterilization process is completed and the ‘Complete’ screen is displayed, open the door and retrieve the object.
    • The sterilization execution screen for each selected mode and the screen displayed after the sterilization is completed. (Remaining time is shown)

How to Take Action in the Event of Problems

Phenomenon Description How to Take Action



| **** Caution| Empty Cartridge| Replace cartridge
Expired Cartridge| Replace cartridge
shutdown occurred / during process|

  • Restart after checking the power cable
  • Call service center in case of recurrence

Emergency stop

(When Purification process is completed)

| ▪ Restart the process
sliding door / was opened

(When Purification process is completed)

|

  • Rdeosotrairst calfoteser dconfirming that the cartridge
  • Call service center in case of recurrence

chamber door / was opened (When Purification process is completed)|

  • Rdeosotrairst calfoteser dconfirming that the chamber
  • Call service center in case of recurrence



| Code. 10| Rebooting or / contact service center|

  • Restart the process
  • Call service center in case of recurrence

Code. 20| Base pressure / issue (When Purification process is completed)|

  • Rmeositsatrutrtehaenpdroocilefsrsoamfttehrerecmhaomvibnegrand sterilization object se of recurrence
  • Call service center in ca

Code. 21| Diffusion pressure / issue (When Purification process is completed)| Restart after checking the cartridge status
Code. 30| Invalid cartridge| Replace cartridge

Please refer to your user manual and contact the manufacturer if the issue is not resolved or recurs.

STERLINK U510 Self Check List (for reference)

NO. Contents Check list Results
Yes No
1 Installation Have you read the user manual thoroughly before the
installation?
2 Check if the product has been installed on a level surface (within 3
degrees)
3 Check if the product has been installed with a minimum clearance of 5cm
from the back/front/left/right sides
4 Check if the power cable provided by Plasmapp is being used Check if the
power strip or the wall outlet is properly grounded
5 Using Are you using a dedicated CARTRIDGE for your product?
6 Check if the CARTRIDGE has been correctly placed under the loading block
7 Check if the products are thoroughly dried before initiating the cycle
8 Clean Check if the power is turned off
9 Check if any prohibited chemicals or solutions have been used for cleaning
the product

STERLINK-U510-Veterinary-Equipment- \(12\)

How to Clean

Do not use organic solvents other than aqueous alcohol for exterior and interior cleaning.

If there is dust, lightly wipe the contaminated area with a wet tissue.
If the stain is severe or cannot be removed with a wet wipe, use a cleaning sponge with water and wipe off any remaining stains with a cleaning wipe or soft cloth.
If foreign substances including oil are present, spray disinfectant ethanol on the contaminated area and wipe off the remaining stain with a cleaning wipe or soft cloth.
In the case of crayons and colored pencils, wipe by gently rubbing the contaminated area with a clean eraser.

Cautions

  • Clean only when the machine is turned off.
  • Do not use cleaners containing sodium hypochlorite-based or chlorine-based solutions.
  • Do not use ketone substances and substances such as acetone, ethyl alcohol, toluene, methyl acid or methyl chloride.

Manufacturer

Company Plasmapp Co., Ltd.
Address (41061) 102, Cheombok-ro, Dong-gu, Daegu, Republic of Korea
Website www.plasmapp.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals