Vision TechShop VNHB Balance Scale Instruction Manual

August 9, 2024
Vision TechShop

Vision TechShop VNHB Balance Scale Instruction Manual
Vision TechShop VNHB Balance Scale

CALIBRATION

Linear Calibration in g Mode
  • Turn on the scale.
  • Press key during self checking, display will be show
  • Press key until display will be show tech
  • Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale
  • Press key to enter calibration, display will be show Icon
  • In succession press and keys to enter the function, display will be show
  • Press key to enter , display will be show
  • Press key to enter calibration, display will be show Icon
  • In succession press , , keys to enter the function display will be shown
  • Ensure the platform is empty
  • 3 seconds later, display will be show 1/6 of the capacity
  • Place the first calibration mass weight on the pan.
  • 3 seconds later, display will be show
  • Remove the first calibration mass weight from the pan.
  • 3 seconds later, display will be show 1/3 of the capacity
  • Place the second calibration mass weight on the pan.
  • 3 seconds later, display will be show
  • Remove the second calibration mass weight from the pan.
  • 3 seconds later, display will be show 1/2 of the capacity
  • Place the third calibration mass weight on the pan.
  • 3 seconds later, display will be show
  • Remove the third calibration mass weight from the pan.
  • 3 seconds later, display will be show 2/3 of the capacity
  • Place the fourth calibration mass weight on the pan
  • 3 seconds later, display will be show
  • Remove the fourth calibration mass weight from the pan.
  • 3 seconds later, display will be show 5/6 of the capacity
  • Place the fifth calibration mass weight on the pan.
  • 3 seconds later, display will be show
  • Remove the fifth calibration mass weight from the pan.
  • 3 seconds later, display will be show full capacity
  • Place the sixth calibration mass weight on the pan.
  • 3 seconds later, scale will be show  and back to setting mode.
  • Remove the sixth calibration (full capacity) mass weight from the pan.

Table 1: Required calibration weights

| 500 g| 600 g| 1200 g| 5000 g| 6000 g
---|---|---|---|---|---
Load1| 100 g| 100 g| 200 g| 1000 g| 1000 g
Load2| 200 g| 200 g| 400 g| 2000 g| 2000 g
Load3| 300 g| 300 g| 600 g| 3000 g| 3000 g
Load4| 500 g| 400 g| 800 g| 5000 g| 4000 g
Load5| /| 500 g| 1000 g| /| 5000 g
Load6| /| 600 g| 1200 g| /| 6000 g

For capacity of 500g & 5000g, it’s only required to load 4 calibration weights.

Normal Calibration in g Mode
  • Turn on the scale

  • Press key during self checking, display will be show

  • Press key until display will be show

  • Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale

  • Press key to enter calibration, display will be show Icon

  • In succession press   and keys to enter the function, display will
    be show

  • Press key to select

  • Press key to enter, display will be show

  • Press key to confirm, display will be show

  • Remove all the weight from the platform.

  • 3 seconds later, display will be show the calibration weight.

  • If necessary press key to select the weight as you desired.

  • Place the calibration weight on the platform

  • A few seconds later, scale will be show and back to setting mode.

Linear Calibration in lb Mode
  • Turn on the scale.
  • Press key during self checking, display will be show
  • Press key until display will be show
  • Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale.
  • Press   key to enter calibration, display will be show Icon
  • In succession press and keys to enter the function, display will be show
  • Press key to enter , display will be show
  • Press ** key to select Lb**
  • Press key to enter calibration, display will be show Icon
  • In succession press   keys to enter the function display will be shown
  • Ensure the platform is empty
  • 3 seconds later, display will be show 1/2 of the capacity
  • Place the first calibration mass weight on the pan.
  • 3 seconds later, display will be show
  • Remove the first calibration (1/2 of the capacity) mass weight from the pan
  • 3 seconds later, display will be show full capacity
  • Place the second calibration (full capacity) mass weight on the pan.
  • 3 seconds later, scale will be show and back to setting mode.
  • Remove the second calibration (full capacity) mass weight from the pan.

Table 1: Required calibration weights

  500 g 600 g 1200 g 5000 g 6000 g
Load1 0.5 lb 0.5 lb 1.0 lb 5.0lb 5.0lb
Load2 1.0 lb 1.0 lb 2.0 lb 10.0lb 10.0lb
Normal Calibration in lb Mode
  • Turn on the scale.
  • Press  key  during self checking, display will be show
  • Press key until display will be show
  • Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale
  • Press key to enter calibration, display will be show Icon
  • In succession press and keys to enter the function, display will be show
  • Press key to select
  • Press key to enter, display will be show
  • Press key to select Lb
  • Press key to confirm, display will be show
  • Remove all the weight from the platform.
  • 3 seconds later, display will be show the calibration weight.
  • If necessary press key to select the weight as you desired.
  • Place the calibration weight on the platform.
  • A few seconds later, scale will be show and back to setting mode

Note: Incase display will show any error message or incorrect measurement, repeat the calibration again.