chois TECHNOLOGY XST200 Electronic Thermometer User Manual
- June 6, 2024
- chois TECHNOLOGY
Table of Contents
chois TECHNOLOGY XST200 Electronic Thermometer User Manual
Introduction
Summary
This product is Class II, Non-Invasive Device. The electronic thermometer is a device that measures the body temperature using the thermistor resistance change. The body temperature display must be connected to the mobile app installed on tie smartphone, so that it can be displayed. Smartphones are rot provided as components of this product.
Intended Use
Measuring instrument for measuring the body temperature of a patent by using the electrical properties
Intended Operator
Education
At least 11 years – 5 years of intensive reader experience (school)
Knowledge
Be able to distinguish body position (measurement See, chest, armpit)
Understanding languages
To understand English or user manual.
Experience
No experience required
Acceptable faults
No specific information
Indications
No record.
Side Effects and Contraindications
No record.
Before Use
- Be sure to read and use the user manual.
- Do not use for other purposes.
- Stabilize the state of the person before the measurement and make the measurement in the correct stale.
- The temperature detecting Lit of the body temperature sensor is brought into close contact with the skin. Use in body cavity is prohibited.
Safety Caution and Warning
Danger
- Be sure to use the specified rated battery and check the polarity of the battery.
Warning
- If an abnormality is found in the instrument, immediately stop the operation of the instrument, remove the instrument from the human body and ask the manufacturer for inspection.
- Do not exchange battery arbitrarily.
Caution
- This electronic thermometer can be used for all ages. However, please be sure to use as directed by your doctor if you have severe high fever.
- Be careful not to use the device by not intended operator.
- Be careful not to cause damage to the device.
- Be careful not to let liquid or foreign matter get into the body.
- Please contact the manufacturer if any of the following situations occur.
- If you have a skin reaction (urticaria, allergies, etc.)
- If something is wrong with the whole device
- In case of using in parallel with other medical devices, change of measured body temperature occurs remarkably
- If the unit detects a heat, be sure to remove it from your body.
- It should not be used in the same place where there is equipment with strong electromagnetic wave(e.g. X-ray, MRI etc.)
Notice
- Since this electronic thermometer should be used in conjunction with a smartphone, operation may not be smooth depending on the location and power status of the smartphone. Please observe the followings.
- Maintain the power supply status of the smartphone by using auxiliary battery or adapter for continuous measurement of body temperature.
- Since the normal temperature measurement range is within 15m from the thermometer, be careful not to let the smartphone deviate more than 10m.
- When using in two connected rooms, distance to use may be significantly reduced. Use in one independent space if possible.
- This electronic thermometer can be used for all ages, but please be aware of the following when using Young or Infant.
- Measures should be taken after 30 minutes or more after the infant enters the room from outside or moves the thermometer to a different temperature environment.
- The sensor part on the back of the product is very sensitive and should always be kept clean and intact for accurate measurement.
- In case of body temperature measurement for infants, in case of severe strangulation, distance between skin and sensor may not be kept constant, so measurement error may occur.
- Do not allow infants or children to touch the thermometer sensor and battery with their mouths or with wet hands.
- This electronic thermometer attached to the body, please note the following points.
- limitation for contact duration : 24hour – 30days
Specifications
System, Controls, Feature Description
NO | Name | Function |
---|---|---|
A | Case | Protect PCB ASSY, shape product |
B | Temperature sensor | Detect human body temperature and convert it to digital |
signal
C| Button| ON / OFF button to operate the product
D| Battery cover| So that the battery can be easily attached and detached.
E| LED indicator| Display the Device status LED OFF : Sleep mode LED ON(every
3sec blinking) : Active mode
Product Performance
Section | Specification |
---|---|
Battery | CR2032 (DC 3V) |
Measurement Range | 50.36°F – 139.82°F |
Accuracy | ± 0.18°F (93.2°F – 107.6°F) |
± 0.54°F (50.36°F – 93.02°F 107.7°F -139.8299
Minimum unit| 0.01°F
Measurement Method| Thermistor / Direct mode only
Measurement site| armpit
Measurement Display| Bluetooth communication with your smartphone to show in
your mobile app
Measurement Transient time| After 15-20 minute
---|---
Use Distance| Max 15m
Lift time| Max 5days
Software Version| Firmware : Ver, 1,0,0 Mobile App : Ver. 2.0.9
Product characteristics
- Protection for electric shock Internal Power device, BF type applied part
Waterproof grade
IP22 (according tolEC529)
- Protected against a solid object grater than 12.5mm, such as a finger.
- Protected against water drops at a 15degree angle.
Label
Pictogram
Environmental condition ( Storage, transfer, operation)
-
Storage Condition
Temperature : 50°F – 158°F
Humidity : 0 – 93 % RH
Air pressure : 700 – 1,060 hPa -
Transfer Condition
Temperature : 50°F – 158°F
Humidity : 0 – 93 % RH
Air pressure : 700 – 1,060 hPa -
Operation Condition
Temperature : 41°F – 140°F
Humidity : 15 – 93 % RH
Air pressure : 700 – 1,060 hPa
Operation
Preparation before use (before operation)
- Stabilize the state of the person before the measurement and make the measurement in the correct state.
- The temperature detecting unit of the body temperature sensor is brought into close contact with the skin. Use in body cavity is prohibited.
How to operation
-
Install a mobile app (name: Thermosafer) on you smartphone.
-
Activate the Bluetooth function of the smartphone.
-
Race the operating button of the device on and check Whether the LED indicator is on, If LED is blinking every 3sec then operation is active.
-
Attach the thermometer to the side using the patch.
-
Rut the mobile app and select the thermometer you searched tor.
-
Check that the temperature is displayed on the mobile app when the thermometer is normally connected.
-
Changes in body temperature information will be stabilized after about 15 to 20 minutes.
Precautions when operation
- If there is any problem with the device, stop operation and contact the manufacturer.
After operation management methods
- Avoid contact with water. Do not store near water.
- Do not use in places subject to adverse influence such as air pressure, temperature. humidity. wind.
- Pay attention to the stable state such as vibration and shock including in transportation).
- Do not store in chemical storage area or gas generating place.
Routine maintenance
Routine check
- Please check the battery remaining in the mobile app and replace the battery.
-
Battery replacement is as follows
- Be sure to clean the surface of the product with a cloth
- Please check the storage conditions and usage conditions of the product.
- You do not need to take a separate calibration.
Trouble shootings
Symptom | Possible Reason | Solution |
---|---|---|
The temperature data cannot be displayed properly |
- Low battery of Instrument
|
- Tum on again
- Please replace battery
Cannot turn on Instrument|
- May the battery power is shortage or no electricity
- There are some problems of the instrument
|
- Turn on again
- Please replace battery
- Contact customer service
Unusual temperature data|
- The position of instrument is not correct
- There are some problems of the instrument
|
- Check the position of instrument.
- Contact customer service
Mobile App error
- User data not saved
- Wrong data received
- Measurement error
- Alarm error (High temp, Low temperature, Low battery, Disconnection
|
- The app may not install properly or may not be compatible with your smartphone
|
- Contact customer service
Shop for supplies, A/S
Supplies list and Purchase path
Item | Purchase path |
---|---|
Patch | Contact customer service |
Battery | Contact customer service |
A/S explanation
If you have any problems while using the product, or if you need to improve
things, please contact the following contact details and we will kindly
consult you.
Email: cs@kmact.com
Fax: 281-405-0891
Homepage: https://kmact.com
Warranty
- The warranty period for this product is 3 years. Any defect of the product caused by materials and mechanical problems will be covered for free for a year. If our model you purchased is found to be defective within that time, we will promptly repair or replace it. However, this warranty does not cover problems or damages resulted from unauthorized repair, modification or disassembly. In case of defect or failure of the product that occurred after one year of purchase during the warranty period, paid repair is possible.
- If any of the following situations occur within 1 year during the warranty period, it will be charged.
- In case of malfunction due to user’s mistake
- failure or damage caused by the arbitrary disassembly, modification of structure, performance, and function
- failure or damage caused by the use of the product other than its intended purpose
- In case of malfunction due to carelessness such as dropping and flooding after purchase
- failure due to natural disaster(earthquake, fire, flood, lightning)
EMC Compliance
Electromagnetic Emission The product is suitable for use in an specific electromagnetic environment. The customer and/or the user of the product should assure that it is used in an electromagnetic environment as described below.
Emission Test | Compliance | Electromagnetic Environment Guidance |
---|---|---|
RF- emission CISPR 11 | Group 1 | The product use RF energy only for its |
internal function. Therefore, its RF emissions are very low and not likely to
cause any interference in nearby electronic equipment.
RF emission CISPR 11| Class A| The product is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used
for domestic purpose
Harmonic emissions IEC 6100-3-2| Class A
Voltage fluctuations / flicker emissions IEC 61000-3-3| Complies
Electromagnetic Immunity
The product is suitable for use in a specific electromagnetic environment. The
customer and/or the user of the product should assure that it is used in an
electromagnetic environment as described below.
Immunity Test| IEC 60601 – level| Compliance| Electromagnetic Environment
Guidance
---|---|---|---
Electrostatic dischargeable SD IEC61030-4-2| ± 6kV contact ± 1 5kV air| ± 6kV
contact ± 15kV air| Floor should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity should be at least
30 %
Electrical fast transient/1)11.st IEC61000-4-4| ± 2kV for power supply lines ±
1 kV for input/output lines| ± 2kV for power supply lines ± 1 kV for
input/output lines| Mains power quality should be that of a typical commercial
and/or hospital environment
Surge IEC610)0-4-5| ± 1 kV differential mode ± 2kV common mode| ± 1 kV
differential mode ± 2kV common mode| Mains power quality should be that of a
typical commercial and/or hospital environment
Voltage dips, short interruption and voltage variations on power supply Input
Ines IEC61030-4-1| <5% U (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT)
for 5 cycles 70% UT (30% DIP IN UT) for 25 cycles <5% UT (>95% dip in UT) for
5 sec| <5% U-(>95% dip in UT) for 0.5 cycle 40% U-(60% dip in UT) for 5 cycles
70% U-(30% DIP IN LI) for 25 cycles <5% U-(>95% dip in UT) for 5 sec| Mains
power quality should be that of a typical commercial and/or hospital
environment. If the user of the product requires continued operation during
power mains interruptions, it is recommended that the product be powered from
an uninterruptible power supply or a battery
Power frequency 50/60 Hz) magnetic field IEC 61000-4-8| 3 A/m| 3 A/m| Power
frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
Conducted RF IEC 61000 4-6| 3 Vrms 150 kHz to 80 MHz| 3 Vrms| Portable and
mobile RF communications equipment should be used no closer to any part of the
product, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance: d = 1.21/P
d = 1.2./P for 80 MHz to 800 MHz
d = 2.3VP
for 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in Watt (Vs./) according to the transmitter manufacturer and d is the re-commended separation distance in meters (m) Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey’, should be less than the compliance ) level° in each (( (i)) frequency range Interference may occur in the vicinity of equipment marked with the symbol described lateral.
---|---|---|---
Radiated RF IEC 61000-4-3| 3 V/m 80 MHz to 2.5 GHz| 3 Vrms
| |
Note 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 : These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects, people and animals.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered, if the measured field strength in the location in which the product is used exceeds the applicable RF compliance level above, the product should be observed, additional measures may be necessary, such as reorienting or relocating the product. ‘Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended Separation Distances between portable and mobile HF- communications equipment and the product
The product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the product can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the product – according on output power and frequency of the communications equipment – as recommended in the following table.
Rated maximum output power of transmitter in watts (W)| Separation
distance according to the frequency of transmitter in meter (m)
---|---
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, people and animals.
Wastes
Instructions for waste procedure
Disposal of Electronic Thermometers
- Follow local laws, standards and guidelines for disposal of used electrical equipment.
- Do not allow parts to be contaminated during disposal.
Disposal of packaging materials
- All packaging materials can be disposed of or recycled in an environmentally friendly manner.
- Please send the old packing materials to the recycling collection companies for the prevention of waste and recycling of raw materials.
Radio Regulations :
FCC Part 15 standard(s) ISO/IEC 60601-1-2 Electromagnetic Compatibility
FCC Statement and Legal Notices
This device complies with part 15 of the FCC rules. Operation is subject to
the following two conditions (1) this device may not cause harmful
interference, and (2) this device must accept any interference received
including interference that may cause undesired operation. FCC WARNING: any
changes or modifications not expressly approved by the party responsible for
compliance could void the users authority to operate this equipment under FCC
regulations.
FCC RF Radiation Exposure Statement: This equipment complies with FCC RF Radiation exposure limits set forth for an uncontrolled environment. This device and its antenna must not be co-located or operating in conjunction with any other antenna or transmitter.
FCC ID: RVBXST200-BT
Model: XST200 Responsible Party: KM HEALTHCARE 7811 N. Shepherd Dr.
SteIt215-A, Houston, Texas 77088, USA, TEL:
281-405-0888
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>