DAM IND308/B520 Terminal User Guide

June 6, 2024
DAM

IND308/B520 Replacement Quick Guide

Hardware Replacement

  1. Power supply NoticeC30 requires 24VDC power supply
    Pin| Description
    ---|---
    GND| Negative terminal
    +24VDC| Positive terminal
  2. Replacement of analog load cell interface follows the sequence of B520 wire
  3. Serial interface
    IND308 PIN| B520 PIN| Description
    ---|---|---
    485A| 485A| RS485 RxD+
    485B| 485B| RS485 RxD-
    COM| COM| Common
    SHD| | Shield

Scale calibration

**
Enter “Scale”, set “Capacity & Increment”

Enter “Capacity & Increment”, set
F1.2 → “Capacity”
F1.3 → “Increment”|
Enter “Calibration”,”Linearity” set to “Disabled”
1. Enter “Set Zero”, empty the scale and press”“to do the zero calibration. When successful, the message “Capture Zero OK” displays and press “” to return.
2. Enter 
“Span Calibration”** , enter “Test load 1”, and press ““, place Test load 1 on the scale and press”“. When successful, the message “Capture Span OK” displays and press”“to return.
---|---

PLC Mode DI/O Definition

Input replacement Output replacement
IND308 PIN 8520 PIN
PIN** Description
IN1 IN1
1N2 1N2
1N3 1N4
1N4 1N5
COM1 COM1
+24VDC +24VDC
1N5 1N6
1N6 IN7/IN8
COM2 COM2
+24VDC +24VDC

PLC Mode Parameters setting


Enter “Application”, and set
F2.1:0 → “Work Mode” : “PLC Mode”|
Enter “Control Mode”, and set
F2.3: 0 → “Output Type” : “Concurrent”
F2.3: 1 → “Output Type” : “Independent”
F2.6: 0 → “Judge Tolerance” : “Disabled”
F2.6: 1 → “Judge Tolerance” : “Enabled”
F3.18: 0 → “Discharge Near Zero” : “Disabled”
F3.18: 1 → “Discharge Near Zero” : “Enabled”
F2.5: 0 → “Power Up Mode Continue Bagging” : “Continue Bagging”
F2.5: 1 → “Power Up Mode Continue Bagging” : “Empty Reset”
---|---

Enter “Time”, and set
F3.1.1 → “SP1 Inhibit Time”
F3.1.2 → “SP2 Inhibit Time”
F3.1.3 → “Spill Inhibit Time”
F3.2 → “Stable Time”
F3.3 → “Before Discharge Off”
F3.4 → “After Discharge Off”
F3.16 → “Jog Time”|
Enter “Input” , if any input has no connection, set the input as “None”. If any input function changes, set the input as corresponding function.

Enter “Output” , if any output has no connection, set the output as “None”. If any output function changes, set the output as corresponding function.

Bulk Grain Mode DI/O Definition

Input replacement Output replacement
IND308 PIN 8520 PIN
PIN** Description
IN1 IN1
IN2 IN2
IN3 IN3
IN4 IN4
IN5 IN5
COM1 COM1
+24VDC +24VDC
IN6 INC
IN7 IN7/IN8
IN8 IN7/IN8
COM2 COM2
+24VDC +24VDC

Bulk Grain Mode Parameters Setting

Enter “Application”, and set
F2.1: 1 → “Work Mode”: “Bulk Grain”


Enter “Sequence”, and set
F3.12 → “Bulk Weight”

|
Enter “Control Mode”, and set
F2.3: 0 → “Output Type” : “Concurrent”
F2.3: 1 → “Output Type” : “Independent”
F3.10 → “Periodic Zero Cycles”
F2.6: 0 → “Judge Tolerance”: “Disabled”
F2.6: 1 → “Judge Tolerance” : “Enabled”
F2.5: 0 → “Power Up Mode Continue Bagging” : “Continue Bagging”
F2.5: 1 → “Power Up Mode Continue Bagging” : “Empty Reset”
Enter “Signal Mode”, and set
F2.7: 1 → “Upper Hopper Low” : “Default On”
F2.7: 2 → “Upper Hopper Low” : “Default Off”
F2.8: 1 → “Lower Hopper High” : “Default On”
F2.8: 2 → “Lower Hopper High” : “Default Off”
---|---

Enter “Time”, and set
F3.1.1 → “SP1 Inhibit Time”
F3.1.2 → “SP2 Inhibit Time”
F3.1.3 → “Spill Inhibit Time”
F3.2 → “Stable Time”
F3.3 → “Before Discharge Off”
F3.4 → “After Discharge Off”
F3.16 → “Jog Time”|

Enter “Input” , if any input has no connection, set the input as “None”. If any input function changes, set the input as corresponding function.

Enter “Output” , if any output has no connection, set the output as “None”. If any output function changes, set the output as corresponding function.

No Hopper Mode DI/O Definition

Input replacement Output replacement
IND308 PIN 8520 PIN
PIN** Description
IN1 INI
IN2 IN2
IN3 IN3
IN4 IN4
COM1 +24VDC COM1 +24VDC
OUT6 OUT5
OUT6 Clamp Flap/Drop
IN5 INS
IN6 IN6
IN7 IN7/IN8
IN8 IN7/IN8
COM2 COM2
+24VDC +24VDC

Have Hopper Mode Parameters Setting


Enter “Application”, and set
F2.1: 2 → “Work Mode”: “Have Hopper Mode”Enter “Sequence”, and set
F2.16: 0 → “Flap/Drop Mode” : “Flap”
F2.16: 1 → “Flap/Drop Mode” : “Drop”
F3.7 → “Final Flap/Drop Times”
F3.5 → “Flap Interval”
F3.11 → “Preset Packages”
F3.13 → “Ton Bag Weight”|
Enter “Control Mode”, and set

F2.3: 0 → “Output Type” : “Concurrent”
F2.3: 1 → “Output Type” : “Independent”
F3.10 → “Periodic Zero Cycles”
F2.6: 0 → “Judge Tolerance” : “Disabled”
F2.6: 1 → “Judge Tolerance” : “Enabled”
F3.18: 0 → “Discharge Near Zero” : “Disabled”
F3.18: 1 → “Discharge Near Zero” : “Enabled”
F2.5: 0 → “Power Up Mode Continue Bagging” : “Continue Bagging”
F2.5: 1 → “Power Up Mode Continue Bagging” : “Empty Reset”
Enter “Signal Mode”, and set
F2.7: 1 → “Upper Hopper Low” : “Default On”
F2.7: 2 → “Upper Hopper Low” : “Default Off”
F2.8: 1 → “Lower Hopper High”: “Default On”
F2.8: 2 → “Lower Hopper High” : “Default Off”
---|---


Enter “Time”, and set
F3.1.1 → “SP1 Inhibit Time”
F3.1.2 → “SP2 Inhibit Time”
F3.1.3 → “Spill Inhibit Time”
F3.2 → “ Stable Time”
F3.3 → “Before Discharge Off”
F3.4 → “ After Discharge Off”
F3.16 → “Jog Time”
F3.14 → “Clamp Delay”
F3.15 → “Delay to Declamp”|
Enter “Input”, if any input has no connection, set the input as “None”. If any input function changes, set the input as corresponding function.
Enter “Output”, if any output has no connection, set the output as “None”. If any output function changes, set the output as corresponding
function.

No Hopper Mode DI/O Definition

Input replacement Output replacement
IND308 PIN 8520 PIN
PIN** Description
IN1 IN1
1N2 IN2
1N3 IN3
IN4 IN4
COM1 COM 1
+24VDC +24VDC
INS INS
1N6
IN? IN7/IN8
IN8 IN7/IN8
COM2 COM2
+24VDC +24VDC

No Hopper Mode Parameters Setting

Enter “Application”, and set
F2.1: 3 → “Work Mode” : “No Hopper Mode”

Enter “Sequence”, and set
F2.16: 0 → “Flap/Drop Mode” : “Flap”
F2.16: 1 → “Flap/Drop Mode” : “Drop”
F3.6 → “Middle Flap/Drop Times”
F3.7 → “Final Flap/Drop Times”
F3.5 → “Flap Interval”
F3.11 → “Preset Packages”
F3.9 → “Rising Threshold”
F3.8 → “Rising Time”

|
Enter “Control Mode”, and set
F2.3: 0 → “Output Type” : “Concurrent”
F2.3: 1 → “Output Type” : “Independent”
F3.10 → “Periodic Zero Cycles”
F2.6: 0 → “Judge Tolerance” : “Disabled”
F2.6: 1 → “Judge Tolerance” : “Enabled”
F3.18: 0 → “Discharge Near Zero” : “Disabled”
F3.18: 1 → “Discharge Near Zero” : “Enabled”
F2.5: 0 → “Power Up Mode Continue Bagging”: “Continue Bagging”
F2.5: 1 → “Power Up Mode Continue Bagging” : “Empty Reset”
Enter “Signal Mode”, and set
F2.7: 1 → “Upper Hopper Low” : “Default On”
F2.7: 2 → “Upper Hopper Low” : “Default Off”
F2.8: 1 → “Lower Hopper High” : “Default On”
F2.8: 2 → “Lower Hopper High” : “Default Off”
---|---

Enter “Time”, and set
F3.1.1 → “SP1 Inhibit Time”
F3.1.2 → “SP2 Inhibit Time”
F3.1.3 → “Spill Inhibit Time”
F3.2 → “Stable Time”
F3.16 → “Jog Time”
F3.14 → “Clamp Delay”
F3.15 → “Delay to Declamp”|
Enter “Input”, if any input has no connection, set the input as “None”. If any input function changes, set the input as corresponding function.

Enter “Output”, if any output has no connection, set the output as “None”. If any output function changes, set the output as corresponding
function.

Dual Scale Mode DI/O Definition

Input replacement Output replacement
IND308 PIN 8520 PIN
PIN** Description
IN I NiI
IN2 IN2
IN3 INS
IN4 IN4
COMI COMI
+24VDC +24VDC
INS INS
IN6 IN6
IN7 INI/IN8
INS IN7/IN8
COM2 COM2
+24VDC +24VDC

Dual Scale Mode Parameters Setting


Enter “Application”, and set
F2.1: 5 → “Work Mode”: “Dual Scale Mode”


Enter “Sequence”, and set
F3.11 → “Preset Packages”
F2.11: 0 → “Master/Slave Mode”: Master Scale
F2.11: 1 → “Master/Slave Mode”: Slave Scale
F2.12: 0 → “Dual Scale Discharge”: “Combination”
F2.12: 1 → “Dual Scale Discharge”: “Interlock”| **
Enter “Control Mode”, and set
F2.3: 0 → “Output Type”: “Concurrent”
F2.3: 1 → “Output Type”: “Independent”
F3.10 → “Periodic Zero Cycles”
F2.6: 0 → “Judge Tolerance”: “Disabled”
F2.6: 1 → “Judge Tolerance”: “Enabled”
F3.18: 0 → “Discharge Near Zero”: “Disabled”
F3.18: 1 → “Discharge Near Zero”: “Enabled”
F2.5: 0 → “Power Up Mode Continue Bagging”: “Continue Bagging”
F2.5: 1 → “Power Up Mode Continue Bagging”: “Empty Reset”
Enter “Signal Mode”, and set
F2.7: 1 → “Upper Hopper Low”: “Default On”
F2.7: 2 → “Upper Hopper Low”: “Default Off”
F2.8: 1 → “Lower Hopper High”: “Default On”
F2.8: 2 → “Lower Hopper High”: “Default Off”**
---|---


Enter “Time”, and set
F3.1.1 → “SP1 Inhibit Time”
F3.1.2 → “SP2 Inhibit Time”
F3.1.3 → “Spill Inhibit Time”
F3.2 → “Stable Time”
F3.3 → “Before Discharge Off”
F3.4 → “After Discharge Off”
F3.16 → “Jog Time”
F3.14 → “Clamp Delay”
F3.15 → “Delay to Declamp”|

Enter “Input” , if any input has no connection, set the input as “None”. If any input function changes, set the input as corresponding function.


Enter “Output” , if any output has no connection, set the output as “None”. If any output function changes, set the output as corresponding
function.

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

DAM User Manuals

Related Manuals