FISHER PAYKEL OPT970 Optiflow+ Tracheostomy Direct Connection Instructions

June 4, 2024
Fisher & Paykel

FISHER-PAYKEL-Ologo

FISHER PAYKEL OPT970 Optiflow+ Tracheostomy Direct Connection

INTENDED USE

Tracheostomy patient interface for delivery of humidified respiratory gases.

SETUP
Airvo™ or myAirvo™ series Humidifier with an AirSpiral™ heated breathing tube and chamber kit (e.g 900PT561). Flow Range: 10–60 L/min

OR
MR850 Respiratory Humidifier, RT series kit with 22 mm heated inspiratory tube and chamber (e.g. RT232). Flow range: 5–60 L/min

OR
F&P 950 ™ Respiratory Humidifier in Optiflow™ mode with F&P 950 Adult Heated Circuit kit (e.g. 950A40). Flow Range: 5–70 L/min

Note: F&P 950 Respiratory Humidifier may not be available in all countries.

Tracheostomy tube connection: 15 mm female conical connector. Spare part: Sputum Guard Spare (OPT971).

SETUP INSTRUCTIONS

Step 1(a), (b)
To ensure loading and movement on tracheostomy tube is kept to a minimum, make sure the lanyard is secured properly. Lanyard position as shown in 1a is not suitable for patients with damaged skin.

Step 2
Connect to the humidification system. Make sure all connections are secure. If using an MR850 Respiratory Humidifier, attach the tubing clip to the breathing circuit but ensure probe cable is not crushed by tubing clip.

Step 3
Orientate the sputum guard so that sputum drains downwards. The sputum guard can be removed if required.FISHER-PAYKEL-OPT970-Optiflow+-Tracheostomy-
Direct-Connection-fig-1

WARNINGS

  • Appropriate patient monitoring must be used at all times. Failure to monitor the patient may result in loss of therapy, serious injury or death.
  • Do not obstruct and or modify any part of the interface to avoid hypoxia from loss of therapy.
  • Do not use with an air entrainer, to avoid over-pressure.
  • Do not use system near a naked flame or any ignition source, including electrosurgery, electrocautery, or laser surgery instruments. Exposure to oxygen increases the risk of fire that may result in patient injury or death.
  • Do not crush or stretch tube, to prevent loss of therapy.
  • For single patient use only. Reuse may result in transmission of infectious substances. Attempting to reprocess will result in degradation of materials and render the product defective.
  • The lanyard poses a potential risk of strangulation to vulnerable patients (e.g. children).
  • The lanyard is designed to minimize loading and movement of the tracheostomy tube. Secure the lanyard properly to avoid accidental decannulation or airway damage.
  • The sputum guard is intended to be detachable which may pose
    • a choking risk.
  • California residents please be advised of the following, pursuant to Proposition 65: This product contains chemicals known to the State of California to cause cancer, birth defects and other reproductive harm. For more information, please visit: www.fphcare.com/prop65.

CAUTIONS

  • Failure to use the set-up described above can compromise performance and affect patient safety.
  • Achievable flow rates may depend on individual patient and/or flow source.
  • Before connecting the interface, check for adequate gas flow and ensure that the system has warmed up.
  • Check for condensate regularly. Drain as required.
  • Do not use if packaging is not sealed.

HOSPITAL USE

  • This product is intended to be used for a maximum of 14 days.
  • Do not soak, wash or sterilise.

HOME USE

This product is intended to be used for a maximum of 30 days providing daily and weekly cleaning instructions are followed (see the myAirvo User Manual).

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Fisher & Paykel User Manuals

Related Manuals