NGPOD-01 Handheld Device Instruction Manual
- June 1, 2024
- NGPOD
Table of Contents
- NGPOD-01 Handheld Device
- Section 1: General and Safety Information
- Section 2: NGPOD® and Sensor Product Description
- Section 3: NGPOD® and Sensor User Instructions
- Clear the NG tube
- Insert sensor into tube
- Document test result
- Section 4: NGPOD® Cleaning, Maintenance, Storage and Disposal
- Section 5: Troubleshooting
- Product Specifications
- FAQ
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
NGPOD-01 Handheld Device
INSTRUCTION MANUAL
The NGPOD® System is to be used ONLY by personnel trained in its operation
Section 1: General and Safety Information
1.1 Preface
This manual and any information contained within, may not be reproduced, distributed, or transmitted in any form, or by any means, for any purpose, without the express written permission of NGPOD Global Ltd. To order additional copies of this manual, contact NGPOD Global Ltd by using the contact number provided on page 2. You can also access the digital copy on our website www.ngpodglobal.com/IFU using password: showmethepod1 or scanning the QR code on the Sensor and POD labels.
The information in this document has been carefully compiled and is believed
to be entirely accurate. Furthermore, NGPod Global Ltd reserves the right to
make changes to any products herein to improve readability, function, or
design. NGPOD Global Ltd does not assume any liability arising from the mis-
use of any product described herein; neither does it cover any license under
its patent rights nor the rights of others.
Any serious incident that has occured in relation to the device should be
reported to the manufacturer and the competent authority of the Member state
in which the user is established.
1.2 About This Manual
Prior to use, the operator must be trained in the use of the NGPOD®, a system to assist in the placement confirmation of Nasogastric (NG) feeding tubes. The system consists of a detection device (hereinafter referred to as the NGPOD®) and a single-use disposable Sensor (hereinafter referred to as the Sensor).
If a software change occurs and the operation/specification for the device changes, new or additional operating instructions will be issued, if required.
1.3 NGPOD® and Sensor Introduction
Device Description and Applications
The NGPOD® is a device to assist in position confirmation of a NG tube and
consists of a re-usable, portable handheld detection device and a single use
fibre-optic Sensor tipped with a pH indicator.
It has been designed to be an alternative to conventional feeding tube
position confirmation tests, delivering a visual indication of the pH
environment of the NG tube tip. The result is a YES/NO indication, reducing
the risk of clinical interpretation error.
The Sensor attaches to the NGPOD® handheld device. The NGPOD® handheld device couples light into the fibre-optic Sensor. Light reflected from the sensor tip is received by the pod. Data from the pod sensors is processed to assess whether the colour of the indicator has passed a threshold that corresponds to the sensor tip being in an acidic environment.
The NGPOD® can only be used with the NGPOD® fibre optic sensor. The system is
intended to aid in the correct placement of NG tubes by detecting an acidic
environment through a colour change in the pH indicator.
As the consequences of feeding tube misplacement can be very serious, the
quick, clear indication delivered by the NGPOD® can facilitate safer and more
efficient NG placement in the clinical environment.
This device is fully compliant with NPSA (National Patient Safety Agency) feeding tube positioning confirmation guidelines.
Intended Use
NGPOD® is designed to signal to the clinical practitioner that the presence of
a substance with a pH below 5.5 has been detected. When the NGPOD® device is
used in the correct manner, a positive signal can typically be understood to
mean that the presence of stomach acid has been detected.
It is the responsibility of the practitioner, using the result from the NGPOD®
device, together with their own skills and experience, to make the final
judgement as to the successful emplacement of the feeding tube and therefore
whether feeding may safely proceed. These instructions DO NOT replace local
guidelines for NGT placement. NGPOD® has been designed to replace aspirate
testing as a 1st line, bedside test to assist in the confirmation of
nasogastric tubes.
If a user chooses to use the NGPOD® system following an attempt to obtain aspirate from the nasogastric tube,the Operating Precaution ‘If using after an attempt to obtain aspirate..’ should be observed – see below.
Contraindications
The NGPOD® system should not be used where a pH test is clinically contraindicated for assisting in nasogastric tube placement, such as when a patient’s condition may result in acid being present in the lungs, or where the respiratory tract may have a pH below pH 6.
1.4 Advisory Terms and Symbols
This section describes standard terms used to signal advisory statements and other messages in this manual and system labelling. Please read and understand this manual and all warnings and precautions before operating the
NGPOD® system.
Warnings, Precautions and Notes
The information may be a warning, a precaution or a note. These are described
as:
- WARNING: Signals a warning statement that alerts the reader to a potential safety hazard which, if not avoided, could result in serious injury or death
- PRECAUTION: Signals a precaution statement that alerts the reader to a potential safety hazard or unsafe practice which, if not avoided, may result in minor or moderate injury to the user or patient, or damage to the device or other property
- NOTE: Indicates that the information that follows is additional information, a tip that will help you recover from an error or point you to related information within the manual
Operating Precautions
-
The NGPOD® has been designed and manufactured to exact specifications; however, it is not intended to replace trained personnel in the supervision of NG tube placement
-
The NGPOD® Sensor is designed for use with specific lengths of NG Tube. Please ensure that the Sensor labelling has been checked prior to use to ensure compatibility and safety
IF USING NGPOD® AFTER AN ATTEMPT TO OBTAIN ASPIRATE FROM A NASOGASTRIC TUBE the following should be observed: -
Insufflate air into the nasogastric tube prior to inserting the NGPOD sensor into the lumen of the nasogastric tube. This should be done even if the attempt to obtain aspirate has failed.
-
The fibre-optic Sensor is packaged separately to the NGPOD® handheld device. The Sensor is a single-use item and the NGPOD® handheld device is a semi re-usable item. A new Sensor must be used for each patient
-
THE TIP OF THE SENSOR MUST NOT BE TOUCHED BEFORE INSERTING DOWN THE NG TUBE, AS IT COULD POTENTIALLY CONTAMINATE THE pH INDICATOR
-
If significant resistance is felt when advancing the sensor down an NGT, the test should be stopped and the NGT confirmed as being validated with the NGPOD system using the Appendix to the NGPOD Instructions for use.
-
If the sensor cannot be removed easily, remove the NGT with the sensor in situ and repeat the test using a new NGT and NGPOD Sensor.
-
NGPOD Global Ltd will assume no responsibility for incidents which may occur if the product is not used, stored or transported in accordance with the environmental conditions stipulated in this document and on the package labelling
-
The NGPOD® is tested to IP21. Wipe clean only
-
The NGPod system has been tested and certified (to IEC-60601-1-2) for EMC (electromagnetic compatibility) to safely function in general and professional healthcare environments. The system should not be used within extreme electric or magnetic fields, for example in close vicinity of active MRI scanners, powerful radio transmitters or X-Ray equipment. In the unlikely event that such fields affect operation of the system, the pod should detect anomalous behaviour, safely block test progression and power down
-
The NGPOD® is designed for regular use and should withstand everyday handling. If the NGPOD® is dropped onto a hard surface, or is suspected of being dropped, the operation should be checked by running the self-test.
-
For self-test process, please see page 13 of this document
-
Using the NGPOD® near operating equipment which radiates high-energy radio frequencies (such as electrosurgical/ cauterising equipment, two-way radios, or cellular telephones) may cause failure to operate correctly. If this happens, reposition the NGPOD® away from the source of interference, or turn off the NGPOD® or other equipment as appropriate
-
An explosion hazard exists if the NGPOD® is used in the presence of flammable anaesthetics, such as Desflurane, Enflurane, Halothane etc. Exercise care to locate the NGPOD® away from such hazardous sources
BEFORE USE, ENSURE THAT YOU HAVE PERFORMED THE SENSOR SAFETY CHECK OUTLINED IN SECTION 3.1.2
PAGE 13 OF THIS USER GUIDE -
Caution should be used in patients who have significantly altered anatomy of the mouth, oropharynx or upper gastrointestinal tract due to trauma or surgery, particularly where this affects the ability to clear secretions from the upper gastrointestinal tract. In these patient groups, a clinical assessment should be made as to whether x-ray rather than a bedside pH test is the most appropriate method of confirmation
-
Do not use acidic lubricator with the sensor
General Precautions
- Do not use hard or sharp objects to press buttons
- The specified accuracy of the NGPOD® can only be achieved if the NGPOD® is used and maintained in accordance with the instructions given in this manual
- Follow the cleaning schedule and methods defined under Section 4 to ensure proper operation of the NGPOD®.
- Wipe off spills immediately. Although the NGPOD® is splash-proof, do not allow fluid or residues to remain on the NGPOD® as these could cause biological hazards or equipment malfunction
- The NGPOD® has been designed to be as safe as possible to handle; however, care should always be exercised in order, to avoid user injury or device breakage
1.5 System Symbols
The following symbols are used on the NGPOD® and Sensor. Labels on the device or statements in this manual preceded by any of the following words and/or symbols are of special significance and/or are intended to help you to operate the device in a safe and successful manner.
NGPOD® Symbols
Symbols | Description |
---|---|
CE mark indicates conformance to the Medical Device Regulation EU/2017/745 | |
Indicates conformance to the UK Medical Device Regulations 2002 | |
Not for disposal in domestic or municipal waste: this product is compliant |
with the WEEE (Waste Electrical and Electronic Equipment) directive
| Manufacturer
| Part number
| Indicates the need for the user to consult the instructions for use
| Serial number
Sensor Symbols
Symbols | Description |
---|---|
Sterilized with Ethylene Oxide | |
Safety symbol. There are specific warnings and precautions associated with |
the device which are not otherwise found on the label
| Indicates the need for the user to consult the instructions for use
| The use of single-use disposable components on more than one patient is a
biological hazard. Do not reuse single-use disposable components
| Do not use if package is damaged. If the sensor packaging looks like it has
been damaged/opened, contact the storage department for a replacement
| Phthalate (DEHP) free
Symbols| Description
---|---
| Latex free
| Manufacturer reference/part number
| Keep out of direct sunlight. To indicate that transport package shall not be
exposed to sunlight
| Keep dry
| CE mark indicates conformance to Medical Device Directive 93/42/EEC
| UKCA marked in accordance with the UK Medical Device Regulations 2002 0120
| The unique lot code (batch number) for the device (The lot number YYYY is
year of manufacture and NNNN is a unique number)
| Date of manufacture
| Expiry date. This symbol is accompanied by a date to indicate that the
device should not be used after the end of the year, month or day shown
| Manufacturer
1.6 NGPOD® Technical Specifications
Component | Description |
---|---|
Housing Materials | Casing – Polycarbonate ABS (fire retardant). |
Front & Rear Labels – Polyester
Dimensions| 170mm x 40mm x 45mm
Weight| 118g
Power Supply| Battery type:
Non-rechargable AA 1.5Volt Alkaline
Battery Operation| At least 5000 tests under normal use conditions
Classification| Class 1 device
Protection against current leakage| Type CF equipment
Standards| Manufactured in accordance to ISO 9001 and ISO 13485.
CE Marked in accordance to Medical Devices Regulation EU/2017/745
UKCA marked in accordance with the UK Medical Device Regulation
2002
Electrical safety| Electromagnetic compatibility (EMC)
BS EN 60601-1-2:2015
Transport| Must be transported in the original packaging or equivalent to
maintain protection
1.7 Device Inspection and Unpacking
Inspecting The NGPOD® Before Use
Remove the NGPOD® from the packaging and inspect for damage that could have
occured during shipment or storage.
Make sure you have the following items:
NGPOD® Handheld Device & Quick User Guide (QUG)
If any items are missing or damaged, please contact your supplies department.
- PRECAUTION: Visually inspect packaging and contents before each use
- PRECAUTION: Do not drop as this may cause damage to the NGPOD® handheld device.Do not use the NGPOD® and accessories if there are any obvious signs of damage. Return for inspection to NGPod Global Ltd
Section 2: NGPOD® and Sensor Product Description
2.1 NGPOD® Feature Recognition
| Feature| Description
---|---|---
1.| Self-Test Cap| Protects internal mechanisms. Remove cap to insert sensor
connector to the NGPOD®
2.| Lock-Ring| To guide the locking and unlocking of the collar to NGPOD®
3.| Green Tick and Red Cross pH status indicators| Displays test result:
Green indicates pH reference in range Red indicates pH reference out of range/
undetected
4.| Test button| Press to start pH test
5.| Amber LED indicator light| Flashing amber light indicates NGPOD® is in a
ready
state for pH testing
6.| Tether| Attaches the NGPOD® cap to the body
7.| Power on/off button| Powers the NGPOD® on and off
2.2 Sensor Feature Recognition
Sensors are currently available with a length of 92cm
The sensor packaging states the dimensions for nasogastric tube compatibility.
Section 3: NGPOD® and Sensor User Instructions
Introduction
These instructions for use do not include instructions for insertion or management of a Nasogastric (NG) feeding tube; please refer to local policy and guidance for inserting and managing NG tubes.
Local guidance will indicate the frequency and circumstances of pH checks which can include;
- Following initial insertion
- Before administering feed and/or fluids
- Before administering medication
- Once daily check if on continuous feeding
- Following episodes of vomiting, retching or coughing
3.1 Carrying Out the NGPOD® Test
Sequence for pH Testing
The key steps for carrying out pH testing are:
3.1.1 Prepare the patient and equipment
3.1.2 Carry out the NGPOD® pH test
3.1.3 Remove and dispose of sensor
3.1.1 Prepare the patient and equipment
Prepare patient
Follow local policy for infection prevention + control
Explain the procedure to patient, gain consent and position the patient to
carry out the procedure.
Prepare equipment
Assemble equipment:
- NGPOD® Handheld device
- NGPOD® Sensor
- Gloves
- Apron
- pH neutral, non-acidic lubricant
Check NGPOD®
- Examine the outside of the NGPOD® for any signs of damage, including cracked housing
- Do not use the NGPOD® if it has been damaged to the extent that cracks are evident in the casing or parts of it are missing
- Ensure the lock ring is fully engaged
Check Sensor
Check that the sensor packaging is sealed, intact and undamaged before use. DO
NOT USE if packaging is open or damaged.
Always ensure pre-test Sensor check is carried out prior to inserting the
Sensor down the NG tube
Visually check that the sensor tip is BLUE
- WARNING: ENSURE NGPOD AND SENSOR COMBINED SAFETY CHECK ON PAGE 13 IS PERFORMED PRIOR TO
- INSERTING THE SENSOR DOWN NG TUBE
- WARNING: Prior to use of the NGPOD® system, compatibility with the nasogastric tube(s) (NGT) to be used should be established. Please refer to Appendix 1 of this IFU for a list of validated NGTs, and the correct sensor for use with each NGT . If the nasogastric tube you wish to use is not included in Appendix 1 please contact NGPod Global for advice or to arrange a validation
- PRECAUTION: If the tip of the sensor is green/yellow, then it must not be used. Yellow discolouration indicates contamination
3.1.2 Carry out the NGPOD® test
Power on NGPOD® Handheld Device to perform Self-Test
Ensure the NGPOD® self-test cap is fully seated.
To power on, press and release the power button
When the device is powered on, the self- test begins automatically and takes a
few seconds to complete. This checks the LED’s, internal sensors and the
battery.
During the self-test, observe the LED light sequence (Red – Amber – Green) :
The LED lights will cycle through the above light sequence, if self-test is successful then the device will enter Sensor Check mode. This is indicated by a repeating sequence of three flashes of the amber LED followed by a single flash of the red LED.
DO NOT proceed if any of the indicator LEDs fail to illuminate during self-
test.
If the NGPOD detects an error during self-test, the red LED will flash 4 times
and the device will power off.
If the NGPOD detects low battery voltage, the amber LED will flash 10 times
after the self test is complete. The NGPOD will continue to operate but
arrangements should be made to replace the NGPOD at the earliest opportunity.
NOTE: The NGPOD® will remain in ‘Sensor-Check’ mode until the device has been connected to a new, unused Sensor and the software has validated all components are working
Sensor Check Mode
Remove NGPOD® handheld device self-test cap.
Open the Sensor package slightly from the corner (not fully) to expose the
Sensor connector cap and retain the end of the Sensor in the packaging for
protection.
Push the connector of the NGPOD® sensor into the collar of the NGPOD handheld
device and rotate until you hear it click.
With the Sensor connected the device automatically runs through the Sensor
Check.
If successful, the green indicator light will briefly illuminate and the
device will enter pH test mode, signified by a repeated flashing amber light
every second. Disconnect NGPOD® handheld device from the Sensor.
If the NGPOD fails to enter pH test mode with the sensor connected correctly,
the sensor has failed the sensor check.
Discard the sensor and try agin with a new sensor.
THE DEVICE IS NOW READY TO BE USED
NOTE : The NGPOD® will power off after five minutes to conserve battery power. If the unit powers off, repeat the self- test as above
Clear the NG tube
When the NGPOD system is used for a repeat NG Tube position test consider insufflating air in the NG Tube before attempting to insert the NGPOD sensor.
- NOTE: This will clear any debris from the end of the tube and dislodge the tip of the NG tube if it is against/adjacent to the gastric mucosa. The patients medical condition must be considered prior to insufflating air down the NG tube, and if there is any doubt as to whether this is appropriate, air should not be insufflated
- NOTE: In order to maintain the ability to successfully pass the NGPOD Sensor it is important that local or national guidelines in relation to the care of the NG Tube are followed to prevent the obstruction of the lumen of the NG Tube
Insert sensor into tube
Remove sensor from packaging, Insert the Sensor tip into the NG tube and
gently advance the Sensor slowly down the NG tube until the NGPOD® Sensor
connector is reached.
NB: The sensor is not attached to the POD during insertion
- WARNING: Do not touch the tip of the sensor or allow it come into contact with any liquid prior to insertion into the nasogastric tube
- NOTE: NG Tubes differ in exact length, the NGPOD® Connector may not come into direct contact with the NG Tube Connector. However, if the sensor cannot be advanced to closer than 5mm from the NG Tube Connector the sensor tip is likely to be at the distal tip of the NG Tube
- NOTE: Take care not to create any sharp or permanent bends in the sensor
- PRECAUTION: Do not use force to advance the Sensor, the NG tube may be displaced, blocked or kinked.
Follow local policy for instructions on checking tube patency and position
When the amber LED light is flashing intermittently, press and release the button. The result will display in 4 seconds
GREEN LED light indicates that the sensor tip is in a pH environment
associated with the stomach. Follow local policy to commence feeding or
administer medications.
RED LED light indicates that the sensor tip is in a pH environment NOT
associated with the stomach.
When a red LED light displays, DO NOT proceed with feeding or administration
of liquids. Whilst the sensor is in place the NG tube may be advanced or
withdrawn in small increments.
Refer to National/local policies and guidelines on repositioning and re-
testing
WARNING: It may take up to 5 minutes following initial insertion of the NGPOD Sensor and performance of a pH test for a Green Tick result to occur. This should be anticipated in patients where a higher pH may be expected (eg patients on acid reducing medication and/or patients with low residual stomach volume due to continuous feeding regimes)
WARNING: Medications affecting pH
The pH can be affected by factors such as; medication and frequency of feeds.
Check whether the patient is on an acid inhibiting medication. Medications
such as antacids, h2 antagonists and proton pump inhibitors can affect the
gastric pH. The results of the NGPOD® test should be documented. If the
patient is on medication that could elevate the pH level of gastric contents,
refer to local policy and instructions for checking gastric pH
WARNING: Dilution of gastric acid by enteral feed
The presence of enteral feed/food in the stomach may alter the pH. If the test
fails four times, remove and discard the sensor, wait for up to an hour to
allow time for the stomach to empty and the pH to fall and then retest. The
time interval will depend on the clinical need of the patient and if they are
on continuous or bolus feeds
PRECAUTION: Lens dirty/ not clean
Check the NGPOD® lens. If the lens appears dirty or debris is visible, clean
the lenses according to the cleaning instructions on page 16, then repeat the
NGPOD® test from the beginning
WARNING: Excessive/abnormal gastric secretions
The NGPOD® device works by responding to a colour change when the tip of the
sensor comes into contact with acidic liquids. Please note that patients
experiencing excessive or abnormal gastric secretions may have higher levels
of acidic liquids in the Oesophagus and Pharynx. The NGPOD® should not be used
if there is a risk of a positive result through contact with acidic liquids in
areas other than the stomach
Document test result
Follow local policy for recording test results in patient’s medical records or on a test monitoring chart.
WARNING: If the patient is subsequently moved before feed is
administered, then a re-test should be conducted to ensure that the NG tube
has not become dislodged from the position that it was in when the test was
conducted
WARNING: If the patient vomits during or immediately prior to the NGPOD®
test then the user should revert to X-ray for NG placement confirmation as
there is a possibility of gastric contents in the lung which could result in a
positive result for a tube placed in the lung
NOTE: After feeding, the NG tube should be flushed as per local policy
3.1.3 Removal and disposal of equipment
Remove the Sensor connector from the NGPOD®
Replace the cap onto the NGPOD®
Gently withdraw the Sensor from the NG tube
NOTE: AFTER USE, ON VISUAL INSPECTION THE SENSOR TIP SHOULD BE GREEN/YELLOW. IF THE SENSOR TIP IS STILL BLUE, DO NOT COMMENCE FEED AND REPEAT TEST USING A NEW SENSOR
Dispose of Sensor in clinical waste
WARNING : During pH testing, the Sensor has contact with gastric
contents.
Therefore, used Sensors carry a risk of infection
WARNING : On removal of the sensor from the nasogastric tube, dispose of immediately and avoid any cross-contamination with the NGPOD hand-held device (POD)’
Following completion of every NGPOD test cycle, i.e. a green or persistent red result has been obtained and testing stopped, the device should be turned off by pressing and holding the power button until the amber light flashes five times. This procedure resets the device ready for a test on the next patient.
Section 4: NGPOD® Cleaning, Maintenance, Storage and Disposal
4.1 NGPOD® Handheld Device Cleaning
The NGPOD® requires no maintenance or servicing and contains no user serviceable parts.
WARNING : Do not attempt to open or modify the NGPOD® as this will void
warranty and may cause the unit to malfunction and/or give incorrect readings
and/or may cause irreparable damage
PRECAUTION : If the NGPOD® is dropped, subjected to excessive moisture,
humidity or high temperature, or otherwise suspected to have been damaged,
remove it from service
PRECAUTION : The NGPOD® is designed for regular use, and should withstand
everyday handling. If the NGPOD® is dropped onto a hard surface, or is
suspected of being dropped, the operation and calibration should be checked by
running the self-test
4.2 Cleaning and disinfecting the NGPOD®
Only the external parts of the NGPOD® should be regularly cleaned/disinfected.
Before routine cleaning/disinfecting the NGPOD®, ensure the cap is in place.
There may be infrequent occasions where it is necessary to remove the POD
collar assembly for cleaning the collar and cleaning the area around the lens.
See 4.4.2 for cleaning of the POD lens.
Cleaning and Disinfecting Procedure
- Clean and disinfect the NGPOD® with the cap in position
- The exterior of the NGPOD® can be cleaned with soft disinfecting wipes
- Test any cleaner on a small area of the cover before using it on the NGPOD®
- Always remove any soapy residue after cleaning.
- Do not use hard instruments for cleaning
- Always clean/disinfect the NGPOD® after each use to avoid cross-infection
- For a NGPOD® that has been in an isolation area, select those agents that both clean and disinfect
- Refer to infection prevention + control local policy for cleaning of non-invasive medical devices.
FOR FURTHER DETAILS ON CLEANING/DISINFECTING RECOMMENDATIONS AND SUITABLE
CLEANING AGENTS, PLEASE
REFER TO THE ‘NGPOD CLEANING GUIDELINES’ ON THE NGPOD WEBSITE.
(www.ngpodglobal.com)
WARNING : Avoid cleaning the inside of the cap when cleaning the optical
lens. Do not allow scratching or obscuring of the optical pathway or allow
objects or debris such as dust, dirt, hair, fluff, etc to interfere with the
optical pathway. Device performance can be severely degraded and may cause
incorrect operation. Keep the NGPOD® cap fitted when the Sensor is removed to
prevent this PRECAUTION: Avoid abrasive cleaners or harsh chemicals which
could affect the surface finish. e.g. scouring powder, wire wool, acetone
cleaner
PRECAUTION: Do not allow fluid or residues to remain on the NGPOD® as
these could cause biological hazards or equipment malfunction. While the
product design safeguards against minor fluid spillage, if fluid enters the
NGPOD®, this will cause device malfunction and may void the warranty
PRECAUTION: Do not autoclave or use ethylene oxide gas, this will cause
device malfunction and may void the warranty
4.3 NGPOD® Handheld Device and Sensor Storage and Disposal NGPOD® Storage
- Clean and disinfect the NGPOD® after use before storage
- Store the NGPOD® between use in a designated storage area
- Do not store the NGPOD® in an environment of high temperature, humidity, explosive or inflammable atmosphere or strong magnetic fields
NGPOD® Disposal
To dispose of the device, accessories or packaging, please ensure disposal
minimises any negative impact on the environment. You may be able to use
special recycling or disposal schemes. To find out about these, contact your
technical service department or local waste disposal service. Existing
national or local regulations concerning waste disposal must take precedence
over the above advice.
In the event of difficulty with disposal, please return your device for
disposal to NGPod Global Ltd at the address on the label. NGPod Global Ltd
will dispose of the device in accordance with the WEEE directive.
Sensor Storage
- Do not store Sensor’s in direct sunlight
- Do not store in high humidity
- Do not store near a strong electromagnetic field
Sensor Disposal
- Sensor is SINGLE USE ONLY
- Single use items must be disposed of following local policy and procedures
- Dispose in clinical waste as per local policy
- Do not clean or disinfect the sensor
4.4 POD Collar Assembly – cleaning, removal and refitting
The collar assembly consists of the pod collar (which accepts the sensor connector), the lock-ring and the test cap (connected to the lock-ring by a rubber tether). At the far end of the collar is a slot that engages with a feature of the sensor connector. When the collar is fitted this slot should be in line with the arrow on the pod casing.
Removal of the collar assembly is not recommended as part of normal use of the pod.
WARNING : Before each use, the collar lock-ring should be inspected to ensure that it is fully engaged in the locked position – the D-shaped boss should be at the end of the slot in the lock-ring and the slot at the end of the collar should be in line with the arrow on the pod casing
4.4.1 Removal and refitting
If it is suspected that the pod lens is dirty (this could be indicated by repeated failure of the pod self-test), the collar assembly can be removed to allow inspection and cleaning Removal
- Firmly hold the lock-ring by the grip features on either side of the lock-ring
- Twist the lock-ring in an anti-clockwise direction (when viewed from the front of the pod)
- Pull the collar assembly away from the pod
Refitting
- Taking care not to touch the pod lens, push the collar assembly on to the front of the pod. The D- shaped boss should fit easily into the slot in the lock-ring. Do not force the lock-ring into an incorrect orientation
- Firmly hold the lock-ring by the grip features on either side of the lock-ring
- Twist the lock-ring in a clockwise direction (when viewed from the front of the pod) until the end of the slot is in contact with the D-shaped boss
NOTE: The collar assembly can only be fitted in one orientation. The slot at the end of the collar should be in line with the arrow on the pod casing. Do not attempt to force the collar assembly into a different orientation
4.4.2 Inspecting and cleaning the Pod lens
- Remove the collar as in 4.4.1
- Wipe the interior and exterior of the collar assembly with disinfectant wipes
- Carefully wipe the pod lens and surrounding area with disinfectant wipes. Wipe away any liquid remaining on the lens. Take care not to leave dust, lint or any other contaminants on the lens surface
- Replace the collar assembly as in 4.4.1
Section 5: Troubleshooting
Problem | Symptoms | Remedies |
---|---|---|
Pod fails self-test | Red LED flashes 4 times before pod powers down | • Ensure |
test cap is fitted and fully seated.
• Ensure inside of test cap is clean
•Ensure pod collar is fitted correctly and lock-ring is fully engaged – see
section 4.4
•Ensure pod lens is clean – see section 4.4
• Do not operate pod under very bright direct lighting
• Replace pod
LEDs are not functioning| Red, Amber or Green LED does not flash during self-
test| • Replace pod
Low battery| Amber LED flashes 10 times during self-test| • Replace the NGPOD®
at the earliest opportunity
Sensor fails sensor-check| Pod will not enter pH Test mode from Sensor Check
mode| • Check that the sensor connector is correctly fitted to pod
• Remove sensor from pod and long-press power button to switch pod off. Ensure
test cap is fitted and switch pod on to repeat self-test. Remove test cap and
reconnect sensor to pod
• Sensor is damaged or defective – dispose of sensor. Use a new sensor
Product Specifications
- Model: MD-NG-V11
- Issue Date: February 2024
- Revision: V11
- Product Codes: NGPOD-01, NGPOD Sensor-01
FAQ
Q: What should I do if the device malfunctions?
A: If you encounter any issues with the device, please contact our customer support team for assistance.
Q: Can the device be used with pediatric patients?
A: The device can be used with pediatric patients under appropriate clinical supervision.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>