ottobock 757M20 Myo Cuff Instruction Manual
- June 5, 2024
- ottobock
Table of Contents
ottobock 757M20 Myo Cuff Instruction Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a practitioner licensed by law of the State in which he/she practices to use or order the use of the device.
Foreword
INFORMATION
Date of last update: 2022-03-29
- Please read this document carefully before using the product and observe the safety notices.
- Instruct the user in the safe use of the product.
- Please contact the manufacturer if you have questions about the product or in case of problems.
- Report each serious incident related to the product to the manufacturer and to the relevant authority in your country. This is particularly important when there is a decline in the health state.
- Please keep this document for your records.
The product “757M20=* Myo cuff” is referred to as the product/cuff below.
These instructions for use provide you with important information on the use,
adaptation and handling of the product.
The product may not be transferred to the patient without prior instruction.
Only put the product into use in accordance with the information contained in
the accompanying documents supplied.
According to the manufacturer (Otto Bock Healthcare Products GmbH), the
patient is the operator of the product according to the IEC
60601-1:2005/A1:2012 standard.
Product description
Function
The product is intended to determine the suitability of a patient for using
Myo Plus control. The product is a flexible armband that is temporarily
applied to the patient’s forearm. This makes it possible to evaluate the
patient’s signal patterns using the Myo Plus app, without fabricating a check
socket. In case of problems with separation of the signal patterns, the
patient can use the product to practise at the facilities of the O&P
professional or in the rehabilitation centre.
The product is a flexible armband that is applied to the end user’s sound
limbs.
The product measures the patient’s control signals and assigns them to the
prosthesis movements.
Design
The product consists of the following components:
- Remote electrode dome
- Textile cuff snap fasteners
- Charging receptacle and battery
- Textile cuff
- Housing for remote electrodes and battery
- Housing and MyoPlus TR
- Nameplate
Explanation of symbols
-
Donning direction (left/right)
The arrow has to point towards the hand. -
Charging receptacle
-
Alignment for the patient’s ulna
-
Electrode position (1–8)
Intended use
Indications for use
The product is to be used exclusively for exoprosthetic fittings of the upper limbs.
Conditions of use
The product is to be used exclusively for the period of end user evaluation and training.
Indications
- For patients with unilateral and bilateral amputations.
- For patients with transcarpal and transradial amputation.
Contraindications
- For patients with transhumeral amputation.
- For patients with partial hand amputation.
- For patients with shoulder disarticulation.
Qualification
Fitting a patient with the product may only be carried out by O&P
professionals, therapists or nursing staff who have been authorised by
Ottobock upon completion of a corresponding training course. Furthermore, the
O&P professional must have the technical qualifications required for the
alignment of a prosthesis together with all necessary settings and
adjustments.
The “O&P professional mode” of the adjustment app may only be used by
qualified personnel, therapists and nursing staff after participating in the
relevant product training and obtaining certification for the application.
Additional product training courses may become necessary to qualify for app
updates.
Safety
Explanation of warning symbols
WARNING Warning regarding possible serious risks of accident or injury.
CAUTION Warning regarding possible risks of accident or injury.
NOTICE Warning regarding possible technical damage.
Structure of the safety instructions
WARNING
The heading describes the source and/or the type of hazard
The introduction describes the consequences in case of failure to observe the
safety instructions. Consequences are presented as follows if more than one
consequence is possible:
- E.g.: Consequence 1 in the event of failure to observe the hazard
- E.g.: Consequence 2 in the event of failure to observe the hazard
- This symbol identifies activities/actions that must be observed/carried out in order to avert the hazard.
General safety instructions
WARNING
Non-observance of safety notices
Injury due to faulty control or malfunction of the product.
Destruction of the product.
- Observe the safety notices and the stated precautions in this accompanying document.
WARNING
Operating the product near medical devices critical for safety
Interference with life-sustaining medical devices (e.g., pacemaker,
defibrillator, heart-lung machine, etc.) due to electromagnetic interference
of the product.
- When operating the product in the immediate vicinity of life-sustaining medical devices, ensure that the minimum distances stipulated by the manufacturer are observed.
- Make sure to observe the operating conditions stipulated by the manufacturer and the safety notices.
WARNING
Use of inappropriate components
Injury due to malfunction of the product.
- Operate the product only with components prescribed by the manufacturer. You can find the list of components in the section “Combination possibilities”.
- Operate the product only with accessories recommended by the manufacturer. You can find the list of components in the section “Scope of delivery and accessories”.
WARNING
Changes or modifications to the product made independently
Injury due to faulty operation or malfunction of the product.
- Have any changes or modifications to the product carried out only by authorised, qualified Ottobock personnel.
- The battery may only be handled by authorised, qualified Ottobock personnel (replacement by the user is not permitted).
- The product and any damaged components may only be opened and repaired by authorised, qualified Ottobock personnel.
CAUTION
Proximity to sources of strong magnetic or electrical interference (e.g.
theft prevention systems, metal detectors)
Injury due to unexpected behaviour of the product caused by interference with
internal data communication.
- Avoid remaining in the vicinity of visible or concealed theft prevention systems at the entrance/exit of stores, metal detectors/body scanners for persons (e.g. in airports) or other sources of strong magnetic and electrical interference (e.g. high-voltage lines, transmitters, transformer stations, computer tomographs, magnetic resonance tomographs, etc.).
- When walking through theft prevention systems, body scanners or metal detectors, watch for unexpected behaviour of the product.
CAUTION
Skin irritation due to inadequate cleaning of the product
Skin irritation due to contact with soiled electrode domes.
- Clean the product only as described in the section “Cleaning and care” (Cleaning and care).
CAUTION
Reuse on another user without corresponding cleaning of the product
Skin irritation due to soiling of the product by a previous user.
- Only use the product on one user for the duration of a fitting.
- Between two fittings, send the product to the Ottobock Service Centre for cleaning.
CAUTION
Insufficient distance to HF communication devices (e.g. mobile phones,
Bluetooth devices, WiFi devices)
Injury due to unexpected behaviour of the product caused by interference with
internal data communication.
- Therefore, keeping a minimum distance of 30 cm from HF communication devices is recommended.
CAUTION
Operating the product in very close proximity to other electronic devices
Injury due to unexpected behaviour of the product caused by interference with
internal data communication.
- Do not operate the product in the immediate vicinity of other electronic devices.
- Do not stack the product with other electronic devices during operation.
- If simultaneous operation cannot be avoided, monitor the product and verify proper use in the existing setup.
CAUTION
Use of a damaged product
Injury due to material damage of the product.
- Inspect the product each time for sharp-edged and pointy parts before putting it on.
- Have the product repaired by an authorised Ottobock Service Centre in case of damage.
NOTICE
Insufficient skin contact of the electrodes
Insufficient information about the suitability of the muscle signals to use
MyoPlus
- Makes sure that the contact surfaces of the electrode domes are placed on intact skin only and have full-surface contact during a test.
NOTICE
Shifting of the electrodes
Insufficient information about the suitability of the muscle signals to use
MyoPlus
- Make sure that the electrodes do not shift during a test, or restart the test.
NOTICE
Incorrect electrode settings due to muscle fatigue
Insufficient information about the suitability of the muscle signals to use
MyoPlus
- Make sure that breaks are taken during calibration.
NOTICE
Penetration of dirt and humidity into the product
Reduction of the informative value of the muscle signal due to product
malfunction.
Skin irritation due to dirt or chemical reactions in combination with
perspiration.
- Make sure that neither solid particles nor liquids can penetrate into the product.
NOTICE
Improper product disposal
Environmental damage due to improper disposal.
- After the end of the service life or in case of total loss of the product, dispose of the product according to the applicable national legal regulations.
Scope of Delivery and Accessories
Scope of Delivery
Items in the scope of delivery or accessories marked with ▀ are application
parts according to the IEC 60601-1:2005/A1:2012 standard.
- 1 pc. 646C107 Bluetooth PIN card
- 1 pc. 757M20=* Myo gaiter includes: 13Z161 electrode dome, flat ▀ or 13Z162 electrode dome, medium ▀ or 13Z163 electrode dome, high ▀
- 1 pc. battery charger 757L35 MyoCharge Integral
- 1 pc. Instructions for use (qualified personnel)
Accessories
- Myo Plus app
- 1 pc. 757P41-1 Y-cable
Charging the battery
We recommend charging on a daily basis. If the device is not used for an extended period, the battery charge level should be checked regularly and the battery recharged.
- Connect the charging plug to the charging receptacle of the product.
- A correct connection between the battery charger and the product is indicated by feedback (see page 14).
- The product is switched off automatically.
- The charging process starts.
- Disconnect the battery charger from the product after the charging process is complete.
Preparing the product for use
Determine electrode position
- Pull the product over the end user’s forearm.
- Position the product 6 to 7 cm distal of the olecranon.
- Position the charging pill laterally.
INFORMATION
Make sure that the electrode domes are in contact with the end user’s skin.
Establishing a connection to a demo prosthesis
Prerequisite for intended use
- Base
- 10S17 Active rotation unit
- Compatible hand prosthesis
- Compatible battery
Note the following steps for establishing a connection:
-
Open the product’s fabric far enough so the cables (two-pin cable and 757P41-1 Y-cable) can be brought to the outside.
-
Refasten the product’s snap fasteners.
-
Connect the projecting cables to the supplied extension cables.
Make sure the cables are respectively connected with the same marking.
-
Connect the other ends of the extension cable to the demo prosthesis.
See the information for connecting the rotation unit in the Myo Plus TR instructions for use. -
Switch on the product and connect to the app.
-
In the menu of the Myo cuff, press the “Connect demo prosthesis” button.
-
Switch the prosthesis on within ten seconds.
INFORMATION
Only use the 757P41-1 Y-cable for the connection.
Myo Plus app
With the Myo Plus app, the user can view and evaluate their signals and the
frequently used functions together with the O&P professional. The
repeatability of movements can be trained with the app. When combined with the
product, the app permits the simulation of bebionic hands or MyoBock hands.
INFORMATION
The Myo Plus app can be downloaded free of charge from the respective online store. To download the Myo Plus app, the QR code on the supplied Bluetooth PIN card can be read with the mobile device (requirement: QR code reader and camera).
System Requirements
See the information in the Apple App Store or Google Play Store regarding compatibility with mobile devices and versions.
Initial connection between MyoPlus app and product
The following points should be observed before the initial connection:
- Bluetooth on the mobile device must be switched on.
- The Bluetooth ID and Bluetooth PIN of the product being connected must be known.
The Bluetooth PIN and Bluetooth ID are found on the enclosed Bluetooth PIN card.
The Bluetooth ID is found on the Myo Plus TR (see fig. 1, item 6) and starts with the letters “BT ID”.
Starting the Myo Plus app for the first time
INFORMATION
Keep mobile app up to date.
- Please contact the manufacturer if you suspect cybersecurity problems.
INFORMATION
Bluetooth on the prosthesis must be turned on in order to use the MyoPlus app.
After turning on Bluetooth, it remains active for approx. 5 minutes. The app
must be started and the connection established during this time.
- Tap the icon for the Myo Plus app .
- The end user licence agreement (EULA) is displayed.
- If rights are requested the first time the app is started, these have to be granted. Otherwise, the app cannot be started.
- Accept the end user licence agreement (EULA). If the end user license agreement (EULA) is not accepted, the Myo Plus app cannot be used.
- The “Bluetooth connection” menu item is opened.
- Tap to establish a connection.
- Select the desired component from the list.
- Enter the Bluetooth PIN code and tap “Connect”.
- The icon is displayed when the connection has been established.
- Once the connection has been established, the data are read from the component. Then the radar chart is displayed.
Use
Evaluating muscle signals in the MyoPlus app
To perform an evaluation of the muscle signals or for training, the following steps have to be completed:
- Press the charging receptacle button and hold for one second.
- The product is switched on.
- Two short beeps are emitted.
- In the app, tap the “Myo Cuff” menu.
- Select the desired hand that was chosen in the planned fitting.
- Now the muscle signals can be evaluated and trained.
Connecting to the demo prosthesis
- Press the Myo cuff button and hold for one second.
- The product is switched on.
- Two short beeps are emitted.
- Establish a connection between Myo Plus and the Myo cuff.
- In the app, tap the “Myo cuff” menu.
- Press the button to establish a connection to the demo prosthesis.
- Switch on the demo prosthesis within 10 seconds.
- Now the muscle signals can be evaluated and trained.
Display of the current charge level
The battery charge level can be queried at any time.
- With the product switched on, press the button on the charging receptacle and hold for less than one second.
- The LED display on the charging receptacle provides information on the current battery charge level (see page 14).
Safety Shutoff
The purpose of the safety shutoff is to protect the battery; it is triggered in case of:
- Excessively high or low temperature during the charging process
- Short circuit
- Overvoltage or undervoltage
INFORMATION
After a safety shutdown, the battery charger must be connected to reactivate the electronics.
Cleaning and Care
CAUTION
Reuse on another user without corresponding cleaning of the product
Skin irritation due to soiling of the product by a previous user.
- Only use the product on one user for the duration of a fitting.
- Between two fittings, send the product to the Ottobock Service Centre for cleaning.
Cleaning the electrode domes
- Clean the electrode domes with a cleaning cloth and 634A58 isopropyl alcohol after each application.
- Dry the electrode domes with a cloth.
Legal information
All legal conditions are subject to the respective national laws of the country of use and may vary accordingly.
Liability
The manufacturer will only assume liability if the product is used in accordance with the descriptions and instructions provided in this document. The manufacturer will not assume liability for damage caused by disregarding the information in this document, particularly due to improper use or unauthorised modification of the product.
Trademarks
All product names mentioned in this document are subject without restriction
to the respective applicable trademark laws and are the property of the
respective owners.
All brands, trade names or company names may be registered trademarks and are
the property of the respective owners.
Should trademarks used in this document fail to be explicitly identified as
such, this does not justify the conclusion that the denotation in question is
free of third-party rights.
CE conformity
Otto Bock Healthcare Products GmbH hereby declares that the product is in
compliance with applicable European requirements for medical devices.
The product meets the requirements of the RoHS Directive 2011/65/EU on the
restriction of the use of certain hazardous substances in electrical and
electronic devices.
This product meets the requirements of the 2014/53/EU directive.
The full text of the regulations and requirements is available at the
following Internet address: http://www.ottobock.com/conformity
Local Legal Information
Legal information that applies exclusively to specific countries is written in the official language of the respective country of use in this chapter.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- This device must accept any interference received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/ TV technician for help.
Any changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Caution: Exposure to Radio Frequency Radiation.
This device must not be co-located or operating in conjunction with any other
antenna or transmitter.
Responsible party:
Otto Bock Health Care, LP
3820 West Great Lakes Drive
Salt Lake City, Utah 84120-7205 USA
Phone + 1-801-956-2400
Fax + 1-801-956-2401
This device complies with RSS 210 of Industry Canada.
Operation is subject to the following two conditions:
- this device may not cause interference, and
- this device must accept any interference, including interference that may cause undesired operation of this device.
Caution: Exposure to Radio Frequency Radiation.
The installer of this radio equipment must ensure that the antenna is located
or pointed such that it does not emit RF field in excess of Health Canada
limits for the general population; consult Safety Code 6, obtainable from
Health Canada’s website
http://www.hc-sc.gc.ca/rpb.
Responsible party:
Otto Bock Healthcare Canada Ltd.
5470 Harvester Road
L7L 5N5 Burlington, Ontario
Canada
Phone + 1-800-665-3327
Caution: Federal law (USA) restricts this device to sale by or on the order of a practitioner licensed by law of the State in which he/she practices to use or order the use of the device
Technical data
Ambient conditions|
---|---
Storage in original packaging| +5 °C/41 °F to +40 °C/104 °F
Max. 85% relative humidity, non-condensing
Transport in original packaging| -25 °C/-13 °F to +35 °C/95 °F
15% to 90% relative humidity, non-condensing
+35 °C/95 °F to +70 °C/158 °F
Water vapour pressure up to 50 hPa
Storage between subsequent applications| -25 °C/-13 °F to +35 °C/95 °F
15% to 90% relative humidity, non-condensing
+35 °C/95 °F to +70 °C/158 °F
Water vapour pressure up to 50 hPa
Operation| +5 °C/41 °F to +40 °C/104 °F
15% to 90% relative humidity, non-condens ing; air pressure 533 hPa to 1060
hPa
Charging the battery| 0°C/+32°F to +45°C/+113°F
General|
---|---
Reference number| 757M20=*
Inner diameter| 45 mm
Exterior diameter| 110 mm
Weight| 210 g
Textiles used| 80% polyamide (PA)
20% elastane (EL)
Service life| 5 years
Battery of the product| 757B35=0
---|---
Battery type| Li-Po
Output voltage| Approx. 7.4 V DC
Charging voltage| Approx. 8.2 V DC
Capacity| 300 mAh
Dimensions of battery cells| 35 x 20 x 6.5 mm per cell
Dimensions of protection electronics| 22 x 17 x 5 mm
Weight| 25 g (battery without accessories)
Operating time| 7.5 h
Charging time until battery is fully charged| 2.0 h
Product behaviour during the charging process| The product is non-functional
Data transfer|
---|---
Wireless technology| Bluetooth Smart Ready
Range| min. 3 m/9.84 ft
Frequency range| 2402 MHz to 2480 MHz
Modulation| GFSK, π/4 DQPSK, 8DPSK
Data rate (over the air)| 2178 kbps (asymmetrical)
Maximum output power (EIRP):| +8.5 dBm
Appendices
Symbols Used
Type BF applied part
Declaration of conformity according to the applicable European directives
Manufacturer
In some jurisdictions it is not permissible to dispose of these products with
unsorted household waste. Disposal that is not in accordance with the
regulations of your country may have a detrimental impact on health and the
environment.
Please observe the instructions of your national authority pertaining to
return and collection.
Serial
number (YYYY WW NNN)
YYYY – year of manufacture
WW – week of manufacture
NNN – sequential number
Article number
Operating States
Status signals
Charging receptacle | Event |
---|---|
LED shows green light | Battery fully charged |
LED shows yellow light | Battery 50% charged |
LED shows orange light | Battery empty |
Beep signals
The following beep signals are set by default:
Beep signal | Additional display | Event |
---|---|---|
1 x long | – |
- Turn off product on charging receptacle
- Charging begins (charging plug connected to charging receptacle)
- Charging ends (charging plug disconnected from charging receptacle)
2 x short| LED on charging receptacle lights up briefly| Switching on the
product
3 x short| –| Battery voltage too low, product shuts off auto matically
Troubleshooting
Event | Cause | Required action |
---|
Unexpected display of the recognised movement in various arm positions, e.g. during overhead use| Use in this arm posi tion was not taken into account during calib ration, or was incor|
- Note arm position during calibration
- Take the weight into account during calibration
- Adjust the values in the app under “Settings > Advanced Settings”
| rectly configured in the app.|
–| General fault|
- Turn the Myo gaiter off and back on again
Unexpected display of the recognised movement in case of unfavourable resid ual limb conditions| No contact between the electrodes and skin|
- The Myo gaiter has to fully contact the residual limb
Myo gaiter slipping out of place
Directives and manufacturer’s declaration
Electromagnetic environment
This product is designed for operation in the following electromagnetic environments:
- Operation in a professional healthcare facility (e.g. hospital, etc.)
- Operation in areas of home healthcare (e.g. use at home, use outdoors)
Observe the safety notices in the section “Information on proximity to certain areas” (Information on Proximity to Certain Areas).
Electromagnetic emissions
Interference meas urements| Compliance| Electromagnetic
environment directive
---|---|---
HF emissions accord ing to CISPR 11| Group 1/class B| The product uses HF
energy exclusively for its internal functioning. Its HF emissions are
therefore very low, and interference with neigh bouring electronic devices is
unlikely.
Harmonics according to IEC 61000-3-2| Not applicable – power below 75 W| –
Voltage fluctu ations/flicker accord ing to IEC 61000-3-3| Product meets the
requirements of the standard.| –
Electromagnetic interference immunity
Phenomenon| EMC basic standard or test procedure|
Interference immunity test level
---|---|---
Electrostatic dis charge| IEC 61000-4-2| ± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
High-frequency elec tromagnetic fields| IEC 61000-4-3| 10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
Magnetic fields with rated power frequen cies| IEC 61000-4-8| 30 A/m
50 Hz or 60 Hz
Electrical fast transi ents/bursts| IEC 61000-4-4| ± 2 kV
100 kHz repetition rate
Surges
Line against line| IEC 61000-4-5| ± 0.5 kV, ± 1 kV
Conducted interfer ence induced by high- frequency fields| IEC 61000-4-6| 3 V
0.15 MHz to 80 MHz
6 V in ISM and amateur frequency bands between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Voltage drops| IEC 61000-4-11| 0% UT; 1/2 period
At 0, 45, 90, 135, 180, 225, 270 and 315 degrees
0% UT; 1 period and
70% UT; 25/30 periods Single phase: at 0 degrees
Voltage interruptions| IEC 61000-4-11| 0% UT; 250/300 periods
Interference resistance against wireless communication devices
Test fre quency [MHz]| Frequency band [MHz]| Radio
ser vice| Modulation| Maximum power [W]| Distance [m]|
Interfer ence immunity test level [V/m]
---|---|---|---|---|---|---
385| 380 to 390| TETRA 400| Pulse modu lation
18 Hz| 1.8| 0.3| 27
450| 430 to 470| GMRS 460,
FRS 460| FM
± 5 kHz deviation
1 kHz sine| 1.8| 0.3| 28
710| 704 to 787| LTE band 13,
17| Pulse modu lation
217 Hz| 0.2| 0.3| 9
745
780
810| 800 to 960| GSM 800/90 0,
TETRA 800,
iDEN 820,
CDMA 850, GSM 800/90 0,
LTE band 5| Pulse modu lation
18 Hz| 2| 0.3| 28
870
930
1,720| 1,700 to
1,990| GSM 1800;
CDMA 1900;
GSM 1900; DECT;
LTE band 1,
3, 4, 25; UMTS| Pulse modu lation
217 Hz| 2| 0.3| 28
1,845
1,970
2,450| 2,400 to 2,570| Bluetooth WLAN 802.1
1 b/g/n,
RFID 2450
LTE band 7| Pulse modu lation 217 Hz| 2| 0.3| 28
5,240| 5,100 to 5,800| WLAN 802.1
1 a/n| Pulse modu lation 217 Hz| 0.2| 0.3| 9
5,500
5,785
The product is covered by the following patents:
USA: US 9,566,016 B2
Patents pending in: Germany
Otto Bock Healthcare Products GmbH
Brehmstraße 16
1110 Wien
Austria
T +43-1 523 37 86
F +43-1 523 22 64
info.austria@ottobock.com
www.ottobock.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>