Dentis SQUVA100 SQ UV Activator User Manual
- June 5, 2024
- Dentis
Table of Contents
Dentis SQUVA100 SQ UV Activator
Purpose of Use & Conditions for Use
Purpose of Use
The SQ UV ACTIVATOR is a product designed to increase hydrophilicity on the
implant surface.
Conditions for Use
- Condition of Use
- Temperature : 0 – 40 ℃
- Relative humidity : 30 – 90 %RH
- Atmospheric pressure : 0 – 2,000 m (106 – 80 Kpa)
- Transport environment
- Temperature : 0 – 40 ℃
- Relative humidity : 0 – 80 %RH
CE & FCC
FCC warning statement
- FCC Part 15.19:This device complies with part15 of FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
- FCC Part 15.105:This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help
- FCC Part 15.21:Changes or modifications not expressly approved by the manufacturer (or party responsible) for compliance could void the user’s authority to operate the equipment
- RF Exposure: This equipment should be installed and operated with a minimum 20 cm between the radiator and your body.
CE statement
- RED Article 10.2 This device can be operated in at least one member state without infringing applicable requirements on the use of radio spectrum.
- RF Range: 13,535 kHz ~13,587 kHz
- Max Output Power(E.I.R.P.): -27.46 dBm
Warnings
Warnings :
- Please read the USER manual before using the product.
- Check all components after opening this product.
- Please install this product in a well ventilated place.
- Do not use it in uneven, sloping, or vibrating area.
- Do not use in water or wet areas.
- Do not touch the body and power plugs with wet hands.
- Do not pull the power cable when disconnecting the power plug, but hold the power plug and disconnect.
- Please connect a grounded wall outlet.
- Use input power in the range of AC 100 – 240 V, 50/60 Hz.
- Do not use adapters other than those provided.
- This product has been certified by the Electromagnetic Compatibility KC, CE and FCC
- This product complies with EM standards. However, electromagnetic interference may cause performance degradation when used improperly.
- Do not disassemble or separate except for experts designated by supplier. (No warranty can be obtained if arbitrarily disassembled or separated)
- If the product displays an ozone warning on the LCD or smells strange (ozone smell), ventilate it immediately, remove the power plug, and contact your supplier.
- Do not put any substances other than FIXTURE in the lifting section. (Risk of failure or electric shock)
- After reforming the FIXTURE, after the pump stops (approximately 30 seconds), change the other FIXTURE.
- Do not insert FIXTURE driver during product operation.
- Make sure to use products within 8mm diameter of FIXTURE.
- Use FIXTURE in combination with the FXIUTRE support.
- Use the FIXTURE support in the correct position on the TRAY.
- Do not place objects on the TRAY.
- Do not place items on top of the product
- When operating the instrument, be careful not to cause external shock
- If the wire is damaged, stop using it and contact your supplier to replace it
- If your product smells of smoke or burning, please remove the power plug and contact your supplier.
- Be careful when installing the product
- Contact your supplier if there is a fault or damage
- If it is not used for a long time, remove it from the outlet and store it
- In case of failure, do not forcibly separate the closed lifting part and contact the place of purchase
- Do not remove the RFID CARD during use
- Please contact the supplier when the RFID CARD exceeds the limit.
- Dispose of the product in accordance with the laws of the country concerned
- Never use this product for any purpose other than its intended use
- The temperature of the FIXTURE driver is high after reforming, so be sure to wear latex gloves
Warranty
- UV LAMP : 700 hrs
- O3 FILTER : 1 years
- Product : 1 years
Components and Specifications
Components
No. | Description |
---|---|
① | SQ UV Activator |
② | Adapter & Power Plug |
③ | FIXTURE Driver KIT |
④ | RFID CARD |
Specifications
Category | Unit | Specification | Remark |
---|---|---|---|
Rated power input | – | AC 100 – 240 V , 50/60 Hz | |
Power consumption | W | 137 |
Adapter Output
| V| DC 24|
W| 250|
Product size| mm| W336 x L185 x H330|
Product weight| kg| 6|
Ozone accrual volume| ppm| Less than 0.05|
Reforming time| s| 10|
Wavelength| nm| 172|
RF Range| kHz| 13535 ~ 13587|
RF Max Output Power(E.I.R.P.)| dBm| – 27.46|
Use
Components
**How to use
**
Control part
Sensor Explanation
※ The “non-contact sensor” at the top of the LCD can detect the user’s hand at a distance of 2 cm and perform “START” action. (Prevent infection by non- contact method)
Storage Method
- Close and store the TRAY after use.
- Don’t twist the power cord and store it.
- Please keep it away from the water.
- In case of an earthquake, do not store on a high shelf.
- After use, turn the power button off.
- If not used for a long time, keep the adapter and main body separately.
Cleaning and Sterilization
Cleaning and Sterilizing Fixture Driver
- The used kits and instruments should be cleaned with distilled or flowing water with a the temperature of 30~40℃, using a brush for 20 seconds to wash away any remaining blood and external materials.
- Tools that have inner parts needs to be cleaned up, such as a torque wrench, should be dismantled to wash away any remaining blood and foreign materials
- Make 1% TergazymeTM mixture mixed 1L 30~40℃ water and TergazymeTM 10mL in the cleaning water tank and keep the kits and instruments in the 1% TergazymeTM mixture for 10 minutes. (using an ultrasonic cleaning machine is more effective)
- The kits and instruments should be washed using distilled or flowing water (30 ~ 40℃) using a brush for cleaning 3 times. Use air blower, wipers or fabrics in order named to remove water on the kits and instruments. Finally, use alcohol cotton to remove water stain on the kits and instruments.
Cleaning and Disinfection the Main Body
- Before and after using the product, sterilize the contents of FIXTURE KIT (FIXTURE support, etc.) and FIXTURE CASE.
- The following substances cannot be used :
- High concentration organic and inorganic acids, chlorinated hydrocarbons
Sterilization
- Before and after using the product, sterilize the FIXTURE driver and FIXTURE case.
- Autoclave by adding FIXTURE case and FIXTURE driver
- Sterilization condition (Please refer to the sterilization methods for each country)
Troubleshooting
No. | Problem | Cause | Solution |
---|---|---|---|
1 | The screen does not turn on | Power failure | Make sure AC power is supplied. |
(AC 100 – 240 V, 50/60 Hz)
| | The adapter is not connected to the
main body
| ****
Connect the adapter to the main body
---|---|---|---
If the power cord is not properly engaged| Check the connection between the
power cord and the adapter.
Connect the power cord to a multi-tap or
AC power source.
Etc.| Please contact your supplier and AS representative.
2| the button does not
work
| LCD issue| Please contact your supplier and AS
representative.
3| The sensor is not operating| Operational Distance Error| When operating the sensor, operate it within 2cm.
4
| ****
The lifting part does not work
| ****
–
| If the TRAY does not work after operating the “OPEN” and “CLOSE” buttons, contact your place of purchase and AS
representative.
5
| ****
“Ozone Detection” error message
| ****
Ozone generation
| Please stop the product and ventilate it. Check that the product is fully ventilated when reusing, stop the product if the same symptoms occur, and contact your
supplier and AS representative.
Please contact the place of purchase and the person in charge of AS for
replacement when the ozone filter has reached its end of life.
6
| “Lamp communication” error message| ****
Nonoperating
| ****
Please contact your supplier and AS representative.
7
| “Actuator location” error message| –| Please contact your supplier and AS
representative.
Etc.| Please contact your supplier and AS representative.
8| Product damage| –| Please contact your supplier and AS representative.
9
| ****
If the RFID CARD is not recognized
| RFID CARD error| Please put the RFID CARD back in.
When CARD availability has
exhausted
| Please recharge your CARD. Please contact supplier and AS
representative.
Etc| Please contact your supplier and AS representative.
10
| ****
Etc.
| ****
–
| Please contact your supplier and AS representative.
References
- 덴티스 | 의료기기 전문기업
- DENTIS | Dental Implant, Surgical LED Lights, Dental 3D Printer
- 덴티스 임플란트 | 행복한 미소, 덴티스 임플란트
- Dentis Implant | Dental Implant Solution