CALIBRATION
Linear Calibration in g Mode
- Turn on the scale.
- Press key during self checking, display will be show
- Press key until display will be show tech
- Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale
- Press key to enter calibration, display will be show
- In succession press and keys to enter the function, display will be show
- Press key to enter , display will be show
- Press key to enter calibration, display will be show
- In succession press , , keys to enter the function display will be shown
- Ensure the platform is empty
- 3 seconds later, display will be show 1/6 of the capacity
- Place the first calibration mass weight on the pan.
- 3 seconds later, display will be show
- Remove the first calibration mass weight from the pan.
- 3 seconds later, display will be show 1/3 of the capacity
- Place the second calibration mass weight on the pan.
- 3 seconds later, display will be show
- Remove the second calibration mass weight from the pan.
- 3 seconds later, display will be show 1/2 of the capacity
- Place the third calibration mass weight on the pan.
- 3 seconds later, display will be show
- Remove the third calibration mass weight from the pan.
- 3 seconds later, display will be show 2/3 of the capacity
- Place the fourth calibration mass weight on the pan
- 3 seconds later, display will be show
- Remove the fourth calibration mass weight from the pan.
- 3 seconds later, display will be show 5/6 of the capacity
- Place the fifth calibration mass weight on the pan.
- 3 seconds later, display will be show
- Remove the fifth calibration mass weight from the pan.
- 3 seconds later, display will be show full capacity
- Place the sixth calibration mass weight on the pan.
- 3 seconds later, scale will be show and back to setting mode.
- Remove the sixth calibration (full capacity) mass weight from the pan.
Table 1: Required calibration weights
| 500 g| 600 g| 1200 g| 5000 g| 6000 g
---|---|---|---|---|---
Load1| 100 g| 100 g| 200 g| 1000 g| 1000 g
Load2| 200 g| 200 g| 400 g| 2000 g| 2000 g
Load3| 300 g| 300 g| 600 g| 3000 g| 3000 g
Load4| 500 g| 400 g| 800 g| 5000 g| 4000 g
Load5| /| 500 g| 1000 g| /| 5000 g
Load6| /| 600 g| 1200 g| /| 6000 g
For capacity of 500g & 5000g, it’s only required to load 4 calibration weights.
Normal Calibration in g Mode
-
Turn on the scale
-
Press key during self checking, display will be show
-
Press key until display will be show
-
Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale
-
Press key to enter calibration, display will be show
-
In succession press and keys to enter the function, display will
be show -
Press key to select
-
Press key to enter, display will be show
-
Press key to confirm, display will be show
-
Remove all the weight from the platform.
-
3 seconds later, display will be show the calibration weight.
-
If necessary press key to select the weight as you desired.
-
Place the calibration weight on the platform
-
A few seconds later, scale will be show and back to setting mode.
Linear Calibration in lb Mode
- Turn on the scale.
- Press key during self checking, display will be show
- Press key until display will be show
- Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale.
- Press key to enter calibration, display will be show
- In succession press and keys to enter the function, display will be show
- Press key to enter , display will be show
- Press ** key to select Lb**
- Press key to enter calibration, display will be show
- In succession press keys to enter the function display will be shown
- Ensure the platform is empty
- 3 seconds later, display will be show 1/2 of the capacity
- Place the first calibration mass weight on the pan.
- 3 seconds later, display will be show
- Remove the first calibration (1/2 of the capacity) mass weight from the pan
- 3 seconds later, display will be show full capacity
- Place the second calibration (full capacity) mass weight on the pan.
- 3 seconds later, scale will be show and back to setting mode.
- Remove the second calibration (full capacity) mass weight from the pan.
Table 1: Required calibration weights
500 g | 600 g | 1200 g | 5000 g | 6000 g | |
---|---|---|---|---|---|
Load1 | 0.5 lb | 0.5 lb | 1.0 lb | 5.0lb | 5.0lb |
Load2 | 1.0 lb | 1.0 lb | 2.0 lb | 10.0lb | 10.0lb |
Normal Calibration in lb Mode
- Turn on the scale.
- Press key during self checking, display will be show
- Press key until display will be show
- Note: Before enter the tech menu, it’s required to press calibration switch, which is locating below the scale
- Press key to enter calibration, display will be show
- In succession press and keys to enter the function, display will be show
- Press key to select
- Press key to enter, display will be show
- Press key to select Lb
- Press key to confirm, display will be show
- Remove all the weight from the platform.
- 3 seconds later, display will be show the calibration weight.
- If necessary press key to select the weight as you desired.
- Place the calibration weight on the platform.
- A few seconds later, scale will be show and back to setting mode
Note: Incase display will show any error message or incorrect measurement, repeat the calibration again.