ottobock 4R84 Double Adapter Pyramid Instructions
- June 5, 2024
- ottobock
Table of Contents
ottobock 4R84 Double Adapter Pyramid
Product description
INFORMATION
Date of last update: 2022-03-07
- Please read this document carefully before using the product and observe the safety notices.
- Instruct the user in the safe use of the product.
- Please contact the manufacturer if you have questions about the product or in case of problems.
- Report each serious incident related to the product to the manufacturer and to the relevant authority in your country. This is particularly important when there is a decline in the health state.
- Please keep this document for your records.
Construction and Function
The 4R84 double adapter features a pyramid adapter and pyramid receiver. It
serves as a connector between prosthetic components. Proximal and distal angle
adjustments are possible in the frontal and sagittal planes.
Combination possibilities
This prosthetic component is compatible with Ottobock’s system of modular
connectors. Functionality with components of other manufacturers that have
compatible modular connectors has not been tested.
Intended use
Indications for use
The product is intended exclusively for lower limb exoprosthetic fittings.
Area of application
Approved for a body weight of up to 150 kg.
Environmental conditions
-
Storage and transport
Temperature range –20 °C to +60 °C (–4 °F to +140 °F), relative
humidity 20 % to 90 %, no mechanical vibrations or impacts -
Allowable environmental conditions
- Temperature range: –10 °C to +45 °C (14 °F to 113 °F)
- Chemicals/liquids: fresh water, salt water, perspiration, urine, soapsuds, chlorine water
- Moisture: Submersion: max. 1 h in 3 m depth, relative humidity: no restrictions
- Solids: Dust, occasional contact with sand
- Clean the product after contact with humidity/chemicals/ solids, in order to avoid increased wear and damage (see page 8).
-
Unallowable environmental conditions
- Solids: Highly hygroscopic particles (e.g. talcum), continuous contact with sand
- Chemicals/liquids: Acids, continuous use in liquid media
Lifetime
This product was tested by the manufacturer with 3 million load cycles.
Depending on the user’s activity level, this corresponds to a maximum lifetime
of 5 years.
Safety
Explanation of warning symbols
CAUTION: Warning regarding possible risks of accident or injury.
NOTICE: Warning regarding possible technical damage.
General safety instructions
CAUTION!
Risk of injury and risk of product damage
- Comply with the product’s field of application and do not expose it to excessive strain (see page 6).
- Note the combination possibilities/combination exclusions in the instructions for use of the products.
- Observe the maximum lifetime of the product.
- To prevent mechanical damage, use caution when working with the product.
- If you suspect the product is damaged, check it for proper function and readiness for use.
- Do not use the product if its functionality is restricted. Take suitable measures (e.g. cleaning, repair, replacement, inspection by the manufacturer or a specialist workshop).
NOTICE!
Risk of product damage and limited functionality
- Do not expose the product to prohibited environmental conditions.
- Check the product for damage if it has been exposed to prohibited environmental conditions.
- Do not use the product if it is damaged or in a questionable condition. Take suitable measures (e.g. cleaning, repair, replacement, inspection by the manufacturer or a specialist workshop).
Signs of changes in or loss of functionality during use
Among other factors, changes in functionality can be indicated by an altered
gait pattern, a change in the positioning of the prosthetic components
relative to each other and by the development of noises.
Scope of delivery
Quantity | Designation | Reference num ber |
---|---|---|
1 | Instructions for use | – |
1 | Adapter | – |
4 | Set screw | 506G3=M8x12-V |
Preparing the product for use
CAUTION
Incorrect alignment or assembly
Risk of injury due to damaged prosthetic components
- Observe the alignment and assembly instructions.
CAUTION
Improper assembly of the screw connections
Risk of injury due to breakage or loosening of the screw connections
- Clean the threads before every installation.
- Apply the specified torque values.
- Follow the instructions regarding the length of the screws and about how to secure the screws.
Installation in the modular prosthesis
The pyramid adapter is fixed with the set screws of the pyramid receiver.
- Required materials: torque wrench (e.g. 710D20), 636K13 Loctite 241
- Fitting:
- Screw in the set screws.
- Use the torque wrench to tighten the set screws (10 Nm).
- Definitive assembly:
- Secure the set screws with Loctite.
- Screw in the set screws.
- Pre-tighten the set screws with the torque wrench (10 Nm) and then tighten them (15 Nm).
-
Replace any set screws that are protruding or are recessed too much with appropriate set screws (see selection chart).
Selection table for set screwsReference number| Length (mm)
506G3=M8X12-V| 12
506G3=M8X14| 14
506G3=M8X16| 16
Alignment
The set screws in the pyramid receiver can be used to make static
adjustments during alignment, trial fittings and after the prosthesis is
finished.
Replacement and disassembly
The set position of the prosthetic component can be maintained dur ing
replacement or disassembly. In order to do this, unscrew the two set screws
that are screwed in the furthest and located next to each other.
Cleaning
- Clean the product with a damp, soft
- Dry the product with a soft
- Allow to air dry in order to remove residual
Maintenance
- A visual inspection and functional test of the prosthetic components should be performed after the first 30 days of use.
- Inspect the entire prosthesis for wear during normal consultations.
- Conduct annual safety inspections.
Disposal
In some jurisdictions it is not permissible to dispose of the product with unsorted household waste. Improper disposal can be harmful to health and the environment. Observe the information provided by the responsible authorities in your country regarding return, collection and disposal procedures.
Legal information
All legal conditions are subject to the respective national laws of the country of use and may vary accordingly.
Liability
The manufacturer will only assume liability if the product is used in
accordance with the descriptions and instructions provided in this document.
The manufacturer will not assume liability for damage caused by disregarding
the information in this document, particularly due to improper use or
unauthorised modification of the product.
CE conformity
The product meets the requirements of Regulation (EU) 2017/745 on medical
devices. The CE declaration of conformity can be downloaded from the
manufacturer’s website.
Warranty
The manufacturer warrants this device from the date of purchase. The warranty
covers defects that can be proven to be a direct result of flaws in the
material, production or construction and that are reported to the manufacturer
within the warranty period.
Further information on the warranty terms and conditions can be obtained from
the competent manufacturer distribution company.
Technical data
Reference number | 4R84 |
---|---|
Weight [g] | 115 |
System height [mm] | 36 |
Build height [mm] | 36 |
Material | Titanium |
Max. body weight [kg] | 150 |
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>