Promega J3190 Cholesterol Ester Glo Assay
Product Information
Description
The Cholesterol/Cholesterol Ester-GloTM Assay is a biochemical assay used to measure the amount of cholesterol in a sample. The assay utilizes cholesterol esterase, cholesterol dehydrogenase,reductase substrate, NADH, luciferin, Ultra-GloTM Luciferase, ATP,and Mg2+ to convert cholesterol esters to cholesterol and generate light proportional to the amount of cholesterol present in the sample.
Product Components and Storage Conditions
The Cholesterol/Cholesterol Ester-GloTM Assay is available in two sizes:
- 5ml (Catalog Number J3190): Contains sufficient reagents to perform 100 reactions in 96-well plates. Includes 10ml Cholesterol Lysis Solution, 5ml Cholesterol Detection Solution, 55l Reductase Substrate, 500l Cholesterol Standard (20mM), and 50l Cholesterol Esterase.
- 50ml (Catalog Number J3191): Contains sufficient reagents to perform 1000 reactions in 96-well plates. Includes 50ml Cholesterol Detection Solution, 500l Cholesterol Standard (20mM), and 500l Cholesterol Esterase.
Product Usage Instructions
Assay Principle
The Cholesterol/Cholesterol Ester-GloTM Assay utilizes a series of enzymatic reactions to measure the amount of cholesterol in a sample. The steps involved in the assay are as follows:
- Prepare the Cholesterol Lysis Solution.
- Add the Cholesterol Lysis Solution to the samples.
- Prepare the Reductase Substrate.
- Add the Cholesterol Detection Reagent to the samples.
- Incubate the samples at room temperature for 60 minutes.
- Read the samples on a plate-reading luminometer.
Note: For measurement of total cholesterol, include cholesterol esterase in the cholesterol detection reagent. For measurement of free cholesterol, omit cholesterol esterase from the cholesterol detection reagent. Perform measurements in separate wells simultaneously to calculate the amount of cholesterol esters from the difference between free and total cholesterol.
All technical literature is available at:
www.promega.com/protocols/
Visit the web site to verify that you are using the most current version of
this Technical Manual.
E-mail Promega Technical Services if you have questions on use of this system:
techserv@promega.com
Description
The Cholesterol/Cholesterol Ester-Glo™ Assay(a) provides a luminescent method for measuring cholesterol and cholesterol esters in cultured cell lysates and other biological samples such as lipoprotein fractions, cell culture medium, serum and tissue homogenates. Cholesterol is an essential lipid involved in steroidogenesis, bile acid synthesis, cell signaling and maintenance of membrane structure. However, excessive levels of cholesterol contribute to inflammation, atherosclerosis, metabolic disease and cancer. In fact, the cholesterol content of lipoproteins is routinely used as an indicator of cardiovascular risk.
The Cholesterol/Cholesterol Ester-Glo™ Assay measures cholesterol using a
cholesterol dehydrogenase that links the presence of cholesterol to the
production of NADH and the activation of a proluciferin that produces light
with luciferase (Figure 1). For measuring Cholesterol Ester, the assay
includes a cholesterol esterase that will remove the fatty acid from
cholesterol esters to produce one molecule of cholesterol per molecule of
ester. The amount of cholesterol esters is determined from the difference of
cholesterol measured in the absence (free cholesterol) and presence (total
cholesterol) of esterase.
The Cholesterol/Cholesterol Ester-Glo™ Assay is part of a bioluminescent
metabolite assay platform that offers rapid and sensitve metabolite detection
in many sample types (1–2). The assay can be performed in 96- and 384-well
plates, and the detection reagents are added directly to samples without the
need for organic extraction, making it convenient for high-throughput
applications (Figure 2).
Figure 1. Schematic diagram of the Cholesterol/Cholesterol Ester-Glo™ Assay principle. Cholesterol esterase converts cholesterol esters to cholesterol. Cholesterol Dehydrogenase is used to generate NADH. In the presence of NADH, Reductase enzymatically reduces a pro-luciferin Reductase Substrate to luciferin. Luciferin is detected in a luciferase reaction using Ultra-Glo™ Luciferase and ATP, and the amount of light produced is proportional to the amount of cholesterol in the sample.
Figure 2. Cholesterol/Cholesterol Ester-Glo™ Assay reagent preparation and protocol. The protocol above is for measurement of total cholesterol. For measurement of free cholesterol, omit cholesterol esterase from the cholesterol detection reagent. Measurements in the absence and presence of esterase are performed in separate wells simultaneously to calculate the amount of cholesterol esters from the difference between free and total cholesterol.
Product Components and Storage Conditions
This system contains sufficient reagents to perform 100 reactions in 96-well plates (50μl of sample + 50μl of cholesterol detection reagent). Includes:
- 10ml Cholesterol Lysis Solution
- 5ml Cholesterol Detection Solution
- 55μl Reductase Substrate
- 500μl Cholesterol Standard (20mM)
- 50μl Cholesterol Esterase
This system contains sufficient reagents to perform 1000 reactions in 96-well plates (50μl of sample + 50μl of cholesterol detection reagent). Includes:
- 100ml Cholesterol Lysis Solution
- 50ml Cholesterol Detection Solution
- 2 × 275μl Reductase Substrate
- 500μl Cholesterol Standard (20mM)
- 500μl Cholesterol Esterase
Storage Conditions: Store complete kits at less than –65°C. Alternatively, store the Cholesterol Detection Solution and Reductase Substrate at less than –65°C, store the Cholesterol Esterase at less than –10°C and store the Cholesterol Standard and Cholesterol Lysis Solution at less than +10°C.The kit components can be freeze-thawed three times with no effect on assay performance. As needed, dispense kit components into single- use aliquots to minimize freeze-thaw cycles.
Note: Use personal protective equipment and adhere to your institution’s safety guidelines and disposal requirements when working with biohazardous materials such as cells and cell culture reagents.
Measuring Total and Free Cholesterol
A. Materials to Be Supplied By the User
- PBS
- 96- or 384-well assay plates (opaque white-walled with clear or white bottom) compatible with a standard plate reader
- single- and multichannel pipettors, tips and reagent reservoirs
- plate-reading luminometer (e.g., GloMax® Discover, Cat.# GM3000)
- water bath
B. Reagent Preparation
This protocol is for a reaction with 50μl of prepared sample and 50μl of
cholesterol detection reagent in a 96-well plate. The assay can be adapted to
other volumes provided the 1:1 ratio of cholesterol detection reagent volume
to sample volume is maintained (e.g., 20μl of sample and 20μl of cholesterol
detection reagent in a 384-well format).
-
Thaw all components in a 22°C water bath and mix to ensure homogeneous solutions prior to use. Place the
Reductase Substrate and Cholesterol Esterase on ice; all other components can be held at 22°C until use. -
Determine the amount of cholesterol detection reagent necessary for your current experiment. Use the reagent on the day it is prepared; do not store it for later use.
-
To prepare cholesterol detection reagent, add 10μl of Reductase Substrate per ml of Cholesterol Detection Solution and mix by inversion. Cholesterol detection reagent without Cholesterol Esterase is used to measure free cholesterol.
-
To prepare cholesterol detection reagent with Esterase, add 10μl of Reductase Substrate and 10μl Cholesterol Esterase per ml of Cholesterol Detection Solution and mix by inversion. Cholesterol detection reagent with Esterase is used to measure total cholesterol.
C. Assay Protocol
This protocol is for measuring free (without esterase) and total (with
esterase) cholesterol. The quantity of cholesterol esters can be calculated as
the difference between total and free cholesterol, determined in separate
wells simultaneously.
Since cholesterol levels can vary significantly in different samples (see
Table 2), samples may need to be diluted to a concentration below 80μM to fit
within the linear range of the assay. Twice as much lysis solution as
detection solution has been included in the assay for this purpose.
This kit includes a 20mM Cholesterol Standard that can be used to generate a standard curve (Figure 3) to confirm that samples are within the linear range of the assay and to calculate cholesterol concentration. If the sample relative light unit (RLU) values fall outside the linear range of the cholesterol standard curve, the sample dilutions should be adjusted and re- assayed. It is important to prepare standards in the same buffers used for preparing samples and to follow the same assay protocol. Since the esterase reaction releases one molecule of cholesterol per molecule of ester, free and total cholesterol can be calculated from the cholesterol standards in the presence and absence of esterase; a cholesterol ester standard is not required.
-
For each sample, standard or background control, prepare wells for measurement with and without esterase as needed.
For Medium, Serum and Homogenized Tissue Samples:-
Dilute samples in Cholesterol Lysis Solution to bring their cholesterol concentrations below 80μM.
Transfer 25μl of sample, standard or background control to a 96-well plate. Refer to Table 2 for dilution recommendations. -
Add 25μl of Cholesterol Lysis Solution, shake briefly and incubate for 30 minutes at 37°C.
For Adherent Cells and 3D Cultures: -
Remove medium from cells in a 96-well plate. Wash cells twice with 100μl of PBS.
-
Add 50μl of Cholesterol Lysis Solution, shake briefly and incubate for 30 minutes at 37°C.
-
If needed, dilute samples in Cholesterol Lysis Solution to bring their cholesterol concentrations below 80μM. Transfer 50μl of any diluted samples, standards or background controls to empty wells in a 96-well white-walled assay plate.
-
-
Add 50μl of cholesterol detection reagent with or without Esterase as prepared in Section 3.B to all wells.
-
Shake the plate for 30–60 seconds by hand or at a low rpm on a plate shaker.
-
Incubate at room temperature for 1 hour.
-
Record luminescence using a plate-reading luminometer such as the GloMax® Discover.
Note: The light signal continues to increase until all cholesterol is consumed and the signal plateaus. At any time point the signal is directly proportional to cholesterol concentration. -
Calculate free and total cholesterol by comparison of the luminescence of samples and standards assayed under the same conditions (See Figure 3).
Figure 3. Cholesterol standard curves. Dilutions of the provided Cholesterol Standard (20mM, in ethanol) were prepared in Cholesterol Lysis Solution, beginning with a 2μl aliquot of Cholesterol Standard into 498μl of Cholesterol Lysis Solution. (Refer to Table 1 for concentrations of Cholesterol Standard used in standard curve.) A 50μl aliquot of cholesterol detection reagent was added to 50μl of each standard in triplicate and the luminescence was read after 1 hour. Concentration was plotted against average RLU at each standard point and a linear curve was fit. The Cholesterol/Cholesterol Ester-Glo™ Assay can detect less than 1μM cholesterol and has an upper limit of 80μM cholesterol. See Sections 4.A and 4.B for other ways to convert luminescence into concentration.
Table 1. Cholesterol Titration Data.
Cholesterol
Cholesterol
(µM)
| ****
80
| ****
60
| ****
40
| ****
20
| ****
10
| ****
5
| ****
2
| ****
1
| ****
0
---|---|---|---|---|---|---|---|---|---
RLU
(thousands)
| 8,343| 6,685| 4,825| 2,537| 1,389| 745| 444| 293| 134
STDEV
(thousands)
| 229| 277| 285| 31| 35| 8| 7| 2| 3
CV| 2.7%| 4.1%| 5.9%| 1.2%| 2.5%| 1.0%| 1.6%| 0.7%| 2.5%
S/B| 62.5| 50.1| 36.1| 19.0| 10.4| 5.6| 3.3| 2.2| 1.0
S/N| 2,416.5| 1,928.5| 1,381.2| 707.4| 369.6| 180.1| 91.5| 46.8| –
Note: Coefficient of variation (CV) is 100 × STDEV/RLU. Signal-to- background ratio (S/B) is mean signal from samples divided by the mean signal from negative controls. Signal-to-noise ratio (S/N) is the net signal (mean signal minus mean negative control) divided by the standard deviation of the negative control.
Table 2. Recommendations for Sample Preparation.
Sample Type
| Expected Cholesterol
Concentration in Sample
| ****
Preparation Recommendations
---|---|---
Cell culture medium
(extracellular)
| Variable| • Dilute medium samples in Cholesterol Lysis Solution
Cell lysates (intracellular)| Variable| • Remove media from cells and wash
with PBS
• Add Cholesterol Lysis Solution
• Further dilution may be necessary
Differentiated adipocytes| 10–100s of μM| • Remove media from cells and wash with PBS
• Add Cholesterol Lysis Solution
• Further dilution may be necessary
Tissues| e.g., 100s of μM in a 25 mg/ ml mouse liver homog- enate| • Homogenize tissue with Tissue Tearor or other homogenization device
• Add Cholesterol Lysis Solution
• Further dilution (e.g., 10-fold) may be necessary
Plasma and serum| 100s of μM| • Dilute plasma or serum samples (e.g., 10-fold or more)
in Cholesterol Lysis Solution
Purified lipoproteins| mM levels in mg/ml
amounts of lipoprotein
| • Dilute lipoproteins (e.g., 200-fold) in Cholesterol Lysis
Solution
Example Experiments
A. Measuring Cholesterol and Cholesterol Esters in Lipoproteins
It is important to note that if the amount of free cholesterol is
significantly greater than the amount of esterified cholesterol, it may not be
possible to detect the ester. We have found this to be true with a number of
cancer cell lines.
In contrast, lipoproteins have significant amounts of cholesterol esters, and
it is easy to detect them.
In Figure 4, esterified and unesterified cholesterol concentrations were
measured using a single standard for quantification.
Signals from a lipoprotein sample (sample RLU), a cholesterol standard
(standard RLU) of known concentration (STD) and a negative control (background
RLU) containing no cholesterol were applied to the following formula:
Considering the data in Figure 4, and 250- or 500-fold dilutions of high density lipoprotein (HDL) or low density lipoprotein (LDL), respectively, this formula yields free and total cholesterol concentrations of 2.2mM and 6.8mM for HDL and 7.4mM and 17.7mM for LDL, respectively. The difference between free and total corresponds to cholesterol ester concentrations of 4.6mM in HDL and 10.3mM in LDL.
Figure 4. Cholesterol detection in human lipoproteins using a single
cholesterol standard. Human high density lipoprotein (HDL, 10mg/ml) and
human low density lipoprotein (LDL, 5mg/ml) were purchased from Kalen
Biomedical.
The HDL and LDL samples were diluted 250- and 500-fold into Cholesterol Lysis
Solution, respectively, and 50μl aliquots were assayed per the protocol, with
and without esterase. 40μM and 0μM of cholesterol in Cholesterol Lysis
Solution were assayed as the standard and background controls, respectively.
The RLU value (in millions) is listed at the top of each bar.
The relative amounts of lipoproteins in serum samples can also be determined by quantifying total cholesterol. An important feature of such an experiment is to have a way to separate the different lipoprotein fractions; an example using PEG8000 is shown in Figure 5.
Figure 5. Relative amounts of high vs. low and very low density lipoproteins (VLDL) in sera. Samples of bovine and human sera were mixed 5:2 with 20% PEG8000, incubated for 20 minutes at room temperature and then centrifuged for 30 minutes at 10,000 × g. HDL in the supernatant was transferred to a fresh tube and the LDL/VLDL in the precipitate was resuspended in PBS. Each lipoprotein fraction was diluted 100-fold into Cholesterol Lysis Solution and then assayed for total cholesterol (unesterified + esterified) as per the standard protocol.
Measuring Cholesterol in Mouse Liver Tissue
Cholesterol concentrations in a mouse liver homogenate were measured using a
cholesterol standard spike for quantification (Figure 6). Signals from mouse
liver samples minus (sample RLU) or plus (sample + spike RLU) of an added
amount of cholesterol (spike) and from a negative control (background RLU)
containing no cholesterol were applied to the following formula:
Using the data in Figure 6, and accounting for a 10-fold dilution in Cholesterol Lysis Solution, this formula yields a value of 102µM cholesterol in a 25mg/ml mouse liver homogenate.
Note that the homogenate itself does not interfere with the detection chemistry because the ΔRLU ± 10µM cholesterol is the same in the absence (standard RLU – background RLU = 5.12 million RLU) or presence (sample + spike RLU – sample RLU = 5.30 million RLU) of the diluted homogenate.
Figure 6. Cholesterol detection in mouse liver using a cholesterol spike. 25mg of mouse liver was placed in 1ml of PBS and homogenized with a Tissue Tearor (BioSpec Cat.# 985370-07) for 30 seconds. A 5µl aliquot of the homogenate was diluted into 45µl of Cholesterol Lysis Solution in the presence (sample + spike) or absence (sample) of an additional 10µM of cholesterol. These diluted samples were incubated for 30 minutes at 37°C and assayed per the standard protocol for free cholesterol. Cholesterol (10µM and 0µM) in Cholesterol Lysis Solution were assayed as the standard and background controls, respectively. The RLU value (in millions) is listed at the top of each bar.
Extraction of Cholesterol with β-Cyclodextrins
Some researchers choose to manipulate the cholesterol levels in their
biological systems prior to starting an experiment by incubation with
β-cyclodextrins. These molecules have a hydrophobic core that is roughly the
size of cholesterol, and hence they can “extract” cholesterol from cells.
However, extraction of too much cholesterol can lead to cell death, so it is
important to control the amount of β-cyclodextrin and the time of incubation.
In the experiment shown in Figure 7, HCT116 colon cancer cells were incubated
with and without methyl-β-cyclodextrin, and samples of the medium were removed
over time and assayed with the Cholesterol/Cholesterol Ester-Glo™ Assay. To
monitor cell viability, RealTime-Glo™ MT Cell Viability Reagent was added to
the cells and luminescence was measured over time. Cytotoxicity was also
measured by removing samples of medium over time and assaying for lactate
dehydrogenase (LDH) release with the LDH-Glo™ Cytotoxicity Assay.
Figure 7. Multiplexing viability and cytotoxicity measurements with the Cholesterol/Cholesterol Ester-Glo™ Assay during extraction of cholesterol by β-cyclodextrin. Twenty-thousand HCT116 colon cancer cells per well were plated in a 96-well plate and incubated overnight. The cells were washed twice with PBS, and 100μl of medium was added containing DMEM, 2% BSA and 1X RealTime-Glo™ Reagent, in the absence (blue) or presence (red) of 20mM methyl-β-cyclodextrin. At the indicated times, luminescence from the RealTime- Glo™ Assay was measured, and aliquots of medium were collected to measure the amount of extracted cholesterol. These aliquots were diluted 20-fold into Cholesterol Lysis Solution for the Cholesterol/Cholesterol Ester-Glo™ Assay and 100-fold into LDH Storage Buffer for the LDH-Glo™ Assay. The latter two assays were performed once all samples were collected per their standard protocols. Viability was normalized to the luminescent value measured at 2 hours, cholesterol was calculated from a standard curve and toxicity was normalized to a maximum release sample (via addition of 0.2% Triton X-100).
Appendix
Signal Stability
After addition of the cholesterol detection reagent to a sample, the
Cholesterol/Cholesterol Ester-Glo™ Assay yields a signal that is stable over
several hours (Figure 8).
Figure 8. Signal Stability of the Cholesterol/Cholesterol Ester-Glo™ Assay. An equal volume of cholesterol detection reagent was added to 40μM of cholesterol in Cholesterol Lysis Solution and the luminescence was read over time. After 5 hours, the signal was >50% of the signal at 1 hour, which is convenient for use in high-throughput applications.
Cholesterol Solutions
Cholesterol is not very soluble or stable in aqueous solution, and it can
adhere to plastic surfaces. Thus, it is important to store cholesterol in a
nonaqueous solution and in glass containers. Our Cholesterol Standard is
supplied as an ethanol solution in a glass vial. Significant concentrations of
detergents are included in our lysis and detection solutions to maintain
cholesterol concentrations while performing the assays, but cholesterol is not
stable in these solutions over extended periods of time.
Temperature and Reagent Compatibility
The intensity and stability of the luminescent signal is temperature
sensitive. For consistent results, equilibrate the reagents and samples to
room temperature before use. Avoid the use of DTT and other reducing agents in
the samples to be tested. Reducing agents will react with the Reductase
Substrate and increase background.
Plates and Equipment
Most standard plate readers designed for measuring luminescence are suitable
for this assay. Some instruments do not require gain adjustment, while others
may require optimization of the gain settings to achieve sensitivity and
dynamic range. An integration time of 0.5–1 second per well should serve as
guidance. For exact instrument settings, consult your instrument manual. For
optimum performance, use opaque-walled, white multiwell plates that are
compatible with your luminometer (e.g., Corning Costar® #3917 96-well or
Costar® #3570 384-well plates). Luminescence signal is diminished in black
plates and increased well-to-well crosstalk is observed in clear plates. The
RLU values shown in the figures of this technical manual vary depending on the
plates and luminometers used to generate the data. Although relative
luminescence output will vary with different instruments, this variation does
not affect assay performance.
Multiplexing and Normalization
The Cholesterol/Cholesterol Ester-Glo™ Assay can be multiplexed to normalize
for changes in viability and to account for well-to-well variation. If a
sample of media is removed for assay, the remainder of the sample can be
assayed with RealTime-Glo™, CellTiter-Fluor™ or CellTiter-Glo® Cell Viability
Assays, following the protocols provided with the respective assays. If
multiplexing with intracellular cholesterol detection, RealTime-Glo™ and/or
CellTiter-Fluor™ reagents can be added to the medium and measured prior to
media removal.
To determine if a treatment is toxic to cells, the LDH-Glo™ Cytotoxicity Assay
may be multiplexed with the Cholesterol/Cholesterol Ester-Glo™ Assay. A small
(2–5μl) sample of medium can be removed for the LDH-Glo™ Cytotoxicity Assay to
a separate plate, and the remaining cells and medium can be used for
cholesterol detection. Refer to Section 4.C for example data.
References
- Vidugiriene, J. et al. (2014) Bioluminescent cell-based NAD(P)/NAD(P)H assays for rapid dinucleotide measurement and inhibitor screening. Assay Drug Dev. Technol. 12, 514–56.
- Leippe, D. et al. (2017) Bioluminescent assays for glucose and glutamine metabolism: High-throughput screening for changes in extracellular and intracellular metabolites. SLAS Discovery. 22, 366–77.
- Glucose Uptake-Glo™ Assay Technical Manual #TM467.
Related Products
Energy Metabolism Assays
Product Size | Cat.# |
---|
Glycerol-Glo™ Assay
5ml| J3150
Triglyceride-Glo™ Assay
5ml| J3160
Glucose Uptake-Glo™ Assay
5ml| J1341
Glutamate-Glo™ Assay
5ml| J7021
Glutamine/Glutamate-Glo™ Assay
5ml| J8021
Glucose-Glo™ Assay
5ml| J6021
Lactate-Glo™ Assay
5ml| J5021
NAD(P)H-Glo™ Detection System
10ml| G9061
NAD/NADH-Glo™ Assay
10ml| G9071
NADP/NADPH-Glo™ Assay
10ml| G9081
Mitochondrial ToxGlo™ Assay
10ml| G8000
Other sizes are available.|
Viability Assays
Product Size | Cat.# |
---|
RealTime-Glo™ MT Cell Viability Assay
100 reactions| G9711
CellTiter-Glo® 2.0 Assay
10ml| G9241
CellTiter-Glo® Luminescent Cell Viability Assay
10ml| G7570
CellTiter-Glo® 3D Cell Viability Assay
10ml| G9681
CellTiter-Fluor™ Cell Viability Assay
10ml| G6080
Other sizes are available.|
Cytotoxicity Assays
Cytotoxicity Assays
Product
|
Size
|
Cat.#
---|---|---
LDH-Glo™ Cytotoxicity Assay| 10ml| J2380
CellTox™ Green Cytotoxicity Assay| 10ml| G8741
CytoTox-Glo™ Cytotoxicity Assay| 10ml| G9290
CytoTox-Fluor™ Cytotoxicity Assay| 10ml| G9260
Other sizes are available.| |
Multiplex Viability and Cytotoxicity Assays
Multiplex Viability and Cytotoxicity Assays Product|
Size
|
Cat.#
---|---|---
MultiTox-Glo Multiplex Cytotoxicity Assay| 10ml| G9270
MultiTox-Fluor Multiplex Cytotoxicity Assay| 10ml| G9200
Other sizes are available.| |
Oxidative Stress Assays
Oxidative Stress Assays Product|
Size
|
Cat.#
---|---|---
ROS-Glo™ H2O2 Assay| 10ml| G8820
GSH-Glo™ Glutathione Assay| 10ml| V6911
GSH/GSSG-Glo™ Assay| 10ml| V6611
Other sizes are available.| |
Detection Instruments
Detection Instruments Product|
Size
|
Cat.#
---|---|---
GloMax® Discover System| 1 each| GM3000
GloMax® Explorer System| 1 each| GM3500
Summary of Changes
The following changes have been made to the 6/23 revision of this document:
- The legend for Figure 3 was updated.
- Equations were reformatted and miscellaneous text edits were made.
- The font was updated.
(a)U.S. Pat. Nos. 9,273,343 and 9,951,372, European Pat. No. 2751089, Japanese
Pat. No. 6067019 and other patents pending.
© 2020–2023 Promega Corporation. All Rights Reserved.
CellTiter-Glo and GloMax are registered trademarks of Promega Corporation.
CellTiter-Fluor, CellTox, CytoTox-Fluor, CytoTox-Glo, Cholesterol/Cholesterol
Ester-Glo™, Glucose‑Glo, Glucose Uptake-Glo, Glutamate-Glo, Glutamine
/Glutamate-Glo, Glycerol-Glo, GSH-Glo, GSH/GSSG-Glo, Lactate-Glo, LDH-
Glo,Mitochondrial ToxGlo, NAD/NADH-Glo, NAD(P)H-Glo, NADP/NADPH-Glo, RealTime-
Glo, ROS-Glo and Triglyceride-Glo are trademarks of Promega Corporation.
Products may be covered by pending or issued patents or may have certain
limitations. Please visit our Web site for more information.
All prices and specifications are subject to change without prior notice.
Product claims are subject to change. Please contact Promega Technical
Services or access the Promega online catalog for the most up-to-date
information on Promega products.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA · Toll Free in USA 800-356-9526 · 608-274-4330 · Fax 608-277-2516 TM601 · Revised 6/23
All technical literature is available at:
www.promega.com/protocols/
Visit the web site to verify that you are using the most current version of
this Technical Manual.
E-mail Promega Technical Services if you have questions on use of this system: