APx500 Software User Guide
- June 5, 2024
- Audio Precision
Table of Contents
- Summary
- The APx500 Plugin Framework
- Installing a Plugin Measurement
- Initiating the API Reference
- Adding a Measurement
- Showing a Measurement
- Generator Settings
- Analyzer Settings
- Results
- Deriving a Custom Result
- Custom Result Settings
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Controlling APx500 Plugin
Measurements Via the API
Audio Precision —Tech Guide // Benjamin Kilada
Summary
The APx500 software comes with many built-in measurements. Additional
measurements, known as plugin measurements, can be added to the software by
running installers made available on the Audio Precision (AP) website. Plugin
measurements integrate directly with the APx500 software, allowing users to
take advantage of built-in APx500 features such as pass/fail limits,
sequencer control, derived results, and reporting.
Like built-in measurements, plugin measurements can be controlled
programmatically via the software’s Application Programming Interface (API).
However, unlike built-in measurements, all plugin measurements derive from the
Signal Acquisition measurement and are controlled via the Signal Acquisition
APIs. This document provides an overview of how to do this.
Note: All API examples are provided in the Visual C# programming
language. For the example code to function, the related plugin must first be
installed on the PC on which the code is executing.
The APx500 Plugin Framework
In the 4.2 release of the APx500 software, AP added a preliminary plugin framework. This framework consists of infrastructure within the APx software that allows custom measurements and analysis features to be added independently of an official APx software release. The first AP product to take advantage of this framework was the Speech Transmission Index (STI) plugin, introduced in October 2015. Further enhancements to the plugin framework were added in the APx 4.3 software release, and then again in APx500 v4.5, when the application transitioned from 32-bit to 64-bit.
Note: Plugins of version “2.x” are 64-bit and are incompatible with APx500
software version 4.4 and earlier.
The plugin framework allows for both custom measurements and custom-derived
results to be added to the APx500 software. (In this context, custom refers to
measurements and derived results that are not natively part of the APx500
software).
Custom measurements derive from the Signal Acquisition measurement and
typically have some of the same generator and analyzer controls (for example,
generator level or acquisition length) and primary results (Acquired
Waveform). In addition, plugin measurements may have their own unique
generator and analyzer controls (custom controls) and primary results (custom
results). The controls and results that are shared with Signal Acquisition can
be accessed using standard Signal Acquisition APIs, whereas custom controls
and results can be modified by indexing into a collection of properties using
the control or result name. This is described with examples in the following
sections. Custom-derived results can be of type meter or XY and often have
custom controls (controls that appear below a derived result graph) that can
be accessed using “get” and “set” APIs. This is described in sections eight
and nine.
Installing a Plugin Measurement
Download and install the desired plugin(s) using the installer(s) provided on the AP website, ensuring that the APx500 application is first closed. Plugin compatibility with Px500 software versions is discussed on the download page of each plugin.
Initiating the API Reference
When controlling a plugin measurement from the API, it is necessary to first instantiate the APx500 class that represents the APx500 software application. Doing so will launch the APx500 software if closed. To access this class, you must first reference the Audio Precision API Dynamic Link Library (DLL) that is installed with the APx500 application. For more information on referencing the API DLL, please refer to the materials provided in the API developer tools, available on the AP website. Note that in most cases, it is necessary to initialize the application into Sequence Mode, as most plugins are only available in this operating mode. An example of how to do this is shown below. APx500 APx = new APx500(APxOperatingMode.SequenceMode);
Adding a Measurement
To add a plugin measurement to the current project, the add measurement method must be called using the string name of the measurement as the second argument. Plugin measurement names can be found in the add-measurement dialog in the APx500 graphical user interface (GUI). This dialog can be opened using the button highlighted in the image shown below.
An example of how to add the STI Speech Level measurement (highlighted in the
image above) is shown below.
APx.AddMeasurement(0, “Speech Level (IEC 60268-16J)”);
Note: Adding a measurement also makes it visible.
Showing a Measurement
To show a measurement that already exists in the project, the “show measurement” method must be called using the string name of the plugin measurement as demonstrated below. APx.ShowMeasurement(0, “Speech Level (IEC 60268-16J)”); With the plugin measurement added and visible, we can now modify the measurement settings.
Generator Settings
The Signal Acquisition measurement has several built-in generator settings
that can be configured via the
API. These settings include:
- Waveform
- Level
- Offset
- Loop Waveform
- Channel enable/disable
When these settings are shown in a plugin measurement, they can be gotten or
set via the built-in APx.SignalAcquisition.Generator APIs, are discoverable
via the API browser application. An example of how to use these APIs is shown
below.
APx.SignalAcquisition.Generator.Waveform = “Default 48 kHz.wav”;
APx.SignalAcquisition.Generator.Levels.SetValue(OutputChannelIndex.Ch1, “1
Vrms”);
Generator settings that are unique to plugin measurements (and not available
in Signal Acquisition) can be gotten or set via the Signal Acquisition
“GeneratorProperties” collection. To do so, the appropriate get or set method
must be called based on the type of control that is being modified. A
breakdown of when to use each type of method is included below.
Note: All set methods take the string name of the control as the first
argument, and the value to be set as the second argument. Get methods to take
only the string name of the control as the argument
A drop-down pick list is of type string and can be set to any value in the
pick list via the SetString method.
Similarly, the current value can be obtained using the GetString method. For
example, the Frequency control in the Attack and Recovery plugin measurement
can be gotten and set in the following manner: string freq =
APx.SignalAcquisition.GeneratorProperties.GetString(“Frequency”); APx. Signal
Acquisition Generator Properties Set String (“Frequency”, “1.6 kHz”);
A numerical field that accepts fractional numbers is of type double and can be set to any whole or fractional number that is within the range of accepted values via the Set Double method. Similarly, the current value can be obtained using the Get Double method. For example, the High Time control in the Attack and Release measurement can be gotten and set in the following manner: double high time = APx.SignalAcquisition.GeneratorProperties.get double(“High Time”); APx Signal Acquisition Generator Properties Set Double (“High Time”, 1.5);
Note: There are currently no plugin measurements with generator properties of type Integer, Bool, Bool Array or Enum.
Analyzer Settings
The Signal Acquisition measurement has several built-in analyzer settings that can be configured via the API. These settings include:
- Acquisition Type
- Acquisition Length
- Trigger Type
- Delay Time
- Save To File
When these settings are shown in a plugin measurement, they can be modified
via the built-in
APx.SignalAcquisition APIs, are discoverable via the API browser application.
An example of how to use these
APIs is shown below.
APx.SignalAcquisition.AcquisitionType = AcqLengthType.Seconds;
APx.SignalAcquisition.AcquisitionSeconds = 10;
APx.SignalAcquisition.TriggerType = TriggerType.Generator;
APx.SignalAcquisition.DelayTimeInSeconds = 0.1;
APx.SignalAcquisition.SaveToFileSettings.SaveAcquisitionToFile = true;
Analyzer settings that are unique to a plugin measurement (and not available
in Signal Acquisition) can be gotten or set via the Signal Acquisition
properties collection. As with generator properties, the appropriate get or
set method must be called based on the type of control that is being modified.
A breakdown of when to use each type of method is included below.
A checkbox is of type boolean and can be set to either true (checked) or false
(unchecked) via the SetBool method. Similarly, a checkbox value can be
obtained using the GetBool method. For example, the Auditory Effects checkbox
in the Speech Transmission Index (STI) measurement can be gotten and set in
the following manner:
bool use effects = APx.SignalAcquisition.Properties.GetBool(“Auditory
Effects”);
APx.SignalAcquisition.Properties.SetBool(“Auditory Effects”, true);
A drop-down pick list is of type string and can be set to any value in the
pick list via the SetString method.
Similarly, the value can be obtained using the GetString method. For example,
the Frame Length control in the STI Speech Level plugin measurement can be
gotten and set in the following manner: string frame length =
APx.SignalAcquisition.Properties.GetString(“Frame Length”, “20 ms”);
APx.SignalAcquisition.Properties.SetString(“Frame Length”, “20 ms”);
A numerical field with no decimal places is of type integer and can be set to
any whole number that is within the range of accepted values via the setting
method. Similarly, the current integer value can be obtained using the GetInt
method. For example, the Number of Trials controlled in the ABC-MRT plugin
measurement can be gotten and set in the following manner:
int numTrials = APx.SignalAcquisition.Properties.GetInt(“Number of Trials”);
APx.SignalAcquisition.Properties.SetInt(“Number of Trials”, 16);
A numerical field that accepts fractional numbers is of type double and can be
set to any whole or fractional number that is within the range of accepted
values via the double method. Similarly, the value can be obtained using the
GetDouble method. For example, the Attack Settling Tolerance control in the
Attack and Release measurement can be gotten and set in the following manner:
double tol = APx.SignalAcquisition.Properties.GetDouble(“Attack Settling
Tolerance”);
APx.SignalAcquisition.Properties.SetDouble(“Attack Settling Tolerance”,
1.5);
Note: There are currently no plugin measurements with analyzer properties of type Bool Array or Enum.
Results
Results in a plugin measurement can be modified by indexing into the Signal
Acquisition graphs collection via the string name of the graph (or the integer
index). For example, the Speech Level measurement’s Frame Levels result can
be included in the active sequence using the following API.
APx.SignalAcquisition.Graphs[“Frame Levels (A-weighted)”].Checked = true;
To access the settings that pertain to a particular result type (for example,
y-axis settings on an XY graph), the result must first be cast to the
appropriate type using the “Result.As[GraphType]” APIs. An example of how to
cast a result to type XY graph and modify the y-axis unit is shown below.
IXYGraph frameLevelsXY = APx.SignalAcquisition.Graphs[“Frame Levels
(A-weighted)”].
Result.AsXYGraph(); frameLevelsXY.YAxis.Unit = “dBv”;
Similarly, a meter graph can be cast using the AsMeterGraph() method shown
below:
IMeterGraph rmsLevelMeter = APx.SignalAcquisition.Graphs[“RMS Level
(A-weighted)”].
Result.AsMeterGraph(); RMS level meter.Axis.Unit = “dBV”;
If the plugin has an Acquired Waveform result, the result can also be accessed
using the standard Signal
Acquisition APIs as shown below. No casting is necessary.
APx.SignalAcquisition.AcquiredWaveform.YAxis.Unit = “V”;
Deriving a Custom Result
There are a few plugins that add derived result types to the picklists of measurement results. For example, installing the Octave Analysis plugin enables you to derive an “Octave Spectrum” result from a Power Spectral Density result in an FFT or Transfer Function measurement.
Plugin-derived results can be added using the AddDerivedResult method by
providing the string name of the derived result as the argument. An example of
how to do so is shown below.
APx. Benchmade Measurements transfer function Power Spectral Density.
AddDerivedResult(“Octave Spectrum”);
Custom Result Settings
Although the API shown above does add a derived result, a custom result’s
settings can only be accessed after casting the result to the appropriate
custom processing result type. For example, the Octave Spectrum-derived
result (which is an XY graph) should be cast to type CustomProcessingXyResult.
An example of how to do this is indicated below.
I Custom Processing Result octave Spectrum = APx.BenchMode.Measurements.
TransferFunction.PowerSpectralDensity.AddDerivedResult(“Octave
Spectrum”).Result.
AsCustomProcessingXyResult();
Once cast to the appropriate type, a result’s settings can be modified using
its properties collection and the string name of the property. An example of
how to get and set the Octave Bandwidth property of the
The octave Spectrum result is shown below.
string bandwidth = octaveSpectrum.Properties.GetString(“Octave Bandwidth”);
octaveSpectrum.Properties.SetString(“Octave Bandwidth”, “1/12 Octave”);
For more information on what get and set methods to use for each result set,
please refer to sections five and six, above.
© 2022 Audio Precision, Inc/All Rights Reserved.