Hantek Electronic HBT3000 Series Internal Resistance Tester User Guide

June 17, 2024
Hantek Electronic

Electronic HBT3000 Series Internal Resistance Tester
User Guide

Warranties and Declarations

Copyright
The copyright of this document belongs to Qingdao Hantek Electronic Co., LTD.
Statement
Qingdao Hantek Electronic Co., Ltd. reserves the right to amend this document without prior notice. Qingdao Hantek Electronic Co., Ltd. promises that the information provided is correct  and reliable but does not guarantee that this document is free from errors. Before using this product, please make sure that the specifications of relevant technical documents are the latest effective version. If you use documents or products of Qingdao Hantek Electronic Co., LTD and need products, patents or works of third parties to cooperate with them, you shall be responsible for obtaining the consent and authorization of the third parties. The above consent and authorization shall not be the liability of Hantek.
Product certification
The Hantek certified HBT3000 series internal resistance tester meets Chinese national and industry standards, and has passed CE certification and UKCA certification.
Contact Us
If you have any questions or uncertainties while using the products of Qingdao Hantek Electronics Co., Ltd., you can obtain service and support through the following methods:
Email: service@hantek.com , support@hantek.com Website: http://www.hantek.com

Safety requirement

1.1 Summary of general security issues
Read the following safety precautions carefully to avoid injury and to prevent damage to this product or any product connected. To avoid possible dangers, please use this product in accordance with the regulations.

  • Only professionally authorized personnel can perform repairs.
  • Use the correct power cord.

Only use the dedicated power cord recognized by the country where the product is located.

  • Ground the product.

To avoid electric shocks, the product is grounded through a grounding conductor of the power cable. The grounding conductor must be connected to the ground before connecting the input or output terminals of the product.
Ensure that the product is properly grounded.

  • View all terminal rating values.

To avoid fire or excessive current, please check all rating values and signs on the product. Please consult the product manual for details of the rating values before connecting the product.

  • Do not operate with the cover open.

Do not run the product with the cover or panel open.

  • Avoid circuit exposure.

Do not touch exposed connectors and components after power is switched on.

  • Do not operate if the product is suspected to be faulty.

If you suspect that the product has been damaged, please ask qualified maintenance personnel to check it.

  • Maintain proper ventilation.
  • Do not operate in a humid environment.
  • Do not operate in inflammable or explosive environment.
  • Please keep the product surface clean and dry.

Warning:
Equipment that meets Class A requirements may not provide adequate protection for broadcast services in residential environments.
1.2 Security terms and signs
Security terms in this manual:
Warning:
Indicates that the operation may not cause immediate damage to you.
Warning:
Indicates that if you perform this operation, it may not immediately cause damage to you.
Note:
Indicates that the operation may cause damage to the product or other property.
Safety terms on products:
Warning:
Indicates a potential hazard may be caused to you if you do not perform this operation.
Safety signs on the product: 1.3 Measurement category
Measurement category
This instrument can be used for measurement under class I.
Warning:
This instrument is only allowed to be used in the specified measurement class.
Measurement class definition

  • Class I refers to measurements taken on a circuit not directly connected to the main power supply. For example, measurements made on circuits that  are not exported from a main power supply, especially from a protected (internal) main power supply. In the latter case, the instantaneous stress will change. Therefore, the user should understand the instantaneous capacity of the instrument.
  • Class II refers to measurements taken on a circuit directly connected to lowvoltage instruments. For example, measurements made on household appliances, portable tools, and similar equipment.
  • Class III refers to measurements taken on construction equipment. For example, measurements made on switchboards, circuit breakers, circuits (including cables, busbars, junction boxes, switches, sockets) in fixed equipment, as well as equipment for industrial use and certain other equipment (for example, fixed motors permanently connected to fixed instruments).
  • Class IV refers to measurements taken at the source of low-voltage equipment. For example, measurements made on electricity meters, primary overcurrent protection equipment, and pulse control units.

1.4 Ventilation Requirement
To ensure sufficient ventilation, when using the instrument in a workbench or rack, please ensure that there is a gap of at least 10 centimeters on both sides, above, and behind it.
Note:
Poor ventilation can cause an increase in instrument temperature, leading to instrument damage. Good ventilation should be maintained during use, and ventilation openings should be regularly checked.
1.5 Work Environment
Operating temperature and humidity range
0 ℃ -40 ℃, below 80% RH (without condensation)
Storage temperature and humidity range
-10 ℃ -50 ℃, below 80% RH (without condensation)
Accuracy assurance temperature and humidity range
23 ° C ± 5 ° C, below 80% RH (without condensation)
Warning:
To avoid the risk of short circuits or electric shock inside the instrument, do not operate the instrument in a damp environment.
Altitude
Operating and Non-operating: 2,000m.
This product is powered by mains conforming to installation (overvoltage) category II.
Warning:
Ensure that no overvoltage (e.g. from lightning) reaches the product. Otherwise, the operator may be in danger of receiving electric shock.
Installation (Overvoltage) Category Definitions.
Installation (overvoltage) category I refers to the signal level, which is suitable for measuring terminals connected to equipment in the source circuit, where measures have been taken to limit instantaneous voltage to the corresponding low level.
Installation (overvoltage) category II refers to the local distribution level, which is suitable for equipment connected to the mains (AC power supply).
Pollution Degree
Pollution Degree 2
Pollution Degree Definition

  • Pollution Degree 1: No pollution or only dry, nonconductive pollution occurs. The pollution has no effect. For example, a clean room or airconditioned office environment.
  • Pollution Degree 2: Normally only nonconductive pollution occurs. Temporary conductivity caused by condensation is to be expected. For example, indoor environment.
  • Pollution Degree 3: Conductive pollution or dry nonconductive pollution that becomes conductive due to condensation occurs. To be found in industrial environment or construction sites (harsh environments). For example, sheltered outdoor environment.
  • Pollution Degree 4: The pollution generates persistent conductivity caused by conductive dust, rain, or snow. For example, outdoor areas.

Security level
Class 1 – Grounded products
1.6 Care and Cleaning
Care:
When storing or placing the internal resistance tester, do not expose the LCD display to direct sunlight for a long time.
Cleaning:
According to the requirements of operating conditions, regularly check the internal resistance tester and test line. Please clean the outer surface of the instrument according to the following steps:

  1. Use a lint free cloth to remove floating dust from the outside of the internal resistance tester and test line. Please be extremely careful to avoid scratching the smooth display filter material.
  2.  Clean the internal resistance tester with a soft cloth soaked in water. To clean more thoroughly, a water solvent of 75% isopropanol can be used.

Note:
To avoid damaging the surface of the internal resistance tester or test line, do not use any corrosive or chemical cleaning agents.
Warning:
Before re powering on, please confirm that the instrument has dried thoroughly to avoid electrical short circuits or even personal injury caused by moisture.
1.7 Environmental precautions
The following symbols indicate that this product complies with the requirements established by WEEE Directive 2002/96/EC.
Equipment recycling:
The production of this equipment requires the extraction and utilization of natural resources. If the scrapping of this product is not handled properly, certain substances contained in the equipment may be harmful to the environment or human health. To avoid the release of harmful substances into the environment and reduce the use of natural resources, it is recommended to use appropriate methods to recycle this product to ensure that most materials can be correctly reused.

Overview

This document is used to guide users to quickly understand the front and rear panels, user interface, and basic operating methods of the HBT3000 series internal resistance tester.
Tip: The latest version of this manual can be downloaded at (http://www.hantek.com).
Document number:202312
Software version: Software upgrade may change or increase product functionalities, please pay attention to Hantek website for the latest version. Document format conventions:

  1. button
    Use “square brackets+text (bold)” to represent the front panel buttons, for example, [Utility] represents the “Utility” button.

  2. menus
    Use “menu text (bold)+blue” to represent a menu option, such as “other” to click on the “other” option on the current operating interface of the instrument and enter the “other” function configuration menu。

  3. Operation steps
    Use “->” to represent the next step. For example, [Utility] -> Other means click Utility label before clicking Other menu.

  4. Button icon

Icon Button
Direction keys
Confirm key
Screenshot key
Menu softkeys

Document content agreement:
The HBT3000 series internal resistance tester includes the following models.
Unless otherwise specified, this manual uses HBT3563H as an example to explain the HBT3000 series and its basic operations.

Model| Resistance resolution| Voltage resolution| Measuring range| Interface
---|---|---|---|---
HBT3561A| 0.1μΩ| 1μV| 0~3.6kΩ/
0~110V| RS232/485,USB,LAN,I
O,ANALOG OUTPUT
HBT3562A| 0.1μΩ| 1μV| 0~3.6kΩ/
0~210V| RS232/485,USB,LAN,I
O,ANALOG OUTPUT
HBT3563A| 0.1μΩ| 1μV| 0~3.6kΩ/
0~360V| RS232/485,USB,LAN,I
O,ANALOG OUTPUT
HBT3564A| 0.1μΩ| 10μV| 0~3.6kΩ/
0~1100V| RS232/485,USB,LAN,I
O,ANALOG OUTPUT
HBT3565A| 0.1μΩ| 10μV| 0~3.6kΩ/
0~1600V| RS232/485,USB,LAN,I
O,ANALOG OUTPUT
HBT3566A| 0.1μΩ| 10μV| 0~3.6kΩ/
0~2000V| RS232/485,USB,LAN,I
O,ANALOG OUTPUT

General inspection

Check transportation packaging
After receiving the internal resistance tester, the user should check the equipment according to the following steps: check whether there is any damage caused by transportation: if the packaging carton or foam plastic protective pad is seriously damaged, please keep it until the whole machine and accessories pass the electrical and mechanical tests.
Check attachments
The details of the provided attachments are explained in Appendix A: Models and Accessories at the end of this manual. If missing or damaged attachments are found, please contact the dealer responsible for this business.
Inspect the entire machine
If you find that the appearance of the instrument is damaged, the instrument is not working properly, or it fails the performance test, please contact the dealer responsible for this business.

Product Introduction

The HBT3000 series internal resistance tester is a battery internal resistance tester with high precision, high resolution, and fast measurement characteristics. Using the AC four terminal testing method to more accurately test the internal resistance and voltage of the battery; 0.1μΩ Resistance resolution, minimum voltage resolution 1μV. Ensure the credibility of the measurement; Built in comparator function to automatically determine whether battery parameters  meet standards; Built in multiple high-speed communication interfaces, suitable for more testing scenarios; Easy to operate, precise and fast, stable and reliable, this battery internal resistance tester will be your best choice.
4.1 Introduction to Front Panel

1 Key switch 2 Menu soft keys
3 Directional keys (up/down/left/right), conf key 4 Plus keys, range keys,

and minus keys
5 Input ports| 6 Function keys
7 LCD display screen| 8 Restore default settings
9 Help| 10 USB HOST interface

4.2 Introduction to Rear Panel

1 AC power socket 2 Shell grounding terminal
3 Fuse holder 4 Safety lock holes
5 USB DEVICE interface 6 LAN interface
7 EXT I/O interface 8 ANALOG OUTPUT interface
9 RS-232/485 interface

4.3 User Interface Introduction

1 Range Quick Identification 2 Trigger sources
3 Sampling rates 4 Delay
5 USB icon display 6 USB device icon display
7 Network icon display 8 lock screen icons
9 Local/Remote Display 10 Voltage measurement values
11 Resistance measurement value 12 Auxiliary menu bar
13 Resistance range 14 Voltage range

Preparation before use

5.1 Connecting the power supply
This series of internal resistance testers can input AC power with specifications of 90-240 VAC, 47/63Hz, and 30VA. Please connect the internal resistance tester to the power supply using the power cord provided in the attachment. Press the power switch in the lower left corner of the front panel to turn on the instrument. If the instrument is not turned on, please confirm that the power cord is securely connected and ensure that the instrument is connected to a powered power source.
Warning:
To avoid electric shock, please ensure that the instrument is properly grounded.
5.2 Adjusting the handle
To adjust the handle of the instrument, hold the handles on both sides of the instrument and pull them outward, then rotate the handles. 5.3 Setting System Language
The internal resistance tester supports both Chinese and English menus, and provides corresponding help information, prompt information, and interface display.
Press [Utility] ->Other ->English to select the desired language. When selecting “Chinese” or “English”, the menu, help information, prompt messages, and interface are displayed in Chinese or English, respectively.

Using the built-in help system

To obtain help information for any front panel buttons or menu soft keys, on the measurement display interface, press the [?] button on the front panel, and then  press the button you need help with to obtain the operation prompt for that button. Press the [?] button again to exit the help system.

Set parameter values

The parameter settings of this series of internal resistance testers support numeric keyboard input. Parameter settings can be completed by left clicking, right clicking, and confirmation keys. Move the cursor position by clicking the left and right buttons, and click the confirm button [OK] to input. Move the cursor to the enter key   , then click [OK] to save and exit. Note: If no settings are made on the numeric keyboard interface for more than 30 seconds, the instrument will automatically jump to the measurement display interface.

Start using

  • To prevent electric shock accidents, please do not short circuit the top of the test line and the line with voltage.
  • Do not measure AC voltage, AC current, or DC current. Otherwise, it may cause instrument damage or personal injury accidents.
  • To prevent electric shock, please confirm the rated value of the test wire before measurement, and do not measure voltage higher than the rated value.

8.1 Measurement Settings
Click the [Range] button to enter the measurement display interface. Click the Type menu soft key to select categories such as V&R, Voltage, and Resistance.
8.2 Set voltage range

  1.  Press the [Range] key to enter the measurement page;
  2. Click on the Range (V) menu soft key;
  3. Use the menu soft key to select the range and complete the voltage range setting.

8.3 Setting the Resistance Range

  1. Press the [Range] key to enter the measurement page;
  2. Click on the Range (R) menu soft key;
  3. Use the menu soft key to select the range and complete the setting of the resistance range.

8.4 Zeroing
Click on the [0 ADJ] key, or click on [Utility] ->Cal in sequence. After entering the password, click on the [0 ADJ] soft key to execute zero adjustment. After successful zero adjustment, the interface will pop up as shown in the following figure. If zero adjustment fails, please check if the short circuit is correct, and then perform zero adjustment until successful.

Remote Control

The internal resistance tester can be remotely controlled in the following three methods:

  1. User-defined programming:
    The Standard Commands for Programmable Instruments (SCPI) command can be used to control the internal resistance tester. Please refer to HBT3000-SCPI for detailed instructions on commands and programming.

  2. Use IO software:

The user can use the IO software to send commands for remote control of the internal resistance tester. PC software IO provided by Keysight is recommended. You can log in to Keysight official website (www.keysight.com) to download the software.
operating steps:

  • Establish communication between internal resistance tester and computer
  • Run IO and search for internal resistance tester resources
  •  Open the Remote Command Control panel and send the command

The internal resistance tester can communicate with PC through the following interfaces:

  1. Controlled through USB
  2. Controlled through LAN
  3. Controlled through RS232/485

The manual will provide detailed instructions on how to use the IO software provided by Keysight to remotely control the internal resistance tester through various interfaces.

Default value

The default setting of the machine factory, press the [P] key can restore the instrument state to the factory default value.

Function Default value

Range
Measurement type| V&R
Voltage range| Auto
Resistance range| Auto
MEA|
Sample| Slow
Average| 1
Trigger| Internal
Absolute| Open
Delay(ms)| 1
SAVE/L| OAD
Boot Set| Default
STA| T
Switch| OFF
Data Temp| 1000
COMP|
Switch| Close
Comp Mode| Auto
Function| Default value
---|---
Buzzer| Close
Resistance| Limit
Voltage| Limit
Utility
Brightness| 50%
GPIB Address| 2
Key tone| OFF
Language| Chinese
IP| 192.168.1.127
MASK| 255.255.255.0
IP MODE| Manual
Serial-Baud| 9600
Serial-Data| 7
Serial-Check| None
Serial-Stop| 1

More Product Information

  1. More Product Information
    Click on [Utility] ->Sys Info, you can obtain the machine model, serial number, software version, and hardware version.

  2.  View the option information and the option installation

For more information on this product, please refer to the relevant manual (you can download it from Hantek’s official website (www.hantek.com).

  • HBT3000 Series – User Manual: provides the function of this product to introduce the operation method, remote control method, possible faults in the use process and processing methods, and order information.
  • HBT3000-SCPI Programming Manual: provides the SCPI command set for this product.

Warranty Summary
Qingdao Hantek Electronics Co., Ltd. (hereinafter referred to as Hantek) promises that the host and accessories of its production instruments will have no material or process defects during the product warranty period.
During the warranty period, if the product is proven to be defective, Hantek will provide free repair or replacement for the user. Please refer to the official website of Hantek or the product warranty card for detailed warranty regulations. To obtain the full text of repair services or warranty instructions, please contact the Hantek repair center or local offices.
Except for the warranties provided in this summary or other applicable warranty cards, Hantek Company does not provide any other express or implied warranties, including but not limited to any implied warranties of product tradability and suitability for special purposes. Under no circumstances shall Hantek be liable for indirect, special or consequential losses.

http://www.hantek.com.cn
http://weixin.qq.com/r/WTk7I5XE2cimrdFA92xL
Address:shandong
Management:400-036-7077
Emil: service@hantek.com
0532-55678770, 55678772, 55678773
266000
www.hantek.com

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals