External Cavity SAF Laser
CES Cateye
Revision 1.0
CES Cateye External Cavity SAF Laser
Limitation of Liability
MOG Laboratories Pty Ltd (MOGLabs) does not assume any liability arising out
of the use of the information contained within this manual. This document may
contain or reference information and products protected by copyrights or
patents and does not convey any license under the patent rights of MOGLabs,
nor the rights of others. MOGLabs will not be liable for any defect in
hardware or software or loss or inadequacy of data of any kind, or for any
direct, indirect, incidental, or consequential damages in connections with or
arising out of the performance or use of any of its products. The foregoing
limitation of liability shall be equally applicable to any service provided by
MOGLabs.
Copyright
Copyright © MOG Laboratories Pty Ltd (MOGLabs) 2014 2023. No part of this
publication may be reproduced, stored in a retrieval system, or transmitted,
in any form or by any means, electronic, mechanical, photocopying or
otherwise, without the prior written permission of MOGLabs.
Contact
For further information, please contact:
MOG Laboratories P/L
49 University St
Carlton VIC 3053
AUSTRALIA
+61 3 9939 0677
info@moglabs.com| MOGLabs USA LLC
419 14th St
Huntingdon PA 16652
USA
+1 814 251 4363
www.moglabs.com
---|---
Preface
Diode lasers can be wonderful things: they are efficient, compact, low cost,
high power, low noise, tunable, and cover a large range of wavelengths. They
can also be obstreperous, sensitive, and temperamental, particularly external
cavity diode lasers (ECDLs). In combination with advanced electronics such as
the MOGLabs DLC external cavity diode laser controller, the CES cateye laser
described here provides a robust, stable, acoustically inert, low linewidth
and highly tunable laser system. We hope that the MOGLabs CES works well for
your application. Please let us know if you have any suggestions for
improvement in the laser or in this document, so that we can make life in the
laser lab easier for all, and check our website from time to time for updated
information.
MOGLabs, Melbourne, Australia
www.moglabs.com
Safety Precautions
Safe and effective use of this product is very important. Please read the
following laser safety information before attempting to operate the laser.
Also please note several specific and unusual cautionary notes before using
MOGLabs lasers, in addition to the safety precautions that are standard for
any electronic equipment or for laser-related instrumentation.
CAUTION USE OF CONTROLS OR ADJUSTMENTS OR PERFORMANCE OF PROCEDURES OTHER
THAN THOSE SPECIFIED HEREIN MAY RESULT IN HAZARDOUS RADIATION EXPOSURE
Laser output from the CES can be dangerous. Please ensure that you implement
the appropriate hazard minimizations for your environment, such as laser
safety goggles, beam blocks, and door interlocks. MOGLabs takes no
responsibility for safe configuration and use of the laser. Please:
- Avoid direct exposure to the beam.
- Avoid looking directly into the beam.
- Note the safety labels (examples shown in figure below) and heed their warnings.
- When the laser is switched on, there will be a short delay of two seconds before the emission of laser radiation, mandated by European laser safety regulations (IEC 60825-1).
- The STANDBY/RUN keyswitch must be turned to RUN before the laser can be switched on. The laser will not operate if the keyswitch is in the STANDBY position. The key cannot be removed from the controller when it is in the clockwise (RUN) position.
- To completely shut off power to the unit, turn the keyswitch anticlockwise (STANDBY position), switch the mains power switch at rear of unit to OFF, and unplug the unit.
- When the STANDBY/RUN keyswitch is on STANDBY, there cannot be power to the laser diode, but power is still being supplied to the laser head for temperature control.
WARNING
The internal circuit board and piezoelectric transducers are at high voltage
during operation. The unit should not be operated with covers removed.
CAUTION
Although the CES is designed and priced with the expectation that the end-user
can replace the diode and change the alignment, some components are fragile.
In particular the filter, piezo actuator, and mirror are very easily damaged.
Please take care of these items when working inside the laser. The filter and
mirror are hard-coated and can be cleaned but great care is needed as with any
intracavity laser optics.
NOTE
MOGLabs products are designed for use in scientific research laboratories.
They should not be used for consumer or medical applications.
Label identification
The International Electrotechnical Commission laser safety standard IEC
60825-1:2007 mandates warning labels that provide information on the
wavelength and power of emitted laser radiation, and which show the aperture
where laser radiation is emitted. Figure 1 shows examples of these labels, and
figure 2 shows their location on the CES laser.
Model number: CES
Serial number: A72211001
Manufactured: NOVEMBER 2022
Complies with 21 CFR 1040.10, and 1040.11 except for deviations pursuant to
Laser Notice No.50, dated 24 June 2007 MOG Laboratories Pty Ltd, 49 University
St Carlton VIC 3053, AUSTRALIA| US FDA compliance
---|---
AVOID EXPOSURE
LASER RADIATION IS EMITTED FROM THIS APERTURE| Aperture label engraving
| Warning and advisory label
Class 3B
Figure 1: Warning advisory and US FDA compliance labels.
Figure 2: Schematic showing location of laser warning labels compliant with International Electrotechnical Commission standard IEC 60825-1:2007, and US FDA compliance label. Aperture label engraved on the front of the CES laser near the exit fibre; warning advisory label on the rear and compliance label on side.
Protection Features
MOGLabs lasers includes a number of features to protect you and your laser.
Protection relay
When the power is off, or if the laser is off, the laser diode is shorted via
a normally-closed solid-state relay at the laser head board.
Emission indicator
The MOGLabs controller will illuminate the emission warning indicator LED
immediately when the laser is switched on. There will then be a delay of at
least 2 seconds before actual laser emission.
Interlock
It is assumed that the laser power supply is keyed and interlocked for safety.
The laser head board also provides connection for an interlock (see appendix
B), if used with a power supply which does not include such an interlock.
RoHS Certification of Conformance
MOG Laboratories Pty Ltd certifies that the MOGLabs External Cavity Diode Laser does not fall under the scope defined in RoHS Directive 2002/95/EC, and is not subject to compliance, in accordance with DIRECTIVE 2002/95/EC Out of Scope; Electronics related; Intended application is for Monitoring and Control or Medical Instrumentation. MOG Laboratories Pty Ltd makes no claims or inferences of the compliance status of its products if used other than for their intended purpose.
Extending laser diode and piezo lifetime
At night, switch to standby:
-
If using the CES to seed an amplifier, first turn off the amplifier.
-
Switch the laser diode current off.
If using a MOGLabs DLC controller, don’t adjust the current, just switch the toggle up (off). -
Switch from RUN to STANDBY.
For a MOGLabs DLC controller in standby mode, the temperature controller will continue to operate, so the laser is ready for quick startup the next day. But the laser diode current and piezo voltage will be zero, extending their operating life.
In the morning, switch back on:
- Switch from STANDBY to RUN.
- Switch the laser diode toggle down (on).
You don’t need to adjust the current, just wait a few minutes for the diode temperature to equilibrate.
You should switch your MOGLabs DLC into STANDBY mode at nights and weekends and whenever the laser is not being used for more than a few hours. Many lasers tend to operate only 40 hours during a 168 hour week; thus switching to standby mode can extend the diode and piezo lifetime by a factor of four.
Introduction
Semiconductor laser diodes are compact, efficient and low-cost, but usually
have poor wavelength control, linewidth and stability. The addition of an
external frequency-selective cavity allows control of the operating wavelength
over a few nm range, with sub-MHz linewidth and stability. The MOGLabs CES
(see Fig. 1.1) is machined from a solid aluminium block, so that the laser is
stable, robust, and insensitive to acoustic disturbances. The cavity is
hermetically sealed for additional suppression of environmental fluctuations
and drift.
The MOGLabs CES is a “cat-eye” design (see Fig. 1.2), in which an external
cavity is formed between the exit surface of the semiconductor diode and a
cat-eye reflector at several centimetres from the rear of the diode [1 3].
Rather than the customary diffraction grating of Littrow-configuration ECDLs,
a high efficiency ultranarrow filter is used to select a single external
cavity mode. Without the need for illuminating a large area of a grating for
feedback, a cat-eye retroreflector can be used to form the external cavity.
The cateye reflector is inherently self-aligning, so that the laser is extremely insensitive to mechanical disturbance, and also ensures high feedback coupling efficiency and consequently narrow linewidth.
Figure 1.2: Schematic of the cateye external cavity diode laser (ECDL)
using single-angled facet (SAF) diode. The external cavity is formed between
the facet of the laser diode at the output (fibre) end, and the mirror. One
longitudinal cavity mode is selected by an ultranarrow intracavity bandpass
filter. A cateye reflector is formed by the piezo-mounted mirror and
intracavity lens.
The output beam from the rear facet of a laser diode is collimated with a high
numerical aperture lens and incident on the filter. The filter transmission
wavelength depends on the rotation angle. Transmitted light is back-reflected
by the cateye lens/mirror combination which efficiently couples light back
into the laser diode. More details can be found in references [13]. The
usable output is directly and permanently coupled at the front diode facet to
a polarisation-maintaining fibre with FC/APC connector at the output.
1.1 External cavity
Semiconductor laser diodes normally have a high reflectivity rear facet and a
front facet with reflectivity of only a few percent. The diode cavity is
called the intrinsic or internal cavity. The external cavity is formed by the
mirror and the diode rear facet, and when the external feedback is greater
than that of the front facet, the external cavity determines the lasing
wavelength. The external cavity is typically about 40 mm long from rear facet
of semiconductor to mirror, giving a cavity mode spacing determined by the
free spectral range FSR = /2L = 4 GHz.
The laser diode is butterfly-mounted and fixed to the laser barrel. The
collimating lens is rigidly held in a focusing tube, itself mounted in an –
translation stage. The filter is fixed to a bearing-mounted rotation assembly
with a fine thread screw to adjust the angle. The wavelength adjuster is
opposed by a spring-loaded screw, and the two can be locked against each other
to further reduce the effects of mechanical vibration. The filter angle is
used for coarse wavelength and mode selection, within the gain bandwidth of
the laser diode.
1.2 Piezo-electric frequency control
Small changes to the laser frequency are achieved by controlling the external
cavity length with a piezo electric actuator. For the MOGLabs CES, the
frequency change is about 20 GHz over the full 150 V range of the piezo. The
bandwidth is limited by mechanical resonances, typically 25 kHz.
1.3 Temperature and current
The laser frequency is also dependent on temperature and injection current;
the sensitivities are typically 3 MHz/µA and 30 GHz/K [4]. Thus, low-noise
stable electronics, such as the MOGLabs DLC external cavity diode laser
controller, are essential (see Ref. [5]) to achieve sub-MHz linewidth and
stability.
An important aspect of an ECDL is temperature control of the cavity, since the
laser frequency depends on the cavity length and hence on the thermal
expansion coefficient of the cavity material [6]. The cavity can be machined
from materials with low thermal expansion coefficient but even then the
passive stability is inadequate for research applications. Active feedback of
the cavity temperature and piezo cavity length provide flexible and stable
control. The MOGLabs CES uses a negative temperature coefficient (NTC)
thermistor to sense the cavity temperature and Peltier thermoelectric cooler
(TEC) to heat and cool the cavity material.
First light
The MOGLabs CES laser has a fibre output permanently attached. Great care
should be taken when handling the laser and being mindful of stresses applied
to the fibre (e.g. bending at tight radii, clamping), stresses applied to the
rubber boot connecting the fibre to the butterfly mount and contamination of
the FC/APC connector end where the fibre is exposed.
Mount your laser to an optical table using the screws provided. Your laser has
been carefully tuned to the specifications given in your laser test report.
Please make sure as you continue with this manual, that the diode injection
current, temperature and piezo offset (FREQUENCY) match those of the test
report.
It is assumed that a MOGLabs DLC controller has been provided with your laser.
If a third party controller is used, please set a current limit according to
the maximum safe operating current stipulated in your test report.
For longer wavelength lasers, an IR upconversion card or CCD camera without IR
filter can be very helpful. Common low-cost security cameras, computer USB
cameras, and home movie or still cameras are also good options, although they
often have infra red filters which may need to be removed.
2.1 Standby/Run
Please first check that the MOGLabs DLC has been set to the correct mains
supply voltage by inspecting the red voltage selector above the rear panel IEC
power inlet. Then turn the main power switch on. Check the DLC internal DIP
switches match the laser test report specification. Make sure that the laser
diode current supply (CURRENT knob) is turned fully anticlockwise, and that
the OFF/MOD, SLOW and FAST lock switches are off (up), and then turn the
keyswitch from STANDBY to RUN. The LED status indicator should be yellow
indicating that the thermistor and TEC elements
2.2 Current
Turn the laser diode CURRENT adjust to zero (fully anti-clockwise). Note that
it is not recommended to turn the current to zero when turning off the laser:
the soft-start function of the DLC ensures that the current is ramped up
slowly and safely to the required current. When first aligning the laser to
your experiment, it is important to set the laser output power to a low value
for safety.
The usable output is directly and permanently coupled at the front diode facet
to a polarisation-maintaining fibre with FC/APC connector at the output. To
measure the power output, use a power meter and sensor with a suitable fibre
connector adapter, preferably an integrating sphere type sensor to avoid the
saturation typically observed with Si photodiode based sensors and small spot
sizes.
Adjust the diode current to 5 10 mA and check that the diode voltage
(VOLTAGE selection on main control knob on DLC) is the same as listed in the
diode data sheet. If in doubt, please contact MOGLabs before continuing.
The laser threshold current is defined as the current at which the output is 1
mW. Adjust CURRENT to achieve 1 mW output, and if the threshold differs from
that in the test report by more than 10 mA please refer to section 4.3.
Above threshold, the laser power vs. injection current is well approximated by
a linear curve function (see figure 2.1). Initially the current should be set
above threshold, but well below the maximum operating current, until the laser
is fully aligned with your experiment.
2.3 Temperature
The optimum temperature has been set by MOGLabs and should not require
adjustment. Once the diode current is set, allow 5 minutes for the temperature
to reach equilibrium. When the laser is not needed for extended periods, for
example overnight, turn the laser diode off and the keyswitch to STANDBY. In
standby mode, the temperature controller remains active, so that stable
operation can be achieved more quickly than when the DLC is powered off. It is
not necessary to change the CURRENT setting when turning the laser diode on
and off.
Operation
Your laser has been carefully tuned to the specifications provided in the laser test report. In most cases, the laser will perform as expected once the current, temperature and piezo settings are adjusted to those in the test report.
3.1 Power
Adjust CURRENT and compare the output power from the fibre to the laser power
vs. injection current curve provided on the first page of your laser test
report. Use a power meter and sensor with a suitable fibre connector adapter,
preferably an integrating sphere type sensor to avoid the saturation typically
observed with Si photodiode based sensors and small spot sizes. The threshold
current and slope above threshold should be similar; if not, refer to section
4.3.
3.2 Wavelength
Once the power measured is comparable with the test report, the wavelength can
be adjusted with reference to a wavemeter or spectrometer. Increase CURRENT to
the value recorded in the laser test report. If the measured wavelength is
within 0.1 nm of the desired wavelength, the current and piezo (FREQUENCY) can
be used to make small changes. If the precise wavelength cannot be reached
with current and piezo adjustments, then the wavelength should be adjusted
using the adjustment screw (see figure 3.1). Ensure that the spring plunger is
not locked against the arm of the filter spindle. For larger wavelength
changes, release the spindleclamp lock screw, then rotate the filter spindle
and re-clamp.
Note that the cateye lens focus is wavelength dependent, so whenever the laser
wavelength is changed substantially (more than say 5-10 nm), we recommened
also optimising the current threshold (see sec. 4.3).
Figure 3.1: CES key components for wavelength and focus adjustment.
Filter angle adjustment, showing the primary (fine) wavelength adjustment
screw and counter-acting spring plunger; filter alignment mark and spindle
locking screw for coarse wavelength adjustment; rear cateye focus adjustment
for current threshold optimisation; fibre output.
After the target wavelength has been achieved, SPAN can be adjusted to
increase the width of the piezo scan. Adjustments of the SPAN knob should be
gradual, and careful adjustments of the diode current may be required in order
to maintain single-mode operation. Confirm that your laser is capable of
reaching the mode-hop free scan range (MHFR) specified in the laser test
report. If the MHFR is less than specified, proceed to sections 3.3 and 4.1.
3.3 Mode-hops
Mode-hops are a frequent occurrence with external cavity diode lasers. A mode-
hop is a discontinuity when tuning or scanning the laser wavelength. As the
laser wavelength is varied, usually by changing the cavity length with a
piezo, competition between the wavelength determined by the different
wavelength-dependent cavity elements can lead to a mode hop: a jump in laser
wavelength to a different external cavity mode. Wavelength-dependent elements
include the external cavity, the laser diode internal cavity between the rear
and front facets of the diode, the filter, and the gain bandwidth of the laser
diode.
The different wavelength-dependent characteristics are shown schematically in
figure 3.2. The net gain is the product of semiconductor gain, filter
response, and internal and external cavity interference. The net gain can be
very similar at adjacent external cavity modes. A small change in the laser
cavity optical path length, the diode internal cavity mode frequency, or the
filter angle, can lead to the overall gain being greater at a mode adjacent to
the mode in which the laser is oscillating, and the laser then hops to that
higher-gain mode. See Ref. [6] for a detailed discussion.
Figure 3.2: Schematic representation for the various frequency-dependent
factors of an ECDL, adapted from Ref. [6], for wavelength = 780 nm and
external cavity length Lext = 15 mm.
3.4 Scanning
The external cavity length is controlled by a piezo actuator moving the
external cavity mirror (a partially reflective output coupler for MOGLabs CEL
or a high reflectivity mirror for MOGLabs CES). The cavity length changes with
piezo voltage, and for a large change, the laser will usually hop to a
neighbouring cavity mode. Figure 3.3 is a schematic of the net gain variation
with laser frequency, showing two adjacent modes of very similar gain. Figure
3.4 is a measurement of the frequency of a laser scanning properly, and with a
mode-hop at one edge of the scan.
Figure 3.3: Combined gain for an external cavity diode laser, including
the internal and external modes, the diode laser gain, and the filter
response. The broad feature is the frequency selectivity of the filter, and
the smaller peaks are the external cavity modes (see fig. 3.2). The laser will
easily hop between the two highest external modes with similar net gain.
Figure 3.4: Frequency of a laser scanning properly (left) and with a
mode-hop at one edge (right).
The mode-hop-free scan range (MHFR) can be optimised by careful adjustment of
the injection current, which affects the optical path length of the diode and
hence the frequency of the cavity mode.
Troubleshooting
4.1 Scanning adjustment
Achieving a wide continuous mode-hop free wavelength scan requires careful
optimization of the control parameters, including laser diode current, focus,
and feed-forward current bias. The laser has been carefully tuned to the
wavelength and scan range specified in the laser test report. The instructions
below describe how to achieve the maximum mode-hop free range (MHFR).
The first step is to confirm that the laser is indeed scanning, for example
with a wavemeter such as the MOGLabs Fizeau wavemeter (FZW), MOGLabs
economical wavemeter (MWM), or a Fabry-Perot cavity.
- Slowly increase SPAN while monitoring the frequency change with a wavemeter.
- If there is no change in scan range, then disable feed-forward current bias using DIP switch 4 in the DLC. If there is no evidence that the laser is scanning with bias disabled, then the piezo has failed; please contact MOGLabs for assistance.
- If the laser scans with bias disabled, adjustment of CURRENT and BIAS (once DIP switch 4 is enabled again) can be used to increase the scan range. Slowly increase SPAN, and use small adjustments of CURRENT and piezo offset FREQUENCY to suppress mode-hops.
- Increase SPAN to the maximum before mode-hops are evident.
- Observe the CHAN B Current output to measure the BIAS; that is, the change in diode current over the scan range. If the peak-to-peak voltage of CHAN B Current is significantly less than that specified in the laser test report, BIAS adjustment is required. For a CES laser, the BIAS is typically set to maximum negative value (fully anti-clockwise).
4.1.1 BIAS optimisation
Ideally the frequencies of the external cavity mode and the intrinsic laser
diode mode are identical (see figure 3.2). The external mode frequency is
controlled by the piezo. The intrinsic diode mode frequency can be controlled
by adjusting the laser diode current.
The diode injection current can be “automatically” adjusted as the laser
frequency is changed, using a “feed-forward” or current bias which changes as
the piezo voltage is changed. Feed-forward current bias adjustment is a
feature of MOGLabs DLC controllers. Each laser requires a different change in
diode current for a given change in piezo, and the ratio can be adjusted with
the BIAS trimpot on the DLC controller.
Optimisation is straightforward. With the laser frequency scanning, the BIAS
control is adjusted until the maximum mode-hop-free scan range is observed.
Small changes to the injection current optimise the scan range near the
nominal centre frequency. A fast Fizeau wavemeter, an atomic absorption
spectroscopy signal, or a Fabry-Perot cavity can be used to monitor the laser
frequency while varying the different control parameters.
-
Ensure that BIAS is enabled (DIP switch 4).
-
Referring to the laser test report CH A FREQ in the Tuning tests (final tuned wavelength) section, adjust SPAN while monitoring CHAN A Freq and CHAN B Current on a dual-channel oscilloscope. Confirm that the peak-to-peak voltages on the two sawtooth outputs match the values specified in the laser test report by first matching CHAN A Freq using SPAN only. Now adjust BIAS until the peak-to-peak voltage for CHAN B Current agrees with the CH B CURRENT in the laser test report, noting also the sign/gradients should agree with the laser test report.
-
Reduce SPAN to zero. Adjust the laser diode CURRENT to find the required laser wavelength and approximate output power.
-
If the wavelength is close but not quite correct, small adjustments of either CURRENT or FREQUENCY may be required to find a better lasing mode. If more significant wavelength adjustment is required, mechanically rotate the filter (see section 4.2).
-
If the wavelength is within a few picometres (GHz) of the target, increase SPAN while observing the wavelength scan as shown in figure 3.4.
-
Increase SPAN until a mode-hop is evident. If using absorption spectroscopy to monitor the laser wavelength, it can be helpful to observe the derivative, for example the demodulated error signal (CHAN B Error on a MOGLabs DLC).
The mode hop should be at one edge of the scan; if so, adjust FREQUENCY so that the scan no longer ‘clips’ this mode hop (i.e. the scan is free of mode hops), and continue adjusting in the same direction until a mode hop is observed on the other edge of the scan. -
Adjust FREQUENCY to the mid-point between the two extremes.
-
Increase SPAN and adjust FREQUENCY until mode hops are evident at both edges of the scan.
-
Adjust diode CURRENT by small amounts to suppress the mode hops. Increase SPAN and adjust CURRENT and FREQUENCY until the mode hops cannot be suppressed.
-
Adjust the BIAS trimpot to suppress the mode hops. Typically for a CES laser the BIAS trimpot should be fully anti-clockwise. Repeat the steps above: increase SPAN, adjust CURRENT, FREQUENCY and BIAS, and repeat until no further improvements can be made.
-
If the MHFR is substantially less than expected (refer to the factory test report), it may be helpful to optimise the focus; see section 4.3.
-
If the MHFR is still less than expected, it may be helpful to change to a different intrinsic diode mode by increasing or decreasing CURRENT. Alternately rotate the filter slightly to alter the net gain so that one cavity mode has higher gain than those adjacent.
4.2 Filter adjustment
The primary control of wavelength is the filter rotation angle, which can
be adjusted while the laser is operational. A MOGLabs Fizeau wavemeter or
high-resolution spectrometer is almost essential, though with patience it is
possible to find an atomic resonance by carefully adjusting the filter angle
while scanning the laser.
Note that the cateye lens focus is wavelength dependent, so whenever the laser
wavelength is changed substantially (more than say 5-10 nm), we recommened
also optimising the current threshold (see sec. 4.3).
To change the wavelength:
-
Set the laser current so that the output power is sufficient, taking care to ensure that the internal cavity power is below the maximum rated for the bare diode i.e. without feedback, if applicable (see figure 2.1).
For adjustments of greater than 1nm: -
Unlock the filter so that it can rotate, by turning the spindle-clamp lock screw anti-clockwise, for example one full turn.
-
Rotate the filter assembly spindle using an allen key or hex ball driver in the hex socket in the centre of the shaft, making sure that the notch does not align at 0 to the laser body.
-
Re-lock the spindle-clamp.
-
Optimise the cateye focus (refer sec. 4.3). For adjustments of less than 1nm:
-
Check that the spring-plunger is engaged but not locked against the brass arm of the spindle clamp. That is, the pin should be visibly protruding from the spring-plunger screw, and touching the brass arm.
-
Adjust the filter angle using the fine thread adjust screw, acting against the spring-loaded plunger opposite to the adjust screw.
-
The laser will hop between external cavity modes as the wavelength is adjusted, through cycles of dim and bright output.
-
It may be necessary to adjust the spring plunger, so that pressure is maintained on the brass arm of the spindle clamp, without locking completely.
-
Once the laser is operating near the desired wavelength, adjust the diode current and the piezo voltage to achieve the exact wavelength required. If the CURRENT adjustment has shifted the wavelength to a mode further away from the desired wavelength, use the fine screw adjustment method to correct the wavelength.
The filter transmission wavelength shifts with rotation according to (4.2.1) where is the angle of incidence, 0 is the filter wavelength at normal incidence and eff 1 7 is an effective refractive index. The sensitivity to rotation of the fine tangential wavelength adjustment screw is about 0.5 nm to 1 nm per turn.
4.3 Threshold optimisation
The lasing threshold, at which the overall gain exceeds losses, should be as
low as possible to maximise the output power and also the scan range and
frequency stability. The lowest threshold is achieved by optimising the focus
of the cateye lens (see figure 3.1). The cateye lens focus is wavelength
dependent, so whenever the laser wavelength is changed substantially, we
recommened also optimising the threshold. Note also that the gain curve of the
SAF laser diode has a current-dependent wavelength shift; optmising the cateye
lens focus at low currents does not necessitate optimised feedback at
higher/operating currents.
If the CES output wavelength is near a diode gain edge, the current threshold
can be much higher than expected and hence difficult to optimise. For this
reason, a general approach to threshold optimisation is:
- Increase the laser diode current to the maximum safe current specified in the laser test report.
- Optimise the CES output power by adjusting the cateye lens focus. The cateye lens focus is adjusted by rotating the threaded lens holder (see figure 3.1), using the tool provided with the laser, a wrench or a flat-blade screwdriver.
If it is not possible to obtain the output power expected, first ensure the filter angle is not set to a wavelength at or beyond an edge of the diode gain curve by using e.g. a MOGLabs FZW Fizeau wavemeter (refer to the laser diode manufacturer datasheet for the gain curve). If the output wavelength is well within the gain curve, a more involved procedure to optimise the cateye lens focus is described below:
-
Ensure the DLC is off or on standby.
-
Take note of the filter angle by marking on the laser barrel the alignment of the filter assembly notch (the mark on the aluminium spindle, not the mark on the brass clamp) relative to the laser barrel notch. Take a photo from above, so that you can restore the original filter angle if needed.
-
Loosen the spindle-clamp lock screw and rotate the filter spindle such that the filter alignment mark (the mark on the aluminium spindle, not the mark on the brass clamp) is aligned with the laser barrel notch. Tighten the spindle-clamp lock screw back onto the spindle.
-
Unscrew and remove the two socket head cap screws either side of the filter spindle.
-
Undo the spring plunger and wavelength adjust screws until they are well clear and no longer touching the brass filter adjust clamp.
-
Gently prise the filter assembly from the laser barrel (see figure 4.1). If necessary, use pliers to gently grip the brass adjuster, or flat blade tweezers to lever the brass adjuser, or turn the laser upside down so the assembly drops out (be very careful not to stress the fibre output). Excessive force should not be required. If the filter seems stuck, it’s because the filter angle isn’t quite aligned.
Figure 4.1: Removing the intracavity filter assembly. -
Connect a fibre-coupled power sensor and power meter to monitor the laser output power. An integrating sphere type sensor is much preferred due to the tendency for small spots size to saturate photodiode based sensors.
-
Turn the DLC on and adjust CURRENT to 5-10 mA above the threshold specified in the laser test report. You should have at least 1 mW of output power, otherwise increase the current further, but do not exceed the maximum safe current specified in the laser test report.
-
Adjust the cateye lens focus to optimise the output power by rotating the focus knob using the tool provided with the laser, a wrench or a flat-blade screwdriver.
-
Iterate reducing the injection current to obtain 1 mW, followed by focus adjustment of the cateye to optimise power, until the minimum current threshold is achieved.
-
Turn off the current, reassemble the laser and adjust the filter angle to approximately the angle you require for your desired wavelength, as noted at the start of this enumerated sequence with a photo or mark on the laser barrel.
-
Turn on the current to the maximum safe current specified in the laser test report and fine-adjust the filter angle to achieve the desired wavelength by monitoring on e.g. a MOGLabs FZW Fizeau wavemeter.
-
Increase the laser diode current to the operating current if not already exceeded, then do a final optimisation of the CES output power at this wavelength by adjusting the cateye lens focus.
A. Specifications
Parameter | Specification |
---|
Wavelength and peak power
960 – 1080 nm
1080 – 1200 nm
1170 – 1270 nm
1250 – 1390 nm
1420 – 1520 nm
1510 – 1610 nm
1860 – 1970 nm| 200 mW, no isolator
100 mW, no isolator
10 mW, with isolator
70 mW, with isolator
40 mW, with isolator
40 mW, with isolator
7 mW, no isolator
Linewidth| Typically < 25 kHz
Filter| 0.2 to 0.4 nm bandpass
Tuning range| Diode dependent, up to 100 nm
Full range may require multiple filters
Sweep/scan
Scan range| > 20 GHz
Mode-hop free| > 10 GHz; typically 20 GHz
Piezo| 3 µm @ 150 V, 100 nF (typical)
Fibre
Polarisation Connector| Polarisation maintaining > 20 dB (100:1)
FC/APC
Thermal
TEC| ±14:5 V 3:3 A Q= 23 W standard
Sensor| NTC 10 kΩ standard; AD590, 592 optional
Stability at base| ±1 mK (controller dependent)
Cooling| Optional: 4 mm diam quick-fit connections
Electronics
Protection| Diode short-circuit relay; cover interlock
connection; reverse diode
Indicator| Laser ON/OFF (LED)
Connector| MOGLabs Diode Laser Controller single cable
connect
Modulation input| Active (AC and DC coupled)
Mechanical & power
Dimensions| 108 ×70 ×83 mm (L×W×H), 1 kg
Beam height| 58 mm
Shipping| 420 ×360 ×260 mm (L×W×H), 3.1 kg
A.1 CES mechanical
Figure A.1: Dimensions of CES laser head.
Laser head electronics
Within the laser head there are two electronic circuit boards, the B1080 and B1085/B1086. B1085/B1086 is a flexible circuit board that connects directly to the B1080 laser headboard. B1085/B1086 acts as mount for the SAF butterfly laser diode and has connection for a thermistor. B1080 has a DVI-D DL socket for connection to the MOGLabs DLC laser controller and a connector for the B1085/B1086 flexible circuit board. It also includes passive protection filters, a laser-on LED indicator, and an SMA connection for direct diode current modulation.
Figure B.1: B1080 laser headboard. Jumpers at top left can be configured for AC or DC coupling, and “Buff” for direct or buffered (differential) ground coupling. Buff is shown connected for differential coupling; change to pins 1 and 2 for direct. Modulation input via SMA connector, sensitivity 2.5 mA/V. Connectors are JST S2B-EH.
B.1 SMA input
The SMA input provides AC or DC coupling high bandwidth active current
modulation for wide bandwidth frequency stabilisation and linewidth narrowing,
for example using a high finesse optical cavity or polarisation spectroscopy.
Note that connection to the SMA input will reduce the diode current by about
2.5 mA with zero input voltage. Ground connection is direct or buffered; the
latter is about 10% slower but reduces problems with ground-loop noise.
| B1080
---|---
Input range| ±2.0 V max
Input coupling| DC (direct) AC/DC (buffered)
AC time constant| 15 Res (10 kHz)
Phase delay| < 20ns (direct)
< 30ns (buffered)
Gain bandwidth (-3 dB)| 20 MHz
Input impedance| AC buffered: 1 kΩ at
10 kHz
DC buffered: 1 kΩ
Direct: 1 kΩ
Current gain| 1 mA/V
Laser diode voltage| 2.5 V max
B.2 Headboard connection to controller
Note
The MOGLabs laser cable is a digital DVI-D DL (dual link) cable. There is a
bewildering assortment of apparently similar cables available. Most computer
display DVI cables will not work because they are missing important pins; see
diagram below. Only high quality digital dual-link DVI-D DL cables should be
used.
Pin | Signal | Pin | Signal | Pin | Signal |
---|---|---|---|---|---|
1 | TEC — | 9 | DIODE — | 17 | DISC + |
2 | TEC + | 10 | DIODE + | 18 | DISC- |
3 | Shield | 11 | Shield | 19 | Shield |
4 | TEC — | 12 | DIODE — | 20 | STACK + |
5 | TEC + | 13 | DIODE + | 21 | STACK — |
6 | Tsense — | 14 | Relay GND | 22 | |
7 | Tsense + | 15 | +5V in | 23 | NTC — |
8 | 16 | Interlock out | 24 | NTC + |
Figure B.2: Headboard connector. Note that the pinout is different to
that of the matching connector on the rear of the DLC controller.
A 10 k thermistor should be connected to NTC+ and NTC, but an AD590 or AD592
temperature sensor can be instead be connected to Tsense. Pin 15 should be
connected to a +5 V supply. To activate the laser diode, relay GND (pin 14)
should be grounded to open the relay that otherwise short-circuits the diode
current. +5 V (pin 15) is internally connected to pin 16 (Interlock), normally
with a permanent connection but on some headboards (see above), a connector is
provided to allow connection to a cover-activated microswitch to disable the
laser when the cover is removed.
Bibliography
- Daniel J. Thompson and Robert E. Scholten. Narrow linewidth tunable external cavity diode laser using wide bandwidth filter. Review of Scientific Instruments, 83(2):, 2012. 1,
- X. Baillard, A. Gauguet, S. Bize, P. Lemonde, Ph. Laurent, A. Clairon, and P. Rosenbusch. Interference-filter-stabilized external-cavity diode lasers. Opt. Communic., 266:609, 2006. 1, 2
- M. Gilowski, Ch. Schubert, M. Zaiser, W. Herr, T. Wu¨bbena, T. Wendrich, T. Mu¨ller, E.M. Rasel, and W. Ertmer. Narrow bandwidth interference filter-stabilized diode laser systems for the manipulation of neutral atoms. Optics Communications, 280(2):443 447, 2007. 1, 2
- H. Talvitie, A. Pietila¨inen, H. Ludvigsen, and E. Ikonen. Passive frequency and intensity stabilization of extendedcavity diode lasers. Rev. Sci. Inst., 68(1):1, 1997. 3
- S. D. Saliba and R. E. Scholten. Linewidths below 100 khz with external cavity diode lasers. Appl. Opt., 48(36):6961, 2009. 3
- S. D. Saliba, M. Junker, L. D. Turner, and R. E. Scholten. Mode stability of external cavity diode lasers. Appl. Opt., 48(35):6692, 2009. 3, 11