Peak Tech 5306 PH/EC/Temp-Meter User Manual
- June 5, 2024
- Peak Tech
Table of Contents
5306 PHECTemp-Meter
User Manual
Safety precautions
This product complies with the requirements of the following directives of the
European Union for CE conformity: 2014/30/EU (electromagnetic compatibility),
2011/65/EU (RoHS).
Damage caused by failure to observe the following information is excluded from
claims of any kind. For the operational safety of the device, the following
safety instructions for operating the device must be observed.
- Read these instructions carefully and make them accessible to a subsequent user
- Use this device only within its scope and specifications.
- Take all necessary safety precautions, especially when using acids.
- Wear personal safety equipment, gloves, and safety glasses when handling chemical products.
- Never operate the device if it is not completely closed.
- Do not operate the device near strong magnetic fields (motors, transformers, etc.)
- Avoid strong vibrations of the device Before starting operation, the device should be stabilized to the ambient temperature (Important when transporting cold to warm rooms and vice versa)
- Do not make any technical changes to the device
- Only qualified service technicians may open the device, as well as perform maintenance and repair work
Cleaning the cabinet
Clean only with a damp soft cloth and a commercially available mild household
cleanser. Ensure that no water gets inside the equipment to prevent possible
shorts and damage to the equipment.
Introduction
The PeakTech 5306 is used to measure the pH, conductivity, and temperature of
various liquids and liquids. With the removable probe, it is possible after
the respective measurement to thoroughly free the device and the probe from
residues of the liquid and to clean them. Calibration can be carried out via
two potentiometers using the
attached screwdriver.
To protect the measuring electrode, a protective cap is attached to the end of
the P 5306 (this is removed during the measurement).
- Digital pH meter with illuminated display
- Conductivity measurement in µS/cm
- Conductivity is shown as µS x10
- Liquid temperature measurement (°C/°F)
- Multi-Line, 3½-digit LC-Display
- Can be calibrated with 2 calibration points
- Exchangeable measuring electrode (dry stored)
- Waterproofed case
- Electrode protective cap
Controls
Measuring operation
The pH meter is used to determine the pH values of various liquids. Thanks to
the display, which has a backlight, and the intuitive handling of the device,
the measurement of the pH value can be carried out quickly and in a user-
friendly manner.
Carrying out a pH measurement
In order to carry out the pH measurement, the protective cap protecting the
electrode of the probe must be removed. Then the device must be switched on
with the on / ff switch. The background lighting is permanently on during
operation in order to highlight the measured value. If the pH meter has not
been used for a long time, has been in operation for a long time or if it has
been used in environments in which stronger external influences can occur, it
is advisable to calibrate the device (see point 4.2).
Before measuring, clean the probe with distilled water in order to exclude
possible falsifications of the measured value. Carefully dry the probe with a
cloth.
Now the probe of the pH meter can be immersed in the liquid to be measured.
Make sure that only the bottom 4 cm of the probe is immersed in the liquid.
After completing the measurement, remove the pH meter from the liquid and
clean the probe again with distilled water.
If no further measurements are to be made, put the protective cap of the
electrode back onto the end of the probe to protect it from contamination or
damage.
Proceed as follows:
- Remove the protective cap from the electrode.
- Clean the electrode with distilled water, and dry it well.
- Press the “power” button to turn on the device.
- Immerse the electrode into the test solution, and stir lightly.
- Press EC/pH to choose “EC” or “pH” mode, Take a few seconds for final stable reading.
- The EC value is shown as µS x10, so multiply the shown value with 10 to receive the actual measurement.
- Wash electrode after use. Press “ON/OFF” to turn it off.
Calibration of the device
4.2.1 PH calibration
To calibrate the pH value, a standard buffer solution pH6.86 is required, for
the EC value a standard buffer solution EC 12880 µS / cm, which can be
obtained from specialist dealers. A pH 4.01 and pH 6.86 buffer solution and an
EC 1413 µS / cm solution can be used for the subsequent control. Calibration
is carried out using a potentiometer in the attery compartment. A
corresponding screwdriver is included as an accessory.
- Immerse the electrode into pH 6.86 (under the temperature of 25℃) standard calibration solution. Stir lightly.
- Press EC/pH to choose pH. Turn the “pH7” trimmer (right potentiometer front view) with a screwdriver until the value corresponds to the 6.86pH.
- Recheck: Clean the electrode with distilled water. Immerse the electrode in pH4.01or pH9.18 the standard buffer solution. After about one minute, the value corresponds to 4.01pH or pH9.18.
- Showing value and buffer solution’s value is within the error range.
Note: For even more precise calibration, fill the respective buffer solution into two different beakers. One is only used to clean the electrode and the other is only used for calibration. This allows the contamination by foreign matter to be reduced to a minimum.
Note: Perform a calibration if:
- the electrode has been replaced.
- no calibration has been carried out for a long time.
- the device has been used often and for a long time.
- particularly high measurement accuracy is required
EC calibration
- Press “EC/pH” to choose “EC” mode. Immerse EC electrode into 12880us/cm standard buffer solution, and stir lightly.
- Adjust “EC” button until it shows 12880us/cm when the showing value is stable, Wash the electrode in distilled water and dry it well.
- Immerse the electrode into 1413us/cm standard solution. If the reading is within the allowed error range, calibration is finished.
NOTE:
- Stir the electrode while testing, to avoid bubbles occurring on the surface of the electrode. Bubbles on the electrode have a bad effect on the accuracy of value.
- If the error value is large or the EC value not go back to 0 us/cm. Immerse the electrode into ethyl alcohol and stir for 5 to10 seconds.
Battery replacement
The PeakTech 5306 is a battery-operated measuring device. After a certain period of use of the device, it may happen that the capacity of the batteries is used up and new batteries have to be inserted. This becomes visible when the display gets darker or looks a little blurry. To change the batteries, make sure that the device is switched off. Remove the upper screw cap of the device and remove the batteries from the battery compartment. Now insert the new batteries of the appropriate polarity (which is shown in the housing) into the battery compartment. Make sure to use the correct batteries that the measuring device requires (4 x 1.5V AG13 button cells). After changing the battery, put the screw cap on the device and screw it back on.
Specifications
Display | 3 1/2-digit, LCD-disp ay max 1999 counts |
---|---|
Functions | PH: |
0.00 to 14.00 pH | Temperature: 0.0°C to 50.0°C |
0 to 19999 pS/cm
Resolution| 0,01 pH| 0,1°C| 10 pS/cm
Accuracy| ± 0,1pH| ± 1 °C| +/- 2% full scale
Sampling rate| 1 — 2 x Sec.
Calibration| With potentiometer
Operation temperature| 0°C…+50°C (32°F…122°C); < 80% RH
Power Supply| 4 x 1,5V Batteries (AG-13)
Dimensions (WxHxD)| 188 x 35 x 35 mm
Weight| 80 g
Accessories| Manual, 4 x 1,5V Batteries (AG-13), Screwdriver
Notification about the Battery Regulation
The delivery of many devices includes batteries, which for example serve to
operate the remote control. There also could be batteries or accumulators
built into the device itself. In connection with the sale of these batteries
or accumulators, we are obliged under the Battery Regulations to notify our
customers of the following:
Please dispose of old batteries at a council collection point or return them
to a local shop at no cost. The disposal in domestic refuse is strictly
forbidden according to the Battery Regulations. You can return used batteries
obtained from us at no charge at the address on the last side in this manual
or by posting with sufficient stamps. Contaminated batteries shall be marked
with a symbol consisting of a crossed-out refuse bin and the chemical symbol
(Cd, Hg or Pb) of the heavy metal which is responsible for the classification
as a pollutant:
- “Cd” means cadmium.
- “Hg” means mercury.
- “Pb” stands for lead.
All rights, also for translation, reprinting, and copy of this manual or parts
are reserved.
Reproduction of all kinds (photocopy, microfilm or other) only by written
permission of the publisher.
This manual considers the latest technical knowledge. Technical changings
which are in the interest of progress are reserved.
We herewith confirm that the units are calibrated by the factory according to
the specifications as per the technical specifications.
We recommend calibrating the unit again, after 1 year.