RebornEndo M-Cure 9 LED Dental Cure Lamp Curing Light Instruction Manual

June 1, 2024
RebornEndo

RebornEndo-logo

RebornEndo M-Cure 9 LED Dental Cure Lamp Curing Light

Product Information

Specifications

  • Model: M-Cure 3, M-Cure 8, M-Cure 9
  • Size: 228.52626mm (for all models)
  • Weight: 126g (for all models)
  • Structure and Components: Main unit, top of the main unit, LED, light hood, bracket, battery, adapter, disposable sleeve
  • Battery: 3.7V/1500mAh
  • Display screen: OLED
  • Circuit structure: Control chip, OLED drive circuit
  • Working modes: M-Cure 3 – Five modes, M-Cure 8 – Five modes, M-Cure 9 – Eight modes

Product Usage Instructions

Principle and Usage

The curing light adopts the principle of ray radiation to solidify the light- sensitive resin by shooting at it in a short time.
This device is used to restore teeth. It should only be used by qualified and well-trained dentists on dental patients in hospitals or professional medical sites.

Usage Guidelines

  1. Ensure the device is fully charged before use.
  2. Place the disposable sleeve on the device.
  3. Select the appropriate working mode based on the treatment required.
  4. Aim the light at the target area and activate the curing process.
  5. Follow standard dental procedures for curing and resin solidification.

Frequently Asked Questions

Q: Can this device be used by anyone other than a qualified dentist?

A: No, this device should only be operated by dentists who are qualified and well-trained in dental procedures.

Q: How many times can the device be used on a full battery charge?

A: The device can be used continuously for more than 400 times (based on 10s light per time) with a fully charged battery.

Product Description

Product Introduction

1s Curing Light (M-Cure3/M-Cure6/M-Cure8/M-Cure9) adopts the principle of ray radiation to solidify the light-sensitive resin by shooting at it in a short time.

The main features of 1s Curing Light are:

  1. Four models of 1s Curing Light with eight working modes in total.
  2. Constant output optical power. And the curing effect will not be affected due to the decrease in battery power.
  3. 1s fast curing light Efficient electro-optic conversion efficiency. The device can be used continuously for more than 400 times (based on 10s light per time) with a fully charged battery.

Model and Specification

Model: M-Cure 3、M-Cure 8、M-Cure 9

Type M-Cure 3 M-Cure 8 M-Cure 9
Size 228.52626mm 228.52626mm 218.52626mm
Weight 126g 126g 126g
Structure and components Mainly composed of the main unit, the top of the
main unit, LED, light hood, bracket, battery, adapter, and disposable sleeve.

Mainly composed of the main unit, the top of the main unit, LED, light hood, bracket, battery, adapter, and disposable sleeve.| Mainly composed of the main unit, top of the main unit, LED, light hood, bracket, battery, adapter, and disposable sleeve.
Specification| Battery:3.7V/1500mAh Display screen: OLED| Battery:3.7V/1500mAh Display screen: OLED| Battery:3.7V/1500mAh Display screen: OLED
Circuit structure| Circuit Control chip, OLED drive circuit| Circuit Control chip, OLED drive circuit| Circuit Control chip, OLED drive circuit
Working mode| Five working modes: Efficient curing mode: M1 Deep curing mode: M2 High power mode: M3、Fast curing mode Caries checking mode)| Five working modes: Fast curing mode Soft mode、 Pulse mode Ortho mode Caries checking mode)| Eight working modes: Efficient curing mode: M1 Deep curing mode: M2 High power mode: M3 Fast curing mode Soft mode Pulse mode Ortho mode Caries checking mode)

Top of the main unit

| | |

Structure and Components

  • The curing light is mainly composed of the main unit, the top of the main unit, the light hood, the bracket, the battery, the adapter, disposable sleeve.

Principle and usage

  • 1s curing light adopts the principle of ray radiation to solidify the light-sensitive resin by shooting at it in a short time. This device is used to restore teeth.
  • The device can only be used by a dentist who is qualified and well-trained. This device is used on dental patients in the place of a hospital or professional medical site.

Contraindications

  • It is forbidden to use this instrument in the following situations:
  • The doctor with a pacemaker is disabled.
  • Patients with cardiac pacemakers (or other electrical equipment) are warned not to use small appliances (such as Electric razors, hair dryers, etc.) patients are disabled.
  • The patient is sensitive to light.
  • Hemophilia patients are banned.
  • Use with caution in patients with heart disease, pregnant women, and young children.
  • Product Configuration
  • Appearance IdentificationRebornEndo-M-Cure-9-LED-Dental-Cure-Lamp-Curing-Light-fig-4

Components and AccessoriesRebornEndo-M-Cure-9-LED-Dental-Cure-Lamp-
Curing-Light-fig-5

Warning

  • This section contains descriptions of serious side effects and potential safety risks for the product itself and users/patients. The following warning clauses must be read before use:
  • The instrument can only be used by professional dentists under national regulations.
  • The instrument can only be used in designated places and cannot be used outdoors.
  • The instrument requires special precautions regarding electromagnetic compatibility (EMC) and must be in strict accordance with the EMC information for installation and use.
  • Do not use this equipment, especially in the vicinity of fluorescent lamps, radio transmitting devices, remote control devices, and handheld and mobile high-frequency communication devices.
  • It is forbidden to expose the instrument directly or indirectly to a heat source. The instrument must be operated and stored in a safe environment.
  • It is forbidden to use it under conditions with flammable anesthetic mixtures.
  • Do not immerse it in liquid.
  • Only suitable for original accessories.
  • Please use an original power adapter. Another power adapter will result in damage to the lithium battery and control circuit.
  • Do not use damaged or defective instruments.
  • Do not carry out any repairs or modifications without authorization. In case of failure, please contact your local dealer and do not allow unauthorized personnel to repair.
  • Please do not make any changes to the device. Any changes may violate safety regulations, causing harm to the patient. There will be no promises of any modification.
  • Please confirm whether the file is well installed and locked before starting the motor handpiece.
  • The instrument complies with electromagnetic compatibility standards, but it is still necessary to prevent possible risks caused by electromagnetic interference.
  • Please remove the battery if the motor handpiece is not likely to be used for some time.

Instrument Safety Classification

  • Type of operation mode: Continuous operating device.
  • Type   of protection against electric shock: Class II equipment with internal power supply.
  • Degree of protection against electric shock: B-type applied part.
  • Degree of protection against harmful ingress of water: Ordinary equipment (IPX0).
  • Degree of safety application in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide: Equipment cannot be used in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide.

The Main Technical Parameters

  • Lithium battery: 3.7V/ 1500mAh
  • Power Adapter Input ~100V-240V 50Hz/60Hz 0.5A
  • Output: DC 5V/ 2A
  • Light source:10W high power
  • Wavelength: 385nm~515nm
  • Class: class 2
  • Check method: When the operating machine is properly, the LED luminous means the LED is in good condition.
  • The wavelength of our curing light machine can match with dental resin material which is commonly used in clinical, such as 3M and Dentsply.

Work condition:

  1. Environment temperature: 0°C to +40°C
  2. Relative humidity: 30%~80%
  3. Atmosphere pressure: 70kPa to106kPa

Product Installation

a) The installation of the top of the main unit

  • Aim for the mounting holes on the upper of the main unit, and make the top of main unit inserted into it (must screw the top of main unit to the end, don’t inclined inserted).
  • The top of the main unit can rotate 360 degrees.

b) The installation of the disposable sleeve

Take out the disposable sleeve and put it on the top of the main unit.

WARNING:

  • The disposable sleeves are single patient use only.
  • Discard used disposable sleeves in standard waste after each patient.

c)The installation of a light hood

  • Clip the notch of the light hood into the head of the top of the main unit, and then clamp the light hood down on the main unit.

d)Replace the battery

  • When replacing the battery, please open the battery cover, first remove the lithium battery, and then gently pull out the plug-in, put the new battery connector plug correctly into the host, and then put the new battery in.
  • Put the battery cover on the host mounting port, push it up, and cover the battery.

e) Battery charging

  • The original charge must be used for charging. Plug the connector of the power adapter into the charging connector of the host when charging.
  • The host sends out a ‘DI’prompt tone, and the power indicator lights up at the same time. The status of the breathing lamp is gradually brightened and faded flashing.
  • The charging symbol on the host screen is dynamically displayed in the state of charge.
  • When the battery is fully charged, the power symbol on the host screen shows a full charge state and the breathing lamp in the stop flashing.
  • It takes about 4 hours to fully charge the batter. Before using the equipment for the first time, the equipment shall be fully charged for at least 4 hours.

Product operation

RebornEndo-M-Cure-9-LED-Dental-Cure-Lamp-Curing-Light-
fig-7

  1. Switch key
  2. Mode setting key ‘M
  3. Time and power setting key

The default settings of the eight working modes:

  1. Efficient curing mode: M1(only suitable for M-Cure 3、M-Cure 9
    • The default light intensity is 1000mW/cm².
    • When the output light intensity is 1000、1500、1800W/cm², short press the S key to select of time which could be 5 10 15 20s.
    • When the output light intensity is 2300mW/cm², short press the S key to select the work time which could be 1 2 3s.
  2. Deep curing mode: M2(only suitable for M-Cure 3、M-Cure 9 The default light intensity is 1500mW/cm².
    • When the output light intensity is 1000、1500、1800W/cm², short press the S key to select work time which could be 5 10 15 20s.
    • When the output light intensity is 2300mW/cm², short press the S key to select the work time which could be 1 2 3s.
  3. High power mode: M3(only suitable for M-Cure 3、M-Cure 9 The default light intensity is 1800mW/cm².
    • When the output light intensity is 1000、1500、1800W/cm², short press the S key to select work time which could be 5 10 15 20s.
    • When the output light intensity is 2300mW/cm², short press the S key to select the work time which could be 1 2 3s.
  4. Fast curing mode(only suitable for M-Cure 3、M-Cure 8、M-Cure 9
    • The output light intensity can not be chosen, and it is 2300mW/cm² with the default setting.
    • Short press the S key to select the work time which could be 1、2、3s.
  5. Soft mode(only suitable for M-Cure 8、M-Cure 9
    • The output light intensity can not be chosen, and it will gradually increase the output power up to 1200mW/cm² with the default setting.
    • Short press the S key to select the work time which could be 5 10 15 20
  6. Pulse mode(only suitable for M-Cure 8、M-Cure 9
    • The intermittent output light intensity can not be chosen, and it is 1200mW/cm² with the default setting. Short press the S key to select the work time which could be 5 10 15 20s.
  7. Ortho mode(only suitable for M-Cure 8、M-Cure 9
    • The output light intensity can not be chosen and it is 2000mW/cm² with the default setting.
    • Short press the S key to select the work time which could be 3、5s.
    • The curing light will work 3/5s with a 1s interval, and cycle for 10 times.
  8. Caries checking mode(only suitable for M-Cure 3、M-Cure 8 M-Cure 9
    • The output light intensity can not be chosen, and it is 600 mW/cm ² with the default setting.
    • Short press the S key to select the work time which could be 5 10 15 20s .
    • When purple light irradiates caries or dental plaque, it will produce orange red fluorescence reaction to detect surface caries or dental plaque.

Enter the sleep status

  • When the device is in a shutdown state, press the switch key accompanied by a ‘DI’prompt tone. And then the indicator lights up .
  • The machine will automatically enter the sleep status.

Switching power mode

  • When the equipment is in the sleep status or working status, it can quickly cycle to the next working mode with a short press M key.
  • The light power and the curing time both can be set in M1、M2、M3. Only the curing time can be set in another work mode.

Notice

  • Short press the S key to select the work time. Then press to confirm the current setting.
  • When the work mode is in M1、M2、M3, it can enter the light power setting interface with a short press of the M key. Then, it can choose the output power by short-pressing the S key.

Start/stop working status

  • When the device is in the sleep status, it can start working status by a short time press .
  • In this case, the LED star works under the selected mode. Then it counts down to 0 seconds to end the solidification.
  • When the device is in the working status, it can stop with a short time press .
  • In this case, the device returns to sleep status.

Power off

  • In the case of sleep/working status, the device can be shut down by a long-time press .
  • In the case of sleep status, the device also can automatically shut down by waiting 2 min.

Attention

  • At the end of a working cycle, the next working cycle can be started immediately by short-pressing the button.
  • If the main unit gets hot obviously, please turn off the device until the main unit becomes cool. Please don’t make it continuously illuminate more than 10 times.
  • When operating finished, please clean the top of the main unit with calico to avoid infecting the light intensity.
  • When operating, put the disposable sleeve on the main unit, clip the notch of the light hood into the head of the top of the main unit, and then clamp the light hood down on the main unit.
  • The low power detection circuit is fixed inside of the main unit, when low power is detected, the battery level icon on the main unit screen displays 0 grid power and shows a flashing status.
  • The effective light intensity of this equipment is much higher than Halogen Lamp, The solidified depth of the curing light composite resin for 10 seconds will not be less than 4mm.

Advanced mode setting

  • When the device is under the shutdown status/ sleep status, it can enter the advanced mode setting interface by pressing the S key at the same time.RebornEndo-M-Cure-9-LED-Dental-Cure-Lamp-Curing-Light-fig-8

Cleaning, disinfection & maintenance

  • The cleaning of the main unit and link hood are as follows.
  • The main unit can not be cleaned with clean flowing water. It only can be wiped with a soft cloth dipped with a little alcohol.
  • After each use, check whether the dental composite material remains on the surface of the link hood to avoid affecting the service life of the link hood or the curing effect.
  • It is forbidden to use the following methods to disinfect the main unit and link hood:
  • Cook in the water or liquid solution;
  • Soak with iodine, alcohol, or disinfectant;
  • Bake in an oven.
  • Maintenance instructions for rechargeable lithium battery Please keep the battery in half-charged status and store it in dry conditions if the battery has not been used for a long time.
  • Be sure to use the matching original adapter to charge the battery. The recommended environmental temperature range is 0-40 ℃. Battery performance will degrade if the environmental temperature range exceeds the above value.

Routine maintenance

  • Any maintenance of the equipment shall be carried out by professionals or authorized dealers designated by the equipment manufacturer.
  • Turn off the power switch when the equipment is not in use. The equipment shall be charged once a month for five minutes each time if the equipment is not used for a long time.

Troubleshooting

Fault Possible cause Solution
No indication, no response 1、Battery out of power. 2、Faulty of battery.

1、Charge the device/Change a new battery.

2、Change a new battery

Low light power| 1、Low power of battery.

2、Resin residue on the top of the main unit.

| 1、Charge the device/Change a new battery.

2、Remove the resin.

Do not charge after connecting to the power adapter| 1、The power supply was not well connected.

2、Faulty of battery/adapter.

| 1、Connect the adapter and Charging interface again.

2、Replace the battery/adapter.

‘BLED’ shows on the screen| 1、The top of the main unit is not inserted or tightly inserted. 2、Poor contact or damage of blue light bulb.

3、Faulty of the top of the main unit.

| 1、Insert the top of the main unit tightly.

2、Send to after-service for repair. 3、Send to after-repair service.

‘PLED ERROR’ shows on the screen| 1、The top of the main unit is not inserted or tightly inserted. 2、Poor contact or damage of violet light bulb.

3、Faulty of the top of the main unit.

| 1、Insert the top of the main unit tightly.

2、Send to after-service for repair. 3、Send to after-repair service.

‘LOW POWER’ shows on the screen| Low power of battery| Charge the device/Change a new battery.

Safety precautions

  • Please carefully read this Instruction Manual before the first operation.
  • This device should be operated by a professional and qualified dentist in a qualified hospital or clinic. The use of the product must comply with the requirements of the relevant operating specifications and relevant regulations of the medical department.
  • This product does not contain toxic or hazardous substances and it is discarded under the relevant laws and regulations on waste medical devices. Please deal with the replaced battery according to national regulations.
  • Before each use, place a new disposable sleeve over the curing light. The disposable sleeve is custom-fitted to the curing light and keeps the surface of the curing light clean.
  • The disposable sleeve helps prevent cross-contamination.
  • Please note that the disposable sleeve must meet the requirements of relevant medical device regulations. To prevent cross-infection, it is forbidden to reuse the disposable sleeve.
  • During operation, the blue light should be aimed straight at the composite resin to ensure the effect of solidification.
  • Do not look directly into the light output. Patients, clinicians, and assistants are always suggested to wear protectors (safety goggles, face shield) to avoid crossing infection.
  • Please use the original light hood and follow the right installation ways to avoid the damage of LED light to the eyes. The light source can only aim at the dental resin.
  • It is forbidden to use the curing light aimed at a patient’s eyes, skin, oral mucosa, or other similar human parts.
  • After a working cycle, an operator can press the ON/OFF switch to start another working cycle. Stop operating if the equipment begins to heat obviously, do not restart until the equipment cools down.
  • Suggest continuing the working cycle less than 10 times.
  • Only the original adapter could be used because other brand adapters are likely to damage the circuit.
  • Do not carry out any repairs or modifications without authorization. In case of failure, please contact your local dealer and do not allow unauthorized personnel to repair.
  • Do not use damaged or defective instruments. Please check the equipment before each use. If there is any damage to the equipment, an operator must stop using it.
  • When charging or using, do not touch the charging interface, battery, and patient at the same time.
  • It is forbidden to turn on the working state during the equipment charging.
  • Keep the machine clean before and after an operation.
  • Please turn off the power switch after each treatment.
  • As a professional manufacturer of medical instruments, we are only responsible for the safety on the following conditions:
  1. The maintenance, repair, and modification are made by the manufacturer or the authorized dealer.
  2. The changed components are original of our company and operated correctly according to the instruction manual.
    • The instrument has electromagnetic interference. Do not use it around the electronic operation, at the same time have a strong electromagnetic interference environment should be careful to using the instrument.

Documents / Resources

| RebornEndo M-Cure 9 LED Dental Cure Lamp Curing Light [pdf] Instruction Manual
M-Cure 9 LED Dental Cure Lamp Curing Light, M-Cure 9, LED Dental Cure Lamp Curing Light, Dental Cure Lamp Curing Light, Lamp Curing Light, Curing Light
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals