TPC PC2800 ProClean Extra Oral Suction System Instruction Manual

June 6, 2024
TPC

TPC PC2800 ProClean Extra Oral Suction System fig
5TPC PC2800 ProClean Extra Oral Suction System

Overview

Thank you very much for choosing the ProClean Extra Oral Suction System. Please read this instruction manual and the accompanying materials in detail. The ProClean is covered by a 2-year warranty (not including replacement filters and UV lamp)

Contents

This instruction manual includes an overview, instructions for use of the suction unit, operation routines, maintenance, and precautions. You can view related content as needed and get corresponding help.

Structure and Composition
The external-oral suction unit is mainly composed of a generally vacuum pressure suction module, a primary filter, high-efficiency filters, ultraviolet sterilization, plasma sterilization, and a free arm.

Ideal Applications

It is used for the suction and purification of bacterial aerosols between medical staff and patients during dental treatment.

Safety Information

Electrical Safety

  •  The power supply for the environment in which this product is used must have a good, standard grounding device, and the grounding wire must be firm. It must use the national standard single-phase three-plug power cord, and cooperate with a well-grounded single-phase three-hole socket to ensure good grounding of this product.

attention

  • To avoid the risk of electric shock, the device can only be connected to a power cable with a protective ground. An incorrect grounding plug can cause a risk of electric shock.
  •  Before powering on this product, please confirm that the power supply you provide can meet the input power information marked near the power input port of this product.
  • Do not share the socket with other electrical appliances to prevent the voltage from being unstable sometimes, which may damage the product
  • When performing maintenance/repair or cleaning this product, be sure to unplug the power cord to ensure that the product is completely powered off before proceeding.
  • Regularly check the power cord and power plug for damage, and make sure that the power cord is not crushed by other objects. not crushed by other objects.

Regular Inspection

  • Regular safety inspections should be performed by trained and skilled personnel, the following items should be checked at least once a year:
  • Check the functional status of the equipment and machines.
  • Check whether the relevant safety signs are clear.
  • Check whether the performance of the device meets the instructions in the manual.
  • Test whether the grounding resistance is less than the resistance value 0.1 required in GB9706.1. 1.4.3 Precautions
  1.  You should read and understand all contents in the manual carefully before operating.

  2.  Follow all warnings and instructions on the instrument during operation.

  3.  During maintenance and repair, the water and electricity sources must be emptied before the water and
    electricity sources are turned off.

  4.  Only trained and skilled medical personnel can operate the equipment. Improper use will cause serious personal injury.

  5.  In the following cases, please do not use the device, nor can you open the device or replace the accessories without authorization, and can only be repaired by the manufacturer or authorized service provider. Damaged power cord.

    • The device does not work properly.
    • The device is damaged.
    • Water entered the device.
    • The device runs loudly or has a harsh sound, the output air is overheated, and an unpleasant odor is emitted.
    •  Product introduction, instruction manual, technical manual, product repair card, circuit diagram, parts list, and other useful information, etc., need to be provided when the technician of an authorized service provider repair.
  6.  The operator should unplug the power plug when leaving.

  7.  In order to avoid damage to machinery and electrical appliances, prevent attachments from falling.

  8.  The disposal of wastes and residues of products shall comply with the requirements of relevant national environmental protection regulations.

  9.  Strong electromagnetic waves will not be generated during the operation of this product, but in order to ensure the normal use of other medical devices, it should be prepared for grounding or try to avoid the use of other highly sensitive instruments and equipment at the same time, so as to avoid the equipment from being interfered by electromagnetic sources.

Cleaning

Please keep the environment of the product clean. Before cleaning the product, please disconnect the power supply, wipe the outer surface of the product with a soft cloth and neutral detergent; ABS parts should be cleaned with a soft cloth dipped in soapy water; metal baking parts are cleaned with a soft cloth dipped in soapy water or wiped with car wax.

Attention Do not use liquid or cleaning agents containing flammable substances.

Symbols and Marks

During installation, operation and maintenance, please pay attention to the important information represented by the symbols on the packaging and product labels:

Technical Specifications

Technical Specifications:

  • Name TPC ProClean Extra Oral Suction
  • Model Model PC2800
  • Voltage/Frequency AC 220V / 50 Or 110V 60
  • Power 1200W
  • Noise 65dB
  • Suction rate / wind pressure 4000L/min 5KPA
  • Suction pressure: 29 KPa
  • Dimensions 368.1323.5762MM (not including the free arm)
  • Filter level HEPA 13 Filter efficiency: 99.97%
  • Classification
  • Classified by type of anti-motor: Class I equipment
  • Classification by electric shock protection: ordinary equipment without application part
  • Classified according to the degree of protection against liquid ingress: IP20 equipment
  • According to the safety level when used in the case of anesthetic gas mixed with air or flammable anesthetic gas mixed with nitrous oxide Classification: Not AP type and APG type equipment
  • Classified by operating mode: intermittent operating equipment
  • Transport, storage, and use conditions

Transport and storage
The machine should be transported and stored under the following conditions: Temperature: -10 ~ + 50 Relative humidity range: 90%, atmospheric pressure range: 50 kPa ~ 106 kPa

Working condition
Temperature: + 5 ~ 40 , relative humidity: 80%, atmospheric pressure: 86kPa ~ 106kPa
There should not be a lot of dust and corrosive substances in the surrounding air.

Electromagnetic Compatibility
Medical electrical equipment requires special tips on electromagnetic compatibility, as well as installation and use according to the electromagnetic compatibility information provided in the random files; portable and mobile radio frequency communication equipment may affect the use of the equipment. In order to ensure the best EMC performance, please use the specified cable. The AC power cord specification is 1.8 meters in length; the transducers and cables sold by the company as spare parts for the components of the internal active oxygen air disinfection machine. The use of accessories, transducers and cables outside the regulations may cause an increase in the emission of the air disinfection machine or reduced immunity.
The suction machine should not be used close to or stacked with other equipment. If it must be used close to or stacked, it should be observed to verify that it can operate normally.

Product Structure and Installation

Structure diagram:

Guidance and manufacturer’s declaration-electromagnetic emissions

Installation procedure:

Open-package Inspection
Open the package, and check whether the device is in good condition, check the complement of the unit, spare parts, and consumable parts according to the packing list. If you have any questions, please contact our distributor or our company directly.

Installation

  1.  Move the machine to the appropriate location on the flat solid floor and press on the brake at the bottom of the wheel.
  2.  Install the free arm onto the mounting joint at the top of the box.
  3.  Install the suction cover onto the designated spot as the above picture shows.
  4.  Using one-phase AC 220V-50hz or other local power (220V 60hz, 240V 50hz, 110V 60hz) would be ok.

Operation mode control
Move the machine to the appropriate location and press on the brake at the bottom of the wheel, plug in the power cord, pull the free arm to an appropriate position according to the demand, switch on the power and device start to work. Suction power could be controlled through the knob on the control panel, turn off the power switch after use, pull back the free arm.
This equipment has a step-less knob, it can meet the needs for most operations when the knob is adjusted to the low power area (green area). If some surgery need larger flow, it can be adjusted clockwise, the maximum power is not recommended.

Working principle for ProClean Extra Oral Suction System

 Maintenance

  1.  Keep the outside of the machine clean and tidy, do not use chemical detergent, alcohol to wipe the machine.
  2.  Clean the primary filter regularly (wash with water once every month).
  3.  Maintain and replace HEPA filter and exhaust filter, ultraviolet lamp once a year.
  4. ProClean Extra Oral Suction System: When not in use, store in a dry and ventilated place.
  5.  Replacement Filters & Items:
  • PC2800-SYNFLTR
  • PC2800-HEPA13SML #PC2800-UVLAMP
  • PC2800-PLSMAGEN #PC2800-HEPA13LR #PC2800-MOTOR
  • Replacement washable synthetic fiber filter Replacement HEPA-13 small filter Replacement UV lamp
  • Replacement Plasma Generator Replacement HEPA-13 large filter Replacement Suction Motor

Trouble-shooting

Electric principle diagram

Disposing notes

Following requirements shall be followed as stored:

• Disconnect the appliance from power.
• The components of the product are non-toxic, dispose of the appliance should follow all environmental regulations. Product name Unit Quantity

 Packing list

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

TPC User Manuals

Related Manuals